intro n theory aas
TRANSCRIPT
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 1/14
ABSTRACT
Atomic Absorption Spectroscopy (AAS) instrument is used to measures the absorption of light at
specific wavelengths by elements in the Free State, such as calcium. The objective of this
eperiment is to detect calcium in the supplement tablets. Then, this eperiment was conducted
to compare the concentration of un!nown sample (commercial calcium supplement) with the
amount of calcium that needed by human. "n!nown sample was prepared and the data obtained
was used to construct calibration curve. The concentration values that were obtained from
calibration curve are #$%.&#, ##'.'%, .'%, $*.& and #+.% ppm which respect to the p- ', ,
, %, and +. This showed that the larger the p- level, the less acidic a substance is. Since the
relationship between the p- level and the concentration of the un!nown solution was
determined, the effectiveness of the p- solubility and the digestion system in the human.
1
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 2/14
INTRODUCTION
The eperiment was conduct to determine the calcium content present in commercial
tablets using Atomic Absorption Spectroscopy (AAS). alcium is an essential mineral found in
great abundance in the body and also essential for living organisms mostly in cell physiology. As
a major material used in minerali/ation of bones and shells, calcium is the most abundant metal
by mass in many animals and in humans. 0inety1 nine percent (##2) of all the calcium in the
body is found in the bones and teeth. The remaining one percent ('2) is in the blood. alcium
plays important roles in nerve conduction, muscle contraction, and blood clotting. 3tracellular
calcium is also important for maintaining the potential difference across ecitable cell
membranes, as well as proper bone formation. 4f calcium levels in the blood drop below normal
levels, calcium is transferred from the bones to the blood in order to maintain blood calcium
levels. Therefore, it is important to consume sufficient calcium to maintain ade5uate blood and
bone calcium levels.
alcium supplements are used to prevent and to treat calcium deficiencies. 6ost eperts
recommend that supplements be ta!en with food and that no more than * mg should be ta!en
at a time because the percent of calcium absorbed decreases as the amount of calcium in the
supplement increases. 4t is recommended to increase doses throughout the day. 4n addition, they
also recommended that daily calcium inta!e for adults7 ranges from ' to '+ mg and
suggested to ta!e supplements with food to support in absorption.
Atomic spectroscopy is the determination of elemental composition by its
electromagnetic or mass spectrum. The study of the electromagnetic spectrum of elements is
called 8ptical Atomic Spectroscopy. 3lectrons eist in energy levels within an atom. These
levels have well defined energies and electrons moving between them must absorb or emit
energy e5ual to the difference between them. 4n optical spectroscopy, the energy absorbed to
move an electron to a more energetic level and9or the energy emitted as the electron moves to aless energetic energy level is in the form of a photon. The wavelength of the emitted radiant
energy is directly related to the electronic transition which has occurred. Since every element has
a uni5ue electronic structure, the wavelength of light emitted is a uni5ue property of each
individual element. As the orbital configuration of a large atom may be comple, there are many
2
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 3/14
electronic transitions which can occur, each transition resulting in the emission of a characteristic
wavelength of light as illustrated in Figure '.
Figure ': The transition resulting in the emission of a characteristic wavelength of light.
The science of atomic spectroscopy has yielded three techni5ues for analytical use:
'. Atomic Absorption.
. Atomic 3mission.
. Atomic Fluorescence.
The process of ecitation and decay to the ground state is involved in all three fields of
atomic spectroscopy. 3ither the energy absorbed in the ecitation process, or the energy emitted
in the decay process is measured and used for analytical purposes.
Atomic Absorption Spectrophotometer (AAS) is designed to determine the amount
(concentration) of an object element in a sample, utili/ing the phenomenon that the atoms in the
ground state absorb the light of characteristic wavelength passing through an atomic vapor layer
of the element. 4n addition, it also provides accurate 5uantitative analyses for metals in water,
sediments, soils or roc!s. Atomic absorption units have four basic parts which are
interchangeable lamps that emit light with element1specific wavelengths, a sample aspirator, a
flame or furnace apparatus for volatili/ing the sample, and a photon detector.
AAS has been used to study structure1function relationships in foods for both li5uid and
solid to improve overall food 5uality, safety and sensory characteristics. Furthermore, it has also
used to investigate fungal infections in plant materials li!e fruits and seeds or to study mobility
of different chemical components in food materials. ;esides that, there are many applications of
AAS in the industries such as in chemical, petrochemical and refineries industries,
3
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 4/14
pharmaceutical industries, recycling and waste management, water analytics and water treatment
and many more. 4n petrochemical and refineries industries, they apply AAS for determination of
heavy metals, sulfur, nitrogen, and chlorine in oil, fuel, paraffin, fatty acids and polymers. 4n
addition, they utili/e the AAS in identification of substance, 5uality control and sewages
analysis. 4n pharmaceutical industries, AAS are use in screening of human body and for active
ingredients. Additionally, they also apply for control of cleaning validation procedures and
identification of substances.
OBJECTIVE
'. To determine the commercial calcium supplement by Atomic Absorption
Spectrophotometry.
. To compare the concentration of un!nown sample (commercial calcium supplement)
with the amount of calcium that needed by human.
4
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 5/14
THEORY
Atomic absorption units have four basic parts which are interchangeable lamps that emit
light with element1specific wavelengths, a sample aspirator, a flame or furnace apparatus for
volatili/ing the sample, and a photon detector. The techni5ue atomic absorption spectroscopy
typically ma!es use of a flame to atomi/e the sample, but other atomi/ers such as a graphite
furnace are also used. Three steps are involved in turning a li5uid sample into an atomic gas:
'. <esolvation = the li5uid solvent is evaporated, and the dry sample remains.
. >aporisation = the solid sample vaporises to a gas.
. Atomi/ation = the compounds ma!ing up the sample are bro!en into free atoms.
4f light of just the right wavelength impinges on a free, ground state atom, the atom may absorb
the light as it enters an ecited state in a process !nown as atomic absorption. This process is
illustrated in the Figure . Atomic absorption ta!es place when unecited atoms absorb energy
and become ecited atoms. 3lectrons can eist in one of two states:
'. ?round State: 4n this state, the electron contains the least energy possible, orbiting as close as
it can to the nucleus.
. 3cited State: 4n this state, the electron contains more energy than in its ground state,
orbiting further from the nucleus of the atom.
Absorption therefore is carried out by unecited atoms, whereas emission arises from ecited
atoms.
Figure : The atomic absorption process
5
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 6/14
Atomic absorption measures the amount of light at the resonant wavelength which is
absorbed as it passes through a cloud of atoms. As the number of atoms in the light path
increases, the amount of light absorbed increases in a predictable way. ;y measuring the amount
of light absorbed, a 5uantitative determination of the amount of analyst element present can be
made. The use of special light sources and careful selection of wavelength allow the specific
5uantitative determination of individual elements in the presence of others. The atom obscure
re5uired for atomic absorption measurements is produced by supplying enough thermal energy to
the sample to dissociate the chemical compounds into free atoms. Aspirating a solution of the
sample into a flame aligned in the light beam serves this purpose. "nder the proper flame
conditions, most of the atoms will remain in the ground state form and are capable of absorbing
light at the analytical wavelength from a source lamp. The ease and speed at which precise and
accurate determinations can be made with this techni5ue have made atomic absorption one of the
most popular methods for the determination of metals.
The use of light absorption measurements to determine concentrations of atoms is called
atomic absorption spectroscopy (AAS). Although this phenomenon has been studied since early
in the nineteenth century, it was about '#* when the techni5ue became part of the analytical
repertoire, in large part as a conse5uence of wor! by @alsh in Australia. ;ecause of the
simplicity of the electronic structure of an atom (compared with a molecule), and the absence of
vibration and rotational transitions, atomic spectra are simpler than molecular spectra. 4nstead of
the band spectra characteristic of molecular absorption in the "> and visible regions, atoms
ehibit line spectra that is, the absorption pea!s are so narrow that they are essentially lines.
Some subtle effects are responsible for producing lines of finite width, but they are etremely
narrow nevertheless, with line widths of .'1.'A being typical. As a conse5uence, AAS is
capable of high specificity, since the narrow lines minimi/e the overlapping absorption so
common in molecular band spectra. As in any form of absorption spectroscopy, the light source
must generate radiation that includes the fre5uency that is absorbed by the sample. 4n AAS,etreme specificity is achieved by using as the light source a lamp whose emitting element is
identical with the element to be determined. The most common AAS flame gas miture is air1
acetylene, which generates a flame temperature of B.
Instrumentation of Atomic absorption spectroscop
6
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 7/14
Figure : Schematic of an atomic1absorption eperiment
The e5uipment used in AAS is basically the same as for other spectroscopic absorption
methods. 4t consists of:
•
Ra!iation source
The light source is usually a hollow1cathode lamp of the element that is being measured.
Casers are also used in research instruments. Since lasers are intense enough to ecite atoms to
higher energy levels, they allow atomic absorption and atomic fluorescence measurements in a
single instrument. The disadvantage of these narrow1band light sources is that only one element
is measurable at a time.
• "onoc#romator
The function of the monochromator is to select radiation of the correct wavelength and
eliminate other radiation from the light path. For many elements such as sodium, potassium, and
copper the spectrum is simple and only low resolution monochromator is re5uired. -owever, for
certain other elements, particularly the transition elements, high resolution is necessary to
prevent unabsorbable emission lines.
• Detectors
Dhotomultipliers are used eclusively on commercial e5uipment. 4n practice, it is the
function of the detector to measure the intensity of radiation before and after absorption by the
sample. From this one can calculate how much radiation has been absorbed from the intense
beam.
7
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 8/14
• Optica$ s$it sstem
Two slits are included in the optical system, an entrance slit and an eit slit. The entrance
slit serves to obtain a narrow, parallel beam of light from the source. The eit slit is used to select
radiation of the correct wavelength after it emerges from the monochromator.
• Atomi%er
The function of the atomi/er is to convert the combined atoms of the li5uid sample into
free atoms. The most common atomi/er is the flame. 4n practice the li5uid sample is introduced
into the flame in the form of a droplet. The droplets evaporate and leave a solid residue. The
residue, which contains the same atoms, is decomposed by the flame and free atoms are
liberated. These free atoms absorb the radiation which is measured in this procedure.
• Ana$tica$ &ariab$es
The 5uantitative application of atomic absorption spectroscopy is based on two
relationships. First, the degree of absorption is a function of how many free atoms are present in
the light path. Second, the number of free atoms is a function of the concentration of the metal on
the original sample. The process of atomi/ation depends on several variables. This are1
a) ;urner design
b) Sample flow rate
c) Type of flame
d) Solvent
e) @avelength choice
8
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 9/14
'ROCEDURES
A stoc! solution with +ppm concentration is prepared using 'mC volumetric flas!.
Then a series of standard solution of +, '.+ and *.+ppm is to be prepared using the stoc!
solution f +ppm. The solutions then are to be used in calibration of the spectrometer. Five
different sample of p- is to be ta!en to be measured using Atomic Adsorption Spectrometer
(AAS). -aving the value of the standard solution, then the calibrating curve is plotted to
determine the concentration of the sample.
9
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 10/14
RESU(TS AND CA(CU(ATION
i) Standard Solutions
Sample 0ame Sample oncentration (ppm) Adsorption
' . .
*.+ .#*
'.+ .%'
% +. .+*#
+ + .#%
10
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 11/14
ii) Sample Solutions
Sample 0ame Sample oncentration (ppm) Adsorption
p- ' #$%.&# '.%%
p- ##'.'% '.%#
p- .*% .* p- % $*.& .'%
p- + #+.% .'++
11
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 12/14
DISCUSSION
From the data obtained, the concentrations of calcium in the un!nown solution were calculated
via calibration curve. The calibration curve plays an important role as the reference to determine
the un!nown solutions. The calibration curve was measured using the standard solution which
had been prepared. The data was then accumulated in present of graph in term of the absorbance
versus the concentration. From the graph obtained, the e5uation of y E .'% .'& is used
to determine the concentration of the un!nown solutions. The un!nown solution of p- ' to p- +
is to be determined. From the graph plotted, the correlation coefficient value is .##$.
For the second graph the value of the concentration is calculated using the first e5uation obtain
from the calibration graph y E .'% .'&. The concentration for p- ' was #$%.&#ppm
with an adsorption of '.%%. Thus the second p- of p- was accumulated to have a
concentration of ##'.'%ppm. 4t adsorption value for p- was '.%#. The concentration of p-
was .*%ppm. The adsorption for the p- was .*. The concentration of p- % was
$*.&ppm. The adsorption for the p- % was .'%. The concentration of p- + was $*.&ppm.
The adsorption for the p- % was .'%ppm. The adsorption for the p- + was .'++.
<ue to the mista!e in calculating the mass need to be measured to prepared the stoc! solution
and the series of standard solution, the calibrating is giving a very slightly value for the edge.
The edge with decimal point (.') giving the concentration to be highly concentrated.
Although the value was in range .'+ to '.%, the highest concentration was turn to almost
'ppm. This is the error related to the second graph measurement for AAS. 3ven the
calibration is slightly wrong, the pattern of the concentration for the second graph is still can be
manually calculated as the value of the adsorption doesn7t correspondent to the calibration value.
As shown on the second graph, the highest concentration stated was in p- . This indicated that
the more acidic the solution, the higher concentration of the calcium. This pattern related that the
acidic solution have the more calcium to dissolved.
There are relationship between the p- and the concentration of the calcium hydroide. The same
scenario happens in the human stomach. The digestion process been divided into some part. The
acidic part was when the gastric acid is release from cells lining the stomach, which are coupled
to systems to increase acid production when needed. ?astric acid is a digestive fluid, formed in
12
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 13/14
the stomach. 4t has a p- of ' to and is composed of hydrochloric acid (-l) (around .+2, or
+ parts per million), and large 5uantities of potassium chloride (Gl) and sodium chloride
(0al). The acid plays a !ey role in digestion of proteins, by activating digestive en/ymes, and
ma!ing ingested proteins unravel so that digestive en/ymes can brea! down the long chains of
amino acids. The acidic p- gain from the gastric acid have the possibility to dissolve the calcium
and other nutritious that the body needed. The same ways goes for the eperiment where the
calcium hydroide is strongly reacted with the lower p-. This indicated that at the lower p-, the
calcium hydroide can dissolved very well comparing to the higher value of p- or the basic
solution.
CONC(USION
4n the nutshell, the analytical techni5ue which Atomic Adsorption Spectroscopy (AAS) was used
was to determine the concentration of the calcium inside the un!nown solution. The relation of
the p- and the concentration was then investigated and to be determine to which p- has the
highest concentration. The relation between the concentration and the p- can be determining
from the calibration curve. As from the calibration curve, the e5uation of y E .'% .'&
was obtained. Thus, from the calculation of the concentration of the calcium content was related
to each of the p- investigated. . The concentration values which correspond to p- ', , , %, +
and * are #$%.&#, ##'.'%, .*%, $*.&, and #+.%ppm respectively. This had proven the
eistence of the relationship between the p- and the concentration of the calcium. . 6aimum
amount of calcium that recommended is *mg9C per day. 4n our stomach, gastric juice appears
at p- ' to p- in order to digest our food. The concentration of un!nown sample from p- ' to
p- is suitable for human body. The total concentration from p- ' to p- can be supported by
stomach. ;esides, this showed the concentration for p- in the range of ' to is ade5uate by
blood and bone calcium levels.
13
8/13/2019 Intro n Theory AAS
http://slidepdf.com/reader/full/intro-n-theory-aas 14/14
RECO""ENDATION
'. 6a!e sure the sample preparation which is the dilution is conducted by a proper
procedure.
. "se digital pipette to ensure the eperiment easy to run and more accurate.
. 4t is better to have the e5uipment chec!ed and maintenance to prevent any error.
%. "se the eact cathode lamp to determine the correct composition.
+. Hinse and dried the apparatus to get a better calibration curve.
RE)ERENCES
http:99dietary1supplements.info.nih.gov9factsheets9calcium.asp
http:99www.calciuminfo.com9about9foodsources.asp
S!oog, <. A., -oller, F. I., J 0ieman, T. A. ('##$). Principles of Instrumental Analysis. ;roo!s
ole.
S!oog, <. A., @est, <. 6., -oller, F. I., J rouch, S. H. (%). Fundamentals of Analytical
Chemistry. ;roo!s ole.
14