Implantable Cardiac Pulse Generators:
Pacemakers&
In The Name of God
&Implantable Cardiac Defibrillators (ICDs)
Omid Aghadavoudi, MDAssociate Professor of Anesthesiology
![Page 2: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/2.jpg)
Synonyms
• Cardiac Implantable Electronic Device(CIED)
• Cardiac Rhythm Management Device(CRMD)( )
![Page 3: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/3.jpg)
Pacemakers
• Overview• Pacemaker Codes• Pacemaker Indications• Pacemaker Magnets• Preanesthesia Evaluation
and Pacemaker• Reprogramming
Implantable Cardioverter-Defibrillators (ICD)
• Overview• Indications• Magnets• Preanesthesia Evaluation
and Reprogramming• Intraoperative or• Reprogramming
• Intraoperative orProcedural Managementof pacemakers
• Pacemaker Failure• Postanesthesia
Pacemaker Evaluation
pProcedural Management of
• PostanesthesiaImplantable Cardioverter –Defibrillator Evaluation
![Page 4: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/4.jpg)
Transcutaneous pacemaker
![Page 5: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/5.jpg)
Miniature temporary cardiac pacemaker
![Page 6: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/6.jpg)
Internal Pacemaker
![Page 7: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/7.jpg)
History
• Pacemaker � In 1958 by CW Lillehei (a cardiothoracicsurgeon) and Earl Bakken (an electrical technician)
• Implantable pacemaker� A few years later by WilsonGreatbatch, a cardiologist
• lCD � 1980 by Michael Mirowski (Baltimore, MD).• First approved by the U.S. FDA � in 1985, required a
thoracotomythoracotomy
![Page 8: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/8.jpg)
Overview
� The complexity of pacemakers and ICDs�Multitude of programmable parameters�Population aging�Enhancements in implantable technology�Enhancements in implantable technology�New indications for implantation�Patients with a pulse generator (PG) often
have significant comorbid diseases
![Page 9: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/9.jpg)
Not all electronic generators implanted in the chest arecardiac devices
• Pulse generator implantation has been approved bythe FDA for:
1. Pain control2. Thalamic stimulation to control Parkinson's disease3. Phrenic nerve stimulation to stimulate the diaphragm in3. Phrenic nerve stimulation to stimulate the diaphragm in
paralyzed patients4. Vagus nerve stimulation to control epilepsy• Vagus nerve stimulation for heart failure & obesity
![Page 10: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/10.jpg)
Overview
• Pacemaker manufacturers: 26 companies• more than 2,500 models to date.• Currently, more than 250,000 adults and children
in the United States undergo new pacemakerl t hplacement each year.
• Nearly 2 million patients have pacemakers today.
![Page 11: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/11.jpg)
![Page 12: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/12.jpg)
Transvenous endocardial VVIpacemaker
![Page 13: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/13.jpg)
A: conventional single lead pacemakerB: implantable cardioverter-defibrillator (ICD)
![Page 14: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/14.jpg)
Radiographic identifiers for somegenerator manufacturers
![Page 15: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/15.jpg)
Pacemaker Indications
• Symptomatic diseases of impulse formation(sinus node disease)
• Symptomatic diseases of impulse conduction(disease of the atrioventricular node)
• Long QT syndrome• Hypertrophic obstructive cardiomyopathy• Dilated cardiomyopathy
![Page 16: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/16.jpg)
COMMON USES FOR PERMANENT PACEMAKER
![Page 17: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/17.jpg)
Pacemaker System
• Impulse generator• Leads
– Transvenous– Epicardial
• Leadsunipolar (one– unipolar (oneelectrode per lead),
– bipolar (twoelectrodes per lead)
– multipolar (multipleelectrodes)
![Page 18: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/18.jpg)
Examples of single-chamber and dual-chamberpacemakers with both unipolar and bipolar leads.
![Page 19: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/19.jpg)
![Page 20: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/20.jpg)
Pacemakers
• Unipolar leads– The second electrode will be the metal generator case– Requires that the generator pocket be devoid of gas– Electrical continuity has reportedly been disrupted by the
use of nitrous oxide.– More sensitive to the effects of electromagnetic
interference (EMI)– Produce larger "spikes" on an analogue recorded
electrocardiogram.• Bipolar pacing
– usually requires less energy– more resistant to interference from muscle artifacts or
stray electromagnetic fields.– Often can be identified on the chest film since they will
have a ring electrode 1 to 3 cm proximal to the lead tip.
![Page 21: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/21.jpg)
![Page 22: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/22.jpg)
pacemaker wire
![Page 23: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/23.jpg)
The pacing lead system in situ
![Page 24: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/24.jpg)
Excitation threshold & Spike
![Page 25: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/25.jpg)
NASPE/BPEG Revised (2002) GenericPacemaker Code (NBG)
Position I: PacingChamber(s)
Position II:SensingChamber(s)
Position III:Response(s) toSensing
Position IV:Programmability
Position V:Multisite Pacing
O= none O= none O= none O= none O= none
A= atrium A= atrium I= inhibitedR= ratemodulation
A= atrium
V= ventricle V= ventricle T= triggered V= ventricleD= dual (A + V) D= dual (A + V) D= dual (T + I) D= dual (A + V)
North American Society of Pacing and Electrophysiology (NASPE) nowcalled Heart Rhythm Society (HRS)
British Pacing and Electrophysiology Group (BPEG)
The NBG code: NASPE and BPEG Generic code
![Page 26: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/26.jpg)
Pacemaker sensing
• VAT• AAI
![Page 27: Implantable Cardiac Pulse Generators and ICD(1).ppt](https://reader031.vdocuments.mx/reader031/viewer/2022012013/6158765ca490961be231f7a7/html5/thumbnails/27.jpg)
Position III NBG Code• D (Dual):
– DDD pacing provides atrioventricular (AV) synchrony. In the DDD setting,atrial pacing will take place in the “inhibited” mode; that is, the pacingdevice will emit an atrial pulse if no sensed atrial event (or interveningventricular event because any ventricular event will reset atrial timing)takes place within the appropriate timeframe. In DDD or VDD devices,once an atrial event has occurred (whether native or paced), the pacingdevice will ensure that a ventricular event follows (up to the uppertracking rate [UTR]).tracking rate [UTR]).
• I (Inhibited):– The appropriate chamber is paced unless intrinsic electrical activity is
detected during the pacing interval. For the DDI mode, AV synchrony isprovided only when the atrium is paced. If intrinsic atrial activity ispresent, no AV synchrony is provided by the pacing function.
• T (Triggered):– The pacing device will emit a pulse only in response to a sensed event.
The triggered mode is used when the device is being tested.