how to choose the right gauge block comparator

14
1 GAUGE BLOCK CALIBRATION BY MECHANICAL COMPARISON (GAUGE BLOCK COMPARATOR) Arrangement by Saraswanto Abduljabbar, Valunteer, specialist and advisor Length Metrology Measurement 1. BACK GROUND International standard ISO 3650 cover the range of accuracy requirements along the traceability chain and calibration method is selected according to accuracy require- ments of the standard and the user. In mechanical comparison methods, the similar nominal size gauge blocks are compared to each other by suitable probing elements. Since the compared gauge blocks are in the same nominal sizes, the inductive probes which have a short measurement range with high accuracy is used in the mechanical comparison technique, figure 2. Factors influencing the measurement are: the length calibration of the standard, factors inherent in the comparator equipment used to measure the length difference such as scale linearity and reading capability, gauge geometry with respect to its effect on probing the length difference, the temperature and other environmental factor, etc. 2 Basis of measurement, traceability, reference condition 1 2.1 Unit of length: metre The metre is defined as the length of the path travelled by light in vacuum in 1/299 792 458 of a second (17th General Conference of Weights and Measures, 1983). The definition is realized by working wavelength standards recommended by the Interna- tional Committee of Weights and Measures (CIPM). 2.2 Traceability of the length of a gauge block The measured length of a gauge block is traceable to national or international length standards, if the measurement result can be related by an unbroken chain of compari- son measurements each with stated uncertainties to a gauge block which has been calibrated by interferometry/using appropriate wavelength standards. 1 EN ISO 3650:1998, Geometrical Product Specification (GPS) - Length Standards – Gauge Blocks, Second Edition, 1998-12-15.

Upload: saraswanto-abduljabbar

Post on 17-Jul-2015

118 views

Category:

Engineering


0 download

TRANSCRIPT

  • 1

    GAUGE BLOCK CALIBRATIONBY

    MECHANICAL COMPARISON(GAUGE BLOCK COMPARATOR)

    Arrangement bySaraswanto Abduljabbar,

    Valunteer, specialist and advisor Length Metrology Measurement

    1. BACK GROUND

    International standard ISO 3650 cover the range of accuracy requirements along

    the traceability chain and calibration method is selected according to accuracy require-

    ments of the standard and the user. In mechanical comparison methods, the similar

    nominal size gauge blocks are compared to each other by suitable probing elements.

    Since the compared gauge blocks are in the same nominal sizes, the inductive probes

    which have a short measurement range with high accuracy is used in the mechanical

    comparison technique, figure 2.

    Factors influencing the measurement are: the length calibration of the standard, factors

    inherent in the comparator equipment used to measure the length difference such as

    scale linearity and reading capability, gauge geometry with respect to its effect on

    probing the length difference, the temperature and other environmental factor, etc.

    2 Basis of measurement, traceability, reference condition1

    2.1 Unit of length: metre

    The metre is defined as the length of the path travelled by light in vacuum in 1/299 792

    458 of a second (17th General Conference of Weights and Measures, 1983). The

    definition is realized by working wavelength standards recommended by the Interna-

    tional Committee of Weights and Measures (CIPM).

    2.2 Traceability of the length of a gauge block

    The measured length of a gauge block is traceable to national or international length

    standards, if the measurement result can be related by an unbroken chain of compari-

    son measurements each with stated uncertainties to a gauge block which has been

    calibrated by interferometry/using appropriate wavelength standards.

    1EN ISO 3650:1998, Geometrical Product Specification (GPS) - Length Standards Gauge Blocks, Second Edition,

    1998-12-15.

  • 2

    2.3 Reference temperature and standard pressure.

    The nominal length and the measured lengths of a gauge block apply at the reference

    temperature of 20 C (see ISO 1) and the standard pressure 101 325 Pa = 1,013 25

    bar2.

    NOTE: The effect on the length of a gauge block caused by deviations from the standard pressure may

    be ignored under normal atmospheric conditions.

    2.4 Reference orientation of gauge blocks

    The length of a gauge block up to and including 100 mm nominal length refers to the

    vertical orientation with the measuring faces horizontal. The length of a gauge block

    over 100 mm nominal length refers to the horizontal orientation with the block supported

    on one of the narrow side faces, without additional stress, by suitable supports each at

    a distance of 0.211 times the nominal length from the ends. When such a gauge block

    is measured by interferometry in horizontal orientation, the weight of the auxiliary plate

    wrung to one of the measuring faces shall be compensated.

    2.5 Dimensional stability

    The maximum permissible changes in length per year of gauge blocks are stated in

    Tabel 1. They apply when the gauge blocks are not exposed to exceptional tempera-

    tures, vibrations, shocks, magnetic fields or mechanical forces.

    Tabel 1 Dimensional Stability

    Grade Maximum permissible cahnge in length per yearK0

    (0.02 + 0.25 10 )

    12

    (0.05 + 0.5 10 )

    NOTE : is expressed in millimeters

    Table 2 Flatness3 Tolerances

    Nominal Length, mm

    Flatness tolerance,

    K 0 1 20.5 150 0.05 0.1 0.15 0.25150 500 0.1 0.15 0.18 0.25500 1000 0.15 1.18 0.2 0.25

    2ISO 1:1975, Standard reference temperature for industrial length measurements.

    3ISO 1101:, Geometrical Product Specifications (GPS) Geometrical tolerancing Generalities,

    definitions, symbols, indication on drawings

  • 3

    Table 3 Limit deviation , of the length at any point of the measuring face from

    and tolerance, , for the variation in length.

    Nominallength,

    Calibration grade K Grade 0 Grade 1 Grade 2limit

    deviationof length

    at anypointfrom

    nominallength

    tolerancefor the

    variationin length

    limitdeviationof length

    at anypointfrom

    nominallength

    tolerancefor the

    variationin length

    limitdeviationof length

    at anypointfrom

    nominallength

    tolerancefor the

    variationin length

    limitdeviationof length

    at anypointfrom

    nominallength

    tolerancefor the

    variationin length

    0.5 1010 2525 5050 75

    0.20.30.40.5

    0.050.050.060.06

    0.120.140.2

    0.25

    0.10.10.1

    0.12

    0.20.30.40.5

    0.160.160.180.18

    0.450.60.81

    0.30.30.3

    0.35

    75 100100 150150 200

    0.60.81

    0.070.080.09

    0.30.40.5

    0.120.140.16

    0.60.81

    0.20.2

    0.25

    1.21.62

    0.350.40.4

    200 250250 300300 400

    1.21.41.8

    0.10.1

    0.12

    0.60.70.9

    0.160.180.2

    1.21.41.8

    0.250.250.3

    2.42.83.6

    0.450.50.5

    400 500500 600600 700

    2.22.63

    0.140.160.18

    1.11.31.5

    0.250.25

    3

    2.22.63

    0.350.4

    0.45

    4.456

    0.60.70.7

    700 800800 900

    900 1000

    3.43.84.2

    0.20.2

    0.25

    1.71.92

    0.30.350.4

    3.43.84.2

    0.50.50.6

    6.57.58

    0.80.91

    3. Measurement by comparison

    3.1 Principle of measurement

    In order to determine the length of a gauge block by compa-

    rison, the difference of its central length from that of a reference

    gauge block is measured and applied algebraically to the length

    of the reference. For the probing, the measuring faces of each

    gauge are touched from opposite directions as shown in Figure

    1, and the length difference is measured by a high resolution

    length indicator.

    algebraically difference (see arrow point 1) = = Measured length

    = Nominal length

    Fig. 1 Measurement of central length by comparison taking the perpendicular

    distance from the centre of a measuring face to the opposite one

  • 4

    3.2 Central length

    A measurement by comparison transfers the central length of a standard gauge block to

    a gauge block under test. The reference gauge block may either directly be measured

    by interferometry or related through one or several stages by comparison to a reference

    gauge measured by interferometry.

    NOTE:

    The effect of one wringing, which is included in the length of the reference gauge block

    measured by interferometry, is transferred by the comparison measurement.

    3.3 Method of determining length by comparison

    The relatively small difference in central length between a reference gauge block of

    known central length and another gauge of unknown central length is measured by a

    high resolution length indicator. (see Figure 2).

    Figure 2 Measurement of central length by comparison taking the perpendicular distance from the

    centre of a measuring face to the opposite one. Schematic diagram of a gauge block comparator with

    Measured Length Difference

    Reading ()

    Calibration ()

    Drift ()

    Temperature

    Reading ()

    Calibration ()

    Drift ()

    = 20

    PenetrationFlatness

    = = +

    InductiveProbe 1

    InductiveProbe 2

    Parallelism

  • 5

    opposing styli and digital readout. The diagram depicts the influence factors in the measurement, and the

    mathematical symbols which will represent them in the text.

    Figure 2 shows the reference gauge block in the reading position between an upper

    contact and one beneath. The anvils are retractable and the weight of the block is

    supported independently. The connection line between the two anvils is perpendicular

    to the measuring faces. A reading of the indicator is taken at the centre of the reference

    gauge block which is then replaced by the gauge block to be measured and a central

    reading is taken on it. The vertical position is used for comparison of gauge blocks of

    nominal lengths up to 100 mm; Gauge blocks of nominal lengths longer than 100 mm

    can also be measured by comparison with a reference gauge block in the horizontal

    position. If a vertical orientation as specified in fig.2 is used, the supports are adjusted

    horizontally and vertically so that one anvil of the comparator contacts the centre of one

    measuring face of the gauge block and the second anvil is moved over the second face

    until the minimum reading is obtained.

    4. HOW CHOOSING THE RIGHT GAUGE BLOCK COMPARATOR

    For mainly economic reasons, most of customers of accredited laboratories prefer their

    standards being calibrated by mechanical comparison because of lower costs and

    shorter calibration time. In the market there are many products of the Gauge Block

    comparators. All of them of course have advantages and disadvantages. This white

    paper will guide you how to excecute the right Gauge Block Comparators.

    As you know many comparisons, especially those in dimensional metrology, cannot be

    done simultaneously. Using a gage block comparator, the standard, control (check

    standard) and test block are moved one at a time under the measurement stylus. For

    those comparisons each measurement is made at different time. The term calibration

    design can be applied to experiments where only differences between nominally equal

    objects or groups of objects can be measured.

    Ordinarily the order in which these measurements are made is of no consequence. The

    usefulness of drift eliminating designs depends on:

    1. The stability of the thermal environment, accuracy required in the calibration. The

    environment has to be stable enough that the drift is linear.

    2. Each measurement must be made in the same amount of time so that the

    measurements are made at fairly regular intervals. Finally, the measurements of

    each block are spread evenly as possible across the design. The designs are

    constructed to:

    Be immune to linear drift.

    Minimize the standard deviations for test blocks (as much as possible)

  • 6

    Spread the measurements on each block throughout the design

    Be completed in 5-10 minutes to keep the drift at the 5 nm level

    4.1 Critical contributions to the uncertainty

    Critical contributions, which could be dimi-

    nished, are temperature deviations, calibra-

    tion of the comparator and calibration of

    reference gauge blocks. Temperature devi-

    ations are influenced by the air temperature

    deviations and by radiation of illuminating

    bodies and of the operator.

    Temperature can not be measured

    on the gauge block and therefore table

    temperature is assumed to be the tempe-

    rature of the gauge block. However, devia-

    tions between these two temperatures can differ for up to 0.05C. Additional problems

    appear by longer gauge blocks (near 100 mm) because temperature is not constant

    along the height. Beside that, thermal expansion coefficient of gauge blocks is usually

    not exactly known. It may differ for 1 10. Thermal expansion correction is

    therefore not accurate.

    These problem happened because difficult to know what is excactly the CTE4 of

    the material gauge block. According to Ted Doiron and John Beers 5, The uncertainty in

    the expansion coefficient of the gauge block is more difficult to estimate.

    Table 4: shows that as the blocks get longer, the thermal expansion coefficient becomes systematically

    smaller. It also shows that the differences between blocks of the same size can be as large as a few

    percent. Because of these variations, it is important to use long length standards as near to 20 C as

    possible to eliminate uncertainties due to the variation in the expansion coefficient.

    4CTE (Coefficient Thermal Expansion).

    5Gage Block Hand Book, Dimensional Metrology Group Precision Engineering Division, National Institute of

    Standards and Technology (NIST-USA).

  • 7

    4.2 UNCERTAINTY BUDGET AND CRITICAL CONTRIBUTIONS

    In order to establish proper environmental conditions, special microclimatic chamber

    with own precise air conditioning system was built in the laboratory. Temperature

    deviations in the measuring space of gauge block comparator are limited to 0,1C.

    Temperature is the most important influence factor on measuring uncertainty in

    calibration beside the accuracy of calibration of reference gauge blocks. The expanded

    uncertainty of the calibration of gauge blocks by mechanical comparison in the machine

    is currently (0.03 + L/3000) m, L in mm*. Using coverage factor k = 2. It was

    calculated according to guide ISO GUM 2 edition, 1995 and comprises all important

    components. Some of them were evaluated statistically by repeated measurements,

    some of them were evaluated by experience and some of them on the basis of different

    data from manuals and calibration certificates. All the measurements including tempera-

    ture measurements were performed with traceable equipment and measuring uncer-

    tainties were evaluated for all measurements.

    5. STRATEGY FOR DECREASING CURRENT LEVEL OF

    UNCERTAINTY 6

    5.1 Study of temperature influences

    The following temperature parameters have been analysed:

    change of gauge block temperature after putting it on the comparator,

    temperature difference between the gauge blocks and the comparator table,

    temperature influence of the illumination and the operator, and

    thermal expansion coefficients of gauge blocks.

    All the above parameters were detected as critical and are

    therefore subject of optimisation. Research was focused into

    experimental simulations of different materials and colours of

    these materials. It was found out that the radiation of different

    materials is quite different and that calibration results are critically

    influenced by radiation. The first resource of radiation are lights

    producing constant radiation that can be eliminated from calibra-

    tion result by calculation. The second source of radiation is the

    operator. This source is not constant and is quite critical because

    the gauge block that is closer to the operator is heated stronger

    that the gauge block standing behind the first one. Shield that is

    put between the operator and the comparator is made of

    transparent plastics in order to enable the operator to look into measuring area.

    6B. Acko, A. Sostar and A. Gusel, REDUCTION OF UNCERTAINTY IN CALIBRATION OF GAUGE BLOCKS

    Laboratory for Production Measurement, Faculty of Mechanical Engineering, University of Maribor, 2000, Maribor,Slovenia.

  • 8

    However, the isolating properties of this material are not good enough to stop the

    radiation of a human body.

    5.2 Surfaces for thermal stabilisation of the gauge blocks

    Surfaces for thermal stabilisation of gauge blocks have a great influence on

    thermal behaviour of gauge blocks during the calibration. If these surfaces get other

    temperature than the air and the comparator table, the gauge block temperature

    changes rapidly. Our analyses have shown that an aluminium plate with bright shining

    or light grey surface gives the best results. Long term temperature measurements are

    shown in Figure 3 as an example. The white plastic surfaces used before these studies

    are always warmer than the air and cause additional heating of gauge blocks. The

    same properties have steel surfaces. The stone that has been used in the past has the

    worst characteristics and is not recommended at all.

    In this research four materials (aluminium, steel, stone and plastic) in four different

    colours (black, white, grey and shining) were tested. Temperature was measured in the

    air, on the tested surface and on the gauge blocks, that were put on these surfaces.

    Long term and short term deviations were measured with the light turned on and with

    the light turned off in the measurement room. When the light was turned off, the

    differences were not so critical because the radiation was very low. In contrary, when

    the lights were turned on, big differences in temperature changes were observed. The

    worst behaviour was detected at dark steel and stone surfaces.

    Figure 3. Temperature comparison between the aluminium base plate and the gauge

    block.

    5.3 Material of the comparator table

    Some simulations with different materials have shown the importance of the

    material and the colour of the table. The existing tables (tested on Mahr and Cary

  • 9

    comparators) do not have ideal properties because of dark colours. The temperature is

    constantly above the air temperature. When gauge block is put on the table, the

    temperature starts to grow and the length of the gauge block is changing.

    Typical temperature behaviour after putting gauge block on the table is shown in figure

    3. In this case gauge block was put on the comparator table at 10:41 and at 10:50

    calibration of this gauge block was finished. From this case we can learn, that the

    gauge block should not be measured immediately after putting it on the table. It should

    be stabilised for approximately three minutes. Probe 1 in Figure 3 was measuring the

    reference gauge block temperature and probe 2 was measuring the temperature of

    measured gauge block. We can see that the temperatures of both gauge blocks after

    stabilisation are not equal.

    5.4 Thermal expansion coefficient of the gauge blocks

    Thermal expansion coefficient must be determined with standard uncertainty

    0.3 10. For this we shall determine thermal expansion coefficient within the

    limits of 0.5 10. Usually this is not a problem, but sometimes we get gauge

    blocks without any data about . In such cases we can not be sure that was

    determined with proper accuracy. Temperature deviations must therefore be dimini-

    shed to minimum possible values.

    Another problem is the difference between thermal expansion coefficients of the

    reference and measured gauge blocks. Uncertainty contribution of his difference in mm

    is expressed by the equation:

    = ()

    Where:

    - uncertainty contribution in mm() - uncertainty of the difference of thermal expansion coefficients ( ( - nominal length of gauge blocks - maximum possible temperature deviation

    If uncertainty of the difference between thermal expansion coefficients is 0.6 10

    and maximum possible temperature deviation is 0.1C, than the uncertainty contribution

    in calibration of 100 mm gauge block is 0.006 m what is the maximum possible limit for

    our demands. If temperature deviation is 0.3C, we get the value of 0.018 m, what is

    far too much.

    From the above example we can conclude that temperature deviations and the

    limits of determination of thermal expansion coefficient must be within very low values.

    These two components are in strong correlation and influence the length dependent

    part of the combined uncertainty.

  • 10

    Figure 4. Thermal behaviour of gauge blocks after putting them on the comparator

    table

    5.5 Indentation of different gauge blocks by the applied measuring force

    Research in this field is made in order to be able to calibrate gauge blocks of

    different materials. The indentation is calculated by the following (Roark's) equation:

    = 1.4

    =1

    +1

    Where:

    - indentation - probing force - probing ball diameter - Poisson number of the probing ball material - Poisson number of the gauge block material - elasticity module of the probing ball material - elasticity module of the gauge block material

    If probing force is 0,1 N, probing ball of d = 2 mm is made of hart metal and gauge block

    is made of steel, than the indentation is 0.116 m. For the gauge block made of hart

  • 11

    metal this value would be 0.081 m. The difference of the both indentations is 0.035

    m. The greatest problem in this calculations is to get right values for and . Every

    deviation of these values causes uncertainty of indentation calculation. The second

    problem is evaluation of probing force, which shall be measured very precisely. There-

    fore, some experimental work has already been performed in order to evaluate uncer-

    tainty of indentation. It was found out that the standard uncertainty can be held in the

    limits of 0.008 m.

    6. DRIFT ELIMINATING DESIGN

    The sources of variation in measurements are numerous. Measurements on gauge

    blocks are subject to drift from heat built-up in the comparator. This effect cannot be

    minimized by additional measurement because it is not generally pseudorandom, but a

    nearly monotonic shift in the readings. In dimensional work the most important cause of

    drift is thermal changes in the equipment during the test.

    The purpose of drift eliminating design is remove the effects of linear instrumental drift,

    but also alowing the measurement of the linear drift itself. This measured drift can be

    used as a process control parameter. For small drift rates an assumption of linear drift

    will certainly be adequate. But, for high drift rates or long measurements time the

    assumption of linear drift may not be true.

    Measurements on gauge blocks are subject to drift from heat build-up in the compa-

    rator. This drift must be accounted for in the calibration experiment or the lengths

    assigned to the blocks will be contaminated by the drift term. The designs are

    constructed so that the solutions are immune to linear drift if the measurements are

    equally spaced over time. The size of the drift is the average of the difference

    measurements. Keeping track of drift from design to design is useful because a marked

    change from its usual range of values may indicate a problem with the measurement

    system.

    7. OVERVIEW OF THE GAUGE BLOCK COMPARATOR PRECIMAR 826 VS

    TESA UPG.

    PRECIMAR 826 TESA UPD TESA UPCDESIGN: rigid cast-iron stand with vertical

    guiderigid cast-iron stand with verticalguide

    Measuring range 0.5 mm to 170 mm 0.5 mm up to 100 mmWeight 37 Kg (comparator only) 23 kg (comparator only)NOTE:Heavier Unit more steady (immune from vibration)Repeatability (with no influenceof external temperature)

    0.01 m 0.04 m

    Measuring uncertainty (.+ )/ , = (. + . ( For instance Gauge Block 100 For instance Gauge Block 100

  • 12

    mm, = (0.03 + 100/3000) = 0.03 + 0.033

    = 0.063

    mm, = (0.07 + 0.5 0.1) = 0.07 + 0.05

    = .NOTE:Following the hierarchy of calibration, calibrator must be 10% or 20% smaller then tolerance of theworking standard. For example, Gauge Block tolerance 0.12 m (Grade 0, length 100 mm), must becalibrated by Gauge Block Comparator with Uncertainty minimum 0.12 x 0.2 = 0.024 m. Following thisreason, methode direct measurement is not acceptable. In the comparison method the stability of theinstrument is 0.02 + 0.25 10 ( Precimar 826).Repeatability (with no influenceof external temperature)

    0.01 m 0.025 m

    Measuring uncertainty (.+ )/ , = (. + . ( For instance Gauge Block 100mm, = (0.03 + 100/3000)

    = 0.03 + 0.033= 0.063

    For instance Gauge Block 100mm, = (0.05 + 0.5 0.1)

    = 0.05 + 0.05= .

    Repeatability (with no influenceof external temperature)

    0.01 m 0.015 m

    Measuring uncertainty (.+ )/ , = (. + . ( For instance Gauge Block 100mm, = (0.03 + 100/3000)

    = 0.03 + 0.033= .

    For instance Gauge Block 100mm, = (0.02 + 0.2 0.1)

    = 0.02 + 0.02= .

    NOTE:Following the Gauge Block tolerance in the Table 3, UPC and UPD can calibrate Gauge Block Grade 1and 2 only. Precimar 826 can calibrate Gauge Block Grade 0, 1 and 2. Only TESA with= (0.02 + 0.2 ( can be used for Grade 0, but this instrument still struggle to get the uncertaintybecause it will depends on the temperature correction. You know that in the paragraph 4.1 explainedThermal expansion correction is therefore not accurate. So that is why the Comparator need the deviationof the temperatur is 0.03C.Material Housing Diecast Iron Diecast AluminumSensors for length dimensions:

    Double Probes Yes Yes Yes

    NOTE:The measuring configuration consisting of two probes aligned opposite one another combined with theconcept and quality of the measuring system is the guarantee for an extra low measurement uncertainty.

    With analog and digitalindication system,

    Yes Digital Only Analog Only

    Measuring force activation Measuring bolt retraction

    YesBy electrical vacuum pum lifted

    YesBy Vacuum

    YesBy electro-motorised

    Measurement Method Comparative Measurement DirectMeasurement

    ComparativeMeasurement

    Advantage:Enhances the measuring condi-tions, thus permitting all measure-ments to be taken with loweruncertainty. NO linear error.Disadvantages:Allows gauge blocks of samenominal lengths to be measuredby comparison. Need 122 GaugeBlock set as a standard compa-rator.

    Advantage:Permits over 90 % of a 122-pieceset to be checked using the samereference gauge block All nominallengths of the full gauge set beingcontained within 0,5 and 25 mm.Disadvantages:In fact the measuring span istherefore exceeded, caused ofLinear Error of the probes.

    NOTE:For the Direct measurement method actually data will found linear error that is not an advantages for themachine due to the machine must immune to linear drift. So the choice of the Comparative methode is thebest solution.

  • 13

    PRECIMAR 826 TESA UPD TESA UPDComputer-aided calibration Measurement of single gage blocks Measurement of complete gage

    block sets Simultaneous calibration of several

    equal gage block sets.

    YesYes

    Yes

    YesYes

    Yes

    Features of the Hardware: Selection and determination of

    measuring sequences.Yes Yes

    Management of testpiece andstandard gauge block sets.

    Yes Yes

    Management of individual gaugeblocks

    Yes Yes

    Measuring program to perform gaugeblock tests

    Yes Yes

    Control of all operations and inputs Yes Yes Automatically assigning the sequenceof nominal dimensions for set tests

    Yes Yes

    Organization of the measurementprocess for testing multiple sets

    Yes Yes

    Printer program for test records andfor the printout of standard gage blocksets

    Yes Yes

    Data management Yes Yes

    ADVANTAGES: Rigid cast-iron stand, making the

    unit temperature stableYes Yes

    Easily adjustable vertical slide withupper probe

    Yes Yes

    Fine adjustment via rigidlyConnected Parallelogram springs

    Yes No

    NOTE:Fine adjustment is carried out by adjusting a rigidly connected spring parallelogram system which isintegrated into the support arm. Electro pneumatic lifting Yes Yes Gentle, precise and extremely smoothmanipulator operation due to slideways,which are impervious to dirt

    Yes Yes

    NOTE:The foot of the base holds the manipulator for correctly positioning the gage blocks. Smooth movement isensured by high precision ball bush guides. Ergonomical operation due to the

    optimally arranged gauge blockmanipulator

    Yes Yes

    NOTE:Whenever the gage blocks are moved, the inductive probes are lifted by means of an electrical vacuumpump. Measurement not influenced by

    manual forceYes Yes

    Easy movement of the gage blocksdue to support consisting ofhardened circular guide bars

    Yes Yes

    No zero setting required, since theset value is automatically related tothe nominal value of the respectivereference gage block

    Yes Yes

    Very effective protection from heat Yes Yes

  • 14

    due to an acrylic glass screen alongthe front and both sides of the unit

    Penetration correction Yes NONOTE:See paragraph 5.5, the correction of the geometric factor of the probe tip penetration. Correction of differing thermal

    expansion coefficientsYes Yes

    Computation of mean values Yes Yes Simultaneous calibration of several

    sets possibleYes Yes

    Inherent stabilization of temperatureand vibration

    Yes Yes

    CONCLUTION:

    All the concept of the Comparator must be supported for the its accuracy, with the

    requirement of the uncertainty measurement factors, such as:

    1. The design, must be immune from the errors, such as measuring base by refelcted

    material to reduce temperature influence.

    2. Comparation method, to avoid linear error of the probes.

    3. Quick measurement to avoid linear drift cause of time of measurement.

    4. Avoid themperature correction due to difference of the CTE, unless the themperature

    difference over 0.4C.

    5. Correction from the elasticity of the material effect from the contact point.

    References:

    1. ISO 3650:1999, Geometrical Product Specifications (GPS) Length standards

    Gauge blocks.

    2. ISO. GUM: 1995, Guide Uncertainty of Measurement.

    3. Ted Doiron and John Beers, Gage Block HandBook, Dimensional Metrology Group

    Precision Engineering Division National Institute of Standards and Technology.

    4. B. Acko, A. Sostar and A. Gusel, reduction of uncertainty in

    calibration of gauge blocks, Laboratory for Production Measurement Faculty of

    Mechanical Engineering University of Maribor, 2000 Maribor, Slovenia.