implantable sensor to measure liquor pressure

2
The human brain is filled with cerebrospinal fluid (CSF), which is produced in the ventric- les. Patients suffering from normal pressure hydrocephalus (NPH) produce more CSF as they resorb. This leads to an increase of the intracranial pressure (ICP) in the skull, what may cause progressive enlargement of the head, convulsion, tunnel vision and mental disability due to shortage of oxygen and nut- rients. To reduce the intracranial pressure, the implantation of a drainage system (shunt) which removes excessive CSF, e.g. into the abdominal cavity, is necessary. Development In corporation with the companies Aesculap AG and Christoph Miethke GmbH & Co KG a system was developed for wireless measu- rements of the cerebrospinal fluid pressure in the shunt. The system is approved for use as a long-term implant in humans. IMPLANTABLE SENSOR TO MEASURE LIQUOR PRESSURE OF A VENTRICULAR DRAINAGE SYSTEM Pressure Sensor and Sensor Readout Electronics The integrated capacitive pressure sensor of the transponder ASIC, as shown in Figure 3, consists of an array of circular pressure sensi- tive and pressure insensitive elements. The capacitor of a pressure sensor is formed by a fixed electrode in the substrate and a second electrode as a deflectable membrane of poly- crystalline silicon above. The cavity under the membrane is obtained by anisotropic etching and later vacuum sealing. The diameter of one element is approx. 100µm and the capa- citive change between minimal and maximal pressure is approximately 50fF. A differential C/V converter transforms the capacitance of the pressure sensor into a voltage signal. This signal is A/D converted by a RSD (Redundant Signed Digit) cyclic A/D converter into a digital 13 × 2 bit data stream. 1 Shunt sensor position in the drainage system. 2 Handheld reader unit measuring. FRAUNHOFER INSTITUTE FOR MICROELECTRONIC CIRCUITS AND SYSTEMS IMS Fraunhofer Institute for Microelectronic Circuits and Systems IMS Finkenstr. 61 D - 47057 Duisburg Phone +49 203 37 83-0 Fax +49 203 37 83-266 www.ims.fraunhofer.de Contact Michael Bollerott Phone +49 203 37 83-227 [email protected]

Upload: others

Post on 18-Dec-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

The human brain is filled with cerebrospinal

fluid (CSF), which is produced in the ventric-

les. Patients suffering from normal pressure

hydrocephalus (NPH) produce more CSF as

they resorb. This leads to an increase of the

intracranial pressure (ICP) in the skull, what

may cause progressive enlargement of the

head, convulsion, tunnel vision and mental

disability due to shortage of oxygen and nut-

rients. To reduce the intracranial pressure, the

implantation of a drainage system (shunt)

which removes excessive CSF, e.g. into the

abdominal cavity, is necessary.

Development

In corporation with the companies Aesculap

AG and Christoph Miethke GmbH & Co KG

a system was developed for wireless measu-

rements of the cerebrospinal fluid pressure in

the shunt. The system is approved for use as

a long-term implant in humans.

Implantable SenSor to meaSure lIquor preSSure of a VentrIcular DraInage SyStem

Pressure Sensor and Sensor

Readout Electronics

The integrated capacitive pressure sensor of

the transponder ASIC, as shown in Figure 3,

consists of an array of circular pressure sensi-

tive and pressure insensitive elements. The

capacitor of a pressure sensor is formed by a

fixed electrode in the substrate and a second

electrode as a deflectable membrane of poly-

crystalline silicon above. The cavity under the

membrane is obtained by anisotropic etching

and later vacuum sealing. The diameter of

one element is approx. 100µm and the capa-

citive change between minimal and maximal

pressure is approximately 50fF.

A differential C/V converter transforms the

capacitance of the pressure sensor into a

voltage signal. This signal is A/D converted by

a RSD (Redundant Signed Digit) cyclic A/D

converter into a digital 13 × 2 bit data stream.

1 Shunt sensor position in the drainage system.

2 Handheld reader unit measuring.

F R A U N H O F E R I N S T I T U T E F O R M I c R O E l E c T R O N I c c I R c U I T S A N D S y S T E M S I M S

Fraunhofer Institute for

Microelectronic circuits

and Systems IMS

Finkenstr. 61

D - 47057 Duisburg

Phone +49 203 37 83-0

Fax +49 203 37 83-266

www.ims.fraunhofer.de

Contact

Michael Bollerott

Phone +49 203 37 83-227

[email protected]

3 Pressure and temperature transponder ASIC.

4 Typical raw pressure curve of a complete in

metal encapsulated sensor.

The cyclic conversion algorithm is based on

the conventional restoring numerical division

principle. The conversion of the capacitance

is adjustable in offset and gain to compensate

fabrication related tolerances of the pressure

sensor parameters. The C/V converter and

the A/D converter are realized in a switched

capacitor technique. A typical pressure curve

for raw sensor data is shown in Figure 4.

Transponder System

The complete transponder system, as shown

in Figure 2, consists of a passive transponder

equipped with the described pressure sensor

and readout electronics, and a reader device

which powers the transponder by an electro-

magnetic field and receives back information

from the transponder. In this transponder

system, load-modulation is used to transmit

data from the transponder to the reader. The

transmitted data package comprises of the

sensor readout data, calibration data, a

unique identification number and a checks-

um for data validation. The reader unit calcu-

lates the measured pressure and temperature

out of the received data by a fixed formula.

To overcome the damping caused by the

metal encapsulation of the implant the hand-

held reader was designed with a high quality

factor of the transmitter coil, to optimize the

performance regarding reading distance and

battery lifetime.

Note: The illustrations and information pre-

sented herein are not the specifications of

a medical product but examples of a Fraun-

hofer IMS pressure sensor system.

Parameter Value

Sensor transponder ASIC size (l, w, h) 9.4 mm, 2.5 mm, 0.73 mm

Pressure range 733 – 1233 hPa

Pressure sensor accuracy 2.66 hPa

Pressure sensor sample rate 44 Hz

Temperature range 20 – 45 °C

Temperature sensor accuracy 0.5 K

Temperature sensor sample rate 22 Hz

Operation distance 0 – 5 cm

Reader HF-frequency 133 kHz