Page 1: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Instruction Manual

optek-sensors AF56optek-Manual--1004-2005-02--AF56-US-2011-01-04

PN: 1004-2005-02 (-52)

optek-Danulat GmbH Emscherbruchallee 245356 Essen

Phone: +49-(0)201-63409-0

Fax: +49-(0)201-63409-999

E-Mail: [email protected]

Internet: www.optek.com

Page 2: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Preface

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Preface

This instruction manual is written to assist the user in proper procedures for

trouble-free operation.

It is explicitly pointed out that optek-Danulat GmbH assumes no responsibility for

loss or damage caused due to improper use of this instruction manual or

products described herein.

This manual is protected by copyright. However, the user may produce copies

and translations if required for correct operation of the products.

On request, this manual is available in other languages as well as in digital

format (Acrobat® Reader 7.0 required).

Our products are being continuously improved. Technical data is subject to

change without notice.

Essen, December 2010

Page 3: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Table of contents

Table of contents

Preface

1 Using the instruction manual .................................................................................................1

1.1 Validity of the instruction manual .........................................................................................1

1.2 Pictograms and signal words ...............................................................................................2

2 Returns and disposal ..............................................................................................................3

2.1 Declaration of decontamination............................................................................................3

2.2 Disposal ...............................................................................................................................3

3 Intended use .............................................................................................................................4

4 Description of AF56-F / AF56-N sensors ...............................................................................5

5 Technical data and exploded views .......................................................................................7

5.1 Exploded view of AF56-F / AF56-N sensor ........................................................................ 10

6 Installation ..............................................................................................................................11

6.1 Standard sensor bodies — installation instructions ............................................................ 11

6.2 Installation of the sensor .................................................................................................... 12

6.3 Airpurge ............................................................................................................................. 13

7 Connection to converter 156 ................................................................................................14

8 Faults ......................................................................................................................................18

- I -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 4: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Table of contents

9 Maintenance ...........................................................................................................................20

9.1 Preventive maintenance..................................................................................................... 20

9.2 Exchange of the lamp module ............................................................................................ 21

9.3 Exchange of detector module............................................................................................. 22

9.4 Return to operation after the exchange of lamp module and detector module.................... 23

9.5 Spare parts and accessories.............................................................................................. 24

9.5.1 Lamp modules / wavelength modules / detector assembly .....................................24

9.5.2 Standard window (see corresponding armature manual) ........................................25

9.5.3 Gaskets (not wetted) ...............................................................................................25

9.5.4 Screw sets ...............................................................................................................26

9.5.5 Small part sets .........................................................................................................26

9.5.6 Installation material .................................................................................................26

9.5.7 Accessories cable sets ............................................................................................27

10 Declaration of conformity .....................................................................................................28

11 Contacts .................................................................................................................................29

- II -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 5: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Using the instruction manual

- 1 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

1 Using the instruction manual

1.1 Validity of the instruction manual

This instruction manual is valid for AF56-F, AF56-N sensors and their variants.

There are the following variants:

• PV

• SF

Abbreviations mean the following sensor options:

• F color (NIR) sensor with application specific wavelength measure-

ment

• N near infrared sensor with range of wavelengths 730 – 970 nm

• SF silicone-free version with restricted temperature ratings

• PV purge Varivent (air purge connections at sensor)

Follow the instruction manual for every operation. If the sensor is not used as

described in this instruction manual, your safety and the sensor function could

be affected.

To maintain reliability of the product, enhance its life cycle, and avoid down

times, follow the instructions given in this manual.

Furthermore, please follow the existing accident prevention and environmental

protection instructions, as well as recognized technical instructions for safe and

professional working.

Page 6: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Using the instruction manual

- 2 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

1.2 Pictograms and signal words

Important information in this instruction manual is marked with the following

pictograms:

Danger!

This pictogram indicates immediate danger to life and health of persons.

The text next to the symbol gives information on how to avoid bodily injuries.

If the possible cause of risk can be specified, the corresponding pictogram pre-

cedes instructions:

Danger!

Electrical voltage.

This pictogram indicates danger due to electrical voltage.

Caution!

This pictogram indicates information on how to avoid material damage.

Note!

This pictogram indicates instructional or general advice.

Page 7: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Returns and disposal

- 3 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

2 Returns and disposal

2.1 Declaration of decontamination

For the safety of our employees and because of legal regulations we need a

signed “declaration of decontamination” before your return can be handled.

This signed declaration must be included with the shipping documents on the

outside of the packaging.

Any returns which were exposed to hazardous substances and were not

professionally decontaminated are not accepted and will be sent back on your

cost.

optek’s declaration of decontamination and contact information can be found on

our website www.optek.com.

2.2 Disposal

Special legal regulations apply to the return and disposal of industrial waste

equipment. However, manufacturer and user can contractually agree on which

party is to fulfill these legal obligations.

Observe current national disposal regulations.

To dispose packaging material, please separate materials into the following

groups:

• Paper / paperboard

• Plastic

For disposal, disassemble the system components and separate them

according to different material groups.

Dispose of materials according to national and local regulations.

If no agreement has been made, products may be shipped to optek for disposal.

Page 8: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Intended use

- 4 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

3 Intended use

The optek-sensors AF56 and their variants are to be applied only as absorption

sensors for liquids and gases in inline applications according to the technical

data.

The use of the sensors in hazardous locations is prohibited. You can purchase

flameproof sensors for hazardous locations from optek. For these sensors,

special instruction manuals are delivered with the sensor.

Unauthorized constructional changes, additional fittings or rebuildings regarding

the sensor are prohibited. The only exception are rebuildings into one of the

variants listed in the chapter "Validity of the instruction manual". Changes to and

interference with the converter program are prohibited as well.

To lay sensor cables underground without protection is prohibited.

The manufacturer is not liable for damage resulting from misuse.

Following this instruction manual is part of the intended use.

Page 9: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Description of AF56-F / AF56-N sensors

4 Description of AF56-F / AF56-N sensors

This sensor is a high precision absorption sensor. It measures attenuation of

light passing through the process medium. The sensor was designed for direct

installation into pipelines.

Fig. 1 AF56-F / AF56-N

The following are the main sensor components:

Fig. 2 Schematic AF56-F / AF56-N

1. Sensor body

2. Windows

3. Filter (VIS or NIR)

4. Detector

5. Optics module

6. Lamp module

Its modular construction offers maximum flexibility in adapting to the process,

without affecting the property of measurement. A precisely defined, constant

light beam penetrates the process medium. A hermetically sealed silicon

photodiode measures intensity of incoming and optically filtered light and

supplies the converter with it in form of photocurrent.

1 22 3 456

- 5 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 10: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Description of AF56-F / AF56-N sensors

Attenuation of light intensity, as a result of absorption and / or scattering of

substances in the process medium, is described by the Lambert-Beer Law. It

states, the logarithm of the transmission loss is proportional to concentration of

the substance. This is true for dissolved substances and undissolved material.

Depending on the particular material properties, concentrations can be detected

in ppm as well as in %.

According to this law, the extinction coefficient, defined through ingredient and

wavelength, affects the logarithm of the transmission loss, too. Therefore, the

choice of wavelength or range of wavelengths is of decisive importance for the

measurement and is defined for each measurement type with the help of optical

filters.

The third factor of influence in the Lambert-Beer Law is the path length of light

passing through the process medium. Equally proportional, the path length, too,

is an important aspect with regard to specification of a photometer. The distance

to be covered by light is defined through the optical path length (OPL), that is the

distance between the windows of a sensor body. In the technical data, OPL is

always given corresponding to window combination A-A.

- 6 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 11: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Technical data and exploded views

5 Technical data and exploded views

Tab. 1 Technical data AF56-F / AF56-N

AF56-F AF56-N

Material sensor body made of stainless steel 1.4571; SS 316 Ti (standard)

Special materials1.4435 (SS 316 L), 1.4539, 1.4462, TFM 4215, Hastelloy® C4, Hastelloy® C22, Titanium, Tantalum, Monel® 400, Inconel® 625, PP, others on request

Line size 1/4" to 8", (DN 6 to DN 200)

Process connectionsASME-Flange, DIN-Flange, Varivent, JIS-Flange, Tri-Clamp, BBS-Clamp,Female Thread NPT, Female Thread DIN ISO 228/1 G,Sanitary Thread (DIN 11851), others on request

Gaskets Viton®, EPDM (FDA), EPDM (USP Cl. VI), Kalrez®, Chemraz®, Fluoraz®, Buna (NBR), Silicone, Viton® / FEP (FDA), others on request

Windows Pyrex®, Sapphire

Optical path length 1 mm - 200 mm

Process pressure 10 mbar to 325 bar (0.15 psi to 4713 psi),depending on process connection, material and design

Process temperature

Data is only valid for appropriate choice of material of sensor body and gaskets! No ice formation on sensor!

permanent:0 – 100 °C (32 – 212 °F)

peak (15 min/day): 0 – 120 °C (32 – 248 °F)

Ambient temperature

operation: 0 – 40 °C (32 – 104 °F)(elevated or reduced ambient temperatures may require restrictions tothe admissible process temperature!)

transport: -20 – 70 °C (-4 – 158 °F)

Airpurge connectors available as standard

Light source incandescent tungsten lamp: 5.0 V DC, 775 mA, typical lamp life 3 to 5 yearsRange of wavelengths specific to application from 400 nm - 480 nm 730 nm - 970 nmDetector silicon photodiode, hermetically sealedCalibration basic calibration in CU (concentration units)Measuring range sensor specific 0 to 3 CU 0 - 0.5 to 4 CUResolution < ± 0.5 % of respective measuring rangeRepeatability < ± 1.0 % of respective measuring rangeLinearity specific to application, < ± 2 % of respective measuring range

Protectionall optical parts protected according to IP65 if sensor is constructed as intended, valid for installation of connector or protection cap

Cable lengthsstandard: 2, 3, 5, 10, 15, 20, 25, 30, 35, 40, 45, 50, 60, 70, 80, 90, 100 m (7, 10, 16, 33, 49, 66, 82, 98, 115, 131, 148, 164, 197, 230, 262, 295, 328 ft.), cable sets length > 100 m (328 ft.) on request; maximum: 250 m (820 ft.)

Cable materialnon-metallic sheathed cable with finely stranded copper conductor according to IEC 228 CL.5, insu-lation PVC (-40 °C … + 70 °C, -40 °F ... +158 °F)

SS plug protector special high shielded cable sets, optional rigid stainless steel connectorCertificates ISO 9001:2008, PED, CE, HP0

- 7 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 12: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Technical data and exploded views

Detailed technical information about the armature is given in the armatures in-

struction manual and the sensor body data sheet of your sensor body.

Tab. 2 Technical data AF56-SF-F / AF56-SF-N

AF56-SF-F AF56-SF-N

Material sensor body made of stainless steel 1.4571; SS 316 Ti (standard)

Special materials1.4435 (SS 316 L), 1.4539, 1.4462, TFM 4215, Hastelloy® C4, Hastelloy® C22, Titanium, Tantalum, Monel® 400, Inconel® 625, PP, others on request

Line size 1/4" to 8", (DN 6 to DN 200)

Process connectionsASME-Flange, DIN-Flange, Varivent, JIS-Flange, Tri-Clamp, BBS-Clamp,Female Thread NPT, Female Thread DIN ISO 228/1 G,Sanitary Thread (DIN 11851), others on request

Gaskets Viton®, EPDM (FDA), EPDM (USP Cl. VI), Kalrez®, Chemraz®, Fluoraz®, Buna (NBR), Silicone, Viton® / FEP (FDA), others on request

Windows Pyrex®, Sapphire

Optical path length 1 mm - 200 mm

Process pressure 10 mbar to 325 bar (0.15 psi to 4713 psi),depending on process connection, material and design

Process temperature

silicone-free version with restricted temperature ratings

permanent:0 – 60 °C (32 – 140 °F)

peak (15 min/day):0 – 80 °C (32 – 176 °F)

Ambient temperature

operation: 0 – 40 °C (32 – 104 °F)(elevated or reduced ambient temperatures may require restrictions tothe admissible process temperature!)

transport: -20 – 70 °C (-4 – 158 °F)

Airpurge connectors available as standard

Light source incandescent tungsten lamp: 5.0 V DC, 775 mA, typical lamp life 3 to 5 yearsRange of wavelengths specific to application from 400 nm - 480 nm 730 nm - 970 nmDetector silicon photodiode, hermetically sealedCalibration basic calibration in CU (concentration units)Measuring range sensor specific 0 to 3 CU 0 - 0.5 to 4 CUResolution < ± 0.5 % of respective measuring rangeRepeatability < ± 1.0 % of respective measuring rangeLinearity specific to application, < ± 2 % of respective measuring range

Protectionall optical parts protected according to IP65 if sensor is constructed as intended, valid for installation of connector or protection cap

Cable lengthsstandard: 2, 3, 5, 10, 15, 20, 25, 30, 35, 40, 45, 50, 60, 70, 80, 90, 100 m (7, 10, 16, 33, 49, 66, 82, 98, 115, 131, 148, 164, 197, 230, 262, 295, 328 ft.), cable sets length > 100 m (328 ft.) on request; maximum: 250 m (820 ft.)

Cable materialnon-metallic sheathed cable with finely stranded copper conductor according to IEC 228 CL.5, insu-lation PVC (-40 °C … + 70 °C, -40 °F ... +158 °F)

SS plug protector special high shielded cable sets, optional rigid stainless steel connectorCertificates ISO 9001:2008, PED, CE, HP0

- 8 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 13: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Technical data and exploded views

Tab. 3 Cable weights

Cable Weight

Lamp cable (1.5 mm2) 0.9 kg / 10 m

Lamp cable (2.5 mm2) 1.2 kg / 10 m

Lamp cable (4.0 mm2) 1.5 kg / 10 m

Detector cable (AF56) 0.6 kg / 10 m

- 9 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 14: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Technical data and exploded views

- 10 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

5.1 Exploded view of AF56-F / AF56-N sensor

Fig. 3 Photo and exploded view AF56-F / AF56-N

Note!

AF56-F / AF56-N sensors are also available as silicone-free version (SF). This

applies to positions 6 and 10 in fig. 3.

13 12 5

6 5

9 8

7 34 22

5

3 4 5

14 13

10 7

8 11

15 17

15 16

1517

1516

1

Tab. 4 Exploded view explanations

No. Explanation No. Explanation

1 Sensor body 10 Detector module AF56-F or AF56-N

2 Window ring M24 x 1.5, incl. 8 screws M5 11 Detector cable AF56

3 O-Ring 25.12 x 1.78, Viton® 12 Lamp cable with SS plug protector

4 Optical housing OH03, M24-AMP11 Purge 134 screws M3 x 12 (DIN 7985), 1.4571 (316Ti), incl. washer

5 O-Ring 21.95 x 1.78, Viton® 14 Detector cable with SS plug protector

6 lamp module AF56 15 O-Ring 4.00 x 1.00, Viton®

7 O-Ring 10.10 x 1.6, Viton® 16 Screw M5 x 06 DIN84, 1.4571 (316Ti)

8 Screws M3 x 6 (DIN 7985), 14571 (316Ti) 17 Purge connection M5, Ms / Ni

9 Lamp cable AF56

Page 15: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Installation

- 11 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

6 Installation

6.1 Standard sensor bodies — installation instructions

Installation instructions are provided in the armatures instruction manual.

Page 16: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Installation

- 12 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

6.2 Installation of the sensor

Tool • not needed

Install sensor as described below:

Note!

O-Rings are delivered separately and are not installed at delivery.

1. Check that there are O-Rings for the sensor assemblies.

2. Place one O-Ring each into the groove of the window rings (see fig. 4).

Fig. 4 O-Ring at window ring

Danger!

Isopropanol is irritating!

Observe the safety regulations when handling Isopropanol!

3. Make sure that the windows of the sensor body are clean. If not, clean them

with Isopropanol.

4. Check the window ring threads for damage and soiling and clean them if

necessary.

5. Screw on sensor assemblies for detector and lamp sides to the sensor body

hand-tight (right-hand thread, fig. 5). It is recommended to use the supplied

installation paste.

Fig. 5 Installed sensor, example AF16-N / AF16-F sensor

Page 17: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Installation

- 13 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

6.3 Airpurge

Should the temperature of the process medium be too low, the temperature of

the air inside the optical housing might fall below dew point. This leads to con-

densation deposits on the window surfaces. For that case, there are airpurge

connections on the window rings of the sensor body or on the optical housings

of the sensor.

Note!

If the product is more than 10 °C (18 °F) cooler than the ambient air of the

sensor, always connect airpurge.

Tool • Screw driver

• wrench 7 mm

At delivery, the drill holes of the air purge connections are sealed with O-Ring

and sealing screws M5 x 6 (DIN 84).

1. Remove the sealing screws and O-Rings.

2. Check if there are O-Rings on air purge connections.

3. Install air purge connections (24, fig. 6) screwing.

4. Place the air purge hoses on the air purge connections (24).

Fig. 6 Window ring air purge connections (left) and optical housing air purge connections (right)

After connecting the sensor to the power supply and putting it into operation,

rinse it as described below:

1. Loosen the two adapters (4,12) turning each 2-3 turns anticlockwise.

2. Rinse the optical housings by aerating them with dry, oil and dust free air for

approx. 10 minutes at a maximum of 0.5 bar (7.25 psi) gauge pressure.

In case you do not have airpurge supply of appropriate quality, you can use

the optek Air Drying System ADS.

3. Reduce air pressure to approx. 0.1 bar (1.45 psi).

4. Retighten the adapter. Keep up the gauge pressure. Air consumption in this

operational state is minimal.

5. Make sure the O-Rings are fitted correctly.

24 241241224244

Page 18: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Connection to converter 156

7 Connection to converter 156

For connecting the sensor cables, observe the following basic conditions:

• Bring the sensor cable to the cable entry from underneath.

• Form a loop with the sensor cable close to the cable entry.

• Do not lay sensor cables in ducts of current-carrying lines.

• Observe cable specifications (s. technical data).

• To lay sensor cables underground without protection is prohibited.

For the connection of the sensor to the converter, there are sensor cables with

plug protector on the sensor side (15, 17; fig. 7) or without plug protector

(11, 14).

Connection to the sensor

On the sensor side, mixing up sensor cables is not possible, as connectors are

distinctive:

• 9-pin connector on the detector side (11, 17),

• 4-pin connector on the lamp side (14, 15)

Fig. 7 Connector with and without plug protector

Tool • not needed

Connection of the sensor cable without plug protector on the detector and lamp

side:

1. Loosen the connection cover of the sensor.

2. Plug in the sensor cable.

3. Screw-tighten the protection cover.

Fig. 8 Connection of the sensor cable without plug protector

11 15 1417

- 14 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 19: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Connection to converter 156

Tool • Screw driver

Connection of the sensor cable with plug protector on the detector and lamp

side:

1. Loosen the connection cover of the sensor.

2. Ensure there is an O-Ring (7, fig. 9) for plug protector.

3. Connect the sensor cable.

4. Fasten each of the four screws of the plug protector (17) with washer.

Fig. 9 Connection of the sensor cable with plug protector

7 17

- 15 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 20: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Connection to converter 156

Note!

Detailed information on the connection of the sensor to the converter 156 is

given in the instruction manual of the converter. There you can as well find the

corresponding wiring plans.

Connection to the converter

The following connections are on the back of the converter:

Fig. 10 Detector inputs and lamp outputs of the converter 156

Letters stand for

• G Detector inputs (5, 1, 2)

• H Lamp outputs (6, 7)

Connections depend on the equipment of your converter and on the number of

sensors you wish to connect.

Note!

Stick to the detector inputs and lamp outputs specified in table 5. Thus, danger

of mixing inputs and outputs up is minimized. The detector input (5, 4, 3) is

inactive.

Tab. 5 Connections

Number of sensorssensor type

Detector inputconnection of the converters

Lamp outputCable set stan-

dard length up to

1 sensor AF56

Detector input (5, 1, 2) Lamp output (6, 7) 100 m / 328 ft.*

*. Cable sets length > 100 m (328 ft.) on request.

ALAR

M II

14

MADE IN GERMANYOPTEK-DANULAT

NPE

L230V

LAM

P AL

1718

1516

3

DETE

CTOR

INPU

TS

554

50/60Hz, 30VA

230VAC, T0,315A115VAC, T0,630A

115/230V !

512

8

ALAR

M I

11

1312

109 m

A

VOLTAGE

67 LA

MP

LAMP -

2120

19

G

H

- 16 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 21: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Connection to converter 156

Tool • Screw driver

To be able to allocate the end splices unequivocally to the terminals of the

converter, each end splice is marked with the number of the correct terminal:

Detector cable to sensor

1 = white (A1) 2 = brown (A2) 5 = black (A5)

Lamp cable to sensor

6 = white or blue (6) 7 = brown (7)

Caution!

Lamp voltage must be adjusted to the cable length in order to compensate volta-

ge loss in the cable. Too low lamp voltage can lead to wrong measuring results.

Too high lamp voltage reduces the life of the lamp module considerably. Follow

the instructions in the instruction manual of the converter. During operation,

voltage at the lamp is 4.8 V DC.

- 17 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 22: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Faults

8 Faults

Among other possibilities, you can detect faults whenever an error message on

the converter appears. Try to clear the fault using the following table and the in-

structions given in the chapter "Software" in the instruction manual of the con-

verter. Should you have any difficulty clearing the fault, feel free to contact our

customer service.

Please find our contact data at Contacts (see chapter 11, page 29).

- 18 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 23: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Faults

Tab. 6 Possible faults and remedies

Possible fault Possible remarks Cause Remedy

Failure of lamp module• "Lamp failure" LED of

converter flashes.• Signal loss

Lamp cable between sensor and converter defective

• Continuity test of lamp cable• Exchange lamp cable for new

one.

Lamp module defective • Exchange lamp module.

Detector failure -

Detector cable between sensor and converter defective

• Continuity test of detector cable

• Exchange detector cable for new one.

Detector defective • Exchange detector.

Condensate formationUnrealistic, random measuring results

Humidity gets into optical housing and forms condensation deposits on windows.

• Use airpurge.

O-Ring missing or defective• Disassemble sensor

assemblies and check O-Rings, exchange if necessary.

Wrong results• Results are fluctuating.• Zero point is drifting.

• Sensor body windows are dirty.

• Sensor body windows are corroded.

• Lamp module near failure, lamp module near the end of its life.

• Clean sensor body window.• Exchange sensor body

window for sapphire window.• Exchange lamp module.

Connection error• No function• No "Lamp failure" LED

message

Detector cable between sensor and converter defective

• Continuity test of detector cable

• Exchange detector cable for new one.

Sensor cable incorrectly connected to converter

• Check and revise connections.

Measuring range exceeded Converter indicates flashing

Process conditions

• Amplify measuring range.• If the measuring range cannot

be amplified, reduce optical path length.

Wavelength-dependent detector module reduces the dynamic measuring range, optical filters reduce wanted signal.

• Reduce optical path length and / or change measuring wavelength.

• Exchange lamp module.

mA-signal (output)

mA-output delivers correct current results if measuring results are low and too low current results if measuring results are high.

Connected load > 600 ohms• Check resistance of wiring.• Use appropriate mA-input.

Small deviations given in %Poor calibration of the receiving mA-input

• Compensation by adjusting calibration of the sending mA-output.

Converter defectiveNone of the above mentioned errors can be detected.

-

• Send system (converter and sensor) to optek for checking purposes. If necessary, the sensor body can remain in the pipeline so that only the optical arms and the converter have to be sent.

- 19 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 24: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Maintenance

- 20 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

9 Maintenance

9.1 Preventive maintenance

Note!

The used detector is not subjected to measurable aging when used properly.

Tab. 7 Preventive maintenance

Component Maintenance activity Maintenance interval Information

Wetted partsinspection with regard to leakages

as part of standard installation maintenance

Possible damaged sealing faces of the sensor body can lead to leakages (see instruction manual of the armature).

Lamp module

exchange 1 - 2 years

Operation of the lamp below its nominal voltage (4.8 V DC instead of 5.0 V DC) enhances lamp life. Strong vibrations, high temperatures or frequent on and off switching of the system can have negative effects on the service life. Statistical service life is 3 years.

Page 25: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Maintenance

- 21 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

9.2 Exchange of the lamp module

Tool • Screw driver

1. Switch the converter voltage-free.

2. Loosen the sensor cable from the lamp module.

3. Loosen the four screws connecting the lamp module to the stainless steel

housing.

4. Pull out the lamp module (8, fig. 11).

5. Check if there is an O-Ring (7) for the lamp module.

6. Insert the new lamp module into the sensor.

7. Fasten the four screws.

8. Re-connect the sensor cable.

9. Switch on the converter.

10. Follow the instructions given in chapter 9.4, page 23.

Fig. 11 Exchange of the lamp module

8 7

Page 26: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Maintenance

- 22 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

9.3 Exchange of detector module

Tool • Screw driver

1. Switch the converter voltage-free.

2. Loosen the sensor cable from the detector module.

3. Loosen the four screws connecting the detector module to the stainless steel

housing.

4. Pull out the detector module (13, fig. 12).

5. Check if there is an O-Ring (7, fig. 12) for the detector module.

6. Insert the new detector module into the sensor.

7. Fasten the four screws.

8. Re-connect the sensor cable.

9. Switch on the converter.

10. Follow the instructions given in chapter 9.4, page 23.

Fig. 12 Exchange of the detector module

7 13

Page 27: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Maintenance

- 23 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

9.4 Return to operation after the exchange of lamp module anddetector module

Before return to operation of the entire measuring system, undertake the follow-

ing activities:

1. Switch on the converter.

2. Wait for approx. 15 minutes, until the converter has reached the working

temperature.

3. Check the system's zero-point (see instruction manual of the converter).

4. Document your settings using the form (see instruction manual of the con-

verter).

5. Check measuring results with regard to plausibility.

6. If settings and measuring results are correct, enable measuring.

Page 28: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Maintenance

9.5 Spare parts and accessories

9.5.1 Lamp modules / wavelength modules / detector assembly

* Recommended spare part for 2-3 years of operation

** other wavelength modules on request

Tab. 8 Spare parts - lamp modules*

Name sensor type Old part number New part number

lamp module AF56 2100-0205-01 1426-3143-1101-01

lamp module AF56-SF 2100-0205-01 1426-3143-1101-02

Tab. 9 Spare parts - detector module AF56-N (NIR)

detector module AF56-N (NIR)(Choose corresponding wavelength module from table 11!)

Old part number New part number

detector module AF56-N 2500-0307-00 1414-3143-8001-00

detector module AF56-SF-N 2500-0307-00 1414-3143-8002-00

Tab. 10 Spare parts - detector module AF56-F**

Detector module AF56-F*(Choose corresponding wavelength module from table 11!)

Old part number New part number

detector module AF56-F 2500-0308-00 1414-3145-8001-00

detector module AF56-SF-F 2500-0308-00 1414-3145-8002-00

Tab. 11 Spare parts — wavelength modules

Name sensor type New part number

wavelength module AF56-N 1430-3143-7300-00

wavelength module AF56-F0430-10 1430-3145-0430-00

wavelength module AF56-F0450-80 1430-3145-0450-00

- 24 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 29: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Maintenance

9.5.2 Standard window (see corresponding armature manual)

* others on request

9.5.3 Gaskets (not wetted)

Tab. 12 Spare parts — standard window

Sensor type / name Name / type Old part number New part number

Standard window* (see corresponding arma-ture manual)

Pyrex® type A 0410-0050-01

type B 0410-0051-01

type C 0410-0052-01

Sapphire type A 0410-0050-03

type B 0410-0051-03

type C 0410-0052-03

type K 0410-0053-03

type AM 0410-0056-03

type BM 0410-0057-03

type CM 0410-0058-03

Tab. 13 Spare parts — gaskets

(4 pieces each) Viton® old part number Viton® new part number

10.10 x 1.60 0208-4004-02 1203-0004-0008-02

21.95 x 1.78 0208-4001-02 1203-0004-0015-02

25.12 x 1.78 0208-4002-02 1203-0004-0016-02

- 25 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 30: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Maintenance

9.5.4 Screw sets

9.5.5 Small part sets

9.5.6 Installation material

Tab. 14 Spare parts — screw sets

(10 pieces each) Old part number New part number

M3 x 6 (DIN 7985) 1206-0010-0023-03

M3 x 12 (DIN 7985) incl. washer M3 (DIN 7980)

1206-0010-0077-01

Tab. 15 Spare parts — small part sets

Small part sets New part number

set purge connection M5 1201-3131-0001-00

Tab. 16 Spare parts — installation material

Installation material New part number

set Klüberpaste UH196-402 4 x 12 g

1201-3131-0002-00

- 26 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 31: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Maintenance

9.5.7 Accessories cable sets

Other cable lengths on request

Other cable lengths on request

Tab. 17 Accessories — cable sets standard

Accessories — cable sets standard New part number

5 m 2311-0115-0005-00

10 m 2311-0115-0010-00

20 m 2311-0115-0020-00

30 m 2311-0115-0030-00

40 m 2311-0115-0040-00

50 m 2311-0115-0050-00

Tab. 18 Accessories — cable sets with SS plug protector

Accessories — cable sets withSS plug protector

New part number

5 m 2311-0315-0005-00

10 m 2311-0315-0010-00

20 m 2311-0315-0020-00

30 m 2311-0315-0030-00

40 m 2311-0315-0040-00

50 m 2311-0315-0050-00

- 27 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

Page 32: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Declaration of conformity

- 28 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

10 Declaration of conformity

Declaration of conformityin accordance with the EC directives

Electromagnetic compatibility 98/336/EEC, 92/31/EEC 93/68/EECLow Voltage Directive 73/23/EEC, 93/68/EEC

We,

the

optek-Danulat GmbH,Emscherbruchallee 2,45356 Essen,

declare in sole responsibility, that the measuring systems

each comprising one converter of the series

X56 with X=1 or 5

and one sensor of the

AF56, AS56, TF56 series

have been developed, constructed and manufactured in conformity with the

mentioned EC directives.

Harmonized standards applied:

• EN 61326:2006 + Corrigendum 1 / June 2008

• EN 61010-1:2001 + Corrigendum 1 / November 2002

+ Corrigendum 2 / January 2004

Manufacturer: optek-Danulat GmbH, Emscherbruchallee 2, 45356 Essen, Ger-

many

Essen, 2009/01/27

Jürgen Danulat Managing Director

Page 33: Instruction Manual - Seita – Soluciones en ... · PDF fileUsing the instruction manual - 1 - optek-Manual--1004-2005-02--AF56-US-2011-01-04 1 Using the instruction manual 1.1 Validity

Contacts

- 29 -

optek-Manual--1004-2005-02--AF56-US-2011-01-04www.optek.com

11 Contacts

For further inquiry, feel free to contact us or our distributing partners at

any time:

Please visit our website for contacts of our local distributors in other countries.

www.optek.com

Germany

optek-Danulat GmbH

Emscherbruchallee 2

45356 Essen / Germany

Phone: +49-(0)201-63409-0

Fax: +49-(0)201-63409-999

E-Mail: [email protected]

Singapore

optek-Danulat Pte. Ltd.

25 Int’l Business Park

#05-109-f German Centre

Singapore 609916

Phone: +65 6562 8292

Fax: +65 6562 8293

E-Mail: [email protected]

Netherlands

optek-Danulat bv

Grote Brugse Grintweg 12a

4005 AH Tiel / Netherlands

Phone: +31-344-683800

Fax: +31-344-653950

E-Mail: [email protected]

Russian Federation

optek-Danulat GmbH

Liter A, Room 911, 26-N

Kolomjazhskij Prospekt 33

197341 St. Petersburg

Phone: +7 963 342 25 66

E-Mail: [email protected]