general calibration/verification procedure for pressure measuring devices
TRANSCRIPT
-
7/29/2019 General Calibration/verification Procedure for Pressure Measuring Devices
1/5
GENERAL PROCEDURE
FOR
CALIBRATION/VERIFICATION
OF
PRESSURE INSTRUMENT MEASURING DEVICES
22 FEBRUARY 2013
-
7/29/2019 General Calibration/verification Procedure for Pressure Measuring Devices
2/5
1
TABLE OF CONTENTS
INSTRUMENTS
PURPOSE AND SCOPE
REQUIRED STANDARD
PREPARATION FOR CALIBRATION
CALIBRATION PROCESS
UNCERTAINTY OF MEASUREMENT
-
7/29/2019 General Calibration/verification Procedure for Pressure Measuring Devices
3/5
2
CALIBRATION / VERIFICATION OF PRESSURE MEASURING DEVICES
1.0 INSTRUMENTS
Pressure Gauges, Test Gauges, Pressure Indicator, Pressure Recorders,
Pressure Meters, Pressure Transmitters, Pressure Transducers, Pressure
Calibrators and any related pressure measuring instruments.
2.0 PURPOSE AND SCOPE
The purpose of this procedure is to provide the calibration/verification method
for pressure measuring instruments.
The document reference(s): BS EN 837-1, 3:1998, EURAMET cg-17 version
2.0(03/2011) and In-house calibration procedure, CP-P001 rev. 5.0 and
CP-P003 rev. 3.0
3.0 ENVIRONMENTAL CONDITION
Ambient Temperature and Relative Humidity
The Ambient Temperature and Relative Humidity for this calibration process will
be determined by taking the average of temperature and humidity during
calibration process at start, middle and end of calibration.
4.0 REQUIRED STANDARDS (OR EQUIVALENT EQUIPMENT)
a) Hydraulic / pneumatic Dead-weight Tester
b) Hydraulic / Pneumatic Pressure Calibrator
c) Hydraulic / Pneumatic Pressure Indicator
d) Hydraulic / Pneumatic Standard Test Gauge
e) Hydraulic / Pneumatic Test Comparatorf) Hydraulic / Pneumatic Hand Pump
g) Field Communicator
-
7/29/2019 General Calibration/verification Procedure for Pressure Measuring Devices
4/5
3
5.0 PREPARATION FOR CALIBRATION
a) Inspect the Unit under Test (UUT) and clean as required.
b) Ensure UUT is not defective and in good working condition.
c) Ensure Standards (e.g. DWT) is level, filled with sufficient oil and ready
for the calibration. (Note: Normally DWT is recommended used for
calibrating of high accuracy of UUT, such as UUT accuracy better than
or equal 0.25% of full scale).
d) The selections of Standard to be used as appropriate manner by the
operator.
e) Ensure the medium used is correct / compatible to UUT.
f) Connect UUT to the Standards (e.g. DWT) and carried out priming as
necessary.
g) Used glove when handling the Dead-weights for pressure Tester.
6.0 THE CALIBRATION PROCESS
a) Exercise UUT slowly, 3 times form minimum and to maximum and
minimum of the scale.
b) Check UUT over entire the scale from 0% to 100%.Taking the reading
both up and down the scale.
c) Taking readings at uniform intervals along its entire e.g. 0%, 10%, 20%,
40%, 60%, 80% and 100%.
d) Hold at each pressure point a few minutes before noting down values.
e) The reading shall be obtained after the gauge has been lightly tapped.
f) If out of accuracy, release pressure and vent testing apparatus to
ambient pressure before UUT removed, and perform adjustment.
g) For the Test Gauges, with both increasing and decreasing pressure atany point on the scale shall not exceed 0.25%of the maximum scale
value.
h) For normal industrial gauges other than 50 mm diameter, the error of
pressure above 0% and below 90% of the maximum scale value shall
not exceed 1% of the maximum scale value.
-
7/29/2019 General Calibration/verification Procedure for Pressure Measuring Devices
5/5
4
i) Close the vent valve; perform a complete calibration by taking the
average of 3 cycles increasing and decreasing pressure readings at
each of the above calibration points.
j) On completion of calibration, release / vent pressure to ambient before
removing UUT.
7.0 UNCERTAINTY OF MEASUREMENT
The uncertainty is at 95% confidence level based on the ISO Guide to the
Expression of Uncertainty in Measurement