evaluation of the leukocyte differential on a new automated flow cytometry hematology analyzer
TRANSCRIPT
Evaluation of the leukocyte differential on a new
automated flow cytometry hematology analyzerJ.-E. KIM, B.-R. KIM, K.-S. WOO, J.-Y. HAN
INTRODUCTION
Differential white blood cell count is a gold standard
for the accurate identification of cells in the peripheral
blood [1]. However, this method is both laborious and
time-consuming. Therefore, complete blood counts
(CBCs) and white blood cell (WBC) differentials con-
ducted using automated hematology analyzers have
replaced the traditional manual differential count
method for the initial screening of hematologic abnor-
malities in clinical laboratories [2, 3]. However,
a majority of automated hematology analyzers have
a limitation in identifying WBCs to only five cell
populations: neutrophils, lymphocytes, monocytes, eo-
sinophils, and basophils.
The Hematoflow (Beckman Coulter, Miami, FL, USA)
is the newest automatic hematology analyzer including
a DxH 800 Coulter Cellular Analysis System and FC 500
Flow Cytometer from Beckman Coulter. The DxH800
uses multiple angles of light scatter, and the FC500
provides a 16-part WBC differential, giving precise
information on WBC subsets [4]. We evaluated the diff-
erential WBC count performance of the FC500 in com-
parison with the Beckman Coulter DxH800 (Beckman
Coulter), Sysmex XE-2100 (TOA Medical Electronics
Co, Kobe, Japan), and the reference manual method.
Departments of Laboratory
Medicine, College of Medicine,
Dong-A University, Busan, Korea
Correspondence:
Jin-Yeong Han, Department of
Laboratory Medicine, Dong-A
University College of Medicine, 1,
3-Ga, Dongdaesin-dong, Seo-gu,
Busan, 602-715, Korea.
Tel.: +82 51 240 5323;
Fax: +82 51 255 9366;
E-mail: [email protected]
doi:10.1111/j.1751-553X.2012.01432.x
Received 14 December 2011;
accepted for publication 10 April
2012
Keywords
Blood cell analyzer, leukocyte
differential, blast, cell lineage
SUMMARY
Introduction: The Hematoflow (Beckman Coulter, USA) is a new
automated hematology analyzer, which provides a 16-part white
blood cell count (WBC) differential.
Methods: We evaluated the differential WBC count performance of
the Hematoflow. 101 blood samples from patients were selected for
comparison analysis.
Results: The methodological comparison of the WBC differential para-
meters of neutrophils, lymphocytes, monocytes and eosinophils
showed good correlations among 4 different analyses. More than 1%
of blast cells were counted in 30 of 101 samples. A good correlation
for blast cell counts obtained by Hematoflow was found with the
reference manual method (r=0.9637, P 0.0001). For blast B, Hemato-
flow shows good correlation with reference method results but did
not identify blast T.
Conclusions: These results demonstrate that the Hematoflow has a
comparable performance with the Sysmex XE-2100 and indicate that
B cell lineage ALL can be identified by the use of the Hematoflow in
an initial evaluation of acute leukemia.
SHORT REPORT INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY
� 2012 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 1
International Journal of Laboratory HematologyThe Official journal of the International Society for Laboratory Hematology
MATERIALS AND METHODS
One hundred one blood samples (54 men and 47
women; age, 1–84 years; mean age, 43 years) in rou-
tine CBC were selected for analysis regardless of previ-
ous diagnosis. All samples were collected in evacuated
3-mL tubes containing EDTA K3 and were analyzed
within 6 h after phlebotomy. The diagnosis and classi-
fication of the patients were made on the basis of the
2008 WHO classification [5] by morphologic, immu-
nophenotypic, cytogenetic, and molecular genetic
findings of the bone marrow. This study was approved
by the Institutional Review Board of the Dong-a
University of College of Medicine.
The FC500 contains a premixed Cytodiff reagent
and analysis software. The Cytodiff panel included six
directly conjugated monoclonal antibodies in a five-
color single reagent. The leukocytes were differenti-
ated into 16 cell populations (B-lymphocytes, CD16)
T-lymphocytes, CD16+ T and NK cells, T and NK lym-
phocytes, total lymphocytes, CD16 monocytes, CD16+
monocytes, total monocytes, immature granulocytes
[IGs], total eosinophils, mature neutrophils, total neu-
trophils, blasts B, blasts T, blasts non-B-non-T, and
total basophils). Analysis procedures were conducted
in accordance with the manufacturer’s manual [6].
The automated WBC differential counts were com-
pared with the manual differential counts. For the
validation of the results from both the analyzers, two
experienced hematopathologist did manual differential
counts by counting 100 cells. The five-part automated
WBC differential was conducted using both the DxH
800 and the XE-2100 within 6 h after blood collec-
tion. The blast suspect flags generated by the DxH 800
apparatus were compared with the results of the man-
ual differential count. The correlation of the blast cells
was made by the number of cells counted by different
methods.
Statistical analysis
Statistical analysis was performed using MEDCALC v11.2
(Mariakerke, Belgium). The results from the different
methods for WBC differential count are compared by
regression analysis. Values of P < 0.05 were regarded
as significant.
RESULTS
The DxH 800 reported CBC results with the five-part
WBC differential count in 96.0% (97 of 101) of the
cases. The XE-2100 provided a valid automated differ-
ential count in 84.2% (85 of 101) of the cases. The
methodological comparison of the WBC differential
parameters of neutrophils, lymphocytes, monocytes,
and eosinophils showed good correlations between
four different analyses (Table 1). The results obtained
by the XE-2100 were better than DxH800 when
Table 1. Comparison of the results from FC500, DxH800, and XE-2100 with manual differential leukocyte count
Neutrophil Lymphocyte Monocyte Eosinophil Basophil
FC 500
Intercept 3.0203 0.1265 2.0742 0.4041 0.3208
Slope 0.9579 0.9907 0.8151 0.9170 0.4989
r 0.9390 0.9666 0.9079 0.9484 0.5028
P <0.0001 <0.0001 <0.0001 <0.0001 <0.0001
DxH 800
Intercept 7.4753 5.7954 5.5057 0.3319 0.9987
Slope 0.8184 0.8993 0.8282 0.9393 1.3110
r 0.8073 0.8084 0.5868 0.9450 0.2684
P <0.0001 <0.0001 <0.0001 <0.0001 0.0093
XE-2100
Intercept 3.1381 6.1907 4.5416 0.1460 0.8843
Slope 0.9492 0.9112 0.7496 0.9566 0.5297
r 0.9574 0.9030 0.6838 0.9739 0.2007
P <0.0001 <0.0001 <0.0001 <0.0001 0.0537
� 2012 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem.
2 J.-E. KIM ET AL. A NEW HEMATOLOGY ANALYZER BY FLOW CYTOMETRY
compared with manual counting and closer to the FC
500, with the exception only of counting basophils.
For basophil count, the correlation was moderate
(r = 0.5028) with the FC500; however, the Sysmex
XE-2100 and DxH800 showed poor correlations
(r = 0.2007 and 0.2684, respectively).
In manual method, more than 1% of blast cells
were counted in 30 of 101 samples. There were seven
cases of acute lymphocytic leukemia (ALL), including
four T-cell ALL (T-ALL) cases and three B-cell ALL
(B-ALL) cases; 20 cases of AML; and three cases of
myelodysplastic syndrome (MDS). A good correlation
for blasts was found with the reference manual
method (r = 0.9637, P < 0.0001). There were five
cases of ALL showing over 10% of blast counts
(Table 2). For B-cell ALL, blast B count of Hemato-
flow shows good correlation with reference method
results. For T-cell ALL, Hematoflow did not identify
blast T and shows complete agreement with manual
counting. For AML and MDS, blast non-B-non-T
counts of Hematoflow shows 100% concordance and
good correlation with reference method results.
DISCUSSION
Modern characterization of acute malignant hemato-
logical disease is a multidisciplinary process. Despite
the advances in diagnostic technologies, the mainte-
nance and improvement of morphological skills still
remain essential requirements in the diagnosis of
acute leukemia. Automated differential white blood
cell count can not count certain cell populations pass-
ing through the blood (immature granulocytes, blasts,
atypical lymphocytes, and lymphoma cells), which
must be analyzed secondarily by optical or digitized
microscopy. In hospitals, these reviews represent 10–
25% of cases according to established criteria [7]. In
this context, Beckman Coulter proposed the new He-
matoFlow concept using flow cytometry of 20 000
cells stained by the CytoDiff monoclonal antibodies
combination prior to microscopy. In the new concept,
the alarms generated by the hematology analyzers
induce the review by flow cytometry before any
check by microscopy.
The DxH 800 reported CBC results with the five-
part WBC differential count in majority (96.0%) of
the cases, whereas the XE-2100 provided a valid auto-
mated differential count in 84.2% of the cases. The
correlation of the FC 500 method was better to that of
the XE-2100 in the counting of lymphocytes, mono-
cytes, and basophils. This means the leukocyte differ-
ential obtained using the Hematoflow method is more
reproducible and sensitive than that obtained using
other methods. It represents a substantial advance in
hematology instrumentation and has considerable
potential for extending the clinical role of hematologic
studies.
We found a moderate correlation (r = 0.5028)
between the FC500 basophil count and reference
counts. Some authors reported a low correlation
between the Cytodiff basophil count and reference
counts [8, 9], whereas some found a high correlation
between these counts (r = 0.8727) [6]. This is owing
to CD2+ CRTH2 expressed in activated T cells, eosin-
ophils, and basophils. Separation between basophils
and non-B-non-T blast populations was not effective
in the Hematoflow method [6].
Whatever the pathology (AML, ALL, and MDS)
and the number of blasts on smear, all patients were
positive for blast detection on the Hematoflow.
Table 2. The comparison between blast counts by manual method and the percentage of each blast subset by
Hematoflow in five cases of ALL
No. of cases Diagnosis Manual (%) Blast B (%) Blast T (%) Blast non-B-non-T (%)
1 T-ALL 57 0.69 0.04 58.54
2 T-ALL 28 2.05 0 28.79
3 B-ALL 84 77.87 0 0.63
4 T-ALL 98 0.12 0 94.25
5 B-ALL 47 51.26 0.01 0.33
ALL, Acute lymphocytic leukemia.
� 2012 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem.
J.-E. KIM ET AL. A NEW HEMATOLOGY ANALYZER BY FLOW CYTOMETRY 3
Additionally, it appears easy to distinguish B-cell line-
age leukemic blasts from T-cell leukemic blasts and
myeloblasts. The Hematoflow was not able to classify
correctly T-cell leukemic blasts as blast T. The recom-
mended procedure in this case is still to use morpho-
logy as the most reliable result. In cases of T-ALL and
AML, it is needed to be reclassified by the hematologic
pathologists and immunotyping using bone marrow
samples is still required for the initial diagnosis. The
detection rate of blast and the correlation of the blast
count by the Hematoflow method were similar to the
previous study [6]. It may be useful to screen samples
for the presence of blasts using Cytodiff. If blasts are
detected by Cytodiff, confirmation of the blasts by slide
review is recommended, especially in new patients.
While acknowledging the limitations of our study,
our data indicate that B-cell lineage ALL can be iden-
tified by the use of the Hematoflow in an initial eval-
uation of acute leukemia. For more accurate
distinctions of lineages, further studies with more
specimens will be necessary to apply for all occasions.
In this study, we became aware of the capabilities
and limitations of the automated Hematoflow for ana-
lyzing patient sample with leukemic blasts. For the
technical staff, the installation of Hematoflow would
have many consequences: the reduction in technical
staff time at the microscope while simultaneously
increasing the efficiency of the workflow, the elimina-
tion of medical technologists facing a difficult diagno-
sis alone. This rapid and quantitative technique may
be helpful in the differential diagnosis of acute leuke-
mias prior to additional investigation.
ACKNOWLEDGEMENTS
This study was supported by the Korea Science and
Engineering Foundation (KOSEF) grant funded by the
Korea government (MEST; R13-2002-044-05002-0).
REFERENCES
1. McClatchey KD, ed. Clinical Laboratory
Medicine, 2nd edn. Philadelphia: Lippincott
Willimas and Wilkins; 2002: 809.
2. Lantis KL, Harris RJ, Davis G, Renner N,
Finn WG. Elimination of instrument-driven
reflex manual differential leukocyte counts.
Optimization of manual blood smear review
criteria in a high-volume automated hema-
tology laboratory. American Journal of
Clinical Pathology 2003;119:656–62.
3. Pierre RV. Peripheral blood film review.
The demise of the eye count leukocyte diff-
erential. Clin Lab Med 2002;22:279–97.
4. Faucher JL, Lacronique-Gazaille C, Frebet
E, Trimoreau F, Donnard M, Bordessoule
D, Lacombe F, Feuillard J. ‘‘6 Markers/5
Colors’’ Extended White Blood Cell Differ-
ential by Flow Cytometry. Cytometry A
2007;71:934–44.
5. Swerdlow SH, Campo E, Harris NL, Jaffe
ES, Pileri SA, Stein H, Thiele J, Vardiman
JW, eds. WHO Classification of Tumours of
Haematopoietic and Lymphoid Tissues, 4th
edn. Lyon: IARC; 2008: 110–78.
6. Jo YJ, Kim SH, Koh KS, Park JM, Shim YB,
Lim JH, Kim YG, Park YJ, Han KJ. Reliable,
Accurate Determination of the Leukocyte
Differential of Leukopenic Samples by
Using Hematoflow Method. Korean Journal
of Laboratory Medicine 2011;31:131–7.
7. Kassengera Z, Penninck D, Pradier O.
Improving and automating the haemato-
logy routine workhorse with Hematoflow
concept using the Cytodiff monoclonal anti-
body combination: the benefits of five
month routine [abstract]. Int J Lab Hematol
2001; 23 (Suppl 1):96.
8. Bjornsson S, Wahlstrom S, Norstrom E,
Bernevi I, O’Neill U, Johansson E, Runs-
trom H, Simonsson P. Total nucleated cell
differential for blood and bone marrow
using a single tube in a five-color flow
cytometer. Cytometry B Clin Cytom
2008;74:91–103.
9. Cherian S, Levin G, Lo WY, Mauck M,
Kuhn D, Lee C, Wood BL. Evaluation of an
8-color flow cytometric reference method
for white blood cell differential enumera-
tion. Cytometry B Clin Cytom
2010;78:319–28.
� 2012 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem.
4 J.-E. KIM ET AL. A NEW HEMATOLOGY ANALYZER BY FLOW CYTOMETRY