instruction manual squareone fixed gas monitor oxygenfinal[1]

16
 OPERATION AND MAINTENANCE MANUAL FOR SQUARE ONE OXYGEN DEPLETION MONITOR Square One Gas & Cryo Equipment, LLC 32818 Walker Road Avon Lake, Ohio 44012 Tel: (440) 930-2150 Fax: (440) 930-0571 www.squareoneequipment.com  

Upload: febo007

Post on 06-Apr-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 1/16

 

OPERATION AND MAINTENANCE MANUALFOR SQUARE ONE

OXYGEN DEPLETION MONITOR

Square One Gas & Cryo Equipment, LLC

32818 Walker RoadAvon Lake, Ohio 44012

Tel: (440) 930-2150Fax: (440) 930-0571

www.squareoneequipment.com 

Page 2: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 2/16

 

OPERATION AND MAINTENANCEMANUAL FOR

SQUARE ONE OXYGEN DEPLETION MONITOR

IMPORTANT:

THE LIFE OF YOUR SQUARE ONE SENSORS WILL EXPIRE IN

 ___________________________ 

PLEASE CALLSQUARE ONE GAS & CRYO EQUIPMENT, LLCTO ORDER YOUR REPLACEMENTS

Page 2 of O & M Manual for Square One Oxygen Depletion Monitor 

Page 3: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 3/16

 

LEFT BLANK INTENTIONALLY

Page 3 of O & M Manual for Square One Oxygen Depletion Monitor 

Page 4: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 4/16

I. OXYGEN DEPLETION MONITOR GENERAL SPECIFICATIONS 

  Gases Detected: Oxygen

  Sensor Technology: Electrochemical Cell

Sensing Method: Diffusion

Electrical Supply: 240 VAC, 50/60 Hz, 50mA (240V Model) or 120 VAC, 50/60 Hz, 100mA (120V Model) or 24 VAC, 50/60 Hz, 500mA (24V Model)

Outputs: Up to 4 SPDT RELAYS, 3A @ 120 VAC

Pollution Degree: 2

Installation Category: II

Maximum Altitude: 2000m

Maximum Humidity: 80% up to 88⁰F (31⁰C)

Operating Temperature Range: 40⁰ to 105⁰F (5⁰ to 40⁰C)

Accuracy: +/- 5% of Calibrated value (electrochemical sensors)

Repeatability: 2% of Signal

Response Time (T90%): 40 seconds

Expected Sensor Lifetime: Dependent on gas type and sensor used

Main Enclosure: NEMA 3 Polystyrene Grey Surface Mounting

Main Enclosure Dimensions: 6” x 6” x 4” (152 mm x 152 mm x 102 mm)

The oxygen depletion monitor is manufactured for indoor use only.

The oxygen depletion monitor is to be connected to a dedicated power source with voltagefluctuations not exceeding +/- 10% of the nominal supply voltage.

Page 4 of O & M Manual for Square One Oxygen Depletion Monitor  

Page 5: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 5/16

II. DESCRIPTION OF EQUIPMENT

The Square One Oxygen Depletion Monitor is a stand-alone single-gas detector designed for single-point monitoring of various toxic and combustible gases.

The sensor can be an integral part of the controller or supplied as a separate/remotepart, to be field connected to the controller by a 2 or 3 conductor low-voltage wire.

With a covering range of up to a 50 feet (15 m) radius for most gases, the Square OneOxygen Depletion Monitor offers a cost effective solution for gas detection in smallenclosed areas.

The Square One Oxygen Depletion Monitor employs a high-precision electrochemicalsensor.

The Square One Oxygen Depletion Monitor is available with up to four (4) SPDT relay

outputs programmed to operate at four (4) different detection levels from either gas; LO,HI, AL1, AL2. These relays can be used for activation of single or multiple-speed fans,dampers and/or remote alarm devices. The Square One Oxygen Depletion Monitor also has a built-in buzzer that can be set off at any desired level(s). The four (4) levelscan also be programmed to have different time delays.

The electronics are housed in a rugged NEMA 3 PVC enclosure with vandal proof screws.

The numbering schematic of the four (4) SPDT relay contacts are as follows:

IMPORTANT NOTE:

The circuit controlled by the dry contacts must be fused at

5A @ 120V or 2A @ 240V

Page 5 of O & M Manual for Square One Oxygen Depletion Monitor  

Page 6: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 6/16

III. FUSE REPLACEMENT

CAUTION

For continued protection against fire, replace only with a fuse of the specified voltageand current ratings:

250V, 1A (BUSSMAN GMA 5 mm x 20 mm or Equivalent)

IV. POWER SUPPLY CONNECTION

The Square One Oxygen Depletion Monitor must be permanently connected to an ACpower supply circuit.

For 120VAC model: (see Fig. 1)Bring AC power to the Monitor from a dedicated uninterruptible 120V, 50/60Hz 5Acircuit either from the top or bottom through the provided ½” pipe knockouts.Run POWER wires inside of the electrical conduits.Make power connection to terminals L & N and make a true ground connection to theground lug provided.

For 24VAC model: (see Fig. 2)Bring AC power to the Monitor from a dedicated uninterruptable 24V, 50/60Hz 5A circuiteither from the top or bottom through the provided ½” pipe knockouts.Run POWER wires inside of the electrical conduits.

Page 6 of O & M Manual for Square One Oxygen Depletion Monitor  

Page 7: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 7/16

V. JUMPER CONFIGURATIONS ON THE MONITOR TO SELECT GASTYPE FOR THE CHANNEL (LK1-LK3 or LK4-LK6)

Please note that your Monitor has been shipped to you with the proper jumper configurations and the illustration on Figure 3 below is provided for your reference only.

Figure 3: Configuration straps and other key components on the

Monitor Board (upper-left quadrant)

Page 7 of O & M Manual for Square One Oxygen Depletion Monitor 

Page 8: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 8/16

VI. INSTALLING THE REMOTE SENSOR

1 2 3 4 5 6L N

G

S+

S+

S-

S-

19.5%OXYGEN

18%OXYGENRUN TRUE

GROUND LINE120V, 1PH, 60 HZ

DEDICATED

G

2 X 18 GA2 CONDUCTORSHIELDED CABLE

MAX. DISTANCE: 500 FT

REMOTE OXYGEN SENSOR

REM-245” X 3” X 1-3/4”MOUNT 4 TO 6 FT

ABOVE FINISHED FLOOR

* * *

** WIRE IN SERIES WITH THEHOLDING COIL OFRESPECTIVE

FAN STARTER

FGM-120

REM-24

 

Figure 4: Remote Sensor Wiring Diagram

Page 8 of O & M Manual for Square One Oxygen Depletion Monitor 

Page 9: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 9/16

The oxygen depletion monitor with remote sensor option consists of one (1) controller and one (1) remote oxygen sensor as shown in Figure 4. Oxygen sensors areelectrochemical cell type and have a measuring range of 0-25%. The oxygen sensorshave a life expectancy of 2 years. The sensor must be discarded and replaced uponexpiry of its life span.

A digital readout displays the reading of the sensor.

There are two (2) on-board relays with status LEDs that are energized at 19.5% and at18% of oxygen. The threshold values for the two (2) on-off levels are user settable (seefollowing section).

The two relay output contacts are SPDT and have a capacity of 3 Amp at 120 VAC.

A built-in buzzer is also activated at 19.5% oxygen.

IMPORTANT

All Square One systems are operated at the factory for a minimum of one week prior tofactory calibration.

The unit must be powered from 120 VAC, 1 PH, 60 Hz dedicated constant power source.

VII. CALIBRATION INSTRUCTIONS FOR THE OXYGEN DEPLETION

MONITOR

Please note that your monitor has been shipped to you factory calibrated and under normal conditions it should not need to be field-calibrated for a period of 12 monthsfollowing initial power-up. It is recommended that your monitor be calibrated annuallyusing the procedure outlined below. The sensors must be discarded and replaced upontheir expiration date.

Please call Square One Gas & Cryo Equipment, LLC to order your replacementsensors.

Page 9 of O & M Manual for Square One Oxygen Depletion Monitor 

Page 10: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 10/16

 

Figure 5: Calibration potentiometer and other key components on theMonitor Board (upper-left quadrant shown)

A. CALIBRATION OF THE BUILT-IN SENSOR

1. Gain access to the circuit board inside the main controller by removing the 4 cover screws.

2. Turn the cover over and secure one corner onto the enclosure using one of thescrews, this way your hands are free. Refer to Figure 5 for orientation.

Page 10 of O & M Manual for Square One Oxygen Depletion Monitor 

Page 11: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 11/16

 

3. Remove existing jumper from the remote sensor’s configuration strap (note positionbefore removing LK4, LK5 or LK6). This way the reading on the 3-digit display will befrom the built-in sensor only.

4. Inject a known concentration of the target gas to the sensor via a nozzle at a flowrate of 1 SCFH. Note that the concentration of the gas injected should be 20.8%oxygen balance nitrogen.

5. Wait about 30 seconds to get the full reaction from the sensor; at the end of the 30seconds the reading on the 3-digit display should be equal to the concentration value of the target gas. If it is NOT, turn the potentiometer P1 to obtain the correct reading.

6. Reinsert the jumper back at its original position. This ends the calibration procedure.

Page 11 of O & M Manual for Square One Oxygen Depletion Monitor 

Page 12: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 12/16

B. CALIBRATION OF THE REMOTE SENSOR 

The oxygen electrochemical sensor is supplied calibrated, and the sensitivity of thedevice should not drift by more than 2% of the full signal per month. A potentiometer for routine adjustment is located on the circuit board (see Fig. 5 below).

-

++ -

“TEST”

INLETWITH PLUG

   4   8 .   3  m  m    P

   C   B

   D   I   A .

CALIBRATION

 

Figure 1: Oxygen Sensor PCB Board

For calibration, the signal may be monitored using a volt meter connected acrossterminals “TEST”. A small nonmetallic screwdriver, with a 1.2 x 0.5 mm tip, is requiredto adjust the potentiometer. Calibration can be carried out in ambient air by simplyadjusting the calibration potentiometer until a reading of 17.4 mA is shown, or 174 mV

across “TEST”, corresponding to 20.9% on a scale of 1-25%.

Page 12 of O & M Manual for Square One Oxygen Depletion Monitor 

Page 13: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 13/16

VIII. STEP-BY-STEP INSTRUCTIONS TO CHANGE SETTINGS, TIMEDELAYS AND BUZZER ACTIVATION OPTIONS ON THE SQUARE ONEOXYGEN DEPLETION MONITOR 

1. Gain access to the circuit board inside the main controller by removing the 4 cover 

screws.

2. Turn the cover over and secure one corner onto the enclosure using one of thescrews, this way your hands are free. Familiarize yourself with the circuit board.Observe the row of 4 mini push-buttons located above the row of 4 relays at the lower left quadrant. These 4 push-buttons, numbered #4 to #1 from left to right, see Figure 6,are all that you need to change the settings. The function of each push-button in theset-up mode is explained as follows:

Push-button #4: Save & Exit from the set-up modePush-button #3: Change setting “DOWN” button

Push-button #2: Change setting “UP” buttonPush-button #1: Advance to next parameter 

3. To enter the set-up mode, press and hold down push-button #4. While push-button#4 is being held down, press push-button #1 once – then release push-button #4.

(Please note that you may save settings and exit anytime by pressing push-button #4 or skip steps by pressing push-button #1 repeatedly.)

4. The first reading displayed on the 3-digit readout is the LOW level on-off output(Relay 1) threshold for the built-in sensor (always CO) expressed in PPM (parts per 

million). Use the “UP” or “DOWN” push-buttons (#2 or #3) to set the desired PPMvalue.

5. Press push-button #1 to advance to the next parameter. This is the TIME DELAY setting associated with the LOW level on-off output expressed in minutes. Again, usethe UP or DOWN push-buttons (#2 or #3) to set the desired time delay on the display.Leave the setting at “00” if no time delay is desired.

6. Press push-button #1 to advance to the next parameter. This is the option thatallows the user to have the buzzer activated at the LOW level. Set the display to “1”using the “UP” push-button (#2) to have the buzzer activated – otherwise leave it at “0”

using the “DOWN” push-button (#3).

7. Press push-button #1 to advance to the HIGH 1 level on-off output (Relay 2)parameters associated with the built-in sensor. Repeat steps 4 through 6.

Page 13 of O & M Manual for Square One Oxygen Depletion Monitor 

Page 14: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 14/16

8. Press push-button #1 to advance to the HIGH 2 level on-off output (Relay 3)parameters associated with the built-in sensor. Repeat steps 4 through 6.

9. Press push-button #1 to advance to the ALARM level on-off output (Relay 4)parameters associated with the built-in sensor. Repeat steps 4 through 6.

10. Press push-button #1 to advance to the parameter settings of the remote sensor.Repeat steps 4 through 9 as required. Please note that if the remote sensor is for combustible gases, the concentration displayed on the 3-digit readout is expressed in% LEL (lower explosive limit) values.

11. Press push-button #4 to save and exit when finished.

Figure 6. Switch position on the Monitor Controller Board

Page 14 of O & M Manual for Square One Oxygen Depletion Monitor 

Page 15: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 15/16

 

Square One Gas & Cryo Equipment, LLC32818 Walker RoadAvon Lake, Ohio 44012Tel.: (440) 930-2150Fax.: (440) 930-0571Web: www.squareoneequipment.com 

LIMITED WARRANTY

The company warrants to the original purchaser that the products manufactured by usare free from defects in material and workmanship when properly stored, installed,operated and maintained. Our obligation under this warranty is limited to repair or replacement, at our option, of defective parts, which in our examination shall disclose toour satisfaction to be thus defective, and when such parts are returned to the factory,freight prepaid, within one year from the date of installation and start-up or eighteenmonths from the date of shipment, whichever occurs first. Notification of the return of any parts must be given and acknowledged by us prior to shipment.

All repairs and replacement parts furnished under this warranty shall be FOB factory.The owner shall pay all necessary freight and delivery charges and also all removal andinstallation costs.

Warranty on Controls and Accessories not manufactured by the Company shall be asoffered by the respective manufacturer.

THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIESEXPRESSED OR IMPLIED INCLUDING THE WARRANTIES OF MERCHANTABILITY

AND FITNESS FOR USE AND OF ALL OTHER OBLIGATIONS OR LIABILITIES ONOUR PART, AND WE NEITHER ASSUME NOR AUTHORIZE ANY OTHER PERSONTO ASSUME FOR US, ANY OTHER LIABILITY IN CONNECTION WITH THE SALEOF OUR PRODUCTS. THIS WARANTY SHALL NOT APPLY TO OUR PRODUCTSOR ANY PART THEREOF WHICH HAS BEEN SUBJECT TO ACCIDENT,NEGLIGENCE, ALTERATION, ABUSE OR MISUSE. WE MAKE NO WARRANTYWHATSOEVER IN RESPECT TO ACCESSORIES OR PARTS NOT SUPPLIED BYUS. THE TERM “ORIGINAL PURCHASER” AS USED IN THIS WARRANTY, SHALLBE DEEMED TO MEAN THAT PERSON FOR WHOM OUR PRODUCTS AREORIGINALLY INSTALLED AND MADE OPERABLE. THIS WARRANTY SHALLAPPLY ONLY WITHIN THE BOUNDARIES OF THE UNITED STATES AND CANADA

AND SHALL NOT ATTACH UNTIL THE PRODUCTS HAVE BEEN PAID FOR IN FULLAND SHALL TERMINATE AT THE END OF THE WARRANTY PERIOD PROVIDEDFOR.

Page 16: Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

8/3/2019 Instruction Manual Squareone Fixed Gas Monitor Oxygenfinal[1]

http://slidepdf.com/reader/full/instruction-manual-squareone-fixed-gas-monitor-oxygenfinal1 16/16

 

DESCRIPTION

This Square One Oxygen Depletion Monitor consists of one (1) built-in O2 sensor as shown on the abovediagram. The O2 sensor is electrochemical cell type and has a measuring range of 0-25%. Also, the O2sensor has a life expectancy of 2 years and must be discarded and replaced upon its expiration date.

There are two (2) on-board relays with status LED’s that are energized at LOW and HIGH O2 levels. Thethreshold values for the levels are user-settable (see the instructions provided in the Operation &Maintenance Manual).

The normal percentage/concentration of O2 in the air is 20.9%. When this concentration drops below19.5%, the LOW level dry contacts 1-2-3 shall energize. Likewise, when the percentage/concentration of O2 in the air increases to over 23.5%, the HIGH level dry contacts 4-5-6 shall energize.

Both conditions shall activate the built-in alarm circuit.

The relay output contacts are SPDT (single pole double throw) and have a capacity of 3 amperes at 120VAC.

IMPORTANT INSTALLATION NOTES 

All Square One systems are operated at the factory for a minimum of one week prior to factory calibration.

The unit must be powered from a 24V, 1 PH, 60 Hz dedicated constant power source.