handbook of sample preparation for scanning electron ...€¦ · electron microscopy and x-ray...

30
Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray Microanalysis

Upload: others

Post on 14-Jun-2020

17 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray Microanalysis

Page 2: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray Microanalysis

Patrick EchlinCambridge Analytical Microscopy, UK

Page 3: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

Patrick EchlinCambridge Analytical Microscopy, [email protected]

ISBN: 978-0-387-85730-5 e-ISBN: 978-0-387-85731-2DOI: 10.1007/978-0-387-85731-2

Library of Congress Control Number: 2008942785

© Springer Science+Business Media, LLC 2009All rights reserved. This work may not be translated or copied in whole or in part without the written permission of the publisher (Springer Science+Business Media, LLC, 233 Spring Street, New York, NY 10013, USA), except for brief excerpts in connection with reviews or scholarly analysis. Use in connection with any form of information storage and retrieval, electronic adaptation, computer software, or by similar or dissimilar methodology now known or hereafter developed is forbidden.The use in this publication of trade names, trademarks, service marks, and similar terms, even if they are not identifi ed as such, is not to be taken as an expression of opinion as to whether or not they are subject to proprietary rights.While the advice and information in this book are believed to be true and accurate at the date of going to press, nei-ther the authors nor the editors nor the publisher can accept any legal responsibility for any errors or omissions that may be made. The publisher makes no warranty, express or implied, with respect to the material contained herein.

Printed on acid-free paper

springer.com

Page 4: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

For Alexander, Charles, Patrick, Francesca, Maeve, and William in the fond hope that some of them might become scientists.

I am dedicating this book to my six grandchildren and thanking a lot of people for their help and, finally for my wife's patience over the past five years.

Page 5: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

Books must follow science, and not science books.

–Francis Bacon, 1657

Page 6: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

Acknowledgments

This book would not have been possible without the help of many people. I am privileged to have been working in this field for the past 45 years and am indebted to the many friends and colleagues from all over the world who have listened to my ideas, corrected my errors, and provided practical advice. I am also grateful to the many people and manufacturers who have provided information and illustrations for this book and who are acknowledged in the text.

I am particularly grateful to my colleagues with whom I have taught at the Lehigh Microscopy School for the past 32 years. They have been a constant source of enlightenment and their constructive and critical analysis of my work and writings has been very supportive.

Finally, gratitude to Joe Michael for the use of one of his micro-graphs as a cover illustration for this book.

Patrick EchlinCambridge, January 2009

ix

Page 7: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

Contents

Acknowledgments ........................................................................ ix

Chapter 1 Introduction ............................................................ 1

Chapter 2 Sample Collection and Selection ....................... 11

Chapter 3 Sample Preparation Tools .................................... 19

Chapter 4 Sample Support ..................................................... 31

Chapter 5 Sample Embedding and Mounting ................... 47

Chapter 6 Sample Exposure ................................................... 65

Chapter 7 Sample Dehydration ............................................. 97

Chapter 8 Sample Stabilization for Imaging in the SEM............................................................... 137

Chapter 9 Sample Stabilization to Preserve Chemical Identity .................................................. 185

Chapter 10 Sample Cleaning ................................................... 235

Chapter 11 Sample Surface Charge Elimination .................. 247

Chapter 12 Sample Artifacts and Damage ............................ 299

Chapter 13 Additional Sources of Information .................... 307

References ....................................................................................... 317

Index ................................................................................................ 323

xi

Page 8: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

Scanning electron microscopy (SEM) and x-ray microanalysis can produce magnified images and in situ chemical information from virtually any type of specimen. The two instruments generally operate in a high vacuum and a very dry environment in order to produce the high energy beam of electrons needed for imaging and analysis. With a few notable exceptions, most specimens destined for study in the SEM are poor conductors and composed of beam sensitive light elements containing variable amounts of water.

In the SEM, the imaging system depends on the specimen being sufficiently electrically conductive to ensure that the bulk of the incoming electrons go to ground. The formation of the image depends on collecting the different signals that are scattered as a consequence of the high energy beam interacting with the sample.

Backscattered electrons and secondary electrons are generated within the primary beam-sample interactive volume and are the two principal signals used to form images. The backscattered electron coefficient ( η ) increases with increasing atomic number of the specimen, whereas the secondary electron coefficient ( δ ) is relatively insensitive to atomic number. This fundamental differ-ence in the two signals can have an important effect on the way samples may need to be prepared. The analytical system depends on collecting the x-ray photons that are generated within the sample as a consequence of interaction with the same high energy beam of primary electrons used to produce images.

1. The use of scanning electron microscopy and x-ray microa-nalysis may be considered under three headings.It is first necessary to understand the actual process of microscopy and analysis. This is not considered here in any detail because my colleagues and I have produced an excellent textbook that covers these processes in great detail (Goldstein et al., 2004).

1 Introduction

P. Echlin, Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray Microanalysis, DOI: 10.1007/978-0-387-85731-2_1,© Springer Science + Business Media, LLC 2009

1

Page 9: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

2 1. Introduction

2. It is necessary to consider how to prepare samples prior to microscopy and analysis. These procedures are the topic of this book.

3. It is necessary to be able to interpret the information obtained from the SEM and attempt to relate the form and structure of the two-dimensional images and the identity, validity, and location of the chemical data back to the three-dimensional sample from which the information was derived. This is a topic of continuing debate.

There are two approaches to dealing with the frequent impasse that may exist between the properties of the sample and the optimal operating conditions of the SEM. We can either modify the instruments so they employ less invasive procedures or we can modify the specimen to make it more robust to the withering beam of high energy electrons. The former approaches are dis-cussed in the book by Goldstein et al. (2004); the latter approach is considered here. With a few exceptions, both approaches are a compromise.

Sample preparation is an absolute prerequisite for microscopy and analysis.

Every specimen that goes into the SEM needs some form of sample preparation. There are no exceptions.

Consider carrying out microscopy and analysis on the compo-nents at our breakfast table. After drinking our fresh orange juice, we use a knife to butter our toast before we drink our coffee. The glass containing our juice is composed of a beam resistant, non-conducting, non-crystalline, light element solid. The orange juice and the coffee are wet, non-conducting, biopolymer composed of light element materials that are very bean sensitive. The buttered toast is a thermally and beam sensitive, non-conducting, damp, light element biopolymer. The knife is made of a beam resistant conducting metal that, in spite of being washed, is dirty. The plastic plate is made of a beam sensitive, non-conducting, light element polymer and the coffee cup is a beam resistant, non-conducting inorganic ceramic. All of these objects need some form of preparation before they may be examined properly by scanning electron microscopy and x-ray microanalysis.

The preceding example shows that there is a very wide range of specimen types. For convenience they are divided into six groups on the basis that each group has distinct characteristics, and as a consequence, may require different approaches to sample prepa-ration. This sixfold division is somewhat artificial because many specimens are composed of material from more than one of these groups. For example, a human tooth is composed of hard dry, inorganic material in a biological matrix, a car tire is a mixture of metal and polymers, and deep sea oil is a mixture of brine, mud, and organic material. The six groups are as follows.

Page 10: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

1.2 Hard, Dry, Inorganic Materials 3

1.1 Metals, Alloys, and Metallic Materials

These types of samples occur naturally either as metals or min-eral ores and as fabricated components such as nanowires and microwires, cables, sheets, girders, and complex structures such as automobiles and televisions. Metals range in atomic weight from the depleted uranium (Z = 92) component of artillery ordnance, to beryllium (Z = 4) used for specimen stubs. Metals are dry and good conductors of electricity and heat ranging from the high conductivity of silver to the low conductivity of alloys such as Nichrome. They are generally chemically stable and resistant to radiation damage in the microscope column (Figure 1.1 ).

1.2 Hard, Dry, Inorganic Materials

This important group of samples are generally chemically sta-ble but are all poor electrical conductors. The type of specimens includes ceramics, rocks, geological samples, minerals, parti-cles, microelectronic devices, optoelectronic communication systems, integrated circuits, package devices, and fibers. The group also contains a wide range of manufactured materials such as microelectronic devices, concrete, cement and other building materials, concentrated ores, and glass. Fossil bones and teeth may fall into this category provided they are devoid of significantly important moisture and/or organic material. Living or recently dead bones and teeth are better considered as

Figure 1.1. A steel rail tie from the old Colorado Midland Railroad track (1883–1918) near the Hageman Tunnel in Colorado at 3600m, 1980

Page 11: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

4 1. Introduction

biological material. Some specimens such as clays, soils, mud, and suspended particles, contain varying amounts of water that must either be removed by drying or converted at sub-zero temperatures to the solid state. This second group of specimens are generally composed of low atomic number materials (Z = 1–20) and are resistant to radiation damage in the microscope column. Provided they are dry, the samples are generally chem-ically stable but are all poor conductors (Figure 1.2 ).

1.3 Hard or Firm, Dry Natural Organic Materials

This category consists of polymers of carbohydrates, proteins, and fat derived from living organisms. Close to 99% of the chemical composition of all living material is formed from the low atomic number material (Z = 1–20) and include the major elements (H, C, N, O) and minor elements (Na, Mg, Si, P, S., Cl, K, and Ca). The remaining 1% is made up primarily of Z = 24 to Z = 30 and includes the trace elements Cr, Co, Cu, Fe, Mn, Zn, and I, which are important in various metalloenzymes and coenzymes.

The type of samples allocated to this group include wood and its derived products including timber, furniture, and paper; dried seeds and their manufactured powders, some foods, and plant fibers that form the basis of rope, string, threads, textiles, and fabrics. This category also includes hair, leather, wool, and skin, which are composed of complex polymers of proteins and solid waxes derived from fats. These are all derived materials,

Figure 1.2. Serpentine sculpture “Warming myself” by Zachariah Njobo, Zimbabwe, 1999

Page 12: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

1.5 Biological Organisms and Materials 5

and although they all are susceptible to varying degrees of hydration, chemical destruction, and biological decay, they do not readily form chemical bonds with small molecules. They are poor conductors and are susceptible to radiation damage. Some specimens are easily wetted; others are very resistant to water (Figure 1.3 ).

1.4 Synthetic Organic Polymer Materials

The primary feed stock for these synthetic polymers are natural oil, gas, and coal, which provide the source for ethylene, meth-ane, alkenes, and aromatic compounds, and which in turn pro-vide the building blocks for polymers, paints, and plastics. They are hydrocarbons composed of hydrogen and carbon (Z = 1–8) and have a wide range of everyday applications in the produc-tion of fibers, films, membranes, adhesives, elastomers, and engineering and construction polymers. Manufactured polymers also may exist as composites containing minerals, glass, and met-als. They are not readily wetted, and are usually more resistant to hydration and chemical destruction than the previous group of dry natural organic materials. As a group they are very poor conductors and are more susceptible to radiation damage than the previous group of samples (Figure 1.4 ).

1.5 Biological Organisms and Materials

These are living organisms, and they are composed primarily of light elements (Z = 1–20), which also contain a large number

Figure 1.3. Nineteenth century Kilim Prayer Rug. Kapadokya, Turkey, 1976

Page 13: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

6 1. Introduction

Figure 1.5. Polyanthus flowers. Cambridge, 2008

of trace elements. Water is an essential component of living bio-logical samples, together with complex organic macromolecules such as proteins, lipids, and carbohydrates. They are susceptible to dehydration, chemical destruction, and biological decay. They are poor conductors and very susceptible to radiation damage (Figure 1.5 ).

Figure 1.4. Plastic orange juicer

Page 14: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

1.6 Wet and Liquid Samples 7

1.6 Wet and Liquid Samples

This group of samples, although unique, is a subset of the pre-vious five samples. They are included here as a separate group because their liquids are an essential part of their form, structure and chemistry.

Wet samples are defined as specimens that do not flow and have a readably recognizable structural integrity at normal temperatures and pressures. For example, a drilled rock sample from an oil field, a fresh leaf, or freshly mixed concrete.

Liquid samples will flow and lack any readably recognizable structure at normal temperatures and pressures; for example, paint, mud, and milk (and beer).

In both types of sample the most prominent liquid is water, but anhydrous oils, emulsions, inorganic and organic solutions, and suspensions are found in the natural state of many different specimens. These are the most challenging group of samples to prepare for although there are specialist devices to directly exam-ine wet samples in the SEM, most microscopy and analysis is car-ried out on dry samples. The liquid components are exclusively composed of light elements and are very susceptible to conven-tional drying and chemical activity. The liquids are poor conduc-tors and very susceptible to radiation damage (Figure 1.6 ).

This book is designed as an introduction to sample prepara-tion and will not contain recipes for every type of specimen. The approach is to bridge the gap between the properties of the specimen and the exigencies of the SEM and provide a general

Figure 1.6. Wet yogurt and and beer, 2008

Page 15: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

8 1. Introduction

way to prepare the sample. This general approach is continually interspersed with additional more specific information to which readers are directed in the scientific literature.

It is intended to discuss in general terms, twelve different activities that are involved in sample preparation arranged in the sequence they would normally be expected to occur during sample preparation. It will be made clear where there are notable differ-ences to this sequence. For example, metals and dry inorganic samples may need repeated cleaning.

Prior to designing any sample preparation, it is necessary to clearly establish the scientific questions to be answered and whether scanning electron microscopy and x-ray microanalysis are the most appropriate instrumentations to use. At this prelimi-nary stage it is useful to assemble as much chemical and struc-tural information as possible about the specimen available from other sources. As most samples are multiphase, it is probably best either to design the preparative procedures to accommodate the most fragile component or set out different procedures for each of the different components. For example, a plastic insert in a metal rod might need two procedures and the metal and plastic knee replacement in my left leg might need three or even four methods.

The first chapter considers the collection and selection of the sample to ensure it is given a unique identity code that will follow it through the whole procedure of preparation, examina-tion, analysis, and eventual storage. It is necessary to know the dimensions of the specimen stage inside the microscope, as this will determine the size, shape, and form of the specimen to be studied.

The second chapter considers what tools and equipment are needed to prepare the specimen. These can vary from large dia-mond saws to fine forceps and from freeze dryers to toothpicks. Usually the specimen and proposed investigation dictate the type of instruments that are needed to image and analyze the sample.

The third chapter discusses the different ways samples may be supported in the microscope.

The fourth chapter considers the ways different types of sam-ples may be embedded in, or mounted on, the chosen specimen support.

The fifth chapter provides information about the different ways the region of interest in the sample may be exposed for microscopy and, if necessary, polished for x-ray microanalysis.

The sixth chapter gives details of the different processes that may be used to dehydrate the sample and ensure it is dry when it is placed in the microscope.

The seventh chapter discusses how delicate samples may be stabilized to enable them to be imaged in the microscope.

Page 16: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

1.6 Wet and Liquid Samples 9

The eighth chapter considers how any of the six types of sam-ples may be treated to preserve their chemical identity as well as their structure.

The ninth chapter describes the different ways to ensure the sample is clean and free of any materials that will either obscure the surface image and/or compromise its chemical identity.

The tenth chapter provides details of all the methods that may be used to ensure the sample is electrically conductive inside the microscope column.

The eleventh chapter addresses the problems associated with recognizing artifacts and damage that may arise during sample preparation.

The final chapter provides a compendium of additional con-tinuing oral and written information about new developments in sample preparation.

Ideally, once sample preparation is completed the specimen should be immediately placed into the microscope for imaging and analysis. If this is not possible, it should be stored in a clean, dry, secure environment for examination at a later date. All that remains now is to study the specimen—but that is another story.

Page 17: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

Before considering the different aspects of specimen preparation, it is first necessary to understand some general features of the sample and the scanning electron microscope (SEM ). This involves collecting and selecting the samples in order to establish their optimal form and dimensions for examination in a particu-lar scanning electron microscope.

2.1 Sample Collection

Usually sample collection is not a problem as investigators bring their samples to the laboratory, but it is necessary to consider the problems involved in collecting and transporting specimens to the microscope. It is important to try to collect a representative sample; for specimens 1 mm 3 or smaller, the best approach is to collect several items. Large specimens may present a problem, and the best that can be done is to take a photograph of the whole sample to show what part of it was collected. For most inert dry samples, such as metals, rocks, geological specimens, wood, paper, fabrics, seeds, and most plastic and polymer speci-mens, it is only necessary to carefully wrap the specimen in clean dry material and place it into a secure well-labeled container. Powders, soils, and fine metal filings are more conveniently collected in clean containers or plastic bottles. More delicate samples should be wrapped in several layers of fine dry tissue. At this stage, no attempt should be made to clean the specimen other than to remove any excess surface liquid.

Biological samples and wet samples present a number of prob-lems, and the aim should be to collect and maintain them in their natural state. Aquatic organisms should be kept in their natural environment, terrestrial specimens are best maintained in a damp environment, and dry material should be loosely wrapped

2 Sample Collection and Selection

P. Echlin, Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray Microanalysis, DOI: 10.1007/978-0-387-85731-2_2,© Springer Science + Business Media, LLC 2009

11

Page 18: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

12 2. Sample Collection and Selection

in soft tissue. Dissected material should be gently covered with surgical gauze. These types of samples should be kept, ventilated or aerated if necessary, in secure well-labeled containers.

Wet and liquid specimens should be maintained in their natural state in sealed containers made of either plastic or glass. Frozen samples, such as ice cream, should be collected and kept under liquid nitrogen or dry ice and brought back to the labo-ratory. One of the most remarkable low temperature collecting and transport devices was that devised by scientists at British Antarctic Survey to enable them to collect polar ice and snow and transport the intact precipitation back to their laboratory in Cambridge, England.

The key to these collection and transport procedures is to main-tain the sample in a condition as close as possible to its natural environment and speedily transport it to the microscope.

2.2 Sample Selection

It is necessary to consider the parameters that govern selecting a suitable sample for examination in the SEM. Size, shape, and proposed sample examination need to be considered.

2.2.1 Internal Dimensions of the SEM Specimen Chamber

The specimen chamber of the SEM is much larger than the speci-men region of a transmission electron microscope (TEM). Some SEM stages can accommodate and image over the whole area of an 8-in. wafer and even a hamburger, as shown in Figure 2.1 .

Figure 2.1. Picture of a hamburger. Picture courtesy of John Mansfield, University of Michigan, Ann Arbor, MI

Page 19: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

2.2 Sample Selection 13

There is one spectacular scanning microscope with a chamber big enough to hold a car tire, parts of a jet engine or even a Terracotta Soldier, as shown in Figure 2.2 . Further details of this large chamber SEM may be found in the recent paper by Hariharan et al. (2008).

The specimen stages of most scanning electron microscopes are limited to examining every point on a specimen 5–10 cm in diameter.

A number of factors limit the size of the specimen that may be examined in the SEM. The entrance to the microscope chamber must be large enough to allow the specimen to be placed on the specimen stage. For many SEM this does not present a problem because the entire specimen chamber is first vented to atmospheric

Figure 2.2. One of the Chinese Terracotta Soldiers inside the Large Chamber SEM made by VisiTech. Picture courtesy of VisiTech of America

Page 20: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

14 2. Sample Collection and Selection

pressure and the stage assembly is exposed in a draw-like fashion. Quite large specimens may then simply be placed onto the stage; the chamber is then closed and evacuated to its working pressure.

In other microscopes, the mechanical parts of the stage assem-bly are maintained at high vacuum within the chamber volume and the specimen is inserted onto the stage through an air lock. This arrangement limits the size of specimens that may be put into the microscope for imaging and analysis.

It is important to have a good understanding of the internal geometry of the specimen chamber of the SEM being used for microscopy and analysis. The dimension, position, and size of the final lens must be known. It is necessary to know the position of the specimen stage and how much it can be moved in the X , Y , Z, and tilt directions. It is important that the operator knows where the secondary and back scattered electron detectors and the energy dispersive x-ray detectors are located in the chamber and how much these may be moved. The secondary electron detector is usually in a fixed position at the edge of the chamber and the demountable solid state backscattered detector usually surrounds the final pole piece. The position of the backscattered detector may limit the shortest working distance between the sample surface and the final lens to 3–5 mm. The movable x-ray detector is usually located to one side of the chamber and only moved close to the sample during microanalysis. The position and subtending angle of the detector must be known to avoid a large sample crashing onto the front of the detector.

Figure 2.3 shows the relative position of the secondary, back-scattered, and x-ray detectors to a flat specimen at three different working distances in a FEI XL-30 SEM. These images were taken using a wide-angle, fixed focus, infrared cameral placed at a fixed point inside the microscope chamber. This a very useful lifetime system to ensure that bits and pieces inside the micro-scope column do not bang into each other. The quality of the recorded images is ideal.

The working distance is measured from the lower surface of the final lens to the top of the specimen. A short working dis-tance is associated with high-resolution images. A mid-working distance is used for most general SEM and, with the FEI XL-30, for x-ray microanalysis. The long working distance is used to provide an increased depth of field in order to view complexly sculptured specimens. The six images in Figure 2.3(A–F) were recorded either using a charge-coupled device (CCD) camera or an infrared camera attached to the inside of the microscope. They show the relative position of the specimen stage to different detectors as a function of working distance.

An alternative and rather unorthodox way of viewing the inside of the chamber is to attempt to record and image an uncoated glass sphere at 15–20 kV. The non-conductive specimen quickly

Page 21: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

2.2 Sample Selection 15

Figure 2.3. A. The settings for a short working distance of 5 mm. The secondary detector in all six figures is out of sight at the top right corner of the images. B. The settings at 7-mm working distance below the back-scattered (BS) detector. C. The settings are at a mid-working distance of 15 mm with no BS detector. D. Settings at 15-mm working distance with the BS detector in position. E. Settings at 15-mm working distance with the BS and the x-ray detector (XR) in position. F. Settings at a long working distance of 30 mm with both the BS and XR detectors in position

FINAL LENS

SPECIMEN STAGE

SPECIMEN

A

B

BS DETECTOR

C

Page 22: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

16 2. Sample Collection and Selection

charges and reflects the incoming electrons to the internal surface of the microscope, which provides a highly distorted image that shows the location and relative position of various chamber components.

These types of images, together with the manufacturer’s instructions and information, let the operator know how much space is available for a specimen inside the microscope column and how far it may be moved. For example, with a 2- to 5-mm thick flat specimen, the standard stage of an FEI XL-30 microscope

Figure 2.3. (continued)

D

X-RAYDETECTOR

E

F

Page 23: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

2.2 Sample Selection 17

can be moved 50 mm in the X and Y direction and the working distance moved 30 to 5 mm. Using a retractable x-ray detector, it is possible to safely examine a 25-mm cube of material.

2.2.2 Microscope Operating Conditions

Having established the spatial parameters of all the components inside the microscope, it is important to establish precisely what sort of information about the specimen is needed. Magnification, resolution, and depth of focus are determined to a large extent by the working distance between the specimen and the final lens and the position of the signal detectors. It is necessary when using a long working distance to examine a large specimen, to avoid touching any internal component inside the microscope. Long working distances reduce resolution but enhance the depth of focus when examining a highly convoluted sample at low magnification. If high-resolution images are required it is necessary to use a much smaller specimen that can be moved closer to the final lens. It may also be necessary to tilt the sample toward the x-ray detector in order to improve the take-off angle of the emitted x-ray photons.

2.2.3 Sample Size

Most specimens examined in the SEM are much smaller than the dimensions mentioned in the previous paragraphs and usu-ally conveniently fit onto the so-called Cambridge specimen stub, which is 12 mm in diameter and 3 mm thick. The general approach is to make the specimen as small as possible without compromising the appearance of the features of interest and the ability of the microscope to image and analyze these features.

2.2.4 Large Specimens

Sometimes it is not possible to cut a very large specimen, such as a piece of rock, down to a convenient size without damaging the material. It can be argued that a large specimen gives a more rep-resentative view of the material being examined. Large biological samples have their own set of problems. Such samples invariably need stabilization prior to examination in the SEM. This stabiliza-tion is time dependent. The larger the sample, the longer it takes to stabilize. The same problem occurs when it is necessary to dry the specimen. Large porous samples such as some plastics, fab-rics, and minerals may take a long time to pump out inside the microscope in order to reach the optimum high vacuum.

2.2.5 Small Specimens

Although smaller samples generally are easier to handle than large specimens, there is one disadvantage to their reduced size. It is important to be certain that the small sample that has been

Page 24: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

18 2. Sample Collection and Selection

separated from a much larger sample is representative of the whole specimen. This problem is not unique to the SEM but is one of the disadvantages of any reductive analysis process. One way around this conundrum is to examine a larger number of smaller samples taken from the same specimen.

2.2.6 Sample Shape

As far as possible the sample should preserve its natural shape. It may be necessary to examine both the natural surface of the sample and/or details of the interior. The procedures associated with exposing the interior of samples are discussed later, but at this early stage of sample selection it is important to know that both the surface and the interior of the specimen will go into the microscope for examination and analysis.

2.2.7 Sample View

The SEM is designed to obtain information from the very surface of specimens. This information may come either from the natural surface of the specimen or from a surface that has been exposed by artificial means to reveal the interior of the sample. A decision should be made early in the process of sample preparation about the proposed examination of the specimen.

2.3 Sample Labeling

It is vital that the SEM specimen is properly labeled. It is rarely possible to label the specimen directly, but this approach can be achieved with an indelible marker on the underside of a metal specimen. There are a number of marking systems, but an alpha numeric code is straightforward. The sample label should be quite unique and should follow the sample throughout the process of sample selection to the final image or data set. The same unique code should be used in the accompanying manda-tory laboratory notebook, which should have recorded details of everything that has happened to the sample.

If it is not immediately possible to label the sample directly, it should be kept in a labeled container. At the stage in the prepa-ration procedure in which the sample is firmly attached to the specimen holder, the unique label should be firmly attached to the underside of the specimen holder. In this way, the identity of the specimen will not be lost, particularly in conditions in which several different samples are placed on the same specimen stage.

Page 25: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

3.1 Introduction

Samples destined for examination and analysis by scanning microscopy come from two different sources. Many specimens are sent from established laboratories that, although fitted with the appropriate equipment designed for the study of all aspects of the particular discipline, may lack the facilities for microscopy and analysis. For example, a materials science laboratory would include a furnace, whereas a biological laboratory would be equipped with an ultra-microtome, although both types of labo-ratory would contain light microscopes and balances.

The other sources of specimens are those collected directly from the natural environment or from manufacturing facilities and com-mercial and public centers that lack any equipment for scientific enquiry. Thus, clothing for forensic study, a failed oil pipe, or a newly discovered meteorite require collection and transport to a facility for preliminary treatment before they are prepared for microscopy.

There are two types of SEM specimen preparation laborato-ries—those that provide a general service for any type of sample, and those dedicated only for the study of a particular type of specimen. It is inappropriate to provide a long list of all the tools, equipment, and processes used to handle all types of specimens. The prospective samples must first be examined in an appropri-ate institution to establish the nature of the problem; make a preliminary assessment in terms of size, shape, and form of the likely sample; and most important of all, must determine whether the SEM is the most appropriate way to provide information. Remember that smart scientists only tackle solvable problems.

The more specialized institutions may have some but not all of the tools and equipment for microscopy. Thus, a metallurgical laboratory would have equipment to handle quite large pieces of

3 Sample Preparation Tools

P. Echlin, Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray Microanalysis, DOI: 10.1007/978-0-387-85731-2_3,© Springer Science + Business Media, LLC 2009

19

Page 26: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

20 3. Sample Preparation Tools

metal using stretching, compression, bending, fracturing, cutting, and melting devices, together with saws, grinders, and polishing equipment. Some of the same types of equipment would be found in a laboratory handling large inorganic specimens such as rocks, fossils, and smaller samples such as particles, dust, and semicon-ductors. Hard organic materials such as wood, paper, and fabric rely on cutting, sawing, and fracturing equipment. Softer pliable material, such as many polymers, may only need to be cut to ini-tiate sample preparation. Studies on biological material require a broad range of specialized tools and equipment: Wet and liquid specimens may require using the specialized equipment devel-oped for low temperature microscopy and analysis.

The starting point for considering the essential tools and equip-ment needed for SEM sample preparation will be with specimens that, by one means or another, are considered appropriate for study in the SEM but that need specialized specimen preparation. The specialized tools and equipment needed to maintain and repair the microscope, x-ray microanalyzer, and image processing and storage devices are not considered.

3.2 Specimen Size

It is important to produce samples that will fit inside a SEM. Chapter 2 shows that some scanning electron microscopes can accommodate very large specimens by using large specimen stages and chambers. Most non-specialized scanning microscopes, although they can handle specimens up to 50 mm diameter and 20 mm thick, usually examine much smaller samples that can fit either on the 12-mm diameter, so-called Cambridge support or stub, or the larger, 25.4-mm (1-in.) stub shown in Figure 3.1 . (This and other images are courtesy Frank Platek, US Food and Drug

Figure 3.1. 12- and 24.5-mm aluminum specimen stubs

12mm and 25.4mm aluminium specimen stubs

Page 27: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

3.3 General SEM Sample Preparation Laboratory 21

Administration, Forensic Chemistry Center, Cincinnati, OH, although depiction of commercial products in any of the figures supplied by Frank Platek does not imply endorsement by the US Food and Drug Administration. (See Figs. 3.3, 3.5, to 3.8, & 3.11))

The metal support stubs shown in Figure 3.1 are suitable for holding specimens that are to be imaged in the SEM. However, as shown in Chapter 9, more specialized support stubs may be required for specimens that are to be analyzed by x-ray microanalysis. Figure 3.2 shows some of these stubs.

3.3 General SEM Sample Preparation Laboratory

If involved in the construction of a new SEM laboratory, it is useful to first consider the various activities that will take place. It is inappropriate to discuss details of laboratory design and construction here, but the following requirements are important.

There must be separate wet and dry areas; a low-vibration area for sensitive equipment; a region set aside for sputter and evaporative coatings; spark-free storage cupboards and fume chambers; drying ovens and ultrasonic cleaners; plenty of cup-boards and shelves; a region set aside for any pieces of special-ized equipment; and secure places to tether high-pressure gas bottles, such as those containing nitrogen, carbon dioxide, and argon. The whole laboratory should be well lit and ventilated, and fitted with plenty of electrical power points and appropriate laboratory furniture, including comfortable chairs. There must be separate waste bins for glass and sharps, toxic and hazardous material, and general waste. All these pieces of equipment and the general arrangement and operation of the laboratory must meet the Health and Safety regulations in force in a particular country,

Figure 3.2. 12- and 25.4-mm porous and pyrolytic carbon support plan-chets used for imaging and analyzing specimens in the SEM

Porous and pyrolytic graphite planchets

Page 28: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

22 3. Sample Preparation Tools

as some of the chemicals and many of the operations involved in sample preparation are potentially hazardous.

As far as appropriate, specimen preparation should be carried out on a firm bench with a comfortable adjustable chair, because small hand-held tools must be held securely. For very small specimens it may be necessary to use a micromanipulator to hold and move the tools.

3.4 Equipment to Facilitate Looking at the Process of Sample Preparation

A few larger pieces of equipment facilitate specimen preparation, particularly for smaller samples. It is useful to have a good quality 100-mm diameter glass hand lens with a magnification of ×5 that can be fixed to an adjustable support and is close to a pair of flexible light sources. A smaller ×10 magnification hand-held lens is useful for inspecting the sample more closely. Easy access to a good binocular light microscope is an additional advantage.

Specialist laboratory suppliers provide tools for SEM, but depending on the type of samples being handled, suitable tools also can be found at DIY and hardware stores, opticians, den-tists, and model building shops. The following two companies provide specialized tools for use with the SEM. Microtools for Microscopists ( www.minitoolinc.com ) and Dorn and Hart Microedge Inc. ( www.dornandhart.com ). Additional information is given in Chapter 13.

3.5 Tools to Expose Samples

3.5.1 Grinders, Saws, and Cutters for Hard Samples

The size of these tools depends very much on the size of the specimen being studied. Choose the size, form, and power for the expected type of specimen. Dental burrs are an excellent way to trim and grind small samples of hard material. Mud, soils, and liquids may contain fine particles of suspended or deposited material, and the filtration equipment shown in Figure 3.3 can be used to separate liquids and solids.

3.5.2 Blades, Knives, and Scissors for Soft Samples

Sharp surgical knives and disposable scalpel blades of various shapes and size are available from a number of different supply houses. Disposable razor blades are very effective for cutting away larger pieces of soft material, but it is important to first wipe the surface with acetone to remove the fine layer of oil. Very thin stainless steel razor blades can be cut to size using scissors and then held in a metal holder.

Page 29: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

3.5 Tools to Expose Samples 23

Tungsten carbide knives are harder and tougher and can be used to cut soft metals. Kyocera ceramic knives ( www.kyocera.co.uk ) have blades made of zirconium oxide, which is second in hardness to diamond. These blades are very sharp and retain their edge for 6 months. Unlike metal knives they are rather brittle and may break when dropped onto a hard surface. The diamond and glass knives used for microtomy generally are not suitable for cutting and trimming samples for SEM, although they are the best way to cut sections of soft specimens. When using thin blades and knives, hold the cutting edge at an angle and as you cut, slowly and con-tinually draw the blade down and toward you to minimize compressing the sample.

Surgical scissors of different size and shape are available; curved blades are particularly useful. One disadvantage of even very sharp scissors is that the blades compress the edges of the sample during the cutting process. If the cutting tools have to be heat sterilized before or after use, it is important to choose steel blades that remain sharp after such treatment. A wide range of microsurgical tools are available that can be used for both cutting and probing very small specimens. It is useful to have spring-loaded cutting edges that automatically retract after the incision is made.

Figure 3.3. Filtration apparatus and filters composed of a system that allows a liquid suspension to pass through filters of various size designed to retain solids of different sizes and compositions

Filtration Apparatusand Filters

Page 30: Handbook of Sample Preparation for Scanning Electron ...€¦ · Electron Microscopy and X-Ray Microanalysis. Handbook of Sample Preparation for Scanning Electron Microscopy and X-Ray

24 3. Sample Preparation Tools

3.6 Tools to Manipulate Samples

3.6.1 Fine-Pointed Probes, Needles, and Brushes

Dental probes are suitable to expose hidden parts of the sample. Figure 3.4 shows that they are not necessarily expensive.

Sewing needles may be used to pick up small specimens and move them to the sample holder. Fractured tungsten wire provides very fine-pointed needles and the fine fresh needles used in acu-puncture are sterile and have very fine points. Small disposable needles can be held conveniently in metal holders.

In addition to the metal needles and probes, wooden applica-tor sticks, toothpicks, fine glass needles, and even some plant prickles, thorns, and spines can be used to manipulate small sam-ples. Figure 3.5 shows a prepared set of disposable cactus needles that may be used to manipulate specimens.

Figure 3.4. A set of stainless steel dental probes

A set of stainless steel dental probes purchases from HarborFreight Tools for $2.99. October 2007