infrared microscopy and microanalyses for hire: analytical lab for small companies
DESCRIPTION
Infrared Microscopy and Micro Analyses for hire, Exact answers and interpretations for lowest prices, Expert chemical analyses and interpretations, I don't just tell you what it is, I tell you why it happened.TRANSCRIPT
John Donohue - [email protected]
Chemical Microanalysis for Industry
State-of-the-Art Analysis for hireMedical Device Problem Solving
Polymer Problem SolvingIndustrial Problem Solving
Asbestos Analysesetc., etc., etc. by
John Donohue201-294-2581
John Donohue - [email protected]
Hello. I’m John Donohue and this is the best Infrared Microscope ever made byany manufacturer: Nicolet’s IR-Plan Research Microscopemounted on a Magna 560 Mainbench
This amazing instrument allows my Labto obtain the Chemical Identity, via the Infrared Spectrum, of an area as small as 10 microns by 10 microns (10μX10μ). That’s as small as 9 red blood cells placed in a 3X3 square.
When you can chemically identify objects that small you can perform such amazing investigations that theycan often depart completely from the expected and traditional uses of Infrared Spectroscopy, as you will see.
These documents will describe the type of analyses I perform for Industry and Publication using this instrument and the other equipment in my Laboratory.
Part 1: Introduction to the Technology and Method
John Donohue - [email protected]
Mirrored “Lampshade”
Mirrored “Flying Saucer”
Hard Silicon Surface for ATR “Contact” Spectra
My IR-Plan is usually set up with Two Reflachromats:One For Reflection And Transmission, The Other Dedicated To ATR
More “Specialized Reflachromats” and Visible Light Objectives are available, if needed
… and Visible Light Objectives forVisible and Polarized Light Microscopy
John Donohue - [email protected]
Microscope Reflection ModeLightpath
The Upper Reflachromat objective projects a conical surface of IR radiation through the sample. It reflects off the shiny metal beneath the sample and follows the same conical surface up through the sample, back to the objective, and from there to the detector.
This mode is very fast and easy to perform. The shiny metal substrate is usually 0.005 inch aluminum sheet taped onto a microscope slide. It is cut from 5 inch by 5 inch sheet that is cheap and disposable. It can also be performed on any flat or curved metal surface such as injection molding tool surfaces, medical steel cannulas, engine valves, gun metal, you name it.
To Detector
From Source
Sampleon metal
Shiny Metal Substrate
Shiny Metal Substrate
John Donohue - [email protected]
>The FTIR Microscope increases greatly the utility of FTIR and allows the successful use of IR in analyses that simply could not be done by a Mainbench alone.
>The Micro - ATR Objective obtains surface spectra (of the top ~micron of material) with almost no sample prep. This is excellent for thin coatings or surface analysis.
>The FTIR Microscope can obtain useful spectra from extremely thin samples as small as a 10μ X 10μ Square. The amount of mass providing such a signal can approach the Detection Limits of GC/MS
>Example - FTIR Microscope’s Limit of Detection is about a 10μ X 10μ Square : If sample is 1μ thick, 10μ X 10μ X 1μ sample of Polyethylene = how many grams? 1cc of PE = 1g = 10mm X 10mm X 10mm = 1000 cubic mm = 103 mm => 10mm X 10mm X 10mm = 104 μ X 104 μ X 104 μ = 1012 cubic microns => So: 1012 cubic microns = 1g
10μ X 10μ X 1μ =100 cubic microns = 102 cubic microns 102 / 1012 = 10-10 g So: 10μ X 10μ X 1μ sample of Polyethylene = 0.0000000001 grams of PE
So, the FTIR Microscope can ID 100 trillionths of a gram of PE
UNIQUE ADVANTAGES OF THE FTIR MICROSCOPEHow sensitive is it? See arithmetic below.
John Donohue - [email protected]
Microscope ATR Mode and LightpathThe ATR Reflachromat objective projects a conical surface of IR radiation onto the inside of the ATR Crystal’s Sample Contact Point. A small part of the IR radiation “tunnels” into the sample touching this Contact Point. Some of it is absorbed and the rest continues on to the detector.
This mode is particularly well-suited to surface analysis (surfaces that are bioactive, drug eluting, bioresorbable, coated, “blooming” additives, lubricious, non-thrombogenic, etc.).
It is also a good choice for highly absorbing materials that are difficult to get an IR beam in and out of such as Black Rubber.
From Source
To Detector
Sample Sample
John Donohue - [email protected]
How Small Can Samples Be?How small IS the 10μ X 10μ Limit of Detection?
John Donohue - [email protected]
Lincoln seated on Penny’s back; mm scale to left
This image is 3.0mm X 2.3mm.
John Donohue - [email protected]
IR-Plan Visible Light 10X Objective view (Glass)Lincoln’s Head and Shoulders
This image is 550μ X 420μ
John Donohue - [email protected]
IR-Plan IR Objective 15X Reflachromat view: Lincoln’s Head
This is the magnification at whichKnife-edge apertures are used to frame
the area to be analyzedand FTIR Spectra are obtained.
10μ X 10μ
This image is 390μ x 300μ
Limit ofDetection:
“Lincoln’s “Eye”
Lincoln’s Face on the previous slide is about 140μ X 120μ.The IR-Plan can obtain good spectra from much smaller samples than this (see asbestos ribbon, below). The thin polymer coating on Lincoln’s Face is easily IDdas Polycarbonate.
John Donohue - [email protected]
John Donohue - [email protected]
Asbestos Identificationby IR Microscopy and/or
Polarized Light Microscopy
Part 2: Asbestos Testing
Friable Asbestos Identification by IR Microscopy of broken cementitious tile:
John Donohue - [email protected]
John Donohue - [email protected]
Zooming in on the Friable Asbestos
John Donohue - [email protected]
John Donohue - [email protected]
Microscopic amount of Asbestos squashed onto aluminum sheet
John Donohue - [email protected]
NOTE: Often Spectra are corrected for humidity in the Laboratory air
John Donohue - [email protected]
The Asbestos sample is Chrysotile
John Donohue - [email protected]
The IR Microscope is so very sensitive that even a 20μ X 65μ areaof a Single “ribbon” of Asbestos is enough for Identification
John Donohue - [email protected]
John Donohue - [email protected]
Asbestos Identification and Quantificationis routinely performed as per
EPA-600-R-93-116 usingPolarized Light Microscopy (PLM)
Examples of PLM in my LabsMaterials Analyses follow
John Donohue - [email protected]
>These barrels were for “EpinephrinePens” needed to counter the threat ofNerve Gas and thus keep Saudi Arabiafrom backing down to Iraq in the GulfWar. This allowed the USA to Stagethe invasion of Kuwait.
>The barrels were breaking 100% uponejection from mold with undercut. >Defect invisible until Crossed Polarizersrevealed “Knit Line” pointing toCrack Initiation Point.
>The molten plastic was too cool whenIt came squeezing around the Core Pinfor the two advancing Melt Fronts tomelt together adequately.
>I told Manufacturer to increaseZone Temperatures by 30 degrees Cand Ejection Breakage ceased. ... and that’s how I won the war.
Device ProductionRestarted and DefectDetermined and EliminatedVia Stress BirefringenceAnalysis
CrackInitiationPoint
John Donohue - [email protected]
Polypropylene Syringe Wall Cross SectionControl of Cooling Rate is a major parameterin Determining Morphology and Properties
This type of work was very important in the implementation of Clear Polypropylene Devices using Milliken Clarifying Agents that have theundesirable effect of increasing the brittlenessof Polypropylene.
Polypropylene Syringe Gate Cross SectionControl of Rheology is a major parameterin Determining Morphology and Properties
This type of work was also very important inthe implementation of Radiation Sterilizationwhich also has the undesirable effect ofincreasing the brittleness of Polypropylene.
Injection Molding Morphology and Physical Properties
John Donohue - [email protected]
Solving Medical Device problems using
IR Microscopy
Visible Microscopy
Part 3: Application to Medical Device issues
John Donohue - [email protected]
-0.30
-0.25
-0.20
-0.15
-0.10
-0.05
0.00
0.05
0.10
0.15
Abso
rbanc
e
1000 1500 2000 2500 3000 3500 Wavenumbers (cm-1)
Silicone Lubricated Gasket
Less Silicone After Surface Scraped
No Silicone After Acetone Wash:Silicate-filled rubber polymer
ATR Spectra of Surface of IR-Opaque Rubber
2.3mm x 3.0mm Photos of an Insulin Needle Point
The Technology of a Needle Point
What can we learn with a fast analysis?John Donohue - [email protected]
420μ x 550μ Photos of Needle’s three cut planes
The “subtle” cutJohn Donohue - [email protected]
John Donohue - [email protected]
IR-PLAN 550μ X 420μ FOV of Point and Enhanced Image
Raw Image Obtained
Digital Enhanced Image
A “Metal Burr” is visibleon this needle point.
John Donohue - [email protected]
- 0. 000
0. 001
0. 002
0. 003
0. 004
0. 005
0. 006
0. 007
0. 008
0. 009
0. 010
0. 011
0. 012
Ab
so
rb
an
ce
1000 1500 2000 2500 3000 3500 W avenum ber s ( cm - 1)
Infrared Microscopy Provides Chemical Identification that showsGelled Lubricant is used on Hypodermic Needles
3.0mm x 2.3mm
550μ X 420μ
390μ x 300μ
Gel is Silicone
8.5mm X 11.2mm
Hoop Stress cracking
Polyetherimide (PEI) Stopcock Outerbody resists splitting caused by fatty Feeding Liquids but costs more than Polycarbonate (PC) (which is cracked by combination of fats and stress). Stopcock InnerBody is pressed into Outerbody and this strong PEI polymer can still crack if it can’t stretch enough under this “Hoop Stress”tensile load (vs PC which is very stiff but also very TOUGH/RUBBERY).
Engineering Resins: ESCR vs “Hoop Stress Failures”
John Donohue - [email protected]
John Donohue - [email protected]
PET Barrels That Failed At Gate
A “Cold Slug” in an Injection Molding “Gate”can initiate breakage at unacceptably low force
John Donohue - [email protected]
Gate A Gate B Ridge/Valley
Cold Slugs
Blade Striations
Devices broke due to “Cold Slug” in Injection Molding Gate
John Donohue - [email protected]
Identification of PET Barrel SkirtContaminant by Micro-FTIR
Pre-Extracted Cellulose on PET Skirt Same Area of PET Skirt After Wiping
Streaks where Greasewas removed
Cellulose
Grease aboutto be removed
PET Barrel Skirt Bloom removed for analysis
John Donohue - [email protected]
5 .S ilic o n e B B L L u b e A T R ,B ra d m o n s
K im w ip e d B lo o m in P E T B B L s k irt,G riffith ,R E F L 3 2
P E T B B L s k irt g re a s e ,K im Wip e d & M C d ,C a s tro ,R E F L 3 2 ,2 /2 7 /0 9
1 2 0 a 2 2 0 9 ,M IC R ,C E L L U L O S E ,A C C U WIP E /A L ,2 /2 4 4
-0 .4 0
-0 .3 5
-0 .3 0
-0 .2 5
-0 .2 0
-0 .1 5
-0 .1 0
-0 .0 5
0 .0 0
0 .0 5
0 .1 0
0 .1 5
0 .2 0
0 .2 5
0 .3 0
0 .3 5
0 .4 0
0 .4 5
Lo
g(1
/R)
10 00 15 00 20 00 25 00 30 00 35 00 40 00
Wa v e nu mb ers (c m-1)
Silicone
Cellulose
Grease Found Previously
Grease Found Recently
PET Barrel Skirt Contaminant is Same as Found previouslyThe Difference Seen below is Cellulose from the Kimwipe.
The Grease is an Acid (1712 cm-1) and an Ester (1737 cm-1).
John Donohue - [email protected]
John Donohue - [email protected]
An unusually profitable Bioresorbable Implant with a brilliant Market-dominating future had its Market Launch endangered by the occurrence of foreign matter inside the tiny molded PolyLactide:Glycolide implant. A CERTAIN fix was required immediately: the cause had to be IDd with certainty and eliminated.
Many Hypotheses were proposed:
>The Resin Supplier suggested that Residual Monomer was boiling during molding. Extensive testing of residual monomer levels in retained lot samples vs amount of bubbles found from molding those lots was proposed.
>Molding Engineers suggested that the bubbles were: Shrink Voids, Entrapped Air, Entrapped Condensation, Etc. Humidity archives were to be examined and production areas were to be desiccated.
>”Shotgun Experiments” were about to be fired in every direction.
Instead a “RIFLE BULLET” EXPERIMENT was aimed right at the bubbles themselves.
>Contaminant was Bubbles Inflated while Flowing “COMET”.>Cut thin Cross-sections of the polymer through the Bubbles.>Micro-FTIR repeatedly found a single deposit of Secondary Amide localized in a small area on each Bubble’s inner surface.
Part 4: Nailing the Answer and Preventing a Recall
FlowDirection
John Donohue - [email protected]
Hazy TranslucentArea on border ofBubble
When a thin slice of the Part, containing the Bubble, was shaved off,the Hole in the shaving had a tiny region on its circumference
that was Hazy Translucent while the rest of the circumference andof the part’s shaving was Clear Translucent
Cross-section of Bubble
Clear translucent polymer slice
John Donohue - [email protected]
1534
1648
3060
3288
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
2.2
2.4
2.6
2.8
3.0
3.2
3.4
Ab
so
rba
nc
e
1000 1500 2000 2500 3000 3500 4000 Wavenumbers (cm-1)
“Zooming in” on Hazy deposit on Bubble’s circumferenceIncreases the Secondary Amide Absorbance Peaks relative to the PLG’S...
PLG
PLG
...therefore the hazy area is a Secondary Amide
FTIR Microscopy
Zooming in
John Donohue - [email protected]
-1.0
-0.5
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
Ab
so
rba
nc
e
1000 1500 2000 2500 3000 3500 4000 Wavenumbers (cm-1)
Three Amide deposits on Aluminum with PLG dissolved away. Two Amide deposits squashed onto Aluminum.
Methylene Chloride removedmost of the PLG
Squashed onto Aluminum
Squashed onto Aluminum
AmideAmide
... so what is this Secondary Amide and what is it doing?
John Donohue - [email protected]
Synergistic Use of
FTIR Microscopeand
Gas Chromatograph / Mass Spectrometer
The two most powerful instrumentsIn Analytical Chemistry
... so what is this Secondary Amide and what did it do?
John Donohue - [email protected]
The SIS Thermal DesorberA Heated Inlet for GC/MS
sold by Scientific Instrument Services
John Donohue - [email protected]
Carrier Gas Flowpath to Mass Spec Detector
Pure He flowing throughglass lined steel tubethat contains sample.
Tube is injected intoGC Inlet and heated.
Volatiles separate oncolumn and are analyzedby the MS.
Excised tiny intact Bubbles melted in THDSBpop and release CO2 (and NOT O2 or N2)
PLG Hydrolysis in heated tube releases Lactide and Glycolide
John Donohue - [email protected]
Hydrolysis of PLG: The Chemistry Occurring
The moisture in the amide flakesforms steam during injectionmolding, inflating the bubbles.
The steam is consumed as ithydrolyzes the PLG, filling thebubbles with CO2.
as per Dr. C. C. Chu of CornellJohn Donohue - [email protected]
John Donohue - [email protected]
Spectral changes of PLG hydrolyzed on Hotplate
Carbonyl and Hydroxyl orbitalsare formed as the controlshydrolyze on the hotplate.
John Donohue - [email protected]
Spectral changes of PLG vs proximity to amide deposit
The concentration of the same Carbonyland Hydroxyl orbitals is seen to increasewith increasing proximity to the bubblesformed by the amide flakes. This largegradient in concentration occurs acrossa microscopic range of distance from thebubbles’ walls.
It is in analyses like this one that theFTIR Microscope is the very best toolthere is. No other instrument coulddo this work: easy and exact spatiallocalization of a chemical identification.
John Donohue - [email protected]
Preventing Recall - Conclusion>Skin Flakes boiled off their water, inflating the “comets”
>As this steam inflated the PLG, it was immediately consumed by the PLG’s Hydrolysis Reaction
>Hydrolyzing PLG emits CO2, the gas remaining inside the “comets” that was detected by GC/MS as the “comets” melted and popped
>Hydrolyzing PLG on a hot plate using drops of water Darkened spots ranging from light brown to nearly black
>FTIR spectra of these darkened regions peaks appearing near 1730 cm-1 and 1620 cm-1
>FTIR spectra obtained from microscopic regions as the area examined approached the hydrolyzed circumference of the “comet” slices peaks appearing near 1730 cm-1 and 1620 cm-1
>The polymer pellets were protected against any further contamination by skin flakes, there was no further problem with “comets”, the product launch was continued, and the implant dominated its market One Billion $ in $ales per year
John Donohue - [email protected]
Ending $100MM/yr Device Factory Shutdowncaused by “wrinkled” Injection Molded Parts
A $100MM per year device factory was shutdown because Injection Molded parts were slightly wrinkled. No one was going to make any money until the problem was fixed.
Possible causes that were proposed for investigation: >Electrical supply fluctuations >Cooling Water Heat Transfer fluctuations >Press Horsepower fluctuations
Many “Shotgun Experiments” were about to begin with no expense to be spared
--- BUT ---
>The Polypropylene (PP) Resin was supposed to contain Sodium Benzoate (NaBZT), an additive that makes the Polymer solidify at a higher temperature (faster) than normal.>This fact suggested that the best place to look for the cause was in the PP pellets themselves.>I developed an Infrared analytical method that measured very quickly the NaBZT concentrations of large numbers of individual pellet.
John Donohue - [email protected]
Plant Shutdown Ended“Wrinkled” Injection Molded Parts caused by Resin Supplier’s Error
FREQ
NaBZT Concentration
This is one example of when to use the FTIR Mainbench instead of the Microscope:When you want to analyze as large of an area as possible per each scan.
>The Resin Supplier by mistake had added different pellets to the Finished Material: pellets that contained a significantly higher concentration of NaBZT than the correct pellets. These wrong pellets were dumped into the Railcar on top of the correct Resin.
>The resulting mixed resins received only the little bit of blending that occurred during transport from the Railcar to the Silo and then to the Hoppers.
>The Parts wrinkled because the varying NaBZT concentrations caused varying degrees of Shrinking and Packing during molding.
...but some had too much
Analytical Method developedto measure RAPIDLY theConcentration of individual pelletsproving Bimodal Distribution
Most Pellets hadRight NaBZT amount
John Donohue - [email protected]
FTIR Microscopy can do a lot morethan the simple chemical
identification of contaminants.
> It can determine the physical morphology of polymers.
> It can measure the oxidative degradation caused by radiation sterilization (dosimetry).
> It can measure the hardness of PVCs, Polyurethanes, and numerous other Thermoplastic Elastomers
> It can measure the Butadiene or Styrene content of Styrenics
Part 5: Special Applications
John Donohue - [email protected]
Measuring Changes in Crystallinity:PET Tube Turned Opaque White
Customer complained that PET catheter was opaque white instead of clearBUT
Had the customer accidently left this catheter in a hot Autoclave overnight?
Polyethylene Terephthalate Crystal Structure
John Donohue - [email protected]
John Donohue - [email protected]
The Mettler Hotstage
Remove the glass window and insert a strip of aluminum sheet or a thin, “homemade” KBr “poker chip”.Then you can obtain spectra while heating the sample.
John Donohue - [email protected]
Spectral Differences Between Clear And Hazy PET TubesCrystallinity Peaks marked in Hazy Return
FullSpectrum
zoom in
The hazy (opaque white)catheter had spectraldifferences whencompared to the clearcatheter (crystallinity peaksare shown at right)
John Donohue - [email protected]
When the clear tube was quenched from the melt by Boiling it in Water, it crystallized, turning white like the Customer’s Hazy returned tube.
Proof that Spectral Differences Between Clear and Hazy Tubesare due to Crystallinity
Crystallized in Boiling Water
Crystallized in Boiling Water
The Spectra of the clear tube whenit is molten at 260C is the same aswhen it is quenched from the meltby Liquid Nitrogen.BUT when it is quenched from themelt by Boiling it in Water, itanneals and grows crystals.
Crystallized by Customer’sUnknown Method
John Donohue - [email protected]
Proof that White catheter has hydrolyzed to Lower Molecular WeightWhen the Clear tube, a PET controlsample, and the Hazy PET weremelted and then instantly quenchedfrom the melt in Liquid Nitrogen, theHazy sample has already Crystallized.and the Clear tube sample has not.
The Hazy PET crystallizes much morereadily because it has Hydrolyzed toa Lower Molecular Weight.
When the clear tube is melted andkept (quenched) in ambient air for4 seconds or 10 seconds before beingquenched in Liquid Nitrogen, NO crystallization begins after the 4seconds in air and only a tiny amountafter the 10 seconds in air.
The Hazy PET crystallizes much more readily because it has Hydrolyzed to a Lower Molecular Weight.The Customer probably left the catheter in a hot Autoclave overnight.
John Donohue - [email protected]
Effects of Radiationand Accelerated Aging
Part 6: Radiation Sterilization Issues
ATR Spectrum
Reflection from smearon aluminum sheet
The bloom visible on this catheterwas analyzed both by contactingits surface with the ATR Objectiveand by smearing the bloom ontoaluminum sheet and analyzingthe smear by Reflectance.The smear shows only the DEHPwhile the ATR sees both theDEHP and the PVC catheter.
Radiation Increases PVC BloomingOf Cytotoxic DEHP Plasticizer
John Donohue - [email protected]
John Donohue - [email protected]
Radiation Generates Free Radicals in Polypropylene
John Donohue - [email protected]
Free Radicals Generate Carbonyl Groups in Polymers
John Donohue - [email protected]
Test Method DevelopmentDecreased Ductility Can Cause Devices To Break During Use.
Radiation may increase the force to strain the device, but the strainat break decreases. Devices seldom fail because they’re too stiff.
They fail because they break.
A Syringe barrel wall may be unbreakable during use despite receiving the highest radiation dose allowed. Syringe tips may be much more fragile. Customer complaint history and product history may indicate what areas of a product are areas of concern; what are the products’ weakest links. It is these modes of failure that must be tested.
Testing must mimic the mode of failure expected during customer use. Breakage tests must be performed at a speed of deformation similar to that experienced by the product during customer use.
Testing samples subjected to Accelerated Aging will provide data quickly that predicts the future behavior of products. The Test Protocols will define the Accelerated Aging to be used.
Measuring Radiation Damage using Micro-FTIRMounting Aluminum Sheet On 3X2 Glass Slide
John Donohue - [email protected]
John Donohue - [email protected]
Shavings off Polypropylene Barrel’s Surface Analyzed Herein:Shavings Obtained with Razor Blade
Close-ups of Polypropylene Shavings Analyzed Herein
John Donohue - [email protected]
John Donohue - [email protected]
0. 2
0. 4
0. 6
0. 8
1. 0
1. 2
1. 4
1. 6
1. 8
2. 0
Ab
so
rb
an
ce
1000 1500 2000 2500 3000 3500 4000 W avenum ber s ( cm - 1)
Same Polypropylene Film Before and After 40 kGYs of Cobalt RadiationBlue is after
John Donohue - [email protected]
0. 10
0. 15
0. 20
0. 25
0. 30
0. 35
0. 40
0. 45
0. 50
0. 55
0. 60
0. 65
0. 70
Ab
so
rb
an
ce
2000 2500 3000 3500 4000 W avenum ber s ( cm - 1)
Same Polypropylene Film Before and After 40 kGYs of Cobalt RadiationBlue is after
Hydroperoxides Carbonyls
John Donohue - [email protected]
0. 08
0. 10
0. 12
0. 14
0. 16
0. 18
0. 20
0. 22
0. 24
0. 26
Ab
so
rb
an
ce
1600 1700 1800 1900 2000 W avenum ber s ( cm - 1)
Same Polypropylene Film Before and After 40 kGYs of Cobalt RadiationBlue is after
Scissioned Polymer Chains Oxidize
John Donohue - [email protected]
IR Spectrum Shows Radiation Induced Oxidation of Polypropylene
as per J. Donohue MDDI
John Donohue - [email protected]
Polypropylene Oxidation from 20 and 40 kGYs of Cobalt Radiation
John Donohue - [email protected]
The “Dark Reaction” of Irradiated PolypropyleneOxidative Degradation Continues Long After Irradiation Has Ceased
This is the Reaction that is Accelerated by Accelerated Aging
John Donohue - [email protected]
Dr. Apostolou and I “wrote the book” on Accelerated Aging Methods that work
John Donohue - [email protected]
Oxygen Can More Easily Penetrate and Reactwith the Polymer in a Thin Film
John Donohue - [email protected]
This Post Rad Oxidation is Not Just Peroxide Scissions. Ambient Oxygen Continues to React with the Polymer.
This is Proven by this Vacuum Oven Aging.
Polyethylene Undergoes Similar Oxidationwhen Irradiated with 20 and 40 kGYs
John Donohue - [email protected]
John Donohue - [email protected]
A Carbonyl Index Can be Defined to Measure this Oxidation
John Donohue - [email protected]
Statistical Results for Micro-FTIR Dosimetry of Gamma vs ControlThe Micro-FTIR Method is Accurate, Precise, and Robust
John Donohue - [email protected]
0.10
0.20
0.30
0.10 0.20 0.300.00
0.00 0.40
X = Thickness of slice (area of 1304 cm-1 Absorbance)
Area of C=OAbsorbancedivided by X
(Carbonyl Index)
Radiation Damage (= Dose) Measured for Thin Surface Shavings of Sterilized Polypropylene Medical Devices
Shavings of samples with 0 Mrads
Shavings of samples with 3.5 Mrads
Shavings of sterile product
Shavings of underdosed product
FTIR Microscopy can determine a competitor’s dose or detect underdosed non-sterile devices.
John Donohue - [email protected]
Stability of Fina Syndiotactic and IsotacticPolypropylenes to Cobalt Radiation
and Accelerated Aging
John Donohue - [email protected]
Ziegler-Natta and Metallocene Catalysts
Controlled Orientation of Monomer Approach To Active Site Yields Controlled Stereoregularity of Polymer Chain Formed.For Z-Ns, Solid Catalysts Control Approach to Active Site.For Metallocenes, Molecular Structure Controls Approach.
John Donohue - [email protected]
Stereostructure of Isotactic Polypropylene
Hydroperoxide Formation by “Backbiting” Oxidation
Strings of Close, Unstable, Pendant Hydroperoxides
Free Radical Degradation of Isotactic Polypropylene
John Donohue - [email protected]
0. 2
0. 4
0. 6
0. 8
1. 0
1. 2
1. 4
1. 6
1. 8A
bs
or
ba
nc
e
1000 1500 2000 2500 3000 3500 4000 W avenum ber s ( cm - 1)
Isotactic Polypropylene Before and After 38 kGy & 17 Days @ 70 C
Hydroperoxides Carbonyls
The Isotactic Polymer is extensively Oxidized by Irradiation. It sizzles like bacon when it is melted.
John Donohue - [email protected]
0. 2
0. 4
0. 6
0. 8
1. 0
1. 2
1. 4
1. 6
1. 8
2. 0
Ab
so
rb
an
ce
1000 1500 2000 2500 3000 3500 4000 W avenum ber s ( cm - 1)
Syndiotactic Polypropylene Before and After 38 kGy & 17 Days @ 70 C
The Syndiotactic Polymer exhibits very little Rad-induced Oxidation
John Donohue - [email protected]
0. 08
0. 10
0. 12
0. 14
0. 16
0. 18
0. 20
0. 22
0. 24
Ab
so
rb
an
ce
1600 1700 1800 1900 2000 W avenum ber s ( cm - 1)
IPP & SPP: 38 kGy & 70 C Aging StudyAccelerated Aging Increases IPP Oxidation
but has Very Little Effect on SPPIPP @ 70 C: 17 Days 88 Hrs 16 Hrs 0 Hrs
IPP 0 Dose
SPP 0 Dose
SPP @ 70 C:0 to 17 Days
John Donohue - [email protected]
0. 08
0. 10
0. 12
0. 14
0. 16
0. 18
0. 20
0. 22
0. 24
Ab
so
rb
an
ce
1600 1700 1800 1900 2000 W avenum ber s ( cm - 1)
IPP & SPP: 38 kGy & 40 C Aging Study
IPP @ 40 C: 17 Days 88 Hrs 16 Hrs 0 Hrs
IPP 0 Dose
SPP 0 Dose
SPP @ 40 C:0 to 17 Days
Accelerated Aging Increases IPP Oxidationbut has Very Little Effect on SPP
John Donohue - [email protected]
Carrier Gas Flowpath to Mass Spec Detector
Pure He flowing throughglass lined steel tubethat contains sample.
Tube is injected intoGC Inlet and heated.
Volatiles separate oncolumn and are analyzedby the MS.
John Donohue - [email protected]
Radiation SterilizedSyndiotactic PolypropyleneGenerates an Order ofMagnitude Less Volatilesthan an Equal Mass ofIrradiated Isotactic Polypropylene
THDSB/GC/MS Analysesof Post-Rad Volatiles
John Donohue - [email protected] Close, Unstable, Pendant Hydroperoxides Explode Like a String of Firecrackers
CO2acetaldehyde
acetoneacetic acid
4-hydroxy4-methylpentanone
2,4-dimethylfuran
cyclopropylacetone
allylacetone
acetic anhydride
3,5,5-trimethylfuranone
acetoacetone
THDSB/GC/MS IDENTIFICATION OF IPP POST-RAD VOLATILES
Heated Irradiated Isotactic Polypropylene Degrades into Volatiles Based on C-C-O Units that Reveal the Chemical Mechanisms of its Oxidation
John Donohue - [email protected]
Thermally Degraded Post Rad Isotactic PolypropyleneEmits Low Mass Scission Products Based on C-C-O Units …
Air & WaterDesorbed OutOf Tube
Acetone
Acetic Acid
… Because Similar Oxidized Structures Degrade into Similar Volatiles
THDSB/GC/IR Analyses of Post-Rad Volatiles
Thermally Degraded Post Rad Syndiotactic PolypropyleneEmits a Far More Random Mix of Scission Products …
Air & WaterDesorbed OutOf Tube
2-Hydroxy-Propionic Acid
3-Methyl-2,4-Pentanediol
4-Butyl-Gamma-Octanolactone
… Because a More Random Dispersion of Oxidized Structures Yieldsa More Random Mix of Volatile Degradation Products
THDSB/GC/FTIR Analyses of Post-Rad Volatiles
John Donohue - [email protected]
John Donohue - [email protected]
Phenolic Antioxidants Protect Against Radiation Damage by Scavenging the Free Radicals Formed in the Polymer by Radiation
Part 7: Additives; their Analysis and Issues
John Donohue - [email protected]
But Phenolic Antioxidants Turn Plastic Yellow When Irradiated
John Donohue - [email protected]
Hindered Amine Light Stabilizers Form Cytotoxic Hydroxylaminesas they Protect the Polymer from Radiation Damage
Without Discoloration
John Donohue - [email protected]
Millad 3988 Clarifies PolypropyleneMolding Heat Causes Hydrolysis, Releasing Benzaldehyde Derivatives
This Causes the Polymer to Emit a “Cherry Candy” SmellMillad makes Polypropylene more brittle
Clarified Polypropylene Crystallized at 130 C from the MeltNucleation Determines Morphology
ClarifiedNot Clarified
Sublimation depletes Boundary of MilladPolypropylene Spherulites grow
Millad preventsSpherulite growth
John Donohue - [email protected]
John Donohue - [email protected]
Millad Forms Thermally Unstable Precipitate if it is Overheated DuringInjection Molding. Precipitate’s Sizzling Decomposition into Gaseous,
Superheated Aldehyde Strips Char out of Molding Press and Into Molded Parts. Heated excised precipitate chunks undergoing thermal decomposition on Hotplate
John Donohue - [email protected]
Sizzling Decomposition into Aldehyde and Arylate
>Tiny orange spots were scattered across the Polypropylene matrix
>Any attempt to get the spectrum of more than one orange spot at a time yielded only a spectrum of the Polypropylene matrix
>But the IR-Plan can can zoom in on a single tiny orange spot to yield the Arylate spectrum shown …
… and Thermal Desorption of the degrading material into the GC/MS can show the formation of the Aldehydes and Alcohol Intermediates.
John Donohue - [email protected]
Thermal Decomposition of Precipitate Forms Aldehyde.Cannizzaro Reaction of Aldehydes Forms Acid and Alcohol.
Condensation of Acid and Alcohol Forms Arylate.
Arylate Forms The Orange Spots
John Donohue - [email protected]
Thin Layer Chromatography (TLC) can Separate Chemical Mixtures that theGC/MS can’t: Chemicals that are non-Volatile or Thermally Unstable
Liquid carries chemicals in spot of extract up the plate, separating them
John Donohue - [email protected]
>TLC is done on a Plate Covered with Fluorescent Silica>UV Light makes the Separated Chemicals from the Mixture Visible>The Regions of Silica Containing these Chemicals are scraped off the plate, separated from the Silica, and Identified using the Analytical Instrumentation
Under Visible Light
Under UV Light
After Scraping
0. 0
0. 5
1. 0
1. 5
2. 0
2. 5
3. 0
3. 5
4. 0
4. 5
Ab
so
rb
an
ce
1000 1500 2000 2500 3000 3500 4000 W avenum ber s ( cm - 1)
Measuring Durometer of FINISHED (competitive) Devices:FTIR of DEHP Plasticized PVC Device vs Pure PVC
Pure PVCDEHP
Part 8: Tricks of the Trade
John Donohue - [email protected]
John Donohue - [email protected]
- 0. 4
- 0. 2
0. 0
0. 2
0. 4
0. 6
0. 8
1. 0
1. 2
1. 4
1. 6
1. 8
2. 0
2. 2
Ab
so
rb
an
ce
1450 1500 1550 1600 W avenum ber s ( cm - 1)
FINISHED Device: Measuring Plasticized PVC Durometer
Hard Endotracheal Tube
Soft Nasogastric Tube
DEHP
PVC
Spectra “normalized” for equal plasticizer (DEHP) content showthat the harder PVC has a higher PVC to Plasticizer RatioSuch a test can tell the Durometer used by a Competitor
HigherPVC
conc.
John Donohue - [email protected]
0. 0
0. 1
0. 2
0. 3
0. 4
0. 5
0. 6
0. 7
0. 8
0. 9
1. 0
1. 1
Ab
so
rb
an
ce
1000 1500 2000 2500 3000 3500 4000 W avenum ber s ( cm - 1)
Reverse engineering competitive cathetersMicro-FTIR Spectrum of Pellethane
zoom in here next slide
John Donohue - [email protected]
0. 15
0. 20
0. 25
0. 30
0. 35
0. 40
0. 45
0. 50
0. 55
0. 60
Ab
so
rb
an
ce
1360 1380 1400 1420 W avenum ber s ( cm - 1)
Micro-FTIR ID: FINISHED Device Pellethane Durometer
80A
90A
75D
>Many catheters have tips made from a softer grade material than the shaft.
>FTIR can measure the Durometer of finished Polyurethane devices quickly and easily
>Multiple runs below show that the method is robust
John Donohue - [email protected]
0. 2
0. 4
0. 6
0. 8
1. 0
1. 2
1. 4
1. 6
Ab
so
rb
an
ce
1000 1500 2000 2500 3000 3500 4000 W avenum ber s ( cm - 1)
Micro-FTIR Spectrum of Tecothane
zoom in here next slide
John Donohue - [email protected]
0. 02
0. 04
0. 06
0. 08
0. 10
0. 12
0. 14
0. 16
0. 18
Ab
so
rb
an
ce
1360 1380 1400 1420 W avenum ber s ( cm - 1)
Micro-FTIR ID: FINISHED Device Tecothane Durometer
74A
85A
55D
75D
John Donohue - [email protected]
0. 05
0. 10
0. 15
0. 20
0. 25
0. 30
0. 35
0. 40
0. 45
0. 50
0. 55
0. 60
0. 65
0. 70
0. 75
0. 80
Ab
so
rb
an
ce
1000 1500 2000 2500 3000 3500 4000 W avenum ber s ( cm - 1)
Micro-FTIR Spectrum of Tecoflex with 20% BaSO4
zoom in here next slide
John Donohue - [email protected]
0. 1
0. 2
0. 3
0. 4
0. 5
0. 6
0. 7
0. 8
0. 9
1. 0
1. 1
1. 2
Ab
so
rb
an
ce
1300 1320 1340 1360 1380 1400 W avenum ber s ( cm - 1)
Micro-FTIR ID: FINISHED Device Tecoflex Durometer
80 A
85 A
100 A
65 D
60 D
NOTE:This method is more accurate than themanufacturer’s ability to control or measuretheir Durometer. The Hardness Pucks thatwere supplied by the manufacturer (aftermeasuring Durometer with a Hardness Tester)and were supposed to be 60 D were insteadHARDER than the supposed 65 D pucks.
John Donohue - [email protected]
HIGH-GLOSSABS CRACKS
LOW-GLOSS ABSRESISTS CRACKING
FTIR Shows Why High-Gloss ABS Has Less ESCR Than Low-Gloss ABS
More Styrene/Butadiene Rubber dispersed in the Acrylonitrile matrixyields High Gloss ABS with less resistance to Environmental Stress Cracking
John Donohue - [email protected]
Packaging Materials
>There are a lot of “Tricks of the Trade” in Materials Analysis.
>For example, Most packaging films are laminates with outer heat-sealable plies and an inner strength ply.
>The FTIR Microscope requires Liquid Nitrogen (LN2) to operate and this LN2 can be used to cryo-fracture materials.
>Cryo-fractured laminated films can be separated easily into their individual plies for material identification and also for accurate thickness measurements free from thickness artifacts caused by cutting techniques that can decrease the measured thicknesses.
John Donohue - [email protected]
The film contains K-Resin…
…and the stretched Heat Seal Ply is EVA
Device package bottom web wascryo-fractured with liquid N2 andthe heat seal ply was stretched over aluminum sheet.
Device package bottom web wascryo-fractured with liquid N2 andthe protruding and overhangingplys show a Surlyn center plysandwiched in EVA heat seal plys.
Surlyn center plyprotruding fromfractured laminatedfilm
Surlyn ply and EVA heatseal ply together
John Donohue - [email protected]
John Donohue - [email protected]
Device package bottom web wascryofractured with liquid N2 andthis two ply film separated intoa K Resin ply and a more flexibleEVA heat seal ply.
K Resin ply is stiffer
EVA heat seal ply
John Donohue - [email protected]
Chemical Degradation of Polycarbonateby Dymax Adhesive
Analysis of White Stain inside Cured Part
Part 9: UV Adhesives and the Chemical Fragility of Polycarbonate
John Donohue - [email protected]
ABSTRACT>Polycarbonate (PC) has poor resistance to attack by a large number of chemicals.>Chemicals that “plasticize” PC cause problems that include clouding, warping, crazing, cracking, breaking, falling apart, etc. For example, Formula 409 contains about 10% “grease cutter” and will quickly destroy PC; especially when the PC has a relatively low Molecular Weight and/or lots of molded-in-stress pulling it apart.>Chemicals that are “Bases” (electron donators) catalyze the polymerization of PC and therefore can also catalyze the reverse reaction, depolymerizing PC into oligomers of Bisphenol A (BPA) monomer or the BPA monomer itself. For example, Amines famously do this to PC.>A major ingredient of the Dymax adhesive is N,N-Dimethylacrylamide (a base). It is well-known that many adhesives will attack PC if they are in contact with it too long prior to being cured. This chemical is one of the reasons for this. When the adhesive is completely cured, this chemical is completely consumed (this fact is demonstrated using the spectra herein).>The white stain on the PC in contact with the “cone” of adhesive is partially depolymerized PC. Its FTIR spectra show the presence of both PC and the endgroups (that appear as BPA and its oligomers) are formed from the scissioning of PC. The adhesive must be quickly and fully cured to prevent this.>The side of the adhesive cone that was peeled free from the PC is coated with a thin shattered film of PC. The side that was in contact with the Polyurethane (PUR) is coated with the PUR. Thus, the adhesive is tougher and more tenacious than either polymer.
John Donohue - [email protected]
What Happened to the Polycarbonate?
John Donohue - [email protected]
White Stain Seen Through the PC Wall Prior to DissectionWhite Stain is Connected to the PC
John Donohue - [email protected]
White Stain Remains on PC After Adhesive is Peeled Back
PolycarbonatePeeled-backDymax Adhesive
Polycarbonate
John Donohue - [email protected]
White Stain Remains on PC Higher Magnification
Polycarbonate Polycarbonate
John Donohue - [email protected]
White Stain Remains on PC
The White Stain is Partially Depolymerized Polycarbonate so FTIR shows thatit contains Absorptions for PC, Oligomers, and BPA Monomer
Polycarbonate
White Stain
Bisphenol A
John Donohue - [email protected]
John Donohue - [email protected]
The Surface of the Adhesive that was against the PC is encasedin a thin shattered PC Film that again contains Oligomers and BPA
Thin Shattered Film on Adhesive that was against the PC contains Polycarbonate, BPA, and Oligomers
Polycarbonate
BPA
John Donohue - [email protected]
The other side of the adhesive was against the Polyurethane and it is encased in that Polyurethane
Polyurethane Tube
John Donohue - [email protected]
John Donohue - [email protected]
What Chemical(s) Attackedthe Polycarbonate?
O
O
NO
OHO
OHO
DYMAX Disclosed 5 Components of this Adhesive
N,N-DimethylacrylamideIsobornyl Acrylate
Irgacure 184 Darocur 1173
Reactive Diluents
Photoinitiators
Ketone
Amide
These 4 are All Ketones and one is also an Amide. Ketones can Craze and CrackPolycarbonate and Amides can Depolymerize it to Bisphenol A. John Donohue -
Pre-UV Adhesive
Post-UV Adhesive
Adhesive Taken Off Device
Adhesive’s Spectrum Pre and Post Polymerization by UVDouble Bond at 1612 cm-1 Completely Consumed by Polymerization
Double Bond
John Donohue - [email protected]
John Donohue - [email protected]
Double Bond @ 1614 cm-1
N, N, Dimethylacrylamide’s Double Bond is Consumed during the PolymerizationThe Two Carbons Become Part of the Cured Adhesive’s “Backbone”
The Absorption Disappears as the Adhesive CuresThis Chemical Attacks the Polycarbonate
John Donohue - [email protected]
The Isobornyl Acrylate’s Double Bond is also Consumed by the Polymerization
Double Bond
John Donohue - [email protected]
Drug Eluting Coronary Stent Coatings
Part 10: Drug Eluting Heart Stent Coatings
John Donohue - [email protected] ©1996 American Heart Association
Edelman, E. R. et al. Circulation 1996;94:1199-1202
Angioplasty and Stenting are Competitive ProceduresDivergent processes of vascular repair after balloon angioplasty
and stenting of an atherosclerotic vessel
John Donohue - [email protected]
Divergent processes of vascular repair after balloon angioplasty and stenting of an atherosclerotic vessel. Balloon angioplasty (top) compresses and fractures the atherosclerotic plaque (light gray) and tunica media (black), slightly enlarging the artery. After a few days, a thin layer of platelet-rich thrombus (dark gray) lines the lumen and fills the dissection plane. The lumen shrinks from combined effects of early elastic recoil and later formation of a fibrocellular neointima (speckled area). Stent deployment after angioplasty (bottom) compresses the dissection plane and enlarges the lumen while stretching the artery with minimal elastic recoil. Within hours to days after stenting, caps of thrombus infiltrated with inflammatory cells (dark gray) form over stent struts (black rectangles), particularly abundant at sites of deep injury. Over ensuing weeks, a neointima forms (speckled area), thicker where injury is more severe. Although intimal growth after stenting is greater than after balloon angioplasty, the residual lumen is also larger, as the scaffolding of the stent maintains luminal dimensions. Late changes in arterial size are not depicted because the contribution of remodeling to restenosis after angioplasty or stenting remains incompletely characterized. (Figure prepared by James Squire.)
Edelman, E. R. et al. Circulation 1996;94:1199-1202
John Donohue - [email protected]
Stents scrape blood vessel walls. This injury causes reblockage.
Edelman and Squire
John Donohue - [email protected]μ X 550μ 300μ X 390μ
A Different Manufacturer’sDrug Eluting Coated Stent
IR Spectrum of PC 1036 from Biomaterials 21 (2000) 1847-1859
Phosphatidyl Choline (PC)coating, invented byBiocompatibles, Ltd., usedon some Medtronic andAbbott Drug ElutingCoronary Stents.
Spectra of coatings obtainedoff the actual stent surfacesshow the PC coating and theanti-restenosis drug it elutes.
All drug was extracted fromthe Abbott stent but PCcoating remains.
John Donohue - [email protected]
John Donohue - [email protected]
IR Spectra of pure PC, PC 1036, Medtronic Stent Coating, and Abbott Stent Coating
The Pure PC’s IR shows most of the vibrations present in the PC 1036 coating.
The PC 1036 IR was published back when Biocompatibles, Ltd. was trying to “drum up” big company interest in their materials. [This particular spectrum is slightly distorted (an “enlarged” 1090 vibration) because it is a surface spectrum (obtained via ATR) of a PC polymer in which the hydrophilic PC moieties have been rotated preferentially to the surface by contact with water. The 1090 is the C-O-P stretching vibration.]
The Medtronic (Endeavor™) and Abbott (BiodivYsio™) stent coating spectra were obtained by reflection off the stents’ surfaces.
The stent coatings are excellent matches to the spectrum of PC 1036. Since this is the coating Biocompatibles, Ltd. developed for the BiodivYsio stent, I believe it is the PC coating on the Medtronic and Abbott stents. EDAX can be run using the SEM to see if silicon is detected from the PC 1036 TSMA component. The literature published by Biocompatibles, Ltd. suggests a TSMA content of 3 to 5%. At this loading EDAX should detect the silicon.
John Donohue - [email protected]
Biocompatibles, Ltd. developed the PC-coatedBiodivYsio™ stent and marketed a number of PCpolymers. One of these polymers, PC1036, appearsto be the PC coating on the BiodivYsio stent,now a product of Abbott. It also appears to bethe coating Abbott has licensed to Medtronic forthe Endeavor stent.
PC 1036 is made using the four acrylic monomersshown here.>2-methacryloyloxyethyl phosphorylcholine (MPC)>lauryl methacrylate (LMA)>hydroxypropyl methacrylate (HPMA)>3-trimethoxysilylpropyl methacrylate (TSMA)
MPC supplies the PC functionality to the polymer.LMA helps the polymer adhere to metal surfaces.TSMA makes the polymer crosslinkable, whichimproves its adhesion and cohesion and helps control its rate of drug elution.HPMA is a co-crosslinker that, used with TSMA at 25% loading, gave good mechanical properties.AIBN (azoisobutyronitrile) initiates polymerization.
PC 1036 Biocompatibles, Ltd.
Biocompatibles, Ltd. paper
John Donohue - [email protected]
TSMA Hydrolysis initiates the Crosslinking of the Coating
The Drug Elution Rate is determined mainly by the extent of crosslinking.The hydroxy groups of the HPMA moiety will crosslink with TSMA also.
Biocompatibles, Ltd. paper
John Donohue - [email protected]
Part 11: Particulate Identification
John Donohue - [email protected]
Cardboard-Colored CelluloseOrganic Coating on one side
John Donohue - [email protected]
John Donohue - [email protected]
John Donohue - [email protected]
John Donohue - [email protected]
Biodegradation of an InsectThis insectdied ofnatural causesin the springof 2006.
(No insect’swere harmedin themaking ofthis document)
After storage in an airtight container for morethan one year, webbing is growing on it.
All 8.5mm X 11.2mm
Part 12: Fish Hooks, “Bait”, and Electronics
John Donohue - [email protected]
Views of Webbing8.5mm X 11.2mm
John Donohue - [email protected]
… yields a better protein Spectrum after it is Squashed
-0.2
-0.1
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
1.3
Abso
rban
ce
1000 1500 2000 2500 3000 3500 Wavenumbers (cm-1)
550u X 420u FOVsSpectra of Webbing on Aluminum Sheet …
Webbing placed on sheet
Webbing squashed on sheet
John Donohue - [email protected]
-3.0
-2.5
-2.0
-1.5
-1.0
-0.5
0.0
0.5
1.0
1.5
2.0
Abso
rban
ce 1000 1500 2000 2500 3000 3500 4000
Wavenumbers (cm-1)
Polyimide
White Epoxy
Green Ink
Blue Ink
8.5mm X 11.2mm
2.3mm X 3.0mm
Epoxy
Polyimide
Metal
Inkjet Cartridge Printhead
Laser Cut Metal “Window”
John Donohue - [email protected]
11.2mm X 8.5mm
3.0mm X 2.3mm
-2.5
-2.0
-1.5
-1.0
-0.5
0.0
0.5
1.0
1.5
2.0
2.5
3.0
Ab
so
rba
nc
e
1000 2000 3000 4000 Wavenumbers (cm-1)
PET + Butylacrylate Adhesive
Butylacrylate Glue
Cu & Glue is between PET & Polyimide
Ceramic beneath slotted Metal
Micaceous Cleavage of Ceramic
Printhead Dismantled – Piezoelectric Ceramic Under Metal
John Donohue - [email protected]
Who made the Chip that’s encased in hard black epoxy?
Original Photo Flipped Image
Toshiba
John Donohue - [email protected]
Fish Hook Manufacturing Processes:Cutting, Bending, Welding, Coating
8.5mm X 11.2mm FOVs
John Donohue - [email protected]
Fish Hook’s Point and Scratched Inorganic Coating:420u X 550u FOVs