installation & maintenance instructions minitrac 31

84
Installation & Maintenance Instructions MINITRAC 31 Foundation Fieldbus Radiation-based sensor for density measurement Reading Office Cutbush Park, Danehill, Lower Earley, Reading, Berkshire. RG6 4UT. UK. Tel: +44 (0)118 9311188 Email: [email protected] Aberdeen Office Unit 6 Airside Business Park, Kirkhill Industrial Estate, Dyce, Aberdeen. AB21 0GT. UK. Tel: +44 (0)1224 725999 Email: [email protected] Internet: www.able.co.uk e-procurement: www.247able.com Registered in England No: 01851002 VAT No: GB 417 2481 61

Upload: others

Post on 23-Mar-2022

1 views

Category:

Documents


0 download

TRANSCRIPT

Installation & Maintenance Instructions

MINITRAC 31Foundation Fieldbus

Radiation-based sensor for density measurement

Reading OfficeCutbush Park, Danehill, Lower Earley,Reading, Berkshire. RG6 4UT. UK.Tel: +44 (0)118 9311188 Email: [email protected]

Aberdeen OfficeUnit 6 Airside Business Park, Kirkhill Industrial Estate, Dyce, Aberdeen. AB21 0GT. UK.Tel: +44 (0)1224 725999Email: [email protected]

Internet: www.able.co.uke-procurement: www.247able.comRegistered in England No: 01851002VAT No: GB 417 2481 61

2

Contents

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Contents1 About this document

1.1 Function ........................................................................................................................... 41.2 Target group ..................................................................................................................... 41.3 Symbolism used ............................................................................................................... 4

2 For your safety2.1 Authorised personnel ....................................................................................................... 52.2 Appropriate use ................................................................................................................ 52.3 Warning about incorrect use ............................................................................................. 52.4 General safety instructions ............................................................................................... 52.5 CE conformity ................................................................................................................... 62.6 NAMUR recommendations .............................................................................................. 62.7 Environmental instructions ............................................................................................... 6

3 Product description3.1 Configuration .................................................................................................................... 73.2 Principle of operation........................................................................................................ 83.3 Packaging, transport and storage ................................................................................... 103.4 Accessories and replacement parts ............................................................................... 103.5 Corresponding source container .................................................................................... 11

4 Mounting4.1 General instructions ....................................................................................................... 134.2 Mounting instructions ..................................................................................................... 14

5 Connecting to power supply5.1 Preparing the connection ............................................................................................... 195.2 Connection-Density,massflowratemeasurement ....................................................... 225.3 Connection - Level detection .......................................................................................... 24

6 Set up with the display and adjustment module6.1 Insert display and adjustment module ............................................................................ 266.2 Adjustment system ......................................................................................................... 276.3 Parameter adjustment - Level measurement .................................................................. 276.4 Parameter adjustment - Density measurement .............................................................. 326.5 Parameter adjustment/Level detection ........................................................................... 446.6 Parameter adjustment - X-ray alarm ............................................................................... 526.7 Parameter adjustment/Real value correction .................................................................. 546.8 Saving the parameter adjustment data ........................................................................... 56

7 Setup with PACTware7.1 Connect the PC .............................................................................................................. 577.2 ParameteradjustmentwithPACTware ............................................................................ 577.3 Saving the parameter adjustment data ........................................................................... 58

8 Set up with other systems8.1 DD adjustment programs ............................................................................................... 598.2 Field Communicator 375, 475 ........................................................................................ 59

9 Diagnostics and service9.1 Maintenance .................................................................................................................. 609.2 Status messages ............................................................................................................ 609.3 Rectify faults ................................................................................................................... 63

3

Contents

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

9.4 Exchanging the electronics module ................................................................................ 659.5 Softwareupdate ............................................................................................................. 659.6 Howtoproceedincaseofrepair .................................................................................... 65

10 Dismounting10.1 Dismounting steps.......................................................................................................... 6710.2 Disposal ......................................................................................................................... 67

11 Supplement11.1 Technical data ................................................................................................................ 6811.2 Supplementary information Foundation Fieldbus ........................................................... 7211.3 Dimensions .................................................................................................................... 77

Safety instructions for Ex areasPleasenotetheEx-specificsafetyinformationforinstallationandop-eration in Ex areas. These safety instructions are part of the operating instructionsmanualandcomewiththeEx-approvedinstruments.Editing status: 2013-04-29

4

1 About this document

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

1 About this document

1.1 FunctionThis operating instructions manual provides all the information you needformounting,connectionandsetupaswellasimportantinstruc-tionsformaintenanceandfaultrectification.Pleasereadthisinforma-tion before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target groupThis operating instructions manual is directed to trained specialist personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism usedInformation, tip, noteThis symbol indicates helpful additional information.Caution:Ifthiswarningisignored,faultsormalfunctionscanresult.Warning:Ifthiswarningisignored,injurytopersonsand/orseriousdamage to the instrument can result.Danger:Ifthiswarningisignored,seriousinjurytopersonsand/ordestruction of the instrument can result.

Ex applicationsThis symbol indicates special instructions for Ex applications.

• ListThedotsetinfrontindicatesalistwithnoimpliedsequence.

→ ActionThisarrowindicatesasingleaction.

1 SequenceNumbers set in front indicate successive steps in a procedure.

Battery disposalThis symbol indicates special information about the disposal of bat-teries and accumulators.

5

2 For your safety

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

2 For your safety

2.1 Authorised personnelAll operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.Duringworkonandwiththedevicetherequiredpersonalprotectiveequipmentmustalwaysbeworn.

2.2 Appropriate useThe MINITRAC 31 is a sensor for density measurement and level detection.Youcanfinddetailedinformationontheapplicationrangeinchapter"Product description".Operational reliability is ensured only if the instrument is properly usedaccordingtothespecificationsintheoperatinginstructionsmanualaswellaspossiblesupplementaryinstructions.

2.3 Warning about incorrect useInappropriate or incorrect use of the instrument can give rise to application-specifichazards,e.g.vesseloverfillordamagetosystemcomponents through incorrect mounting or adjustment.

2.4 General safety instructionsThisisastate-of-the-artinstrumentcomplyingwithallprevailingregulations and guidelines. The instrument must only be operated in a technicallyflawlessandreliablecondition.Theoperatorisresponsiblefor the trouble-free operation of the instrument.During the entire duration of use, the user is obliged to determine the complianceofthenecessaryoccupationalsafetymeasureswiththecurrentvalidrulesandregulationsandalsotakenoteofnewregula-tions.The safety instructions in this operating instructions manual, the na-tionalinstallationstandardsaswellasthevalidsafetyregulationsandaccident prevention rules must be observed by the user.Forsafetyandwarrantyreasons,anyinvasiveworkonthedevicebeyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbi-traryconversionsormodificationsareexplicitlyforbidden.The safety approval markings and safety tips on the device must also be observed.This measuring system uses gamma rays. Therefore take note of the instructions for radiation protection in chapter "Product description". Allworkonthesourcecontainermayonlybecarriedoutunderthesupervisionofaqualifiedradiationprotectionofficer.

6

2 For your safety

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

2.5 CE conformityThedevicefulfillsthelegalrequirementsoftheapplicableECguide-lines.ByaffixingtheCEmarking,VEGAconfirmssuccessfultestingof the product.

Only with class A instruments:The device is a class A instrument designed for use in an industrial environment.Whenusedinadifferentenvironment,e.g.,inalivingarea, the electromagnetic compatibility must be ensured by the user. If necessary, suitable screening measures against conducted and emitted disturbances must be taken.Youcanfindtheconformitycertificateinthedownloadsectionunderwww.vega.com.

2.6 NAMUR recommendationsNAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are acceptedasthestandardinfieldinstrumentation.ThedevicefulfillstherequirementsofthefollowingNAMURrecom-mendations:

• NE21–Electromagneticcompatibilityofequipment• NE 43 – Signal level for malfunction information from measuring

transducers• NE53–Compatibilityoffielddevicesanddisplay/adjustment

components• NE107–Self-monitoringanddiagnosisoffielddevicesFor further information see www.namur.de.

2.7 Environmental instructionsProtection of the environment is one of our most important duties. Thatiswhywehaveintroducedanenvironmentmanagementsystemwiththegoalofcontinuouslyimprovingcompanyenvironmentalpro-tection.Theenvironmentmanagementsystemiscertifiedaccordingto DIN EN ISO 14001.Pleasehelpusfulfillthisobligationbyobservingtheenvironmentalinstructions in this manual:

• Chapter "Packaging, transport and storage"• Chapter "Disposal"

7

3 Product description

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

3 Product description

3.1 ConfigurationThenameplatecontainsthemostimportantdataforidentificationanduse of the instrument:

2

1

99

45

3

6

7

8

Fig. 1: Layout of the type label (example)1 Instrument version2 Product code3 Electronics4 Protection rating5 Ambient temperature6 Hardware and software version7 Order number8 Serial number of the instrument9 ID numbers, instrument documentation

With the serial number of the instrument on the type label you have accesstothefollowingdataonourhomepage:

• Article number of the instrument (HTML)• Delivery date (HTML)• Order-specificinstrumentfeatures(HTML)• Operating instructions at the time of shipment (PDF)• Order-specificsensordataforanelectronicsexchange(XML)• Testcertificate"MeasuringAccuracy"(PDF)For this purpose, move to www.vega.com and "VEGA Tools".As an alternative, you have access to these data via your Smart-phone:

• DownloadtheSmartphone-App"VEGA Tools" from the "Apple App Store " or the "Google Play Store"

• Scan the Data-Matrix-Code on the type label of the instrument or• Enter the serial number manually into the App

Thisoperatinginstructionsmanualappliestothefollowinginstrumentversions:

• Hardwarefrom1.0.4• Softwarefrom1.5.0• Modificationstatus,electronicsasof-01

Type plate

Serial number

Scope of this operating instructions manual

8

3 Product description

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Theinstrumentisavailableindifferentelectronicsversions.Eachver-sioncanbeidentifiedviatheproductcodeonthetypelabel:

• Standard electronics type PT30E-XX

The scope of delivery encompasses:

• Radiation-based sensor• Documentation

– this operating instructions manual – Operating instructions manual "Display and adjustment mod-

ule" (optional) – Ex-specific"Safety instructions"(withExversions) – ifnecessary,furthercertificates

3.2 Principle of operationTheinstrumentissuitableforapplicationsinliquidsandbulksolidsinvesselsunderdifficultprocessconditions.Thereareapplicationpos-sibilities in nearly all areas of industry.The measured value is detected contactlessly right through the vessel wall.Neitheraprocessfittingnoravesselopeningarerequired.Theinstrument is thus ideal for retro installation.Theinstrumentcanbeusedformanydifferentmeasuringtasks.Apartfrom the main applications such as density measurement and level detection, the MINITRAC 31 can also detect residues and the mass flowrateinconjunctionwithaflowmeter.

Electronics versions

Scope of delivery

Application area

9

3 Product description

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

1 2

3

4

Fig. 2: MINITRAC 31 - Application possibilities1 Level measurement - Residue detection2 Point level detection3 Density measurement4 Massflowmeasurement

Further application possibilities are also the use as X-ray alarm or real value correction.If X-ray alarm is selected, the instrument detects radiation from exter-nal sources. Possible external radiation sources can be, for example, aweldjointtestinaneighbouringfacilityorotherradiation-basedinstruments.When the instrument operates as real value correction it transmits the real value to correct another radiation-based sensor. The measure-ment can thus be adapted perfectly to the situation in the vessel.

In radiation-based measurement, a Caesium-137 or Cobalt-60 iso-topeemitsfocussedgammaraysthatareattenuatedwhenpenetrat-ingthetubewallandthemedium.TheNaIdetectorontheoppositeside, on a pipeline for example, receives the radiation. The intensity of the radiation is dependent on the density of the measured media. The measuring principle has proven to be very reliable in conjunction withextremeprocessconditionsbecauseitmeasurescontactlesslyfromoutsidethroughthetubewall.Themeasuringsystemensures

Functional principle

10

3 Product description

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

maximum safety, reliability and plant availability independent of the medium and its properties.

3.3 Packaging, transport and storageYourinstrumentwasprotectedbypackagingduringtransport.Itscapacity to handle normal loads during transport is assured by a test based on ISO 4180.The packaging of standard instruments consists of environment-friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.

Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.

The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or con-cealeddefectsmustbeappropriatelydealtwith.

Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.Unlessotherwiseindicated,thepackagesmustbestoredonlyunderthefollowingconditions:

• Not in the open• Dry and dust free• Not exposed to corrosive media• Protected against solar radiation• Avoiding mechanical shock and vibration

• Storage and transport temperature see chapter "Supplement - Technical data - Ambient conditions"

• Relative humidity 20 … 85 %

3.4 Accessories and replacement partsThe display and adjustment module PLICSCOM is used for measured value indication, adjustment and diagnosis. It can be inserted into the sensor or the external display and adjustment unit and removed at any time.Youcanfindfurtherinformationintheoperatinginstructions"Display and adjustment module PLICSCOM" (Document-ID 27835).

TheinterfaceadapterVEGACONNECTenablestheconnectionofcommunication-capable instruments to the USB interface of a PC. For parameteradjustmentoftheseinstruments,anadjustmentsoftwaresuchasPACTwarewithVEGA-DTMisrequired.Youcanfindfurtherinformationintheoperatinginstructions"Interface adapter VEGACONNECT" (Document-ID 32628).

Packaging

Transport

Transport inspection

Storage

Storage and transport temperature

PLICSCOM

VEGACONNECT

11

3 Product description

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

VEGADIS81isanexternaldisplayandadjustmentunitforsensorswithsingleordoublechamberhousing.For the double chamber housing there is also an interface adapter for VEGADIS81required.Youcanfindfurtherinformationintheoperatinginstructions"VE-GADIS 81" (Document-ID 43814).

The electronics module PT30E.XX is a replacement part for radiation-based sensors MINITRAC 31.TheelectronicsmodulecanonlybeexchangedbyVEGAservicetechnician.

3.5 Corresponding source containerAradioactiveisotopeinasuitablesourceholderistheprerequisitefora radiation-based measurement setup.Thehandlingofradioactivematerialisregulatedbylaw.Theradiationprotectionrulesofthecountryinwhichthesystemisoperatedapplyfirstandforemost.In Germany, for example, the current radiation protection ordinance (StrlSchV)basedontheAtomicEnergyLaw(AtG)applies.Thefollowingpointsareimportantformeasurementwithradiation-based methods:

Ahandlingpermitisrequiredforoperationofasystemusinggammarays.Thispermitisissuedbytherespectivegovernmentofficeortheresponsibleauthority(inGermany,forexample,officesforenviron-mental protection, trade supervisory boards, etc.)Youcanfindfurtherinstructionsintheoperatinginstructionsmanualof the source container.

When handling a radioactive source, unnecessary radiation exposure must be avoided. An unavoidable radiation exposure must be kept as lowaspossible.Takenoteofthefollowingthreeimportantmeasures:

1 2 3

Fig. 3: Measures for protection against radioactive radiation1 Shielding2 Time3 Distance

Shielding-Providegoodshieldingbetweentheradioactivesourceandyourselfaswellasallotherpersons.Specialsourcecontainers

VEGADIS 81

Electronics module

Handling permit

General instructions for radiation protection

12

3 Product description

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

(e.g.VEGASOURCE)aswellasallmaterialswithhighdensity(e.g.lead,iron,concrete,etc.)provideeffectiveshielding.Time: Stay as short a time as possible in radiation exposed areas.Source: Your distance to the source should be as large as possible. The local dose rate of the radiation decreases in proportion to the squareofthedistancetotheradiationsource.

Theplantoperatormustappointaradiationsafetyofficerwiththenecessaryexpertknowledge.Heisresponsiblethattheradiationprotection ordinance is maintained and that all radiation protection measures are implemented.

Controlareasareareasinwhichthelocaldoserateexceedsacertainvalue.Onlypersonswhoundergoofficialdosemonitoringareallowedintothesecontrolareas.Youcanfindtherespectivelyvalidlimitvaluesfor control areas in the guideline of the respective authority (in Ger-many, for example, the radiation protection ordinance).We are at your disposal for further information concerning radiation protection and regulations in other countries.

Radiationsafetyofficer

Control area

13

4 Mounting

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

4 Mounting

4.1 General instructionsThe source container is part of the measuring system. In case the sourcecontainerisalreadyequippedwithanactiveisotope,thesource container must be locked before mounting.

Danger:Before mounting; make sure that the source is securely closed. Use a padlock to secure the source container in the closed condition and prevent it from being inadvertently opened.

Protectyourinstrumentthroughthefollowingmeasuresagainstmoisture penetration:

• Use the recommended cable (see chapter "Connecting to power supply")

• Tighten the cable gland• LooptheconnectioncabledownwardinfrontofthecableglandThis applies particularly to:

• outdoor mounting• installationsinareaswherehighhumidityisexpected(e.g.through

cleaning processes)• installations on cooled or heated vessels

Make sure that all parts of the instrument exposed to the process are suitable for the existing process conditions.These are mainly:

• Active measuring component• Processfitting• Process seal

Process conditions are particularly:

• Process pressure• Process temperature• Chemical properties of the medium• AbrasionandmechanicalinfluencesYoucanfindthespecificationsinchapter"Technical data" and on the nameplate.

Inthecaseofinstrumenthousingswithself-sealingNPTthreads,itisnotpossibletohavethecableentriesscrewedinatthefactory.Theopeningsforthecableglandsarethereforecoveredwithredprotec-tive caps as transport protection.Priortosetupyouhavetoreplacetheseprotectivecapswithap-provedcableglandsorclosetheopeningswithsuitableblindplugs.Thesuitablecableglandsandblindplugscomewiththeinstrument.

Switchoffsource

Protection against mois-ture

Suitability for the process conditions

Protective caps

14

4 Mounting

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

4.2 Mounting instructionsNote:Duringtheplanningstage,ourspecialistswillanalysetheconditionsof the measuring point and dimension the source (isotope) accord-ingly.Yougeta"SourceSizing"documentspecifyingtherequiredsourceactivity and containing all relevant mounting information for your measuring point.Youmustfollowtheinstructionsofthis"SourceSizing"documentinadditiontothefollowingmountinginstructions.Thefollowingmountinginformationisapplicableaslongasthereisnothingelsespecifiedinthe"SourceSizing"document.Youcanfindinformationonprotectivebarriersandthemountingof the corresponding source container in the operating instructions manualofthesourcecontainer,e.g.VEGASOURCE.You can mount the MINITRAC 31 in any position. If you have ordered yourinstrumentwithaleadcoverasaprotectionagainstambientradiation (optionally), the the sensor is shielded laterally against X-ray radiation. In this case, the radiation can only penetrate frontally.Fastenthesensorsinsuchawaythattheycannotfalloutoftheholder.Direct the exit angle of the source container to the MINITRAC 31.Mount the source container as close as possible to the vessel. If there aregaps,securetheareawithasafetyfenceandprotectivegratingso that no one can reach into the dangerous area.

Caution:Makesurethatthetubeisalwayscompletelyfilled.Mainlyinhorizon-tally arranged tube measuring distances, air bubbles or buildup in thetubecaninfluencethemeasuringresult.Measurementshouldbepreferably through the centre of the tube.

1

2

Fig. 4: Installation on a horizontal pipeline1 Air bubbles2 Buildup

Installation position

15

4 Mounting

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Density measurementA density and concentration measurement is possible on pipelines and vessels. The accuracy of the measurement increases in propor-tion to the radiated length (L) of medium. This is particularly important inthecaseofproductswithlowdensityorsmalltubediameters.Therearedifferentwaystoincreasetheradiatedlength(L)ofthemedium.

L

L

L

1

4

a

b

2

3

Fig. 5: Installation possibilities - Density measurement or concentration meas-urement1a Direct radiation - Horizontal mounting1b Direct radiation - Vertical mouting or when the lead cover is used as protec-

tion against ambient radiation2 Inclined radiation for extention of the radiated length (L)3 Extension of the radiated length (L) by adding a tube angle piece as meas-

uring distance4 Integrated lead cover as protection against ambient radiation - the instru-

ment is hence shielded laterally

16

4 Mounting

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Withdensitymeasurement,theradiationdifferencewithdifferingdensityisverylow.Especiallywithsmalltubediameters,thechangeis negligible.Therefore it is important to shield from interfering X-ray radiation. To protecttheinstrumentagainstX-rayradiation,itcanbebeequippedwithanoptionalleadring.Alaterretrofittingoftheleadringisnotpossible.

MassflowrateThemassflowratecanbedeterminedwiththeMINITRAC31inconjunctionwithaflowmeter.

1

2

Fig.6:Massflowratemeasurement1 Flow meter2 MINITRAC 31

Point level detectionForleveldetection,thesensorisgenerallymountedhorizontallyattheheightoftherequestedlimitlevel.Makesurethattherearenostrutsor reinforcements at this position in the vessel.Directtheexitbeamofthesourcecontainerexactlytowardsthemeasuring range of MINITRAC 31.

17

4 Mounting

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

3

2

1

Fig. 7: Mounting position - level detection1 Mounting horizontal2 Mounting vertical3 Mounting horizontally, at right angles to container

Level measurement - Residue detectionThe MINITRAC 31 can be used for residue detection, e.g. in storage tanksforhigh-costliquids.Forthispurpose,theinstrumentmustbemountedatthelowestpointofthevessel.

18

4 Mounting

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Fig. 8: Level measurement - Residue detection on a storage tank

If the max. ambient temperature is exceeded, you must take suitable measures to protect the instrument against overheating.You can protect the instrument by providing a suitable insulation againsttheheatormountingtheinstrumentfurtherawayfromtheheat source.Make sure these measures are taken into account already in the plan-ningstage.Ifyouwanttocarryoutsuchmeasureslateron,contactour specialists to ensure that the accuracy of the application is not impaired.Ifthesemeasuresarenotsufficienttomaintainthemax.ambienttemperature,youcouldconsiderusingthewatercoolingsystemweofferforMINITRAC31.Thewatercoolingmustalsobeincludedinthecalculationsforthemeasuring point. Contact our specialists regarding the dimensioning ofthewatercooling.

Protection against heat

19

5Connectingtopowersupply

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

5 Connecting to power supply

5.1 Preparing the connectionAlwayskeepinmindthefollowingsafetyinstructions:

• Connect only in the complete absence of line voltage• If overvoltage surges are expected, overvoltage arresters should

be installed

Theinstrumentrequiresaoperatingvoltageof9…32VDC.Operat-ingvoltageandthedigitalbussignalarecarriedonthesametwo-wireconnectioncable.PowerissuppliedviatheH1powersupply.

ConnectioniscarriedoutwithscreenedcableaccordingtoFieldbusspecification.Usecablewithroundcross-section.Acableouterdiameterof5…9mm(0.2…0.35in)ensuresthesealeffectofthecablegland.Ifyouareusingcablewithadifferentdiameterorcross-section,exchange the seal or use a suitable cable gland.Make sure that the entire installation is carried out according to the Fieldbusspecification.Inparticular,makesurethatthebusistermi-natedwithsuitableterminatingresistors.

Generallyprovideallunusedcableentrieswithsuitableblindplugs.Thethinfoamrubberwahsersinthecableglandsareonlyusedasdust cover during trasport.

Inthecaseofinstrumenthousingswithself-sealingNPTthreads,itisgenerallynotpossibletohavethecableglandsscrewedinatthefactory.Theopeningsforthecableglandsarethereforecoveredwithred protective caps as transport protection.Youhavetoreplacetheseprotectivecapswithapprovedcableglandsbeforesetuporcoverthemwithsuitablefillerplugs.Unusedcableglandsdonotprovidesufficientprotectionagainstmoistureandmustbereplacedwithfillerplugs.Thesuitablecableglandsandblindplugscomewiththeinstrument.

Make sure that the cable screening and ground is executed accord-ingtotheFielbusspecification.IfelectromagneticinterferenceisexpectedwhichisabovethetestvaluesofEN61326-1forindustrialareas,werecommendtoconnectthecablescreenonbothendstoground potential.Withsystemswithpotentialequalisation,connectthecablescreendirectlytogroundpotentialatthepowersupplyunit,intheconnectionbox and at the sensor. The screen in the sensor must be connected directly to the internal ground terminal. The ground terminal outside onthehousingmustbeconnectedtothepotentialequalisation(lowimpedance).Insystemswithoutpotentialequalisationwithcablescreeningonboth sides, connect the cable screen directly to ground potential at

Safety instructions

Voltage supply

Connection cable

Cable entry

Cable gland ½ NPT

Cable screening and grounding

20

5Connectingtopowersupply

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

thepowersupplyunitandatthesensor.IntheconnectionboxorT-distributor, the screen of the short stub to the sensor must not be connected to ground potential or to another cable screen. The cable screenstothepowersupplyunitandtothenextdistributormustbeconnected to each other and also connected to ground potential via a ceramiccapacitor(e.g.1nF,1500V).Low-frequencypotentialequali-sationcurrentsarethussuppressed,buttheprotectiveeffectagainsthighfrequencyinterferencesignalsremains.

The voltage supply and signal output are connected via the spring-loaded terminals in the housing.The connection to the display and adjustment module or to the inter-face adapter is carried out via contact pins in the housing.

Proceedasfollows:Theprocedureappliestoinstrumentswithoutexplosionprotection.1. Unscrewthebighousingcover2. Loosen compression nut of the cable entry3. Remove approx. 10 cm (4 in) of the cable mantle, strip approx.

1cm(0.4in)ofinsulationfromtheendsoftheindividualwires4. Insert the cable into the sensor through the cable entry

1

Fig. 9: Connection steps 4 and 51 Locking of the terminal blocks

5. Insertasmallslottedscrewdriverfirmlyintotherectangularlockopenings of the respective connection terminal

Connection technology

Connection procedure

21

5Connectingtopowersupply

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

6. Insertthewireendsintotheroundopeningsoftheterminalsac-cordingtothewiringplan

Information:Solidcoresaswellasflexiblecoreswithcableendsleevesareinserteddirectlyintotheterminalopenings.Incaseofflexiblecoreswithoutendsleeves,presstherectangularlockopeningwithasmallscrewdriver;theterminalopeningisfreed.Whenthescrewdriverisreleased, the terminal opening closes again.7. Checktheholdofthewiresintheterminalsbylightlypullingon

themToloosenaline,insertasmallslottedscrewdriverfirmlyintotherectangular locking opening according to the illustration

8. Connect the screen to the internal ground terminal, connect the outergroundterminaltopotentialequalisation

9. Tighten the compression nut of the cable entry. The seal ring must completely encircle the cable

10. ScrewthehousingcoverbackonTheelectricalconnectionishencefinished.

Information:The terminal blocks are pluggable and can be detached from the electronics.Forthispurposeloosenthetwolaterallockingleversoftheterminalblockwithasmallscrewdriver.Whenlooseningthelock-ing,theterminalblockisautomaticallysqueezedout.Itmustsnapinplacewhenre-inserted.

22

5Connectingtopowersupply

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

5.2 Connection-Density,massflowratemeasurement

Non-Ex instruments and instruments with non-intrinsically safe current output

1 1

24

/L

/N

PE

56

2211 12

2122

1920

1817

1413

1615

12

2

6

1

43

5

7

8

910

Fig. 10: Electronics and connection compartment with non-Ex instruments and instruments with non-intrinsically safe current output1 Voltage supply2 Relay output3 Signal output FF bus4 Signal input 4 … 20 mA (active sensor)5 Switching input for NPN transistor6 Switchinginputfloating7 Transistor output8 Interface for sensor-sensor communication (MGC)9 Simulation switch (1 = simulation on)10 Setting the bus address for sensor-sensor communication (MGC)1)

5 6 7 8

2

1

Fig. 11: Adjustment and connection compartment with non-Ex instruments and instruments with non-intrinsically safe current output1 Terminals for the external display and adjustment unit2 Contact pins for the display and adjustment module or interface adapter

Instruments with intrinsically safe current outputYoucanfinddetailedinformationontheexplosion-protectedversions(Ex-ia,Ex-d)intheEx-specificsafetyinstructions.Thesesafety

Electronics and con-nection compartment - Non-Ex instruments and instruments with non-intrinsically safe current output

Adjustment and con-nection compartment - Non-Ex instruments and instruments with non-intrinsically safe current output

1) MGC = Multi Gauge Communication

23

5Connectingtopowersupply

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

instructionsarepartofthescopeofdeliveryandcomewiththeEx-approved instruments.

1 1

24

/L

/N

PE

56

2211 12

2122

1920

1817

1413

1615

12

2

5

1

3

4

6

7

89

Fig. 12: Electronics and connection compartment (Ex-d) with instruments with intrinsically safe current output1 Voltage supply2 Relay output3 Signal input 4 … 20 mA (active sensor)4 Switching input for NPN transistor5 Switchinginputfloating6 Transistor output7 Interface for sensor-sensor communication (MGC)8 Simulation switch (1 = simulation on)9 Setting the bus address for sensor-sensor communication (MGC)2)

5 6 7 8

2

3

4

1 2+( ) (-)

1

Fig. 13: Adjustment and connection compartment (Ex-ia) with instruments with intrinsically safe current output1 Terminals for intrinsically safe signal output FF bus2 Contact pins for the display and adjustment module or interface adapter3 Terminals for the external display and adjustment unit4 Ground terminal

Electronics and connec-tion compartment - In-struments with intrinsi-cally safe current output

Adjustment and connec-tion compartment - In-struments with intrinsi-cally safe current output

2) MGC = Multi Gauge Communication

24

5Connectingtopowersupply

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

5.3 Connection - Level detectionNon-Ex instruments and instruments with non-intrinsically safe current output

1 1

24

/L

/N

PE

56

2211 12

2122

1920

1817

1413

1615

12

2

6

1

43

5

7

8

910

Fig. 14: Electronics and connection compartment with non-Ex instruments and instruments with non-intrinsically safe current output1 Voltage supply2 Relay output3 Signal output FF bus4 Signal input 4 … 20 mA (active sensor)5 Switching input for NPN transistor6 Switchinginputfloating7 Transistor output8 Interface for sensor-sensor communication (MGC)9 Simulation switch (1 = simulation on)10 Setting the bus address for sensor-sensor communication (MGC)3)

5 6 7 8

2

1

Fig. 15: Adjustment and connection compartment with non-Ex instruments and instruments with non-intrinsically safe current output1 Terminals for the external display and adjustment unit2 Contact pins for the display and adjustment module or interface adapter

Instruments with intrinsically safe current outputYoucanfinddetailedinformationontheexplosion-protectedversions(Ex-ia,Ex-d)intheEx-specificsafetyinstructions.ThesesafetyinstructionsarepartofthescopeofdeliveryandcomewiththeEx-approved instruments.

Electronics and con-nection compartment - Non-Ex instruments and instruments with non-intrinsically safe current output

Adjustment and con-nection compartment - Non-Ex instruments and instruments with non-intrinsically safe current output

3) MGC = Multi Gauge Communication

25

5Connectingtopowersupply

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

1 1

24

/L

/N

PE

56

2211 12

2122

1920

1817

1413

1615

122

5

1

3

4

6

7

89

Fig. 16: Electronics and connection compartment (Ex-d) with instruments with intrinsically safe current output1 Voltage supply2 Relay output3 Signal input 4 … 20 mA (active sensor)4 Switching input for NPN transistor5 Switchinginputfloating6 Transistor output7 Interface for sensor-sensor communication (MGC)8 Simulation switch (1 = simulation on)9 Setting the bus address for sensor-sensor communication (MGC)4)

5 6 7 8

2

3

4

1 2+( ) (-)

1

Fig. 17: Adjustment and connection compartment (Ex-ia) with instruments with intrinsically safe current output1 Terminals for intrinsically safe signal output FF bus2 Contact pins for the display and adjustment module or interface adapter3 Terminals for the external display and adjustment unit4 Ground terminal

Electronics and connec-tion compartment - In-struments with intrinsi-cally safe current output

Adjustment and connec-tion compartment - In-struments with intrinsi-cally safe current output

4) MGC = Multi Gauge Communication

26

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

6 Set up with the display and adjustment module

6.1 Insert display and adjustment moduleThe display and adjustment module can be inserted into the sensor and removed again at any time. It is not necessary to interrupt the powersupply.Proceedasfollows:1. Unscrewthesmallhousingcover2. Place the display and adjustment module in the desired position

ontheelectronics(youcanchooseanyoneoffourdifferentposi-tions - each displaced by 90°)

3. Press the display and adjustment module onto the electronics and turn it to the right until it snaps in.

4. ScrewhousingcoverwithinspectionwindowtightlybackonRemoval is carried out in reverse order.Thedisplayandadjustmentmoduleispoweredbythesensor,anad-ditional connection is not necessary.

12

Fig. 18: Insert display and adjustment module

Note:Ifyouintendtoretrofittheinstrumentwithadisplayandadjustmentmoduleforcontinuousmeasuredvalueindication,ahighercoverwithaninspectionglassisrequired.

Mount/Dismount display and adjustment module

27

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

6.2 Adjustment system

1

2

Fig. 19: Display and adjustment elements1 LC display2 Adjustment keys

• [OK] key: – Movetothemenuoverview – Confirmselectedmenu – Edit parameter – Save value

• [->] key: – Presentation, change measured value – Select list entry – Select editing position

• [+] key: – Change value of the parameter

• [ESC] key: – Interrupt input – Jump to next higher menu

The sensor is adjusted via the four keys of the display and adjust-ment module. The LC display indicates the individual menu items. The functionsoftheindividualkeysareshownintheaboveillustration.Approx. 60 minutes after the last pressing of a key, an automatic reset tomeasuredvalueindicationistriggered.Anyvaluesnotconfirmedwith[OK]willnotbesaved.

6.3 Parameter adjustment - Level measurementThrough the parameter adjustment the instrument is adapted to the application conditions. The parameter adjustment is carried out via an adjustment menu.

Key functions

Adjustment system

28

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Information:Inthisoperatinginstructionsmanual,theinstrument-specificparam-eters are described. Further general parameters are described in the operating instructions manual "Display and adjustment module".

Caution:Duringthefirstsetuporafteraninstrumentresettheinstrumentstartswithpresetstandardvalues.Thesevaluearenotsuitableforyourap-plication and must be replaced by real values.Carryoutasetupinthesequencedescribedinthefollowing.

Themainmenuisdividedintofivesectionswiththefollowingfunc-tions:

Setup: Settings, e.g. for measurement loop name, isotope, applica-tion, background radiation, adjustment, signal outputDisplay: Settings, for example language, measured value displayDiagnosis: Information, for example, of device status, peak value, simulationAdditional adjustments: Instrument unit, reset, date/time, copying functionInfo:Instrumentname,hardwareandsoftwareversion,dateofmanu-facture, instrument features

Check if the correct language is already set for the display. If not, you can change the language in the menu item "Display/Language".

StartwiththesetupofMINITRAC31.In the main menu point "Setup", the individual submenu points should beselectedsubsequentlyandprovidedwiththecorrectparametersto ensure the optimum adjustment of the measurement. The proce-dureisdescribedinthefollowing.Possiblykeepthesequenceofthemenuitems.

Adjustment

In this menu item you can adjust the MINITRAC 31 to the integrated isotope in the source container.Forthispurpose,checkwhichisotopeisintegratedinthesourcecontainer.Youcanfindthisinformationonthetypelabelofthesourcecontainer.

Enter here, the respective application.

Instrument start

Main menu

Proceeding

Setup/Isotope

Setup - Application

29

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Thismenuitemenablesadaptationofthesensortotherequestedap-plication.Youcanchoosebetweenthefollowingapplications:"Level", "Density", "Limit level", "X-ray alarm" or "Real value correction".

Thenaturalradiationonearthinfluencestheaccuracyofthemeas-urement.With this menu item the natural background radiation can be faded out.For this purpose, the MINITRAC 31 measures the natural background radiationandsetsthepulseratetozero.Inthefuture,thepulseratefromthisbackgroundradiationwillbeautomatically deducted from the total pulse rate. This means: only thecomponentofthepulserateoriginatingfromthesourcewillbedisplayed.The source container must be closed for this setting.

In this menu item you can select the units of the process value and the temperature.

In this menu item you can enter the measuring range (min. and max. process value) of the sensor.Thesesettingsinfluencethecurrentoutputofthesensor.Enterinthemenuwindow"Max. process value" the max. level (full), for example in "m". This corresponds to an output current of 20 mA.Enterinthemenuwindow"Min. process value" the min. level (empty), for example in "m". This corresponds to an output current of 4 mA.

In this menu item you can carry out the adjustment of the sensor.Duetothemeasuringprinciple,thereisnolinearrelationshipbetweenpulserateandlevel.Hence,thisadjustment(i.e.linearization)mustinany case be carried out.

Note:Ifyoucannotfillthevesselwiththeoriginalmedium,itisalsopossibletocarryouttheadjustmentwithwater.Prerequisites:Theradiationisswitchedon-sourcecontainerissetto"On"Thevesseliseithercompletelyfilled(100%)orcompletelyemptied(0 %).Dependingonthefactifthevesselisfilledoremptied,youcancar-riedoutfirstthefullortheemptyadjustment.TheMINITRAC31sortsthe points automatically according to their level.Select "Show table"todisplayandeditthelinearizationpoints.Select "Linearization - New"toenterthefirstpoint.Select "Determine count rate"toenterthefirstpoint.The determination of the actual count rate lasts 2 minutes. After the count rate has been determined, you can accept the value (ct/s).Enternowthecorrespondinglevel(m).

Setup/Background radia-tion

Setup - Units

Setup - Adjustment

Setup - Linearization

30

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

By doing so, you assign a respective level to the actual count rate.

Acceptthevaluepairwith"OK".Dependingifyouhavestartedwithfulloremptyvessel,youhavetocontinueemptyingorfillingthevessel.Alsocarryoutsuchalinearizationwithseveraldifferentfillingheightsif you have a linear vessel.Amaximumof32linearizationpointsispossible.

In this menu item you can adjust the damping of the sensor. With ityoucansuppressfluctuationsinthemeasuredvalueindication,causede.g.byanagitatedproductsurface.Thistimecanbebetween1 and 1200 seconds. Keep in mind that also the reaction time in-creasesandtheinstrumentreactstoquicklevelchangeswithadelay.Generallyatimeofapproximately60secondsissufficienttosmooththe measured value indication.

Radiationfromexternalsourcescaninfluencethemeasuringresultofthe sensor.Possibleexternalradiationsourcescanbe,forexample,aweldjointtest on a neighbouring facility or other radiation-based instruments.An X-ray alarm is triggered if the impulses (ct/s) are more than 25 % abovethemax.valuefromthelinearizationtable.This fault message is only outputted for the period of the increased X-ray radiation. Then the fault message is automatically reset.In this menu item you can determine the behaviour of the sensor whenexternalradiationsourcesappear.

In this menu item you can activate the relay output and determine its functionaswellastheswitchingpoints.When the output of the process values is set, you can choose be-tweenoverfillanddryrunprotection.The relay outputs of the sensor react accordingly.You can choose "no" reference value. In this case, the relay output operates as fail safe relay.

Caution:Independentoftheselectedreferencevalue,therelaywilldeenergizein case of failure.

Additional settings

Whenaresetiscarriedout,allsettings(withonlyafewexceptions)are reset. The exceptions are: PIN, language and SIL.

Setup - Damping

Setup/External radiation alarm

Setup/Relay

Additional adjustments - Reset

31

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Thefollowingresetfunctionsareavailable:Basic adjustments: Resetting of the parameter adjustmetns to defaultvaluesatthetimeofshipment.Order-specificsettingswillbedeleted.Default settings: Resetting of the parameter adjustment like under "Basic adjustment".Inaddition,specialparameterswillberesettodefaultvalues.Order-specificsettingswillnotbedeleted.Peak values measured value: Resetting of the parameter adjust-ments in the menu item "Setup" to the default values of the respective instrument.Order-specificsettingsremainbutarenottakenoverintothe current parameters.Peak values temperature: Resetting of the measured min. and max. temperatures to the actual measured value.Thefollowingtableshowsthedefaultvaluesoftheinstrument.Thevalues apply for the application "Level". First of all you have to select the application.Depending on the instrument version, not all menu items may be availableortheymaybedifferentlyassigned:

32

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Menu Menu item Default value

Adjustment Measurement loop name Sensor

Isotope Cs-137

Applications Level

Background radiation 0 ct/s

Unit of the process value %

Temperature unit ° C

Adjustment min. Min. process value = 0 %

Adjustment max. Max. process value = 100 %

Linearization 0 ct/s = 100 %

90000 ct/s = 0 %

Damping 60 s

Real value correction 0

Mode Overfillprotection

Upperswitchingpoint-Pro-cess value

0 %

Lowerswitchingpoint-Pro-cess value

0 %

Upperswitchingpoint-Tem-perature

50 °C

Upperswitchingpoint-Tem-perature

25 °C

Reference value - Relay None

Block operation Released

Address - Summation Slave free

Display Language Selected language

Displayed value Pulse rate

Display unit ct/s

Additional set-tings

Temperature unit °C

Linearizationcurve Empty

HART mode StandardAddress 0

6.4 Parameter adjustment - Density measurement

Through the parameter adjustment the instrument is adapted to the application conditions. The parameter adjustment is carried out via an adjustment menu.

Information:Inthisoperatinginstructionsmanual,theinstrument-specificparam-eters are described. Further general parameters are described in the operating instructions manual "Display and adjustment module".

33

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

PrerequisitesThefollowingrequirementsmustbefulfilledforreliableandsafeoperation:

• Thetubemustbefilled.Theremustbenoairbubblesinthetube• Thesourcecontainerisswitchedon• A sample point is close to the measuring point

Warning:Whenthesourcecontainerisswitchedon,thetubemustalwaysbefilled.Ifthetubeisempty,therecanbeanincreasedlocaldoserate.Makesurethatthetubeisfilledevenincaseofplantdowntime,orswitchthesourcecontaineroff.

Carryoutasetupinthesequencedescribedinthefollowing.

Caution:Duringthefirstsetuporafteraninstrumentresettheinstrumentstartswithpresetstandardvalues.Thesevaluearenotsuitableforyourap-plication and must be replaced by real values.

Themainmenuisdividedintofivesectionswiththefollowingfunc-tions:

Setup: Settings, e.g. for measurement loop name, isotope, applica-tion, background radiation, adjustment, signal outputDisplay: Settings, for example language, measured value displayDiagnosis: Information, for example, of device status, peak value, simulationAdditional adjustments: Instrument unit, reset, date/time, copying functionInfo:Instrumentname,hardwareandsoftwareversion,dateofmanu-facture, instrument features

Check if the correct language is already set for the display. If not, you can change the language in the menu item "Display/Language".

StartwiththesetupofMINITRAC31.In the main menu point "Setup", the individual submenu points should beselectedsubsequentlyandprovidedwiththecorrectparametersto ensure the optimum adjustment of the measurement. The proce-dureisdescribedinthefollowing.Possiblykeepthesequenceofthemenuitems.

Instrument start

Main menu

Proceeding

34

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Adjustment

In this menu item you can adjust the MINITRAC 31 to the integrated isotope in the source container.Forthispurpose,checkwhichisotopeisintegratedinthesourcecontainer.Youcanfindthisinformationonthetypelabelofthesourcecontainer.

Through this selection, the sensitivity of the sensor is adapted perfectly to the isotope. The normal reduction of the emitter activity is hence considered through the radioactive decay.TheMINITRAC31requiresthisinformationoftheautomaticdecaycompensation. This ensures an interference-free measurement over the complete life time of the gamma emitter - an annual recalibration is not necessary.Entertherequestedparametersviatheappropriatekeys,saveyoursettingswith[OK]andjumptothenextmenuitemwiththe[ESC] and the [->] key.

Enter here, the respective application.Thismenuitemenablesadaptationofthesensortotherequestedap-plication.Youcanchoosebetweenthefollowingapplications:"Level", "Density", "Limit level", "X-ray alarm" or "Real value correction".

Thenaturalradiationonearthinfluencestheaccuracyofthemeas-urement.With this menu item the natural background radiation can be faded out.

Note:Make sure that some products have a self-radiation. This is for ex-amplethecasewithoilorpotashsaltlye.Thereforethetubemustbefilledtodeterminethebackgroundradiation.For this purpose, the MINITRAC 31 measures the natural background radiationandsetsthepulseratetozero.Inthefuture,thepulseratefromthisbackgroundradiationwillbeautomatically deducted from the total pulse rate. This means: only thecomponentofthepulserateoriginatingfromthesourcewillbedisplayed.The source container must be closed for this setting (OFF).

Setup/Isotope

Setup/Application

Setup/Background radia-tion

35

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

In this menu item you can select the units of the process value and the temperature.

In this menu item you can enter the measuring range (min. and max. process value) of the sensor.Thesesettingsinfluencethecurrentoutputofthesensor.

Enterinthemenuwindow"Max. process value" the max. density value, for example in "g/cm3". This corresponds to an output current of 20 mA.

Enterinthemenuwindow"Min. process value" the min. density value, for example in "g/cm3". This corresponds to an output current of 4 mA.

In this menu item you can enter the inside diameter of the tube or the radiated length (L).Thissettinginfluencestheaccuracyofthesensor.

First of all, select the unit of the inside diameter.

Setup/Units

Setup/Adjustment

Setup/Inside diameter

36

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Enterinthemenuwindow"Inside diameter" the inside diamter of the tube, for example in "cm".Ifthetubeisnotradiatedwith90°,thenyouhavetoentertheradiatedlength (L) instead of the tube inside diameter.Enteralsoheretheradiatedlengthwithoutthewallthicknessofthetube.

L

Fig. 20: With inclined mounting, the radiated length of the tube is applicable

In this menu item you can carry out the adjustment of the sensor.

Caution:Duringthefirstsetuporafteraninstrumentreset,thelinearizationstands at the preset value pair (90000 ct/s and 0.500 g/cm3). These values are not suitable for your application and must be replaced by realvalues.Deletethisvaluepairinthefollowingprocedureandcarryoutthelinearization.Duetothemeasuringprinciple,thereisnolinearrelationshipbetweenpulserateanddensity.Hence,thisadjustment(i.e.linearization)mustbe carried out in any event.Carryoutthisadjustmentwithseveralpointstoincreasetheaccuracyof the measurement.

Setup/Linearization

37

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Note:Waterhasasaknowndensityvalueof1g/cm3. Carry out the adjust-mentwithwater,ifpossible.

• Prerequisites:Theradiationisswitchedon-sourcecontainerissetto"On"Thetubeiscompletelyfilled.Possiblegasbubblesorairinclusionscaninfluencethemeasurement.The MINITRAC 31 sorts the points automatically according to their density.

Select "Show table"todisplayandeditthelinearizationpoints.

Select "Linearization - New"toenterthefirstpoint.

Select "Determine count rate"toenterthefirstpoint.

The determination of the actual count rate lasts 2 minutes. After the countratewasdetermined,youcanacceptthevalue.The count rate is stated in ct/s. This is the number of counts per sec-ond, i.e. the measured radioactive radiation dose actually reaching the sensor.

Enternowthecorrespondingdensityvalue(g/cm3).You thus assign a corresponding density to the actual count rate.

Note:If possible, take a product sample and determine the density simulta-neously in your sampling position.Ithasprovenmarkingtheproductsmapleswithdateandrespec-tivecountrate.Hencethevaluescanbeassignedlateronwithoutproblems.

38

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Acceptthevaluepairwith"OK".Enterasmanylinearizationpointsaspossible.Youcaninfluencethe accuracy of the density measurement by doing this. The more linearizationpointsyouenterandthehigherthedifferencebetweenthe density values, the more reliable the measurement.Valuepairsthatarenotyetcomplete,e.g.duetoamissingdensitydetermination,canbeeditedlaterwiththefunction"Setup/Lineariza-tion" under the item "Change/Edit".Amaximumof32linearizationpointsispossible.

Note:If you cannot change the medium during the adjustment process, it ispossibletocarryoutthelinearizationwithonlyonepoint.However,youshouldenterfurtherlinearizationpointslater,ifpossible.

• ShowdiagramThismenuitemisonlyavailableifalinearizationwasalreadycarriedout.

• ShowtableInthismenuitemyoucanshowtheindividualvaluepairsofthelinearization.

• Linearization-DeleteYoucanalsodeleteindividuallinearizationpoints.Enterthenumberofthepointyouwanttodelete.

• Linearization-ModifyYoucanalsomodifyindividuallinearizationpoints.

39

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

After editing, you have to activate the complete value pair so that the linearizationpointwillbeeffective.

Yourecognizeactivelinearizationpointsbyasmallsquarewithacrossnexttothenumberofthelinearizationpoint.

In this menu item you can adjust the damping of the sensor. With ityoucansuppressfluctuationsinthemeasuredvalueindication,causede.g.byanagitatedproductsurface.Thistimecanbebetween1 and 1200 seconds. Keep in mind that also the reaction time in-creasesandtheinstrumentreactstoquicklevelchangeswithadelay.Generallyatimeofapproximately60secondsissufficienttosmooththe measured value indication.With the setting "Automatic", the instrument itself calculates a suitable damping on the basis of the adjustment and the measured value changes.Thissettingisparticularlysuitableforapplicationwherefastandslowlevelchangesoccur.

The implementation of a real value corretion is only necessary if the requirementsofthemeasurementloophavechanged,forexamplewithabrasioninatube.Ifyouknowthedensityofacertainmedium,youcanenterthedeter-mined real density in this menu item to correct the measured value. Thefunctionshiftsthelinearizationcurvetothisdeterminedpoint.The measurement can thus be adapted exactly to the conditions in the tube.

In this menu item you can activate the relay output and determine its functionaswellastheswitchingpoints.

Setup/Damping

Setup/Real value correc-tion (manually)

Setup/Relay

40

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

When the output of the process values is set, you can choose be-tweenoverfillanddryrunprotection.The relay outputs of the sensor react accordingly.Thefollowingreferencevaluescanbeselected:

• None - Relay operates as fail safe relay• Electronics temperature• Process value

"No" reference value means that the relay output operates as fail safe relay.

Example for the adjustment of the process value

Caution:Independentoftheselectedreferencevalue,therelaywilldeenergizein case of failure.

With this menu item you safeguard the sensor parameters against unauthorizedorunintentionalmodifications.This menu item is described in the operating instructions manual "Display and adjustment module".

Display

With this parameter you can change the display language.This parameter is described in the operating instructions manual "Display and adjustment module".

With this parameter you can change the indication of the display.Youcanchooseifthedisplayshouldshowtheactualpulserate,theelectronics temperature or the percentage value.

Lock setup/adjustment

Display/Language

Display/Displayed value

41

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Diagnostics

Inthismenuitem,youcanenquirethestatusofyoursensor.Innormaloperation, the sensor displays the message "OK". In case of fault, you willfindthecorrespondingfaultcodehere.This parameter is described in the operating instructions manual "Display and adjustment module".

The peak value function holds the max. and min. values during opera-tion.This parameter is described in the operating instructions manual "Display and adjustment module".

Here you can retrieve the adjustment value of the sensor. This is the percentagevalueofthedifferenceofthemin.andmax.adjustmentpoints (Delta I). The value is an indication for the reliability and repro-ducibility of the measurement.Thehigherthedifferencebetweenthetwoadjustmentpoints,thehigherthedifferentialvalue(DeltaI)andthemorereliablethemeasurement.ADeltaIvaluebelow10%isanindicationforacriticalmeasurement.To increase the Delta I value, you have to increase the distance of the min.andmax.adjustmentpointsinthelinearization.

In this menu item you can simulate measured values via the current output.Thisallowsthesignalpathtobetested,e.g.viadownstreamindicating instruments or the input card of the control system.

Note:Tocarryoutasimulationwiththedisplayandadjustmentmodule,youhavetoswitchonthesimulationswitchontheelectronicsmodule(switchposition1).Youcanfindtherotaryswitchontheelectronicsmoduleintheelec-tronics and connection department (large cover).Youcansimulatedifferentvalues:

Pulse rate of the sensor

Process value

Diagnostics/Device status

Diagnosis/Peak value

Diagnosis/Adjustment data

Diagnosis/Simulation

42

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Switchingfunctionoftherelay

Information:The simulation is automatically terminated 10 minutes after the last press of a key.

Additional settings

In this menu item, the PIN is permanently activated/deactivated. Thus youprotectthesensordataagainstunauthorizedaccessandunin-tended changes. The default setting of the PIN is 0000.This parameter is described in the operating instructions manual "Display and adjustment module".

In this menu item you can set the actual date and time.This parameter is described in the operating instructions manual "Display and adjustment module".

Whenaresetiscarriedout,allsettings(withonlyafewexceptions)are reset. The exceptions are: PIN, language, SIL and HART mode.

Thefollowingresetfunctionsareavailable:Basic adjustments: Resetting of the parameter adjustmetns to defaultvaluesatthetimeofshipment.Order-specificsettingswillbedeleted.Default settings: Resetting of the parameter adjustment like under "Basic adjustment".Inaddition,specialparameterswillberesettodefaultvalues.Order-specificsettingswillnotbedeleted.Peak values measured value: Resetting of the parameter adjust-ments in the menu item "Setup" to the default values of the respective instrument.Order-specificsettingsremainbutarenottakenoverintothe current parameters.Peak values temperature: Resetting of the measured min. and max. temperatures to the actual measured value.

Additional adjustments/PIN

Additional adjustments/Date time

Additional adjustments/Reset

43

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Thefollowingtableshowsthedefaultvaluesoftheinstrument.Thevalues apply for the application "Level". First of all you have to select the application.Depending on the instrument version, not all menu items may be availableortheymaybedifferentlyassigned:

Menu Menu item Default value

Adjustment Measurement loop name Sensor

Isotope Cs-137

Applications Density

Background radiation 0 ct/s

Unit of the process value g/cm3

Temperature unit ° C

Adjustment min. process value

0.500 g/cm3

Adjustment max. process value

1.500 g/cm3

Inner diameter 0.20 m

Linearization 90000 ct/s and 0.500 g/cm3

Damping 60 s

Real value correction 0

Reference value - Relay None

Block operation Released

Display Language Selected language

Displayed value Pulse rate

Display unit ct/s

With this function

• Load parameter adjustment data from the sensor into the display and adjustment module

• Write parameter adjustment data from the display and adjustment module into the sensor

This parameter is described in the operating instructions manual "Display and adjustment module".

Info

Inthismenuyouwillfindthefollowingmenuitems:

• Instrumentname-showsinstrumentnameandserialnumber• Instrumentversion-showshardwareandsoftwareversionofthe

instrument• Dateofmanufacture-showscalibrationdateandthedateofthe

last change• DeviceID-showsthedeviceIDandthesensorTAG(PD_TAG)• Instrumentfeatures-showsfurtherinstrumentfeatures

Additional adjustments/Copy instrument settings

Info

44

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

These parameters are described in the operating instructions manual "Display and adjustment module".

6.5 Parameter adjustment/Level detectionThrough the parameter adjustment the instrument is adapted to the application conditions. The parameter adjustment is carried out via an adjustment menu.

Information:Inthisoperatinginstructionsmanual,theinstrument-specificparam-eters are described. Further general parameters are described in the operating instructions manual "Display and adjustment module".

Caution:Duringthefirstsetuporafteraninstrumentresettheinstrumentstartswithpresetstandardvalues.Thesevaluearenotsuitableforyourap-plication and must be replaced by real values.Carryoutasetupinthesequencedescribedinthefollowing.

Themainmenuisdividedintofivesectionswiththefollowingfunc-tions:

Setup: Settings, e.g. for measurement loop name, isotope, applica-tion, background radiation, adjustment, signal outputDisplay: Settings, for example language, measured value displayDiagnosis: Information, for example, of device status, peak value, simulationAdditional adjustments: Instrument unit, reset, date/time, copying functionInfo:Instrumentname,hardwareandsoftwareversion,dateofmanu-facture, instrument features

Check if the correct language is already set for the display. If not, you can change the language in the menu item "Display/Language".

StartwiththesetupofMINITRAC31.In the main menu point "Setup", the individual submenu points should beselectedsubsequentlyandprovidedwiththecorrectparametersto ensure the optimum adjustment of the measurement. The proce-dureisdescribedinthefollowing.Possiblykeepthesequenceofthemenuitems.

Instrument start

Main menu

Proceeding

45

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Adjustment

In this menu item you can adjust the MINITRAC 31 to the integrated isotope in the source container.Forthispurpose,checkwhichisotopeisintegratedinthesourcecontainer.Youcanfindthisinformationonthetypelabelofthesourcecontainer.

Through this selection, the sensitivity of the sensor is adapted perfectly to the isotope. The normal reduction of the emitter activity is hence considered through the radioactive decay.TheMINITRAC31requiresthisinformationoftheautomaticdecaycompensation. This ensures an interference-free measurement over the complete life time of the gamma emitter - an annual recalibration is not necessary.Entertherequestedparametersviatheappropriatekeys,saveyoursettingswith[OK]andjumptothenextmenuitemwiththe[ESC] and the [->] key.

Enter here, the respective application.Thismenuitemenablesadaptationofthesensortotherequestedap-plication.Youcanchoosebetweenthefollowingapplications:"Level", "Point level" or "Summation Slave".

Thenaturalradiationonearthinfluencestheaccuracyofthemeas-urement.With this menu item the natural background radiation can be faded out.For this purpose, the MINITRAC 31 measures the natural background radiationandsetsthepulseratetozero.Inthefuture,thepulseratefromthisbackgroundradiationwillbeautomatically deducted from the total pulse rate. This means: only thecomponentofthepulserateoriginatingfromthesourcewillbedisplayed.The source container must be closed for this setting.

In this menu item you can select the temperature unit.

Setup/Isotope

Setup - Application

Setup/Background radia-tion

Setup/Unit

46

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

inthismenuitemyoucanselectifyouwanttocarryoutasingleordouble point adjustment on the sensor.With the double point adjustment, the Delta I value is selected auto-matically.We recommend selecting the double point adjustment. To use this, you must be able to change the level of the vessel so as to carry out theadjustmentofthesensorwithfullstatus(covered)andwithemptystatus (uncovered).Hence,youwillgetaveryreliableswitchingpoint.Withsinglepointadjustment,youhavetodefinethedifferencebetweenthemin.andmax.adjustmentpoints(DeltaI)yourselfduringthefollowingsetup.

This menu item appears only if you have selected "Single point adjustment" as adjustment mode (Setup/Adjustment mode).InthismenuitemyoudeterminethepointatwhichtheMINITRAC31shouldswitchinuncoveredstatus.Empty the vessel until the sensor is uncovered.Forthisentertherequestedpulseratemanuallyorlettheratebedetermined by MINITRAC 31. Automatic determination of the pulse rate should be given preference.The pulse rate is entered in ct/s. This is the number of counts per second, i.e. the measured gamma radiation reaching the sensor.Prerequisites:

• Theradiationisswitchedon-sourcecontainerissetto"On"• Thereisnomediumbetweensourcecontainerandsensor

You can enter the value for "Adjustment uncovered" (ct/s) manually.

You can have the value for "Adjustment uncovered" determined by MINITRAC 31.

Setup/Adjustment mode

Setup/Adjustment uncovered (single point adjustment)

47

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

This menu item appears only if you have selected "Single point adjustment" as adjustment mode (Setup/Adjustment mode).Inthismenuitemyoucanadjustatwhichpercentagevalueofthemax.pulseratethesensorshouldswitchover.Since in most cases the radiation is almost completely absorbed whenthesensoriscovered,thepulseratewhenthesensoriscov-eredisverylow.Thechangebetweenthetwostatusesissufficientlyclear.Hence a percentage value of 90 % for the Delta I value is recom-mended.Youselectlowervaluesforthesensitivedetectionofmaterialconesorbuildupwhichonlycauseapartialabsorptionoftheradiation.

This menu item appears only if you have selected under adjustment mode (setup/adjustment mode) the "Double point adjustment".Inthismenuitemyoucansetthemin.pulserate(ct/s)atwhichthesensorshouldswitchover.Fill the vessel until the MINITRAC 31 is covered.You thus get the min. pulse rate (ct/s) for the adjustment covered.Entertherequestedpulseratemanuallyorlettheratebedeterminedby MINITRAC 31. Automatic determination of the pulse rate should be given preference.

You can enter the adjustment point (ct/s) manually.

You can let the adjustment point be determined by MINITRAC 31.

Setup/Delta I (single point adjustment)

Adjustment covered (double point adjust-ment)

48

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

This menu item appears only if you have selected under adjustment mode (setup/adjustment mode) the "Double point adjustment".Inthismenuitemyoucansetthemax.pulserate(ct/s)atwhichthesensorshouldswitchover.Empty the vessel until the MINITRAC 31 is uncovered.You thus get the max. pulse rate (ct/s) for the adjustment uncovered.Entertherequestedpulseratemanuallyorlettheratebedeterminedby MINITRAC 31. Automatic determination of the pulse rate should be given preference.

You can enter the adjustment point (ct/s) manually.

You can let the adjustment point be determined by MINITRAC 31.

Inthismenuitemyoucanselectwhichmodethesensorshouldoper-ate in.Youcanchoosebetweenoverfillanddryrunprotection.The relay outputs of the sensor react accordingly.Overfillprotection=therelaywilldeenergise(safecondition)whenthe max. level is reached.Dryrunprotection=therelaywilldeenergise(safecondition)whenthe min. level is reached.Make sure that you have selected the correct characteristics. See menu item "Setup/Current output mode".

With this menu item you safeguard the sensor parameters against unauthorizedorunintentionalmodifications.This menu item is described in the operating instructions manual "Display and adjustment module".

Display

With this parameter you can change the display language.

Adjustment uncovered (double point adjust-ment)

Setup/Relay

Lock setup/adjustment

Display/Language

49

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

This parameter is described in the operating instructions manual "Display and adjustment module".

With this parameter you can change the indication of the display.Youcanselectifthedisplayshouldshowtheactualpulserateoftheelectronics temperature.

Diagnostics

Inthismenuitem,youcanenquirethestatusofyoursensor.Innormaloperation, the sensor displays the message "OK". In case of fault, you willfindthecorrespondingfaultcodehere.This parameter is described in the operating instructions manual "Display and adjustment module".

The peak value function holds the max. and min. values during opera-tion.This parameter is described in the operating instructions manual "Display and adjustment module".

Here, you can retrieve the adjustment value of the sensor. This is the percentagevalueofthemax.pulserateatwhichthesensorswitchesover.If you have carried out a single point adjustment, this is the entered value. With a double point adjustment, this is the calculated value.The value is an indication for the reliability and reproducibility of the switchingpoint.Thegreaterthedifferenceinthepulseratebetweencoveredanduncoveredstatus,thegreaterthedifferentialvalue(DeltaI)andthemore reliable the measurement. The automatically calculated damp-ing is also oriented around the Delta I value. The higher the value, the lowerthedamping.ADeltaIvaluebelow10%isanindicationforacriticalmeasurement.

In this menu item you can simulate measured values via the current output.Thisallowsthesignalpathtobetested,e.g.viadownstreamindicating instruments or the input card of the control system.

Note:Tocarryoutasimulationwiththedisplayandadjustmentmodule,youhavetoswitchonthesimulationswitchontheelectronicsmodule(switchposition1).

Display/Displayed value

Diagnostics/Device status

Diagnosis/Peak value

Diagnosis/Adjustment data

Diagnosis - Simulation

50

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Youcanfindtherotaryswitchontheelectronicsmoduleintheelec-tronics and connection department (large cover).Youcansimulatedifferentvalues:

Pulse rate of the sensor

Switchingfunctionoftherelay

Information:10minutesafterthekeywaspressedforthelasttime,thesimulationisinterruptedautomatically.Youcanalsointerruptthesimulationwiththeswitchontheelectronicsmodule.

The sensor calculates a suitable integration time automatically.

Additional settings

In this menu item, the PIN is permanently activated/deactivated. Thus youprotectthesensordataagainstunauthorizedaccessandunin-tended changes. The default setting of the PIN is 0000.This parameter is described in the operating instructions manual "Display and adjustment module".

In this menu item you can set the actual date and time.This parameter is described in the operating instructions manual "Display and adjustment module".

Whenaresetiscarriedout,allsettings(withonlyafewexceptions)are reset. The exceptions are: PIN, language, SIL and HART mode.

Diagnosis/Calculated damping

Additional adjustments/PIN

Additional adjustments/Date time

Additional adjustments - Reset

51

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Thefollowingresetfunctionsareavailable:Basic adjustments: Resetting of the parameter adjustmetns to defaultvaluesatthetimeofshipment.Order-specificsettingswillbedeleted.Default settings: Resetting of the parameter adjustment like under "Basic adjustment".Inaddition,specialparameterswillberesettodefaultvalues.Order-specificsettingswillnotbedeleted.Peak values measured value: Resetting of the parameter adjust-ments in the menu item "Setup" to the default values of the respective instrument.Order-specificsettingsremainbutarenottakenoverintothe current parameters.Peak values temperature: Resetting of the measured min. and max. temperatures to the actual measured value.Thefollowingtableshowsthedefaultvaluesoftheinstrument.Thevalues apply for the application "Level". First of all you have to select the application.Depending on the instrument version, not all menu items may be availableortheymaybedifferentlyassigned:

Menu Menu item Default value

Adjustment Measurement loop name Sensor

Isotope Cs-137

Applications Limit level

Adjustment mode Single point adjustment

Adjustment - uncovered 90000 ct/s

Adjustment - covered 9000 ct/sonlywithtwo-pointadjust-ment

Delta I 90 %

Background radiation 0 ct/s

Temperature unit ° C

Adjustment mode 1-point

Adjustment uncovered 900000 ct/s

Delta I 90 %

Mode relay Overfillprotection

Block operation Released

Display Language Selected language

Displayed value Pulse rate

Display unit ct/s

52

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

With this function

• Load parameter adjustment data from the sensor into the display and adjustment module

• Write parameter adjustment data from the display and adjustment module into the sensor

This parameter is described in the operating instructions manual "Display and adjustment module".

Info

Inthismenuyouwillfindthefollowingmenuitems:

• Instrumentname-showsinstrumentnameandserialnumber• Instrumentversion-showshardwareandsoftwareversionofthe

instrument• Dateofmanufacture-showscalibrationdateandthedateofthe

last change• DeviceID-showsthedeviceIDandthesensorTAG(PD_TAG)• Instrumentfeatures-showsfurtherinstrumentfeaturesThese parameters are described in the operating instructions manual "Display and adjustment module".

6.6 Parameter adjustment - X-ray alarmX-rayradiation,forexample,causewithweldjointtestscaninfluenceyour measured values unnoticeably. Hence it is importnat to recog-nizedinterferingX-rayradiation.TheMINITRAC31recognizestheX-rayraditationandreactsfromanindividuallydefinableraditiationvolume.The level or density measuring instrument acts as Master and the MINITRAC 31 operates as Slave.Firstofall,definethefunctionoftheSlaveinstrumentbeforeyoudefinetheMasterinstrument.TheMasterinstrumentcanthusimme-diatelyrecognizetheconnectedSlave.Forthispurpose,theSlaveinstrumentmustbedefinedas"X-rayalarm". Select under the menu item "Setup/Application" the function "X-ray alarm".Youcansettheaddress(MGC)ontheSlaveinstrumentatwill.Onlythe address "0 - 0" is reserved for the Master instrument.Set the address setting (MGC) on the Master instrument to "0 - 0".

You have to enter the address of the Slave instrument in the list of the Master instrument. This function is not possible in the display and adjustmentmodule.TodothisyouneedPACTwarewiththerespectiveDTM.

Additional adjustments/Copy instrument settings

Info

X-ray alarm

53

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Themainmenuisdividedintofivesectionswiththefollowingfunc-tions:

Setup: Settings, e.g. for measurement loop name, isotope, applica-tion, background radiation, adjustment, signal outputDisplay: Settings, for example language, measured value displayDiagnosis: Information, for example, of device status, peak value, simulationAdditional adjustments: Instrument unit, reset, date/time, copying functionInfo:Instrumentname,hardwareandsoftwareversion,dateofmanu-facture, instrument features

Check if the correct language is already set for the display. If not, you can change the language in the menu item "Display/Language".

StartwiththesetupofMINITRAC31.In the main menu point "Setup", the individual submenu points should beselectedsubsequentlyandprovidedwiththecorrectparametersto ensure the optimum adjustment of the measurement. The proce-dureisdescribedinthefollowing.Possiblykeepthesequenceofthemenuitems.

Adjustment

In this menu item you can adjust the MINITRAC 31 to the integrated isotope in the source container.Forthispurpose,checkwhichisotopeisintegratedinthesourcecontainer.Youcanfindthisinformationonthetypelabelofthesourcecontainer.

With this selection, the sensitivity of the sensor is adapted in an opti-mumwaytotheisotope.TheMINITRAC31requiresthisinformationforthedecaycompensa-tion. An annual recalibration is hence not necessary.Entertherequestedparametersviatheappropriatekeys,saveyoursettingswith[OK]andjumptothenextmenuitemwiththe[ESC] and the [->] key.

Main menu

Proceeding

Setup/Isotope

54

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Enter here, the respective application.Thismenuitemenablesadaptationofthesensortotherequestedap-plication.Youcanchoosebetweenthefollowingapplications:"Level", "Density", "Limit level", "X-ray alarm" or "Real value correction".

With this menu item you safeguard the sensor parameters against unauthorizedorunintentionalmodifications.This menu item is described in the operating instructions manual "Display and adjustment module".

6.7 Parameter adjustment/Real value correctionTo correct a level measurement, a real value correction can be carried outwhenacertainlevelisreached.Thisisforexamplenecessaryincaseofrecurringbuilduponthevesselwall.The level measurement can thus be reset to the correct real value.The level measuring instrument acts as Master and the MINITRAC 31 operates as Slave.Firstofall,definethefunctionoftheSlaveinstrumentbeforeyoudefinetheMasterinstrument.TheMasterinstrumentcanthusimme-diatelyrecognizetheconnectedSlave.Forthispurpose,theSlaveinstrumentmustbedefinedas"Realvaluecorrection". Select under the menu item "Setup/Application" the func-tion "Real value correction".Youcansettheaddress(MGC)ontheSlaveinstrumentatwill.Onlythe address "0 - 0" is reserved for the Master instrument.The Master instrument must have the function "Level". For this purpose, select under the menu item "Setup/Application" the function "Level".Set the address setting (MGC) on the Master instrument to "0 - 0".

You have to enter the address of the Slave instrument in the list of the Master instrument. This function is not possible in the display and adjustmentmodule.TodothisyouneedPACTwarewiththerespectiveDTM.

Themainmenuisdividedintofivesectionswiththefollowingfunc-tions:

Setup/Application

Lock setup/adjustment

Real value correction

Main menu

55

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Setup: Settings, e.g. for measurement loop name, isotope, applica-tion, background radiation, adjustment, signal outputDisplay: Settings, for example language, measured value displayDiagnosis: Information, for example, of device status, peak value, simulationAdditional adjustments: Instrument unit, reset, date/time, copying functionInfo:Instrumentname,hardwareandsoftwareversion,dateofmanu-facture, instrument features

Check if the correct language is already set for the display. If not, you can change the language in the menu item "Display/Language".

StartwiththesetupofMINITRAC31.In the main menu point "Setup", the individual submenu points should beselectedsubsequentlyandprovidedwiththecorrectparametersto ensure the optimum adjustment of the measurement. The proce-dureisdescribedinthefollowing.Possiblykeepthesequenceofthemenuitems.

Adjustment

In this menu item you can adjust the MINITRAC 31 to the integrated isotope in the source container.Forthispurpose,checkwhichisotopeisintegratedinthesourcecontainer.Youcanfindthisinformationonthetypelabelofthesourcecontainer.

With this selection, the sensitivity of the sensor is adapted in an opti-mumwaytotheisotope.TheMINITRAC31requiresthisinformationforthedecaycompensa-tion. An annual recalibration is hence not necessary.Entertherequestedparametersviatheappropriatekeys,saveyoursettingswith[OK]andjumptothenextmenuitemwiththe[ESC] and the [->] key.

Enter here, the respective application.Thismenuitemenablesadaptationofthesensortotherequestedap-plication.Youcanchoosebetweenthefollowingapplications:"Level", "Density", "Limit level", "X-ray alarm" or "Real value correction".

Proceeding

Setup/Isotope

Setup/Application

56

6Setupwiththedisplayandadjustmentmodule

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

With this menu item you safeguard the sensor parameters against unauthorizedorunintentionalmodifications.This menu item is described in the operating instructions manual "Display and adjustment module".

6.8 Saving the parameter adjustment dataWe recommended noting the adjusted data, e.g. in this operating instructionsmanual,andarchivingthemafterwards.Theyarethusavailable for multiple use or service purposes.Iftheinstrumentisequippedwithadisplayandadjustmentmodule,the data in the sensor can be saved in the display and adjustment module. The procedure is described in the operating instructions manual "Display and adjustment module" in the menu item "Copy sensor data". The data remain there permanently even if the sensor powersupplyfails.Thefollowingdataorsettingsforadjustmentofthedisplayandad-justment module are saved:

• All data of the menu "Setup" and "Display"• In the menu "Additional adjustments" the items "Sensor-specific

units, temperature unit and linearization"• ThevaluesoftheuserprogrammablelinearizationcurveThe function can also be used to transfer settings from one instru-ment to another instrument of the same type. If it is necessary to exchange a sensor, the display and adjustment module is inserted intothereplacementinstrumentandthedataarelikewisewrittenintothe sensor via the menu item "Copy sensor data".

Lock setup/adjustment

57

7SetupwithPACTware

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

7 Setup with PACTware

7.1 Connect the PC

2

1

3

Fig. 21: Connection of the PC directly to the sensor via the interface adapter1 USB cable to the PC2 Interface adapter VEGACONNECT 43 ">Sensor

Information:TheinterfaceadapterVEGACONNECT3isnotsuitableforconnec-tion to the sensor.

7.2 Parameter adjustment with PACTwareForparameteradjustmentofthesensorviaaWindowsPC,thecon-figurationsoftwarePACTwareandasuitableinstrumentdriver(DTM)accordingtoFDTstandardarerequired.Theup-to-datePACTwareversionaswellasallavailableDTMsarecompiledinaDTMCollec-tion. The DTMs can also be integrated into other frame applications according to FDT standard.

Note:To ensure that all instrument functions are supported, you should alwaysusethelatestDTMCollection.Furthermore,notalldescribedfunctionsareincludedinolderfirmwareversions.Youcandownloadthelatestinstrumentsoftwarefromourhomepage.Adescriptionofthe update procedure is also available in the Internet.Further setup steps are described in the operating instructions manu-al "DTM Collection/PACTware" attached to each DTM Collection and whichcanalsobedownloadedfromtheInternet.Detaileddescrip-tionsareavailableintheonlinehelpofPACTwareandtheDTMs.

Via the interface adapter directly on the sensor

Prerequisites

58

7SetupwithPACTware

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Fig. 22: Example of a DTM view

All device DTMs are available as a free-of-charge standard version and as a full version that must be purchased. In the standard version, all functions for complete setup are already included. An assistant for simpleprojectconfigurationsimplifiestheadjustmentconsiderably.Saving/printingtheprojectaswellasimport/exportfunctionsarealsopart of the standard version.In the full version there is also an extended print function for complete projectdocumentationaswellasasavefunctionformeasuredvalueand echo curves. In addition, there is a tank calculation program as wellasamultiviewerfordisplayandanalysisofthesavedmeasuredvalue and echo curves.Thestandardversionisavailableasadownloadunderwww.vega.com/downloads and "Software". The full version is available on CD from the agency serving you.

7.3 Saving the parameter adjustment dataWe recommend documenting or saving the parameter adjustment dataviaPACTware.Thatwaythedataareavailableformultipleuseorservice purposes.

Standard/Full version

59

8Setupwithothersystems

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

8 Set up with other systems

8.1 DD adjustment programsDevice descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™ and PDM.Thefilescanbedownloadedatwww.vega.com/downloads under "Software".

8.2 Field Communicator 375, 475Device descriptions for the instrument are available as EDD for pa-rameteradjustmentwiththeFieldCommunicator375or475.

60

9 Diagnostics and service

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

9 Diagnostics and service

9.1 MaintenanceIfthedeviceisusedcorrectly,nomaintenanceisrequiredinnormaloperation.The corresponding source container must be checked in regular intervals.Youcanfindfurtherinformationintheoperatinginstructionsmanual of the source container.

9.2 Status messagesThe instrument features self-monitoring and diagnostics according toNE107andVDI/VDE2650.Inadditiontothestatusmessagesinthefollowingtablestherearemoredetailederrormessagesavailableunder the menu item "Diagnostics" via the display and adjustment module,PACTware/DTMandEDD.

Thestatusmessagesaredividedintothefollowingcategories:

• Failure• Function check• Outofspecification• Maintenancerequirementand explained by pictographs:

41 2 3Fig. 23: Pictographs of the status messages1 Failure - red2 Outofspecification-yellow3 Function check - orange4 Maintenance - blue

Failure: Due to a malfunction in the instrument, a failure message is outputted.Thisstatusmessageisalwaysactive.Itcannotbedeactivatedbytheuser.Function check: The instrument is in operation, the measured value is temporarily invalid (for example during simulation).This status message is inactive by default. It can be activated by the userviaPACTware/DTMorEDD.Outofspecification: The measured value is unstable because the instrumentspecificationisexceeded(e.g.electronicstemperature).This status message is inactive by default. It can be activated by the userviaPACTware/DTMorEDD.Maintenance:Duetoexternalinfluences,theinstrumentfunctionislimited.Themeasurementisaffected,butthemeasuredvalueis

Status messages

61

9 Diagnostics and service

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup).This status message is inactive by default. It can be activated by the userviaPACTware/DTMorEDD.

Thefollowingtableshowstheerrorcodesandtextmessagesinthestatus message "Failure"andprovidesinformationoncausesaswellas corrective measures.Example for a failure message

CodeText message

Cause Rectification

F008Error multi sensor communication

– Additionalsensorsnotswitchedon – EMCinfluences – No other sensor available

– Checkwiringbetweenthesensors – Connect the sensors correctly and make

them ready for operation

F013Sensor signals a fault

– Error on the current input – No valid measured value – Connectedinstrumentswithoutfunction

– Check current input – Check connected instruments (Slaves)

F016Adjustment data ex-changed

– Valuesofthemin.andmax.adjustmentexchanged

– Correct adjustment data

F017Adjustment span too small

– The values of the min. and max. adjust-ment are too close together

– Correct adjustment data

F025Invalidlinearizationtable

– Emptylinearizationtable – Wrongvalueinthelinearizationtable

– Createlinearizationtable – Correctlinearizationtable

F030Process value out of limits

– Processvaluesarenotwithintheadjusted measuring range

– Repeat adjustment

F034EPROMhardwareerror

– Electronics defective – Restart instrument – Exchanging the electronics

F035EPROM data error

– Error in the internal instrument com-munication

– Carry out a reset – Exchanging the electronics

F036Faulty program memory

– Errorduringsoftwareupdate – Repeatsoftwareupdate – Exchanging the electronics

F037RAMhardwareerror

– Error in RAM – Restart instrument – Exchanging the electronics

Failure

62

9 Diagnostics and service

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

CodeText message

Cause Rectification

F038Slave signals failure

– Connection cable to the Slave instrument interrupted

– InstrumentnotdefinedasSlaveinstru-ment

– Check the connection cable to the Slave instrument

– DefineinstrumentasSlave

F040Hardwareerror

– Sensor defective – Restart instrument – Exchanging the electronics

F041Photomultiplier error

– Error in the measured value recording – Restart instrument – Exchanging the electronics

F052Faultyconfiguration

– Invalid parameter adjustment – Carry out a reset

F066Faulty adjustment

– Adjustment not yet carried out – Errorduringadjustmentorwhenenteringthelinearizationtable

– Carrying out adjustment – Carryoutlinearization

F068Count rate too high

– Faulty instrument settings – Carry out a reset

F072Limit exceeded

– Faulty instrument settings – Carry out a reset

F073Error real value cor-rection

– Faulty real value correction – Carry out real value correction again

F080System error

– Instrument error – Restart instrument – Call our service

F086Communication error

– Error in the Fieldbus communication – Restart instrument – Call our service

F114Error real time clock

– Discharge accumulator – Readjust real time clock

F120Filter time error

– Faulty or missing instrument adjustment – Carrying out adjustment

F121Faulty participant list on the multisensor communication bus

– Slave instruments not found – Check Slave instruments – Check Slave list in Master instrument – Slaveinstrumentwithwrongaddress

F122Double addresses on the multisensor com-munication bus

– Instrumentaddressewasassignedseveral times

– Change instrument addresses

F123X-ray alarm

– External instruments cause radiation – Determine reason for X-ray alarm – In case of brief X-ray radiation: Monitor switchingoutputsforthistimemanually

F124Alarm due to in-creased radiation

– Radiation dose too high – Determine reason for increased radiation

63

9 Diagnostics and service

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

CodeText message

Cause Rectification

F125Ambient temperature too high

– Ambient temperature on the housing outsidethespecification

– Cooltheinstrumentorprotectitwithisolating material against radiation heat

Thefollowingtableshowstheerrorcodesandtextmessagesinthestatus message "Function check" and provides information on causes aswellascorrectivemeasures.

CodeText mes-sage

Cause Rectification

C029Simulation

– Simulation active – Finish simulation – Wait for the automatic end

after 60 mins.

Thefollowingtableshowstheerrorcodesandtextmessagesinthestatus message "Outofspecification" and provides information on causesaswellascorrectivemeasures.

CodeText mes-sage

Cause Rectification

S017Accuracy outside the specification

– Accuracy outside the specification

– Correct adjustment data

S025Badlineariza-tion table

– Badlinearizationtable – Carryoutlinearization

S038Slave outside thespecifica-tion

– Slaveoutsidethespecifica-tion

– Check Slave

S125Ambient tem-perature too high/toolow

– Ambient temperature too high/toolow

– Protectinstrumentwithisolating material against extreme temperatures

The instrument has no status messages to the section "Maintenance".

9.3 Rectify faultsThe operator of the system is responsible for taking suitable meas-ures to rectify faults.

Thefirstmeasuresare:

• Evaluation of fault messages via the indicating and adjustment module

• Checking the output signal

Function check

Outofspecification

Maintenance

Reaction when malfunc-tions occur

Procedure for fault recti-fication

64

9 Diagnostics and service

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

• Treatment of measurement errors

FurthercomprehensivediagnosispossibilitiesareofferedbyaPCwiththesoftwarePACTwareandsuitableDTM.In many case, the failure reasons can thus be determined and faults rectified.

Thefollowingtabledescribespossibleerrorswiththeoutputsignaland helps to remove them:

Error Cause Rectification

Output signal not stable Levelfluctuations Set damping according to the instrument via the display and adjustment module or PACT-ware/DTM

Output signal missing Electrical connection faulty Check connection according to chapter "Connection steps" and if necessary, correct ac-cording to chapter "Wiring plan"

Voltagesupplymissing Check cables for breaks; repair if necessary

Operatingvoltagetooloworload resistance too high

Check, adapt if necessary

Thefollowingtabledescribespossiblefaultsthatmaynotgenerateanerror message:

Error Cause Rectification

The instrument signals cov-eredwithoutcoveringwiththemediumThe instrument signals cov-eredwithcoveringwiththemedium

Voltagesupplymissing Check cables for breaks; repair if necessary

Operatingvoltagetooloworload resistance too high

Check, adapt if necessary

Electrical connection faulty Check connection according to chapter "Connection steps" and if necessary, correct ac-cording to chapter "Wiring plan"

Electronics defective Changetheswitchingbehaviourofthesensorsunder "Diagnosis/Simulation". If the instrument doesnotswitchover,senditinforrepair.

Buildupontheinnerwallofthe vessel

Remove buildupCheck the Delta I value.Improvetheswitchingthreshold-carryoutadouble point adjustment

Depending on the reason for the fault and the measures taken, the steps described in chapter "Setup" must be carried out again or must be checked for plausibility and completeness.

Should these measures not be successful, please call in urgent cases theVEGAservicehotlineunderthephoneno.+49 1805 858550.Thehotlineisalsoavailableoutsidenormalworkinghours,sevendaysaweekaroundtheclock.Sinceweofferthisserviceworldwide,thesupportisprovidedinEnglish. The service itself is free of charge, the only costs involved are the normal call charges.

Check output signal (level measurement)

Check output signal (level detection)

Reaction after fault recti-fication

24 hour service hotline

65

9 Diagnostics and service

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

9.4 Exchanging the electronics moduleIf the electronics module is defective, it can be replaced by the user.In Ex applications only one instrument and one electronics module withrespectiveExapprovalmaybeused.If there is no electronics module available on site, the electronics module can be ordered through the agency serving you. The electron-icsmodulesareadaptedtotherespectivesensoranddifferinsignaloutput or voltage supply.Thenewelectronicsmodulemustbeloadedwiththedefaultsettingsof the sensor. These are the options:

• In the factory• Or on site by the user

In both cases, the serial number of the sensor is needed. The serial numbers are stated on the type label of the instrument, on the inside ofthehousingaswellasonthedeliverynote.Whenloadingonsite,firstofalltheorderdatamustbedownloadedfrom the Internet (see operating instructions manual "Electronics module").

9.5 Software updateThefollowingcomponentsarerequiredtoupdatethesensorsoft-ware:

• Sensor• Voltagesupply• InterfaceadapterVEGACONNECT4• PCwithPACTware• CurrentsensorsoftwareasfileYoucanfindtheactualsensorsoftwareaswellasdetailedinfor-mation of the procedure under "www.vega.com/downloads" and "Software".Youcanfindinformationabouttheinstallationinthedownloadfile.

Caution:Instrumentswithapprovalscanbeboundtocertainsoftwarever-sions.Thereforemakesurethattheapprovalremainseffectivewithasoftwareupdate.Youcanfinddetailedinformationonwww.vega.com/downloads and "Approvals".

9.6 How to proceed in case of repairThefollowingprocedurerefersonlytothesensor.Shouldarepairofthesourcecontainerbenecessary,youcanfindtherespectiveinstructions in the operating instructions manual of the source con-tainer.Youcanfindarepairformaswellasdetailedinformationonhowtoproceed under www.vega.com/downloads and "Formsandcertifi-cates".

66

9 Diagnostics and service

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Bydoingthisyouhelpuscarryouttherepairquicklyandwithouthav-ing to call back for needed information.Ifasensorrepairisnecessary,pleaseproceedasfollows:

• Printandfilloutoneformperinstrument• Clean the instrument and pack it damage-proof• Attach the completed form and, if need be, also a safety data

sheet outside on the packaging• Please ask the agency serving you for the address of your return

shipment.Youcanfindthecompetentagencyonourwebsitewww.vega.com.

67

10 Dismounting

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

10 Dismounting

10.1 Dismounting stepsWarning:Beforedismounting,beawareofdangerousprocessconditionssuchas e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

10.2 DisposalTheinstrumentconsistsofmaterialswhichcanberecycledbyspe-cialised recycling companies. We use recyclable materials and have designed the parts to be easily separable.Correctdisposalavoidsnegativeeffectsonhumansandtheenviron-mentandensuresrecyclingofusefulrawmaterials.Materials: see chapter "Technical data"Ifyouhavenowaytodisposeoftheoldinstrumentproperly,pleasecontact us concerning return and disposal.

WEEE directive 2002/96/EGThis instrument is not subject to the WEEE directive 2002/96/EG and therespectivenationallaws.Passtheinstrumentdirectlyontoaspe-cialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.

68

11 Supplement

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

11 Supplement

11.1 Technical dataGeneral data316L corresponds to 1.4404 or 1.4435Materials,non-wettedparts

Ʋ Scintillation material NaI (sodium iodide) Ʋ Aluminium die-casting housing Aluminiumdie-castingAlSi10Mg,powder-coated-basis:

Polyester Ʋ Stainless steel housing 316L Ʋ Sealbetweenhousingandhousingcover

NBR (stainless steel housing, investment casting), silicone (Aluminium housing)

Ʋ Inspectionwindowinhousingcover(optional)

Polycarbonate

Ʋ Ground terminal 316LProcessfittings

Ʋ Fastening lugs ø 9 mm (0.35 in), hole centre distance 119 mm (4.69 in)Weight

Ʋ Aluminiumhousing,withelectronics 4.1 kg (9 lbs) Ʋ Stainlesssteelhousing,withelectron-ics

9.1 kg (20.1 lbs)

Max.torque,mountingscrews-fasteninglugs on the sensor housing

50 Nm (36.88 lbf ft)

Max.torqueforNPTcableglandsandConduittubes Ʋ Aluminium/Stainless steel housing 50 Nm (36.88 lbf ft)

Input variableMeasured variable The measured variable is the intensity of the gamma ra-

diation of an isotope. If the radiation intensity decreases, e.g. due to increasing product density, the measured value of MINITRAC 31 changes in proportion to the density.

Analogue input Ʋ Input type 4 … 20 mA, passive Ʋ Internal load 250Ω

Switchinginput Ʋ Type of input - Open Collector 10 mA Ʋ Type of input - Relay contact 100 mA

Output variable - Level measurementOutput

Ʋ Signal digital output signal, Foundation Fieldbus protocol Ʋ Physical layer according to IEC 61158-2

Damping (63 % of the input variable) 0 … 999 s, adjustable

69

11 Supplement

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Channel Numbers Ʋ Channel 1 Process value (level) Ʋ Channel 8 Electronics temperature Ʋ Channel 9 Pulse rate

Transmission rate 31.25 Kbit/sCurrent value 10 mA, ±0.5 mAResolution, digital > 0.1 mm (0.004 in)

Output variable - Density measurementOutput

Ʋ Signal digital output signal, Foundation Fieldbus protocol Ʋ Physical layer according to IEC 61158-2

Damping (63 % of the input variable) 0 … 999 s, adjustableChannel Numbers

Ʋ Channel 1 Process value (density) Ʋ Channel 3 Processtemperature(withexternaltemperaturesensor) Ʋ Channel 8 Electronics temperature Ʋ Channel 9 Pulse rate

Transmission rate 31.25 Kbit/sCurrent value 10 mA, ±0.5 mAResolution, digital > 0.1 mm (0.004 in)

Output variable - Level detectionOutput

Ʋ Signal digital output signal, Foundation Fieldbus protocol Ʋ Physical layer according to IEC 61158-2

Damping (63 % of the input variable) 0 … 999 s, adjustableChannel Numbers

Ʋ Channel 1 Processvalue(switchingstatus) Ʋ Channel 8 Electronics temperature Ʋ Channel 9 Pulse rate

Transmission rate 31.25 Kbit/sCurrent value 10 mA, ±0.5 mAResolution, digital > 0.1 mm (0.004 in)

Output variable - X-ray alarmOutput

Ʋ Signal digital output signal, Foundation Fieldbus protocol Ʋ Physical layer according to IEC 61158-2

Damping (63 % of the input variable) 0 … 999 s, adjustableChannel Numbers

Ʋ Channel 1 Process value

70

11 Supplement

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Ʋ Channel 8 Electronics temperature Ʋ Channel 9 Pulse rate

Transmission rate 31.25 Kbit/sCurrent value 10 mA, ±0.5 mAResolution, digital > 0.1 mm (0.004 in)

Output variable - Real value correctionOutput

Ʋ Signal digital output signal, Foundation Fieldbus protocol Ʋ Physical layer according to IEC 61158-2

Damping (63 % of the input variable) 0 … 999 s, adjustableChannel Numbers

Ʋ Channel 1 Process value Ʋ Channel 8 Electronics temperature Ʋ Channel 9 Pulse rate

Transmission rate 31.25 Kbit/sCurrent value 10 mA, ±0.5 mAResolution, digital > 0.1 mm (0.004 in)

Relay outputOutput Relayoutput(SPDT),floatingspdtSwitchingvoltage

Ʋ Min. 10mV Ʋ Max. 253VAC,253VDC

Switchingcurrent Ʋ Min. 10 µA Ʋ Max. 3 A AC, 1 A DC

Breaking capacity Ʋ Min. 50 mW Ʋ Max. 750VAAC,40WDC

Ifinductiveloadsorstrongercurrentsareswitchedthrough,thegoldplatingontherelaycontactsurfacewillbe permanently damaged. The contact is then no longer suitableforswitchinglow-levelsignalcircuits.

Contact material (relay contacts) AgNi or AgSnO and Au plated

Transistor outputOutput Floating transistor output, permanently shortcircuit-proofLoad current < 400 mAVoltageloss <1VSwitchingvoltage <55VDCBlocking current < 10 µA

71

11 Supplement

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Accuracy (according to DIN EN 60770-1)Process reference conditions according to DIN EN 61298-1

Ʋ Temperature +18 … +30 °C (+64 … +86 °F) Ʋ Relative humidity 45 … 75 % Ʋ Air pressure 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)

Repeatability ≤0.1%Deviationwithbulksolids The values depend to a great extent on the application.

Bindingspecificationsarethusnotpossible.DeviationunderEMCinfluence ≤1%

VariablesinfluencingmeasurementaccuracyTemperature drift - Digital output ±3 mm/10 K relating to the max. measuring range or

max. 10 mmAdditional deviation through electromag-netic interference acc. to EN 61326

< ±50 mm

Characteristics and performance dataStep response time5) ≤5s(withdamping1s)

Ambient conditionsAmbient, storage and transport tempera-ture

-40 … +60 °C (-40 … +140 °F)

Process conditionsFortheprocessconditions,pleasealsonotethespecificationsonthetypelabel.Thelowervaluealwaysapplies.Process pressure UnpressurizedProcess temperature (measured on the detector tube)

-40 … +60 °C (-40 … +140 °F)Withtemperaturesofmorethan60°Cwerecommendtheuseofwatercooling

Vibrationresistance6) mechanicalvibrationsupto1ginthefrequencyrange5…200Hz

Electromechanical data - version IP 66/IP 67Cable entry

Ʋ M20 x 1.5 2 x cable gland M20 x 1.5 (cable: ø 6 … 12 mm), 4 x blind plug M20 x 1.5Included: 1 x cable gland M20 x 1.5

Ʋ ½ NPT 5 x closing cap (red) ½ NPTIncluded: 3 x cable gland ½ NPT (cable: ø 6 … 12 mm), 4 x blind plug ½ NPT

Spring-loadedterminalsforwirecross-section Ʋ Massivewire,cord 0.2 … 2.5 mm² (AWG 24 … 14) Ʋ Strandedwirewithendsleeve 0.2 … 1.5 mm² (AWG 24 … 16)

5) Timespanafterasuddenmeasuringdistancechangebymax.0.5minliquidapplications,max2mwithbulksolidsapplications,untiltheoutputsignalhastakenforthefirsttime90%ofthefinalvalue(IEC61298-2).

6) Tested according to the guidelines of German Lloyd, GL directive 2.

72

11 Supplement

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Display and adjustment moduleDisplay element DisplaywithbacklightMeasured value indication

Ʋ Number of digits 5 Ʋ Sizeofdigits W x H = 7 x 13 mm

Adjustment elements 4 keysProtection rating

Ʋ unassembled IP 20 Ʋ mountedintothehousingwithoutcover

IP 40

Material Ʋ Housing ABS Ʋ Inspectionwindow Polyester foil

Integrated clockDate format Day.Month.YearTime format 12 h/24 hTimezoneExfactory CET

Measurement electronics temeratureResolution 1 °C (1.8 °F)Accuracy ±1 °C (1.8 °F)

Voltage supplyOperating voltage

Ʋ Sensorpowersupply 20…72VDCor20…253VAC,50/60Hz Ʋ FF bus 9…32VDC

Electrical protective measuresProtection, depending on housing ver-sion

IP 66/IP 677)

Overvoltage category IIIProtection class I

ApprovalsInstrumentswithapprovalscanhavedifferenttechnicaldatadependingontheversion.For that reason the associated approval documents of these instruments must be carefully noted. Theyarepartofthedeliveryorcanbedownloadedunderwww.vega.com and "VEGA Tools" as wellasunder"Downloads" and "Approvals".

11.2 Supplementary information Foundation FieldbusThefollowingtablegivesyouanoverviewoftheinstrumentversionsandthecorrespondingdevicedescriptions,theelectricalcharacteristicsofthebussystemaswellastheappliedfunctionblocks.

7) Asuitablecableistheprerequisiteformaintainingtheprotectionrating.

73

11 Supplement

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Revisions Data DD-Revision Rev_01

CFF-File 010101.cff

Device Revision 0101.ffo0101.sym

Cff-Revision xx xx 01

Devicesoftwarerevision > 1.5.0

ITK (Interoperability Test Kit) Number 6.0.1

Electricial Characteristics Physicial Layer Type Low-powersignaling,bus-pow-ered, FISCO I.S.

Input Impedance >3000Ohmsbetween7.8KHz-39KHz

Unbalanced Capacitance < 250 pF to ground from either in-put terminal

Output Amplitude 0.8VP-P

Electrical Connection 4 Wire

Polarity Insensitive Yes

Max. Current Load 10 mA

Device minimum operating voltage 9V

Transmitter Function Blocks Resource Block (RB) 1

Transducer Block (TB) 1

Standard Block (AI) 3

Execution Time 30 ms

Diagnostics Standard Yes

Advanced Yes

Performance No

Function Blocks Instantiable No

General Information LAS (Link Active Scheduler) No

Master Capable No

NumberofVCRs(VirtualCommunicationRelationships)

24

Function block Analog Input (AI)The function block "Analog Input (AI)" takes the original measured value selected by a Channel Number and makes it available to additional function blocks on its output.

74

11 Supplement

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Fig. 24: Schematic presentation function block Analog Input (AI)

Parameter listThefollowingtablegivesyouanoverviewoftheparametersused.

FF descriptor Rel. In-dex

Description Unit

PRIMARY_VALUE 13 PRIMARY_VALUE(Linearizedvalue).Thisistheprocess value after min/max adjustment and Lin-earizationwiththestatusofthetransducerblock.Theunitisdefinedin"PRIMARY_VALUE_UNIT"

FF_PRIMARY_VAL-UE_UNIT

FF_PRIMARY_VALUE_UNIT

14 Selected unit code for "PRIMARY_VALUE" -

FF_VAPOR_DENSITY 15 DensitywithTemperaturecorrection FF_VAPOR_DENSI-TY_UNIT

FF_VAPOR_DENSITY_UNIT

16 Selected unit code for “FF_VAPOR_DENSITY” -

FF_PROCESS_TEM-PERATURE

17 Process temperature FF_PROCESS_TEM-PERATURE_UNIT

FF_PROCESS_TEM-PERATURE_UNIT

18 Selected unit code for “FF_PROCESS_TEMPER-ATURE”

-

FF_DENSITY 19 Density FF_DENSITY_UNIT

FF_DENSITY_UNIT 20 Selected unit code for “FF_DENSITY” -

FF_VOLUMETRIC_FLOW

21 Volumetricflow FF_VOLUMETRIC_FLOW_UNIT

FF_VOLUMETRIC_FLOW_UNIT

22 Selected unit code for “FF_VOLUMETRIC_FLOW” -

FF_WEIGHT 23 Weight on belt FF_WEIGHT_UNIT

FF_WEIGHT_UNIT 24 Selected unit code for “FF_WEIGHT” -

FF_BELT_SPEED 25 Belt speed FF_BELT_SPEED_UNIT

FF_BELT_SPEED_UNIT

26 Selected unit code for “FF_BELT_SPEED” -

FF_ELECTRONIC_TEMPERATURE

27 Electronics temperature FF_ELECTRONIC_TEMPERATURE_UNIT

75

11 Supplement

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

FF descriptor Rel. In-dex

Description Unit

FF_ELECTRONIC_TEMPERATURE_UNIT

28 Selected unit code for “FF_ELECTRONIC_TEM-PERATURE”

-

FF_COUNT_RATE 29 Count rate FF_COUNT_RATE_UNIT

FF_COUNT_RATE_UNIT

30 Selected unit code for “FF_COUNT_RATE” -

DEVICE_TAG 31 Tagname -

DEVICE_NAME 32 Device type -

DEVICE_STATE 33 Error code -

PEAK_MEAS_VAL_MIN 34 Pulse rate (min.) -

PEAK_MEAS_VAL_MAX

35 Pulse rate (max.) -

PEAK_TEMP_VAL_MIN 36 Electronics temperature (min.) -

PEAK_TEMP_VAL_MAX

37 Electronics temperature (max.) -

APPLICATION_TYPE_SEL

38 Selected application -

TEMP_COMP_UNIT 39 Selected unit code for process temperature -

DELTA_I 40 Calculated percent delta I -

GAUGE_TEMPERA-TURE

41 Electronics temperature -

DECAY_COMPENSA-TION_FACTOR

42 Factor for the decay compensation -

PMT_VOLTAGE_CALI-BRATION

43 Photomultiplier voltage on delivery -

CORRELATION_CO-EFF

44 Correlationcoefficientforlinearizertable -

ERROR_TEXT 45 Error text -

PMT_VOLTAGE_AC-TUAL

46 Current photomultiplier voltage -

STANDARDIZATION_FACTOR

47 Factor for the real value correction -

SERIALNUMBER 48 Serial number -

NAMUR_STATE 49 NAMUR state -

NULL_COUNT_RATE 50 Zero count rate -

COUNT_RATE_PMT 51 Pulseratephotomultiplier(rawvalues) -

ADJ_DENSITY_ABS_COEFF

52 Processabsorptioncoefficient -

DEV_SW_VER_ASCII 53 Softwareversion -

POINT_LEVEL_AD-JUST_MODE

54 Point level adjustment mode -

RELAY_VALUE_SEL 55 Relay basic value -

76

11 Supplement

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

FF descriptor Rel. In-dex

Description Unit

DIGITAL_IN 56 Frequencyofdigitalinput -

DIGITAL_IN_BOOL 57 State of digital input -

ANALOG_IN 58 Current on analog input -

FF_CHANNEL_AVAIL-ABLE

59 Available channels -

FF_CHANNEL_USED 60 Used channels -

Mapping of Process Value StatusHex Quality Sub-Status Condition

0x00 Bad Non-specific Unexpected error

0x01 Bad ConfigurationError ItwastriedtosetawrongunitwithFFinterfaceOn user error codes: 16, 17, 25, 52, 57, 66, 72, 117, 120

0x0C Bad Device Failure Five or more internal communications have failed while"ProcessDataUpdate".LastusablevalueisdisplayedOn user error codes: 8, 34, 35, 36, 37, 38, 73, 80, 86, 121, 122, 141

0x10 Bad Sensor Failure On user error codes: 40, 41, 53, 68, 123, 124, 125

0x1C Bad Out of Service Transducer block is in mode "Out of Service"Channel is not assigned to an AIFBChannel is not available in running application

0x20 Bad Transducer in MAN Transducer block is in mode "Manual"

0x44 Uncertain LastUsableValue Three or more internal communications have failed while"ProcessDataUpdate".Lastusablevalueisdisplayed

0x48 Uncertain Substitute On user error codes: 29

0x4C Uncertain InitialValue After startup of device or channel assignment in AIFBwaschanged.Thechannelunitmaybeun-knownuntilnext"ProcessDataUpdate"On user error codes: 13

0x80 Good (NC) Non-specific No errors concerning to channel handlingOn user error codes: 0, 33, 45, 71, 126, 127

Mapping of User Error Codes to FF Field DiagnosticsPriority Description User Error Codes NE-107 Status

31 Hardwarefailure 40, 41 FAILURE

30 Memory failure 34, 35, 36, 37 FAILURE

29 Softwarefailure 80 FAILURE

28 Paramererizationcorrupt 72 FAILURE

27 Undefined27 - -

77

11 Supplement

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Priority Description User Error Codes NE-107 Status

26 Undefined26 - -

25 Paramererizationerror 16, 17, 25, 52, 53, 57, 66, 117, 120

FAILURE

24 ConflictinMGC 121, 122, 141 FAILURE

23 Communication error in MGC 8 FAILURE

22 MGC slave reports error 38 FAILURE

21 Undefined21 - -

20 Undefined20 - -

19 Undefined19 - -

18 Detector temperature critical 125 OUT_OF_SPEC

17 Errorwhileauto-standardization 73 FAILURE

16 Excessive radiation 123, 124 FAILURE

15 Input out of bounds 13 OUT_OF_SPEC

14 Errorwhilesignalprocessing 68 FAILURE

13 Undefined13 - -

12 Undefined12 - -

11 Undefined11 - -

10 Undefined10 - -

9 Undefined9 - -

8 Undefined8 - -

7 AITB simulated 29 FUNCTION_CHECK

6 Undefined6 - -

5 Undefined5 - -

4 Undefined4 - -

3 Undefined3 - -

2 Undefined2 - -

1 Errorwhiletrendrecording 126, 127 GOOD

0 Reserved - Reserved

Not displayed 33, 45, 71, 86 -

11.3 DimensionsThefollowingdimensionaldrawingsrepresentonlyanextractofallpossibleversions.Detaileddimensionaldrawingscanbedownloadedatwww.vega.com/downloads under "Drawings".

78

11 Supplement

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

Aluminium and stainless steel housing

242,

5 m

m (9

.55"

)

54,5

mm

(2.1

5")

9 m

m(0

.35"

)9

mm

(0.3

5")

175 mm(6.89")

90 mm(3.54")

100 mm(3.94")

143,5 mm(5.65")

169 mm(6.65")

116,5 mm(4.59")

119 mm(4.69")

M20x1,5/½ NPT

Fig. 25: Aluminium housing or stainless steel housing - Precision casting

79

11 Supplement

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

MINITRAC 31

57 m

m(2

.24"

)24

2,5

mm

(9.5

5")

Fig. 26: MINITRAC 31L Measuring range

80

11 Supplement

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

11.4 Industrial property rightsVEGA product lines are global protected by industrial property rights. Further information see www.vega.com.Only in U.S.A.: Further information see patent label at the sensor housing.VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.Nähere Informationen unter www.vega.com.Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellec-tuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad indus-trial. Para mayor información revise la pagina web www.vega.com.Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность. Дальнейшую информацию смотрите на сайте www.vega.com.VEGA系列产品在全球享有知识产权保护。进一步信息请参见网站<www.vega.com>。

11.5 TrademarkAllthebrandsaswellastradeandcompanynamesusedarepropertyoftheirlawfulproprietor/originator.

81

INDEX

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

INDEX

AAccessories

– Display and adjustment module 10 – External display and adjustment unit 11 – Interface adapter 10

Adjustment 29, 35Adjustment data 41, 49Adjustment mode 46Adjustment point 46Adjustment system 27Applications 28, 34, 45, 54, 55

BBackground radiation 29, 34, 45Block operation 40, 48, 54, 56

CCable entry 19Check signal 64Connection procedure 20Connection technology 20Control areas 12

DDamping 30, 39, 50Date 42, 50Date of manufacture 43, 52Default values 31, 43, 51Delta I 47Device name 43, 52Device settings Copy 43, 52Displayed value 40, 49

EEDD (Enhanced Device Description) 59Error messages 60

FFaultrectification63FF parameter 74Functional principle 9Function blocks

– Analog Input (AI) 73

HHandling permit 11Heat 18

IInner diameter 35

Installation position 14Instrument features 43, 52Isotope

– Co-60 28, 34, 45, 53, 55 – Cs-137 28, 34, 45, 53, 55

KKey function 27

LLanguage 40, 48Linearization29,36

MMain menu 28, 33, 44, 53, 54Moisture 13

NNAMUR NE 107

– Failure 61 – Function check 63 – Maintenance 63 –Outofspecification63

PPackaging 10Peak value 41, 49PIN 42, 50

RRadiation protection 11Radiationsafetyofficer12Radiation source 28, 34, 45, 53, 55Real value correction 39, 54Relay 30, 39, 48Repair 65Replacement parts

– Electronics module 11Reset 30, 42, 50

SSensor status 41, 49Service hotline 64Simulation 41, 49Source container 11Status messages 60Storage 10

TTime 42, 50Type plate 7

82

INDEX

MINITRAC 31 • Foundation Fieldbus

41782-EN-130430

UUnit 45Units 29, 35

VVoltagesupply72

WWater cooling 18

XX-ray alarm 30, 52

83

Notes

MINITRAC 31 • Foundation Fieldbus

4178

2-EN

-130

430

Printing date:

VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermany

4178

2-E

N-1

3043

0All statements concerning scope of delivery, application, practical use and operat-ing conditions of the sensors and processing systems correspond to the information available at the time of printing.Subject to change without prior notice

© VEGA Grieshaber KG, Schiltach/Germany 2013

Phone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected]