epi-illumination is form of kohler illumination: objective is also condenser lamp or laser detector...

44
Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly increases S/N Light is focused At back aperture Of the objective, Conjugate to condenser aperture Different illumination And image paths White light (regular Kohler) White light (regular Kohler) Brightfield, phase, etc Brightfield, phase, etc lens

Upload: shonda-dorsey

Post on 17-Jan-2016

225 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Epi-illumination is form of Kohler Illumination:Objective is also condenser

Lamp orlaser

detector

Detect at 90 degreesSplit with dichroic mirrorGreatly increases S/N

Light is focusedAt back aperture Of the objective,Conjugate tocondenseraperture

Different illuminationAnd image paths

White light (regular Kohler)White light (regular Kohler)Brightfield, phase, etcBrightfield, phase, etc

lens

Page 2: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

First barrier filterSelects excitation

dichroicmirror

Secondbarrier filterSelects signalFrom background

objective lens

specimen

Epi-illumination separates light source,Fluorescence signal

Arclamp

Page 3: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

•Excitation filter typically interference bandpassExcitation filter typically interference bandpass•Dichroic is longwave passDichroic is longwave pass•For one dye-maybe no emission filterFor one dye-maybe no emission filter

Page 4: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Dielectric layers or Metallic layers used as filter coating

Reflect, transmit colors of choice by using multilayers

Page 5: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Coatings work by interference

Reflectance depends onWavelength, film thickness material (index*length), incident angle.

Fabry-Pérot interferometer

Page 6: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Block 3-6 OD outside of band Transmit 10-50% (worse for UV)

Use of bandpass interference filters in wavelength selection

Page 7: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Dichroic Mirrors: separate colors by using coatings

Beam separator:Separate different colors (fluorescence)At right angles: used in microscopes

Beam combiner:Multiple lasers

Transition should be sharp

Page 8: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

How CCD Camera Works

Serial readout limit speed. A partial solution is using Frame-Transfer.

Page 9: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Comparison: Detector Quantum Yield

Page 10: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Efficiency & Signal/Noise?

• Collection efficiency of microscopy: ~25%

• Detector quantum yield: ~70-90%

• Thermal noise

• Shot noise (quantum noise):

• Read noise (A/D conversion)

Page 11: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

CCD Dark counts

Liquid Nitrogen

Thermal Electric

Thermal Electric in ultrahigh vacuum

Cooling methods:

Page 12: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

EM-CCD

- Largely eliminate read noise

- Introduces amplification noise

- Net effect is S/N improvement for extremely low light level situation

Page 13: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly
Page 14: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Detecting A Single Fluorescent Molecule?

• Size: ~ 1nm

• Absorption Cross-section: ~ 10-16 cm2

• Quantum Yield: ~1

Absorbance of 1 molecule = ?

How many fluorescence photons per excitation photons?

Page 15: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Single Molecule “Blinks”

Page 16: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

How to Analyze Single Molecule Measurements (I)-- Histograms

Most Probable Value vs Average value

Page 17: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

• Emission– Wavelength dependenceof detectors– Spectral separation fromexcitation– Efficient detection optics– Autofluorescence andcontaminantfluorescence– Photobleaching and ISC– Scatter:• elastic (Rayleigh)• inelastic (Raman)

Single molecule fluorescence:experimental considerations

• Excitation– High NA objective lens– “Bright” fluorophores• High extinctioncoefficient• High quantum yield– Exclude quenchers• particularly molecularoxygen!• O2 scavengers includeβ-mercaptoethanol(BME), catalase

Page 18: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Back to Single Protein Detection

Myosin V -- a motor protein.

Page 19: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly
Page 20: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

De-convolution Microscopy

Thompson, RE; Larson, DR; Webb, WW, Biophys. J. 2002,

Page 21: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Paul Selvin

Page 22: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Photodiode

PMT: photomultiplier

APD: Avalanche Photodiode

CCD

Page 23: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

PMT

APD

Both can work under Single-photon Countingmode

Page 24: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Typical Dark Counts

CCD APD

0.001 e/sec/pixel 10-100 e/sec/pixelDark Counts

Temperature -70 C -20 C

Sensitive Area 10-20 m 100-500 m

Page 25: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Total internal reflection: the reflectionthat occurs when light, in a higherrefractive-index medium, strikes aninterface with a medium that has a lowerrefractive index, at an angle of incidence(α1) greater than the critical angle.

Total Internal Reflection Fluorescence Microscopy

TIRFM

Page 26: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Snell’s law

Page 27: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly
Page 28: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

1)sin(/)sin(4 212

0

g

pn

d

Page 29: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Application Example 1 – Cytoskeleton

TIRF Epi

Page 30: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Prism-TIRF Objective-TIRF

Setting up the TIRF microscope

Page 31: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

A little History: EVDLS

Page 32: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Daniel Axelrod

1980s: start to apply TIR principleto fluorescence and bio-imaging.

Page 33: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Prism Based TIRF Setup 1

Page 34: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Spherical Aberration from Aqueous Sample

Sample near glass coverslip Sample in the bulk water

Page 35: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Water Immersion Objective

Fully water immersion Water immersion with coverslip

Page 36: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Prism-TIRF Objective-TIRF

Page 37: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

• NA requirement

• Oil immersion

• Size of the beam

Key Points:

柳田敏雄 Toshio Yanagida

Page 38: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Through Objective TIR Design 1: direct coupling

Page 39: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Through Objective TIR Design 2: Fiber Optics

Optical fiber based light delivery

Easy conversion from non-TIR to TIR

Compatible with Arc lamp

Page 40: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly
Page 41: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Other Practical Concerns:

• Upright or inverted microscope?

• Light sources?

• Polarization?

Page 42: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Arc Lamp TIRF

Page 43: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Fresnel equations

Page 44: Epi-illumination is form of Kohler Illumination: Objective is also condenser Lamp or laser detector Detect at 90 degrees Split with dichroic mirror Greatly

Polarization Control