Fundamentals of modern UV-visible spectroscopy

Presentation Materials

Page 2: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy2Figure :

The Electromagnetic Spectrum

= c / E = h

Page 3: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy3Figure :

Electronic Transitions in Formaldehyde

Page 4: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy4Figure :

Electronic Transitions and Spectra of Atoms

Page 5: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy5Figure :

Electronic Transitions and UV-visible Spectra in Molecules

Page 6: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy6Figure :

Derivative Spectra of a Gaussian Absorbance Band

1st Derivative:

2nd Derivative:

Absorbance:

)('

fd

dA

)(fA

)(''2

2

fd

Ad

Page 7: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy7Figure :

Resolution Enhancement

• Overlay of 2 Gaussian bands with a NBW of 40 nm separated by 30 nm

• Separated by 4th derivative

Page 8: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy8Figure :

Transmission and Color

The human eye sees the complementary color to that which is absorbed

Page 9: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy9Figure :

Absorbance and Complementary Colors

Page 10: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy10Figure :

Transmittance and Concentration

The Bouguer-Lambert Law

PathlengthConsteIIT 0/

Page 11: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy11Figure :

Transmittance and Path LengthBeer’s Law

ionConcentratConsteIIT 0/

Concentration

Page 12: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy12Figure :

The Beer-Bouguer-Lambert Law

cbIIIITA /log/loglog 00

Page 13: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy13Figure :

Two-Component Mixture

Example of a two-component mixture with little spectral overlap

Page 14: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy14Figure :

Two-Component Mixture

Example of a two-component mixture with significant spectral overlap

Page 15: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy15Figure :

Influence of 10% Random Error

Influence on the calculated concentrations• Little spectral overlap: 10% Error• Significant spectral overlap: Depends on similarity, can be much higher (e.g. 100%)

Page 16: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy16Figure :

Absorption Spectra of Hemoglobin Derivatives

Page 17: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy17Figure :

Intensity Spectrum of the Deuterium Arc Lamp

• Good intensity in UV range• Useful intensity in visible range• Low noise• Intensity decreases over lifetime

Page 18: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy18Figure :

Intensity Spectrum of the Tungsten-Halogen Lamp

• Weak intensity in UV range• Good intensity in visible range• Very low noise• Low drift

Page 19: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy19Figure :

Intensity Spectrum of the Xenon Lamp

• High intensity in UV range• High intensity in visible range• Medium noise

Page 20: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy20Figure :

Dispersion Devices

• Non-linear dispersion• Temperature sensitive

• Linear Dispersion• Different orders

Page 21: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy21Figure :

Photomultiplier Tube Detector

Anode

• High sensitivity at low light levels• Cathode material determines spectral sensitivity• Good signal/noise• Shock sensitive

Page 22: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy22Figure :

The Photodiode Detector

• Wide dynamic range• Very good signal/noise at high light levels• Solid-state device

Page 23: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy23Figure :

Schematic Diagram of a Photodiode Array

• Same characteristics as photodiodes• Solid-state device• Fast read-out cycles

Page 24: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy24Figure :

Conventional Spectrophotometer

Schematic of a conventional single-beam spectrophotometer

Page 25: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy25Figure :

Diode-Array Spectrophotometer

Schematic of a diode-array spectrophotometer

Page 26: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy26Figure :

Diode-Array Spectrophotometer

Optical diagram of the HP 8453 diode-array spectrophotometer

Page 27: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy27Figure :

Conventional Spectrophotometer

Optical system of a double-beam spectrophotometer

Page 28: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy28Figure :

Diode-Array Spectrophotometer

Optical system of the HP 8450A diode-array spectrophotometer

Page 29: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy29Figure :

Conventional Spectrophotometer

Optical system of a split-beam spectrophotometer

Page 30: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy30Figure :

Definition of Resolution

Spectral resolution is a measure of the ability of an instrument to differentiate between two adjacent wavelengths

Page 31: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy31Figure :

Instrumental Spectral Bandwidth

The SBW is defined as the width, at half the maximum intensity, of the band of light leaving the monochromator

Page 32: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy32Figure :

Natural Spectral Bandwidth

The NBW is the width of the sample absorption band at half the absorption maximum

Page 33: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy33Figure :

Effect of SBW on Band Shape

The SBW/NBW ratio should be 0.1 or better to yield an absorbance measurement with an accuracy of 99.5% or better

Page 34: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy34Figure :

Effect of Digital Sampling

The sampling interval used to digitize the spectrum for computer evaluation and storage also effects resolution

Page 35: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy35Figure :

Wavelength Resettability

Influence of wavelength resettability on measurements at the maximum and slope of an absorption band

Page 36: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy36Figure :

Effect of Stray Light

Effect of various levels of stray light on measured absorbance compared with actual absorbance

Page 37: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy37Figure :

Theoretical Absorbance Error

The total error at any absorbance is the sum of the errors due to stray light and noise (photon noise and electronic noise)

Page 38: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy38Figure :

Effect of Drift

Drift is a potential cause of photometric error and results from variations between the measurement of I0 and I

Page 39: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy39Figure :

Transmission Characteristics of Cell Materials

Note that all materials exhibit at least approximately 10% loss in transmittance at all wavelengths

Page 40: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy40Figure :

Open-topped rectangular standard cell (a) and apertured cell (b) for limited sample volume

Cell Types I

Page 41: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy41Figure :

Cell Types II

Micro cell (a) for very small volumes and flow-through cell (b) for automated applications

Page 42: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy42Figure :

Effect of Refractive Index

Changes in the refractive index of reference and sample measurement can cause wrong absorbance measurements

Page 43: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy43Figure :

Non-planar Sample Geometry

Some sample can act as an active optical component in the system and deviate or defocus the light beam

Page 44: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy44Figure :

Averaging of data points reduces noise by the square root of the number of points averaged

Effect of Integration Time

Page 45: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy45Figure :

• Wavelength averaging reduces also the noise (square root of data points)• Amplitude of the signal is affected

Effect of Wavelength Averaging

Page 46: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy46Figure :

Selection of a wavelength in the slope of a absorption band can increase the dynamic range and avoid sample preparation like dilution

Increasing Dynamic Range

Page 47: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy47Figure :

Scattering causes an apparent absorbance because less light reaches the detector

Scattering

Page 48: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy48Figure :

Scatter Spectra

• Rayleigh scattering: Particles small relative to wavelength• Tyndall scattering: Particles large relative to wavelength

Page 49: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy49Figure :

Absorbance at the reference wavelength must be equivalent to the interference at the analytical wavelength

Isoabsorbance Corrections

Page 50: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy50Figure :

Background modeling can be done if the interference is due to a physical process

Background Modeling

Page 51: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy51Figure :

Corrects for constant background absorbance over a range

Internal Referencing

Page 52: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy52Figure :

Three-Point Correction

• Uses two reference wavelengths• Corrects for sloped linear background absorbance

Page 53: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy53Figure :

Discrimination of Broad Bands

• Derivatives can eliminate background absorption• Derivatives discriminate against broad absorbance bands

Page 54: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy54Figure :

Scatter Correction by Derivative Spectroscopy

Scatter is discriminated like a broad-band absorbance band

Page 55: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy55Figure :

Effect of Fluorescence

The emitted light of a fluorescing sample causes an error in the absorbance measurement

Page 56: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy56Figure :

Acceptance Angles and Magnitude of Fluorescence Error

• Forward optics: Absorbance at the excitation wavelengths are too low• Reversed optics: Absorbance at the emission wavelengths are too low

Page 57: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy57Figure :

Inadequate Calibration

• Theoretically only one standard is required to calibrate• In practice, deviations from Beer’s law can cause wrong results

Page 58: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy58Figure :

Calibration Data Sets

• Forward optics: Absorbance at the excitation wavelengths are too low• Reversed optics: Absorbance at the emission wavelengths are too low

Page 59: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy59Figure :

Wavelength(s) for Best Linearity

• A linear calibration curve is calculated at each wavelength• The correlation coefficient gives an estimate on the linearity

Page 60: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy60Figure :

Wavelength(s) for Best Accuracy

• The quantification results are calculated at each wavelength• The calculated concentration are giving an estimate of the accuracy

Page 61: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy61Figure :

Precision of an Analysis

Precision of a method is the degree of agreement among individual test results when the procedure is applied repeatedly to multiple samplings

Page 62: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy62Figure :

Wavelength(s) for Best Sensitivity

• Calculation of relative standard deviation of the measured values at each wavelength• The wavelength with lowest %RSD likely will yield the best sensitivity

Page 63: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy63Figure :

Wavelength(s) for Best Selectivity

Selectivity is the ability of a method to quantify accurately and specifically the analyte or analytes in the presence of other compounds

Page 64: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy64Figure :

Ideal Absorbance and Wavelength Standards

• An ideal absorbance standard would have a constant absorbance at all wavelengths• An ideal wavelength standard would have very narrow, well-defined peaks

Page 65: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy65Figure :

Ideal Stray Light Filter

An ideal stray light filter would transmit all wavelengths except the wavelength used to measure the stray light

Page 66: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy66Figure :

Holmium Perchlorate Solution

The most common wavelength accuracy standard is a holmium perchlorate solution

Page 67: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy67Figure :

Potassium Dichromate Solution

The photometric accuracy standard required by several pharmacopoeias is a potassium dichromate solution

Page 68: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy68Figure :

Stray Light Standard Solutions

The most common stray light standard and the respectively used wavelengths

Page 69: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy69Figure :

Toluene in Hexane (0.02% v/v)

The resolution is estimated by taking the ratio of the absorbance of the maximum near 269 nm and minimum near 266 nm

Page 70: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy70Figure :

Confirmation Analysis

In confirmation analysis, the absorbance at one or more additional wavelengths are used to quantify a sample

Page 71: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy71Figure :

Spectral Similarity

Comparative plots of similar and dissimilar spectra

Page 72: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy72Figure :

Precision and Accuracy

Precision – Precision + Precision – Precision +

Accuracy – Accuracy – Accuracy + Accuracy +

Page 73: Fundamentals of modern UV-visible spectroscopy Presentation Materials

Fundamentals of modern UV-visible spectroscopy73Figure :

Hydrolysis of Sultone

Wavelength [nm]

Abso

rban

ce [

AU

]