gas flow measurement

Upload: anuranjan

Post on 04-Jun-2018

229 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/13/2019 Gas Flow Measurement

    1/71

    UNIT-5

    GASFLOWMEASUREMENT

  • 8/13/2019 Gas Flow Measurement

    2/71

    It is required to enable the determination of the amount ofgas being produced or sold, and also as a basicparameter for almost all of the design procedures.

    The produced gas stream is in a continuous state of flowfrom the instant it leaves the reservoir until it is consumedat the delivery end.

    Gas measurements must be done mostly on a flowingstream of gas.

    Gas is most commonly measured in terms of volumebecause of the simplicity of procedure.

  • 8/13/2019 Gas Flow Measurement

    3/71

    MEASUREMENTFUNDAMENTALS

    Flow is one of the most difficult variables to

    measure because it cannot be directly measured

    like temperature.

    It must be inferred by indirect means such as the

    pressure differential over a specified distance,

    speed of rotation of a rotating element etc.

    Many flow measurement techniques and devices

    have been developed for a wide range applications.

  • 8/13/2019 Gas Flow Measurement

    4/71

    PARAMETERSTOCHARACTERIZETHEFLOW

    METER

    Accuracy

    Rangeability

    Repeatability

    Linearity

  • 8/13/2019 Gas Flow Measurement

    5/71

    ACCURACY

    This is a measure of a flow meters ability to indicatethe actual flow rate within a specified flow raterange.

    It is defined as the ratio of the difference betweenthe actual and measured rates to the actual rate.

    Accuracy = Actual rate-Measured rate

    ----------------------------- X 100%Actual rate

    Accuracy is represented in either two ways :percent of full scale, or percent of reading.

  • 8/13/2019 Gas Flow Measurement

    6/71

    RANGEABILTY

    A flow meters rangeability is the ratio of themaximum flow rate to the minimum flow rate at thespecified accuracy.

    Rangeability= Maximum rate that can bemeasured

    ---------------------------------------------------

    Minimum rate that can be measured

    Rangeability is usually measured as a ratio x : 1

  • 8/13/2019 Gas Flow Measurement

    7/71

    REPEATABILITY

    It is also known as reproducibility or precision

    ,repeatability is the ability of a meter to reproduce

    the same measured readings for identical flow

    conditions over a period of time.

    It is computed as the maximum difference between

    measured readings.

  • 8/13/2019 Gas Flow Measurement

    8/71

    LINEARITY

    This is a measure of the deviation of the calibration

    curve of a meter from straight line.

    It can be specified over a given flow-rate range, or

    at a given flow rate.

    A linear calibration curve is desirable because it

    leads to a constant metering accuracy.

  • 8/13/2019 Gas Flow Measurement

    9/71

    SELECTIONOFMEASUREMENT

    Accuracy desired

    Expected useful life of the measuring device

    Range of flow ,temperature

    Maintenance requirements

    Power availability Liquid or gas

    Cost of operation

    Initial cost

    Availability of partsAcceptability by others involved

    Purpose for which measurements are to be used

    Susceptibility to theft

  • 8/13/2019 Gas Flow Measurement

    10/71

    DIFFERENTIALPRESSUREMETHOD

    There are basically two types of differential

    pressure devices

    The pressure difference is measured across a flowrestriction.

    Eg: Orifice, venturi

    The difference in pressure measured upon impactEg: Pitot tube

  • 8/13/2019 Gas Flow Measurement

    11/71

    ORIFICEMETER

  • 8/13/2019 Gas Flow Measurement

    12/71

  • 8/13/2019 Gas Flow Measurement

    13/71

  • 8/13/2019 Gas Flow Measurement

    14/71

    ORIFICEMETER

    This is the most commonly used device formetering natural gas.

    It consists of a metal plate with a circular hole

    ,centered in a pair of flanges in a straight pipesection.

    The pressure differential is measured across thisplate to yield the flow rate .

    This is a rugged, accurate, simple, and economicaldevice and can handle a wide range of flow rates.

  • 8/13/2019 Gas Flow Measurement

    15/71

    Orifice flow meters are used to determine a liquid or gasflow rate by measuring the differential pressure (P1 - P2)across the orifice plate.

    They are generally less expensive to install andmanufacture than the other commonly used differentialpressure flow meters;

  • 8/13/2019 Gas Flow Measurement

    16/71

  • 8/13/2019 Gas Flow Measurement

    17/71

    VENTURIMETER

  • 8/13/2019 Gas Flow Measurement

    18/71

  • 8/13/2019 Gas Flow Measurement

    19/71

    VENTURIMETER

    A venturi is a point in a pipe that has been narrowed sothat the flow is restricted slightly.

    The venturi is widely used because it has no movingparts and the small amount of restriction it produces toinduce a pressure drop does not disturb the fluid flowtoo much.

    The change in cross-sectional area in the venturi tubecauses a pressure change between the convergentsection and the throat, and the flow rate can bedetermined from this pressure drop.

    Although more expensive that an orifice plate; theventuri tube introduces substantially lower non-recoverable pressure drops.

  • 8/13/2019 Gas Flow Measurement

    20/71

    ADVANTAGES

    The pressure recovery is much better for the venturi

    meter than for the orifice plate.

    The venturi tube is suitable for clean, dirty and viscousliquid and some slurry services.

    The rangeability is 3.5:1.

    Pressure loss is low.

    Typical accuracy is 1% of full range.

  • 8/13/2019 Gas Flow Measurement

    21/71

    PITOT TUBE

  • 8/13/2019 Gas Flow Measurement

    22/71

    The pitot tube measures the difference between the

    static pressure at the wall of the flow conduit and the

    flowing pressure at its impact tip where the kinetic

    energy of the flowing stream is converted intopressure.

    It gives the flow velocity only at a point.

    The tip can be easily clogged by liquids or solids.

    Because of the relatively poor accuracy of this deviceit is not used very often.

  • 8/13/2019 Gas Flow Measurement

    23/71

    Turbine flow meter

  • 8/13/2019 Gas Flow Measurement

    24/71

    These meters are sometimes classified as positive

    displacement meters.

    They consist of a turbine or propeller that turns at aspeed proportional to the velocity of the gas ,

    converting linear velocity to rotational speed.

    Turbine meters have been used for measuring liquid

    flow rates rather than gas flow rates.

    Fluctuations in velocity, caused by pressure

    fluctuations, turbulence or unsteady state flow

    conditions, will cause the turbine meter to give a

    higher than actual value.

  • 8/13/2019 Gas Flow Measurement

    25/71

    ROTAMETER

    A Rotameter is a device that measures the flow rate ofliquid or gas in a closed tube.

    It belongs to a class of meters called variable area meters.

    A rotameter consists of a tapered tube, typically made ofglass, with a float inside that is pushed up by flow andpulled down by gravity.

    The fluid entering at the base of the tube causes the floatto rise until the annular area between the float and the tubewall is such that the pressure drop across this constrictionis just sufficient to support the float.

  • 8/13/2019 Gas Flow Measurement

    26/71

    ROTAMETER

  • 8/13/2019 Gas Flow Measurement

    27/71

  • 8/13/2019 Gas Flow Measurement

    28/71

    ADVANTAGES

    A rotameter requires no external power or fuel, ituses only the inherent properties of the fluid, alongwith gravity, to measure flow rate.

    A rotameter is also a relatively simple device that

    can be mass manufactured out of cheap materials,allowing for widespread use in places such as thirdworld countries

  • 8/13/2019 Gas Flow Measurement

    29/71

    DISADVANTAGES Due to its use of gravity, a rotameter must always be vertically

    oriented and right way up, with the fluid flowing upwards.

    Due to its reliance on the ability of the fluid or gas to displace thefloat, the graduations on a given rotameter will only be accurate for agiven substance. The main property of importance is the density ofthe fluid. Either separate rotameters for different substances must beused, or the read out adjusted.

    Rotameters normally require the use of glass (or other transparentmaterial), otherwise the user cannot see the float. This limits theiruse in many industries to benign fluids, such as water.

    Rotameters are not easily adapted for reading by machine: althoughmagnetic floats that drive a follower outside the tube are available.

    http://en.wikipedia.org/wiki/Waterhttp://en.wikipedia.org/wiki/Water
  • 8/13/2019 Gas Flow Measurement

    30/71

    ULTRASONICMETERS

    A typical transit-time flow measurement system utilizes two

    ultrasonic transducers that function as both ultrasonic transmitter

    and receiver. The flow meter operates by alternately transmitting and

    receiving a burst of sound energy between the two transducers andmeasuring the transit time that it takes for sound to travel between

    the two transducers. The difference in the transit time measured is

    directly and exactly related to the velocity of the liquid in the pipe.

  • 8/13/2019 Gas Flow Measurement

    31/71

    To be more precise, let's assume that Tdown is the transit-

    time (or time-of-flight) of a sound pulse traveling from the

    upstream transducer A to the downstream transducer B, and

    Tup is the transit-time from the opposite direction, B to A. The

    following equations hold:

    Tdown = ( D / sin) / ( c + V*cos ), (1)

    Tup = ( D / sin) / ( c - V*cos), (2)

    where c is the sound speed in the liquid, D is the pipe

    diameter and V is the flow velocity averaged over the soundpath. Solving the above equations leads to

    V = ( D / sin2 ) * T / (Tup * Tdown), (3)

    where T = Tup - Tdown. Therefore, by accurately measuringthe upstream and downstream transit-time Tup amd Tdown,

    we are able to obtain the flow velocity V. Subsequently, theflow rate is calculated as following,

    Q = K *A* V, (4)

    where A is the inner cross-section area of the pipe and K is

    the instrument coefficient.

  • 8/13/2019 Gas Flow Measurement

    32/71

    ORIFICEMETERS

    Because orifice meters are simple, accurate, relativelyinexpensive they are most widely used of the flow metersfor gases.

    An orifice meter consists of a thin plate ,0.115-0.398 in.thick depending upon the pipe size and pressure, held

    perpendicular to the direction of flow by a pair of flanges,with a circular sharp square edged orifice accuratelymachined to the required size in the centre of the plate.

    Pressure taps are provided on the upstream and downstream end in the fitting that holds the orifice plate.

    A pressure measuring and recording device is connectedto the pressure taps.

  • 8/13/2019 Gas Flow Measurement

    33/71

    ORIFICETYPESDifferent kinds of orifice plates include concentric, eccentric,

    and segmental, each of which has different shapes andplacements for measuring different processes.

  • 8/13/2019 Gas Flow Measurement

    34/71

    The concentric type is the most common ,because of itslow cost, ease of fabrication and ease of calibration.

    The eccentric and segmental types are very useful fortwophase flow streams and for flow streams withsuspended solids such as dirty gases or slurries.

    Rangeability -3:1

    Accuracy; +/- 1.5-2%

  • 8/13/2019 Gas Flow Measurement

    35/71

    LOCATIONOFPRESSURETAPS

    The magnitude of the measured pressure differential is

    obviously affected by the location of the points across theorifice between which it is measured.

    Pressure taps are designated as P1 and P2. "D" is thediameter of the pipe and "d" is the diameter of the orifice.

    The four types of pressure tap locations that are used:

    Flange type

    Pipe tap

    Corner type

  • 8/13/2019 Gas Flow Measurement

    36/71

    LOCATIONOFTAPS

  • 8/13/2019 Gas Flow Measurement

    37/71

    Flange type: In this type the pressure is measured 1 in.

    from the upstream face of the plate and 1 in from the

    down stream face of the orifice plate. This is the mostcommon type of pressure tap.

    Pipe taps: 2.5 IDs from the upstream, and 8 Pipe IDs

    from the down stream.

    Corner types: In this type the pressure taps are located

    immediately adjacent to the upstream and down stream

    faces of the orifice plate.

  • 8/13/2019 Gas Flow Measurement

    38/71

    STRAIGHTENINGVANES

    Straightening vanes are used to minimize the flowdisturbances in meters.

    Flow eddies, rotation swirls and other undesirable

    flow patterns are minimized as the flow passesthrough the relatively small tubes.

    Straightening vanes are available as pin type or

    flange type in carbon steel or stainless steel.

  • 8/13/2019 Gas Flow Measurement

    39/71

  • 8/13/2019 Gas Flow Measurement

    40/71

  • 8/13/2019 Gas Flow Measurement

    41/71

    MEASUREMENTCALCULATIONS

    The relationship for orifice meters can be derived fromthe general energy equation written between two points

    in the flowing stream point1 being some pointupstream of the orifice plate & point-2 the orifice throat.

    1

    2

    Vdp+1/gc 12

    v dv + g/gc12

    dz = wslw---------(1)For most meters ,change in elevation between points 1and 2 , dz is zero, and no work is done by the flowing

    fluid stream. Therefore equation 1 is written as

    12

    Vdp+1/gc 12

    v dv + lw=0 -----------(2)The lost work term expresses the frictional losses due to

    viscosity and turbulence of the fluid.

  • 8/13/2019 Gas Flow Measurement

    42/71

    These losses can be handled in a manner convenient

    for meter calculations without reference to friction

    factor.

    The basic orifice equation can be written in the form:C212Vdp+1/gc 12v dv=0 -----------(3)

    C= Empirical constant that takes care of friction and

    other irreversibilities.

    Multiplying with

    C212 dp+1/gc 12 v dv =0 -----------(4)

    Assuming a constant , average density avfor simplicityand integrating the equation (4) we get:

    C2(p2- p1) =( av/ 2gc)(v22-v12)=0 -------- (5)

    Converting to commonly used pressure units of psia

    (lbf/in2) and rearranging we get:

  • 8/13/2019 Gas Flow Measurement

    43/71

    (v22-v1

    2) = 2(144) C2(p1- p2) / av-------------(6)

    The mass flow rate ,m (lbm/sec),is given by

    m = vA

    where A= cross sectional area of flow ,ft2

    This analysis assumes steady-state flow conditions , for

    which the mass flow rate is constant. Equation (6)

    can now be written as:

    (m2/ av2 )[1/A22-1/A12]=2(144)gcC2(p1-p2)/ av -------(7)

    Taking A22as common

    (m2/ A22 )[1-(A2/A1)

    2]=2(144) gcC2 (p1-p2) av -------- (8)

    Let d1 and d2 be the diameters of the pipe and theorifice , respectively, in inches. Defining =d2/d1andsolving equation (8) for m:

  • 8/13/2019 Gas Flow Measurement

    44/71

    m= C A2[(2(144)gc(p1-p2) av) /(1- 4) ]0.5

    Or

    m= C d22 [gc(p1-p2) av) / 1152(1- 4) ]0.5 -------(9)Using the gas law, the gas density can be expressed as

    :

    av=28.97 g p av / Zav R Tav ------- (10)The pressure differential (p1-p2) is generally expressed

    in terms of inches of water. This conversion can be

    achieved ,using the relation

    p= g h/gc and is written as

    (p1-p2) = {62.43 h/(144)(12)} -------(11)

    h=pressure differential in inches of water.

  • 8/13/2019 Gas Flow Measurement

    45/71

    Using equations (10) and (11) ,(9) becomes :

    m = C d2

    2 [ (28.97) (62.43)gc

    gh p

    av) / (1,152)(144)(12) (1- 4) Z

    avR Tav]0.5 --------------------(12)

    Gas flow is generally reported in terms of the flow rate qsc

    in scf /hr at standard conditions , which is related to the

    mass flow rate m in lbm/sec as follows (m=q)

    m=(qsc /3600) {(28.97) gpsc/ ZscRT sc}

    Using standard conditions of psc=14.73psia,Tsc=520oR

    and Zsc=1 we obtain:

    m= {(28.97)(14.73)/(3600)(520)R} g

    qsc

    -----------(13)

    Using equation (13) in (12) substituting R=10.732 psia-

    ft3/lb mole-oR and solving for qscwe obtain:

    qsc= {7,717.96 Cd22 }/ {(1- 4) gZ avTav }0.5 {hpav}0.5

  • 8/13/2019 Gas Flow Measurement

    46/71

    In equation 14

    qscis in scf/hr

    d2is in inches

    h is in inches of water

    Pav is in psia

    Tav is inoR

    C, , g,Z av are dimensionlessThe equation 14 is commonly expressed as:

    qsc = Ko {h pav}0.5

    The constant Ko is given by

    Ko= 7717.96 Cd22 / [ (1- 4) g Zav Tav]0.5 ------------(15)

  • 8/13/2019 Gas Flow Measurement

    47/71

    In metering practice ,the average pressure pavis replaced

    by a measurable gauge pressure pf.

    Factors are provided to account for this pf being

    ,measured at the upstream or down stream, or being

    measured as the mean of upstream and downstream

    static pressures and for the type of pipe tap.

    Equation is then written in the following form:

    qsc=K [hwpf]0.5-----------(16)

    hw=differential pressure at 60oF , inches of water

    pf=absolute static pressure of the flowing fluid, psia

    And the constant is expressed as a product of severaldifferent as follows.

    K F F Y F F F F F F F F (17)

  • 8/13/2019 Gas Flow Measurement

    48/71

    K= FbFrY Fpb FtbFtf FgFpvFmFlFa ------------(17)

    Fb= basic orifce factor, scf/hr

    Fr = Reynolds number factor

    Y = expansion factor

    Fpb = pressure-base factor

    Ftb =Temperature base factor

    Ftf=flowing temperature factor

    Fg= specific gravity factor

    Fpv =supercompressibility factor

    Fm = manometer factor

    Fl=gauge location factorFa = orifice thermal expansion factor

  • 8/13/2019 Gas Flow Measurement

    49/71

    1) Basic Orifice Factor , Fb :

    The factor is simply the constant in equation (16) Its value depends upon the type of pressure taps and

    the pipe and orifice diameters.

    Fbcan be obtained from tables 10-2 and 10-7 for flange

    taps and pipe taps respectively.

  • 8/13/2019 Gas Flow Measurement

    50/71

    2) Reynolds Number Factor, Fr:

    This factor accounts for the variation of the orifice

    discharge coefficient with Reynolds number.

    Tables 10-3and 10-8 show the value of Fr for flange

    taps and pipe taps, respectively.

    If Fris small it can be neglected.

  • 8/13/2019 Gas Flow Measurement

    51/71

    3) Expansion Factor , Y:

    This factor accounts for the change in gas density with

    the pressure changes across the orifice. The expansion factor can be obtained from Tables 10-

    4,10-5,and 10-6 for flange taps, and Tables10-9and 10-

    10 for pipe taps.

    These tables indicate the pressure tap from which theabsolute static pressure pf is measured Y1 for

    upstream,Y2 for down stream and Ym for static pressure

    recorded as the mean of the upstream and down stream

    static pressures.

  • 8/13/2019 Gas Flow Measurement

    52/71

    4) PressureBase Factor ,Fpb:

    This factor corrects for cases where the base (standard)

    pressure , pb

    in psia , at which flow is to be measured is

    other than 14.73 psia:

    Fpb=14.73/pb

    5) Temperature- Base Factor, Ftb:

    This factor corrects for cases where the base

    (standard) temperature ,TbinoR at which flow is to be

    measured is other than 520oR:

    Ftb=Tb/520

  • 8/13/2019 Gas Flow Measurement

    53/71

    6) Flowing Temperature Factor, Ftf:

    The flowing temperature factor corrects for cases

    where the flowing temperature Tf

    (oR) , is not 520oR, using

    the fact that the gas flow rate varies inversely as the

    square root of the absolute flow temperature:

    Ftf=[520/Tf]0.5

    7) Specific Gravity factor, Fg:

    The basic orifice factor ,Fb, is determined assuming a

    gas gravity of 1.0 .So, a correction for gas gravity is

    required, as follows:

    Fg=1/g0.5

    8) Supercompressibility Factor F :

  • 8/13/2019 Gas Flow Measurement

    54/71

    8) Supercompressibility Factor , Fpv:

    This factor corrects for the deviation of an actual gas

    from ideal-gas behavior . It is calculated as follows:

    Fpv=Zb/Z0.5

    Zb assumed to be equal to 1.0

    Z at operating conditions

    Due to variations in gas compressibility factors with

    gas composition, pressure and temperature, Fpv isdetermined experimentally or through empirical

    techniques.

    Specific gravity method

    Heating value method

    1) S ifi it th d

  • 8/13/2019 Gas Flow Measurement

    55/71

    1) Specific gravity method:

    This uses the specific gravity and the carbon

    dioxide(CO2)and nitrogen (N2) contents of the gas to

    calculate the pressure and temperature adjustmentindices, fpgand ftgrespectively, as follows:

    fpg = g- 13.8yC+5.420yNftg= g- 0.472yC0.793yN

    g=specific gravity of the gas (air=1)yC=mole fraction of the CO2in the gas

    yN=mole fraction of N2in the gas

  • 8/13/2019 Gas Flow Measurement

    56/71

    These values are used to determine the pressure and

    temperature correction factors from Tables 10-11a and

    10-11b , that are added to the actual flowing pressure

    and temperature of the gas, respectively. these correctedpressure and temperature values are used in Table 10-11

    e to estimate the supercompressibilty factor, Fpv.

    2) H ti V l M th d

  • 8/13/2019 Gas Flow Measurement

    57/71

    2) Heating Value Method:

    The heating value method uses the specific gravity , total

    heating value (Ht in Btu/scf), and the CO2content of the

    gas to calculate the pressure and temperatureadjustment indices, fph and fthas follows:

    fph=g-0.0005688Ht+3.690yC

    fth=g-0.001814Ht+2.641yC

  • 8/13/2019 Gas Flow Measurement

    58/71

    These values are used to determine the pressure and

    temperature correction factors for Tables 10-11c and 10-

    11d, that are added to the actual flowing pressure andtemperature of the gas respectively. These corrected

    pressure and temperature values are used in Table 10-

    11e to estimate the super compressibility factor ,Fpv.

  • 8/13/2019 Gas Flow Measurement

    59/71

    9) Manometer Factor, Fm:

    This factor is required only where mercury manometer is

    used for measuring the differential pressure. It

    compensates for the different heads of gas above the two

    mercury columns of the manometer.

    It is generally negligible and is totally ignored for

    pressures below 500 psia. Table 10-12 gives this

    correction factor as a function of gas gravity, flowingpressure, and ambient temperature.

  • 8/13/2019 Gas Flow Measurement

    60/71

    10) Gauge Location Factor, Fl:

    The gauge location factor fl given in Table 10-13 is used

    where orifice meters are installed at locations other than

    sea level elevation and 45o latitude. This is also a verysmall correction.

  • 8/13/2019 Gas Flow Measurement

    61/71

    11) Orifice Thermal Expansion factor, Fa:

    This factor accounts for the expansion or contraction of

    the orifice hole with flowing temperature, calculated as

    follows:

    Fa=1+[0.0000185(Tf-528)] for stainless steel

    Fa=1+[0.0000159(T

    f-528)] for monel

    Tf= gas flowing temperature at the orifice,oR

  • 8/13/2019 Gas Flow Measurement

    62/71

    ORIFICEMETERSELECTION

    Several factors need to be considered in choosing anorifice metering system:

    Flow rate : flow rate uniformity, maximum and minimumflow rates expected.

    Pressure: expected static and differential pressures and

    their range ; permissible pressure variations. the size of the orifice affects the range of flow rates that

    can be measured and the pressure differential that will beobtained.

    A well designed metering system can only be achieved ifall these factors are carefully considered in choosing thesize and type of orifice and the pressure measuringdevices.

  • 8/13/2019 Gas Flow Measurement

    63/71

    Question: An orifice meter with a 2-inch orifice

    equipped with pipe taps using upstream static

    connections in a 6-in nominal(6.065-in. internal

    diameter) pipe line shows an average differential

    head=60in.water and an average upstream static

    pressure=90psia. The flowing temperature is50oF and the gas gravity is 0.65. Using a base

    pressure of 14.9 psia and base temperature of

    50oF, calculate the gas flow rate indicated by the

    meter.

  • 8/13/2019 Gas Flow Measurement

    64/71

    FACTORSAFFECTINGORIFICEMETER

    ACCURACY

    Following are the sources of constant errors:

    1. Incorrect estimate of orifice size

    2. Convex or concave contouring of the orifice plate

    3. Thick or dull orifice edge

    4. Eccentricity of orifice with respect to the pipe

    5. Incorrect estimate of pipe diameter

    6. Excessive recess between the end of pipe and

    the face of the orifice plate7. Excessive pipe roughness.

    1 Flow disturbances

  • 8/13/2019 Gas Flow Measurement

    65/71

    1. Flow disturbances

    2. Imprecise location of the pressure taps

    3. Pulsating flow

    4. Buildup of solids

    5. Liquid accumulation

    6. Differences or changes in operating conditions

    7. Incorrect zero adjustment

    8. Non-uniform calibration

    9. Corrosion or deposits in the metal internals

    10. Emulsification of liquids with mercury

    11. Leakage around the orifice plate12. Formation of hydrates in meter piping

    13. Over dampening of the meter response

  • 8/13/2019 Gas Flow Measurement

    66/71

    COMMONMEASUREMENTPROBLEMS

    Some of the common measurement problemsencountered in gas metering are:

    Hydrate formation

    Pulsating flow

    Slugging

    Sour gas

    H

  • 8/13/2019 Gas Flow Measurement

    67/71

    HYDRATEFORMATION

    Hydrates may be formed at the orifice ,or in the meter-

    piping or internals, whenever the gas temperature fallsbelow the hydrate-forming temperature for the gas.

    Hydrate formation can be prevented using any of the

    following :

    Gas dehydration Use of hydrate inhibitors

    Installation of heaters along the line or near the meter

    Other methods

  • 8/13/2019 Gas Flow Measurement

    68/71

    PULSATINGFLOW

    Pulsating flow is flow comprising suddenchanges in pressure and flow rate of the flowing

    fluid.

    Common sources of such flow in gas

    measurement are:Reciprocating systems-compressors ,or engines

    Improperly sized, loose, or worn valves and

    regulators Two-phase flow conditions

    Intermitters on wells and automatic drips.

    Prevention:

  • 8/13/2019 Gas Flow Measurement

    69/71

    Prevention:

    1. Locate the meter along the flow line in a position where

    pulsations are minimized.

    2. Reduce the amplitude of the pulsations by placing avolume capacity ,flow restriction, or specially designed

    filter between the pulsation source and the meter.

    3. Operate at pressure differentials as high as possible by

    using a smaller diameter orifice.

  • 8/13/2019 Gas Flow Measurement

    70/71

    SLUGGING

    1. Slugging refers to the accumulation of liquids inthe gas flow line.

    2. Liquid accumulation

    3. Liquid is swept through to the orifice and beyond

    4. Prevention is the installation of liquid

    accumulators in the flow line.

  • 8/13/2019 Gas Flow Measurement

    71/71

    SOURGAS

    Sour gas is detrimental for two reasons:

    1. Corrosion

    2. Accelerated hydrate formation

    Preventive measures are:

    1. To ensure proper gas metering include using H2S

    resistant components in the meters.

    2. Sealing the meters against the atmosphere.