field notes #2 validation of a quality control method for ... · 7/1/2017 · field notes #2...
TRANSCRIPT
Field Notes #2 Validation of a Quality Control Method
For a Clinical PET Tracer
Amy L. Vāvere, PhD Head Radiochemist
Department of Diagnostic Imaging St. Jude Children’s Research Hospital
Memphis, TN
Highlights
• Release Criteria for [11C]Methionine – routine clinical tracer at our institution
• Details of QC Testing for [11C]MET
• Issues with QC of [11C]MET along the way...and how we overcame them
• Best Practices in QC of MET – pointers and helpful tools
• Issues with QC of [18F]Fluorodopamine
L-[methyl-11C]Methionine Synthesis
• Trapping of [11C]CH3I bubbled through of chilled acetone (< 0°C).
• L-homocysteine thiolactone HCl dissolved in 32 mM NaOH (0.1 mg @ 2 mg/mL) is added to the reactor vial, sealed and heated to 100°C.
• Acetone is evaporated with heat and argon flow and solution is neutralized with 10 mM acetic acid and passed through a C18 Sep-Pak® Plus followed by sterile water.
• Average yield is 25-30% (uncorrected) in 17-25 minutes.
HSOH
NH2
OS
OHNH2
O
H311C
[11C]CH3I
100°C
Quality Control – 20 minutes Test Acceptance Criteria
Appearance and Color Clear and Colorless Radiochemical purity &
identity ≥ 90% C-11 Methionine
RT within 15% of standard Chemical Purity
(annually) ≤ 4% D-methionine
Radionuclidic Purity (annually) ≥ 95%
Volatile Organic Impurities
Acetone < 0.5% Tetrahydrofuran < 0.072%
Ethanol < 0.5% Diacetone Alcohol < 0.04%
pH 4.5 – 7.5 Filter Integrity ≥ 50 psi breaking pressure
Radionuclidic Identity 18.36 to 22.44 min (20.4 ± 10%)
Bacterial endotoxin (LAL) ≤ 5 EU/mL
Sterility No colony growth out to 14 days
Quality Control Details
• Visual inspection for clear and colorless with no evidence of particulate matter.
Appearance and Color
• HPLC of [11C]MET by coinjection and comparison to standards (2 before, 1 after – system suitability)
• Confirm [11C]MET represents at least 90% of total radioactivity and < 15% RT difference between RAD and UV
Radiochemical Purity & Identity
Chemical Purity (annually) • Chiral HPLC by comparison to standards of L- and D-MET (2 before,
1 after – system suitability)
• Confirm D-MET is ≤ 4% of total D- + L-MET
QC (2)
• [11C]MET is counted in a MCA using 4 - 4 min analyses over 20 min from 40-2500 keV.
• ≥ 95% of the observed gamma emissions should correspond to the 511 keV, 1022 keV, or Compton scatter peaks of 11C.
Radionuclidic Purity (annually)
Volatile Organic Impurities • GC on Carbowax column with comparison to standards (2 before, 1
after – system suitability).
• Confirm presence of solvents are below limits for release – more details to follow.
QC (3)
• Confirm in range of 4.5 – 7.5 by spotting on pH paper pH
Filter Integrity • Sterilizing filter is placed on a compressed air line with a pressure
gauge and the outlet of the filter is placed under water.
• The gas pressure is increased slowly until a steady stream of bubbles is observed. The pressure when the bubble stream begins must be ≥ 50 psi to pass.
• Activity measurement taken in dose calibrator every 5 min over a period of 20 min. Calculated t½ must be 20.4 min ± 10%.
Radionuclidic Identity
QC (4)
• Endosafe®-PTS™ unit utilizing cartridges (pre-filled with LAL) reagents as controls) to spectrophotometrically determine the endotoxin concentration.
• Final product will contain < 5 EU/mL which, assuming injection of the total 10 mL of final product, would remain well within the USP bacterial endotoxin limit for radiopharmaceuticals of < 175 EU/V.
Bacterial Endotoxin
• Approximately 2 mL [11C]MET reserved for sterility testing with 1 mL each innoculated (within 24 hours) into two media types for incubation – thioglycollate and trypticase.
• Over a 14 day period, samples are monitored for bacterial growth (cloudiness).
Sterility
Radiochemical Purity
• HPLC analysis of a co-injection of L-methionine reference standard (100 µg/mL) with [11C]MET.
• Luna 10 μm SCX 100Å column (4.6 x 250 mm) run at 2 mL/min with 0.10 M potassium phosphate, pH 2.6
-1
1
3
5
7
9
11
0 1 2 3 4 5 6 7
Abs
orba
nce
(mA
U)
Time (minutes)
UVRadioactivity
Radiochemical Purity - Sensitivity
• If multiple HPLC systems are in use, confirm similar sensitivity settings and results between systems.
• Establish the accuracy threshold for injected activity and your radioactive detector.
99.76% pure
98.44% pure
SAME PRODUCT, DIFFERENT
INJECTED ACTIVITY
HPLC - Injected Activity
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
0 2 4 6 8 10 12 14 16 18 20
MET
Pea
k Ar
ea (m
AU)
Approximate Radioactivity Injected (µci)
10 uL5 uL3 uL2 uL1 uL
Volatile Organic Impurities
• Testing for organic solvents used in synthesis OR for cleaning/sterilization.
• Used to be called RESIDUAL SOLVENTS….this is misleading and doesn’t encompass all possible volatile organic impurities
• Keep in mind other molecules are volatile – not only typical solvents.
User Experience - GC
• While validating the QC method for [11C]Methionine…..
Carbowax column with oven temp of 60°C for 1 min then ramped to 140°C at 40°C/min for total of 4 min. Injection volume is 0.5 uL and the front inlet heater is at 250°C. Split ratio is 15:1, hydrogen flow is 40 mL/min and the air flow is 400 mL/min.
Acetone
Ethanol
THF
STANDARD
???
FINAL PRODUCT
Mystery Peak
• Very well defined peak – it’s real and must be identified.
• Tested rinses of multiple components looking for something volatile, but not extremely so – possibly residue from manufacturing?
• Considered side products that might be forming amongst the relatively few chemicals involved in synthesis and cleaning, that is also a volatile compound.
• Possible reactants: Homocysteine thiolactone Water Sodium hydroxide Ethanol Acetone
Troubleshooting
Homocysteine + NaOH + Acetone Homocysteine + NaOH + Acetone – 100° NaOH + Acetone – 100°
Possible Byproducts
O
H2C
O
O
OH
O
CH2
O
O
H2C
O
OHH
OO
- H2O
Self-Aldol Condensation
base
Michael addition
- H2O
Acetone MW = 58.08
Diacetone alcohol MW = 116.6
Mesityl oxide MW = 98.14
Isophorone (MW = 138.21)
Robinson Annulation?
Mass Spec
• Confirmation by mass spec and also GC standard.
Diacetone alcohol – H2O
Diacetone alcohol + H+
Diacetone alcohol + Na+
New Standard and Patient GC
SolventAcetone 0.086% AcetoneTHF 0.000% THFEthanol 0.009% EthanolDAA 0.002% Diacetone Alcohol
Release Criteria≤ 0.5%
≤ 0.04%
RESULTS
≤ 0.072%≤ 0.5%
Residual Hexane (needle rinse)
Acetone
THF
Ethanol
DAA
Ramp temperature after 2 min to shorten run to 4 min
STANDARD
FINAL PRODUCT
Setting Release Limit
• Wording added to our CMC section of the IND:
“There is no dose limit listed by the USP for diacetone alcohol, however it has an HMIS rating of 2 and is WHMIS class D-2B, similar to acetone, acetonitrile and THF. Based on the available toxicity data (rat oral LD50 = 2520 mg/kg) diacetone alcohol it seems closest in hazard to acetonitrile (rat oral LD50 = 2460 mg/kg), but less toxic than THF (rat oral LD50 = 1650 mg/kg). To be conservative, we have set the release criterion for diacetone alcohol at 0.04%, equal to the USP limit for acetonitrile.“
• Accepted by the FDA
Comparison of Solvents
0
50
100
150
200
2501.
171.
241.
311.
391.
461.
541.
611.
681.
761.
831.
911.
982.
062.
132.
202.
282.
352.
432.
502.
572.
652.
722.
802.
872.
953.
023.
093.
173.
243.
323.
393.
463.
543.
613.
693.
763.
843.
913.
984.
064.
134.
214.
284.
354.
434.
504.
584.
654.
734.
804.
874.
95
pA
Time (minutes)
ETHA
NO
L -1
.73
ACET
ON
ITRI
LE-1
.99
TOLU
ENE
-2.1
6
AMM
ON
IUM
ACE
TATE
-3.
99
DMSO
-4.8
9
MET
HAN
OL
-1.6
2
DMF
-3.4
4
BUTY
L AC
ETAT
E -2
.25
DIG
LYM
E -3
.36
NIT
ROM
ETHA
NE
-2.6
0
TRIF
LUO
ROTO
LUEN
E -1
.87
ETHY
L BE
NZE
NE
-2.4
7
THF
-1.5
5
ANIS
OLE
-3.
41
ETHY
L AC
ETAT
E -1
.59
IPA
-1.7
0
HEXA
NE
• Isopropyl alcohol elutes 0.03 min before ethanol.
• IPA can enter the FPV if sterile needles are introduced before the IPA-swabbed septum is dry
• Also possible if the QC dose needle punctures the FPV before the septum is dry.
GC Calculations Volatile Organic Impurity Analysis by GC
Tracer: METBatch #:
Performed by: ERB
CALCULATIONS(patient samples)
SolventAcetone 0.086% AcetoneTHF 0.000% THFEthanol 0.009% EthanolDAA 0.002% Diacetone Alcohol
SYSTEM SUITABILITY STANDARD VALUES(Relative standard deviation of standards) Concentration in MET GC Standards
Solvent Value
Solvent Release Criteria Rel Std Dev Acetone 0.10%
Acetone < 10% 0.504% THF 0.01%
THF < 10% 0.999% Ethanol 0.10%
Ethanol < 10% 0.991% DAA 0.01%
DAA < 10% 2.715%
RAW DATAEnter peak areas from the Area Percent Report
PEAK AREA (pA*s)Solvent RT(min) Standard 1 Standard 2 Patient 1 Patient 2 Standard 3
Acetone 1.42 101.42296 101.96733 88.52475 87.13136 102.45029THF 1.53 12.22383 12.25668 0.00000 0.00000 12.45126Ethanol 1.7 101.51244 103.22704 9.58524 9.43543 103.31979DAA 3.51 9.42761 9.55293 1.66103 1.68544 9.93029
20170530METa
SYSTEM SUITABILITY PASSED
Release Criteria≤ 0.5%
≤ 0.04%
RESULTS
≤ 0.072%≤ 0.5%
Volatile Organic Impurity Analysis by GCTracer: MET
Batch #:Performed by: ERB
CALCULATIONS(patient samples)
SolventAcetone 0.081% AcetoneTHF 0.000% THFEthanol 0.009% EthanolDAA 0.002% Diacetone Alcohol
SYSTEM SUITABILITY STANDARD VALUES(Relative standard deviation of standards) Concentration in MET GC Standards
Solvent Value
Solvent Release Criteria Rel Std Dev Acetone 0.10%
Acetone < 10% 10.188% THF 0.01%
THF < 10% 0.999% Ethanol 0.10%
Ethanol < 10% 0.991% DAA 0.01%
DAA < 10% 2.715%
RAW DATAEnter peak areas from the Area Percent Report
PEAK AREA (pA*s)Solvent RT(min) Standard 1 Standard 2 Patient 1 Patient 2 Standard 3
Acetone 1.42 101.42296 121.02598 88.52475 87.13136 102.45029THF 1.53 12.22383 12.25668 0.00000 0.00000 12.45126Ethanol 1.7 101.51244 103.22704 9.58524 9.43543 103.31979DAA 3.51 9.42761 9.55293 1.66103 1.68544 9.93029
20170530METa
SYSTEM SUITABILITY FAILED
Release Criteria≤ 0.5%
≤ 0.04%
RESULTS
≤ 0.072%≤ 0.5%
Volatile Organic Impurity Analysis by GCTracer: MET
Batch #:Performed by: ERB
CALCULATIONS(patient samples)
SolventAcetone 0.086% AcetoneTHF 0.000% THFEthanol 0.009% EthanolDAA 0.002% Diacetone Alcohol
SYSTEM SUITABILITY STANDARD VALUES(Relative standard deviation of standards) Concentration in MET GC Standards
Solvent Value
Solvent Release Criteria Rel Std Dev Acetone 0.10%
Acetone < 10% 0.504% THF 0.01%
THF < 10% 0.999% Ethanol 0.10%
Ethanol < 10% 0.991% DAA 0.01%
DAA < 10% 2.715%
RAW DATAEnter peak areas from the Area Percent Report
PEAK AREA (pA*s)Solvent RT(min) Standard 1 Standard 2 Patient 1 Patient 2 Standard 3
Acetone 1.42 101.42296 101.96733 88.52475 87.13136 102.45029THF 1.53 12.22383 12.25668 0.00000 0.00000 12.45126Ethanol 1.7 101.51244 103.22704 9.58524 9.43543 103.31979DAA 3.51 9.42761 9.55293 1.66103 1.68544 9.93029
20170530METa
SYSTEM SUITABILITY PASSED
Release Criteria≤ 0.5%
≤ 0.04%
RESULTS
≤ 0.072%≤ 0.5%
Volatile Organic Impurity Analysis by GCTracer: MET
Batch #:Performed by: ERB
CALCULATIONS(patient samples)
SolventAcetone 0.575% AcetoneTHF 0.000% THFEthanol 0.009% EthanolDAA 0.002% Diacetone Alcohol
SYSTEM SUITABILITY STANDARD VALUES(Relative standard deviation of standards) Concentration in MET GC Standards
Solvent Value
Solvent Release Criteria Rel Std Dev Acetone 0.10%
Acetone < 10% 0.504% THF 0.01%
THF < 10% 0.999% Ethanol 0.10%
Ethanol < 10% 0.991% DAA 0.01%
DAA < 10% 2.715%
RAW DATAEnter peak areas from the Area Percent Report
PEAK AREA (pA*s)Solvent RT(min) Standard 1 Standard 2 Patient 1 Patient 2 Standard 3
Acetone 1.42 101.42296 101.96733 587.26654 584.56642 102.45029THF 1.53 12.22383 12.25668 0.00000 0.00000 12.45126Ethanol 1.7 101.51244 103.22704 9.58524 9.43543 103.31979DAA 3.51 9.42761 9.55293 1.66103 1.68544 9.93029
20170530METa
SYSTEM SUITABILITY PASSED
Release Criteria≤ 0.5%
≤ 0.04%
RESULTS
≤ 0.072%≤ 0.5%
Excel spreadsheet with conditional
formatting provides double check
Enantiomeric Purity
• Ishiwata and colleagues* have shown that by limiting the concentration of NaOH to less than 250 mM, less than 4% D-methionine is formed.
-10
0
10
20
30
40
50
0 2 4 6 8 10 12UV
Abs
orba
nce (
mA
U)
TIme (minutes)
UV @ 254 nm
Radioactivity
L D
Phenomenex Chirex-33126 D-penicillamine column (4.6 x 50 mm) with a mobile phase of 1 mM copper sulfate : 2-propanol, 95:5. The flow rate is 0.5 mL/min
* Ishiwata et al. Int J Rad Appl Instrum A (1988) 39:311-314.
Chiral Purity
• Routine annual validation revealed increase in D-MET in product.
• Performed synthesis on two different radiosynthesizers with similar results.
• Synthesis is successful, but what about the chiral purity of the precursor – L-homocysteine?
5.4% D-MET!!
Results with Pure Precursor
• We have added routine testing of any new lot of precursor prior to use.
• We confirmed that the vendor DOES NOT test chiral purity. They assume all is L because their precursor is L. Purity listed is purity of homocysteine, not chiral purity.
6-[18F]Fluorodopamine - Synthera
Use commercially available synthesis modules and kits to make fluorinated compounds quickly and easily.
F-DA Synthesis – 2 Syntheras
K18F, Kryptofix 2.2.2 Azeotrope evap: 90 sec @ 110°
12 mg, 600 µL CH3CN Evap @ 50°C
HI, 600 µL 4 min @160°
NBoc
Boc
IO
O
OOTf
NBoc
Boc
IO
O
O
18F
NH2
18FHO
HO
Toluene, 900 µL 4 min @155°
NBoc
Boc
18FO
O
O
I+
NBoc
Boc
18FO
O
SYNTHERA 1
SYNTHERA 2
25% uncorrected, HPLC-purified yield
in 60 min
Acid-Labile Precursor = 1 Synthera
• Residual levels of Kryptofix were higher than the release limit established by the USP (< 50 µg/mL) even after HPLC purification.
• One-pot synthesis removed intermediate purification that eliminated Kryptofix
K18F, Kryptofix 2.2.2 Azeotrope evap: 2 min @ 110°
15 mg 10/90 MeCN/Toluene
Decomposition 8 min @ 140°
4 M HCl 20 min @ 95°
NBoc
Boc
IO
O
OOTf
O
O
NBoc
Boc
IO
O
O
18F
O
O
NBoc
Boc
18FO
OO
O
O
I+
NH2
18FHO
HO
Replacement?
• Any attempt to remove K2.2.2 also removed F-DA. It is believed they coelute on the HPLC purification.
• Tetraethylammonium bicarbonate (TEA HCO3) and tetrabutylammonium bicarbonate (TBA HCO3) were selected as possible replacements based on previous successes.*
• Fluorination testing demonstrated incorporation by either tetralkylammonium salt comparable to Kryptofix.
* Brichard et al., Eur J Org Chem (2014), 28:6145-6149; Reed et al. J Fluor Chem (2012) 143:231-237.
Excellent Yield
(n = 19) (n = 9) (n = 8) (n = 7)
22.9% 23.0% 22.5% 26.0%
0%
5%
10%
15%
20%
25%
30%[18
F]F-
DA Y
ield
(unc
orre
cted
)
Kryptofix/K2CO3 2.5 mg TEA HCO3
3.4 mg TEA HCO3
2-2.5 mg TBA HCO3
Residual TBA HCO3
• TBA HCO3 standards and F-DA samples were spotted on silica and overspotted with a methanol/25% ammonium hydroxide solution (9:1).
• Development in an iodine chamber for one minute allowed visual comparison for interpolation of TBA HCO3 concentration.
• < 100 µg/mL residual TBA HCO3, and possibly < 50 µg/mL, when visually compared to reference standards (n = 8)
0 50 100 150 200
TBAB (µg/mL)
Patient Sample
Kuntzsch, M. et al., Pharmaceuticals (2014) 7: 621-633
Justification
• USP does not provide a recommended release limit for TBA HCO3, however, the European Pharmacopoeia has an established release limit of 2.6 mg/injected volume.*
• Synthesis requires only 2.0 mg of TBA HCO3 - even if all TBA HCO3 were carried through to the final product, the release limit would not be reached.
• Majority of the TBA HCO3 will be purified out of the final product via HPLC during the final purification.
• Residual TBA HCO3 in several [18F]F-DA preparations were consistently lower than 0.1 mg/mL.
* The European Pharmacopoeia, EDQM (2014) Strassbourg, France.8.0..
Volatile Excipient?
F-DA FINAL PRODUCT
Residual Hexane (needle rinse)
Ethanol
Ammonium Acetate
STANDARD
Acetonitrile
Toluene
Not an Impurity
• Ammonium acetate is added as a neutralization buffer to the F-DA product with a final concentration of 0.15%.
• This is below the approved level of 0.4% for intravenous injection of ammonium acetate as an excipient for sterile dosages in the formulation of pharmaceuticals.*
• Not included in Volatile Organic Impurity calculations because the percentage is known and the addition is intentional, not an impurity.
* Niazi, SK. Chapter 13: Approved Excipients in Sterile Dosage Forms. Handbook of Pharmaceutical Manufacturing Formulations: Sterile Products. Volume Six, 2nd Ed. 2009. Informa Healthcare USA, New York, NY.