introduction gliomas are the most common primary brain tumour. their infiltrative nature makes...

1
Introduction Gliomas are the most common primary brain tumour. Their infiltrative nature makes complete resection difficult, and yet ‘gross total’ resection is believed to confer a survival advantage of around 3 months over a subtotal resection. Infrared absorption spectroscopy probes the molecular make- up of tissues based on their absorption at different wavenumbers; a ratio at 2850/1655cm -1 representing lipid:protein content has been shown to decrease incrementally from normal brain tissue to grade IV astrocytoma (figure 1), and may offer a means of differentiating grades of tumour and normal tissue. However, studies to date use small sample numbers that are not paired with normal brain. Spectroscopic differences, particularly at tumour borders, offer the potential for real-time analysis dependent on the ability to easily and rapidly analyse tissue. A method of practical application might be through the use of the ultrasonic aspirator, used to fragment and remove tissue during surgery. Analysis of the tissue stream in real-time would offer immediate feedback to surgeons (figure 2), but depends on the assurance that the spectroscopic properties of tissue are not altered by ultrasonic fragmentation. Aims See if the infrared absorbance spectra of tissue collected through the ultrasonic aspirator closely resembles that of tissue resected en bloc. See if these differences are maintained within tumours, and between patients. Reproduce the differences in infrared absorption reported between differing tumour grades (including LGGs and non- astrocytic tumours) and tumour margins at 2850/1655cm -1 . See if other reference peaks offer discriminatory value. Materials and methods Samples: were taken during surgery from consented patients undergoing craniotomy for suspected glioma in Wessex Neurological Centre. Ethical approval was granted under REC08/H0505/165. Sample preparation: Bloc-resected samples were cut from tumour destined for the histopathology lab. Aspirated samples were collected from the stream of tissue drawn through the Stryker Sonopet Ultrasonic Surgical System (UST- 2001). Tissue was processed for spectroscopy in KBr pellets, with blank KBr pellets prepared in the same manner to establish a reference. Data acquisition: Data were collected through an FTIR spectrometer, Varian 600-IR in transmission mode, and assembled in Agilent Resolutions Pro v.5. Background was recorded at 20 scans/min, samples at 15 scans/min. Resolution of 4cm -1 throughout. The spectrometer was continuously purged with nitrogen, with regular background spectra recorded. Data analysis: Data were normalised to the peak around 1655cm -1 as per previous literature. Peak maxima were identified. Corresponding values from a blank KBr pellet Acknowledgements Mr. Jonathan Duffill and the Wessex Neurological Centre theatre teams. Tess de Leon, Optoelectronics Department, University of Southampton. Dr. Sandrine Willaime-Morawek, Elodie Siney, Alex Holden, Clinical Neurosciences, University of Southampton. Conclusions We have shown that the effects of high intensity, low frequency ultrasound on the spectroscopic signature of gliomas are limited: We showed a high correlation between aspirated and bloc- resected tissue. Our spectra demonstrate the presence of the same lipid and protein bands as bloc- resected samples (as in figure 5). Pending the production of statistically significant data demonstrating large enough differences between tissue types, it is possible that a device could be developed to analyse tissue in an ultrasonic aspirator as a means of supplying real-time feedback to operating surgeons (figure 2). Our results suggest that there are differences in absorption from paired samples taken from solid HGG and their margins. These differences were observed at 2850/1655cm -1 as well as 2959/1655cm -1 . We were unable show any previously described differences between HGG and low-grade tumours or normal brain. However with n values of 2, further investigation is required. Elizabeth Casselden, 1 Harry Bulstrode, 1 Paul Grundy, 1 Liam Gray, 2 Harvey Rutt, 3 Diederik Bulters 1 1 Wessex Neurological Centre, Southampton, 2 Cardiff University, 3 University of Southampton Optoelectronics Department References Sanai N, Berger MS. Glioma extent of resection and its impact on patient outcome. J Neurosurg 2008;62(4):753-766. Krafft C, Sobottka SB, Schackert G, Salzer R. Analysis of human brain tissue, brain tumors and tumor cells by infrared spectroscopic mapping. Analyst 2004;129:921- 925. Krafft C, Thummler K, Sobottka SB, Schackert G, Salzer R. Classification of malignant gliomas by infrared spectroscopy and linear discriminant analysis. Biopolymers 2006;82:301- 305. Gaigneaux A, Decaestecker C, Camby I, Mikatovic T, Kiss R, Ruysschaert JM et al. The infrared spectrum of human glioma cells is related to their in vitro and in vivo behaviour. Exp Cell Res 2004;297:294-301. Ahmadi F, McLoughlin IV, Chauhan S, ter-Haar G. Bio-effects and safety of low-intensity, low-frequency ultrasonic exposure. Progress in Biophysics and Molecular Biology 2012;108:119-138. Infrared spectroscopy in the resection of gliomas Figure 5: Example absorbance spectra from various tissues demonstrating the consistent presence of the various peaks, and our ability to at least reproduce spectra broadly comparable to those in the literature. Figure 2: A possible configuration of a device to attach to the ultrasonic aspirator. As tissue is drawn through the device, it is assessed by infrared light. There is potential for immediate feedback. Figure 4: Graph of absorption at 2850/1655cm -1 . There appears to be a trend to an increasing value at the 2850/1655cm -1 ratio with increasing malignancy (n=2 for normal tissue; n=2 for grade II tissue; n=4 for grade III tissue; n= 16 For grade IV tissue). Mean value for samples taken from the margins of HGG (n=4) appears lower than of solid tumour. Figure 3: Scatter graph showing correlation between paired aspirated and bloc-resected samples. Pearson’s correlation coefficient r=0.928, p<0.001. n=12 in each group. Figure 1: Previous results published by Krafft et al, who demonstrated a decrease in the 2850/1655cm -1 absorption ratio with malignancy.

Upload: nathan-higgins

Post on 28-Mar-2015

221 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Introduction Gliomas are the most common primary brain tumour. Their infiltrative nature makes complete resection difficult, and yet gross total resection

IntroductionGliomas are the most common primary brain tumour. Their infiltrative nature makes complete resection difficult, and yet ‘gross total’ resection is believed to confer a survival advantage of around 3 months over a subtotal resection. Infrared absorption spectroscopy probes the molecular make-up of tissues based on their absorption at different wavenumbers; a ratio at 2850/1655cm-1 representing lipid:protein content has been shown to decrease incrementally from normal brain tissue to grade IV astrocytoma (figure 1), and may offer a means of differentiating grades of tumour and normal tissue. However, studies to date use small sample numbers that are not paired with normal brain.Spectroscopic differences, particularly at tumour borders, offer the potential for real-time analysis dependent on the ability to easily and rapidly analyse tissue. A method of practical application might be through the use of the ultrasonic aspirator, used to fragment and remove tissue during surgery. Analysis of the tissue stream in real-time would offer immediate feedback to surgeons (figure 2), but depends on the assurance that the spectroscopic properties of tissue are not altered by ultrasonic fragmentation.

Aims See if the infrared absorbance spectra of tissue collected through the ultrasonic aspirator closely resembles that of tissue resected en bloc. See if these differences are maintained within tumours, and between patients. Reproduce the differences in infrared absorption reported between differing tumour grades (including LGGs and non-astrocytic tumours) and tumour margins at 2850/1655cm-1. See if other reference peaks offer discriminatory value.

Materials and methodsSamples: were taken during surgery from consented patients undergoing craniotomy for suspected glioma in Wessex Neurological Centre. Ethical approval was granted under REC08/H0505/165.Sample preparation: Bloc-resected samples were cut from tumour destined for the histopathology lab. Aspirated samples were collected from the stream of tissue drawn through the Stryker Sonopet Ultrasonic Surgical System (UST-2001). Tissue was processed for spectroscopy in KBr pellets, with blank KBr pellets prepared in the same manner to establish a reference.Data acquisition: Data were collected through an FTIR spectrometer, Varian 600-IR in transmission mode, and assembled in Agilent Resolutions Pro v.5. Background was recorded at 20 scans/min, samples at 15 scans/min. Resolution of 4cm-1 throughout. The spectrometer was continuously purged with nitrogen, with regular background spectra recorded.Data analysis: Data were normalised to the peak around 1655cm-1 as per previous literature. Peak maxima were identified. Corresponding values from a blank KBr pellet were subtracted. Spectra were checked for ‘hidden’ peaks in the region of our maxima. Ratios were compared according to histopathological diagnosis. SPSS was used for statistical analyses.

AcknowledgementsMr. Jonathan Duffill and the Wessex Neurological Centre theatre teams.Tess de Leon, Optoelectronics Department, University of Southampton.Dr. Sandrine Willaime-Morawek, Elodie Siney, Alex Holden, Clinical Neurosciences, University of Southampton.

ConclusionsWe have shown that the effects of high intensity, low frequency ultrasound on the spectroscopic signature of gliomas are limited:• We showed a high correlation between

aspirated and bloc-resected tissue. • Our spectra demonstrate the presence

of the same lipid and protein bands as bloc-resected samples (as in figure 5).

Pending the production of statistically significant data demonstrating large enough differences between tissue types, it is possible that a device could be developed to analyse tissue in an ultrasonic aspirator as a means of supplying real-time feedback to operating surgeons (figure 2).

Our results suggest that there are differences in absorption from paired samples taken from solid HGG and their margins. These differences were observed at 2850/1655cm-1 as well as 2959/1655cm-1.

We were unable show any previously described differences between HGG and low-grade tumours or normal brain. However with n values of 2, further investigation is required.

Elizabeth Casselden,1 Harry Bulstrode,1 Paul Grundy,1 Liam Gray,2 Harvey Rutt,3 Diederik Bulters1

1 Wessex Neurological Centre, Southampton, 2 Cardiff University, 3 University of Southampton Optoelectronics Department

ReferencesSanai N, Berger MS. Glioma extent of resection and its impact on

patient outcome. J Neurosurg 2008;62(4):753-766.Krafft C, Sobottka SB, Schackert G, Salzer R. Analysis of human brain

tissue, brain tumors and tumor cells by infrared spectroscopic mapping. Analyst 2004;129:921- 925.

Krafft C, Thummler K, Sobottka SB, Schackert G, Salzer R. Classification of malignant gliomas by infrared spectroscopy and linear discriminant analysis. Biopolymers 2006;82:301-305.

Gaigneaux A, Decaestecker C, Camby I, Mikatovic T, Kiss R, Ruysschaert JM et al. The infrared spectrum of human glioma cells is related to their in vitro and in vivo behaviour. Exp Cell Res 2004;297:294-301.

Ahmadi F, McLoughlin IV, Chauhan S, ter-Haar G. Bio-effects and safety of low-intensity, low-frequency ultrasonic exposure. Progress in Biophysics and Molecular Biology 2012;108:119-138.

Infrared spectroscopy in the resection of gliomas

Figure 5: Example absorbance spectra from various tissues demonstrating the consistent presence of the various peaks, and our ability to at least reproduce spectra broadly comparable to those in the literature.

Figure 2: A possible configuration of a device to attach to the ultrasonic aspirator. As tissue is drawn through the device, it is assessed by infrared light. There is potential for immediate feedback.

Figure 4: Graph of absorption at 2850/1655cm-1. There appears to be a trend to an increasing value at the 2850/1655cm-1 ratio with increasing malignancy (n=2 for normal tissue; n=2 for grade II tissue; n=4 for grade III tissue; n= 16 For grade IV tissue). Mean value for samples taken from the margins of HGG (n=4) appears lower than of solid tumour.

Figure 3: Scatter graph showing correlation between paired aspirated and bloc-resected samples. Pearson’s correlation coefficient r=0.928, p<0.001. n=12 in each group.

Figure 1: Previous results published by Krafft et al, who demonstrated a decrease in the 2850/1655cm-1 absorption ratio with malignancy.