electronic supplement material10.1007/s00134... · web view- a face mask specifically...
TRANSCRIPT
ELECTRONIC SUPPLEMENT MATERIAL
1. Face masks evaluated
Each of the three face masks evaluated has its own characteristics and is representative
of the current market offers:
- A face mask specifically commercialized as designed to deliver a Helium - Oxygen
mixture: Heliox21® face mask with a non-rebreathing valve, two flap side valves, a
reservoir bag and a Y-piece to possibly connect a nebulizer (Heliox21® nebulizer kit
adult n°1420; Intersurgical, UK).
- A face mask designed to achieve high inspired O2 fractions (FIO2) with low delivered
flow rates: Pulmanex® Hi-OX80 face mask with a non-rebreathing valve, no flap side
valves and a reservoir bag (Ref. 50-8000, Viasys, Netherlands). The Hi-OX80 is
supported by two elastic straps and has a manifold with three one-way valves: one
valve on the inhalation tube, a second valve on the exhalation tube, a third valve which
interconnects inhalation and exhalation tubes. A fourth valve located just beside the
inlet of the oxygen supply hose acts as an anti suffocation valve.
- An adult standard high concentration non-rebreathing face mask for oxygen delivery
(Ref. 106, Unomedical, Italy): this flexible face mask is supported by one elastic strap,
with a non-rebreathing valve, two flap side valves and a reservoir bag. This model or a
similar is the mask the most commonly used for the delivery of high FIO2.
2. Helium-Oxygen mixture delivered
The study product is the inhaled medicinal gas mixture He-O2 , which contains:
- Oxygen 22% ± 1% (Volume)
- Helium 78% ± 1% (Volume).
This gas mixture has been produced and released by Air Liquide Deutschland GmbH
according to standard procedures.
He-O2 78:22 has been supplied by the sponsor in B10 gas cylinders pressured at at 200
bars with a content of ± 2 m3 of gas at 1 bar and 15°C.
A dynabloc (including a pressure reducer and a flow meter) was fitted on the cylinders
to insure a fixed regulation and an easy and adjustable flow on the whole range of the
dial.
3. Devices associated to the measurements
He analyser (Prototype by Viasys, Netherlands)
The method of analysis is based on thermal conductivity.
The specifications of the device were as follows:
- Side stream sampling 200-300 mL/min,
- F[He] measurements: linear between 20 and 80% Helium; response time from
0 to 80 % Helium<800 msec,
- Cross sensitivity of He for O2: +1.5 % He from 21 to 100%
Calibration procedure was performed as follows before the measurements:
- Switch on system and let it stabilise for 15 minutes,
- Switch on computer and start the StampDAQ® program (based on Excel),
- He calibration: let the system acquire and sample room air for zero calibration
(no He in room air); Start recording of data for 10 seconds; Let the system sample via
side port of 2 L/min gas flow from the He/O2 78:22 cylinder; Start recording for 10
seconds; In Excel®, the readings (counts) are then recalculated to calibrate He values.
- Calculation of the calibration factor: sample room air in the device to obtain
Signal 0 (S0) in AD Counts which correspond to 0 % He; sample He/O2 78:22 to obtain
Signal 1 (S1) in AD Counts which correspond to 78% He. As measurements are linear,
the He percentages in between can be extrapolated.
After the measurements were performed, the stability of system was checked as follows:
The gas calibration procedure was repeated to check stability of system. The device was
considered stable if differences between calibration values were below 5%.
Nasopharyngeal sampling catheter
The sampling catheter (Tracheal catheter P/N 10635, 5FR, Viasys, Netherlands) is a
small-bore flexible perforated catheter used originally to measure the pressure in the
endotracheal tube during mechanical ventilation. The small diameter of the catheter and
its flexibility will allow inserting it into the face mask through the lateral side of the
mask. The catheter has been fixed to the nose and cheek by a medical adhesive tape to
avoid its displacement. The catheter has been connected to the He analyser via a heated
capillary line of about 1 meter long (Tracheal catheter adaptor Nº50000-40034, 5FR,
Viasys, Netherlands).
Plethysmograph (Respitrace)
Respiratory rate (RR) and tidal volume (VT), which allowed to calculate minute
ventilation (MV=RR x VT) have been measured and recorded continuously by the
respiratory inductive plethysmograph (Respitrace by Viasys, Netherlands). It is
composed of two bands (one to be placed around the ribcage and the other one around
the abdomen) and a pneumotachometer connected to a differential pressure transducer.
The plethysmograph has been calibrated using the qualitative diagnostic calibration
(QDC) method [1] [2].
SenTec Digital Monitor®
The SenTec Digital Monitor (SenTec, Therwil, Switzerland), placed at the ear lobe, has
been used to monitor continuously, simultaneously and non-invasively the
transcutaneous PCO2 (tcPaCO2) and SpO2 [3].
tcPaCO2 is measured with a Severinghaus type pH-sensitive electrode submerged in
bicarbonate containing electrolytic solution enclosed by a permeable membrane. The
sensor warms up skin temperature to +42ºC to produce capillary hyperaemia. Under this
condition, the membrane allows capillary and skin carbon dioxide diffusion into the
chamber, resulting in a modification of pH proportional to transcutaneous PCO2. The
monitor calculates an estimate of +37ºC corrected arterial PCO2. The sensor is
calibrated after reaching the equilibrium with an external 8% CO2 gas mixture. This
process is automated and takes 5 minutes. In vivo calibration (adjustment to measured
PaCO2) is not available in this device. The system was recalibrated every 8 hr of use.
SpO2 was measured by pulse oximetry using spectrophotometry.
4. Data Collection and Processing
The signal from the He analyser were acquired directly by Acknowledge® program
(Biopac systems, US). The Fp (He) was estimated before the sharp increase in He%,
which was considered as the start of inspiration.
5. Additional results
All statistical analyses were performed using the SAS software version 8.02 (SAS®
Institute, North Caroline, US).
Summary of the average end-expiratory percentage of contamination with room air (% Air Cont). Mean (SD).
Flow rate Ventilatory pattern Heliox 21 Hi-Ox 80 Standard mask
7 L.min-1
Resting 63 (7) 44 (13) 64 (9)Hyperventilation 78 (5) 56 (14) 76 (8)
10 L.min-1
Resting 52 (7) 40 (15) 54 (9)Hyperventilation 69 (8) 52 (16) 66 (6)
12 L.min-1
Resting 44 (12) 32 (9) 46 (10)Hyperventilation 65 (9) 49 (20) 66 (10)
15 L.min-1
Resting 32 (17) 24 (8) 38 (11)Hyperventilation 54 (16) 37 (18) 52 (10)
Summary of average Fp(He)% at end-expiration. Median (Range).
Flow rate Ventilatory pattern Heliox 21 Hi-Ox 80 Standard mask
7 L.min-1
Resting 27 (23-37) 48 (27-52) 29 (19-38)Hyperventilation 17 (11-23) 40 (16-43) 20 (9-25)
10 L.min-1
Resting 37 (30-46) 44 (32-61) 36 (25-45)Hyperventilation 24 (15-34) 38 (20-55) 27 (19-33)
12 L.min-1
Resting 46 (28-52) 49 (48-64) 44 (30-50)Hyperventilation 26 (19-40) 38 (22-65) 29 (14-34)
15 L.min-1
Resting 57 (33-66) 59 (54-69) 47 (38-59)Hyperventilation 33 (24-60) 46 (28-70) 37 (27-48)
Summary of average Fp(He)% at end-expiration. Mean (SD).
Flow rate Ventilatory pattern Heliox 21 Hi-Ox 80 Standard mask
7 L.min-1
Resting 29 (6) 44 (10) 28 (4)Hyperventilation 17 (5) 35 (11) 19 (6)
10 L.min-1
Resting 38 (6) 47 (12) 36 (7)Hyperventilation 24 (6) 38 (12) 26 (5)
12 L.min-1
Resting 44 (9) 53 (7) 42 (8)Hyperventilation 27 (7) 40 (15) 27 (8)
15 L.min-1
Resting 53 (13) 59 (6) 49 (9)Hyperventilation 36 (13) 49 (14) 37 (7)
Figure. Median and range of the average Fp(He)% at end-expiration.
Fp(H
e) %
End
-ex
pira
tion
Flow rate (L.min-1)
7
0.2
010 12 15 7 10 12 15
0.4
0.6
0.2
0.4
0.6
Resting pattern Hyperventilation pattern
Heliox 21
HiOx 80
Standard mask
0
10
20
30
40
50
60
70
80
6 7 8 9 10 11 12 13 14 15 160
10
20
30
40
50
60
70
80
6 7 8 9 10 11 12 13 14 15 16
7
80
10 12 15 7 10 12 15
60
40
20
0
80
60
40
20
0Fp(H
e) %
End
-ex
pira
tion
Flow rate (L.min-1)
7
0.2
010 12 15 7 10 12 15
0.4
0.6
0.2
0.4
0.6
Resting pattern Hyperventilation pattern
Heliox 21
HiOx 80
Standard mask
0
10
20
30
40
50
60
70
80
6 7 8 9 10 11 12 13 14 15 160
10
20
30
40
50
60
70
80
6 7 8 9 10 11 12 13 14 15 16
7
80
10 12 15 7 10 12 15
60
40
20
0
80
60
40
20
0
Summary of minute ventilation (L.min-1). Median (Range).
Flow rate Ventilatory pattern Heliox 21 Hi-Ox 80 Standard mask
7 L.min-1
Resting 13 (6-26) 14 (6-24) 16 (7-20)Hyperventilation 24 (19-38) 26 (14-43) 21 (18-32)
10 L.min-1
Resting 16 (6-19) 17 (5-22) 15 (4-20)Hyperventilation 23 (17-38) 32 (14-43) 22 (20-37)
12 L.min-1
Resting 16 (6-19) 13 (6-22) 15 (5-28)Hyperventilation 23 (19-43) 31 (14-39) 26 (17-40)
15 L.min-1
Resting 17 (5-22) 16 (6-26) 15 (5-24)Hyperventilation 24 (20-42) 30 (14-39) 24 (19-34)
REFERENCES ELECTRONIC SUPPLEMENT MATERIAL
1. De Groote A, Paiva M, Verbandt Y, (2001) Mathematical assessment of qualitative diagnostic calibration for respiratory inductive plethysmography. J Appl Physiol 90: 1025-1030
2. Leino K, Nunes S, Valta P, Takala J, (2001) Validation of a new respiratory inductive plethysmograph. Acta Anaesthesiol Scand 45: 104-111