free residual chlorine sensor assembly - automation … rosem… ·  · 2015-02-05the free...

24
Model 450 Free Residual Chlorine Sensor Assembly Instruction Manual PN 51-450/rev.B March 2006

Upload: lethien

Post on 23-Apr-2018

225 views

Category:

Documents


0 download

TRANSCRIPT

Model 450

Free Residual Chlorine Sensor Assembly

Instruction ManualPN 51-450/rev.B

March 2006

CAUTION

SENSOR/PROCESS

APPLICATION COMPATIBILITY

The wetted sensor materials may not

be compatible with process com-

position and operating conditions.

Application compatibility is entirely the

responsibility of the user.

DANGER

HAZARDOUS AREA INSTALLATION

Installations near flammable liquids or in haz-

ardous area locations must be carefully eval-

uated by qualified on site safety personnel.

This sensor is not Intrinsically Safe or

Explosion Proof.

To secure and maintain an intrinsically safe

installation, the certified safety barrier,

transmitter, and sensor combination must

be used. The installation system must com-

ply with the governing approval agency (FM,

CSA or BASEEFA/CENELEC) hazardous

area classification requirements. Consult

your analyzer/transmitter instruction manual

for details.

Proper installation, operation and servicing

of this sensor in a Hazardous Area Instal-

lation is entirely the responsibility of the user.

ESSENTIAL INSTRUCTIONSREAD THIS PAGE BEFORE PRO-

CEEDING!

Rosemount Analytical designs, manufactures, and tests itsproducts to meet many national and international stan-dards. Because these instruments are sophisticated tech-nical products, you must properly install, use, and maintainthem to ensure they continue to operate within their normalspecifications. The following instructions must be adheredto and integrated into your safety program when installing,using, and maintaining Rosemount Analytical products.Failure to follow the proper instructions may cause any oneof the following situations to occur: Loss of life; personalinjury; property damage; damage to this instrument; andwarranty invalidation.

• Read all instructions prior to installing, operating, and

servicing the product. If this Instruction Manual is not the

correct manual, telephone 1-800-654-7768 and the

requested manual will be provided. Save this Instruction

Manual for future reference.

• If you do not understand any of the instructions, contact

your Rosemount representative for clarification.

• Follow all warnings, cautions, and instructions marked

on and supplied with the product.

• Inform and educate your personnel in the proper instal-

lation, operation, and maintenance of the product.

• Install your equipment as specified in the Installation

Instructions of the appropriate Instruction Manual and

per applicable local and national codes. Connect all

products to the proper electrical and pressure sources.

• To ensure proper performance, use qualified personnel

to install, operate, update, program, and maintain the

product.

• When replacement parts are required, ensure that qual-

ified people use replacement parts specified by

Rosemount. Unauthorized parts and procedures can

affect the product’s performance and place the safe

operation of your process at risk. Look alike substitu-

tions may result in fire, electrical hazards, or improper

operation.

• Ensure that all equipment doors are closed and protec-

tive covers are in place, except when maintenance is

being performed by qualified persons, to prevent electri-

cal shock and personal injury.

Emerson Process Management

Liquid Division2400 Barranca Parkway

Irvine, CA 92606 USA

Tel: (949) 757-8500

Fax: (949) 474-7250

http://www.RAuniloc.com

© Rosemount Analytical Inc. 2006

MODEL 450 TABLE OF CONTENTS

MODEL 450

FREE RESIDUAL CHLORINE SENSOR

TABLE OF CONTENTS

Section Title Page

1.0 DESCRIPTION AND SPECIFICATIONS ................................................................ 1

1.1 Features and Applications ....................................................................................... 1

1.2 Specifications........................................................................................................... 1

1.3 Ordering Information................................................................................................ 2

2.0 INSTALLATION....................................................................................................... 3

2.1 General .................................................................................................................... 3

2.2 Inspection and Unpacking ....................................................................................... 3

2.3 Application Requirements........................................................................................ 3

2.4 Preparation for Installation and Startup ................................................................... 3

2.5 Installation................................................................................................................ 4

3.0 TROUBLESHOOTING ............................................................................................ 6

3.1 General .................................................................................................................... 6

3.2 Troubleshooting the Model 450 Sensor .................................................................. 6

4.0 MAINTENANCE ...................................................................................................... 9

4.1 General .................................................................................................................... 9

4.2 Scheduled Preventative Maintenance..................................................................... 9

4.3 Sensor Disassembly, Rebuild and Assembly Procedure ........................................ 9

5.0 PARTS LISTS ......................................................................................................... 11

6.0 RETURN OF MATERIAL ........................................................................................ 15

i

MODEL 450 TABLE OF CONTENTS

MODEL 450

FREE RESIDUAL CHLORINE SENSOR

LIST OF FIGURES

Figure Title Page

1-1 Model 450 Dimensional Data 2

2-1 Model 450-01 and 450-02 Sensor Installation Diagrams 4

2-2 Handrail Mounting Installation for Model 450-03 and 450-04 Sensor Assemblies 5

5-1 Models 450-01 and 450-02 Free Residual Chlorine Flow Through Sensors 11

5-2 Model 450-03 Free Residual Chlorine Submersion Sensor 12

5-3 Model 450-04 Free Residual Chlorine Submersion Sensor/Agitator 13

5-4 Model 450 Free Residual Chlorine Sensor Assembly 14

LIST OF TABLES

Table Title Page

3-1 Temperature Resistance Chart 7

3-2 Model 1054A TFC Quick Troubleshooting Chart 8

1

MODEL 450 SECTION 1.0

DESCRIPTION AND SPECIFICATIONS

SECTION 1.0DESCRIPTION AND SPECIFICATIONS

1.1 FEATURES AND APPLICATIONS

The Rosemount Analytical Model 450 Free ResidualChlorine Sensor, used with the Model 1054A TotalFree Chlorine Analyzer and the Models 1181 or 853free residual chlorine analyzer/transmitter, measuresthe free residual chlorine level in a wide variety ofapplications, such as: Potable water treatment,sewage plant effluent, cooling tower chlorination con-trol, swimming pools, and industrial waste treatmentplant effluent.

The Model 450 sensor is constructed of PVC withpolymer membrane, hardwood liquid junction*, andstainless steel hardware. The sensor is designed forflow (Codes 01 and 02), submersion in a flowing sam-ple (Code 03), and submersion in a stagnant pond(Code 04). All sensors are furnished with an integral20-foot cable. The Code Option 04 incorporates anagitator and impeller, with a 20-foot power cord, foragitation of a stagnant solution so a fresh sample isalways present at the sensor membrane in submer-sion applications.

The membrane retainer** is designed for ease ofinstallation, and provides adequate tension on themembrane to insure an accurate measurement. Inaddition, the electrolyte chamber incorporates a pres-sure compensator to maintain equal pressure on bothsides of the membrane, despite sample pressurechanges. A thermistor or an RTD for automatic tem-perature compensation is incorporated in the tip of thesensor to adjust for temperature change effects on thediffusion rate of free residual chlorine across themembrane.

The sensor requires very little maintenance andrequires no special tools or fixtures.

1.2 SPECIFICATIONS

Temperature: 0 to 50°C (32 to 122°F)

Pressure: 0 to 345 kPa abs (0 to 50 psig)

Material: Housing: PVCHardware: Stainless steelLiquid Junction: HardwoodO-Rings: Buna-NMembrane: PolymerCathode: GoldAnode: Silver

Minimum Sample Flow Rate:Code 01 & 02: 0.13 liter/sec (2 GPM)Code 03: 0.3 meter/sec (1 ft/sec)Code 04: None

Agitator Power Requirement:(Model Code 04 only): 115 VAC ±10% 60Hz (0.2amp)

Weight/Shipping Weight:Code 01, 02 & 03: 0.9 kg/1.8 kg (2lbs/4 lbs)Code 04: 2.3 kg/3.2 kg (5 lbs/7lbs)

• ELIMINATES THE NEED FOR WET CHEMICAL ANALYSIS.

• LOW MAINTENANCE with up to 12 months between rechargings.

• THREE CONFIGURATIONS AVAILABLE to meet most applications.

• COMPACT DESIGN provides for ease of installation and service.

*U.S. Patent No. 4,187,162

**U.S. Patent No. 3,887,194Canadian Patent No. 1,025,496

2

MODEL 450 SECTION 1.0

DESCRIPTION AND SPECIFICATIONS

1.3 ORDERING INFORMATIONModel 450 Free Residual Chlorine Sensor Includes PVC sensor housing with integral chlorine permeablemembrane and temperature compensator.

MODEL450 FREE RESIDUAL CHLORINE SENSOR (3 lbs./1.5 kg)

CODE MOUNTING HARDWARE

01 1-1/2 inch PVC cell, 90° flow

02 1-1/2 inch PVC cell, 180° flow

03 PVC union with 3/4 inch FNPT for submersion service

04 PVC union with 1 inch FNPT, and agitator with 20-ft. cable for submersions. Agitator motor requires 115 VAC power only.

450 01 17 EXAMPLE

CODE OPTIONS

11 Stainless steel tags (specify marking)

14 Handrail mounting bracket (P/N 1000857) for use with J-box (see Code 17 below)

15 Handrail mounting bracket (P/N 1000856) for use with Model 803

17 Weatherproof NEMA 4X junction box (P/N 22719-02)

54 For use with Models 1054A TFC

NOTE: Recommended cable from sensor to transmitter is Belden 8434 or equivalent, available from Rosemount Analytical (P/N 9200074).

Specify length.

FIGURE 1-1. MODEL 450 DIMENSIONAL DATA

*CODE 02 SAME AS CODE 01 EXCEPT SENSOR MOUNTED IN TOP OF TEE, WITH FLOW IN EITHER DIRECTION

3

MODEL 450 SECTION 2.0

INSTALLATION

SECTION 2.0

INSTALLATION

2.1 GENERAL. This section provides installationinstructions for the Model 450 Sensor Assembly.

2.2 INSPECTION AND UNPACKING. Inspect the ship-ping container for any evidence of damage. If damageis evident, notify the carrier immediately. If no damageis evident, open the container and inspect the sensor.Save the container for shipment to the factory for serv-ice.

2.3 APPLICATION REQUIREMENTS. The Model 450is designed for typical clean water application. It isNOT designed for the following:

1. Processes containing ammonia (municipal waste water).

2. Processes containing sea water or bromine chem-ical additions.

3. Low conductivity processes, less than 50 micro-siemens (reverse osmosis).

4. Post-dechlorination processes (example: afteradding sodium bisulfite).

2.4 SENSOR PREPARATION FOR INSTALLATIONAND START-UP. The plumbing fittings provided on theModel 450 Sensor differ slightly, depending on the pipingrequirements. However, the main body, or chlorine sens-ing portion of each of the different codes, is identical. Eachsensor is shipped from the factory dry and without amembrane installed. The recharge kit which accompa-nies the sensor contains one bottle of dry electrolyte andfive membranes. The sensor should be charged utilizingthe following procedure:

CAUTIONClean rubber gloves should be worn duringsensor preparation and maintenance toavoid skin contact and contaminating themembrane and cathode.

1. Prepare sensor electrolyte by filling the bottle ofcrystals supplied with the recharge kit (Ag Cl/KCI)to the neck with distilled or deionized water. Stir orshake to wet crystals thoroughly. Some crystalsshould remain undissolved.

NOTEPlease see Figure 5-4 for Model 450 FreeResidual Sensor assembly exploded view.

2. Remove sensor “cover” and membrane “Retainer”(remove 6 screws at sensor tip).

3. Sand the gold cathode with the 400 grit sandpa-per, in one direction only until it is shiny. Rinsecathode with clean water.

4. Remove the fill-plug.

5. Fill the syringe (included in the recharge kit) withthe AgCl/KCl electrolyte solution. Fill the sensorbody with the electrolyte, shaking the sensor toremove air. Ensure that about 1 cc (or 1 ml) ofcrystals get inside the sensor body.

6. Install the fill plug (wrapped with one or two turnsof TEFLON®1 tape in the side of the sensor. As theplug is screwed into place, electrolyte will beforced through the wood junction. This is impor-tant, since the electrical continuity will not beestablished unless the wood junction is thorough-ly wetted.

Continue to screw the fill plug into place until it is flushwith the sensor body. Do not remove excess electrolytefrom the end of the sensor because it will wet the mem-brane and help hold it in place during installation.

7. Place the O-ring in the groove and press it into posi-tion around the wood junction. If there is no solutionremaining over the gold cathode after installing the O-ring, add a few drops of electrolyte (no crystals).IMPORTANT: No crystals underneath the mem-brane or damage will occur.

8. Place the white-opaque “membrane” on top of thecathode. Place the white-translucent “protector”over the membrane. DO NOT use the blue-paperprotective separators found in the membrane kit!

9. Lay the membrane “retainer” over the membraneand protector. Screw down the (6) screws applyingeven pressure. The membrane should be taut, notspongy, when the retainer is secure.

10. Keep the sensor tip in water if installation is delayedover an hour. Membrane may adhere to the cath-ode if allowed to dry out.

11. Sensor requires approximately (2) hours to polar-ize and (12-24) hours to obtain a stable zero inchlorine-free water.

NOTEWater should be agitated for zeroing andspanning (magnetic stirrer is recommended)

CAUTIONInsufficient zeroing time may cause diffi-culties in calibration and readings thatdrift.

1 Reg. U.S. Pat. Office for DuPont's fluorocarbon resins

4

MODEL 450 SECTION 2.0

INSTALLATION

12. Chlorine concentrations of 1 PPM or above are rec-ommended for the most accurate chlorine calibration.IMPORTANT: Process flow requirement is 1 ft./sec.or if in a “Tee” 2.0-5.0 gal./min. If unable to adjustprocess chlorine concentration to this level, performcalibration in a beaker with a magnetic stirrer for agi-tation. Adding a drop or two of chlorine bleach todeionized water (buffered to 7 pH) will create a chlo-rine residual.

13. Calibrate the chlorine reading to the titrated valueof the process or the chlorine residual sample.

14. The sensor is now ready to be installed.

2.5 INSTALLATION.

2.5.1 Mechanical Installation.

1. Model 450-01 and 450-02 Sensor Assemblies.These sensors are available in two configurations:Angle flow (Code 01) and straight flow (Code 02).The PVC tee is for plumbing into 1-1/2 inch PVC(SCH 80) tubing. Direction of flow should be asshown in Figure 2-1.

2. Models 450-03 and 450-04 Sensor Assemblies.These sensors are for submersion applications. If ahandrail mounting bracket (Code 14) is to be used,refer to Figure 2-2 for installation details. The Model450-03 Sensor Assembly should be installed in astream having a minimum flow of 1.5 ft./sec. toobtain rated response time. Attach the union adapterto a 3/4 inch FPT PVC (SCH 80) tube or metal pipe.If PVC to metal connection is made, use RTV sili-cone rubber adhesive/sealant (RosemountAnalytical P/N 9220106 or 9220018). Check the con-nection frequently for looseness, until no more loose-ness is observed. The sensor may be submerged toa depth of up to 6.1 meters (20 feet).

CAUTIONUse extreme care when attaching PVC tubing orcomponents to metal pipe. Over-tightening cancause damage to the PVC or components, sotighten only until pressure is felt.

3. Model 450-04 Sensor Assembly. The Model 450-04Sensor Assembly is for installation in processeswhere insufficient flow or agitation exists. Attach theunion adapter to a 1 inch FPT PVC (SCH 80) tubeor metal pipe. If PVC to metal connection is made,refer to Paragraph 2 and the CAUTION which fol-lows. The sensor may be submerged to a depth ofup to 6.1 meters (20 feet). The agitator cable iswater tight and should be fastened to the standpipewith tiewraps every five feet.

2.5.2 Electrical Installation. All sensors are wired thesame. Connect the sensor wiring as follows:

NOTEThe sensor is supplied with 20 feet of cable. Ifmore is needed, use Belden 8434 cable orequivalent, available from RosemountAnalytical as P/N 9200074 (specify length).Junction box (P/N 22719-02) may be used forconnection if extended length of cable is used.

FIGURE 2-1. Model 450-01 and 450-02 Sensor Installation Diagrams

Transmitter1054RCI/

Sensor Lead 1054ARCI/1054ATFC 1181RCI 853RCI

Shield (Guard) to TB2-6 TB2-2 TB1-15

White (TC) to TB2-5 TB2-1 TB1-18

Green (TC) to TB2-7 TB2-3 TB1-17

Red (Anode) to TB2-1 TB2-4 TB1-16

Black (Cathode) to TB2-3 TB2-3 TB1-14

5

MODEL 450 SECTION 2.0

INSTALLATION

FIGURE 2-2. Handrail Mounting Installation for Model 450-03 and 450-04 Sensor Assemblies

WHEN INCH AND METRIC DIMS

ARE GIVEN

MILLIMETER

INCH

DWG. NO. REV.

40045019 A

6

3.1 GENERAL. This section contains troubleshooting

data for the Model 450 Sensor Assembly.

3.2 TROUBLESHOOTING THE MODEL 450

SENSOR. The majority of problems encountered in

RCI systems are caused by improper maintenance, or

from internal leakage within the sensor.

Disassemble the sensor completely by removing the

membrane retainer and sensor body (see Section 4.3

and Figure 5-4). The membrane may have dried on to

the gold cathode. Soak this loose with water as scraping

will damage the gold tip. A wet wood junction will make

the sensor body very difficult to remove but a firm twist-

ing, pulling motion should loosen it.

Clean all parts with clean water and dry them. Check for

visible damage (i.e., cracks, deep cuts, broken silver

wires). Perform the following checks with an ohmmeter.

Check T.C. resistance between the green and white

wires. Check continuity between the red wire and the sil-

ver anode, and between the black wire and the gold

cathode (be careful not to scratch the gold). Perform a

high meg check between the following (100 megohms

minimum): Shield; Black; Red; Green or White. If every-

thing checks O.K., the sensor should function after a

proper recharge. If the silver anode appears oxidized, it

may be cleaned with wet or dry 400 sandpaper. Also

brush the gold tip two or three times with 400 sandpaper

in one direction only.

Assemble the body of the RCI sensor and calibrate per

Section 2.4. NOTE

BENCH LOOP CHECK. A bench check of

the sensor in a beaker of distilled water,

with a capful of 7 pH buffer added to it.

Place beaker and sensor on a ring stand

with a magnetic stirrer to create flow. Allow

the sensor to zero for a minimum of 12-24

hours as specified. Caution: zeroing too

early will distort scaling and confuse the

calibration procedure

MODEL 450 SECTION 3.0

TROUBLESHOOTING

SECTION 3.0

TROUBLESHOOTING

WAIT FOR SENSOR TO ZERO OUT. It is

typical for the “Reverse Current Input (rci)”

message to come up while zeroing. When

the input current drops below zero it trig-

gers this message to be displayed. It is also

normal for the chlorine readings to rise and

fall, while zeroing. The input current from

the sensor may even be negative at times.

Add some drops of chlorine to check for a

response. To determine accuracy you

should maintain a pH of 7 and check the

chlorine level with a free chlorine test such

as DPD-50.

NOTE

Record the current that is developed by the

Model 450 sensor. If using a 1054A RC or

1054 A TFC, the input current can be

viewed by selecting “jo” under the “Set

Menu”. To access the “Set Menu” press the

ACCESS/ENTER key twice in rapid suc-

cession. Then press the up-arrow key until

“jo” is displayed then press SELECT key.

The current from the sensor will then be

displayed. NOTE: If using a 1054A RC the

reading is in microamps. If using a 1054A

TFC the reading is in nanoamps. (1000

nanoamps = 1 microamp). If using an 1181

RC, place an ammeter in series with the

sensor's black wire to read the current.

Convert the current reading to nanoamps.

Now divide the nanoamp reading, from the

sensor, by the chlorine ppm concentration

to determine the nanoamps per ppm value

of the sensor. Example:

200 nanoamps ÷ 2.0 ppm = 100 nanoamps

per ppm

A good nanoamp per ppm value is between

70 and 135. If value drops below 70, the

sensor should be rebuilt and the cathode

sanded.

CAUTION

Care should be taken not to damage the

TEFLON®1 membrane.

450Sensor Magnetic

Stirrer

RingStand

7

MODEL 450 SECTION 3.0

TROUBLESHOOTING

If the resistance as specified above cannot be

achieved, replace the sensor.

4. To check for shorted sensor leads, measure the

resistance between one temperature compensator

wire (green or white) and the cathode (black) lead

and anode (red) lead. The meter should indicate

an open circuit (100 megohms or more). Repeat

the measurement using the other temperature

compensator wire. If the sensor shows a short cir-

cuit, the sensor is defective and should be

replaced.

1 Reg. U.S. Pat. Office for DuPont's fluorocarbon resins

1. Low readings and inability to calibrate are gener-

ally the result of a coated membrane. Replace the

coated membrane and return the sensor to serv-

ice.

2. Readings offscale and inability to calibrate are

usually caused by a damaged membrane. If this is

the case, replace the membrane.

3. To check the resistance of the temperature com-

pensator (thermistor or PT100 RTD), connect an

ohmmeter to the green and white leads of the sen-

sor. The resistance should be as indicated at the

temperature listed below:

TABLE 3-1. Temperature Resistance Chart

Sensor Resistance (ohms)Temp. PT 100 Thermistor

Mdl 450-54 Mdl 450

0°C 100.00 371.40K

10°C 103.90 214.50K

20°C 107.70 128.00K

25°C 109.62 100.00K

30°C 111.67 78.00K

40°C 115.54 49.80K

50°C 119.40 32.36K

NOTE

Allow enough time for the T.C. element to

stabilize to the surrounding temperature.

MODEL 450 SECTION 3.0

TROUBLESHOOTING

TABLE 3-2. Model 1054A TFC Microprocessor Analyzer Quick Troubleshooting Chart

Symptoms Action

Fault Code �sdj� 1. Typical message during the zeroing process; if sensor is wired correctly. Messageshould go away as the sensor zeros out (12-24 hours).

Fault Code �cbe� 1. Too littIe current to calibrate to: verify that chlorine concentration is a minimum of1-2 ppm. Increase the chlorine concentration if it is below I ppm.

2. Verify that the chlorine sensor is putting out at least 70 nanoamps per ppm;check �jo� under the SEt menu.

3. Recharge the chlorine sensor, ensure the gold cathode is sanded and a newmembrane is installed.

Fault Code �udI�0�udM� 1. Check wiring between the sensor and analyzer.

2. Verify analyzer’s temperature reading has been standardized properly (250 mayhave been entered instead of 25 degrees)

3. Use the MANUAL temperature compensation mode, to get by if the sensors T.C. is bad.

4. Replace the chlorine sensor, if the T.C is bad.

Fault Code �Pso0�Fdj� 1. Remove sensor from process; clean & check membrane.

2. Replace membrane if suspected of being damaged.

3. Recalibrate.

Fault Code �TFo� 1. Check wiring at TB2-6 and TB2-7 particularly; sense wire and shield must be connected correctly.

2. Check for bad lug crimps or open wires, from sensor.

Chlorine Sensor won’t zero 1. Check sensor membrane for damage (pin-hole).

2. Allow sufficient time for sensor to zero: 12-24 hrs.

3. Perform Sensor Electrical Checks.

4. Verify Analyzer: perform Electronic Bench Check.

Chlorine Sensor has little or 1. Verify proper flow rate past sensor minimum 1 ft/sec.no response to chlorine 2. Clean and check sensor membrane: on the Model 450 Sensor make sure that

thechanges blue paper separator has not been installed with the white membrane and/or the

translucent protector.

3. Verify electrolyte is fresh (change every 5-6 months).

4. Perform Chlorine Sensor Electrical Checks.

5. Rebuild sensor if Electrical Checks are OK.

6. If the Analyzer passes the Electronics Bench Check and sensor continues to failafter rebuilding; replace sensor.

Chlorine reading is all 1. Reading will typically fluctuate when a sensor is charged and put in chlorine-free

over the place water to polarize and zero out. A stable reading, near zero, should occur in 12-24hours.

2. If using automatic pH correction, verify that the pH reading is not fluctuating, thiswould cause the chlorine reading to fluctuate. Place analyzer in Manual pH cor-rection mode and check for a stable chlorine reading.

3. Perform pH Checks if the pH reading is fluctuating.

4. Verify analyzer. Perform the Electronic Bench Check.

5. Perform Chlorine Sensor Checks.

6. Rebuild chlorine sensor, if Sensor Checks are OK

8

9

MODEL 450 SECTION 4.0

MAINTENANCE

SECTION 4.0

MAINTENANCE

4.1 GENERAL. This section provides maintenance

instructions for keeping the sensor assembly in good

operating condition.

4.2 SCHEDULED PREVENTATIVE MAINTENANCE.

For best results the sensor should be clean. Examine

the sensor at periodic intervals to make sure it is clean.

Determine this interval period by examining the sensor

after seven days service, and progressively longer

periods, until recalibration or cleaning of the sensor is

required.

Clean the sensor by washing with clean water, or if it is

coated, wipe gently with soft tissue or cloth.

CAUTION

Make sure the TEFLON membrane is not

damaged during cleaning.

4.3 SENSOR DISASSEMBLY, REBUILD AND

ASSEMBLY PROCEDURE. Recharge the sensor and

replace the membrane as follows:

CAUTION

Clean rubber gloves should be worn during

sensor preparation and maintenance to

avoid skin contact and contaminating the

membrane and cathode.

1. Prepare sensor electrolyte from sensor recharge

kit (P/N 2002402) by filling the bottle of crystals

supplied with the recharge kit (AgCl/KCl) to the

neck with distilled or deionized water. Stir or shake

to wet crystals thoroughly. Some crystals should

remain undissolved.

NOTEPlease see Figure 5-4 for Model 450 FreeResidual Sensor assembly exploded view.

2. Remove sensor “cover” and membrane “Retainer”(remove 6 screws at sensor tip).

3. Sand the gold cathode with the 400 grit sandpa-per, in one direction only until it is shiny. Rinsecathode with clean water.

NOTEThe gold element is very thin and caremust be taken that penetration through theelement does not occur.

4. Remove the fill-plug.

5. Fill the syringe (included in the recharge kit) withthe AgCl/KCl electrolyte solution. Fill the sensorbody with the electrolyte, shaking the sensor toremove air. Ensure that about 1 cc (or 1 ml) ofcrystals get inside the sensor body.

6. Install the fill plug (wrapped with one or two turnsof TEFLON®1 tape in the side of the sensor. As theplug is screwed into place, electrolyte will beforced through the wood junction. This is impor-tant, since the electrical continuity will not beestablished unless the wood junction is thorough-ly wetted.

Continue to screw the fill plug into place until it is flushwith the sensor body. Do not remove excess elec-trolyte from the end of the sensor because it will wetthe membrane and help hold it in place during instal-lation.

7. Place the O-ring in the groove and press it into posi-tion around the wood junction. If there is no solutionremaining over the gold cathode after installing the O-ring, add a few drops of electrolyte (no crystals).IMPORTANT: No crystals underneath the mem-brane or damage will occur.

8. Place the white-opaque “membrane” on top of thecathode. Place the white-translucent “protector”over the membrane. DO NOT use the blue-paperprotective separators found in the membrane kit!

9. Lay the membrane “retainer” over the membraneand protector. Screw down the (6) screws applyingeven pressure. The membrane should be taut, notspongy, when the retainer is secure.

10. Keep the sensor tip in water if installation is delayedover an hour. Membrane may adhere to the cath-ode if allowed to dry out.

1 Reg. U.S. Pat. Office for DuPont's fluorocarbon resins

10

MODEL 450 SECTION 4.0

MAINTENANCE

11. Sensor requires approximately (2) hours to polar-ize and (12-24) hours to obtain a stable zero inchlorine-free water.

NOTEWater should be agitated for zeroing andspanning (magnetic stirrer is recommended)

CAUTIONInsufficient zeroing time may cause diffi-culties in calibration and readings thatdrift.

12. Chlorine concentrations of 1 PPM or above are rec-ommended for the most accurate chlorine calibration.

IMPORTANTProcess flow requirement is 1 ft./sec. or if in a“Tee” 0.5-5.0 gal./min. If unable to adjust processchlorine concentration to this level, perform cali-bration in a beaker with a magnetic stirrer for agi-tation. Adding a drop or two of chlorine bleach todeionized water (buffered to 7 pH) will createa chlorine residual.

13. Calibrate the chlorine reading to the titrated value ofthe process or the chlorine residual sample.

14. The sensor is now ready to be installed.

15. Refer to the Model 1054A TFC, 1181 RC or 853instruction manual for standardization procedure.

11

Figure 5-1 Model orItem No. Part No. Description Qty.

Model 450 Sensor Assembly (Models 1181 RCl and 853 RCI)11

Model 450-54 Sensor Assembly (Models 1054 RCl, 1054A RCl and 1054A TFC)

2 2002566 1¼ inch Tee Assy. - Angle Flow (Code 01) 1

3 2002567 1¼ inch Tee Assy. - Straight Flow (Code 02) 1

4 9320057 Union Assembly 1

5 9550090 O-ring (Buna-N, 2-029) 1

6 ––– 1¼ inch Tee (Straight Flow) 1

7 ––– 1¼ inch Tee (Angle Flow) 1

MODEL 450 SECTION 5.0

PARTS LISTS

SECTION 5.0

PARTS LISTS

GENERAL. This section contains the replacement parts

lists and illustrations for the Model 450 Sensor Assembly.

Parts are keyed with an item number. To locate a part,

look on the illustration for that parts next higher assembly

and look for the part. When you have found the part, fol-

low the arrow to the item number. This will give you the

part number, description and quantity of that part on the

accompanying parts list. The Usage Code column, where

used, identifies parts used on more than one assembly.

Determine which assembly you are concerned with and

then check its code letter in the Usage Code column. All

items with that code letter are used on that assembly. If

there is no code letter, the item is used on all assemblies

in that parts list.

Figure 5-1.

Models 450-01 and 450-02 Free Residual Chlorine Flow Through Sensors

12

Figure 5-2 Model orItem No. Part No. Description Qty.

Model 450 Sensor Assembly (Models 1181 RCl and 853 RCl)11

Model 450-54 Sensor Assembly (Models 1054 RCl, 1054A RCl and 1054A TFC)

2 2002568 Union/Gasket Assy 1

3 3001617 Gasket 1

Figure 5-2.

Model 450-03 Free Residual Chlorine Submersion Sensor

MODEL 450 SECTION 5.0

PARTS LISTS

13

Figure 5-3 Model orItem No. Part No. Description Qty.

Model 450 Sensor Assembly (Models 1181 RCI and 853 RCl)11

Model 450-54 Sensor Assembly (Models 1054 RCl, 1054A RCl and 1054A TFC)

2 2002569 Pump, submersible Assy. 1

3 Screw, SPH SS (#8-32 X 3/8) 4

4 Screw, SPH SS (#8-32 X 3/8) 2

5 Holder Assy 1

6 ––– Union 1

7 3001197 Gasket 1

Figure 5-3.

Model 450-04 Free Residual Chlorine Submersion Sensor/Agitator

MODEL 450 SECTION 5.0

PARTS LISTS

14

Figure 5-4 Model orItem No. Part No. Description Qty.

1 ––– Screw (#4-40 X 1.0) 4

2 ––– Base Assy. Sensor (Model 450/450-54) 1

3 9550144 O-ring (Viton®1,2-222) 1

4 9550056 O-ring (Viton®1, 2-210) 1

5 3002010 Junction, wood 1

6 ––– Body, sensor 1

7 33209-00 Plug, fill 1

8 9550110 O-ring (Viton®1, 2-016) 1

9 2001996 Membrane kit contains five membrane, five polyester filter and

two O-rings (Item 8). Discard blue paper between mesh and membrane. 1

10 ––– Retainer, membrane 1

11 ––– Screw (#4-40 X 0.37) 6

12 ––– Cap, protective 1

2002402 Kit, recharge, chlorine sensor contains the following items: 1

3002010 Junction, wood (same as Item 5) 1

2001996 Membrane kit (same as Item 9) 1

9210001 Solution (AgCI/CKI) 1

––– Syringe (for filling sensor) 1

Figure 5-4. Model 450 Free Residual Chlorine Sensor Assembly

MODEL 450 SECTION 5.0

PARTS LISTS

1 Viton is a registered trademark of DuPont Performance Elastomers.

NOTE

blue protective separator between mem-

brane and mesh is to be discarded

15

Model 450 SECTION 6.0

RETURN OF MATERIAL

SECTION 6.0RETURN OF MATERIAL

6.1 GENERAL. To expedite the repair and return of

instruments, proper communication between the cus-

tomer and the factory is important. A return material

authorization number is required. Call 1-800-654-

7768 or (949) 757-8500. The “Return of Materials

Request” form is provided for you to copy and use in

case the situation arises. The accuracy and complete-

ness of this form will affect the processing time of your

materials.

6.2 WARRANTY REPAIR. The following is the proce-

dure for returning products still under warranty.

1. Contact the factory for authorization.

2. Complete a copy of the “Return of Materials

Request” form as completely and accurately as

possible.

3. To verify warranty, supply the factory sales order

number or the original purchase order number. In

the case of individual parts or sub-assemblies, the

serial number on the mother unit must be sup-

plied.

4. Carefully package the materials and enclose your

“Letter of Transmittal” and the completed copy of

the “Return of Materials Request” form. If possi-

ble, pack the materials in the same manner as it

was received.

IMPORTANT

Please see second section of “Return of

Materials Request Form”. Compliance to

the OSHA requirements is mandatory for

the safety of all personnel. MSDS forms

and a certification that the instruments

have been disinfected or detoxified are

required.

5. Send the package prepaid to:

Rosemount Analytical Inc.

2400 Barranca Parkway

Irvine, CA 92606

Attn: Factory Repair

Mark the package:

Returned for Repair RMA No. _______________

Model No. ______________

6.3 NON WARRANTY REPAIR.

1. Contact the factory for authorization.

2. Fill out a copy of the “Return of Materials Request”

form as completely and accurately as possible.

3. Include a purchase order number and make sure

to include the name and telephone number of the

right individual to be contacted should additional

information be needed.

4. Do Steps 4 and 5 of Section 6.2.

NOTE

Consult the factory for additional infor-

mation regarding service or repair.

MODEL CFA 3000 RETURN OF MATERIAL

RETURN OF MATERIAL

GENERAL.

To expedite the repair and return of instruments, proper communication between the customer and the factory is

important. Before returning a product for repair, call 1-949-757-8500 for a Return Materials Authorization (RMA)

number.

WARRANTY REPAIR.

The following is the procedure for returning instruments still under warranty:

1. Call Emerson Process Management for authorization.

2. To verify warranty, supply the factory sales order number or the original purchase order number. In the case

of individual parts or sub-assemblies, the serial number on the unit must be supplied.

3. Carefully package the materials and enclose your “Letter of Transmittal” (see Warranty). If possible, pack the

materials in the same manner as they were received.

4. Send the package prepaid to:

Emerson Process Management

Liquid Division

2400 Barranca Parkway

Irvine, CA 92606

Attn: Factory Repair

RMA No. ____________

Mark the package: Returned for Repair

Model No. ____

NON-WARRANTY REPAIR.

The following is the procedure for returning for repair instruments that are no longer under warranty:

1. Call Emerson Process Management for authorization.

2. Supply the purchase order number, and make sure to provide the name and telephone number of the individ-

ual to be contacted should additional information be needed.

3. Do Steps 3 and 4 of the Warranty Repair section above.

NOTE

Consult the factory for additional information regarding service or repair.

A Worldwide Network of Sales and Service

Emerson Process Management’s field sales offices are your source for more information on the fill line of Rosemount Analyticalproducts. Field sales personnel will work closely with you to supply technical data and application information.

For more information, please contact your nearest Emerson Process Management sales office.

Immediate, Reliable Analytical Support Now there’s a way to quickly get the right answers for your liquid analytical instrumentation ques-tions: the Analytical Customer Support Center.

Our staff of trained professionals is ready to provide the information you need. If you are placing anorder, verifying delivery, requesting application information, or just want to contact a RosemountAnalytical representative, a call to the Customer Support Center will provide you with the right peo-ple, the right answers, right now.

The right people, the right answers, right now.

THE AMERICAS -HEADQUARTERS

Emerson Process Management

Rosemount Analytical Inc.

Liquid Center of Excellence

2400 Barranca Parkway

Irvine, CA 92606

Phone: +1.949.757.8500

Toll Free: +1.800.854.8257

Fax: +1.949.474.7250

ASIA-PACIFIC

Emerson Process Management

Asia Pacific Private Ltd.

1 Pandan Crescent

Singapore 0512

Republic of Singapore

Phone: 65.777.8211

Fax: 65.777.0947

EUROPE

Emerson Process Management AG

Blegistrasse 21

CH-6341 Baar-Walterswil

Switzerland

T 41.41.768.6111

T 41.41.761.8740

VISIT OUR WEBSITE ATwww.rosemountanalytical.com

GERMANY

Emerson Process Management Process

Gas Analyzer Center of Excellence

GmbH & Co. OHG

Industriestrasse 1

63594 Hasselroth

Germany

T 49.6055.884.0

F 49.6055.884.20

LATIN AMERICA

Emerson Process Management

Rosemount Analytical

10241 West Little York, Suite #200

Houston, TX 77040 USA

T 713.467.6000

F 713.827.3328

MIDDLE EAST AND AFRICA

Emerson Process Management

EPM Building

P. O. Box 17033

Jebe Ali Free Zone

Dubai, United Arab Emirates

T 971.4.8835235

F 971.4.8835312

FROM: RETURN BILL TO:

_____________________________ _____________________________ _____________________________

_____________________________ _____________________________ _____________________________

_____________________________ _____________________________ _____________________________

CUSTOMER/USER MUST SUBMIT MATERIAL SAFETY SHEET (MSDS) OR COMPLETE STREAM COMPOSITION, AND/OR

LETTER CERTIFYING THE MATERIALS HAVE BEEN DISINFECTED AND/OR DETOXIFIED WHEN RETURNING ANY PROD-

UCT, SAMPLE OR MATERIAL THAT HAVE BEEN EXPOSED TO OR USED IN AN ENVIRONMENT OR PROCESS THAT CON-

TAINS A HAZARDOUS MATERIAL ANY OF THE ABOVE THAT IS SUBMITTED TO ROSEMOUNT ANALYTICAL WITHOUT

THE MSDS WILL BE RETURNED TO SENDER C.O.D. FOR THE SAFETY AND HEALTH OF OUR EMPLOYEES. WE THANK

YOU IN ADVANCE FOR COMPLIANCE TO THIS SUBJECT.

SENSOR OR CIRCUIT BOARD ONLY:

(Please reference where from in MODEL / SER. NO. Column)

1. PART NO.__________________________1. MODEL_________________________________1. SER. NO.________________

2. PART NO.__________________________2. MODEL_________________________________2. SER. NO.________________

3. PART NO.__________________________3. MODEL_________________________________3. SER. NO.________________

4. PART NO.__________________________4. MODEL_________________________________4. SER. NO.________________

PLEASE CHECK ONE:

�� REPAIR AND CALIBRATE �� DEMO EQUIPMENT NO. __________________________

�� EVALUATION �� OTHER (EXPLAIN) _______________________________

�� REPLACEMENT REQUIRED? �� YES �� NO _________________________________________________

DESCRIPTION OF MALFUNCTION:

______________________________________________________________________________________________________

______________________________________________________________________________________________________

______________________________________________________________________________________________________

WARRANTY REPAIR REQUESTED:

�� YES-REFERENCE ORIGINAL ROSEMOUNT ANALYTICAL ORDER NO. ________________________________________

CUSTOMER PURCHASE ORDER NO. _________________________________________________

�� NO-PROCEED WITH REPAIRS-INVOICE AGAINST P.O. NO. _________________________________________________

�� NO-CONTACT WITH ESTIMATE OF REPAIR CHARGES: LETTER �� __________________________________________

PHONE �� ___________________________________________

NAME ____________________________________________________ PHONE _________________________________________

ADDRESS ___________________________________________________________________________________________________

______________________________________________________________ ZIP _________________________________________

RETURN AUTHORITY FOR CREDIT ADJUSTMENT [Please check appropriate box(s)]

�� WRONG PART RECEIVED �� REPLACEMENT RECEIVED

�� DUPLICATE SHIPMENT REFERENCE ROSEMOUNT ANALYTICAL SALES ORDER NO.__________

�� RETURN FOR CREDIT RETURN AUTHORIZED BY: ______________________________________

WARRANTY DEFECT____________________________________________________________________________________

_____________________________________________________________________________________________________

24-6047

RETURN OF MATERIALS REQUEST •IMPORTANT!This form must be completed to ensure expedient factory service.

REPAIR

STATUS

REASON

FOR

RETURN

CUSTOMER

NOTICE

TO

SENDER

Emerson Process Management

Liquid Division2400 Barranca Parkway

Irvine, CA 92606 USA

Tel: (949) 757-8500

Fax: (949) 474-7250

http://www.RAuniloc.com

© Rosemount Analytical Inc. 2006

WARRANTY

Seller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufactured

or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira-

tion of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen

(18) months from the date of shipment by Seller, whichever period expires first. Consumables, such as glass electrodes,

membranes, liquid junctions, electrolyte, o-rings, catalytic beads, etc., and Services are warranted for a period of 90

days from the date of shipment or provision.

Products purchased by Seller from a third party for resale to Buyer ("Resale Products") shall carry only the warranty extended by

the original manufacturer. Buyer agrees that Seller has no liability for Resale Products beyond making a reasonable commercial

effort to arrange for procurement and shipping of the Resale Products.

If Buyer discovers any warranty defects and notifies Seller thereof in writing during the applicable warranty period, Seller shall, at

its option, promptly correct any errors that are found by Seller in the firmware or Services, or repair or replace F.O.B. point of man-

ufacture that portion of the Goods or firmware found by Seller to be defective, or refund the purchase price of the defective por-

tion of the Goods/Services.

All replacements or repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuit-

able environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other

cause not the fault of Seller are not covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obli-

gated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance by

an authorized Seller representative. All costs of dismantling, reinstallation and freight and the time and expenses of Seller's per-

sonnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller.

Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty peri-

od or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be amended only

in a writing signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement, THERE

ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FIT-

NESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF THE GOODS OR SERVICES.

RETURN OF MATERIAL

Material returned for repair, whether in or out of warranty, should be shipped prepaid to:

Emerson Process Management

Liquid Division

2400 Barranca Parkway

Irvine, CA 92606

The shipping container should be marked:

Return for Repair

Model _______________________________

The returned material should be accompanied by a letter of transmittal which should include the following information (make a

copy of the "Return of Materials Request" found on the last page of the Manual and provide the following thereon):

1. Location type of service, and length of time of service of the device.

2. Description of the faulty operation of the device and the circumstances of the failure.

3. Name and telephone number of the person to contact if there are questions about the returned material.

4. Statement as to whether warranty or non-warranty service is requested.

5. Complete shipping instructions for return of the material.

Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional charges

for inspection and testing to determine the problem with the device.

If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.

Credit Cards for U.S. Purchases Only.

The right people,the right answers,right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE

http://www.raihome.com

Specifications subject to change without notice.

Emerson Process Management

Liquid Division2400 Barranca Parkway

Irvine, CA 92606 USA

Tel: (949) 757-8500

Fax: (949) 474-7250

http://www.raihome.com

© Rosemount Analytical Inc. 2006