instru compensate

Upload: sharmaine-manahan

Post on 06-Apr-2018

244 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/3/2019 Instru Compensate

    1/41

    1

    I. About Graphs

    What is a graph?In mathematics, a graph is an abstract representation of a set of objects where some

    pairs of the objects are connected by links. The interconnected objects are represented by

    mathematical abstractions called vertices, and the links that connect some pairs of

    vertices are called edges. Typically, a graph is depicted in diagrammatic form as a set of

    dots for the vertices, joined by lines or curves for the edges.[1]

    What are the types of graphs?There are many types of graphs and charts that are commonly used for showing

    different reports. These are listed as follows.

    Bar Graphs Line Graphs Scatter Plot Graphs Pie Charts

    Bar GraphsThese are one of the most common types of graph used to display data.

    Sometimes known as "column charts", bar graphs are most often used to show amounts

    or the number of times a value occurs. The amounts are displayed using a vertical bar

    or rectangle. The taller the bar, the greater number of times the value occurs. Bar

    graphs make it easy to see the differences in the data being compared.[2]

  • 8/3/2019 Instru Compensate

    2/41

    2

    Figure 1.1 A Bar Graph [3]

    Line GraphsThese are often used to plot changes in data over time, such as monthly

    temperature changes. They can also be used to plot data recorded from scientific

    experiments, such as how a chemical reacts to changing temperature or atmospheric

    pressure.

    Similar to most other graphs, line graphs have a vertical axis and a horizontal axis.

    If you are plotting changes in data over time, time is plotted along the horizontal or x-

    axis and your other data. When the individual data points are connected by lines, they

    clearly show changes in your data and you can use these changes to predict future

    results.[2]

  • 8/3/2019 Instru Compensate

    3/41

  • 8/3/2019 Instru Compensate

    4/41

    4

    Pie Charts or Circle GraphsThese graphs are a little different from the other three types of graphs discussed. For

    one, pie charts do not use horizontal and vertical axes to plot points like the others. They

    also differ in that they are used to chart only one variable at a time. As a result, it can

    only be used to show percentages.

    The circle of pie charts represents 100%. The circle is subdivided into slices

    representing data values. The size of each slice shows what part of the 100% it

    represents.[2]

    Figure 1.4 A Circle Graph[4]

  • 8/3/2019 Instru Compensate

    5/41

    5

    II. Line Graphs Importance of constructing and interpreting line graphs

    Line Graphs play an important role in the modeling and understanding data in

    analytical chemistry. Graphs are visual representations of numerical systems and

    equations. Graphs help us to visualize the relationship of one bit of data to another. The

    relationship can also be translated into a mathematically meaningful equation. The

    equation for a line (y = mx + b) is one such equation.

    In constructing a line graph, plotting known data can help us to visualize the

    behavior of systems in situations that have not been measured. Thus, we can predict

    whether the data gathered are decreasing or increasing with respect to its variables.

    Graph Variables Independent and Dependent variables

    DefinitionsIndependent Variables these variables are ones that are more or less controlled.

    Scientists manipulate these variables as they see fit. They still vary, but the variation isrelatively known or taken into account.

    Dependent Variables these not controlled or manipulated in any way, but

    instead are simply measured or registered. These vary in relation to the independent

    variables, and while results can be predicted, the data is always measured. There can be

    any number of dependent variables, but usually there is one to isolate reason for variation.

  • 8/3/2019 Instru Compensate

    6/41

    6

    y vs. x Convention in plotting variablesBy convention, the independent variable is plotted along the horizontal (x) axis

    and the dependent variable is plotted up the vertical (y) axis. To plot a point, you simply

    move along thex axis to the point that corresponds to the value of your independent

    variable, then find the point up the y that corresponds to your dependent variable's score,

    and mark a spot on the graph where lines from these points meet.

    When choosing a title for a chart, say y plotted againstx. That is, 'dependent

    variable' plotted against 'independent variable'.

    Equation of a Line Mathematical equation

    The equation of a line with a given slope m and the y-intercept b is

    y = mx + b

    This is obtained from the point-slope equation by setting a = 0. It must be

    understood that the point-slope equation can be written for any point on the line, meaning

    that the equation in this form is not unique. The slope-intercept equation is unique

    because if the uniqueness for the line of the two parameters: slope and y-intercept.[5]

    Parameters: slope and y-interceptThe slope of a line is a number that measures its "steepness", usually denoted by

    the letterm. It is the change in y for a unit change in x along the line.[6]

    The y-intercept of a line is the point at which is crosses either the y axis. If we do

    not specify which one, the y-axis is assumed. It is usually designated by the letterb.

    Unless that line is exactly vertical, it will always cross the y-axis somewhere, even if it is

    way off the top or bottom of the chart.[7]

  • 8/3/2019 Instru Compensate

    7/41

    7

    Plotting data of a linear curve ProcedureA standard linear curve is a graph relating a measured quantity (radioactivity,

    fluorescence, or optical density, for example) to concentration of the substance of interest

    in "known" samples. You prepare and assay "known" samples containing the substance in

    amounts chosen to span the range of concentrations that you expect to find in the

    "unknown" samples. You then draw the standard curve by plotting assayed quantity (on

    the Y axis) vs. concentration (on the X axis). Such a curve can be used to determine

    concentrations of the substance in "unknown" samples.

    Line of best fitA straight line drawn through the center of a group of data points plotted on a

    scatter plot. Scatter plots depict the results of gathering data on two variables; the line of

    best fit shows whether these two variables appear to be correlated.

    A more precise method for determining the line of best fit is a mathematical

    calculation called the least squares method. The line of best fit is used in regression

    analysis, and is a key input in statistical calculations such as the sum of squares.[19]

    Method ofLeast Squares (Linear Regression)The method of least squares is a standard approach to the approximate solution

    of overdetermined systems, i.e., sets of equations in which there are more equations than

    unknowns. "Least squares" means that the overall solution minimizes the sum of the

    squares of the errors made in solving every single equation.

    [8]

  • 8/3/2019 Instru Compensate

    8/41

    8

    Using Linear Regression (LR) function of a scientific calculatoro Procedure of encoding/key-in of data

    1. Clear all the data firstPress SHIFT > Press CLR > Press 3(ALL) > Display: Reset All > Press

    = > Press AC.

    2. Set calculator to Linear Regression modePress MODE twice > Press 2 (REG) > Press 1 (Lin)

    3. Entry of dataDo data entry one by one with x comma y format

    Time (min) Distance (m)

    1 3.5

    3 5.6

    5 7.4

    Then do entry of data:

    Press 1 > Press , > Press 3.5 > Press M+ > Display n=1

    Press 3 > Press , > Press 5.6 > Press M+ > Display n=2

    Press 5 > Press , > Press 7.4 > Press M+ > Display n=3

    o Determination of linearity of data (linearity check)Press SHIFT > Press "2" > Press >" (next button) twice >>> Press "3" (r) >

    Press "=" > Display (for the above example: r = 0.998

    ). If you need R square, just use thesquare button.

  • 8/3/2019 Instru Compensate

    9/41

    9

    o Determination of slope and y-interceptFor slope (B)

    Press "S" > Press "2" > Press ">" (next button) twice > Press "2" (B) >>> Press "=" >Display (for the above example: B =0.975)

    Fory- intercept (A)

    Press "SHIFT" > Press "2" > Press ">" (next button) twice > Press "1" (A) >>> Press "="

    > Display (for the above example: A=2.575)

    Now, you get the equation: y = Bx+A

    y = 0.975x +2.575 ( r = 0.988)

    o Calculating value of x for a given value of yGiven that y = 5.0, x can be calculated using the equation y = 0.975x +2.575 that was

    made earlier. Substitute y from the equation and find x;

    y = 0.975x +2.575

    x = 5.0 2.575 / 0.975

    x = 2.49

    o Calculating value of y given the value of xGiven that x = 4.0, y can be calculated using the equation y = 0.975x +2.575 that was

    made earlier. Substitute x from the equation and find y;

    y = 0.975x +2.575

    y = 0.975(4.0)+2.575 = 6.475

  • 8/3/2019 Instru Compensate

    10/41

    10

    III. Calibration Curve Definition (what is a calibration curve?)

    In analytical chemistry, a calibration curve is a general method for determining

    the concentration of a substance in an unknown sample by comparing the unknown to a

    set of standard samples of known concentration. A calibration curve is one approach to

    the problem of instrument calibration; other approaches may mix the standard into the

    unknown, giving an internal standard.

    The calibration curve is a plot of how the instrumental response, the so-

    called analytical signal, changes with the concentration of the analyte (the substance to

    be measured).[9]

    Procedure (how to prepare a calibration curve?)A calibration curve is prepared by plotting the data or by fitting them to a suitable

    mathematical equation, such as the linear relationship used in the method of least squares.

    The operator prepares a series of standards across a range of concentrations near the

    expected concentration of analyte in the unknown. Analyzing each of these standards

    using the chosen technique will produce a series of measurements. For most analyses a

    plot of instrument response vs. analyte concentration will show a linear relationship. The

    operator can measure the response of the unknown and, using the calibration curve,

    can interpolate to find the concentration of analyte.[9]

  • 8/3/2019 Instru Compensate

    11/41

    11

    Determination of value of unknown concentration of an analyte from a calibrationcurve

    One of the most common applications of spectrophotometry is to determine the

    concentration of an analyte in a solution. The experimental approach exploits Beer's Law,

    which predicts a linear relationship between the absorbance of the solution and the

    concentration of the analyte (assuming all other experimental parameters do not vary).

    In practice, a series of standard solutions are prepared. A standard solution is a

    solution in which the analyte concentration is accurately known. The absorbances of the

    standard solutions are measured and used to prepare a calibration curve, which is a graph

    showing how the experimental observable varies with the concentration. Thus, the points

    on the calibration curve should yield a straight line (Beer's Law). The slope and intercept

    of that line provide a relationship between absorbance and concentration:

    A = slope c + intercept

    The unknown solution is then analyzed. The absorbance of the unknown

    solution,Au, is then used with the slope and intercept from the calibration curve to

    calculate the concentration of the unknown solution, cu.[10]

    cu =

    Au- intercept

    slope

  • 8/3/2019 Instru Compensate

    12/41

    12

    Illustration (graph)

    Figure 3.1 A Calibration Curve

    Sample problems Applications to Spectroscopy

    Standard solutions of iron complex were prepared and the absorbances of

    these solutions were measured by spectronic 21. The results are presented in table

    3.1

    Table 3.1 Absorbance reading of iron complex

    Solutions Concentration (ppm) Absorbance

    Blank 0.00 0.00

    1 1.00 0.21

    2 3.00 0.433 5.00 0.59

    4 7.00 0.70

    5 9.00 0.89

    Unknown ? 0.46

  • 8/3/2019 Instru Compensate

    13/41

  • 8/3/2019 Instru Compensate

    14/41

    14

    Applications to ElectrochemistryStandard solutions of AR grade sodium chloride were prepared and the conductance

    of each solution was measured. The gathered data were presented in table 3.2.

    Table 3.2 Measured conductances of various solutions with respect to its concentration

    Concentration (ppm) Conductance (mS/cm)

    0.00 0.00

    10.0 356

    20.0 663

    30.0 955

    40.0 1265

    Unknown 781

    a. Construct a calibration curve of conductance versus concentration.b. From the calibration curve, calculate the concentration of the unknown solution.

    Solution:

    Figure 3.3 Calibration curve of AR grade NaCl standard solutions

    y = 31.29x + 22

    R = 0.9987

    0

    200

    400

    600

    800

    1000

    1200

    1400

    0 10 20 30 40 50

    Conductance(mS/cm)

    Concentration (ppm)

  • 8/3/2019 Instru Compensate

    15/41

    15

    The equation of the line was found to be y = 31.29x + 22, let x = unknown

    concentration. To solve for x:

    y = 31.29x + 22

    x= 781 22/31.29

    x = 24.26ppm

    Applications to ChromatographyStandard solutions of AR graded CCl4 of known concentration were prepared

    (0.00, 20.00, 40.00, 60.00 and 80.0) and each concentration gave a peak area of (0.00,

    325, 635, 910 and 1230). Construct a calibration curve of peak versus concentration and

    calculate for the concentration of the unknown if its peak area is 721.

    Solution:

    Figure 3.4 Calibration curve of AR graded cyclohexane

    y = 15.225x + 11

    R = 0.9993

    0

    200

    400

    600

    800

    1000

    1200

    1400

    0 20 40 60 80 100

    Peakarea

    Concentration

  • 8/3/2019 Instru Compensate

    16/41

  • 8/3/2019 Instru Compensate

    17/41

    17

    Standard Addition Curve Procedure (how to prepare a standard addition curve)

    The instrument response is then measured for all of the diluted solutions and the

    data is plotted with volume standard added in the x-axis and instrument response in the y-

    axis. Linear regression is performed and the slope (m) and y-intercept (b) of the

    calibration curve are used to calculate the concentration of analyte in the sample.[11]

    Interpretation of data (how to determine the unknown concentration of ananalyte from a standard addition curve)

    If the response is linear, then when the total analyte concentration is 0, the

    response is 0. We can extrapolate the line back to a point at which Response = 0 (as

    presented in figure 2.2). The Concentration value at which Response = 0 is the x-intercept

    of the regression line. Here it happens to be at 0.0015 units of added analyte. What does

    this mean in the context of the analysis? In order to have a total of 0 analyte causing 0

    response, we would have to add 0.0015 units analyte, or remove 0.0015 units. Why?

    Because the sample of unknown that was placed in every container has 0.0015

    concentration units. Remember, [Analyte]total = [added] + [unknown]. So, 0 = -0.0015added

    + [unknown], and thus 1.7 = [unknown].[13]

    Another approach is possible using the linear regression to determine the

    unknown concentration of the analyte,

    S = mVS+ b [Equation 1]

    Where: S = instrument response (signal)

    VS = volume of standard

    Conceptually, if the curve started where the instrument response is zero, the volume of

    standard [(Vs)0] from that point to the point of the first solution on the curve (x = 0)

    contains the same amount of analyte as the sample. So:

    Vxcx = |(VS)0|cS [Equation 2]

  • 8/3/2019 Instru Compensate

    18/41

    18

    Where: Vx = volume of the sample aliquot

    cx = concentration of the sample

    cs = concentration of the standard

    Combining Equation 1 and Equation 2 and solving for cx results in:

    Cx = bCs/mVx [Equation 3]

    And one can then calculate the concentration of analyte in the sample from the slope and

    intercept of the standard addition calibration curve.[12]

    Illustration (graph)

    Figure 2.1 Standard Addition Method Plot (absorbance vs. concentration of standard)

  • 8/3/2019 Instru Compensate

    19/41

  • 8/3/2019 Instru Compensate

    20/41

    20

    Multiple Addition Method (multiple data)

    o Identify variables in equationWhere,

  • 8/3/2019 Instru Compensate

    21/41

    21

    Sample problems Application to spectroscopy

    The single-point standard addition method was used in the determination of

    phosphate by the molybdenum blue method. A 2.00-mL urine sample was treated with

    molybdenum blue reagents to produce a species absorbing at 820 nm, after which the

    sample was diluted to 100-mL. A 25.00-mL aliquot of this solution gave an absorbance

    of 0.428 (solution 1). Addition of 1.00-mL of a solution containing 0.0500 mg of

    phosphate to a second 25.0-mL aliquot gave an absorbance of 0.517 (solution 2). Use

    these data to calculate the concentration of phosphate in milligrams per millilitre of the

    specimen.

    Solution:

    C = A1c1Vs / A2Vt A1Vx [eq. 4.2]

    Here we substitute into eq. 4.2, and obtain

    = 0.00780 mg PO43-

    mL

    This is the concentration of the diluted sample. To obtain the concentration of the

    original urine sample, we need to multiply by 100.00/2.00. Thus,

    [PO43-]= 0.390 mg/mL

  • 8/3/2019 Instru Compensate

    22/41

    22

    Application to electrochemistryA cell containing of a saturated calomel electrode and a silver ion developed a

    potential of -0.4706 V when immersed in 50.00-mL of a sample. A 5.00-mL addition of

    standard 0.02000 M silver solution caused the potential to shift to -0.4490 V. Calculate

    the molar concentration of lead in the sample.

    Solution:

    [eq. 4.1]We shall assume that the activity of Pb

    2+is approximately equal to [Pb

    2+] and

    apply eq. 4.1. Thus,

    whereEcell is the initial measured potential (-0.4706 V).

    After the standard addition is added, the potential becomes Ecell(-0.4490 V), and

    Subtracting this equation from the first leads to

    = 0.7297

    [Ag+]= 4.08 x 10

    -3

  • 8/3/2019 Instru Compensate

    23/41

  • 8/3/2019 Instru Compensate

    24/41

    24

    Figure 5.1 A Continuous Variation Plot (Absorbance vs. Mole fraction)

    Sample ProblemA series of Ni2+ and dimethylglyoxime (DMG) solutions of equal volume was

    prepared to determine the stoichiometry of the complex.The absorbance of each solution

    was measured.

    Table 5.1 Absorbance reading of nickel complex solution

    Sample 0.0050M

    Ni2+

    (mL)

    0.0050M

    DMG

    (mL)

    X Ni2+

    X DMG Absorbance

    1 1.0 9.0 0.1 0.9 0.65

    2 2.0 8.0 0.2 0.8 1.25

    3 3.0 7.0 0.3 0.7 1.39

    4 4.5 5.5 0.45 0.55 1.19

    5 5.5 4.5 0.55 0.45 1.07

  • 8/3/2019 Instru Compensate

    25/41

    25

    Solution:

    Figure 5.2 Plot Continuous Variation

    Beer-Lambert's law states that the absorbance is proportional to the concentration

    of the absorbing species. Since the sum, nM+ nL, is kept constant, the maximum amount

    of complex is formed when the ratio nM/nL corresponds to the stoichiometry of the

    complex.

    Maximum absorbance is observed when the x/y =XNi2+

    /XDMG. The maximum

    occurs at aroundXNi2+ = 0.3. Thus x/y = 0.3/0.70 = 2/4.7 = 2/5.

    Mole-Ratio Method Description ofMethod

    In the mole-ratio method, a series of solutions is prepared in which the analytical

    concentrations of one reactant (usually the cation) is held constant while that of the other

    is varied. A plot of absorbance versus mole-ratio of the reactants is then made. If the

    formation constant is reasonably favourable, we obtain two straight lines of different

    slopes that intersect at a mole ratio corresponding to the combining ratio in the

    complex.[14]

  • 8/3/2019 Instru Compensate

    26/41

    26

    GraphTypical mole-ratio plot is shown in figure 5.2. Formation constants can be

    evaluated from the data in the curved portion of mole-ratio plot where the reaction is least

    complete.[14]

    Figure 5.2 Mole-ratio Method Plot (1:1)

    Sample ProblemThe accompanying data were obtained on a slope ratio investigation of the

    complex formed between NH4 and Quinol. The measurements were made at 420nm

    wavelength.

  • 8/3/2019 Instru Compensate

    27/41

    27

    Slope-Ratio Method Description of the Method

    This approach is particularly is useful for weak complexes but is applicable only

    to systems in which a single complex is formed. The method assumes (1) that the

    complex-formation reaction can be forced to completion by a large excess of either

    reactant, (2) that Beers Law is followed under these circumstances, and (3) that only the

    complex absorbs at the wavelength chosen for the experiment.[14]

    Consider the reaction in which the complex MxLy is formed by the reaction ofx

    moles of the cation M withy moles the ligand.

    xM+yL = MxLy

    Mass-balance expressions for this system are

    cM= [M] +x[MxLy]

    cL = [L] +y[MxLy]

    where cM and cL are the molar analytical concentrations of the two reactants. We now

    assume that at very high analytical concentration ofL, the equilibrium is shifted far to the

  • 8/3/2019 Instru Compensate

    28/41

    28

    right and [M]

  • 8/3/2019 Instru Compensate

    29/41

    29

    Graph

    Figure 5.3 A Plot of Slope-ratio Method

    Sample Problem

  • 8/3/2019 Instru Compensate

    30/41

    30

    VI. SpectrophotometricTitration Description

    In analytical chemistry, Spectrophotometric titration is an analytical method in

    which the radiant-energy absorption of a solution is measured spectrophotometrically

    after each increment of titrant is added.[15]

    Procedure Techniques forEnd-point determination

    The spectrophotometer is an optical device that responds only to radiation within

    a selected very narrow band of wavelengths in the visual, ultraviolet, or infrared regions

    of the spectrum. The response can be made both quantitatively and linearly related to the

    concentration of a species that absorbs radiation within this band. Titrations at

    wavelengths within the visual region are by far the most common.

    An example is the titration of iron(II) in dilute sulfuric acid solution with a

    standard solution of potassium permanganate. The spectrophotometer is adjusted to

    measure the absorbance of the highly colored titrant. Very little absorbance is observed

    until the end point is reached; iron(II) and the reaction products are essentially colorless

    and are not seen by the spectrophotometer. However, the absorbance rises as titration is

    continued beyond the end point because the titrant is no longer being destroyed.[16]

    The

    titration curve is shown in figure 6.1.

  • 8/3/2019 Instru Compensate

    31/41

    31

    Graph

    Figure 6.1 Spectrophotometric titration curve of iron(II) in dilute sulfuric acid solution

    with potassium permanganate as the titrant. The curve shows the response when the

    titrant absorbs beyond the end point.

    Sample Problem

    The concentrations of Fe3+ and Cu2+ in a mixture can be determined following their

    reaction with hexacyanoruthenate (II), Ru(CN)64, which forms a purple-blue complex

    with Fe3+ (max = 550 nm), and a pale green complex with Cu2+ (max = 396 nm) . The

    mo l a r absorptivities (M 1

    cm 1) for the metal complexes at the two wavelengths are

    summarized in the following table.

    Table 6.1 Molar absorptivities of Copper andIron

    550 396

    Fe3+

    9970 84

    Cu2+

    34 856

  • 8/3/2019 Instru Compensate

    32/41

    32

    When a sample containing Fe3+ and Cu2+ is analyzed in a cell with a pathlength

    of 1.00 cm, the absorbance at 550 nm is 0.183, and the absorbance at 396 nm is 0.109.

    What are the molar concentrations of Fe3+ and Cu2+ in the sample?

    Solution:

    (Am)1 = (X)1bCX+ (Y)1bCY [eq. 6.1]

    (Am)2 = (X)2bCX+ (Y)2bCY [eq. 6.2]

    Substituting known values into equations 6.1 and 6.2 gives

    A550 = 0.183 = 9970CFe + 34CCu

    A396 = 0.109 = 84CFe + 856CCu

    To determine the CFe and CCu we solve the first equation for CCu

    and substitute the result into the second equation.

    Solving for [Fe] gives the concentration of Fe3+

    as 1.80 105

    M. Substituting this

    concentration back into the equation for the mixtures absorbance at a wavelength of 396

    nm gives the concentration of[Cu] as 1.26 104

    M.

  • 8/3/2019 Instru Compensate

    33/41

    33

    VII. PotentiometricTitration Description

    In analytical chemistry, potentiometric titration is a titration in which an electrode

    reversible to one of the ionic components of the analyte or titrant is immersed in the

    analyte solution and the electric potential of the electrode (relative to that of an inert

    reference electrode) is measured during addition of titrant. Because the potential of an

    electrode is a logarithmic function of the activity of the ion to which the electrode is

    reversible, an abrupt change in its potential is observable as an equivalence point is

    approached.[17]

    Procedure Techniques forEnd-point determination

    If a pH meter is used, acid-base titration curves can be plotted or recorded. The

    meter and its associated electrodes are first standardized by use of a buffer solution of

    known pH. The electrodes are then immersed in the well-stirred solution to be titrated,

    and the titration is begun. For routine purposes, interest is in rapid and reasonably close

    end-point location. Titration is first carried out quite quickly until the meter shows signs

    of rapid response, and then is slowed so that it can be stopped at the pH jump or fall that

    marks the end point. Obviously, potentiometric titration can be used for a highly colored

    titrand solution, in which the response of a color-change indicator could not be seen.

    There are two approaches to higher precision. In the first the pH at which the

    desired end point occurs is determined, and the titration of the actual sample is then

    arrested exactly at, or very close to, this pH value. The other approach is to stop at the

    point of steepest slope of the titration curve. This can be done by plotting (or, with

    suitable instrumentation, sensing) the first derivative (pH/V) or the second derivative

    (2pH/V

    2) against the volume Vof titrant added. In theory this method is applicable to

    any potentiometric titration, without the need to predetermine the end-point conditions.

  • 8/3/2019 Instru Compensate

    34/41

    34

    By suitable choice of electrodes, these potentiometric methods can also be applied

    to combination titrations and to oxidation-reduction titrations. The advent of modern ion-

    selective electrodes has greatly extended the scope of potentiometric titration and other

    branches of titrimetry.[16]

    Graph

    Figure 7.1 Measurements, first and second derivative in a potentiometric titration.

  • 8/3/2019 Instru Compensate

    35/41

    35

    Sample ProblemCalculate the titration curve for the titration of 50.0 mL of 0.0500 MSn2+ with

    0.100 M Tl3+

    Both the titrand and the titrant are 1.0 M in .HCl. The titration reaction is

    Sn2+

    (aq)+ Tl3+

    (aq) Sn4+

    (aq) + Tl+(aq)

    Solution:

  • 8/3/2019 Instru Compensate

    36/41

    36

    VIII. Molecular Weight Determination using HPLC Description ofLiquid Chromatographic Methods employed in molecular weight

    determination

    High-performance liquid chromatography, HPLC, is a chromatographic

    technique that can separate a mixture of compounds and is used in biochemistry and

    analytical chemistry to identify, quantify and purify the individual components of the

    mixture. HPLC typically utilizes different types of stationary phases, a pump that moves

    the mobile phase(s) and analyte through the column, and a detector to provide a

    characteristic retention time for the analyte. Analyte retention time varies depending on

    the strength of its interactions with the stationary phase, the ratio/composition of

    solvent(s) used, and the flow rate of the mobile phase.

    The widely used technique for the molecular weight determination (usually

    polysaccharides) is the Size-exclusion chromatography (SEC), also known as gel

  • 8/3/2019 Instru Compensate

    37/41

    37

    permeation chromatography or gel filtration chromatography. It separates particles on

    the basis of size. It is generally a low resolution chromatography and thus it is often

    reserved for the final, "polishing" step of purification. SEC is used primarily for the

    analysis of large molecules such as proteins or polymers. SEC works by trapping these

    smaller molecules in the pores of a particle. The larger molecules simply pass by the

    pores as they are too large to enter the pores. Larger molecules therefore flow through the

    column quicker than smaller molecules, that is, the smaller the molecule, the longer the

    retention time.[19]

    ProcedureThe sample to be analyzed is introduced, in small volumes, into the stream of

    mobile phase. The solution moved through the column is slowed by specific chemical or

    physical interactions with the stationary phase present within the column. The velocity of

    the solution depends on the nature of the sample and on the compositions of the

    stationary (column) phase. The time at which a specific sample elutes (comes out of the

    end of the column) is called the retention time; the retention time under particular

    conditions is considered an identifying characteristic of a given sample. The use of

    smaller particle size column packing (which creates higher backpressure) increases the

    linear velocity giving the components less time to diffuse within the column, improving

    the chromatogram resolution. Common solvents used include any miscible combination

    of water or various organic liquids (the most common are methanol and acetonitrile).

    Water may contain buffers or salts to assist in the separation of the sample components,

    or compounds such as trifluoroacetic acid which acts as an ion pairing agent.

    A further refinement ofHPLC is to vary the mobile phase composition during theanalysis; gradient elution. A normal gradient for reversed phase chromatography might

    start at 5% methanol and progress linearly to 50% methanol over 25 minutes; the gradient

    depends on how hydrophobic the sample is. The gradient separates the sample mixtures

    as a function of the affinity. This partitioning process is similar to that which occurs

    during a liquid-liquid extraction but is continuous, not step-wise. In this example, using a

  • 8/3/2019 Instru Compensate

    38/41

    38

    water/methanol gradient, more hydrophobic components will elute (come off the column)

    when the mobile phase consists mostly of methanol (giving a relatively hydrophobic

    mobile phase).

    The choice of solvents, additives and gradient depend on the nature of the column

    and sample. Often a series of tests are performed on the sample together with a number of

    trial runs in order to find the HPLC method which gives the best peak separation.[19]

    Graph

    Figure 8.1 An example of a Plot of log mol. wt. versus Ve/Vo of a series of molecular

    weight standards

    Sample ProblemStandard solutions of polystyrene of known molecular weight were prepared and

    the retention time of each solution was measured using size exclusion chromatography.

    Prepare a plot of log MW versus retention time. Find the molecular weight of the

    unknown with a retention time of 14.55.

  • 8/3/2019 Instru Compensate

    39/41

    39

    Table 8.1 Retention time of Polystyrene

    MW Log MW Retention time (min)

    6.75x106

    6.829 12.78

    5.99 x106 6.777 13.28

    5.04 x106

    6.702 13.77

    4.27 x106

    6.630 14.20

    3.55 x106

    6.550 14.76

    Solution:

    Figure 8.2

    The equation of the line was found to be y = -0.1443x + 8.6836, let y = unknown

    molecular weight. To solve for y:

    y = -0.1443x + 8.6836

    y = -0.1443(14.55)+ 8.6836

    y = 6.5840

    y = 106.5840

    y = 3.84x106

    y = -0.1443x + 8.6836

    R = 0.9944

    6.5

    6.55

    6.6

    6.65

    6.7

    6.75

    6.8

    6.856.9

    12.5 13 13.5 14 14.5 15

    logMW

    Retention time (tR)

  • 8/3/2019 Instru Compensate

    40/41

    40

    LIST OF REFERENCES

    [1] http://en.wikipedia.org/wiki/Graph_%28mathematics%29

    [2] http://spreadsheets.about.com/od/spreadsheetlessons/ss/excel_graph_use.htm

    [3] http://www.ncsu.edu/scivis/chemistry.html

    [4] http://www.ground-water-

    models.com/products/aquachem_details/aquachem_details.html

    [5] http://www.cut-the-knot.org/Curriculum/Calculus/StraightLine.shtml

    [6] http://www.mathopenref.com/coordslope.html

    [7] http://www.mathopenref.com/coordintercept.html

    [8] http://en.wikipedia.org/wiki/Least_squares

    [9] http://en.wikipedia.org/wiki/Calibration_curve

    [10] http://www.chm.davidson.edu/vce/spectrophotometry/unknownsolution.html

    [11] http://docs.google.com/viewer?a=v&q=cache:scrY_M8zXxkJ:www.asi-

    sensors.com/ASI/learning/standard_addition.pdf+Standard+Addition+Method&hl=tl&gl

    =ph&pid=bl&srcid=ADGEESjAQVV1Lot9tOocsmrK3wHb1t8_tfz8WJq0dS3eBYGS0X

    o_xrOg8DvKxHpc-F9KB-wFlyKWo4cowgtHTLmgF02P6CaoHt6vGHuSvpXopPjRB9qT9-

    oagcQAs4LMWlAIZ6hLW7aX&sig=AHIEtbS1_9diVy7uWdJBWyyGn0kIHaAATA

    [12] http://en.wikipedia.org/wiki/Standard_addition

    [13]http://www.google.com.ph/url?sa=t&rct=j&q=how%2Bto%2Bprepare%2Ba%2Bstan

    dard%2BADDITION%2Bcurve&source=web&cd=4&ved=0CDMQFjAD&url=http%3A

    %2F%2Fwww.asdlib.org%2FonlineArticles%2Fecourseware%2FAnalytical%2520Chem

    istry%25202.0%2FText_Files_files%2FChapter5.pdf&ei=roStTuroDof9iQLZprWICw&

    usg=AFQjCNHhMgB5ZFVPsIcNIkPWZiEk-_6FlQ

    [14] Fundamentals of Analytical Chemistry, 8th

    ed., D. A. Skoog, D. M. West, F. J.

    Holler, and S. R. Crouch. Brooks/cole, a division of Thomson Learning, 2004.

    [15] http://www.answers.com/topic/spectrophotometric-titration

    [16] http://accessscience.com/content/Titration/699100

  • 8/3/2019 Instru Compensate

    41/41

    [17] http://www.answers.com/topic/potentiometric-titration-1

    [18] http://en.wikipedia.org/wiki/High-performance_liquid_chromatography#Size-

    exclusion_chromatography

    [19] http://www.investopedia.com/terms/l/line-of-best-fit.asp#ixzz1cLBcCTQD