in-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer...

10
Test method In-line monitoring ow in an extruder die by rheo-optics Jorge Silva a, * , Adillys C. Santos b , Sebasti ~ ao V. Canevarolo a a DEMa e UFSCar, Department of Materials Engineering, Federal University of S~ ao Carlos, S~ ao Paulo 13565-905, Brazil b PPG-CEM, Graduate Program in Materials Science and Engineering, Federal University of S~ ao Carlos, S~ ao Paulo 13565-905, Brazil article info Article history: Received 12 September 2014 Accepted 17 October 2014 Available online 28 October 2014 Keywords: Birefringence Slit die Inline monitoring Rheo-optics Stress-optical rule Extrusion abstract A new modular slit die with optical windows in two different positions and three pressure transducers ush-mounted along the wall was built and coupled to the exit of a twin- screw extruder. Thus, the birefringence and the pressure drop of polystyrene were monitored inline during extrusion. Two experimental procedures were tested: steady- state and cessation of extruder feeding. The latter proved to be very useful in the case of polystyrene since the ratio between the birefringence and the pressure drop can be quantied for a wide range of steady-state conditions with a single experiment. In fact, down to relatively lower values of pressure drop, the birefringence proved to be a function of shear stress at the wall only, depending neither on the initial feeding rate nor on the aspect ratio of the slit die, for W/h down to 5, at least. © 2014 Elsevier Ltd. All rights reserved. 1. Introduction Understanding the ow character of polymers during processing is of extreme importance to control the nal properties of the material. Most rheological analysis to assess the material functions, aiming at the prediction of polymer processability, has been undertaken from ofine experiments, either by the use of rotational or capillary rheometers, at low and high shear stresses, respectively. Despite the indisputable usefulness of these bench tech- niques, the quest for in-process assessment of thermo- mechanical histories imposed on the material and by extension structural changes has been a concern. Over recent years, the inline monitoring of polymer extrusion by attaching a slit channel to the extruder die exit has received considerable attention. A relatively large number of experimental techniques have been used to monitor polymers and polymer composites owing through a slit die [1,2]. It is well-known that the installation of ush-mounted pressure gauges along the slit die wall allows the assessment to viscous features up to relatively high shear stresses with no need of Bagley corrections. It is well established in the literature that slit dies with aspect ratios higher than 10 (W/h > 10) fall within the concept of parallel plates of innite width. Thus, the ow may be looked on as unidimensional and the edge corrections can be neglected. Thus, the wall shear stress is customarily obtained by the equation [3] s w ¼ h 2ð1 þ h=WÞ dp dx (1) where h and W are, respectively, the height and the width of the slit, and dp/dx is the gradient of the outward-acting total normal stress at the wall measured with pressure gauges in the fully developed ow region of slit-die ge- ometries. The corresponding wall shear rate, _ g w , is given by _ g w ¼ _ g a 3 2 þ d ln _ g a d ln s w (2) where _ g a is the apparent shear rate, which is expressed by _ g a ¼ 6Q Wh 2 (3) with Q being the volumetric ow rate. * Corresponding author. E-mail address: [email protected] (J. Silva). Contents lists available at ScienceDirect Polymer Testing journal homepage: www.elsevier.com/locate/polytest http://dx.doi.org/10.1016/j.polymertesting.2014.10.007 0142-9418/© 2014 Elsevier Ltd. All rights reserved. Polymer Testing 41 (2015) 63e72

Upload: others

Post on 11-Feb-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: In-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer composites flowing throughaslitdie[1,2].Itiswell-knownthattheinstallation of flush-mounted

ilable at ScienceDirect

Polymer Testing 41 (2015) 63e72

Contents lists ava

Polymer Testing

journal homepage: www.elsevier .com/locate/polytest

Test method

In-line monitoring flow in an extruder die by rheo-optics

Jorge Silva a, *, Adillys C. Santos b, Sebasti~ao V. Canevarolo a

a DEMa e UFSCar, Department of Materials Engineering, Federal University of S~ao Carlos, S~ao Paulo 13565-905, Brazilb PPG-CEM, Graduate Program in Materials Science and Engineering, Federal University of S~ao Carlos, S~ao Paulo 13565-905, Brazil

a r t i c l e i n f o

Article history:Received 12 September 2014Accepted 17 October 2014Available online 28 October 2014

Keywords:BirefringenceSlit dieInline monitoringRheo-opticsStress-optical ruleExtrusion

* Corresponding author.E-mail address: [email protected] (J. Silva).

http://dx.doi.org/10.1016/j.polymertesting.2014.10.000142-9418/© 2014 Elsevier Ltd. All rights reserved.

a b s t r a c t

A new modular slit die with optical windows in two different positions and three pressuretransducers flush-mounted along the wall was built and coupled to the exit of a twin-screw extruder. Thus, the birefringence and the pressure drop of polystyrene weremonitored inline during extrusion. Two experimental procedures were tested: steady-state and cessation of extruder feeding. The latter proved to be very useful in the caseof polystyrene since the ratio between the birefringence and the pressure drop can bequantified for a wide range of steady-state conditions with a single experiment. In fact,down to relatively lower values of pressure drop, the birefringence proved to be a functionof shear stress at the wall only, depending neither on the initial feeding rate nor on theaspect ratio of the slit die, for W/h down to 5, at least.

© 2014 Elsevier Ltd. All rights reserved.

1. Introduction

Understanding the flow character of polymers duringprocessing is of extreme importance to control the finalproperties of the material. Most rheological analysis toassess the material functions, aiming at the prediction ofpolymer processability, has been undertaken from offlineexperiments, either by the use of rotational or capillaryrheometers, at low and high shear stresses, respectively.Despite the indisputable usefulness of these bench tech-niques, the quest for in-process assessment of thermo-mechanical histories imposed on the material and byextension structural changes has been a concern.

Over recent years, the inline monitoring of polymerextrusion by attaching a slit channel to the extruder die exithas received considerable attention. A relatively largenumber of experimental techniques have been used tomonitor polymers and polymer composites flowingthrough a slit die [1,2]. It is well-known that the installationof flush-mounted pressure gauges along the slit die wallallows the assessment to viscous features up to relatively

7

high shear stresses with no need of Bagley corrections. It iswell established in the literature that slit dies with aspectratios higher than 10 (W/h > 10) fall within the concept ofparallel plates of infinite width. Thus, the flow may belooked on as unidimensional and the edge corrections canbe neglected. Thus, the wall shear stress is customarilyobtained by the equation [3]

sw ¼ h2ð1þ h=WÞ

dpdx

(1)

where h and W are, respectively, the height and the widthof the slit, and dp/dx is the gradient of the outward-actingtotal normal stress at the wall measured with pressuregauges in the fully developed flow region of slit-die ge-ometries. The correspondingwall shear rate, _gw , is given by

_gw ¼ _ga

3

�2þ d ln _ga

d ln sw

�(2)

where _ga is the apparent shear rate, which is expressed by

_ga ¼6QWh2 (3)

with Q being the volumetric flow rate.

Page 2: In-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer composites flowing throughaslitdie[1,2].Itiswell-knownthattheinstallation of flush-mounted

J. Silva et al. / Polymer Testing 41 (2015) 63e7264

Under flow, most polymer melts become opticallyanisotropic, creating the possibility of using non-invasiveoptical techniques to understand the chain conformationand orientation during flow. Moreover, for homopolymersand a relatively broad range of experimental conditions, it ispossible to establish a relation between the stresses in thematerial and the optical anisotropyas given by the refractiveindex tensor, n. Therefore, assuming that the principal axesof the stress tensor, s, and the refractive index tensor arecollinear, the stress-optical rule may be written as [4,5]

n� ⟨n⟩I ¼ Cfsþ pIg (4)

where C is the stress-optical coefficient, ⟨n⟩ is the meanrefractive index, p is the hydrostatic pressure and I is theunit tensor. One of the most commonways tomeasure thebirefringence of a polymer in shear flow is to send a lightbeam in the velocity gradient direction (y-direction). Itshould be noted, however, that the application of equation(4) is not straightforward since the velocity gradient di-rection is neither principal axes of the refractive indextensor nor principal axes of the stress tensor. Neverthe-less, it is possible to show that, for small values of bire-fringence, the following expression is valid [6]

Dn13 ¼ C½s11 � s33� (5)

Additionally, for flow in a slit die, the light traveling in y-direction passes layers of varying degrees of molecularorientation and, therefore, only an average birefringence,⟨Dn13⟩, is measured. However, Wales [7] was able to deducea relation between the average birefringence and thebirefringence at the wall, Dn13,w,

Dn13;w ¼ ⟨n13⟩

�1þ d log⟨n13⟩

d log sw

�(6)

which is analogous to the Rabinovich correction [Equa-tion (2)].

Early measurements of polymer flow birefringence in aslit die were carried out in the 1960s by Wales [7]. In hismaster thesis [8], under supervision of Janeschitz-Kriegl, hereported no influence of the aspect ratio (W/h) of the slit dieon the birefringence for W/h > 10. The influence of theentrance was also studied by Wales. He observed that, for alow density polyethylene flowing through a slit die thebirefringence, ⟨Dn13⟩, decreases with increases distance tothe entrance. Indeed, a steady flow condition, as character-ized by a birefringence not dependent to the distance to theentrance, is not even reached for reduced distances (l/h, withl being the distance to slit entrance) as large as 100 and atlow apparent shear rates. However, for polystyrene, Walesobserved no variation of the birefringence along the slitchannel even for a reduced distance of 15. Obviously, thedifferent behavior of LDPE and PS could stem from the largediscrepancy between the relaxation times of the polymers.

From equation (5), it follows that, if one disregards thedifference between s11�s22 and s11�s33, the birefringencemeasured in 13 plane can be directly related to the firstnormal stress difference, N1. This quantity is an indicationof the elasticity of thematerial and, in general, relates to theshear stress by the following expression [9]

N1 ¼ Asa12 (7)

in which A and a are constants. It is well-known that, in thelimit of the linear viscoelastic range, the value of the slope ais 2. Since the zero-shear viscosity of monodisperse, linearpolymers varies with the 3.4 or 3.5 power of the molecularweight, the zero shear-rate first normal stress difference isexpected to increase with molecular weight to about thepower 7. Measurements of the birefringence on slit dieshave shown that the constant A increases with the molec-ular weight increase even at relatively high shear rates[8,10]. Regarding the influence of molecular weight distri-butions, it appears that broadening the distribution of alinear polymer, while keeping the average molecularweight and the shear stress unchanged, results in an in-crease in the normal stress difference but a decrease in theexponent a [9,11,12]. Furthermore, if Equation (7) holds, thefirst normal stress difference versus shear rate or shearstress must follow timeetemperature superposition. Thatsuperposition was observed for PS in a slit die up to shearstresses of 105 Pa by using the birefringence as ameasure ofN1 [13].

Several authors [14e24] have used the birefringencefringes to visualize the stress state in a slit die. In fact, bycounting the birefringence fringes it is possible to spatiallymap the stresses along a slit die. This technique has beenextensively employed in order to study the flow in theentrance and exit of slit dies. However, the quantification ofbirefringence by counting the fringes is not practical and,more important, it cannot be realized at high stresses. Inaddition to spatial visualization, other authors [25e27]have done a pointwise quantification of the birefringenceby employing a laser as the light source and a detector.

In this paper we aim to show that the more traditionalapproach of controlling the extrusion [28] or injectionmolding [29] processes by inline monitoring the pressuredrop in a slit die can be significantly improved if the bire-fringence is simultaneously measured. Here, a new versionof the rheo-optical system developed in a previous work[10] is presented. The flow birefringence of polystyrenewasmeasured. The three flush-mounted pressure transducersand two optical windows allow establish correlations be-tween the pressure drop and birefringence in two differentregions of the slit channel. With this data, it should beeasier to infer about variations in the temperature due toviscous heating. Also, the “cessation of extruder feeding”approach is presented and its potentialities analyzed. Insome situations, it can be more convenient to use rectan-gular ducts withW/h< 10 aspect ratio. It will be shown thatthe rheo-optical information obtained in those channelscan be easily extrapolated for channels that meet thetraditional “infinitely wide condition”, W/h > 10.

2. Experimental

2.1. Materials

A commercial grade of polystyrene (PS), referred to asN2560, from Innovawas used in this work. Throughout thispaper a value of 0.94 g.cm�3 [30] will be considered for the

Page 3: In-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer composites flowing throughaslitdie[1,2].Itiswell-knownthattheinstallation of flush-mounted

Fig. 1. Schematic of slit die, locations of pressures holes, positions of opticaldetectors and their main parts; A ¼ D ¼ 11.25, B ¼ C ¼ 53.5 mm.

J. Silva et al. / Polymer Testing 41 (2015) 63e72 65

average melt density of PS in the temperature rangestudied.

2.2. Rotational rheometry

To perform the rheological experiments, 25 mm diam-eter disks were prepared by compression molding. Rheo-logical measurements were carried out on a rotationalrheometer (ARES), equipped with parallel plates of 25 mmdiameter, at three different temperatures: 200, 230 and260 �C. The distance between the plates was set to 1 mm.All rheological experiments were conducted under a ni-trogen atmosphere in order to preclude, or at least severelydelay, oxidative degradation of the samples.

2.3. Slit die

The rheo-optical system described here is a largemodification of a previous work presented elsewhere[10]. A slit die of special design has been constructedwhich is coupled to the end of a twin-screw extruder. It ismade of tool steel and is composed of two modulesseparated by calibrated spacers, as shown in Fig. 1. Thisgives us the possibility to create ducts of different widthsjust by changing the spacers in the die. Hence, slitchannels with different aspect ratios, or even convergentor divergent channels, can be studied. In this work, twopairs of spacers were used such that two slit channelswith dissimilar aspect ratios were attained. The variationin the aspect ratio is obtained by varying the width of therectangular cross section; in one case that width is7.5 mm and in the other one 15 mm. Both slit channelsare 130 mm long with the height of the rectangular crosssections being 1.5 mm. This originates slit dies withaspect ratios of 5 and 10.

Three pressure transducers are flush mounted alongthewall of the bottommodule in order to avoid dead spacebetween the tip of the transducers and the flow channel.Doing so, prevents flow disturbances, stagnation of themelt and hole effects that could possibly result in erro-neous readings. The pressure transducers measure thewall-acting normal stress, important to assess the shearstress at the wall, as depicted in Equation (1). The distancefrom the entrance of the die to the first pressure trans-ducer is about 11.3 mm; as the rectangular channel ex-tends 2 mm down the extrusion die, thus the final reduceddistance (l/h) is approximately 9. To prevent stagnation atthe entry of the slit die, the extrusion die was constructedwith an entrance angle of 60 degrees. Both top and bottommodules possess bores with diameters of approximately10 mm and depth of 1 mm where borosilicate opticalwindows can fit flush to the duct walls for further opticalmeasurements. Also, through holes of 4 mm machinedconcentrically to the bores to allow the monochromaticlight beam to pass orthogonally to the windows and flow,as will be explained in more detail later. The simple ge-ometry of the bores permits easy assembly of opticalwindows, avoids leak of melt since the windows arecompressed against the bottom of the bore and allowsstress birefringence to be ruled out. The reduced distancesof the first set of windows is c.a. 27 (l/h).

The slit die temperature is set with the use of a pro-portional derivative integral controller (PID) that furnishesthe necessary power to cartridge heaters placed on eachmodule according to the set-point required. The actualtemperature is measured with the help of a J-type ther-mocouple adjacent to the central pressure transducer andplaced at the vicinity of the channel.

2.4. Optical detectors

In each of the two pairs of bores of the slit die, an opticalarrangement was assembled such that the birefringence inthe 1-3 plan can be measured. This means that an averagebirefringence is probed due to the depth-dependence ofstresses when viewing from the duct wall to the center line,as already discussed in the introductory section. Alsoshown in Fig. 1 are the main parts of the optical detectorsbased on a simple polariscopic arrangement. Here, the axialdirection of the slit die channel is interposed to crosspolarizers bisecting the flat angle, i.e., the principal axis ofthe polarizers is at 45 degrees to the flow direction.

A red light provided by a light emitting diode e LED ofnarrow wavelength range l ¼ 620e645 nm is radiatedorthogonally to the melt (no considerable light interferenceis seen within the experimental conditions applied; thus it

Page 4: In-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer composites flowing throughaslitdie[1,2].Itiswell-knownthattheinstallation of flush-mounted

J. Silva et al. / Polymer Testing 41 (2015) 63e7266

may be considered monochromatic and an average wave-length of 632.5 nm is used for the sake of calculation). First,the light beam passes the polarizer, enters the flowingpolymer which rotates its polarization state due to theanisotropic character of themedium, then is analyzed by theopposite polarizer before reaching the photocell (lightdependent resistore LDR). The optical systemwas designedfor fast assemblyand all parts aremechanically independent.

2.5. Optical set-up

The whole set-up system is presented in Fig. 2. The slitdie described above is attached to a twin-screw extruder(model Werner-Pfleiderer ZSK e 30). A signal handling boxforms an electrical circuit with the photocells capable ofconverting the incident light (non-electrical information)into an electric voltage, V. This electrical signal is convertedin an A/D interface (USB data acquisition device NI-DAQ6812) and sent to a personal computer (read-out device).In order to handle the signal, software was designed inLabView 8.6 platform (from National Instruments) for datacollection, real-time calculation, screen presentation anddata saving. In addition, a hard key placed in the extrudercontrol panel may be set to enable the control of theextruder screw speed via software.

In order to normalize the voltage, V, sent by the pho-tocells, the transmitted light intensity is measured with thepair of polarizers crossed (Vc) and in parallel (Vp). Thus, anormalized voltage, Vn, is calculated by

Vn ¼ V � Vc

Vp � Vc(8)

For the parallel position, maximum light intensity istransmitted and so Vn ¼ 1 whereas, if the pair of polarizersare crossed, minimum light intensity is transmitted Vn ¼ 0.

2.6. Experimental procedure and signal analysis

For result analysis purposes, the slit die will be divided intwo sections. The pressure drop in the first section will be

Fig. 2. Schematic diagram of the experimental apparatus to inline monitor the

given by the difference between the pressures measured bythe first (nearest to the extruder) and the second (middle)transducers divided by the corresponding distance, whereasthe pressure drop in the second section will be calculatedwith values obtained in the second and third (nearest to theexit) transducers. Similarly, the birefringence measured inthe first window (nearest to the extruder) set at mid dis-tance between the transducers will be attributed to the firstsection, while the birefringencemeasured in the second onewill be attributed to the second section.

The extruder was fed at constant rates ranging between2 and 8 Kg/h. The barrel temperature was set to 200 or230 �C and the screw speed was kept constant at 70 rpm. Inthe steady-state, data were recorded and, thereafter, thefeeding was stopped (cessation of extruder feeding) whilekeeping the data recording and the screw rotating at thesame constant speed. During this procedure, both pressureslopes and birefringence decrease continually. Thus,depending on the initial value of the birefringence, theintensity of transmitted light can pass throughmaxima andminima while the pressure drop approaches zero. Anexperimental result where the second order is reached isshown in Fig. 3.

The average birefringence, ⟨n13⟩, is calculated accordingto [10]

⟨n13⟩ ¼ l

h:parcsinðVNÞ1=2 (9)

3. Results and discussion

3.1. Steady-state extruder feeding

In order to evaluate the stability of the flow though theslit channel, the values of the birefringence in the firstwindow are compared with the values in the second one.Similarly, the pressure drops between the first and secondtransducers and between the second and the third are alsocompared. Fig. 4 shows that the birefringence measured inthe second window is almost always higher than thatmeasured in the first window. Likewise, the pressure drop

pressure drop and birefringence of the polymer flow during extrusion.

Page 5: In-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer composites flowing throughaslitdie[1,2].Itiswell-knownthattheinstallation of flush-mounted

Fig. 3. Normalized voltage versus pressure slop as measured in the 1st and2nd sections of the slit die (W/h ¼ 10). These results were obtained intransient mode, during extruder discharge, at 200 �C and 70 rpm. The initialfeeding rate was 4 kg/h.

J. Silva et al. / Polymer Testing 41 (2015) 63e72 67

between the first and second pressure transducers is al-ways lower that the pressure drop between the second andthird ones. There is, however, a significant dissimilaritybetween the results obtained at different temperatures. At230 �C, the relative difference in the birefringence plottedas a function of the apparent shear rate follows the sametrend as the relative difference in the pressure drop, withthe values never exceeding 15%. On the other hand, at200 �C, the difference of birefringence measured in thedifferent windows increases sharply with the apparentshear rate increase, not following the difference in thepressure drop.

It is well known that several effects can contribute to thenon-uniformity of the pressure gradient through the slitchannel. Depending on the material and flow conditions,the formation of a full developed flow may require hugevalues of l/h. However, as already cited in the Introduction,for polystyrene, Wales [8] observed no variation of thebirefringence along the slit channel, even for short reduced

Fig. 4. Difference between the steady-state birefringence (filled symbols)measured in the first and second windows and between the steady-statepressure drops (open symbols) in the first and second sections. Poly-styrene at several apparent shear rates and temperatures (200 �C and230 �C) was tested.

distances and at relatively high shear rates. Also, theentrance effects, the viscous heating and the influence ofpressure on the viscosity can contribute to the non-uniformity of the velocity profile along the channel. It isnoteworthy, however, that the viscous heating and theviscosity dependence on the pressure induce a decrease inthe pressure gradient along the slit die (see p.e. [31,32])which seems to be in contradiction to the present results. Infact, they indicate a decrease of the melt temperature alongthe slit die.

In order to explain a possible decrease of the melttemperature along the slit die, it should be noted that theextruder barrel and the slit die were set at the same tem-perature. Because of viscous heating in the extruder, thetemperature of the material entering the slit die will beseveral degrees above the set temperature. Therefore, if theviscous heating in the slit is not high enough to keep thetemperature and the residence time is high enough, a sig-nificant decrease of the melt temperature can occur alongthe channel. A crude estimation of viscous heating can bedone by comparing the viscosity measured offline withviscosity calculated from the pressure slopes in the slit die(Fig. 5). In this way, for a set temperature of 230 �C, thetemperature of the melt polystyrene is estimated to be 25and 20 �C (Fig. 5a) above in the first and second sections ofthe slit die, respectively, whereas for 200 �C the DT isestimated to be 35 and 30 �C (Fig. 5b). It should beemphasized that the viscosities calculated from the pres-sure drops in the slit die are not true viscosities. Instead,they should be viewed as apparent viscosities, mainlybecause the temperature is not uniform along the slitchannel. At this point, we should note that, whereas theaverage residence time in the slit die ranges between 1.2and 4.9s for flow rates varying between 8 and 2Kg/h, thematerial near to the walls will take longer to cross thechannel.

From Fig. 5, it follows that a t-T superposition with theshift factors calculated from the rotational rheometryshould hold for the curve of shear stress versus shear rate atthe wall, if the real, estimated, temperatures are considered(Fig. 6). It should be noted, however, that for a set tem-perature of 200 �C the points corresponding at 2 Kg/hseems to be out of the general trend (Fig. 5b). This appearseven more clearly in Fig. 6, indicating that at 2 Kg/h theviscous heating and heat transfer phenomena lead tosignificantly different temperatures than those reached forthe other flow rates studied. Remarkably, if the same aTvalues are used to shift the birefringence a superpositionseems to occur. Thus, a curve of Dn13,w versus s12,w shouldbe independent of the temperature and flow rate in therange studied. In fact, Fig. 7a seems to shows that the datasuperpose on a master curve, as expected, even the pointfor 2Kg/h lies on this curve. Nevertheless, the data for thefirst section of the slit die do not seem to superpose on amaster curve (Fig. 7b).

The importance of the superposition of the data in amaster curve that does not depends on the flow rate andtemperature should be highlighted since it could, in prin-ciple, allow detection of changes in the material properties,such as molecular weight and molecular weightdistribution.

Page 6: In-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer composites flowing throughaslitdie[1,2].Itiswell-knownthattheinstallation of flush-mounted

Fig. 6. Birefringence and shear stress at the wall as a function of _g12;w=aT forthe 2nd section of the slit die. The real temperatures as estimated in theprevious graphs were considered. Therefore the aT(230 �C) and aT(250 �C)shift factors calculated from rotational rheometry were used with a Tref of230 �C.

Fig. 5. Comparison between the viscosity of the polystyrene as measuredoffline in a rotational rheometer and inline in the slit die attached to thetwin screw extruder. The temperatures of extruder barrel and slit die wereset to (a) 200 and (b) 230 �C. To calculate the viscosity in the 1st section ofthe slit die the pressure drop between the first and second pressure trans-ducers was used whereas to determine the viscosity in the 2nd section thepressure drop between the second and third pressure transducers was takeninto account.

J. Silva et al. / Polymer Testing 41 (2015) 63e7268

3.2. Cessation of extruder feeding

To perform steady-state experiments at several flowrates requires great amounts of material and is time-consuming. In this section, it will be shown that thecessation of feeding experiments can, in principal, be agood alternative and/or complement the steady-state ex-periments. The purpose is to scan thematerial behavior in arelatively wide range of flow conditions with a singleexperiment. The inline monitoring during the cessation offeeding experiments have already been performed for PSand PS/PMMA blends [33]. In these work, a deeper insightof these kinds of experiments is given, namely, by con-ducting experiments at different temperatures and initialflow rate with the new slit die described above. The pro-cedure consists of stopping the feeding of the extruderwhen a steady-state is reached, keeping the extruder screwrotating at the same speed. As time passes, a continuouslyreducing amount of molten polymer remains inside theextruder, smoothly dropping the pressure and the shearrate as it exits the slit-die.

The evolution of the pressure drop and birefringenceafter the cessation of feeding on the 2nd section of the slitdie, for a temperature of 230 �C, is shown on Fig. 8. The twosignals change in a similar way. For the highest flow rates,there is first a decrease before a shoulder which is followed

Fig. 7. Birefringence at the wall as a function of the shear stress at the wallfor the (a) 2nd and (b) 1st sections of the slit die.

Page 7: In-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer composites flowing throughaslitdie[1,2].Itiswell-knownthattheinstallation of flush-mounted

J. Silva et al. / Polymer Testing 41 (2015) 63e72 69

by a (probably) exponential decrease. Another noteworthyfact is that all the curves converge to the same values atlong times. In the limit, the time necessary to empty theextruder does not depend on the initial feeding rate. Theresults from the 1st section of the slit die are similar and,therefore, will not be shown here. We think that the profileof the pressure drop and birefringence decreases should benot only a function of the rheological properties of thematerial, but also a function of the extruder screw profile.In fact, the shoulders observed for the highest flow ratesseem to be related to the kneading block of 45� screwelement placed at some distance before the extruder exit.This idea will be developed in further work. The shape ofthe curves of pressure drop and birefringence versus timefor 200 �C are slightly different, but most of the observa-tions done for 230 �C are still valid (Fig. 9).

One of the most remarkable observations of this work isthe collapse of the average birefringence versus pressuredrop curves, obtained at different flow rates, in a single,master, curve (Fig. 10). It should be emphasized that, asdiscussed below, the conditions at the slit die are far fromideal since the polymer is entering the slit die at temper-atures significantly above the set temperature. Neverthe-less, the collapse in a master curve occurs for both sectionsof the slit die and at both temperatures. For the sametemperature, the curves for the 1st and 2nd sections of the

Fig. 8. (a) Pressure drop and (b) birefringence as a function of the time at230 �C in the 2nd section of the slit die after cessation of feeding.

Fig. 9. (a) Pressure drop and (b) birefringence as a function of the time at200 �C in the 2nd section of the slit die after cessation of feeding.

slit die have the same slope but they are shifted. In Fig. 11,the results obtained at different temperatures for the 2nd

section of the slit die are compared. The effect of thetemperature is very small since the curves are only slightlyshifted. This phenomenon should be a consequence of thet-T superposition, which holds for polystyrene; a variationin the temperature leads to similar shifts in both N1 andshear stress. Obviously, in this discussion we are assumingthe validity of the stress-optical rule and the temperature-independence of the stress-optical coefficient which,indeed, has been observed to be nearly temperature-independent for PS [8,34].

3.3. Aspect ratio W/h ¼ 5

All the previous results were obtained in a slit die withW/h ¼ 10. In that case, a significant contribution of thelateral walls to the pressure drop is not expected. However,in some situations, it can be more convenient to use rect-angular ducts with lower W/h aspect ratios. Thus, it isimportant to verify if the conclusions above outlined for“infinite width” slit dies can be extended to rectangularducts where the height is comparable to the width.

Firstly, it should be noted that, for the W/h ¼ 5 rectan-gular duct, a range of feeding rates as large as those testedfor W/h ¼ 10 could not be achieved since that would

Page 8: In-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer composites flowing throughaslitdie[1,2].Itiswell-knownthattheinstallation of flush-mounted

Fig. 10. Birefringence as a function of pressure drop for (a) 230 and (b)200 �C.

J. Silva et al. / Polymer Testing 41 (2015) 63e7270

overload the extruder torque. Nevertheless, the dataattained permit drawing some conclusions. In the sameway as observed for W/h ¼ 10, for a W/h ¼ 5 channel thecurves of birefringence versus pressure drop obtained aftercessation of feeding superpose regardless of the initialfeeding rate (Fig. 12). Furthermore, those curves are nearlytemperature independent in the range studied, which isalso an observation common to the infinite width case. This

Fig. 11. Master curves of birefringence as a function of pressure drop for the2nd section of the slit die at 200 �C and 230 �C.

can be seen in Fig. 13, which compares the results for bothdies, at both temperatures, in the second section of thechannel.

Also, Fig. 13a shows the same slope for the curves ofboth rectangular ducts. However, they are shifted with theW/h ¼ 10 slit die exhibiting higher values of birefringencefor the same pressure drop. In order to explain these re-sults, it should be recalled that a rectangular duct cannot betreated as a slit die with infinite aspect ratio.

Numeric calculations for power-law fluids in rectan-gular channels have been performed by several authors[35,36]. Son [37] proposed an approach to calculate theshear viscosity and shear rate with the pressure drop andflow rate data from channels having aspect ratios down to1. Based on previous work of Hartnett and Kostic [38], hederived an expression to calculate the shear rate at thewall.To compute the shear stress at the wall he used Equation(1). By following that approach, he obtained a super-position of the data for rectangular ducts with aspect ratiosranging between 1 and 10. In the present work, if the pa-rameters a and A of Equation (7) do not depend on theshear rate and, providing that C is not a function of shearrate in the range investigated, no correction is necessary totake into account the different shear rates in the differentaspect ratio channels. Thus, if the birefringence is repre-sented as a function of the shear stress at the wall, a

Fig. 12. Birefringence as a function of pressure drop for (a) 230 and (b)200 �C in the slit die with aspect ratio (W/h) ¼ 5.

Page 9: In-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer composites flowing throughaslitdie[1,2].Itiswell-knownthattheinstallation of flush-mounted

Fig. 13. Birefringence as a function of a) pressure drop and b) shear stress atthe wall. The results were obtained in the 2nd section of the rectangularchannels W/h ¼ 5 and W/h ¼ 10 at both temperatures 200 and 230 �C. In thegraphic are represented the curves for all feeding rates studied. It should behighlighted that the data for both rectangular channels are represented ingraph b).

J. Silva et al. / Polymer Testing 41 (2015) 63e72 71

superposition of the data measured for both rectangularchannels is observed (Fig. 13b).

4. Conclusions

A new slit die with optical windows in two differentpositions and three pressure transducers flush-mountedalong the wall was built. It was designed such that, bysimply changing the spacers, ducts of different geometriesor with dissimilar aspect ratios can be implemented.

The birefringence and the pressure drop of a poly-styrene flowing through rectangular channels weremeasured. In fact, two rectangular channels with the sameheight, 1.5 mm, and different widths, 15 and 7.5 mm,coupled one at a time to the exit of a twin-screw extruder.

The birefringence and pressure drop measured in thefirst section of the slit die are different to thosemeasured inthe second section. In addition, a comparison of the vis-cosities obtained offline in a rotational rheometer andinline in the slit die revealed large differences. Those dif-ferences between inline and offline measurements arehigher for the first section of the slit die. The results wereexplained by the occurrence of viscous heating in the

extruder, which makes the temperature of the polymerentering the slit die significantly higher than that set.Nevertheless, it was shown that the t-T superposition holdsfor the birefringence measured in the 2nd section of the slitdie.

An innovative cessation of extruder feeding procedurewas developed. It proved to be very useful in the case ofpolystyrene since the ratio between the birefringence andthe pressure drop can be quantified for a wide range ofsteady-state conditions with a single experiment. In fact,down to relatively lower values of pressure drop, thefunction Dn13/(Dp/Dx) was proved to be independent of theinitial feeding rate. Furthermore, for rectangular channelswith aspect ratios (W/h) down to at least 5, there is su-perposition of the curves of Dn13 versus sw.

In summary, it was shown that rheo-optics can be usedto monitor inline the extrusion process. For example, whilethe curves of logDn13 versus logsw are independent of thefeeding rate and nearly independent of the temperature,their slopes should change and vertical shifts should beobserved if the molecular weight and/or molecular weightdistribution are altered.

Some questions remain open, however. The first relatesto the melt temperature. A crude estimation of viscousheating was made in this paper but a more direct mea-surement of the polymermelt temperature is desirable. It iswell-known, however, that those measurements are diffi-cult to perform. A second question concerns to the timeevolution of the birefringence and pressure drop in cessa-tion of extruder feeding experiments. Probably, that timeevolution is a function of the screw profile. These importantpoints will be addressed in future work.

Acknowledgments

Financial support from CAPES through Grant BJT 019/2012 to J. S. and from CNPq through a PQ scholarship311790/2013-5 to S. V. C. and 142325/2011-2 to A. C. S. aregratefully acknowledged.

References

[1] I. Alig, B. Steinhoff, D. Lellinger, Monitoring of polymer melt pro-cessing, Meas. Sci. Technol. 21 (2010).

[2] S.E. Barnes, M.G. Sibley, H.G.M. Edwards, P.D. Coates, Processmonitoring of polymer melts using in-line spectroscopy, T. I. Meas.Control 29 (2007) 453e465.

[3] H.M. Laun, Polymer melt rheology with a slit die, Rheologica Acta 22(1983) 171e185.

[4] G.G. Fuller, Optical rheometry, Annu. Rev. Fluid Mech. 22 (1990)387e417.

[5] H. Janeschitz-Kriegl, Polymer Melt Rheology and Flow Birefringence,Springer-Verlag, Berlin ; New York, 1983.

[6] F.A. Morrison, Understanding Rheology, Oxford University Press,New York, 2001.

[7] J.L.S. Wales, Apparatus for the measurement of flow birefringence ofpolymer melts at high shear stresses, Rheologica Acta 8 (1969)38e44.

[8] J.L.S. Wales, The Application of Flow Birefringence to RheologicalStudies of Polymer Melts, Delft University Press, Delft, 1976.

[9] R.K. Gupta, Polymer and Composite Rheology, second ed., MarcelDekker, New York, 2000.

[10] K. Soares, A.M.D. Santos, S.V. Canevarolo, In-line rheo-polarimetry: amethod to measure in real time the flow birefringence duringpolymer extrusion, Polym. Test 30 (2011) 848e855.

Page 10: In-line monitoring flow in an extruder die by rheo-optics · monitor polymers and polymer composites flowing throughaslitdie[1,2].Itiswell-knownthattheinstallation of flush-mounted

J. Silva et al. / Polymer Testing 41 (2015) 63e7272

[11] K. Oda, J.L. White, E.S. Clark, Correlation of normal stresses inpolystyrene melts and its implications, Polym. Eng. Sci. 18 (1978)25e28.

[12] R. Racin, D.C. Bogue, Molecular-weight effects in die swell and inshear rheology, J. Rheol. 23 (1979) 263e280.

[13] J.L.S. Wales, H. Janeschi, Flow birefringence in macromolecular liq-uids, Kolloid Z. Z. Polym. 250 (1972) 1142e1149.

[14] P. Beaufils, B. Vergnes, J.F. Agassant, Characterization of the shark-skin defect and its development with the flow conditions, Int.Polym. Proc. 4 (1989) 78e84.

[15] N. Checker, M.R. Mackley, D.W. Mead, On the flow of molten polymerinto, within and out of ducts, Philos. T R Soc. A 308 (1983) 451.

[16] N. ElKissi, J.M. Piau, F. Toussaint, Sharkskin and cracking of polymermelt extrudates, J. Non-Newton Fluid 68 (1997) 271e290.

[17] C.D. Han, L.H. Drexler, Studies of converging flows of viscoelasticpolymeric melts .1. Stress-Brefringent measurements in entranceregion of a sharp-edged slit die, J. Appl. Polym. Sci. 17 (1973)2329e2354.

[18] C.D. Han, L.H. Drexler, Studies of converging flows of viscoelasticpolymeric melts .3. Stress and velocity distributions in entranceregion of a tapered slit die, J. Appl. Polym. Sci. 17 (1973) 2369e2393.

[19] C.D. Han, L.H. Drexler, Stress distribution of polymer melts in exitregion, T Soc. Rheol. 17 (1973) 659e665.

[20] D.G. Kiriakidis, H.J. Park, E. Mitsoulis, B. Vergnes, J.F. Agassant, Astudy of stress-distribution in contraction flows of an Lldpe melt, J.Non-Newton Fluid 47 (1993) 339e356.

[21] F. Legrand, J.M. Piau, Spatially resolved stress birefringence and flowvisualization in the flow instabilities of a polydimethylsiloxaneextruded through a slit die, J. Non-Newton Fluid 77 (1998)123e150.

[22] M.R. Mackley, R.P.G. Rutgers, D.G. Gilbert, Surface instabilities dur-ing the extrusion of linear low density polyethylene, J. Non-NewtonFluid 76 (1998) 281e297.

[23] A.J. Mchugh, M.E. Mackay, B. Khomami, Measurement of birefrin-gence by the method of isoclinics, J. Rheol. 31 (1987) 619e634.

[24] Gv Vinograd, Ek Borisenk, B.B. Boiko, N.I. Insarova, Critical regimesof shear in linear polymers, Polym. Eng. Sci. 12 (1972) 323.

[25] J. Barone, S.Q. Wang, Flow birefringence study of sharkskin andstress relaxation in polybutadiene melts, Rheologica Acta 38 (1999)404e414.

[26] J.R. Barone, S.Q. Wang, Rheo-optical observations of sharkskin for-mation in slit-die extrusion, J. Rheol. 45 (2001) 49e60.

[27] S.R. Galante, P.L. Frattini, Spatially-resolved birefringence studies ofplanar entry flow, J. Non-Newton Fluid 47 (1993) 289e337.

[28] B.K. Nguyen, G. McNally, A. Clarke, Real time measurement andcontrol of viscosity for extrusion processes using recycled materials,Polym. Degrad. Stabil. 102 (2014) 212e221.

[29] A. Fernandez, M. Muniesa, C. Javierre, In-line rheological testing ofthermoplastics and a monitored device for an injection mouldingmachine: application to raw and recycled polypropylene, Polym.Test 33 (2014) 107e115.

[30] T. Ougizawa, G.T. Dee, D.J. Walsh, Pvt properties and equations ofstate of polystyrene e molecular-weight dependence of the char-acteristic parameters in equation-of-state theories, Polymer 30(1989) 1675e1679.

[31] D.M. Binding, M.A. Couch, K. Walters, The pressure dependence ofthe shear and elongational properties of polymer melts, J. Non-Newton Fluid 79 (1998) 137e155.

[32] H.M. Laun, Pressure dependent viscosity and dissipative heating incapillary rheometry of polymer melts, Rheologica Acta 42 (2003)295e308.

[33] A.M.D. Santos, C.A. C�aceres, L.S. Calixto, L. Zborowski,S.V. Canevarolo, In-line optical techniques to characterize thepolymer extrusion, Polym. Eng. Sci. 54 (2013) 386e395.

[34] T. Matsumoto, D.C. Bogue, Stress birefringence in amorphouspolymers under nonisothermal conditions, J. Polym. Sci. Pol Phys. 15(1977) 1663e1674.

[35] W. Kozicki, C.H. Chou, C. Tiu, Non-newtonian flow in ducts ofarbitrary cross-sectional shape, Chem. Eng. Sci. 21 (1966) 665.

[36] J.A. Wheeler, E.H. Wissler, Friction factor-Reynolds number relationfor steady flow of pseudoplastic fluids through rectangular ducts. I.theory, Aiche J. 11 (1965) 207.

[37] Y. Son, Determination of shear viscosity and shear rate from pres-sure drop and flow rate relationship in a rectangular channel,Polymer 48 (2007) 632e637.

[38] J.P. Hartnett, M. Kostic, Heat transfer to Newtonian and non-Newtonian fluids in rectangular ducts, in: P.H. James, F.I. Thomas(Eds.), Advances in Heat Transfer, Elsevier, 1989, pp. 247e356.