enhanced detection in capillary electrophoresis: example determination of serum mycophenolic acid
TRANSCRIPT
Research Article
Enhanced detection in capillaryelectrophoresis: Example determination ofserum mycophenolic acid
In order to overcome the poor absorbency detection limits in CE, two simple strategies
were combined to increase the amount of the sample injected: a long capillary to hold
extra sample while leaving adequate room for the separation and acetonitrile stacking,
which concentrated the sample based on transient pseudo-ITP. The combination of these
two strategies yielded sensitivity comparable or better than that of the HPLC with good
separation and with better theoretical plate number. The analysis of mycophenolic acid, a
common immunosuppressant drug, in serum was used here as an example to illustrate
this enhanced detection and its applicability to therapeutic drug monitoring. Acetonitrile
was used to remove serum proteins followed by direct injection filling 5–21% of the
capillary volume and separation in a borate buffer. The overall CE method compared well
to an assay by HPLC as far as sample preparation, correlation coefficient, and especially
sensitivity of detection.
Keywords:
HPLC / Immunosuppressant / Mycophenolic acid / Stacking / Therapeutic druglevel DOI 10.1002/elps.200800587
1 Introduction
One of the main practical shortcomings of CE in routine
analysis is the poor absorbency detection limits, which are
about 20–60 times less sensitive compared with HPLC. One
of the simplest ways to improve this is by concentration on
the capillary-stacking. Many types of stacking methods have
been described; however, the majority of them are limited in
applicability to a few compounds [1]. Two types of stacking
are of general use. One is the high field strength [1]. This
type is simple and can be applied to most compounds;
however, in practice it yields modest, 3- to 5-fold,
concentration on the capillary. A second type of stacking is
the acetonitrile or transient pseudo-ITP [1, 2]. This type is
also a general stacking, applicable to many different
compounds, easy to accomplish; but it can yield 5- to
25-fold increase in sensitivity. It occurs in samples contain-
ing acetonitrile and salts [2].
Acetonitrile is often used in CE and HPLC as a simple
method for serum deproteinization. Interestingly, acetoni-
trile and other organic solvents, present in the sample, exert
an opposite effect in HPLC compared with that in CE [3]. In
HPLC they cause band broadening due to the formation of a
short gradient; while in CE they cause increased apparent
theoretical plate number because of the stacking effect.
Acetonitrile (in the sample) in CE serves the function of a
terminating ion by supplying the high field strength
necessary to drive or increase the velocity of the analytes
(thus acting as a pseudo-terminating ion in ITP). The
common chloride ion (or other small ions) present naturally
in the deproteinized sample migrates rapidly, acting as a
leading ion producing a low field strength region. The
analytes stack temporarily between the leading ion and the
acetonitrile, analogous to that in ITP, before they enter
the separation buffer [2]. Several workers have confirmed
this type of stacking for different compounds varying from
drugs to peptides [4–10]. However, this stacking lacks the
basic and theoretical studies necessary to maximize its full
potential. One aim of this work here is to draw attention to
the importance of this simple stacking for routine analysis
using mycophenolic analysis as an example. The stacking by
itself improves greatly the detection limits, but it falls short
compared with that of HPLC. However, it is demonstrated
here that the sensitivity of detection, using mycophenolic
acid (MPA) as an example, can be enhanced much further
by combining this stacking with a longer capillary to hold an
extra large volume of the sample, while leaving ample room
in the capillary for adequate separation.
MPA is an immunosuppressant drug derived from the
fungus Penicillium stoloniferuman. It prevents rejection in
Zak K. Shihabi
Department of Pathology, WakeForest University, Baptist MedicalCenter, Winston-Salem, NC, USA
Received September 10, 2008Revised November 18, 2008Accepted December 20, 2008
Abbreviation: MPA, mycophenolic acid
Correspondence: Dr. Zak K. Shihabi, Department of Pathology,Wake Forest University School of Medicine, Winston-Salem, NC27157, USAE-mail: [email protected]: +1- 336 -716-9944
& 2009 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
Electrophoresis 2009, 30, 1516–15211516
organ transplantation. It is given as mycophenolate mofetil
(Cell Cept) or as mycophenolate sodium (Myfortic). Both are
metabolized in the liver to the active moiety MPA. MPA
undergoes glucuronidation as the first step in its metabo-
lism. The glucuronide derivative is not pharmacologically
active. However, there are other minor metabolites that may
be active. MPA acts as an uncompetitive, selective and
reversible inhibitor of inosine monophosphate dehy-
drogenase type II expressed on activated lymphocytes, the
enzyme that controls the rate of synthesis of guanine
monophosphate in the de novo pathway of purine synthesis
in the proliferation of B and T lymphocytes. MPA is exten-
sively and tightly bound to human serum albumin, with the
average bound fraction in individuals with normal renal
function being 97.5%. Its clearance is highly dependent on
protein binding (restrictive clearance) with an apparent half-
life of about 18 h.
MPA is administered mostly in combination with other
immunosuppressive drugs such as cyclosporine , tacrolimus
and prednisolone. This drug is also used as a steroid sparing
treatment in immune-mediated disorders including immu-
noglobulin A nephropathy, lupus nephritis and small vessel
vasculitides. The side effects of MPA therapy include diar-
rhea, nausea, vomiting, infections, leukopenia and/or
anemia . Many physicians simply decrease a prescribed dose
of this drug until the gastrointestinal symptoms disappear.
Thus the routine monitoring of this drug remains under
debate [11]. On the other hand, because of the lack of a
relationship between the dose and its serum levels and a
lack of an association of MPA levels with adverse symptoms,
therapeutic monitoring of MPA is recommended by many
investigators. Under such conditions the risk for acute
rejection is minimized and the patient outcome after renal
transplantation is improved [12].
The therapeutic level in most studies when MPA is used
in combination with other immunosuppressive drugs is
1–3.5 mg/L (3.2–11.4 mmol/L) for liver and kidney trans-
plants for trough concentrations and 30–60 mg/h/L for area
under the curve. At levels above the therapeutic range, the
relative risk of developing infection or leucopenia increases
more than threefold [13, 14].
Several methods have been described to measure total
and free MPA concentration such as HPLC with ultraviolet
detection, as well as mass spectrometric detection [15–17].
An enzyme-multiplied immunoassay technique that can be
automated on the general chemistry analyzers has also been
described and is used in some centers for total MPA
concentration [18].
Because the therapeutic levels are relatively low, the CE
has not been well investigated for the assay of MPA. Unsalan
et al. [19] have used CZE after sample evaporation to improve
the sensitivity, while Carlucci et al. [20] have used MEKC for
MPA determination after protein removal with acetonitrile or
acids. To improve the detection level of this drug in serum,
stacking on the capillary is desirable as a simple method.
Unfortunately stacking in MEKC is difficult for routine
analysis [21]. Furthermore, the addition of organic solvents,
such as acetonitrile in the sample to remove the excess of
proteins, limits the amount of sample in MEKC to be injected
on the capillary to a few seconds or less than 0.5% of the
capillary volume [21, 22]. On the other hand, the presence of
acetonitrile in the sample in CZE allows much larger sample
to be injected leading to stacking. This also eliminates the
tedious steps of evaporation and reconstitution to concentrate
the drug in the sample. Here we describe a simple and
enhanced assay for MPA in serum based on simultaneous
stacking by acetonitrile and the use of a long capillary with
separation by CZE. It illustrates the stacking property and
simplicity of acetonitrile for analysis of small molecules by CE.
2 Materials and methods
2.1 Chemicals
MPA, hippuric acid and boric acid were obtained from
Sigma Chemicals (St. Louis, MO, USA).
2.2 Instrument
Quanta 4000 (Waters, Milford, MA, USA) equipped with
untreated capillary 50 cm � 75 mm (id) (or as specified).
The voltage was set 14.5 kV and the wavelength was at
214 nm. Samples were injected hydrodynamically (10 cm
height gravity) for 40 s (or as specified) and electrophoresed
for 12 min.
2.3 Electrophoresis buffer
Two grams of boric acid was dissolved in 100 mL of water,
pH 7.5. Aliquots containing 10 mL of methanol and 100 mg
of SDS were added to the buffer.
2.4 ZnSO4 (for HPLC analysis)
One volume of ZnSO4 (0.1 M) was mixed with one volume
of acetonitrile (stable at room temperature).
2.5 Standard
MPA, 3 mg/L of 1% sodium chloride.
2.6 Procedure of CE
An aliquot of 100 mL of serum (or standard)was vortex-mixed
with 200 mL of acetonitrile containing 2 mg/L of hippuric
acid (as an internal standard) in 500 mL micro-centrifuge
tube and centrifuged for 15 s at 13 000g. The supernatant
was injected on the capillary.
Electrophoresis 2009, 30, 1516–1521 CE and CEC 1517
& 2009 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
2.7 HPLC
For comparison, MPA was analyzed by HPLC too on a CN
Zorbax column 150� 4.6 mm packed with 5 mm particles
(Agilent Technologies, Wilmington, DE, USA), which was
eluted with a solvent of 25% acetonitrile in 0.1% phosphoric
acid. Serum samples were vortex-mixed with two volumes of
ZnSO4 solution and centrifuged for 40 s at 13 000g and
20 mL of the supernatant was injected on the column, with
detection at 254 nm (Model 441, WatersAssociates). This
method is similar to that by Svensson et al. [17] and Bolon
et al. [15] except that a CN column in place of C18 was used
to reduce the need for acetonitrile in the elution solvent.
This method was linear between 0.5 and 8 mg/L (r 5 0.99)
with minimum detection (3� baseline noise) for standards
of 0.125 mg/L. The average recovery from serum at 2 mg/L
was 88 %, and the RSD for nine replicates at 2.0 mg/L was
3.4%.
3 Results and discussion
Figure 1 illustrates electropherograms of an MPA
standard prepared in 1% sodium chloride (2.7 mg/L), a
serum of a patient free from this drug and the same serum
spiked with an added standard of 2.7 mg/L. Spiking
with a value within the therapeutic range gives a peak
height with good sensitivity (Fig. 1). The absorbency at
214 nm was about 2.3 times higher than that at 254 nm.
The test was linear between 0.5 and 6.0 mg/L with
r 5 0.99. The RSD based on nine replicates was 1.9%. The
recovery of MPA 2.7 mg/L in serum was 93% relative to a
standard prepared in 1% NaCl (n 5 6). Because of the
stacking, standards should be prepared in 1% saline or in
serum free from this drug rather than in water. An increase
in the buffer pH increased the migration time of the
MPA by CZE without affecting appreciably the peak height.
The SDS was added, below its critical micellar concentra-
tion, to speed up the migration. Checking 12 patients
free from this drug as well as several pools of serum, there
were no appreciable interfering peaks with the same
migration of MPA. Some peaks or noise close to the MPA
peak, at less than 0.3 mg/L, can be detected in some
patients. Thus, values below 0.5 mg/L are reported as less
than 0.5 mg/L.
The analysis of MPA by the CE was compared
with that by HPLC. The correlation between the CZE and
the HPLC for 19 patients is good with r 5 0.96 (Fig. 2). The
analytical aspects of both methods are comparable
with those of sample handling and analysis time in this
study and in a previous one [20]. Figure 3 illustrates the
analysis of a standard and the same patient by both
CE and HPLC. Some patients have a peak close to the MPA
peak. Such patients demonstrated a similar peak in
HPLC (Fig. 3D). Whether this peak represents a
metabolite of MPA is not clear. It is known that MPA
undergoes metabolism into several metabolites. In order to
separate the compound close to the MPA in the CE
method, addition of 5% of methanol to the electrophoresis
buffer gives a slight separation, while the addition
of 10% methanol was necessary to give separation close to
baseline (Fig. 3). Low values were more difficult to analyze
by HPLC due to interferences in the serum and loss line-
arity by peak height in the isocratic separation. Unsalan et al.[19] have used CZE for MPA analysis but after
solvent concentration and evaporation. Here in this method
we avoided all these extra steps with much improved
sensitivity.
Figure 1. Electropherogram of (A) MPA standard 2.7 mg/L in 1%NaCl, (B) patient sample free from MPA, (C) the same serumafter addition of 2.7 mg/L MPA (M 5 mycophenolic acid;H 5 hippuric acid as an internal standard); 60 s injection.
Electrophoresis 2009, 30, 1516–15211518 Z. K. Shihabi
& 2009 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
Contrary to the MEKC, acetonitrile in the sample
improves the peak height and also improves the separation
in the CZE through stacking [1,2]. Figure 4 shows the peak
height of a standard with 4, 40 and 80 s injection (filling 0.5,
4.8 and 9.7% of the capillary volume, respectively).
Compared with that with 4 s injection (close to that used in
the MEKC), about 10- and 20-folds increase in sensitivity,
respectively, can be observed due to the stacking (Fig. 4).
The capillary at 80 s is filled with 9.7% of its volume with
sample. Injecting large volumes of sample decreases the
effective length of the capillary resulting in less-efficient
separation for serum.
In order to inject a larger amount of sample on the
capillary, with adequate separation, a longer capillary
75 mm� 80 cm (1.7� the effective length of the previous
one) was tried (Fig. 5). Here, a serum sample spiked with
just 0.3 g/L of the MPA standard (Fig. 5B) can be
separated and detected, after filling the capillary to 8% of its
y = 0.9557xR20.9699 = Sy.x=0.31
0
1
2
3
4
5
6
7
0 1 2 3 4 5 6
CE mg/L
HP
LC
mg
/L
Figure 2. Correlation of the CZE with the HPLC for MPA.
Figure 3. Comparisons of the CZE and HPLC for a standard andpatient on MPA: (A) CE of the standard 2.5 mg/L, (B) a patient,1.7 mg/L MPA, (C) standard by HPLC and (D) the same patient inB by the HPLC (M 5 mycophenolic acid; H 5 hippuric acid as aninternal standard). Figure 4. Stacking of 2 mg/L standard of MPA: (A) 4 s injection
(0.5% of the capillary volume), (B) 40 s injection (4.8 % of thecapillary) and (C) 80 s injection (9.7% of the capillary) (M 5 myco-phenolic acid; H 5 hippuric acid as an internal standard).
Electrophoresis 2009, 30, 1516–1521 CE and CEC 1519
& 2009 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
volume, much better than on the previous capillary. In pure
standards, this capillary can be filled further to 21% of its
volume with sample. Thus, values below the therapeutic
range can be detected (Fig. 6). The limits of detection
depend on many factors such as the electronics, sample
loading, solvents, etc.; however, here the CE can attain the
same, if not better, sensitivity, compared with that of HPLC
(Fig. 6) (minimum detection level is 95 mg/L for CE versus125 mg/L for HPLC). In this case, the length of the injected
sample segment on the capillary was initially about 200 mm.
Remarkably, at the end of the electrophoretic step it
concentrated down to a few millimeters at the detector. The
concentration factor on this long capillary at 21% of the
capillary volume compared with the shorter one with 0.5%
injection represents about 70-fold. Also, in this case, the
apparent theoretical plate number for MPA peak is much
better by the CE compared with that by the HPLC, 53 265
versus 4620, respectively (Fig. 6).
4 Concluding remarks
Neither a long capillary by itself nor stacking by itself is
sufficient to bring along the detection limits of the CE to
that of HPLC. However, the combination of these two
simple strategies did, through an increase in the injected
sample volume. Traditionally, longer capillaries are used
mainly to improve the resolution in CE. However, they also
give increased theoretical plate number in addition to
allowing loading more sample volume. Here, a longer
capillary contributed better to the stacking process by
holding larger sample volume, while leaving enough room
for separation of the complex serum samples. An increase
Figure 5. MPA separation from serum on a long capillary(75 mm� 80 cm) with injection filling 8% of the capillary volume,(16 kV): (A) serum sample free from MPA (M), (B) the samesample but spiked with 300 mg/L of MPA and (C) standard of 700mmg/L MPA (M).
Figure 6. Comparison of the CE (top) and the HPLC (bottom) fora standard of 400 mg/L MPA (M) on a long capillary(75 mm� 80 cm). The standard in the CE was injected for 500 sfilling 21% of the capillary volume (16 kV).
Electrophoresis 2009, 30, 1516–15211520 Z. K. Shihabi
& 2009 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
in the injected sample volume alone, under non-stacking
conditions, causes a rapid deterioration of the theoretical
plate number and the resolution too [23]. Acetonitrile in the
sample overcomes that, while inducing �20-fold concentra-
tion based on the described mechanism of transient pseudo-
ITP [2]. The wide capillary (75 mm id) chosen here,
compared with the more common 50 mm, also doubled the
signal. The combination of these two (or three) strategies
here offered suddenly a greater sensitivity, which matches
that of the HPLC (Fig. 6) (the minimum detection level is
95 mg/L for CE versus 125 mg/L for HPLC), also with better
theoretical plate number. The combination of acetonitrile
stacking with other techniques such as bubble cells
or Z cells probably will improve further the detection limits
too.
Acetonitrile stacking [13, 14], unfortunately,
has not been investigated very much. Many aspects of this
technique remain unknown and deserve further studies. In
order to fully optimize this technique, uncoupling
the stacking step from the separation step is needed.
Imaging techniques in this respect will be helpful. The
mathematical modeling of this stacking has not been
approached either.
After this paper has been submitted, another study [24]
showed that MPA and its glucauronide can be determined
by CE based on the acetonitrile stacking described here.
However, the authors utilized a short capillary with small
diameter (25–35 cm� 50 mm). As a result of this, the
stacking was modest but adequate.
Injecting large sample volumes also has its own short-
comings in terms of increasing the analysis time. The
present CE instrumentations are not designed to qinject
large volumes of samples rapidly in few seconds. The
injection time here, in some cases, increased too long to
500 s, which is not suitable in routine analysis. The capillary
wash and filling steps also require longer time. In most
instances, as for the MPA analysis, acetonitrile stacking in a
short capillary offers enough sensitivity for the needed
therapeutic range and offers the desired high speed. On the
other hand, some analytes are present in very low concen-
trations so that a longer capillary becomes necessary.
Hopefully, the next generation of CE instruments can be
designed to take advantage of the stacking methods needs.
In summary, the combination of a long capillary
and acetonitrile injection directly yielded an enhanced
sensitivity, which matches that of the HPLC, and with
better theoretical plate numbers; while eliminating
the use of expensive columns and avoiding the consump-
tion of high concentrations of organic solvents needed in
HPLC.
The author has declared no conflict of interest.
5 References
[1] Shihabi,Z., J. Chromatogr. A 2000, 902,107–117.
[2] Shihabi, Z., Electrophoresis 2002, 23, 1612–1617.
[3] Shihabi, Z., J. Liq. Chromatogr. Relat. Technol. 2008, 31,3159–3168.
[4] Choy, T. M., Chan, W. H., Lee, A. W., Huie, C. W., Elec-trophoresis 2003, 24, 3116–3123.
[5] Kong, Y., J. Liq. Chromatogr. Relat. Technol. 2006, 29,1561–1573.
[6] Kong, Y., Zheng, N., Zhang, Z., Gao, R., J. Chromatogr.B 2003, 795, 9–15.
[7] Sarah, Y. C., Nan-Yan, Z., Chee-Shan, C., Int. J. Appl.Sci. Eng. 2004, 2, 277–285.
[8] Kubalczyk, P., Bald, E., Anal. Bioanal. Chem. 2006, 384,1181–1185.
[9] Galeano-Diaz, T., Acedo-Valenzuela, M. I., Mora-Diez, N.,Silva- Rodriguez, A., Electrophoresis 2005, 26, 3518–3527.
[10] Shihabi, Z., J. Chromatogr. A. 1996, 744, 231–240.
[11] West-Thielke, P., Kaplan, B., Am. J. Transplant. 2007, 7,2441–2442.
[12] Meur, Y. L., Buchler, M., Thierry, A., Caillard, S., Ville-main, F., Lavaud, S., Etienne, I. et al., Am. J. Transplant.2007, 7, 2496–2503.
[13] Tredger, J. M., Brown, N. W., Adams, J., Gonde, C. E.,Dhawan, A., Rela, M., Heaton, N., Liver Transplantation2004, 10, 492–502.
[14] Van Gelder, T., Le Meur, Y., Shaw, L. M., Oellerich, M.,DeNofrio, D., Holt, C., Kaplan, B. et al., Ther. Drug Monit.2006, 28, 145–154.
[15] Bolon, M., Jeanpierre, L., El Barkil, M., Chelbi, K.,Sauviat, M., Boulieu, R., J. Pharm. Biomed. Analysis2004, 36, 649–651.
[16] Annesley, T. M., Clayton, L. T., Clin. Chem. 2005, 51,872–877.
[17] Svensson, J. O., Brattstrom, C., Sawe, J., Ther. DrugMonit. 1999, 21, 322–324.
[18] Shipkova, M., Schutz, E., Armstrong, V. W., Niedmann,P. D., Wieland, E., Oellerich, M., Transplant. Proc. 1999,31, 1135–1137.
[19] Unsalan, S., Hempel, G., Fobker, M., Wurthwein, G.,Boos, J., Chromatographia 2006, 64, 419–421.
[20] Carlucci, F., Anzini, M., Rovini, M., Cattaneo, D., Merlini,S., Tabucchi, A., Electrophoresis 2007, 28, 3908–3914.
[21] Shihabi, Z., in: Cazes, J. (Ed.), Encyclopedia ofChromatography, Taylor & Francis, Boca Raton 2005,pp. 1490–1498.
[22] Shihabi, Z., in: Landers, J. P. (Ed.), Handbook of Elec-trophoresis, 2nd ed. CRC Press, Boca Raton, FL 1997,pp. 457–477.
[23] Vinther, A., Soeberg, H., J. Chromatogr. 1991, 559, 3–26.
[24] Ohyama, K., Kinoshita, N., Kishikawa, N., Kuroda, N.,Electrophoresis 2008, 29, 3658–3664.
Electrophoresis 2009, 30, 1516–1521 CE and CEC 1521
& 2009 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com