epa remedial project manager: jane dolan deputy … · 2018-05-28 · episode will provide current...

64
(^JUJ^AU.VC- SDMSDocID 451884 QUALITY ASSURANCE PLAN ADDENDUM \\^ ^\ ^<^y CHESHIRE GROUNDWATER CONTAMINATION SUPERFUND SITE U.S. EPA REGION I EPA Remedial Project Manager's Signature: -JeH ixu. m iy\ EPA Remedial Project Manager: Jane Dolan Deputy Program Quality Assurance Officer's Signature: Deputy Program Quality Assurance Officer: Meg Himmel ARCS Lead Chemist's Signature: i^if/f- /^^^/'^ p. V^ ARCS Lead Chemist: Dr. Brian Tucker M«&E Project Manager: Martha Zirbel Amended By: Bruce Livingston >«****. W

Upload: ngocong

Post on 22-Jun-2018

212 views

Category:

Documents


0 download

TRANSCRIPT

(^JUJ^AUVC-

SDMSDocID 451884

QUALITY ASSURANCE PLAN ADDENDUM ^ ^lt^y

CHESHIRE GROUNDWATER CONTAMINATION SUPERFUND SITE

US EPA REGION I

EPA Remedial Project Managers Signature -JeH ixu miy

EPA Remedial Project Manager Jane Dolan

Deputy Program Quality Assurance Officers Signature

Deputy Program Quality Assurance Officer Meg Himmel

ARCS Lead Chemists Signature i^iffshy ^ p V ^

ARCS Lead Chemist Dr Brian Tucker

MlaquoampE Project Manager Martha Zirbel

Amended By Bruce Livingston

gtlaquo

W

QUALITY ASSURANCE PLAN ADDENDUM

v CHESHIRE GROUNDWATER CONTAMINATION SUPERFUND SITE

US EPA REGION I

EPA Remedial Project Managers Signature

EPA Remedial Project Manager Jane Dolan

Deputy Program Quality Assurance Officers Signature

Deputy Program Quality Assurance Officer Meg Himmel

ARCS Lead Chemists Signature S^^^ J 2 Kff^^

ARCS Lead Chemist Dr Brian Tucker

MampE Project Manager Martha Zirbel

Amended By Bruce Livingston

This Quality Assurance Plan (QAP) Addendum is an update to the original QAP for this ^

project which was filed on September 30 1994 Only those portions of the original QAP

that required updating are presented in this Addendum Unless updated with this document

all quality control procedures will be conducted in accordance with the original September

1994 QAP for this project

1 Project Name Cheshire Groundwater Contamination Superfund Site - Groundwater Sampling

2 Project Requested By EPA Region I

3 Date QAP Amendment FUed 112795

4 Date of Project Field Initiation 120495

5 EPA Remedial Project Manager Jane Dolan

6 Deputy Quality Assistance Officer Meg Himmel

7 Project Description

A Objective and Scope The objective of this sampling effort is to obtain updated analytical data to determine groundwater quality in the vicinity of the Cheshire Groundwater Contamination Superfund Site Three existing overburden monitoring wells and one abandoned residential well (Fusco well) that were sampled in November 1994 will be sampled again as part of this field effort Two additional overburden and three bedrock wells recently installed by the US Geological Survey will also be sampled All groundwater samples will be analyzed for low concentration volatile organics through the DAS program and for full inorganics through the CLP RAS system Based on previous analytical results from the site purge water generated during well sampling will be discharged to the ground surface away from each well

B Data UseData Quality Objectives The data acquired during this sampling episode will provide current information about the nature and extent of contamination by volatile organic compounds (VOCs) inorganic elements and cyanide

The VOC data generated through the DAS program will be analytical level V data while the CLP RAS full inorganic data will be analytical level IV The analytical data packages will consist of all documentation required to conduct a

^ ^ r f j M -

Tier III data validation however only Tier I validation (evidence audit) will be conducted The data generated during this field investigation will be used by EPA personnel to conduct a risk assessment for the site

1

c Monitoring Parameters and Sample Frequency of Collection A total of nine wells will be sampled and analyzed using both RAS CLP and DAS methodologies Full inorganics including the twenty three target analyte list (TAL) metals and cyanide will be analyzed using RAS procedures through the CLP system Low concentration VOCs will be analyzed using MampEs EPA approved DAS specification which is based on EPA Method 5242

In addition to the laboratory analytical procedures pH temperature conductivity turbidity and depth to groundwater will be measured at each well during purging and following sample collection

D Parameter Table

Number Analytical of Method Sample Container Field QC

Parameter Samples Matrix Reference Preservation Holding Time Samples

DAS - VOCs 9 GW 5242 2x40 ml VOA vial 14 days V HCI t opHlt2 Ice

RAS - Full Inorganics

TAL Metals 9 GW TTM030 1 liter Plastic 6 months 4 (except HN03 topHlt2

Vaa Mercury)

Mercury 9 GW ILM030 1 liter Plastic 28 days 4 HN03topHlt2

Cyanide 9 GW ILM030 1 liter Plastic 14 days A NaOHtopHgt10

Field Measurements

pH 9 GW 1501 Disposable Plastic Cup Field Analyze None

Temperature 9 GW 1701 Disposable Plastic Cup Field Analyze None

Conductivity 9 GW 1201 Disposable Plastic Cup Field Analyze None

Turbidity 9 GW 1801 Disposable Plastic Cup Field Analyze None

1) Low concentration VOCs will be analyzed using the MampE DAS specification based on EPA Method 5242 A copy of the DAS Specification is provided at the end of this QAP Addendum

2) QC saniples for VOCs include one MSMSD pair (2) one field duplicate two trip blanks one equipment blank and one performance evaluation (PE) sample

3) QC samples for inorganics include one matrix spike one equipment blank one field duplicate and one PE sample

4) Methods for Chemical Analysis of Water and Wastes US EPA March 1983 f ^

Schedule of Tasks and Products

Samples are anticipated to be collected the week of December 4 1995 Following the completion of the field work MampE will prepare a draft trip memo detailing the work completed The field measurements and procedures conducted along with any deviations from the approved project plans will be provided in the draft trip memo The final trip memo will address any comments on the draft and will include all of the field and laboratory analytical data The results wUl be tabulated to facilitate evaluation

Project Organization and Responsibility

Sampling Operations Bruce Livingston Sampling QC Bruce Livingston Laboratory Analysis RASDAS Laboratories Laboratory QC RASDAS Laboratories Data Quality Review RASDAS LaboratoriesMetcalf amp Eddy Performance Auditing None Systems Auditing None MampE Deputy QA Officer Meg Himmel MampE Project Manager Martha Zirbel

10 Laboratory and Field Data Quality RequirementsAssessments W

Estimated Estimated Parameter Matrix Detection Limit Accuracy Precision

VOCs GW 1-5 ngV 80-120 lt30 RPD

TAL Metals GW 2-100 MgL 80-120 lt30 RPD

Cyanide GW 10 MgL 75-125 lt30 RPD

pff GW NA 005 units plusmn01 units

Temperature^ GW NA 005degC plusmn01degC

Conductivity GW NA lOMmhocm plusmn10

Turbidity GW NA plusmn10 plusmn20

1) The detection limits presented are achievable for the referenced methods Due to interferences requiring dilutions these detection limits may not always be met

2) The detection limits for the majority of VOCs are 1-5 MgL Tetra hydrofuran and 14-Dioxane have detection limits of 50 MgL

3) The field measurements accuracy and precision are based on the equipment manufacturer specifications and MampE field experience

10AB No changes

11 No changes except that VOCs will be analyzed through DAS

12 13 14 15A No changes

15B Data Reduction and Reporting Data reduction and reporting will be conducted by the RAS and DAS laboratories

After receiving the data packages MampE will conduct Tier I (evidence audit only) validation The data will be reported in a spreadsheet format

16 Data Validation Tier I data validation will be conducted However all documentation necessary to conduct a Tier III data validation will be provided to meet the project data quality objectives

17 18 19 No changes

20 Reports to EPA Following the completion of groundwater sampling MampE will prepare a draft trip memo identifying the work completed and presenting the field analytical data The draft trip memo will be completed within two weeks of the conclusion of sampling Laboratory data should arrive at MampE five weeks after the conclusion of sampling The fmal trip memo will include a tabulation of all of the laboratory analytical data The final trip memo will be completed within three weeks of receipt of laboratory analytical data by MampE

Attaclunents to Original QAP No changes

ANALYTICAL SPECIFICATION FOR

METHOD 5242

SCOPE

This specification is for the analysis of groundwater surface water or other aqueous matrices for low concentration volatile organic compounds water using EPA Method 5242 The additional analytes added to the list of 5242 compounds are acetone 2-butanone 2-hexanone 4-methyl-2-pentanone carbon disulfide tetrahydrofuran and 14-dioxane The complete list of target compounds is presented in Attachment A In addition to the target compounds up to ten tentatively identified compounds must also be reported

Method 5242 requires the laboratory to perform a method detection limit (MDL) smdy prior to sample analysis to determine the theoretical lowest statistically valid detection limit This study must follow the requirements described in 40 CFR Part 136 Appendix B using at least seven replicate analyses of the lowest calibration concentration The study must meet the detection limits described in Section 6 The continuing calibration must be conducted on an eight hour basis no other time span is acceptable The laboratory performing the work will be subject to audit by Metcalf amp Eddy at the discretion of the Metcalf amp Eddy lead chemist

PURPOSE

The data derived from these analyses will be used to further delineate groundwater contamination to determine if treatment has met human health and ecological risk objectives to determine the quality of data generated by PRP consultants and possibly for other applications

This method is intended for trace analysis of volatiles in waters that are only slightly impacted by pollution and is not intended for use in analysis of industrial effluent landfill leachate or any moderately to highly contaminated surface waters contaminated groundwater or treatability study influents If samples are collected from areas with historical data concentration ranges will be provided to the laboratory in the work order If the samples are found to effervesce during addition of acid for sample preservation the samples will not be preserved and the laboratory will be notified concerning the absence of preservative

DEFINITION OF WORK

Surface water groundwatier or other aqueous matrices are to be analyzed by Method 5242 with the modifications presented within The modifications are described in Section 7 Analytical Procedures

c Performance evaluation samples (PEs) will be provided with up to one in each sample delivery group to be analyzed along with the field samples No more than twenty samples will be submitted in a sample delivery group Field samples will include blind trip bottle and equipment blanks as well as blind duplicates

SCHEDULE

Samples will be shipped at most one day after collection Saturday delivery may be required An overnight delivery service will be used Contacts for shipping will be provided in each work order Data delivery inquiries may be made to Dr Brian Tucker Metcalf amp Eddy Inc Wakefield MA 01880 (617) 246-5200 extension 6433

Holding Time

Analyses are required to be performed within fourteen (14) days of sample collection for preserved samples and seven (7) days for unpreserved samples Sample preservation will be noted on the chain of custodies When samples are not preserved due to effervescence this will be indicated on the chain of custody

Delivery of Data

Data is required to be delivered to MampE within thirty-five (35) days of laboratory receipt of the last sample per sample delivery group of ten (10) samples or less The data must be delivered to Dr Brian Tucker Metcalf amp Eddy with the mailing address 30 Harvard Mill Square PO Box 4071 Wakefield MA 01880-5371

ANALYTICAL REFERENCES

EPA Method 5242 Methods for the Determination of Organic Compounds in Drinking Water EPA-6004-88039 December 1988 This method is provided in Attachment C

SAMPLE PRESERVATION AND DETECTION LIMITS ^ bull bull i t a i ^

Parameter

All analytes listed in Table 1 of EPA Method 5242

Acetone

2-Butanone

2-Hexanone

4-Methyl-2-pentanone

Carbon disulfide

Tetrahydrofuran

14-Dioxane

ANALYTICAL PROCEDURE

Detection Limit

lt 1 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 1 ugL

lt 50 ugL

lt 50 ugL

Preservation

pH lt2 with HCI at 4degC

EPA Method 5242 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas ChromatographyMass Spectrometry from Methods for the Determination of Organic Compounds in Drinking Water EPA-6(X)4-88039 December 1988 modified as instructed in the following paragraphs

A The following compounds must be included for analysis in all applicable aspects of the method

Compound

Acetone 2-Butanone 2-Hexanone 4-Methyl-2-pentanone Carbon disulfide Tetrahydrofuran 14-Dioxane

CAS Number

67-64-1 78-93-3 591-78-6 108-10-1 75-15-0 109-99-9 123-9-1

Primary Ion

43 72 43 43 76 42 88

Secondarv lon(s)

58 57 57 58 78

7172 58

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

QUALITY ASSURANCE PLAN ADDENDUM

v CHESHIRE GROUNDWATER CONTAMINATION SUPERFUND SITE

US EPA REGION I

EPA Remedial Project Managers Signature

EPA Remedial Project Manager Jane Dolan

Deputy Program Quality Assurance Officers Signature

Deputy Program Quality Assurance Officer Meg Himmel

ARCS Lead Chemists Signature S^^^ J 2 Kff^^

ARCS Lead Chemist Dr Brian Tucker

MampE Project Manager Martha Zirbel

Amended By Bruce Livingston

This Quality Assurance Plan (QAP) Addendum is an update to the original QAP for this ^

project which was filed on September 30 1994 Only those portions of the original QAP

that required updating are presented in this Addendum Unless updated with this document

all quality control procedures will be conducted in accordance with the original September

1994 QAP for this project

1 Project Name Cheshire Groundwater Contamination Superfund Site - Groundwater Sampling

2 Project Requested By EPA Region I

3 Date QAP Amendment FUed 112795

4 Date of Project Field Initiation 120495

5 EPA Remedial Project Manager Jane Dolan

6 Deputy Quality Assistance Officer Meg Himmel

7 Project Description

A Objective and Scope The objective of this sampling effort is to obtain updated analytical data to determine groundwater quality in the vicinity of the Cheshire Groundwater Contamination Superfund Site Three existing overburden monitoring wells and one abandoned residential well (Fusco well) that were sampled in November 1994 will be sampled again as part of this field effort Two additional overburden and three bedrock wells recently installed by the US Geological Survey will also be sampled All groundwater samples will be analyzed for low concentration volatile organics through the DAS program and for full inorganics through the CLP RAS system Based on previous analytical results from the site purge water generated during well sampling will be discharged to the ground surface away from each well

B Data UseData Quality Objectives The data acquired during this sampling episode will provide current information about the nature and extent of contamination by volatile organic compounds (VOCs) inorganic elements and cyanide

The VOC data generated through the DAS program will be analytical level V data while the CLP RAS full inorganic data will be analytical level IV The analytical data packages will consist of all documentation required to conduct a

^ ^ r f j M -

Tier III data validation however only Tier I validation (evidence audit) will be conducted The data generated during this field investigation will be used by EPA personnel to conduct a risk assessment for the site

1

c Monitoring Parameters and Sample Frequency of Collection A total of nine wells will be sampled and analyzed using both RAS CLP and DAS methodologies Full inorganics including the twenty three target analyte list (TAL) metals and cyanide will be analyzed using RAS procedures through the CLP system Low concentration VOCs will be analyzed using MampEs EPA approved DAS specification which is based on EPA Method 5242

In addition to the laboratory analytical procedures pH temperature conductivity turbidity and depth to groundwater will be measured at each well during purging and following sample collection

D Parameter Table

Number Analytical of Method Sample Container Field QC

Parameter Samples Matrix Reference Preservation Holding Time Samples

DAS - VOCs 9 GW 5242 2x40 ml VOA vial 14 days V HCI t opHlt2 Ice

RAS - Full Inorganics

TAL Metals 9 GW TTM030 1 liter Plastic 6 months 4 (except HN03 topHlt2

Vaa Mercury)

Mercury 9 GW ILM030 1 liter Plastic 28 days 4 HN03topHlt2

Cyanide 9 GW ILM030 1 liter Plastic 14 days A NaOHtopHgt10

Field Measurements

pH 9 GW 1501 Disposable Plastic Cup Field Analyze None

Temperature 9 GW 1701 Disposable Plastic Cup Field Analyze None

Conductivity 9 GW 1201 Disposable Plastic Cup Field Analyze None

Turbidity 9 GW 1801 Disposable Plastic Cup Field Analyze None

1) Low concentration VOCs will be analyzed using the MampE DAS specification based on EPA Method 5242 A copy of the DAS Specification is provided at the end of this QAP Addendum

2) QC saniples for VOCs include one MSMSD pair (2) one field duplicate two trip blanks one equipment blank and one performance evaluation (PE) sample

3) QC samples for inorganics include one matrix spike one equipment blank one field duplicate and one PE sample

4) Methods for Chemical Analysis of Water and Wastes US EPA March 1983 f ^

Schedule of Tasks and Products

Samples are anticipated to be collected the week of December 4 1995 Following the completion of the field work MampE will prepare a draft trip memo detailing the work completed The field measurements and procedures conducted along with any deviations from the approved project plans will be provided in the draft trip memo The final trip memo will address any comments on the draft and will include all of the field and laboratory analytical data The results wUl be tabulated to facilitate evaluation

Project Organization and Responsibility

Sampling Operations Bruce Livingston Sampling QC Bruce Livingston Laboratory Analysis RASDAS Laboratories Laboratory QC RASDAS Laboratories Data Quality Review RASDAS LaboratoriesMetcalf amp Eddy Performance Auditing None Systems Auditing None MampE Deputy QA Officer Meg Himmel MampE Project Manager Martha Zirbel

10 Laboratory and Field Data Quality RequirementsAssessments W

Estimated Estimated Parameter Matrix Detection Limit Accuracy Precision

VOCs GW 1-5 ngV 80-120 lt30 RPD

TAL Metals GW 2-100 MgL 80-120 lt30 RPD

Cyanide GW 10 MgL 75-125 lt30 RPD

pff GW NA 005 units plusmn01 units

Temperature^ GW NA 005degC plusmn01degC

Conductivity GW NA lOMmhocm plusmn10

Turbidity GW NA plusmn10 plusmn20

1) The detection limits presented are achievable for the referenced methods Due to interferences requiring dilutions these detection limits may not always be met

2) The detection limits for the majority of VOCs are 1-5 MgL Tetra hydrofuran and 14-Dioxane have detection limits of 50 MgL

3) The field measurements accuracy and precision are based on the equipment manufacturer specifications and MampE field experience

10AB No changes

11 No changes except that VOCs will be analyzed through DAS

12 13 14 15A No changes

15B Data Reduction and Reporting Data reduction and reporting will be conducted by the RAS and DAS laboratories

After receiving the data packages MampE will conduct Tier I (evidence audit only) validation The data will be reported in a spreadsheet format

16 Data Validation Tier I data validation will be conducted However all documentation necessary to conduct a Tier III data validation will be provided to meet the project data quality objectives

17 18 19 No changes

20 Reports to EPA Following the completion of groundwater sampling MampE will prepare a draft trip memo identifying the work completed and presenting the field analytical data The draft trip memo will be completed within two weeks of the conclusion of sampling Laboratory data should arrive at MampE five weeks after the conclusion of sampling The fmal trip memo will include a tabulation of all of the laboratory analytical data The final trip memo will be completed within three weeks of receipt of laboratory analytical data by MampE

Attaclunents to Original QAP No changes

ANALYTICAL SPECIFICATION FOR

METHOD 5242

SCOPE

This specification is for the analysis of groundwater surface water or other aqueous matrices for low concentration volatile organic compounds water using EPA Method 5242 The additional analytes added to the list of 5242 compounds are acetone 2-butanone 2-hexanone 4-methyl-2-pentanone carbon disulfide tetrahydrofuran and 14-dioxane The complete list of target compounds is presented in Attachment A In addition to the target compounds up to ten tentatively identified compounds must also be reported

Method 5242 requires the laboratory to perform a method detection limit (MDL) smdy prior to sample analysis to determine the theoretical lowest statistically valid detection limit This study must follow the requirements described in 40 CFR Part 136 Appendix B using at least seven replicate analyses of the lowest calibration concentration The study must meet the detection limits described in Section 6 The continuing calibration must be conducted on an eight hour basis no other time span is acceptable The laboratory performing the work will be subject to audit by Metcalf amp Eddy at the discretion of the Metcalf amp Eddy lead chemist

PURPOSE

The data derived from these analyses will be used to further delineate groundwater contamination to determine if treatment has met human health and ecological risk objectives to determine the quality of data generated by PRP consultants and possibly for other applications

This method is intended for trace analysis of volatiles in waters that are only slightly impacted by pollution and is not intended for use in analysis of industrial effluent landfill leachate or any moderately to highly contaminated surface waters contaminated groundwater or treatability study influents If samples are collected from areas with historical data concentration ranges will be provided to the laboratory in the work order If the samples are found to effervesce during addition of acid for sample preservation the samples will not be preserved and the laboratory will be notified concerning the absence of preservative

DEFINITION OF WORK

Surface water groundwatier or other aqueous matrices are to be analyzed by Method 5242 with the modifications presented within The modifications are described in Section 7 Analytical Procedures

c Performance evaluation samples (PEs) will be provided with up to one in each sample delivery group to be analyzed along with the field samples No more than twenty samples will be submitted in a sample delivery group Field samples will include blind trip bottle and equipment blanks as well as blind duplicates

SCHEDULE

Samples will be shipped at most one day after collection Saturday delivery may be required An overnight delivery service will be used Contacts for shipping will be provided in each work order Data delivery inquiries may be made to Dr Brian Tucker Metcalf amp Eddy Inc Wakefield MA 01880 (617) 246-5200 extension 6433

Holding Time

Analyses are required to be performed within fourteen (14) days of sample collection for preserved samples and seven (7) days for unpreserved samples Sample preservation will be noted on the chain of custodies When samples are not preserved due to effervescence this will be indicated on the chain of custody

Delivery of Data

Data is required to be delivered to MampE within thirty-five (35) days of laboratory receipt of the last sample per sample delivery group of ten (10) samples or less The data must be delivered to Dr Brian Tucker Metcalf amp Eddy with the mailing address 30 Harvard Mill Square PO Box 4071 Wakefield MA 01880-5371

ANALYTICAL REFERENCES

EPA Method 5242 Methods for the Determination of Organic Compounds in Drinking Water EPA-6004-88039 December 1988 This method is provided in Attachment C

SAMPLE PRESERVATION AND DETECTION LIMITS ^ bull bull i t a i ^

Parameter

All analytes listed in Table 1 of EPA Method 5242

Acetone

2-Butanone

2-Hexanone

4-Methyl-2-pentanone

Carbon disulfide

Tetrahydrofuran

14-Dioxane

ANALYTICAL PROCEDURE

Detection Limit

lt 1 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 1 ugL

lt 50 ugL

lt 50 ugL

Preservation

pH lt2 with HCI at 4degC

EPA Method 5242 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas ChromatographyMass Spectrometry from Methods for the Determination of Organic Compounds in Drinking Water EPA-6(X)4-88039 December 1988 modified as instructed in the following paragraphs

A The following compounds must be included for analysis in all applicable aspects of the method

Compound

Acetone 2-Butanone 2-Hexanone 4-Methyl-2-pentanone Carbon disulfide Tetrahydrofuran 14-Dioxane

CAS Number

67-64-1 78-93-3 591-78-6 108-10-1 75-15-0 109-99-9 123-9-1

Primary Ion

43 72 43 43 76 42 88

Secondarv lon(s)

58 57 57 58 78

7172 58

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

This Quality Assurance Plan (QAP) Addendum is an update to the original QAP for this ^

project which was filed on September 30 1994 Only those portions of the original QAP

that required updating are presented in this Addendum Unless updated with this document

all quality control procedures will be conducted in accordance with the original September

1994 QAP for this project

1 Project Name Cheshire Groundwater Contamination Superfund Site - Groundwater Sampling

2 Project Requested By EPA Region I

3 Date QAP Amendment FUed 112795

4 Date of Project Field Initiation 120495

5 EPA Remedial Project Manager Jane Dolan

6 Deputy Quality Assistance Officer Meg Himmel

7 Project Description

A Objective and Scope The objective of this sampling effort is to obtain updated analytical data to determine groundwater quality in the vicinity of the Cheshire Groundwater Contamination Superfund Site Three existing overburden monitoring wells and one abandoned residential well (Fusco well) that were sampled in November 1994 will be sampled again as part of this field effort Two additional overburden and three bedrock wells recently installed by the US Geological Survey will also be sampled All groundwater samples will be analyzed for low concentration volatile organics through the DAS program and for full inorganics through the CLP RAS system Based on previous analytical results from the site purge water generated during well sampling will be discharged to the ground surface away from each well

B Data UseData Quality Objectives The data acquired during this sampling episode will provide current information about the nature and extent of contamination by volatile organic compounds (VOCs) inorganic elements and cyanide

The VOC data generated through the DAS program will be analytical level V data while the CLP RAS full inorganic data will be analytical level IV The analytical data packages will consist of all documentation required to conduct a

^ ^ r f j M -

Tier III data validation however only Tier I validation (evidence audit) will be conducted The data generated during this field investigation will be used by EPA personnel to conduct a risk assessment for the site

1

c Monitoring Parameters and Sample Frequency of Collection A total of nine wells will be sampled and analyzed using both RAS CLP and DAS methodologies Full inorganics including the twenty three target analyte list (TAL) metals and cyanide will be analyzed using RAS procedures through the CLP system Low concentration VOCs will be analyzed using MampEs EPA approved DAS specification which is based on EPA Method 5242

In addition to the laboratory analytical procedures pH temperature conductivity turbidity and depth to groundwater will be measured at each well during purging and following sample collection

D Parameter Table

Number Analytical of Method Sample Container Field QC

Parameter Samples Matrix Reference Preservation Holding Time Samples

DAS - VOCs 9 GW 5242 2x40 ml VOA vial 14 days V HCI t opHlt2 Ice

RAS - Full Inorganics

TAL Metals 9 GW TTM030 1 liter Plastic 6 months 4 (except HN03 topHlt2

Vaa Mercury)

Mercury 9 GW ILM030 1 liter Plastic 28 days 4 HN03topHlt2

Cyanide 9 GW ILM030 1 liter Plastic 14 days A NaOHtopHgt10

Field Measurements

pH 9 GW 1501 Disposable Plastic Cup Field Analyze None

Temperature 9 GW 1701 Disposable Plastic Cup Field Analyze None

Conductivity 9 GW 1201 Disposable Plastic Cup Field Analyze None

Turbidity 9 GW 1801 Disposable Plastic Cup Field Analyze None

1) Low concentration VOCs will be analyzed using the MampE DAS specification based on EPA Method 5242 A copy of the DAS Specification is provided at the end of this QAP Addendum

2) QC saniples for VOCs include one MSMSD pair (2) one field duplicate two trip blanks one equipment blank and one performance evaluation (PE) sample

3) QC samples for inorganics include one matrix spike one equipment blank one field duplicate and one PE sample

4) Methods for Chemical Analysis of Water and Wastes US EPA March 1983 f ^

Schedule of Tasks and Products

Samples are anticipated to be collected the week of December 4 1995 Following the completion of the field work MampE will prepare a draft trip memo detailing the work completed The field measurements and procedures conducted along with any deviations from the approved project plans will be provided in the draft trip memo The final trip memo will address any comments on the draft and will include all of the field and laboratory analytical data The results wUl be tabulated to facilitate evaluation

Project Organization and Responsibility

Sampling Operations Bruce Livingston Sampling QC Bruce Livingston Laboratory Analysis RASDAS Laboratories Laboratory QC RASDAS Laboratories Data Quality Review RASDAS LaboratoriesMetcalf amp Eddy Performance Auditing None Systems Auditing None MampE Deputy QA Officer Meg Himmel MampE Project Manager Martha Zirbel

10 Laboratory and Field Data Quality RequirementsAssessments W

Estimated Estimated Parameter Matrix Detection Limit Accuracy Precision

VOCs GW 1-5 ngV 80-120 lt30 RPD

TAL Metals GW 2-100 MgL 80-120 lt30 RPD

Cyanide GW 10 MgL 75-125 lt30 RPD

pff GW NA 005 units plusmn01 units

Temperature^ GW NA 005degC plusmn01degC

Conductivity GW NA lOMmhocm plusmn10

Turbidity GW NA plusmn10 plusmn20

1) The detection limits presented are achievable for the referenced methods Due to interferences requiring dilutions these detection limits may not always be met

2) The detection limits for the majority of VOCs are 1-5 MgL Tetra hydrofuran and 14-Dioxane have detection limits of 50 MgL

3) The field measurements accuracy and precision are based on the equipment manufacturer specifications and MampE field experience

10AB No changes

11 No changes except that VOCs will be analyzed through DAS

12 13 14 15A No changes

15B Data Reduction and Reporting Data reduction and reporting will be conducted by the RAS and DAS laboratories

After receiving the data packages MampE will conduct Tier I (evidence audit only) validation The data will be reported in a spreadsheet format

16 Data Validation Tier I data validation will be conducted However all documentation necessary to conduct a Tier III data validation will be provided to meet the project data quality objectives

17 18 19 No changes

20 Reports to EPA Following the completion of groundwater sampling MampE will prepare a draft trip memo identifying the work completed and presenting the field analytical data The draft trip memo will be completed within two weeks of the conclusion of sampling Laboratory data should arrive at MampE five weeks after the conclusion of sampling The fmal trip memo will include a tabulation of all of the laboratory analytical data The final trip memo will be completed within three weeks of receipt of laboratory analytical data by MampE

Attaclunents to Original QAP No changes

ANALYTICAL SPECIFICATION FOR

METHOD 5242

SCOPE

This specification is for the analysis of groundwater surface water or other aqueous matrices for low concentration volatile organic compounds water using EPA Method 5242 The additional analytes added to the list of 5242 compounds are acetone 2-butanone 2-hexanone 4-methyl-2-pentanone carbon disulfide tetrahydrofuran and 14-dioxane The complete list of target compounds is presented in Attachment A In addition to the target compounds up to ten tentatively identified compounds must also be reported

Method 5242 requires the laboratory to perform a method detection limit (MDL) smdy prior to sample analysis to determine the theoretical lowest statistically valid detection limit This study must follow the requirements described in 40 CFR Part 136 Appendix B using at least seven replicate analyses of the lowest calibration concentration The study must meet the detection limits described in Section 6 The continuing calibration must be conducted on an eight hour basis no other time span is acceptable The laboratory performing the work will be subject to audit by Metcalf amp Eddy at the discretion of the Metcalf amp Eddy lead chemist

PURPOSE

The data derived from these analyses will be used to further delineate groundwater contamination to determine if treatment has met human health and ecological risk objectives to determine the quality of data generated by PRP consultants and possibly for other applications

This method is intended for trace analysis of volatiles in waters that are only slightly impacted by pollution and is not intended for use in analysis of industrial effluent landfill leachate or any moderately to highly contaminated surface waters contaminated groundwater or treatability study influents If samples are collected from areas with historical data concentration ranges will be provided to the laboratory in the work order If the samples are found to effervesce during addition of acid for sample preservation the samples will not be preserved and the laboratory will be notified concerning the absence of preservative

DEFINITION OF WORK

Surface water groundwatier or other aqueous matrices are to be analyzed by Method 5242 with the modifications presented within The modifications are described in Section 7 Analytical Procedures

c Performance evaluation samples (PEs) will be provided with up to one in each sample delivery group to be analyzed along with the field samples No more than twenty samples will be submitted in a sample delivery group Field samples will include blind trip bottle and equipment blanks as well as blind duplicates

SCHEDULE

Samples will be shipped at most one day after collection Saturday delivery may be required An overnight delivery service will be used Contacts for shipping will be provided in each work order Data delivery inquiries may be made to Dr Brian Tucker Metcalf amp Eddy Inc Wakefield MA 01880 (617) 246-5200 extension 6433

Holding Time

Analyses are required to be performed within fourteen (14) days of sample collection for preserved samples and seven (7) days for unpreserved samples Sample preservation will be noted on the chain of custodies When samples are not preserved due to effervescence this will be indicated on the chain of custody

Delivery of Data

Data is required to be delivered to MampE within thirty-five (35) days of laboratory receipt of the last sample per sample delivery group of ten (10) samples or less The data must be delivered to Dr Brian Tucker Metcalf amp Eddy with the mailing address 30 Harvard Mill Square PO Box 4071 Wakefield MA 01880-5371

ANALYTICAL REFERENCES

EPA Method 5242 Methods for the Determination of Organic Compounds in Drinking Water EPA-6004-88039 December 1988 This method is provided in Attachment C

SAMPLE PRESERVATION AND DETECTION LIMITS ^ bull bull i t a i ^

Parameter

All analytes listed in Table 1 of EPA Method 5242

Acetone

2-Butanone

2-Hexanone

4-Methyl-2-pentanone

Carbon disulfide

Tetrahydrofuran

14-Dioxane

ANALYTICAL PROCEDURE

Detection Limit

lt 1 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 1 ugL

lt 50 ugL

lt 50 ugL

Preservation

pH lt2 with HCI at 4degC

EPA Method 5242 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas ChromatographyMass Spectrometry from Methods for the Determination of Organic Compounds in Drinking Water EPA-6(X)4-88039 December 1988 modified as instructed in the following paragraphs

A The following compounds must be included for analysis in all applicable aspects of the method

Compound

Acetone 2-Butanone 2-Hexanone 4-Methyl-2-pentanone Carbon disulfide Tetrahydrofuran 14-Dioxane

CAS Number

67-64-1 78-93-3 591-78-6 108-10-1 75-15-0 109-99-9 123-9-1

Primary Ion

43 72 43 43 76 42 88

Secondarv lon(s)

58 57 57 58 78

7172 58

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

c Monitoring Parameters and Sample Frequency of Collection A total of nine wells will be sampled and analyzed using both RAS CLP and DAS methodologies Full inorganics including the twenty three target analyte list (TAL) metals and cyanide will be analyzed using RAS procedures through the CLP system Low concentration VOCs will be analyzed using MampEs EPA approved DAS specification which is based on EPA Method 5242

In addition to the laboratory analytical procedures pH temperature conductivity turbidity and depth to groundwater will be measured at each well during purging and following sample collection

D Parameter Table

Number Analytical of Method Sample Container Field QC

Parameter Samples Matrix Reference Preservation Holding Time Samples

DAS - VOCs 9 GW 5242 2x40 ml VOA vial 14 days V HCI t opHlt2 Ice

RAS - Full Inorganics

TAL Metals 9 GW TTM030 1 liter Plastic 6 months 4 (except HN03 topHlt2

Vaa Mercury)

Mercury 9 GW ILM030 1 liter Plastic 28 days 4 HN03topHlt2

Cyanide 9 GW ILM030 1 liter Plastic 14 days A NaOHtopHgt10

Field Measurements

pH 9 GW 1501 Disposable Plastic Cup Field Analyze None

Temperature 9 GW 1701 Disposable Plastic Cup Field Analyze None

Conductivity 9 GW 1201 Disposable Plastic Cup Field Analyze None

Turbidity 9 GW 1801 Disposable Plastic Cup Field Analyze None

1) Low concentration VOCs will be analyzed using the MampE DAS specification based on EPA Method 5242 A copy of the DAS Specification is provided at the end of this QAP Addendum

2) QC saniples for VOCs include one MSMSD pair (2) one field duplicate two trip blanks one equipment blank and one performance evaluation (PE) sample

3) QC samples for inorganics include one matrix spike one equipment blank one field duplicate and one PE sample

4) Methods for Chemical Analysis of Water and Wastes US EPA March 1983 f ^

Schedule of Tasks and Products

Samples are anticipated to be collected the week of December 4 1995 Following the completion of the field work MampE will prepare a draft trip memo detailing the work completed The field measurements and procedures conducted along with any deviations from the approved project plans will be provided in the draft trip memo The final trip memo will address any comments on the draft and will include all of the field and laboratory analytical data The results wUl be tabulated to facilitate evaluation

Project Organization and Responsibility

Sampling Operations Bruce Livingston Sampling QC Bruce Livingston Laboratory Analysis RASDAS Laboratories Laboratory QC RASDAS Laboratories Data Quality Review RASDAS LaboratoriesMetcalf amp Eddy Performance Auditing None Systems Auditing None MampE Deputy QA Officer Meg Himmel MampE Project Manager Martha Zirbel

10 Laboratory and Field Data Quality RequirementsAssessments W

Estimated Estimated Parameter Matrix Detection Limit Accuracy Precision

VOCs GW 1-5 ngV 80-120 lt30 RPD

TAL Metals GW 2-100 MgL 80-120 lt30 RPD

Cyanide GW 10 MgL 75-125 lt30 RPD

pff GW NA 005 units plusmn01 units

Temperature^ GW NA 005degC plusmn01degC

Conductivity GW NA lOMmhocm plusmn10

Turbidity GW NA plusmn10 plusmn20

1) The detection limits presented are achievable for the referenced methods Due to interferences requiring dilutions these detection limits may not always be met

2) The detection limits for the majority of VOCs are 1-5 MgL Tetra hydrofuran and 14-Dioxane have detection limits of 50 MgL

3) The field measurements accuracy and precision are based on the equipment manufacturer specifications and MampE field experience

10AB No changes

11 No changes except that VOCs will be analyzed through DAS

12 13 14 15A No changes

15B Data Reduction and Reporting Data reduction and reporting will be conducted by the RAS and DAS laboratories

After receiving the data packages MampE will conduct Tier I (evidence audit only) validation The data will be reported in a spreadsheet format

16 Data Validation Tier I data validation will be conducted However all documentation necessary to conduct a Tier III data validation will be provided to meet the project data quality objectives

17 18 19 No changes

20 Reports to EPA Following the completion of groundwater sampling MampE will prepare a draft trip memo identifying the work completed and presenting the field analytical data The draft trip memo will be completed within two weeks of the conclusion of sampling Laboratory data should arrive at MampE five weeks after the conclusion of sampling The fmal trip memo will include a tabulation of all of the laboratory analytical data The final trip memo will be completed within three weeks of receipt of laboratory analytical data by MampE

Attaclunents to Original QAP No changes

ANALYTICAL SPECIFICATION FOR

METHOD 5242

SCOPE

This specification is for the analysis of groundwater surface water or other aqueous matrices for low concentration volatile organic compounds water using EPA Method 5242 The additional analytes added to the list of 5242 compounds are acetone 2-butanone 2-hexanone 4-methyl-2-pentanone carbon disulfide tetrahydrofuran and 14-dioxane The complete list of target compounds is presented in Attachment A In addition to the target compounds up to ten tentatively identified compounds must also be reported

Method 5242 requires the laboratory to perform a method detection limit (MDL) smdy prior to sample analysis to determine the theoretical lowest statistically valid detection limit This study must follow the requirements described in 40 CFR Part 136 Appendix B using at least seven replicate analyses of the lowest calibration concentration The study must meet the detection limits described in Section 6 The continuing calibration must be conducted on an eight hour basis no other time span is acceptable The laboratory performing the work will be subject to audit by Metcalf amp Eddy at the discretion of the Metcalf amp Eddy lead chemist

PURPOSE

The data derived from these analyses will be used to further delineate groundwater contamination to determine if treatment has met human health and ecological risk objectives to determine the quality of data generated by PRP consultants and possibly for other applications

This method is intended for trace analysis of volatiles in waters that are only slightly impacted by pollution and is not intended for use in analysis of industrial effluent landfill leachate or any moderately to highly contaminated surface waters contaminated groundwater or treatability study influents If samples are collected from areas with historical data concentration ranges will be provided to the laboratory in the work order If the samples are found to effervesce during addition of acid for sample preservation the samples will not be preserved and the laboratory will be notified concerning the absence of preservative

DEFINITION OF WORK

Surface water groundwatier or other aqueous matrices are to be analyzed by Method 5242 with the modifications presented within The modifications are described in Section 7 Analytical Procedures

c Performance evaluation samples (PEs) will be provided with up to one in each sample delivery group to be analyzed along with the field samples No more than twenty samples will be submitted in a sample delivery group Field samples will include blind trip bottle and equipment blanks as well as blind duplicates

SCHEDULE

Samples will be shipped at most one day after collection Saturday delivery may be required An overnight delivery service will be used Contacts for shipping will be provided in each work order Data delivery inquiries may be made to Dr Brian Tucker Metcalf amp Eddy Inc Wakefield MA 01880 (617) 246-5200 extension 6433

Holding Time

Analyses are required to be performed within fourteen (14) days of sample collection for preserved samples and seven (7) days for unpreserved samples Sample preservation will be noted on the chain of custodies When samples are not preserved due to effervescence this will be indicated on the chain of custody

Delivery of Data

Data is required to be delivered to MampE within thirty-five (35) days of laboratory receipt of the last sample per sample delivery group of ten (10) samples or less The data must be delivered to Dr Brian Tucker Metcalf amp Eddy with the mailing address 30 Harvard Mill Square PO Box 4071 Wakefield MA 01880-5371

ANALYTICAL REFERENCES

EPA Method 5242 Methods for the Determination of Organic Compounds in Drinking Water EPA-6004-88039 December 1988 This method is provided in Attachment C

SAMPLE PRESERVATION AND DETECTION LIMITS ^ bull bull i t a i ^

Parameter

All analytes listed in Table 1 of EPA Method 5242

Acetone

2-Butanone

2-Hexanone

4-Methyl-2-pentanone

Carbon disulfide

Tetrahydrofuran

14-Dioxane

ANALYTICAL PROCEDURE

Detection Limit

lt 1 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 1 ugL

lt 50 ugL

lt 50 ugL

Preservation

pH lt2 with HCI at 4degC

EPA Method 5242 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas ChromatographyMass Spectrometry from Methods for the Determination of Organic Compounds in Drinking Water EPA-6(X)4-88039 December 1988 modified as instructed in the following paragraphs

A The following compounds must be included for analysis in all applicable aspects of the method

Compound

Acetone 2-Butanone 2-Hexanone 4-Methyl-2-pentanone Carbon disulfide Tetrahydrofuran 14-Dioxane

CAS Number

67-64-1 78-93-3 591-78-6 108-10-1 75-15-0 109-99-9 123-9-1

Primary Ion

43 72 43 43 76 42 88

Secondarv lon(s)

58 57 57 58 78

7172 58

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

Schedule of Tasks and Products

Samples are anticipated to be collected the week of December 4 1995 Following the completion of the field work MampE will prepare a draft trip memo detailing the work completed The field measurements and procedures conducted along with any deviations from the approved project plans will be provided in the draft trip memo The final trip memo will address any comments on the draft and will include all of the field and laboratory analytical data The results wUl be tabulated to facilitate evaluation

Project Organization and Responsibility

Sampling Operations Bruce Livingston Sampling QC Bruce Livingston Laboratory Analysis RASDAS Laboratories Laboratory QC RASDAS Laboratories Data Quality Review RASDAS LaboratoriesMetcalf amp Eddy Performance Auditing None Systems Auditing None MampE Deputy QA Officer Meg Himmel MampE Project Manager Martha Zirbel

10 Laboratory and Field Data Quality RequirementsAssessments W

Estimated Estimated Parameter Matrix Detection Limit Accuracy Precision

VOCs GW 1-5 ngV 80-120 lt30 RPD

TAL Metals GW 2-100 MgL 80-120 lt30 RPD

Cyanide GW 10 MgL 75-125 lt30 RPD

pff GW NA 005 units plusmn01 units

Temperature^ GW NA 005degC plusmn01degC

Conductivity GW NA lOMmhocm plusmn10

Turbidity GW NA plusmn10 plusmn20

1) The detection limits presented are achievable for the referenced methods Due to interferences requiring dilutions these detection limits may not always be met

2) The detection limits for the majority of VOCs are 1-5 MgL Tetra hydrofuran and 14-Dioxane have detection limits of 50 MgL

3) The field measurements accuracy and precision are based on the equipment manufacturer specifications and MampE field experience

10AB No changes

11 No changes except that VOCs will be analyzed through DAS

12 13 14 15A No changes

15B Data Reduction and Reporting Data reduction and reporting will be conducted by the RAS and DAS laboratories

After receiving the data packages MampE will conduct Tier I (evidence audit only) validation The data will be reported in a spreadsheet format

16 Data Validation Tier I data validation will be conducted However all documentation necessary to conduct a Tier III data validation will be provided to meet the project data quality objectives

17 18 19 No changes

20 Reports to EPA Following the completion of groundwater sampling MampE will prepare a draft trip memo identifying the work completed and presenting the field analytical data The draft trip memo will be completed within two weeks of the conclusion of sampling Laboratory data should arrive at MampE five weeks after the conclusion of sampling The fmal trip memo will include a tabulation of all of the laboratory analytical data The final trip memo will be completed within three weeks of receipt of laboratory analytical data by MampE

Attaclunents to Original QAP No changes

ANALYTICAL SPECIFICATION FOR

METHOD 5242

SCOPE

This specification is for the analysis of groundwater surface water or other aqueous matrices for low concentration volatile organic compounds water using EPA Method 5242 The additional analytes added to the list of 5242 compounds are acetone 2-butanone 2-hexanone 4-methyl-2-pentanone carbon disulfide tetrahydrofuran and 14-dioxane The complete list of target compounds is presented in Attachment A In addition to the target compounds up to ten tentatively identified compounds must also be reported

Method 5242 requires the laboratory to perform a method detection limit (MDL) smdy prior to sample analysis to determine the theoretical lowest statistically valid detection limit This study must follow the requirements described in 40 CFR Part 136 Appendix B using at least seven replicate analyses of the lowest calibration concentration The study must meet the detection limits described in Section 6 The continuing calibration must be conducted on an eight hour basis no other time span is acceptable The laboratory performing the work will be subject to audit by Metcalf amp Eddy at the discretion of the Metcalf amp Eddy lead chemist

PURPOSE

The data derived from these analyses will be used to further delineate groundwater contamination to determine if treatment has met human health and ecological risk objectives to determine the quality of data generated by PRP consultants and possibly for other applications

This method is intended for trace analysis of volatiles in waters that are only slightly impacted by pollution and is not intended for use in analysis of industrial effluent landfill leachate or any moderately to highly contaminated surface waters contaminated groundwater or treatability study influents If samples are collected from areas with historical data concentration ranges will be provided to the laboratory in the work order If the samples are found to effervesce during addition of acid for sample preservation the samples will not be preserved and the laboratory will be notified concerning the absence of preservative

DEFINITION OF WORK

Surface water groundwatier or other aqueous matrices are to be analyzed by Method 5242 with the modifications presented within The modifications are described in Section 7 Analytical Procedures

c Performance evaluation samples (PEs) will be provided with up to one in each sample delivery group to be analyzed along with the field samples No more than twenty samples will be submitted in a sample delivery group Field samples will include blind trip bottle and equipment blanks as well as blind duplicates

SCHEDULE

Samples will be shipped at most one day after collection Saturday delivery may be required An overnight delivery service will be used Contacts for shipping will be provided in each work order Data delivery inquiries may be made to Dr Brian Tucker Metcalf amp Eddy Inc Wakefield MA 01880 (617) 246-5200 extension 6433

Holding Time

Analyses are required to be performed within fourteen (14) days of sample collection for preserved samples and seven (7) days for unpreserved samples Sample preservation will be noted on the chain of custodies When samples are not preserved due to effervescence this will be indicated on the chain of custody

Delivery of Data

Data is required to be delivered to MampE within thirty-five (35) days of laboratory receipt of the last sample per sample delivery group of ten (10) samples or less The data must be delivered to Dr Brian Tucker Metcalf amp Eddy with the mailing address 30 Harvard Mill Square PO Box 4071 Wakefield MA 01880-5371

ANALYTICAL REFERENCES

EPA Method 5242 Methods for the Determination of Organic Compounds in Drinking Water EPA-6004-88039 December 1988 This method is provided in Attachment C

SAMPLE PRESERVATION AND DETECTION LIMITS ^ bull bull i t a i ^

Parameter

All analytes listed in Table 1 of EPA Method 5242

Acetone

2-Butanone

2-Hexanone

4-Methyl-2-pentanone

Carbon disulfide

Tetrahydrofuran

14-Dioxane

ANALYTICAL PROCEDURE

Detection Limit

lt 1 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 1 ugL

lt 50 ugL

lt 50 ugL

Preservation

pH lt2 with HCI at 4degC

EPA Method 5242 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas ChromatographyMass Spectrometry from Methods for the Determination of Organic Compounds in Drinking Water EPA-6(X)4-88039 December 1988 modified as instructed in the following paragraphs

A The following compounds must be included for analysis in all applicable aspects of the method

Compound

Acetone 2-Butanone 2-Hexanone 4-Methyl-2-pentanone Carbon disulfide Tetrahydrofuran 14-Dioxane

CAS Number

67-64-1 78-93-3 591-78-6 108-10-1 75-15-0 109-99-9 123-9-1

Primary Ion

43 72 43 43 76 42 88

Secondarv lon(s)

58 57 57 58 78

7172 58

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

10AB No changes

11 No changes except that VOCs will be analyzed through DAS

12 13 14 15A No changes

15B Data Reduction and Reporting Data reduction and reporting will be conducted by the RAS and DAS laboratories

After receiving the data packages MampE will conduct Tier I (evidence audit only) validation The data will be reported in a spreadsheet format

16 Data Validation Tier I data validation will be conducted However all documentation necessary to conduct a Tier III data validation will be provided to meet the project data quality objectives

17 18 19 No changes

20 Reports to EPA Following the completion of groundwater sampling MampE will prepare a draft trip memo identifying the work completed and presenting the field analytical data The draft trip memo will be completed within two weeks of the conclusion of sampling Laboratory data should arrive at MampE five weeks after the conclusion of sampling The fmal trip memo will include a tabulation of all of the laboratory analytical data The final trip memo will be completed within three weeks of receipt of laboratory analytical data by MampE

Attaclunents to Original QAP No changes

ANALYTICAL SPECIFICATION FOR

METHOD 5242

SCOPE

This specification is for the analysis of groundwater surface water or other aqueous matrices for low concentration volatile organic compounds water using EPA Method 5242 The additional analytes added to the list of 5242 compounds are acetone 2-butanone 2-hexanone 4-methyl-2-pentanone carbon disulfide tetrahydrofuran and 14-dioxane The complete list of target compounds is presented in Attachment A In addition to the target compounds up to ten tentatively identified compounds must also be reported

Method 5242 requires the laboratory to perform a method detection limit (MDL) smdy prior to sample analysis to determine the theoretical lowest statistically valid detection limit This study must follow the requirements described in 40 CFR Part 136 Appendix B using at least seven replicate analyses of the lowest calibration concentration The study must meet the detection limits described in Section 6 The continuing calibration must be conducted on an eight hour basis no other time span is acceptable The laboratory performing the work will be subject to audit by Metcalf amp Eddy at the discretion of the Metcalf amp Eddy lead chemist

PURPOSE

The data derived from these analyses will be used to further delineate groundwater contamination to determine if treatment has met human health and ecological risk objectives to determine the quality of data generated by PRP consultants and possibly for other applications

This method is intended for trace analysis of volatiles in waters that are only slightly impacted by pollution and is not intended for use in analysis of industrial effluent landfill leachate or any moderately to highly contaminated surface waters contaminated groundwater or treatability study influents If samples are collected from areas with historical data concentration ranges will be provided to the laboratory in the work order If the samples are found to effervesce during addition of acid for sample preservation the samples will not be preserved and the laboratory will be notified concerning the absence of preservative

DEFINITION OF WORK

Surface water groundwatier or other aqueous matrices are to be analyzed by Method 5242 with the modifications presented within The modifications are described in Section 7 Analytical Procedures

c Performance evaluation samples (PEs) will be provided with up to one in each sample delivery group to be analyzed along with the field samples No more than twenty samples will be submitted in a sample delivery group Field samples will include blind trip bottle and equipment blanks as well as blind duplicates

SCHEDULE

Samples will be shipped at most one day after collection Saturday delivery may be required An overnight delivery service will be used Contacts for shipping will be provided in each work order Data delivery inquiries may be made to Dr Brian Tucker Metcalf amp Eddy Inc Wakefield MA 01880 (617) 246-5200 extension 6433

Holding Time

Analyses are required to be performed within fourteen (14) days of sample collection for preserved samples and seven (7) days for unpreserved samples Sample preservation will be noted on the chain of custodies When samples are not preserved due to effervescence this will be indicated on the chain of custody

Delivery of Data

Data is required to be delivered to MampE within thirty-five (35) days of laboratory receipt of the last sample per sample delivery group of ten (10) samples or less The data must be delivered to Dr Brian Tucker Metcalf amp Eddy with the mailing address 30 Harvard Mill Square PO Box 4071 Wakefield MA 01880-5371

ANALYTICAL REFERENCES

EPA Method 5242 Methods for the Determination of Organic Compounds in Drinking Water EPA-6004-88039 December 1988 This method is provided in Attachment C

SAMPLE PRESERVATION AND DETECTION LIMITS ^ bull bull i t a i ^

Parameter

All analytes listed in Table 1 of EPA Method 5242

Acetone

2-Butanone

2-Hexanone

4-Methyl-2-pentanone

Carbon disulfide

Tetrahydrofuran

14-Dioxane

ANALYTICAL PROCEDURE

Detection Limit

lt 1 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 1 ugL

lt 50 ugL

lt 50 ugL

Preservation

pH lt2 with HCI at 4degC

EPA Method 5242 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas ChromatographyMass Spectrometry from Methods for the Determination of Organic Compounds in Drinking Water EPA-6(X)4-88039 December 1988 modified as instructed in the following paragraphs

A The following compounds must be included for analysis in all applicable aspects of the method

Compound

Acetone 2-Butanone 2-Hexanone 4-Methyl-2-pentanone Carbon disulfide Tetrahydrofuran 14-Dioxane

CAS Number

67-64-1 78-93-3 591-78-6 108-10-1 75-15-0 109-99-9 123-9-1

Primary Ion

43 72 43 43 76 42 88

Secondarv lon(s)

58 57 57 58 78

7172 58

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

ANALYTICAL SPECIFICATION FOR

METHOD 5242

SCOPE

This specification is for the analysis of groundwater surface water or other aqueous matrices for low concentration volatile organic compounds water using EPA Method 5242 The additional analytes added to the list of 5242 compounds are acetone 2-butanone 2-hexanone 4-methyl-2-pentanone carbon disulfide tetrahydrofuran and 14-dioxane The complete list of target compounds is presented in Attachment A In addition to the target compounds up to ten tentatively identified compounds must also be reported

Method 5242 requires the laboratory to perform a method detection limit (MDL) smdy prior to sample analysis to determine the theoretical lowest statistically valid detection limit This study must follow the requirements described in 40 CFR Part 136 Appendix B using at least seven replicate analyses of the lowest calibration concentration The study must meet the detection limits described in Section 6 The continuing calibration must be conducted on an eight hour basis no other time span is acceptable The laboratory performing the work will be subject to audit by Metcalf amp Eddy at the discretion of the Metcalf amp Eddy lead chemist

PURPOSE

The data derived from these analyses will be used to further delineate groundwater contamination to determine if treatment has met human health and ecological risk objectives to determine the quality of data generated by PRP consultants and possibly for other applications

This method is intended for trace analysis of volatiles in waters that are only slightly impacted by pollution and is not intended for use in analysis of industrial effluent landfill leachate or any moderately to highly contaminated surface waters contaminated groundwater or treatability study influents If samples are collected from areas with historical data concentration ranges will be provided to the laboratory in the work order If the samples are found to effervesce during addition of acid for sample preservation the samples will not be preserved and the laboratory will be notified concerning the absence of preservative

DEFINITION OF WORK

Surface water groundwatier or other aqueous matrices are to be analyzed by Method 5242 with the modifications presented within The modifications are described in Section 7 Analytical Procedures

c Performance evaluation samples (PEs) will be provided with up to one in each sample delivery group to be analyzed along with the field samples No more than twenty samples will be submitted in a sample delivery group Field samples will include blind trip bottle and equipment blanks as well as blind duplicates

SCHEDULE

Samples will be shipped at most one day after collection Saturday delivery may be required An overnight delivery service will be used Contacts for shipping will be provided in each work order Data delivery inquiries may be made to Dr Brian Tucker Metcalf amp Eddy Inc Wakefield MA 01880 (617) 246-5200 extension 6433

Holding Time

Analyses are required to be performed within fourteen (14) days of sample collection for preserved samples and seven (7) days for unpreserved samples Sample preservation will be noted on the chain of custodies When samples are not preserved due to effervescence this will be indicated on the chain of custody

Delivery of Data

Data is required to be delivered to MampE within thirty-five (35) days of laboratory receipt of the last sample per sample delivery group of ten (10) samples or less The data must be delivered to Dr Brian Tucker Metcalf amp Eddy with the mailing address 30 Harvard Mill Square PO Box 4071 Wakefield MA 01880-5371

ANALYTICAL REFERENCES

EPA Method 5242 Methods for the Determination of Organic Compounds in Drinking Water EPA-6004-88039 December 1988 This method is provided in Attachment C

SAMPLE PRESERVATION AND DETECTION LIMITS ^ bull bull i t a i ^

Parameter

All analytes listed in Table 1 of EPA Method 5242

Acetone

2-Butanone

2-Hexanone

4-Methyl-2-pentanone

Carbon disulfide

Tetrahydrofuran

14-Dioxane

ANALYTICAL PROCEDURE

Detection Limit

lt 1 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 1 ugL

lt 50 ugL

lt 50 ugL

Preservation

pH lt2 with HCI at 4degC

EPA Method 5242 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas ChromatographyMass Spectrometry from Methods for the Determination of Organic Compounds in Drinking Water EPA-6(X)4-88039 December 1988 modified as instructed in the following paragraphs

A The following compounds must be included for analysis in all applicable aspects of the method

Compound

Acetone 2-Butanone 2-Hexanone 4-Methyl-2-pentanone Carbon disulfide Tetrahydrofuran 14-Dioxane

CAS Number

67-64-1 78-93-3 591-78-6 108-10-1 75-15-0 109-99-9 123-9-1

Primary Ion

43 72 43 43 76 42 88

Secondarv lon(s)

58 57 57 58 78

7172 58

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

SCOPE

This specification is for the analysis of groundwater surface water or other aqueous matrices for low concentration volatile organic compounds water using EPA Method 5242 The additional analytes added to the list of 5242 compounds are acetone 2-butanone 2-hexanone 4-methyl-2-pentanone carbon disulfide tetrahydrofuran and 14-dioxane The complete list of target compounds is presented in Attachment A In addition to the target compounds up to ten tentatively identified compounds must also be reported

Method 5242 requires the laboratory to perform a method detection limit (MDL) smdy prior to sample analysis to determine the theoretical lowest statistically valid detection limit This study must follow the requirements described in 40 CFR Part 136 Appendix B using at least seven replicate analyses of the lowest calibration concentration The study must meet the detection limits described in Section 6 The continuing calibration must be conducted on an eight hour basis no other time span is acceptable The laboratory performing the work will be subject to audit by Metcalf amp Eddy at the discretion of the Metcalf amp Eddy lead chemist

PURPOSE

The data derived from these analyses will be used to further delineate groundwater contamination to determine if treatment has met human health and ecological risk objectives to determine the quality of data generated by PRP consultants and possibly for other applications

This method is intended for trace analysis of volatiles in waters that are only slightly impacted by pollution and is not intended for use in analysis of industrial effluent landfill leachate or any moderately to highly contaminated surface waters contaminated groundwater or treatability study influents If samples are collected from areas with historical data concentration ranges will be provided to the laboratory in the work order If the samples are found to effervesce during addition of acid for sample preservation the samples will not be preserved and the laboratory will be notified concerning the absence of preservative

DEFINITION OF WORK

Surface water groundwatier or other aqueous matrices are to be analyzed by Method 5242 with the modifications presented within The modifications are described in Section 7 Analytical Procedures

c Performance evaluation samples (PEs) will be provided with up to one in each sample delivery group to be analyzed along with the field samples No more than twenty samples will be submitted in a sample delivery group Field samples will include blind trip bottle and equipment blanks as well as blind duplicates

SCHEDULE

Samples will be shipped at most one day after collection Saturday delivery may be required An overnight delivery service will be used Contacts for shipping will be provided in each work order Data delivery inquiries may be made to Dr Brian Tucker Metcalf amp Eddy Inc Wakefield MA 01880 (617) 246-5200 extension 6433

Holding Time

Analyses are required to be performed within fourteen (14) days of sample collection for preserved samples and seven (7) days for unpreserved samples Sample preservation will be noted on the chain of custodies When samples are not preserved due to effervescence this will be indicated on the chain of custody

Delivery of Data

Data is required to be delivered to MampE within thirty-five (35) days of laboratory receipt of the last sample per sample delivery group of ten (10) samples or less The data must be delivered to Dr Brian Tucker Metcalf amp Eddy with the mailing address 30 Harvard Mill Square PO Box 4071 Wakefield MA 01880-5371

ANALYTICAL REFERENCES

EPA Method 5242 Methods for the Determination of Organic Compounds in Drinking Water EPA-6004-88039 December 1988 This method is provided in Attachment C

SAMPLE PRESERVATION AND DETECTION LIMITS ^ bull bull i t a i ^

Parameter

All analytes listed in Table 1 of EPA Method 5242

Acetone

2-Butanone

2-Hexanone

4-Methyl-2-pentanone

Carbon disulfide

Tetrahydrofuran

14-Dioxane

ANALYTICAL PROCEDURE

Detection Limit

lt 1 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 1 ugL

lt 50 ugL

lt 50 ugL

Preservation

pH lt2 with HCI at 4degC

EPA Method 5242 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas ChromatographyMass Spectrometry from Methods for the Determination of Organic Compounds in Drinking Water EPA-6(X)4-88039 December 1988 modified as instructed in the following paragraphs

A The following compounds must be included for analysis in all applicable aspects of the method

Compound

Acetone 2-Butanone 2-Hexanone 4-Methyl-2-pentanone Carbon disulfide Tetrahydrofuran 14-Dioxane

CAS Number

67-64-1 78-93-3 591-78-6 108-10-1 75-15-0 109-99-9 123-9-1

Primary Ion

43 72 43 43 76 42 88

Secondarv lon(s)

58 57 57 58 78

7172 58

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

c Performance evaluation samples (PEs) will be provided with up to one in each sample delivery group to be analyzed along with the field samples No more than twenty samples will be submitted in a sample delivery group Field samples will include blind trip bottle and equipment blanks as well as blind duplicates

SCHEDULE

Samples will be shipped at most one day after collection Saturday delivery may be required An overnight delivery service will be used Contacts for shipping will be provided in each work order Data delivery inquiries may be made to Dr Brian Tucker Metcalf amp Eddy Inc Wakefield MA 01880 (617) 246-5200 extension 6433

Holding Time

Analyses are required to be performed within fourteen (14) days of sample collection for preserved samples and seven (7) days for unpreserved samples Sample preservation will be noted on the chain of custodies When samples are not preserved due to effervescence this will be indicated on the chain of custody

Delivery of Data

Data is required to be delivered to MampE within thirty-five (35) days of laboratory receipt of the last sample per sample delivery group of ten (10) samples or less The data must be delivered to Dr Brian Tucker Metcalf amp Eddy with the mailing address 30 Harvard Mill Square PO Box 4071 Wakefield MA 01880-5371

ANALYTICAL REFERENCES

EPA Method 5242 Methods for the Determination of Organic Compounds in Drinking Water EPA-6004-88039 December 1988 This method is provided in Attachment C

SAMPLE PRESERVATION AND DETECTION LIMITS ^ bull bull i t a i ^

Parameter

All analytes listed in Table 1 of EPA Method 5242

Acetone

2-Butanone

2-Hexanone

4-Methyl-2-pentanone

Carbon disulfide

Tetrahydrofuran

14-Dioxane

ANALYTICAL PROCEDURE

Detection Limit

lt 1 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 1 ugL

lt 50 ugL

lt 50 ugL

Preservation

pH lt2 with HCI at 4degC

EPA Method 5242 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas ChromatographyMass Spectrometry from Methods for the Determination of Organic Compounds in Drinking Water EPA-6(X)4-88039 December 1988 modified as instructed in the following paragraphs

A The following compounds must be included for analysis in all applicable aspects of the method

Compound

Acetone 2-Butanone 2-Hexanone 4-Methyl-2-pentanone Carbon disulfide Tetrahydrofuran 14-Dioxane

CAS Number

67-64-1 78-93-3 591-78-6 108-10-1 75-15-0 109-99-9 123-9-1

Primary Ion

43 72 43 43 76 42 88

Secondarv lon(s)

58 57 57 58 78

7172 58

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

SAMPLE PRESERVATION AND DETECTION LIMITS ^ bull bull i t a i ^

Parameter

All analytes listed in Table 1 of EPA Method 5242

Acetone

2-Butanone

2-Hexanone

4-Methyl-2-pentanone

Carbon disulfide

Tetrahydrofuran

14-Dioxane

ANALYTICAL PROCEDURE

Detection Limit

lt 1 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 5 ugL

lt 1 ugL

lt 50 ugL

lt 50 ugL

Preservation

pH lt2 with HCI at 4degC

EPA Method 5242 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas ChromatographyMass Spectrometry from Methods for the Determination of Organic Compounds in Drinking Water EPA-6(X)4-88039 December 1988 modified as instructed in the following paragraphs

A The following compounds must be included for analysis in all applicable aspects of the method

Compound

Acetone 2-Butanone 2-Hexanone 4-Methyl-2-pentanone Carbon disulfide Tetrahydrofuran 14-Dioxane

CAS Number

67-64-1 78-93-3 591-78-6 108-10-1 75-15-0 109-99-9 123-9-1

Primary Ion

43 72 43 43 76 42 88

Secondarv lon(s)

58 57 57 58 78

7172 58

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

c B A Matrix Spike and Matrix Spike Duplicate must be analyzed for each sample

delivery group or twenty (20) field samples whichever is more frequent The spike solution must contain the following compounds Vinyl chloride 12-Dichloroethane Carbon tetrachloride 12-Dichloropropane Trichloroethene 112-Trichloroethane Benzene cis-13-Dichloropropene Bromoform Tetrachloroethene 12-Dibromomethane and 14-Dichlorobenzene with 250 ng25 ml of each The solution must also contain 2-Hexanone at a concentration of 625 ng25 ml

C M O and P xylenes must be reported as a total The dichlorobenzene isomers and cistrans 12-dichloroethene isomers must be resolved and reported as separate isomers

D For l2-dichloroethane-d4 mz 67 can be used as the quantitation ion instead of mz 65 However the same quantitation ion must be used throughout the analytical sequence beginning with the MDL

E- One laboratory fortified blank (LFB) containing all analytes at a concentration of 1 Mg1 5Mgl for ketones and 50 Mg1 for tetrahydrofuran and 14-dioxane must be analyzed daily prior to sample analysis The parent solution must be from a source other than the initial calibration standard and continuing calibration standard All surrogates and intemal standards in the LFB must meet the recovery criteria presented in Section 8

F Up to ten nontarget nonintemal standard nonsurtogate organic compounds of the greatest apparent concentration must be tentatively identified via a forward search of the NIST Mass Spectral Library Compounds with a total ion area greater than 40 percent of the total ion area of the nearest intemal standard are required to be library searched Once the best library matches are obtained a mass spectral interpretation specialist must assign an identification if possible If a specific compound cannot be identified a group or class of compound such as C7 alkane or methyl ethyl benzene isomer may be reported Tentatively identified compounds (TICs) are to be quantitated using total ion areas using the nearest eluting intemal standard and assuming a response factor of one Do not report TICs if the concentration is less than 20 (igL

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

QUALITY CONTROL REQUIREMENTS

The following are the audits required frequency of audits QC limits and required corrective actions

Audits

GCMS Tuning

Intemal Standards

Surrogate Recovery

Initial Calibration

Frequency

Every (8) hours

All samples stds blanks QC samples

All samples cal stds blks QC samples

Initially and when continuing calibration criteria not met

Limits

Table 3 5242 (See 9a)

RT shifts lt 30 seconds Quant Ion area -50 to -1-100 from latest 8 hr std

12-dichoroshyethane-d4 83-143 12-dichloroshybenzene-d4 80shy120 Both surrogates must be in criteria in all samples cal stds blks QC samples

RSD lt 30 RRF gt 005 Ketones 12 dibromoshy3-chloropropane RSD lt30 RFgt002

Cortecfive Action

Remne the GCMS Proper tuning must be verified prior to analysis of stds blanks or samples

Inspect for instmment malfunction Correct problem remn all affected samples blanks QC samples

Remn all samples stds blks or ( C samples with any one surtogate outside criteria Holding tunes must be met If surtogates still out submit both analyses

Determine problem must meet criteria prior to analysis of blanks and samples Contact the MampE lead chemist

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

r - Audits Frequency

Cont Calibration Every 8 hours

Lab Fortified Blank Daily prior to sample analysis

Method Blank Prior to each 8 hour shift

Matrix spike Every 20 samples or Matrix spike dup every SDG

Limits

RRF - Same as initial cal Percent Dlt30

Percent recovery must be 60-140 for all compounds 5 compounds can exceed 140 if not detected in field samples None less than 60 recovery

Target analytes lt 1 Mg1 lt 5 Mg1 for acetone 2-butanone 4-methyl 2shypentanone 2-hexanone THF and 14-dioxane lt 50 MgL

80-120 recovery 13 RPD Maximum (See 7b)

Corrective Action

Reanalyze CC std if still out initial calibrafion must be perfomied CC criteria must be met prior to blank and sample analysis

Reanalyze to meet 60-140 Remn all samples affected LFB criteria must be met prior to sample analysis on a daily basis

Must be met prior to sample analysis

Reanalyze MSMSD once if criteria still not met submit both analyses

9 ANALYTICAL DELIVERABLES

A The data package deliverables must resemble as closely as possible the Routine Analytical Services (RAS) Organic SOW OLM019 or the latest revision data package The fonhs provided in the RAS Organic SOW OLM019 or the latest revision must be utilized where appropriate The data qualifiers utilized in the RAS Organic SOW OLM019 or the latest revision must be applied to the data provided

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

A narrative must be provided describing the procedure performed by the laboratory the volume of sample purged and any deviations from the method Problems encountered during analysis problem resolution and any factors which may affect the validity of the data must be addressed This specification signed and dated chain of custody documentation sample tags shipping airbills and telephone logs must be included The data package must be paginated and good copy quality is required

The raw data must be provided for all blanks spikes standards PE samples and field samples and must include

bull reconstmcted total ion chromatograms

bull instmment quantitation reports containing the following information laboratory sample identification number MampE sample number date and time of analysis RT or scan number of identified Method 5242 or additional compound ion used for quantitation with measured area analyte concentration copy of area table from data system GCMS instmment ID lab file ID column trap composition and operating conditions

bull raw and enhanced spectra for all positive results with standard reference spectra for comparison must be provided

The surtogate results must be provided in tabular format on a modified RAS Form II All standards blanks samples and QC samples which were analyzed must be reported on the form The results must be flagged according to the RAS SOW procedures

The matrix spike and matrix spike duplicate sample results for all non-spiked compounds must be tabulated on RAS Form I modified to include all compounds analyzed for in the Region I Modified Method 5242 The concentration added percent recoveries and relative percent differences must be reported on a modified RAS Form III for the spiked compounds The results must be flagged according to the procedures presented in Section 8 of this specification Raw data must be included Mass spectra for the spiked compounds are not required

The method blanks and associated samples must be summarized on RAS Form IV

^^ ^

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

^

lt-^

bull The mning results must be in tabular format reported on RAS Form V The instmment normalized mass listing and the mass spectmm must be

provided

bull The initial calibration results must also be reported in tabular format on a modified RAS Form VI The relative response factors and the percent relative standard deviation must be calculated for all analytes and surtogates The concentration and source of the standards analyzed must be provided Raw data must be included for all standards analyzed The spectra for all compounds in the initial calibration standards do not have to be included

bull The 10 ugL continuing calibration standard must be reported in a tabular format on modified RAS Form VII The raw data must be included The percent differences and daily response factors must be reported for all analytes and surtogates A RAS Form VII must be provided if the initial calibration standard curve 10 ppb standard was used for quantitation of samples The spectmm for each compound in the continuing calibration standard does not have to be included

bull The Laboratory Fortified Blank (LFB) must be reported in tabular format on a RAS Form I modified to include all compounds presented

in Attachment A Raw data must be included excluding the spectra of all spike analytes The percent recoveries must be calculated and spike

concentrations simimarized on the modified RAS Form I The source of the spike and the acceptable recovery limits must also be reported

bull Provide the intemal standards results on a modified RAS Form VIII The retention time and area counts of the quantification ion for the intemal standards in all blanks samples and QC samples must be reported on this form

bull An example calculation must be provided for positive results and detection limit values reported

bull Provide copies of records (telecons) of communication with field personnel the work order designated project chemist or the MampE lead chemist

bull For TICs provide the spectrum of the TIC and the tiiree best matches obtained from the library search A RAS Form I-TIC must be included for each sample even if no TICs are reported

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

B Complete Sample Delivery Group File (CSF) Audit

Due to the litigative nature of each Superfund site Region I EPA requires that all analytical data quality control data and tabulated raw or supporting data be delivered with each sample delivery group (SDG) With each sample delivery group an EPA Region 1 Complete SDG File Completeness Evidence Audit must be completed The CSF Completeness Evidence Audit Forms are included in Attachment B and must accompany each sample delivery group (data package) The laboratory through these audit forms must demonstrate that all sample data raw data calibration data and any other data requirements of this specification are included by the laboratory in the data package

The forms included in Attachment B are for all types of data packages For this specification the laboratory will use the forms supplied to the best of their ability where deliverable items are applicable

10 EXCEPTIONS

If QC requirements or action limits are exceeded or if analytical samples are destroyed or lost or if matrix interference is suspected contact

Dr Brian Tucker Metcalf amp Eddy Inc (617) 246-5200 ext 6433

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

ATTACHMENT A

Fluorobenzene - Internal Standard

Dichlorodifluoromethane Chloromethane Vinyl Chloride Bromomethane Chloroethane Trichlorofluoromethane 11-Dichloroethene Acetone Carbon Disulfide Methylene Chloride trans-12-Dichloroethene 11-Dichloroethane 22-Dichloropropane 2-Butanone cis-12-Dichloroethene Chloroform Bromochloromethane 111 -Trichloroethane

gtw 11-Dichloropropene Carbon Tetrachloride Benzene 12-Dichloroethane Trichloroethene 12-Dicliloropropane Bromodichloromethane Dibromomethane 4-Methyl-2-Pentanone trans-13-Dichloropropene Toluene cis-13-Dichloropropene 112-Trichloroethane 2-Hexanone Tetrachloroethene 13-Dichloropropane Tetrahydrofuran 14-Dioxane l2-Dichloroethane-d4 (surtogate)

10

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

s Chlorobenzene-d5 - Intemal Standard

Dibromochloromethane Bromoform Isopropylbenzene 12-Dibromoethane Chlorobenzene Ethylbenzene 1112-Tetrachloroethane 1122-Tetrachloroethane Total Xylenes Styrene Bromobenzene N-Propylbenzene 123 -Trichloropropane 2-Chlorotoluene 135-Trimethy Ibenzene 4-Chlorotoluene Tert-Butylbenzene 124-Trimethylbenzene Sec-Butylbenzene 4-Isopropyltoluene 13-Dichlorobenzene 14-Dichlorobenzene N-Butylbenzene 12-Dichlorobenzene 12-Dibromo-3-Chloropropane 124-Trichlorobenzene Hexachlorobutadiene Naphthalene 123 -Trichlorobenzene 12-Dichlorobenzene-d4 (surrogate)

11

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

r - ATTACHMENT B

CSF AUDIT FORMS

bull s laquo raquo ^

12

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

~ ^

ORGANICS COHPLETE SDG FILE (CS7) INVENTORY SHEET (Cont)

SDG NOS TO FOLLOW CASE NO SDC NO SAS NO

PACE NOs CHECK FROM TO LAB EPA

Mlaeallanaotia Pata

Original preparaclon and analysis forms or copies of praparatlon and analysis logbook pages

Internal saaple and saaple extract transfer chain-of-custody records

Screening records All instrument output including strip charts

from screening activities (describe or list)

8 EPA ShlPDlniyRaeelvtng Doetments

Airbills (No of shipments ) Chain-of-Custody Records Sample Tags Sample Log-In Sheet (Lab amp DCl) Miscellaneous ShippingReceiving Records

(describe or list) --i9 ssjg

^ ^ - bullbullbullbullgtbullbullraquobullbullbull

9 Intemal Lab Sample Transfer Records and Tracking Sheets (describe or list)

10 Other Records (describe or list)

Telephone Conmunication Log

11 Conments

ompleted by (CLP Lab) (Signature) (Printed NameTitle) (Dace)

bull bulldi ted by EPA) ( S i g n a t u r e ) ( P r i n t e d N a m e T i d e ) (Dace)

FORM D C - 2 - 4

0LM017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cont) V

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

PAGE NOs CHECK FROM TO LAB EPA

6 Pesticides (cont) b Saaple Data

TCL Results - Organic Analysis Data Sheet (Fora I PEST)

Chromatograms (Primary Column) Chromatograms from second GC column confirmation GC Integration report or data system printout

Manual work sheets For pestieldesAroelors confirmed by GCMS copies of rav spectra and copies of background-subtracted mass spectra of target corapowids (saaples amp standards)

c Standards Data Initial Calibration of Single Component Analytes (Fora VI PEST-1 and PEST-2)

Initial Calibration of Miilticoaponent Analytes (Fora VI PEST-3)

Analyte Resolution Sumaary (Fora VI FEST-4) _^ Calibration Verification Sunoary (Fora VII PEST-1)

x-v Callbratioti Verification Suaaary (Fora VII PEST-2) i^^ Analytical Sequence (Fora VIII PEST)

Florlsll Cartridge Cheek (Fora IX PEST-1) Pesticide 6PC Calibration (Fora DC PEST-2) Pesticide Identification Sumaary for Single Component

Analytes (Form X PEST-1) Pesticide Identification Summary for Multicomponent

Analytes (Fora X PEST-2) Chroaatograos and data systea printouts

A printout of retention tiaes and corresponding peak areas or peak heights

Pesticide CPC calibration data shy UV detector traces

d Raw QC Data Blank Data Matrix SpikeMatrix Spike Duplicate Data

FORM DC-2-3 OLH017 791

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

ORGANICS COMPLETE SDC FILE (CSF) INVENTORY SHEET (Cone)

CASE NO

bull

SDG NO SAS NO

SDG NOS TO FOLLOW

PAGE NOs CHECK FROl TO LAB EPA

Semivolatiles Data a QC Sumaary

bull Surrogate Percent Recovery Sunmary (Fora II SV) MSMSD Summary (Form III SV) Method Blank Summary (Form IV SV)GCMS Instrument Perforaance Check (Fora V SV)

Intemal Standard Area and RT Summary (Fora VIII SV)

~

b Sample Data TCL Results (Fora I SV-l SV2) Tentatively Identified Coapounds (Fora I SV-TIC) Reconstrticted total ion chromatograas (RIG)

for eaeh saaple For each sample Raw spectra and background-subtracted mass spectra o f target compounds

Quantitation reports Mass spectra of TICs vith three best library matches GPC chromatograms (if GPC perfomed)

c Standards Data (All Instruments) Initial Calibration Data (Fora VI SV-l SV-2) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII SV-l SV-2) RICs and Quantitation Reports for all Standards Seaivolatlle GPC Calibration Data-UV detector traces

d Raw QC Data DFTPF Blank DataMatrix SpikeMatrix Spike Duplicate Data

~ ~

Pesticides

a QC Summary Surrogate Percent Recovery Summary (Form II PEST) MSMSD Duplicate Summary (Form III PEST) Method Blank Summary (Form IV PEST)

v FORM DC-2-2

0LN017 791

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET c LABORATORY NAME CITYSTATE

CASE NO SDG NO SDG NOS TO FOLLOW SAS NO

CONTRACT NO SOV MO

All documents delivered in the complete SOG file must be original documents vhere possible (REFERENCE EXHIBIT B SECTION II and SECTION III)

PAGE NOs CHECK FROM TO LAB EPA

1 Inventorr Sheet (Fora DC-2) (Do not nuaber) 2 SDC Case Narrative 3 SDQ Cover SheetTraffic Report 4 Volatiles Data

a QC Summary Systea Monitoring Coapound Summary (Fora II VOA)

Matrix SpikeMatrix Spike Duplicate Summary (Fora III VOA)

Method Blank Summary (Fora IV VOA) GCMS Instrument Performance Cheek (Fora V VOA)

Intemal Standard Area and RT Summary r(Form VIII VOA) NlaquoWlaquo

b Sampla Data TCL Results bull (Fora I VOA) Tentatively Identified Compounds (Fora I VOA-TIC)

Reconstructed total ion chrooatograas (RICgt for each saaple

For each saaple Raw spectra and background-subtracted mass spectra of target coapounds identified

Quantitation reports Mass spectra of all reported TICs with three best library matches

c Standarda Data (All Instruments) Initial Calibration Data (Fona VI VOA) RICs and Quan Reports for all Standards Continuing Calibration Data (Fora VII VOA) RICs and Quantitation Reports for all Standards

d Rav QC Data BFB Blank Data Matrix SpikeMatrix Spike Duplicate Daca

FORM DC-2-I

OLM017 7

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

ATTACHMENT C

EPA METHOD 5242

13 c

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

METWO 5242 NEASUREMEMT OF PURGEABLE ORGANIC COHPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYHASS SPECTROMaRY

Revision 30

A Alford-Stevens J W Eichelberger laquo L Budde - Method 524 Revision 10 (1983)

R U Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT

US ENVIRONNENTAL PROTECTION AGENCY CINCINNATI OHIO 45268

285 V

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

METHOO 5242 MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUWl GAS CHROHATOGRAPHYNASS SPECTROHHRY

Revision 30

A Alford-Stevens J W Elchelberger H L Budde - Method 524 Revision 10 (1983)

R W Slater Jr - Method 5242 Revision 20 (1986)

J W Elchelberger U L Budde - Method 5242 Revision 30 (1989)

ENVIRONMENTAL MONITORING SYSTEMS LABORATORY OFFICE OF RESEARCH ANO DEVELOPNENT US ENVIRONNENTAL PROTECTION AGENCY

CINCINNATI OHIO 45268

285

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

NETHOO S242

MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN UATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHYMASS SPECTROMETRY

1 SCOPE AND APPLICATION

11 This is a general purpose method for the identification and simultaneous measurement of purgeable volatile organic compounds in finished drinking water raw source water or drinking water in any treatment stage (1-2) The method is applicable to a wide range of organic compounds including the four trihalomethane disinfection by-products that have sufficiently high volatility and low water solubility to be efficiently removed from water samples with purge and trap procedures The following compounds can be determined by this method

pylt r^Vi ^ V O Compound

BenzeneBromobenzeneBromochloromethaneBromodichloromethaneBromoformBromomethanen-Butyl benzenesec-Butyl benzenetert-ButylbenzeneCarbon tetrachlorideChlorobenzeneChloroethaneChloroformChloromethane2-Chlorotoluene4-ChlorotolueneDibromochloromethanel2-Dibromo-3-chloropropane12-DibromoethaneDibromomethane12-Dichlorobenzene13-Dichlorobenzene14-DichlorobenzeneDichlorodifluoromethane11-Dichloroethane12-Dichloroethane11-Dichloroethenecis-12-Dichloroethenetrans-12-Dichloroethene12-Dichloropropane13-Dichloropropane

Chemical Abstract Service Registry Number

71-43-2 108-86-1

74-97-5 75-27-4

75-25-2 74-83-9

104-51-8 135-98-8 98-06-6

56-23-5 108-90-7 75-00-3

67-66-3 74-87-3 95-49-8 106-43-4

124-48-1 96-12-8

106-93-4 74-95-3

95-50-1 541-73-1 106-46-7

75-71-8 75-34-3 107-06-2 75-35-4

156-59-4 156-60-5

78-87-5 142-28-9

286

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

22-Oichloropropane11-01chioropropenec1s-l3-D1chloropropenetrans-13-OichloropropeneEthylbenzeneHexachlorobutadieneIsopropylbenzene4-TsopropyltolueneMethylene chlorideNaphthalenen-PropylbenzeneStyrene1112-Tetrachloroethane1122-TetrachloroethaneTetrachloroetheneToluene123-Trlchlorobenzene124-Trichlorobenzene111-Trichloroethane112-TrichloroethaneTrichloroetheneTrichlorofluoromethane123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

590-20-7 563-58-6

10061-01-5 10061-02-6

100-41-4 87-68-3

98-82-8 99-87-6 75-09-2

91-20-3 103-65-1

100-42-5 630-^0-6 79-34-5

127-18-4 108-88-3

87-61-6 120-82-1

71-55-6 79-00-5

79-01-6 75-69-4 96-18-4 95-63-6 108-67-8

75-01-4 95-47-6 108-38-3 laquo 106-42-3

12 Method detection limits (MDLs) (3) are compound and Instrument dependent and vary fron approximately 002-035 i i g l The applicable concentration range of this method is primarily column dependent and is approximately 002 to 200 M9i- for the wide-bore thick-film columns Narrow-bore thin-film columns may have a capacity which limits the range to about 002 to 20 MgL Analytes that are Inefficiently purged from water will not be detected when present at low concentrations but they can be measured with acceptable accuracy and precision when present in sufficient amounts

13 Analytes that are not separated chromatographically but which have different mass spectra and non-interfering quantitation ions can be identified and measured in the same calibration mixture or water sample (Sect 1162) Analytes which have very similar mass spectra cannot be individually Identified and measured in the same calibration mixture or water sample unless they have different retention times (Sect1163) Coeluting compounds with very similarmass spectra typically many structural Isomers must be reported as an isomeric group or pair Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural isomers that may not be resolved on the capillary column and if not must be reported as isomeric pairs

287

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

2 SUWARY OF HETHQD

21 Volatile organic compounds and surrogates with low water solubility are extracted (purged) from the sample matrix by bubbling an Inert gas through the aqueous sample Purged sample components are trapped in a tube containing suitable sorbent materials When purging is complete the sorbent tube is heated and backflushed with helium to desorb the trapped sample components into a capillary gas chromatography (GC) column interfaced to a mass spectrometer (HS) The column is temperashyture programmed to separate the method analytes which are then detected with the HS Compounds eluting from the GC column are identified by comparing their measured mass spectra and retention times to reference spectra and retention times in a data base Reference spectra and retention times for analytes are obtained by the measurement of calibration standards under the same conditions used for samples The concentration of each identified component Is measured by relating the MS response of the quantitation ion produced by that compound to the HS response of the quantitation ion produced by a compound that Is used as an internal standard Surrogate analytes whose concentrations are known in every sample are measured with the same internal standard calibration procedure

3 DEFINITIONS

31 Internal standard shy- A pure analyte(s) added to a solution in known amount(s) and used to measure the relative responses of other method analytes and surrogates that are components of the same solution The internal standard must be an analyte that Is not a sample component

32 Surrogate analyte shy- A pure analyte(s) which Is extremely unlikely tobe found in any sample and which is added to a sample aliquot in knownamount(s) before extraction and Is measured with the same procedures used to measure other sample components The purpose of a surrogate analyte is to monitor method performance with each sample

^ ^

33 Laboratory duplicates (LDl and L02) -- Two sample aliquots taken in the analytical laboratory and analyzed separately with identical proceshydures Analyses of LDl and L02 give a measure of the precision associated with laboratory procedures but not with sample collection preservation or storage procedures

34 Field duplicates (FOI and FD2) -- Two separate samples collected at the same time and place under identical circumstances and treated exactly the same throughout field and laboratory procedures Analyses of FDl and FD2 give a measure of the precision associated with sample collection preservation and storage as well as with laboratory procedures

35 Laboratory reagent blank (LRB) -- An aliquot of reagent water that is treated exactly as a sample including exposure to all glassware equipment solvents reagents internal standards and surrogates that are used with other samples The LRB Is used to determine if method

288

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

c

analytes or other interferences are present In the laboratory environshyment the reagents or the apparatus

v

36 Field reagent blank (FRB) -- Reagent water placed in a sample container in the laboratory and treated as a sample in all respects including exposure to sampling site conditions storage preservation and all analytical procedures The purpose of the FRB is to determine if method analytes or other interferences are present In the field environment

37 Laboratory performance check solution (LPC) -- A solution of one or more compounds (analytes surrogates internal standard or other test compounds) used to evaluate the performance of the Instmment system with respect to a defined set of method criteria

38 Laboratory fortified blank (LFB) mdash An aliquot of reagent water to which known quantities of the method analytes are added In the laboratory The LFB Is analyzed exactly like a sample and its purpose Is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements at the required method detection limit

39 Laboratory fortified sample matrix (LFM) -- An aliquot of an environshymental sample to which known quantities of the method analytes are added in the laboratory The LFM is analyzed exactly like a sample and its purpose is to determine whether the sample matrix contributes bias to the analytical results The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values In the LFM corrected for background concentrashytions

310 Stock standard solution -- A concentrated solution containing a single certified standard that is a method analyte or a concentrated solution of a single analyte prepared In the laboratory with an assayed reference compound Stock standard solutions are used to prepare primary dilution standards

bullj

311 Primary dilution standard solution -- A solution of several analytes prepared in the laboratory from stock standard solutions and diluted as needed to prepare calibration solutions and other needed analyte solutions

312 Calibration standard (CAL) -- a solution prepared from the primary dilution statidard solution and stock standard Solutions of the Internal standards ^d-surrogate analytes The CAL solutions are used to calibrate th^Instrument response with respect to analyte concentrashytion

313 Quality control sam0U (QCS) -- a sample matrix containing method analytes or a solution of method analytes In a water miscible solvent which is used to fortifytreagent water or environmental samples The QCS is obtained from a source external to the laboratory and is used

289

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

to check laboratory performance with externally prepared test materials

4 INTERFERENCES

41 During analysis major contaminant sources are volatile materials in the laboratory and impurities in the inert purging gas and in the sorbent trap The use of non-polytetrafluoroethylene (PTFE) plastic tubing non-PTFE thread sealants or flow controllers with rubber components in the purging device should be avoided since such materials out-gas organic compounds which will be concentrated in the trap during the purge operation Analyses of laboratory reagent blanks provide Information about the presence of contaminants When potential Interfering peaks are noted In laboratory reagent blanks the analyst should change the purge gas source and regenerate the molecular sieve purge gas filter Subtracting blank values from sample results Is not permitted

42 Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after a sample containing relatively high concentrations of volatile organic compounds A preventive technique Is between-sample rinsing of the purging apparatus and sample syringes with two portions of reagent water After analysis of a sample containing high concentrations of volatile organic compounds one or more laboratory reagent blanks should be analyzed to check for cross contamination

43 Special precautions must be taken to determine methylene chloride The analytical end sample storage area should be Isolated from allatmospheric sources of methylene chloride otherwise random backgroundlevels will result Since methylene chloride will permeate through PTFE tubing all gas chromatography carrier gas lines and purge gas plumbing should be constructed of stainless steel or copper tubing Laboratory workers clothing should be cleaned frequently since clothing previously exposed to methylene chloride fumes during common liquidliquid extraction procedures can contribute to sample contaminashytion

^ ^

5 SAFETY

51 The toxicity or carcinogenicity of chemicals used in this method has not been precisely defined each chemical should be treated as a potential health hazard and exposure to these chemicals should be minimized Each laboratory Is responsible for maintaining awareness of OSHA regulations regarding safe handling of chemicals used in this method Additional references to laboratory safety are available (4-6) for the Information of the analyst

52 The following method analytes have been tentatively classified as known or suspected human or mammalian carcinogens benzene carbon tetrachloride 14-dichlorobenzene l2-4ichlorethane^ hexachloroshybutadiene 1122-tetrachloroethane 112-trlchloroethane chloroshy

290 ^

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

form 12-dibromoethanetetrachloroethene trichloroethene and vinyl ^ chloride Pure standard materials and stock standard solutions of V these compounds should be handled in a hood A NIOSHMESA approved

toxic gas respirator should be worn when the analyst handles high concentrations of these toxic compounds

6 APPARATUS AND EQUIPMENT

61 SAMPLE CONTAINERS -- 60-mL to 120-mL screw cap vials (Pierce 19832 or equivalent) each equipped with a PTFE-faced silicone septum (Pierce 12718 or equivalent) Prior to use wash vials and septa with detergent and rinse with tap and distilled water Allow the vials and septa to air dry at room temperature place In a 105degC oven for 1 hr then remove and allow to cool In an area known to be free of organics

62 PURGE AND TRAP SYSTEM -- The purge and trap system consists of three separate pieces of equipment purging device trap and desorber Systems are commercially available from several sources that meet all of the following specifications

621 The all glass purging device (Figure 1) should be designed to accept 25-mL samples with a water column at least 5 cm deep A smaller (5-mL) purging device is recommended If the GCMS system has adequate sensitivity to obtain the method detection limits required Gaseous volumes above the sample must be kept to a minimum (lt 15 mL) to eliminate dead volume effects A glass frit should be Installed at the base of the sample chamber so Cl the purge gas passes through the water column as finely divided bubbles with a diameter of lt 3 mm at the origin Needle spargers may be used however the purge gas must be introduced at a point about 5 mm from the base of the water column

622 The trap (Figure 2) must be at least 25 cm long and have an inside diameter of at least 0105 in Starting from the inlet the trap should contain 10 cm of methyl silicone coated packing and the following amounts of adsorbents 13 of 26-dlphenylene oxide polymer 13 of silica gel and 13 of coconut charcoal If it is not necessary to determine dichlorodifluoromethane the charcoal can be eliminated and the polymer increased to fill 23 of the trap Before Initial use the trap should be conditioned overnight at 180degC by backflushing with an inert gas flow of at least 20 mLmin Vent the trap effluent to the room not to the analytical column Prior to daily use the trap should be conditioned for 10 min at 180degC with backflushing The trap may be vented to the analytical column during dally conditioning however the column must be run through the temperature program prior to analysis of samples

623 The use of the methyl silicone coated packing is recommended but not mandatory The packing serves a dual purpose of protecting the Tenax adsorbant from aerosols and also of

291

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

insuring that the Tenax is fully enclosed within the heated zone of the trap thus eliminating potential cold spots Alternatively sllanized glass wool may be used as a spacer at the trap inlet

624 The desorber (Figure 2) must be capable of rapidly heating the trap to 180degC either prior to or at the beginning of the flow of desorption gas The polymer section of the trap should not be heated higher than 200degC or the life expectancy of the trap will decrease Trap failure is characterized by a pressure drop in excess of 3 pounds per square Inch across the trap during purging or by poor bromoform sensitivities The desorber design Illustrated in Fig 2 meets these criteria

63 GAS CHROMATOGRAPHYMASS SPECTROMETERDATA SYSTEM (6CMSDS)

631 The GC must be capable of temperature programming and should be equipped with variable-constant differential flow controllers so that the column flow rate will remain constant throughout desorption and temperature program operation The column oven must be cooled to lO C therefore a subambient oven controller Is required If syringe injections of BFB will be used a splitsplitless injection port is required bull

632 Capillary Gas Chromatography Columns Any gas chromatography V column that meets the perfonnance specifications of this method

may be used Separations of the calibration mixture must be equivalent or better than those described In this method Three useful columns have been identified

6321 Column 1 -- 60 m x 075 mm ID VOCOL (Supelco Inc) glass wide-bore capillary with a 15 im film thickness

Column 2 -- 30 m X 053 mm ID DB-624 (JampW Scientific Inc) fused silica capillary with a 3 im film thickshyness

Column 3 - 30 m X 032 mm ID DB-5 (JampU Scientific Inc) fused silica capillary with a 1 fun film thickshyness

633 Interfaces between the GC and HS The interface used depends on the column selected and the gas flow rate

6331 The wide-bore columns 1 and 2 have the capacity to accept the standard gas flows from the trap during thermal desorption and chromatography can begin with the onset of thermal desorption Depending on the pumping capacity of the MS an additional interface between the end of the column and the MS may be required An open split interface (7) an all-glass Jet separator or a cryogenic (Sect 6332) device

292

^

^

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

c

are acceptable interfaces Any Interface can be used If the performance specifications described in this method can be achieved The end of the transfer line after the interface or the end of the analytical column if no interface is used should be placed within a few mm of the MS ion source

6332 The narrow bore column 3 cannot accept the thermal desorption gas flow and a cryogenic interface is required This interface (Tekmar Model 1000 or equivalent) condenses the desorbed sample components at liquid nitrogen temperature and allows the helium gas to pass through to an exit The condensed components are frozen in a narrow band on an uncoated fused silica precolumn When all components have been desorbed from the trap the interface Is rapidly heated under a stream of carrier gas to transfer the analytes to the analytical column The end of the analytical column should be placed with a few mm of the HS ion source A potential problem with this interface is blockage of the interface by frozen water from the trap This condition will result in a major loss in sensitivity and chromatographic resolution

634 The mass spectrometer must be capable of electron ionization at a nominal electron energy of 70 eV The spectrometer must be capable of scanning from 35 to 260 amu with a complete scan cycle time (including scan overhead) of 2 sec or less (Scan cycle time bullgt Total HS data acquisition time in seconds divided by number of scans in the chromatogram) The spectrometer must produce a mass spectrum that meets all criteria In Table 3 when 25 ng or less of 4-bromofluorobenzene (BFB) is introduced into the GC An average spectrum across the BFB GC peak may be used to test instrument performance

635 An Interfaced data system is required to acquire store reduce and output mass spectral data The computer software should have the capability of processing stored GCMS data by recognizshying a GC peak within any given retention time window comparing the mass spectra from the GC peak with spectral data in a user-created data base and generating a list of tentatively identified compounds with their retention times and scan

UKJsnxabers The software must allow Integration of the ion abundance of any specific Ion between specified time or scan number limits The software should also allow calculation of response factors as defined in Sect 926 (or construction of a second or third order regression calibration curve) calculashytion of response factor statistics (mean and standard deviashytion) and calculation of concentrations of analytes using either the calibration curve or the equation in Sect 12

293

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

64 SYRINGE ANO SYRINGE VALVES

641 Two S-mL or 25-mL glass hypodermic syringes with Luer-Lok tip (depending on sample volume used)

642 Three 2-way syringe valves with Luer ends

643 One ZS- j i l micro syringe with a 2 in x 0006 in ID 22deg bevel needle (Hamilton I702N or equivalent)

644 Micro syringes - 10 100 n l

645 Syringes - 05 10 and 5-mL gas tight with shut-off valve

65 MISCELLANEOUS

651 Standard solution storage containers -- 15-mL bottles with PTFE-11ned screw caps

7 REAGENTS AND CONSUMABLE MATERIALS

71 TRAP PACKING MATERIALS

711 26-Diphenylene oxide polymer 6080 mesh chromatographic grade (Tenax GC or equivalent)

712 Methyl silicone packing (optional) -- OV-1 (3X) on Chromoshysorb W 6080 mesh or equivalent

713 Silica gel -- 3560 mesh Davison grade IS or equivalent 5 714 Coconut charcoal -- Prepare from Barnebey Cheney CA-580-26

lot IM-2649 by crushing through 26 mesh screen

72 REAGENTS

721 MethanoljengtiiOefflonstrated to be free of analytes

722 Reagent water -- Prepare reagent water by passing tap water through a filter bed containing about 05 kg of activated carbon by using a water purification system or by boiling distilled water for IS min followed by a 1-h purge with inert gas while the water temperature is held at 90degC Store in clean narrow-mouth bottles with PTFE-llned septa and screw caps

723 Hydrochloric acid (1+1) -- Carefully add measured volume of cone HCI to equal volume of reagent water

724 Vinyl chloride -- Certified mixtures of vinyl chloride in nitrogen and pure vinyl chloride are available from several

294

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

-^

y fH- y_

sources (for example Matheson Ideal Gas Products and Scott Gases)

725 Ascorbic acid -- ACS reagent grade granular

73 STOCK STANDARD SOLUTIONS -- These solutions may be purchased as certified solutions or prepared from pure standard materials using the following procedures One of these solutions is required for every analyte of concern every surrogate and the internal standard A useful working concentration is about 1-5 mgmL

731 Place about 98 mL of methanol into a 10-mL ground-glass stoppered volumetric flask Allow the flask to stand unstoppered for about 10 min or until all alcohol-wetted surfaces have dried and weigh to the nearest 01 mg

732 If the analyte Is a liquid at room temperature use a 100-iL syringe and Immediately add two or more drops of reference standard to the flask Be sure that the reference standard falls directly into the alcohol without contacting the neck of the flask If the analyte Is a gas at room temperature fill a 5-mL valved gas-tight syringe with the standard to the 50-mL mark lower the needle to 5 mm above the methanol meniscus and slowly inject the standard Into the neck area of the flask The gas will rapidly dissolve in the methanol

733 Reweigh dilute to volume stopper then mix by inverting the flask several times Calculate the concentration In

n q l i l from the net gain in weight When compound purity is certified at 96X or greater the weight can be used without

correction to calculate the concentration of the stock standard 734 Store stock standard solutions in 15-mL bottles equipped

with PTFE-llned screw caps Methanol solutions prepared from liquid analytes are stable for at least 4 weeks when stored at 4degC Methanol solutions prepared from gaseous analytes are not stable for more than 1 week when stored at lt0degC at room temperature they must be discarded after 1 day

74 PRIMARY DILUTION STANDARDS - Use stock standard solutions to prepare primary dilution standard solutions that contain all the analytes of concern and the surrogates (but not the Internal standard) In methanol The prlm^ifdilution standards should be prepared at concentrations that can be easily diluted to prepare aqueous calibrashytion solutions that will bracket the working concentration range Store the primary dilution standard solutions with minimal headspace and check frequently for signs of deterioration or evaporation especially Just before preparing calibration solutions Storage times described for stock standard solutions in Sect 744 also apply to primary dilution standard solutions

295

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

75 FORTIFICATION SOLUTIONS FOR INTERNAL STANDARD AND SURROGATES

751 A solution containing the Internal standard and the surrogates is required to prepare laboratory reagent blanks (also used as a laboratory performance check solution) and to fortify each sample Prepare a fortification solution containing fluoroshybenzene (internal standard) 12- dichlorobenzene-d4 (surrogate) and BFB (surrogate) in methanol at concentrations of 5 xgmL of each A 5-ML aliquot of this solution added to a 25-mL water sample volume gives concentrations of 1 f t q l of each A S-^L aliquot of this solution added to a 5-mL water sample volume gives a concentration of 5 M Q A of each) Additional Internal standrds and surrogate analytes are optional

752 A solution of the Internal standard alone Is required to prepare calibration standards and laboratory fortified blanks The internal standard should be In methanol at a concentration of 5 MgmL

76 PREPARATION OF LABORATORY REAGENT BLANK shy- Fill a 25-mL (or 5-mL) syringe with reagent water and adjust to the mark (no air bubbles) Inject 10 n l of the fortification solution containing the internal standard and surrogates through the Luer Lok valve Into the reagent water Transfer the LRB to the purging device See Sect 1112

77 PREPARATION OF LABORATORY FORTIFIED BLANK shy Prepare this exactly like a calibration standard (Sect 78) This Is a calibration standard that is treated as a sample

78 PREPARATION OF CALIBRATION STANDARDS i i l0^^

- X 0

781

3

The number of calibration solutions (CALs) needed depends on the calibration range desired A minimum of three CAL solutions is required to calibrate a range of a factor of 20 In concentrashytion For a factor of 50 use at least four standards and for a factor of 100 at least five standards One calibration

1-So 5

standard should contain each analyte of concern and each surrogate at a concentration of 2-10 times the method detection limit (Tables 4-6) for that compound The other CAL standards should contain each analyte of concern and each surrogate at concentrations that define the range of the method Every CAL solution contains the Intemal standard at the same concentrashytion (5 M9L suggested for a 5-mL sample 1 i i g l for a 25-mL sample)

782 To prepare a calibration standard add an appropriate volume of a primary dilution standard (containing analytes and surrogates) to an aliquot of reagent water In a volumetric flask Use a microsyringe and rapidly inject the methanol solutions into the expanded area of the filled volumetric flask Remove the needle as quickly as possible after injection Mix by Inverting the

296 ^ m ^

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

^^flask three times only Discard the contents contained in the

neck of the flask Aqueous standards are not stable In a volumetric flask and should be discarded after 1 hr unless transferred to a sample bottle and sealed immediately

8 SAMPLE COLLECTION PRESERVATION AND STORAGE

81 SAMPLE COLLECTION DECHLORINATION AND PRESERVATION

811 Collect all samples In duplicate If samples contain residual chlorine and measurements of the concentrations of disinfection by-products (trihalomethanes etc) at the time of sample collection are desired add about 25 mg of ascorbic acid to the sample bottle before filling Fill sample bottles to overflowshying but take care not to flush out the rapidly dissolving ascorbic acid No air bubbles should pass through the sample as the bottle is filled or be trapped in the sample when the bottle is sealed Adjust the pH of the duplicate samples to lt2 by carefully adding one drop of 11 HCI for each 20 mL of sample volume Seal the sample bottles PFTE-face down and shake vigorously f o r 1 min

812 When sampling from a water tap open the tap and allow the system to flush until the water temperature has stabilized (usually about 10 min) Adjust the flow to about 500 mLmin and collect duplicate samples from the flowing stream

bull - shy

bdquo ^ 813 When sampling from an open body of water fill a 1-quart wide-mouth bottle or l-11ter beaker with sample from a representative area and carefully fill duplicate sample bottles from the 1-quart container

814 The samples must be chilled to 4degC on the day of collection and maintained at that temperature until analysis Field samples that will not be received at the laboratory on the day of collection must be packaged for shipment with sufficient ice to ensure that they will be at 4degC on arrival at the laboratory

82 SAMPLE STORAGE

821 Store samples at 4degC until analysis The sample storage area must be free of organic solvent vapors

822 Analyze all samples within 14 days of collection Samples not analyzed within this period must be discarded and replaced

83 FIELD REAGENT BLANKS

831 Duplicate field reagent blanks must be handled along with each sample set which is composed of the samples collected from the same general sample site at approximately the same time At the laboratory fill field blank sample bottles with reagent

V 297

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

water seal and ship to the sampling site along with empty sample bottles and back to tht laboratory with filled sample bottles Wherever a set of samples Is shipped and stored it is accompanied by appropriate blanks

832 Use the same procedures used for samples to add ascorbic acid and HCI to blanks (Sect 811)

9 CALIBRATION

91 Demonstration and documentation of acceptable initial calibration is required before any samples are analyzed and Is required intermittently throughout sample analysis as dictated by results of continuing calibration checks After Initial calibration Is successful a continuing calibration check Is required at the beginning of each 8 hr period during which analyses are performed Additional periodic calibration checks are good laboratory practice

92 Initial calibration

921 Calibrate the mass and abundance scales of the MS with calibrashytion compounds and procedures prescribed by the manufacturer with any modifications necessary to meet the requirements in Sect 922

922 Introduce Into the GC (either by purging a laboratory reagent blank o r making a syringe Injection) 25 ng of BFB and acquire mass spectra for ivz 35-260 at 70 eV (nominal) Use the purging procedure andor GC conditions given In Sect Hi If thespectrum does not meet all criteria In Table 2raquo the MS must be retuned and adjusted to meet all criteria before proceeding with calibration An average spectrum across the GC peak may be used to evaluate the performance of the system

923 Purge a medium CAL solution for example 10-20 |tgL using the procedure given In Sect 11

924 Performance criteria for the medium calibration Examine the stored GCMS data with the data system software Figure 3 shows an acceptable total ion chromatogram

9241 GC performance Good column performance will produce symmetrical peaks with minimum tailing for most compounds If peaks are broad or sensitivity poor see Sect 936 for some possible remedial actions

9242 MS sensitivity The GCMSOS peak Identification software should be able to recognize a GC peak in the appropriate retention time window for each of the compounds In calibration solution and make correct tentative Identifications If fewer than 99 of the

^ ^

298

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

compounds are recognized system maintenance is laquo-lt- required See Sect 936

925 If all performance criteria are met purge an aliquot of each of the other CAL solutions using the same GCMS conditions

926 Calculate a response factor (RF) for each analyte surrogate and isomer pair for each CAL solution using the internal standard fluorobenzene Table 1 contains suggested quantitation ions for all compounds This calculation Is supported in acceptable GCMS data system software (Sect 634) and many other software programs RF is a unitless number but units used to express quantities of analyte and Internal standard must be equivalent

R F (Ax)(Qis) (AisXQx)

where Ax - integrated abundance of the quantitation ion of the analyte

A^s - Integrated abundance of the quantitation ion of the Intemal standard

Qx quantity of analyte purged In ng or concentration units

Q^s laquo quantity of internal standard purged in ng or concentration units

W YQ 9261 For each analyte and surrogate calculate the mean RF -gt-^ from the analyses of the CAL solutions Calculate the ^ lty poundgtS0 standard deviation (SD) and the relative standard

^ 70o deviation (RSD) from each mean RSD - 100 (SDM) If the RSD of any analyte or surrogate mean RF exceeds 20 either analyze additional aliquots of appropriate CAL solutions to obtain an acceptable RSD of RFs over the entire concentration range or take action to improve GCMS performance See Sect 927

o gti(^ ^ ^ 927 As an alternative to calculating mean response factors and eS ^ ^ applying the RSD test use the GCMS data system software or

C^^ other available software to generate a second or third order regression calibration curve

93 Continuing calibration check Verify the MS tune and initial calibrashytion at the beginning of each 8-hr work shift during which analyses are performci using the following procedure

931 Introduce into the GC (either by purging alaboratory reagent blank or making a syringe injection) 25 ng of BFB and acquire a mass spectrum that includes data for mz 35-260 If the spectrum does not meet all criteria (Table 2 ) the MS must be

- 299

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

932

933

934

p 935^ J

JL- ^O^ Vgtbull V Qgt^^

f v ^ ^C^C

936

retuned and adjusted to meet all criteria before proceeding with the continuing calibration check

Purge a medium concentration CAL solution and analyze with the same conditions used during the initial calibration

Demonstrate acceptable performance for the criteria shown in Sect 924

Determine that the absolute areas of the quantitation ions of the internal standard and surrogates havenot decreased by more than 30 from the areas measured In the most recent continuing calibration check o r by more than 50 from the areas measured during Initial calibration If these areas have decreased by more than these amounts adjustments must be made to restore system sensitivity These adjustments may require cleaning of the HS ion source or other maintenance as Indicated In Sect 936 and recalibration Control charts are useful aids in documenting system sensitivity changes

Calculate the RF for each analyte and surrogate from the data measured In the continuing calibration check The RF for each analyte and surrogate must be within 30 of the mean value measured In the initial calibration Alternatively If a second

or third order regression Is used the point from the continuing calibration check for each analyte and surrogate must fall

within the analysts Judgement on the curve from the initial calibration If these conditions do not exist remedial action must be taken which may require re-Initial calibration gtlaquo

Some possible remedial actions Major maintenance such as cleaning an Ion source cleaning quadrupole rods etc require returning to the initial calibration step

9361 Check and adjust GC andor MS operating conditions check the NS resolution and calibrate the mass scale

9362 Clean or replace the splitless injection liner silanize a new injection liner

9363 Flush the GC column with solvent according to manushyfacturers instructions

9364 Break off a short portion (about 1 meter) of the column from the end near the Injector o replace GC column This action will cause a change In retention times

9365 Prepare fresh CAL solutions and repeat the initial calibration step

9366 Clean the MS ion source and rods (if a quadrupole)

300

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

9367 Replace any components that allow analytes to come into contact with hot metal surfaces

9368 Replace the MS electron multiplier or any other faulty components

94 Optional calibration for vinyl chloride using a certified gaseous mixture of vinyl chloride in nitrogen can be accomplished by the following steps

941 Fill the purging device with 250 mL (or 5-mL) of reagent water or aqueous calibration standard

942 Start to purge the aqueous mixture Inject a known volume (between 100 and 2000 M L ) of the calibration gas (at room temperature) directly into the purging device with a gas tight syringe Slowly Inject the gaseous sample through a septum seal at the top of the purging device at 2000 Mtm1n If the injection of the standard is made through the aqueous sample inlet port flush the dead volume with several mL of room air or carrier gas Inject the gaseous standard before 5 min of the 11-min purge time have elapsed

943 Determine the aqueous equivalent concentration of vinyl chloride standard In i t q l Injected with the equation

S - 0102 (C)(V)

where S - Aqueous equivalent concentration of vinyl chloride standard in MgL

C - Concentration of gaseous standard in ppm (vv) V - Volume of standard injected In milliliters

10 QUALITY CONTROL

101 Quality control (QC) requirements are the initial demonstration of laboratory capability followed by regular analyses of laboratory reagent blanks field reagent blanks and laboratory fortified blanks The laboratory must maintain records to document the quality of the data generated Additional quality control practices are recommended

102 Initial demonstration of low system background Before any samples are analyzed It must be demonstrated that a laboratory reagent blank (LRB) Is reasonably free of contamination that would prevent the determination of any analyte of concern Sources of background contamination are glassware purge gas sorbants and equipment Background contamination must be reduced to an acceptable level before proceeding with the next section In general background from method analytes should be below the method detection limit

301

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

103 Initial demonstration of laboratory accuracy and precision Analyze five to seven replicates of a laboratory fortified blank containing each analyte of concem at a concentration In the range of 02-5 M9L (see regulations and maximum contaminant levels for guidance on appropriate concentrations)

1031 Prepare each replicate by adding an appropriate aliquot of a quality control sample to reagent water If a quality control sample containing the method analytes is not available a primary dilution standard made from a source of reagents different than those used to prepare the calibration standards may be used Also add the appropriate amounts of internal standard and surrogates If they are being used Analyze each replicate according to the procedures described In Section 11 and on a schedule that results In the analyses of all replicates over a period of several days

1032 Calculate the measured concentration of each analyte in each replicate the mean concentration of each analyte in all replicates and mean accuracy (as mean percentage of true value) for each analyte and the precision (as relative standard deviation RSD) of the measurements for each analyte Calculate the HDL of each analyte using the procedures described in Sect^ 132 (2)

10^33 For each analyte and surrogate the mean accuracy expressed as a percentage of the true value should be 80-120 and the RSD should be lt20 Some analytes particularly the early eluting gases and late eluting higher molecular weight gt-lt compounds are measured wjth less accampracy and precision than ^ other analytes The method detection limits must be sufficient to detect analytes at the required levels If these criteria are not met for an analyte take remedial action and repeat the measurements for that analyte to demonstrate acceptable performance before samples are analyzed

1034 Develop and maintain a system of control charts to plot the precision and accuracy of analyte and surrogate measurements as a function of time Charting of surrogate recoveries is an especially valuable activity since these are present in eyery sample and the analytical results will form a signishyficant record of data quality

104 Monitor the Integrated areas of the quantitation ions of the Internal standards and surrogates In continuing calibration checks These should remain reasonably constant over time A drift of more than 50 in any area Is indicative of a loss In sensitivity and the problem must be found and corrected These Integrated areas should also be reasonably constant In laboratory fortified blanks and samples

302

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

bullIf-J

105 Laboratory reagent blanks Uith each batch of samples processed as a group within a work shift analyze a laboratory reagent blank to determine the background system contamination A FRB (Sect 107) may be used in place of a LRB

106 With each batch of samples processed as a group within a work shift analyze a single laboratory fortified blank (LFB) containing each analyte of concern at a concentration as determined in 103 If more than 20 samples are included in a batch analyze one LFB for every 20 samples Use the procedures described in 1033 to evaluate the accuracy of the measurements and to estimate whether the method detection limits can be obtained If acceptable accuracy and method detection limits cannot be achieved the problem must be located and corrected before further samples are analyzed Add these results to the on-going control charts to document data quality

107 With each set of field samples a field reagent blank (FRB) should be analyzed The results of these analyses will help define contaminashytion resulting from field sampling and transportation activities If the FRB shows unacceptable contamination a LRB must be measured to define the source of the impurities

108 At least quarterly replicates of laboratory fortified blanks should be analyzed to determine the precision of the laboratory measureshyments Add these results to the on-going control charts to document data quality

109 At least quarterly analyze a quality control sample (QCS) from an external source If measured analyte concentrations are not of acceptable accuracy check the entire analytical procedure to locate and correct the problem source

1010 Sample matrix effects have not been observed when this method is used with distilled water reagent water drinking water and ground water Therefore analysis of a laboratory fortified sample matrix (LFM) is not required It is recommended that sample matrix effects be evaluated at least quarterly using the QCS described in 109

1011 Numerous other quality control measures are incorporated into other parts of this procedure and serve to alert the analyst to potential problems

11 PROCEDURE

111 SAMPLE INTRODUCTION AND PURGING

1111 This method is designed for a 25-mL sample volume but a smaller (5 mL) sample volume is recommended if the GCMS system has adequate sensitivity to achieve the required method detection limits Adjust the purge gas (nitrogen or helium) flow rate to 40 mLmin Attach the trap inlet to the

303

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

purging device and open the syringe valve on the purging device t

1112 Remove the plungers from two 25-mL (or 5-mL depending onsample size) syringes and attach aclosed syringe valve toeach Warm the sample to room temperatureopen the sample bottle and carefully pour the sample into one of the syringe barrels to Just short of overflowing Replace the syringe plunger invert the syringe and compress the sample Open the syringe valve and vent any residual air while adjusting the sample volume to 250-mL (or 5-mL) For samples and blanks add 5-ML of the fortification solution containing the Internal standard and the surrogates to the sample through the syringe valve For calibration standards and laboratory fortified blanks add S-ML of the fortification solution containing the Internal standard only Close the valve Fill the second syringe In an Identical manner from the same sample bottle Reserve this second syringe for a reanalysis if necessary

^ ^

1113 Attach the sample syringe valve to the syringe valve on the purging device Be sure that the trap is cooler than 25degC then open the sample syringe valve and inject the sample into the purging chamber Close both valves and initiate purging Purge the sample for 110 min at ambient temperature

112 SAMPLE DESORPTION

1121 Non-cryogenic Interface shy- After the 11-min purge place the purge and trap systen^jn the desorb mode and preheat the trapto IBO C without a flow of desorption gas Then simultanshyeously start the flow of desorption gas at 15-mLm1n for about 4 min begin the temperature program of the gas chromatograph and start data acquisition

^ ^

1122 Cryogenic Interface shy- After the 11-min purge place the purge and trap system In the desorb mode make sure the cryogenic Interface Is a -150lt C or lower and rapidly heat the trap to 180degC while backflushing with an Inert gas at 4 mLmin for about 5 min At the end of the 5 min desorpshytion cycle rapidly heat the cryogenic trap to 250degC and simultaneously begin the temperature program of the gas chromatograph and start data acquisition

1123 Uhlle the trapped components are being Introduced Into the gas chromatograph (or cryogenic Interface) empty the purging device using the sample syringe and wash the chamber with two 25-mL flushes of reagent water After the purging device has been emptied leave syringe valve open to allow the purge gas to vent through the sample Introduction needle

304

^

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

113

^ 1 1 bull

114

115

116

GAS CHROMATOGRAPHYMASS SPECTROMETRY - Acquire and store data over the mass range 35-260 with a total cycle time (Including scan overhead tine) of 2 sec or less Cycle time must be adjusted to measure five or more spectra during the elution of each GC peak Several alternative temperature programs can be used

1131 Single ramp linear temperature program for wide bore columns 1 and 2 with a Jet separator Adjust the helium carrier gas flow rate to about 15 mLmin The column temperature is reduced lO^C and held for 5 min from the beginning of desorption then programmed to 160degC at 6degCmin and held until all components have eluted

1132 Multi-ramp linear temperature program for wide bore column 2 with the open split Interface Adjust the helium carrier gas flow rate to about 46 mLmin The column temperature is reduced lO C and held for 6 min from the beginning of desorption then heated to 70OC at IQOmin heated to 120OC at 50niin heated to 180deg at 8degmin and held at 180deg until all compounds have eluted

1133 Single ramp linear temperature program for narrow bore column 3 with a cryogenic interface Adjust the helium carrier gas flow rate to about 4 mLmin The column temperature is reduced 10degC and held for 5 min from the beginning of vaporization from the cryogenic trap programmed at 6degCmin for 10 min then IS^Cmin for 5 min to 145ltgtC and held until all components have eluted

TRAP RECONDITIONING -- After desorbing the sample for 4 min recondition the trap by returning the purge and trap system to the purge mode Uait 15 sec then close the syringe valve on the purging device to begin gas flow through the trap Maintain the trap temperature at 180degC After approximately 7 min turn off the trap heater and open the syringe valve to stop the gas flow through the trap Uhen the trap Is cool the next sample can be analyzed

TERMINATION OF DATA ACQUISITION -- When all the sample components have eluted from the GC terminate MS data acquisition Use appropriate data output software to display full range mass spectra and appropriate plots of ion abundance as a function of time If any ion abundance exceeds the system working range dilute the sample aliquot In the second syringe with reagent water and analyze the diluted aliquot

IDENTIFICATION OF ANALYTES -- Identify a sample component by comparison of Its mass spectrum (after background subtraction) to a reference spectrum In the user-created data base The GC retention time of the sample component should be within three standard deviations of the mean retention time of the compound in the calibration mixture

305

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

1161 In general all Ions that are present above 10 relative abundance In the mass spectrum of the standard should be present in the mass spectrum of the sample component and should agree within absolute 20 For example if an ion has a relative abundance of 30 In the standard spectrum its abundance in the sample spectrum should be in the range of 10 to 50 Some ions particularly the molecular ion are of special importance and should be evaluated even if they are below 10 relative abundance

1162 Identification requires expert Judgement when sample components are not resolved chromatographically and produce mass spectra containing Ions contributed by more than one analyte Uhen GC peaks obviously represent more than one sample component (Ie broadened peak with shoulder(s) or valley between two or more maxima) appropriate analyte spectra and background spectra can be selected by examining plots of characteristic Ions for tentatively identified components Uhen analytes coelute (Ie only one GC peak is apparent) the identification criteria can be met but each analyte spectrum will contain extraneous ions contributed by the coeluting compound Because purgeable^organic compounds are relatively small molecules and producV comparatively simple mass spectra this Is not a significant problem for most method analytes

1163 Structural Isomers that produce very similar mass spectra can be explicitly Identified only if they have sufficiently different GC retention times Acceptable resolution is achieved If the height of the valley between two peaks is less than 25 of the average height of the two peaks Otherwise structural Isomers are Identified as Isomeric pairs Two of the three Isomeric xylenes and two of the three dichlorobenzenes are examples of structural Isomers that may not be resolved on the capillary columns If unresolved these groups of isomers must be reported as isomeric pairs

1164 Methylene chloride and other background components appear in variable quantities In laboratory and field reagent blanks and generally cannot be accurately measured Subtraction of the concentration In the blank from the concentration in the sample Is not acceptable because the concentration of the background In the blank Is highly variable

12 CALCULATIONS

121 Complete chromatographic resolution Is not necessary for accurate and precise measurements of analyte concentrations If unique Ions with adequate Intensities are available for quantitation

306

Ci

--N

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

1211 Calculate analyte and surrogate concentrations

Ax)(Qis) 1000 Cx

(Ais) RP V where Cx - concentration of analyte or surrogate in MQL in

the water sample Ax - Integrated abundance of the quantitation ion

of the analyte in the sample As bull integrated abundance of the quantitation ion

of the internal standard in the sample Qis - total quantity (in micrograms) of internal

standard added to the water sample V gt original water sample volume In mL RF - mean response factor of analyte from the

initial calibration

1212 Alternatively use the GCMS system software or other available proven software to compute the concentrations of the analytes and surrogates from the second or third order regression curves

1213 Calculations should utilize all available digits of precisshyion but final reported concentrations should be rounded to an appropriate number of significant figures (one digit of uncertainty) Experience indicates that three significant figures may be used for concentrations above 99 M9Lgt two significant figures for concentrations between 1- 99 M9L and one significant figure for lower concentrations

1214 Calculate the total trihalomethane concentration by summing the four individual trihalomethane concentrations In M9t-shy

13 ACCURACY AND PRECISION

131 Single laboratory accuracy and precision data were obtained for the method analytes using laboratory fortified blanks with analytes at concentrations between 1 and 5 MgLgt four sets of results were obtained using the three columns specified (Sect 632) and the open split cryogenic and jet separator interfaces (Sect 633) These data are shown in Tables 4-6

132 Uith these data method detection limits were calculated using the formula (2)

MDL - S t(niialpha - 099)

where

^(n-ll-alpha bull 099) Students t value for the 99 confidence level with n-1 degrees of freedom

307

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

n - number of replicates

S - the standard deviation of the replicate analyses ^J

14 REFERENCES

1 A Alford-Stevens JU Elchelberger UL Budde Purgeable Organic Compounds in Uater by Gas ChromatographyMass Spectrometry Method 524 Environmental Monitoring and Support Laboratory US Environmental Protection Agency Cincinnati Ohio February 1983

2 Madding C Volatile Organic Compounds in Uater by Purge and Trap Capillary Column GCMS Proceedings of the Uater Quality Technology Conference American Uater Works Association Denver CO December 1984

3 Glaser JA OL Foerst GD McKee SA Quave and UL Budde Trace Analyses for Uastewaters Environ Sci Technol 15 1426 1981

4 Carcinogens-Uorking with Carcinogens Department of Health Education and Uelfare Public Health Service Center for Disease Control National Institute for Occupational Safety and Health Publication No 77-206 August 1977

5 OSHA Safety and Health Standards General Industry (29CFR1910) Occupational Safety and Health Administration OSHA 2206 (Revised January 1976)

6 Safety in Academic Chemistry Laboratories American Chemical Society Publication Committee on Chemical Safety 3rd Edition 1979

7 Arrendale RF RF Severson and OT Chortyk Open Split Interface for Capillary Gas ChromatographyMass Spectrometry Anal Chem 1984 5g 1533

8 Flesch JJ PS Fair The Analysis of Cyanogen Chloride in Drinking Water Proceedings of Uater Quality Technology Conference American Water Works Association St Louis MO November 14-16 1988

308 ^

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

c

TABLE 1 MOLECULAR UEIGHTS ANO QUANTITATION IONS FOR NETHOO ANALYTES

Comoound

Internal standard

Fluorobenzene

Surrogate

4-Bromof1uorobenzene l2-Dichlorobenzene-d4

Taraet Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichlpropropane 22-Dichloropropane 11-Dichloropropene

MW

96

174 150

78 156 128 162 250 94 134 134 134 152 112 64 118 50 126 126 206 234 186 172 146 146 146 120 98 98 96 96 96 112 112 112 110

Primary Quantitation

Ion

96

95 152

78 156 128 83 173 94 91 105 119 117 112 64 83 50 91 91 129 75 107 93 146 146 146 85 63 62 96 96 96 63 76 77 75

Secondary Quantitation

Ions

77

174176 115150

77 77158 49130 85127 175252

96 134 134 91 119

77114 66 85 52 126 126 127

155157 109188 95174 111148 111148 111148

87 6583

98 6163 6198 6198 112 78 97

11077

309

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

0 TABLE 1 (continued)

Primary Secondary Quantitation Quantitation

Comoound HW Ion Ions

cis-13-dichloropropene 110 75 110 trans-13-dichloropropene 110 75 110 Ethylbenzene 106 91 106 Hexachlorobutadiene 258 225 260 Isopropylbenzene 120 105 120 4-Isopropyltoluene 134 119 13491 Methylene chloride 84 84 8649 Naphthalene 128 128 n-Propyl benzene 120 91 120 Styrene 104 104 78 1112-Tetrachloroethane 166 131 133119 1122-Tetrachloroethane 166 83 13185 Tetrachloroethene 164 166 168129 Toluene 92 92 91 123-Trichlorobenzene 180 180 182 124-Trichlorobenzene 180 180 182 111-Trichloroethane 132 97 9961 112-Trichloroethane 132 83 9785 Trichloroethene 130 95 130132 Tri chiorof1uoromethane 136 101 103 123-Trichloropropane 146 75 77 124-Trimethylbenzene 120 105 120 135-Trimethylbenzene 120 105 120 Vinyl Chloride 62 62 64

bull

0-Xylene 106 106 91 m-Xylene 106 106 91 p-Xylene 106 106 91

^Monoisotopic molecular weight calculated from the atomic masses of the isotopes with the smallest masses

310

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

TABLE 2 CHROMATOGRAPHIC RETENTION TINES FOR NETHOD ANALYTES ON THREE COLUMNS UITH FOUR SETS OF CONDITIONS

Retention K^^ (minrsec) Compound

Internal standard

Fluorobenzene

Surrocates

4-Bromofluorobenzenel2-Dichlorobenzene-d4

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butyl benzene shysec-Butyl benzene tert-Butylbenzene Carbon Tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride Di bromochloromethane l2-Dibromo-3-Chloropropane 12-Dibromoethane Dibromomethane 12-Dichlorobenzene 13-Di chiorobenzene 14-Dichlorobenzene Dichlorodifluoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12-Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene

Column 1deg Column 2deg Column 1 ^

849 627

1838 1543 2216 1908

814 540 1857 1552 644 423

1035 829 1756 1453 201 058

2213 1929 2047 1805 2017 1734 737 516

1546 1301 205 101 624 448 138 044

1920 1625 1930 1643

1423 1151 2432 2105 1444 1150 1039 756 2231 1910 2113 1808 2133 1823

133 042 451 256 824 550 253 134 611 354 359 222

1005 740 1402 1119 601 348 749 517

1406

2338 2725

1330 2400 1222 1548

- 2246 448

2732 260B 2536 1310 2040

1236 324

2432 2446

1912

1924 1526 2726 2622 2636 308

1048 1338 750

1156 954

1512 1842 1152 1306

Column 3^

803

725 1625 538 920

1542 117

1757 1728 1719 725

1420 127 533 058

1644 1649 103

1248 1802 1336 905

1747 1728 1738 053 402 700 220 504 332 856

1229 519 710

311

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

TABLE 2 (continued)

Retention K^^ (min sec) Compound Column 1^ Column 2deg Column 2^ Column ^^ cis-13-dichloropropenetrans-13-dichloropropeneEthylbenzeneHexachlorobutadiene

1559 2659

1323 2341

1754 1642

2100 3204

1444 1914

Isopropylbenzene4-IsopropyltolueneMethylene ChlorideNaphthalenen-Propyl benzeneStyrene1112-Tetrachloroethane

1804 2112 336

2710 1904

1719 1556

1528 1831 204 2331 1625 1436 1320

2318 2630 916 3212 2420 2224 2052

1625 1738 240 1904 1649 1547 1444

1122-Tetrachloroethane 1843 1621 2404 1547 Tetrachloroethene 1344 1109 1836 1312 Toluene 1226 1000 1724 1131 123-Trichlorobenzene 2747 2411 3258 1914 124-Trichlorobenzene 2633 2305 3130 1850 111-Trichloroethane 716 450 1250 646 112-Trichloroethane 1325 1103 1818 1159 Trichloroethene 935 716 1448 901 Trichlorofluoromethane 216 111 612 146 123-Trichloropropane124-Trimethylbenzene135-TrimethylbenzeneVinyl chlorideo-Xylenem-Xylenep-Xylene

1901 2020 1928

143 1707 1610 1607

1614 1742 1654 047 1431 1341 1341

2408 3130 2450 356 2216 2122 2118

1616 1719 1659 102 1547 1518 1518

^Columns 1-3 are those given In Sect 6321 retention times were measured from the beginning of thermal desorption from the trap (columns 1-2) or from the beginning of thermal release from the cryogenic interface (column 3) degGC conditions given in Sect 1131 ^GC conditions given in Sect 1132 OGC conditions given in Sect 1133

-A

312

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

TABLE 3 ION ABUNDANCE CRITERIA FOR 4-BRaMOFLUOROBENZENE (BFB)

Mass fMzl

50 75 95 96 173 174 175 176 177

Relative Abundance Criteria

15 to 40 of mass 95 30 to 80 of mass 95 Base Peak 100 Relative Abundance 5 to 9 of mass 95 lt 2 of mass 174 gt 50 of mass 95 5 to 9 of mass 174 gt 95 but lt 101 of mass 174 5 to 9 of mass 176

313

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

TABLE 4 ACCURACY ANO PRECISION DATA FRON 16-31 OHERMINATIONS OF THE NETHOO ANALYTES IN REAfiENT HATER USING VIDE BORE CAPILUWY COLUNI 1gt

Compound

Benzene Bromobenzene Bromochloromethane Bromodi chioromethane Bromoform Bromomethane n-ButyIbenzene sec-Butyl benzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene 01bromochloromethane l2-Dibromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Dichloroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Oichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l2-Dichloropropene trans-12-Di chioropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropyltoluene Methylene chloride Naphthalene n-Propyl benzene Styrene

True Cone Range fUQL)

01-10 01-10 05-10 01-10 05-10 05-10 05-10 05-10 05-10 05-10 01-10 05-10 05-10 05-10 01-10 01-10 01-10 05-10 05-10 05-10 01-10 05-10 02-20 05-10 05-10 01-10 01-10 05-10 01-10 01-10 01-10 05-10 05-10

01-10 05-10 05-10 01-10 01-10 01-100 01-10 01-100

Mean Accuracy

( of True Value)

97 100 90 95 101 95 100 100 102 84 98 89 90 93 90 99 92 83 102 100 93 99 103 90 96 95 94 101 93 97 96 86 98

99 100 101 99 95 104 100 102

Rel Std Dev ()

57 55 64 61 63 82 76 76 73 88 59 90 61 89 62 83 70 199 39 56 62 69 64 77 53 54 67 67 56 61 60 169 89

86 68 76 67 53 82 58 72

Method Det Limit (ual)

004 003 004 008 012 011 011 013 014 021 004 010 003 013 004 006 005 026 006 024 003 012 003 010 004 006 012 012 006 004 004 035 010

006 011 015 012 003 004 004 004

314

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

TABLE 4 (Continued)

True Mean Rel Method Cone Accuracy Std Oet Range ( of True Dev Limit

Compound fuoLl Valuel IX) (uoLl

1112-Tetrachl oroethane 05-10 90 68 005 1122-Tetrachloroethane 01-10 91 63 004 Tetrachloroethene 05-10 89 68 014 Toluene 05-10 102 80 011 123-Trichlorobenzene 05-10 109 86 003 124-Trichlorobenzene 05-10 108 83 004 111-Trichloroethane 05-10 98 81 008 112-Trichloroethane 05-10 104 73 010 Trichloroethene 05-10 90 73 019 Tri chiorof1uoromethane 05-10 89 81 008 123-Trichloropropane 05-10 108 144 032 124-Trimethyl benzene 05-10 99 81 013 135-Triraethylbenzene 05-10 92 74 005 Vinyl chloride 05-10 98 67 017 o-Xylene 01-31 103 72 011 m-Xylenc 01-10 97 65 005 p-Xylene 05-10 104 77 013

k ^Data obtained by Robert U Slater using column 1 with a Jet separator-interface and a quadrupole mass spectrometer (Sect 1131) with analytes divided among three solutions

315

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

TABLE 5 ACCURACY ANO PRECISION DATA FRON SEVEN DETERMINATIONS OF THE NETHOO ANALYTES IN REAGENT UATER USING THE CRYOGENIC TRAPPING OPTION AND A NARROW BORE CAPILLARY COLUNN 3

Comoound

Benzene Bromobenzene Bromochloromethane Bromodichloromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Cyanogen chloride^ Di bromochloromethane 12-Di bromo-3-chloropropane 12-Dibromoethane Dibromomethane 12-Oichlorobenzene 13-01chiorobenzene 14-Dichlorobenzene Dichlorodi f1uoromethane 11-01chioroethane 12-Dichloroethane 11-Dichloroethene cis-12 Dichloroethene trans-12-Dichloroethene 12-Dichloropropane 13-Dichloropropane 22-Dichloropropane 11-Dichloropropene cis-l3-01chloropropene trans-13-Dlchloropropene Ethylbenzene Hexachlorobutadiene Isopropylbenzene 4-Isopropytoluene Methylene chloride Naphthalene

True Cone fuaL)

01 05 05 OI 01 01 05 05 05 01 01 01 01 01 01 01

01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

01 01 05 05 05 01

Mean Accuracy ( of True Valuel

99 97 97 100 99 99 94 90 90 92 91 100 95 99 99 96 92 99 92 97 93 97 99 93 99 98 100 95 100 98 96 99 99 98

99 100 98 87 97 98

Rel Std Dev ()

62 74 58 46 54 71 60 71 25 68 58 58 32 47 46 70 106 56 100 56 69 35 60 57 88 62 63 90 37 72 60 58 49 74

52 67 64rr 130 130 72

Method Dect Limit

(UQL)

003 011 007 003 020 006 003 012 033 008 003 002 002 005 005 005 030 007 005 002 003 005 005 004 011 003 002 005 006 003 002 004 005 002

003 004 010 026 009 004

316

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

TABLE 5 (Continued)

Comoound

n-Propylbenzene Styrene 1112-Tetrach1oroethane 1122-Tetrachloroethane Tetrachloroethene Toluene 123-Trichlorobenzene 124-Trichlorobenzene il1-Trichloroethane 112-Trichloroethane Trichloroethene Trichlorofluoromethane 123-Trichloropropane 124-Tri methyl benzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

True Cone

(ual)

01 01 01 05 01 01 01 01 01 01 01 01 01 01 01 01 01 01 01

Mean Accuracy ( of True Valuel

99 96 100 100 96 100 98 91 100 98 96 97 96 96 99 96 94 94 97

Rel Std Dev 1)

66 190 47 120 50 59 89 160 40 49 20 46 65 65 42 02 75 46 61

Method Dect Limit

lUQl)

006 006 004 020 005 008 004 020 004 003 002 007 003 004 002 004 006 003 006

^Data obtained by Caroline A Madding using column 3 with a cryogenic interface and a quadrupole mass spectrometer (Sect 1133)

Reference 8

317

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

TABLE 6 ACCURACY AND PRECISION DATA FRON SEVEN OCTERMINATIONS OF THE MrrHOO ANALYTES IN REAGENT UATER USING UIDE BORE CAPILLARY COLUNN 2

CQmPound NJLL 0

Mean Accuracy ( of True

Value 2 uoL Concl

RSD )

Mean Accuracy ( of True

Value 02 OTL Cone)

RSD )

Internal Standard

Fluorobenzene 1

Surrogates

4-Bromofluorobenzene l2-Dichlorobenzene-d4

2 3

98 97

18 32

96 95

13 17

Target Analvtes

Benzene Bromobenzene Bromochloromethane Bromod i chioromethane Bromoform Bromomethane n-Butylbenzene sec-Butylbenzene tert-Butylbenzene Carbon tetrachloride Chlorobenzene Chloroethane^ Chloroform Chloromethane 2-Chlorotoluene 4-Chlorotoluene Di bromochloromethane l2-Dibromo-3-chloropropane 12-DibromoethaneC Dibromomethane 12-Dichlorobenzene 13-Dichlorobenzene 14-Dichlorobenzene Di chiorodi f1uoromethane 11-Oichloroethane 12-Dichloroethane 11-Oichloroethene cis-12-Dichloroethene trans-12-Dichloroethene

37 38 4 5 6 7 39 40 41 8 42

9 10 43 44 11

|C

13 45 46 47 14 15 16 17 18 19

97 102 99 96 89 55 89 102 101 84 104

97 110 91 89 95

99 93 100 98 38 97 102 90 100 92

44 30 52 18 24 27 48 35 45 32 31

20 50 24 20 27

21 27 40 41 25 23 38 22 34 21

113 101 102 100 90 52 87 100 100 92 103

95 d 108 108 100

95 94 87 94 d 85 100 87 89 85

18 19 29 18 22 67 23 28 29 26 16

21

31 44 30

22 51 23 28

36 21 3 8 29 23

6

318

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

TABLE 6 (Continued)

f v

Mean Accuracy Mean Accuracy ( of True ( of True

Value RSD Value RSD CQmpQund NQ 2 uoL Concl ) 02 ugL Concl fl

12-Dichloropropane 20 102 22 103 29 13-Dichloropropane 21 92 37 93 32 22-Dichloropropane^ 11-Dichloropropene^ cis-13-Dichloropropene^ trans-13-Dichloropropene 25 96 17 99 21 Ethylbenzene 48 96 91 100 40 Hexachlorobutadiene 26 91 53 88 24 Isopropylbenzene 49 103 32 101 21 4-Isopropyltoluene 50 95 36 95 31 Methylene chloride 27 e e Naphthalene 51 93 76 78 83 n-Propylbenzene 52 102 49 97 21 Styrene 1112-Tetrachloroethane 1122-Tetrachloroethane Tetrachloroethene

53 28 29 30

95 99 101 97

44 27 46 45

104 95 84 92

31 38 36 33

Toluene 54 105 28 126 17 123-Trichlorobenzene 55 90 57 78 29 124-Trichlorobenzene 111-Trichloroethane 112-Trichloroethane Trichloroethene

56 31 32 33

92 94 107 99

52 39 34 29

83 94 109 106

59 25 28 25

Tri chiorof1uoromethane 34 81 46 48 13 123-Trichloropropane 124-Trimethylbenzene 135-Trimethylbenzene Vinyl chloride o-Xylene m-Xylene p-Xylene

35 57 58 36 59 60 61

97 93 88 104 97 f 98

39 31 24 35 18

23

91 106 97 115 98 f 103

28 22 32 14 17

14

^Data obtained by James U Elchelberger using column 2 with the open split interface and an ion trap mass spectrometer (Sect 1132) with all method analytes In the same reagent water solution

^Designation in Figures 1 and 2 ^Not measured authentic standards were not available ^Not found at 02 MgL |Not measured methylene chloride was In the laboratory reagent blank ^m-xylene coelutes with and cannot be distinguished from its isomer p-xylene No 61

319

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

OPTIONAL FOAM TRA

X IN 0 0 EXIT

10MM GUASS FRIT MEDIUM JPOROSin

EXIT IN 00

14MU 0 D

WLETXIN 00

SAMPLE INLET

2WAY SYRINGE VALVE bull17CM 20 GAUGE SYRINGE NEEDLE

MM 0 0 RUBBER S9TUH

^10MM 0 D 116 IN 00 ftmdashINLET ]1STAINLpoundSS STEEL

X IN o D

i n MOLECULAR SIEVE PURGE GASRLTEX

PURGE GAS I FLOW CONTROL

FIGURE 1 PURGING DEVICE

320

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

c

PACXIN6 PROCEDURE CONSTRUCTION

COMPRESSION GLASS g n i 7AioOT RTTING NUT WOOL I w RESISTANCE

WIRE WRAPPED AND FERRULES ACTIVATED I J ^i (DOUBLE LAYER)

THERMOCOUPLE

CHARCOAL77CM^ SOLID

I CONTROLLER SeiSOR

15CM JGRADE 1 5 - Q ^ NIC BATUREl

NTROL 7^FOOT D

[PYROMETER RESISTANCE WIRE WRAPPED TUBING 25CM TBVAX 77 Of l^^ SOLID 0105 IN ID

(SINGLE LAYER) 0125 IN 00 3X0V-1 8CM-H STAINLESS STEa ICM GLASS WOOL

TRAP INLET

FIGURE 2 TRAP PACKINGS ANO CONSTRUCTION TO INCLUDE DESORB CAPABILITY

321

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

FIGURE 3 NORMALIZED TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 na (5 MgL) OF MOST COMPOUNDS THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

109

ro ro

TOTshy

JO

m322

7 4 A

280 (42

Y

m 1082

400 1322

30

t l

1A^^^Al^ l l y^ l^

(00 2902

AJ VJ LI ^ laquolaquo Li L^l r i i raquo r f Il raquo r r I raquo raquo raquo

Mj f II II I-j II I I r j II I I II 11 II I II I IIll 1111 II raquol

m m m im 2322 2(42 391)2 3322

u o o

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822

lt ^

FIGURE A AMPLIFIED FIRST EIGHT MINUTES OF A TOTAL ION CURRENT CHROMATOGRAM FROM A VOLATILE COMPOUND CALIBRATION MIXTURE CONTAINING 25 ng (5 wgL) OF EACH COMPONENT THE COMPOUND IDENTIFICATION NUMBERS ARE GIVEN IN TABLE 6

mi

lOT-CO ro

_ i _ - I M I I I I I I I 11111 I bull I I I I I I I

50 m ISO 200 250 142 322 502 (42 822