Efficient Fourier transformation of unstructured meshes and application to MRI simulation by Alejandro Martinez A thesis submitted in conformity with the requirements for the degree of Master of Mechanical Engineering Graduate Department of Mechanical and Industrial Engineering University of Toronto Copyright c 2014 by Alejandro Martinez

Upload: others

Post on 13-Feb-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

  • Efficient Fourier transformation of unstructured meshes

    and application to MRI simulation

    by

    Alejandro Martinez

    A thesis submitted in conformity with the requirementsfor the degree of Master of Mechanical Engineering

    Graduate Department of Mechanical and Industrial Engineering

    University of Toronto

    Copyright c© 2014 by Alejandro Martinez

  • Abstract

    Efficient Fourier transformation of unstructured meshes and application to MRI

    simulation

    Alejandro Martinez

    Master of Mechanical Engineering

    Graduate Department of Mechanical and Industrial Engineering

    University of Toronto

    2014

    This thesis demonstrates different ways to compute the MRI signal in finite element

    magnetic resonance imaging or FE-MRI. For linear elements the MRI signal can be found

    analytically. The error metric for linear elements to calculate the signal with a fixed error

    in k-space using an adaptive Gaussian quadrature is derived using finite element function

    derivatives and is then extended for use in quadratic elements. The feasibility of using

    the numerical steepest descent (NSD) method, which is designed to compute oscillatory

    integrals like the MRI signal equation, was also studied. The method was found to

    compute the MRI signal faster than adaptive Gauss quadrature for quadratic surface

    elements using typical MRI imaging parameters. The feasibility of NSD is discussed for

    volumetric quadratic elements and for different mesh geometries.

    ii

  • Dedication

    For my love, Helena Violet Kelly, with gratitude for your love and support

    iii

  • Acknowledgements

    I would like to thank prof. Steinman. and prof. Huybrechs and the other members of

    the examination committee prof. Nachman and prof. Mandelis for their support and

    discussions. I would also like to thank my family for all their love and support.

    iv

  • Contents

    1 Introduction 1

    1.1 Motivation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

    1.2 Introduction to MRI theory and simulation . . . . . . . . . . . . . . . . . 1

    1.2.1 Signal generation . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

    1.2.2 Slice selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    1.2.3 Spatial encoding . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

    1.2.4 Signal acquisition and reconstruction . . . . . . . . . . . . . . . . 10

    1.2.5 Image resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    1.2.6 Image artefacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

    1.3 Literature review . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

    1.3.1 Analytic solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

    1.3.2 Simulators using magnetization maps . . . . . . . . . . . . . . . . 20

    1.3.3 Simulators using T1, T2 and proton density maps . . . . . . . . . 20

    1.3.4 Finite element based simulators . . . . . . . . . . . . . . . . . . . 21

    1.3.5 Bloch equation simulators . . . . . . . . . . . . . . . . . . . . . . 23

    1.4 MRI acquisition within the FE-MRI library . . . . . . . . . . . . . . . . 23

    1.5 Research objectives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

    2 Adaptive Gauss quadrature for FE-MRI elements 26

    2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

    v

  • 2.2 Gauss quadrature to compute oscillatory integrals . . . . . . . . . . . . . 29

    2.3 Computing oscillatory integrals in higher dimensions . . . . . . . . . . . 30

    2.4 Error estimation for tri3 elements . . . . . . . . . . . . . . . . . . . . . . 31

    2.5 Error estimation for tri6 elements . . . . . . . . . . . . . . . . . . . . . . 36

    2.6 Error estimation for tet4 elements . . . . . . . . . . . . . . . . . . . . . . 38

    2.7 Error estimation for tet10 elements . . . . . . . . . . . . . . . . . . . . . 40

    3 Algorithm details and implementation 41

    3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

    3.2 FE-MRI software architecture . . . . . . . . . . . . . . . . . . . . . . . . 42

    3.3 K-space acquisition algorithm . . . . . . . . . . . . . . . . . . . . . . . . 43

    3.4 Cycles per element quadrature table . . . . . . . . . . . . . . . . . . . . . 46

    3.5 Calculation of quadrature rules . . . . . . . . . . . . . . . . . . . . . . . 48

    3.6 Slice profile simulation for volumetric meshes . . . . . . . . . . . . . . . . 48

    3.7 Additions and improvements to current FE-MRI functions . . . . . . . . 49

    4 Results and discussion 52

    4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

    4.2 Validation of optimal algorithm for tri3 elements . . . . . . . . . . . . . . 52

    4.3 Validation of optimal algorithm for tri6 elements . . . . . . . . . . . . . . 56

    4.4 Validation of optimal algorithm for tet4 elements . . . . . . . . . . . . . 56

    4.5 Validation of optimal algorithm for tet10 elements . . . . . . . . . . . . . 58

    4.6 Computation time comparison between linear and quadratic surface dis-

    cretizations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

    4.7 Simulation of MRI acquisition of a carotid artery . . . . . . . . . . . . . 61

    5 Feasibility of NSD for faster signal calculation 62

    5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

    5.1.1 Approximate evaluation of oscillatory integrals . . . . . . . . . . . 62

    vi

  • 5.1.2 Adaptive quadrature . . . . . . . . . . . . . . . . . . . . . . . . . 63

    5.1.3 Filon method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

    5.1.4 Levin method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

    5.1.5 Numerical steepest descent . . . . . . . . . . . . . . . . . . . . . . 65

    5.2 Experiment using NSD for quadratic surface elements . . . . . . . . . . . 66

    5.2.1 NSD decomposition for quadratic surface elements . . . . . . . . . 68

    5.2.2 Algorithm details . . . . . . . . . . . . . . . . . . . . . . . . . . . 71

    5.3 Results and discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

    6 Conclusions 75

    6.1 Summary of work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

    6.2 Future directions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75

    A Generation of Gauss quadratures 77

    A.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

    A.2 The Golub-Welsch algorithm . . . . . . . . . . . . . . . . . . . . . . . . . 82

    A.3 Sample C++ quadrature code . . . . . . . . . . . . . . . . . . . . . . . . 83

    B Mesh generation with same discretization error 88

    B.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88

    B.2 Discretization error when using quadratic elements . . . . . . . . . . . . 89

    B.3 Discretization error when using linear elements . . . . . . . . . . . . . . . 91

    B.4 MATLAB implementation . . . . . . . . . . . . . . . . . . . . . . . . . . 92

    C Implementation of NSD algorithm 97

    C.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97

    C.2 G1 decomposition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99

    C.3 Steepest descent paths for G . . . . . . . . . . . . . . . . . . . . . . . . . 105

    C.4 Steepest descent paths for F . . . . . . . . . . . . . . . . . . . . . . . . . 106

    vii

  • Bibliography 108

    viii

  • List of Tables

    1.1 Values used for fig. 1.1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

    1.2 Finite element types used in this thesis . . . . . . . . . . . . . . . . . . . 25

    4.1 K-space error error maps and histograms for all the element types imple-

    mented. Notice that the x-axis for the histograms in in terms of n, where

    error = 10n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

    4.2 Test meshes for each element type and their simulated MRI images. For

    the meshes with quadratic elements one of the elements is selected to show

    the blue edges shown are between element nodes and do not delineate a

    quadratic element . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

    5.1 Total computation times in seconds and time ratios for different k-space

    extents for an error threshold of 10−2 . . . . . . . . . . . . . . . . . . . . 72

    5.2 Total computation times in seconds and time ratios for different k-space

    extents for an error threshold of 10−3 . . . . . . . . . . . . . . . . . . . . 72

    A.1 Fejer quadrature points and weighs . . . . . . . . . . . . . . . . . . . . . 81

    ix

  • List of Figures

    1.1 Measured (top) and simulated (bottom) magnetization and T2 maps of an

    artery wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

    2.1 Transformation of a plane wave in local coordinates in the standard trian-

    gle to the unit square . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

    2.2 tri3 domain s and vertices in global and local coordinate systems with a

    given K space vector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

    2.3 The plane formed by the gradient of g with respect to ξ1 and the location

    of its possible maximum absolute values in domain s. In this case the

    maximum absolute value of the gradient is at ξ1 = ξ2 = 0 . . . . . . . . . 37

    3.1 Flowchart of the main FE-MRI algorithm . . . . . . . . . . . . . . . . . 44

    4.1 K-space error as a function of k-space location for a tet4 unit cube with a

    quadrature order of 14 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

    4.2 Number of k-space points as a function of n where error = 10n for a tet4

    unit cube with a quadrature order of 14 . . . . . . . . . . . . . . . . . . . 58

    4.3 Mesh of segmented artery (a) and its MRI reconstruction (b) . . . . . . . 61

    5.1 K-space computation time maps in milliseconds for different adaptive al-

    gorithms and error thresholds. The time ratios between the algorithms is

    also shown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74

    x

  • B.1 Discretization of a circle using quadratic curve segments. The quantity to

    be minimized is proportional to the area between the lines . . . . . . . . 89

    B.2 Discretization of a circle using line segments. The quantity to be mini-

    mized is proportional to the area between the lines . . . . . . . . . . . . 91

    xi

  • Chapter 1

    Introduction

    1.1 Motivation

    MRI simulators are primarily used to determine the origin of acquisition errors and to

    design and optimize new pulse sequences. For example in [1], an MRI simulator is used

    to design and optimize a MRI pulse sequence for the characterization of artherosclerotic

    plaque. They are also used for education and training [2]. The motivation of this thesis

    is to design an adaptive Gauss quadrature to evaluate the MRI signal equation for un-

    structured meshes and to design and test an algorithm using numerical steepest descent

    (NSD) [3], [4], [5], [6]. Using NSD one can compute the Fourier transform of a mesh

    faster than other approaches like numeric quadrature [7]. The algorithm was tested us-

    ing parameters needed for simulating an MRI acquisition. These results are then used

    to implement and improve upon new functions in the FE-MRI library [7] which can be

    used to simulate MRI acquisition.

    1.2 Introduction to MRI theory and simulation

    In this section, an introduction to magnetic resonance imaging (MRI) will be given with

    emphasis on signal acquisition, its Fourier transform approximation (k-space) and image

    1

  • Chapter 1. Introduction 2

    artefacts. The introduction will have an emphasis on imaging and reconstruction methods

    that are based on the Fourier transform and for 2D imaging. This introduction will also

    aid the reader in understanding how in some cases an MRI acquisition of an object can be

    simulated by the Fourier transform and inversion of a set of points or elements weighted

    by the actual material properties of said element such as in the FE-MRI library.

    The MRI acquisition process can be separated into the following steps:

    1. Signal generation: generating a net magnetization in an object

    2. Slice selection: selectively exciting a plane with a given thickness if a 2D acquisition

    is desired

    3. Spacial encoding: spatially encoding the signal in that plane

    4. Signal acquisition and reconstruction: acquiring a radio frequency signal and re-

    constructing the image

    These steps will be described in detail in the subsequent subsections. This introduc-

    tion is based on the following sources: [8], [9], [10]. An overview is given next. MRI

    works by aligning nuclei in an object along their spin axes more in one direction than

    another. This spin is a magnetic and quantum effect but it can be described in the same

    way as a spinning top. This is done by a very strong fixed magnetic field B0. Then

    another moving magnetic field that is perpendicular to the fixed field called B1 tips the

    spins and they precess around the original fixed field. The amount of tipping depends on

    how long the B1 field is applied. The speed of precession is dependant on the strength

    of B0. The B1 field can be chosen in a way that allows for selective excitation of a slice

    perpendicular to the z axis. Another smaller fixed gradient field Gx (i.e B = B0 +Gxx)

    is superimposed on B0 to cause a gradient in the total fixed field. The precessing spins

    emit radio waves that can be picked up by a coil. The signal’s frequency and magnitude

    are related to the spatial location of the spins along the x axis and the amount of spins

  • Chapter 1. Introduction 3

    in that given location. Hence an image of the spin distributions can be reconstructed by

    taking the inverse Fourier transform of the signal. To obtain a 2D distribution of spins,

    the locations along the y-axis can be encoded by using a gradient field Gy applied before

    the measurement is taking place, so that the phase of each spin depends on the location

    along the y-axis. The readout coils which do the measurement can measure the phase,

    magnitude and frequency of the signal which means the 2D distribution of spins in a slice

    can be determined.

    1.2.1 Signal generation

    The coordinate axes that will be used in this and subsequent sections are only used for

    convenience and consistency with MRI literature. Some nuclei in living things, like in

    hydrogen atoms (1H), can have nuclear spin angular momentum. These types of nuclei

    will be referred to as spins. The most common spin used for MRI is 1H since it is a

    component of water and the human body is mostly water. Spins can be described as

    small bar magnets oriented randomly in a given material. When the spins are exposed

    to a static magnetic field, some spins will align with the magnetic field and others align

    in the opposite direction to it and in both orientations they will rotate around the z axis

    with an angular frequency given by the Larmor equation:

    ω = γB0 (1.1)

    Where γ is the gyromagnetic ratio which for a particle is the ratio between its magnetic

    dipole moment and its angular momentum. For 1H, γ/2π = 42.48 MHz/T [10]. A group

    of spins in an object with the same γ value are commonly referred to as isochromats. B0

    is the applied magnetic field referred to as the static field and it points in the positive

    z-axis. Slightly more spins will align in the same direction as B0 because a spin is more

  • Chapter 1. Introduction 4

    likely to take the lower energy state than a higher energy state. The net magnetization

    is simply the spins oriented towards the positive z-axis minus the ones oriented towards

    the negative z-axis and it is given by the following equation:

    M0z = | ~M | =γ2~2B0Ns4KTs

    (1.2)

    Where ~ is Plank’s constant divided by 2π, Ns is the number of spins, K is Boltzmann

    constant and Ts is absolute temperature of the spins. The derivation of equation 1.2

    is similar to the one used in [11] to determine the properties of a simple paramagnetic

    solid. More information is given in [10] and [11]. Equation 1.2 only applies to so called

    spin-1/2 systems like 1H isochromats. The spins also can be made to rotate with other

    magnetic fields with respect to any line in the x-y plane but only by discrete angles;

    it is a quantum mechanical effect. However since the net magnetization is a sum of

    billions of spins in the whole object, the angle of rotation for the magnetization can be

    assumed to be continuous and a classical description analogous to a precessing spinning

    top can be used. After applying the static field, the net magnetization will point in the

    same direction. The equation that describes the time-dependent behaviour of ~M in the

    presence of an applied magnetic field is called the Bloch equation and it is given by:

    d ~M

    dt= γ ~M × ~B − Mxî−My ĵ

    T2− Mz î−M0k̂

    T1(1.3)

    Where ~B is the total applied magnetic field, Mx−z are the components of ~M and T1 and

    T2 are the longitudinal and transverse relaxation times respectively. The range of values

    for the relaxation times for 1H are T1 ∈ [100, 1500] ms and T1 ∈ [100, 1500] ms [10]. Let

    us introduce a transverse magnetic field that oscillates at the Larmor frequency:

  • Chapter 1. Introduction 5

    ~B1(t) = 2Be1(t)cos(ωt+ ϕ)̂i (1.4)

    Where Be1(t) is called an envelope function that for example can have a value of B1 from

    time 0 to some time τp and 0 otherwise. τp is simply the amount of time the B1 field is

    applied. It is also referred to as the pulse width. It will be shown later in this Section

    that the useful factor is only the integral of Be1(t) from 0 to τp and not its shape. The

    value ϕ is an initial phase angle. This longitudinal magnetic field can be decomposed

    into fields rotating in opposite directions about the z-axis. One of the terms rotates in

    the opposite direction as the processing spins and has a negligible effect. Therefore the

    effective field that needs to be considered is:

    ~B1(t) = Be1(t)[cos(ωt+ ϕ)̂i− sin(ωt+ ϕ)ĵ] (1.5)

    The magnetization vector 1.5 can be transformed into a vector in the Larmor-rotating

    frame with the following coordinate transformations:

    î′ ≡ cos(ωt)̂i− sin(ωt)ĵ

    ĵ′ ≡ sin(ωt)̂i+ cos(ωt)ĵ

    k̂′ ≡ k̂ (1.6)

    Using the transformation 1.6, the magnetic field 1.5 becomes:

    ~B1,rot(t) = Be1(t)î

    ′ (1.7)

  • Chapter 1. Introduction 6

    Which is expected since the vector ~i′+~j′ is spinning around the z-axis in the same direction

    and angular speed as the original ~B1(t) vector. If the pulse width is small compared to

    both relaxation times, the effect of ~B1(t) on the magnetization ~M is described by the first

    term of the Bloch equation. As will be shown below, it is also advantageous to transform

    that term into the Larmor-rotating frame. The transformed Bloch equation for a short

    exitation using the transformation 1.6 is:

    ∂ ~Mrot∂t

    = γ ~Mrot × ~Beff (1.8)

    Where ~Beff = ~Brot + ~ω/γ is referred to as the effective magnetic field. Since the total

    rotating field is B0~k′ + Be1(t)î

    ′, the motion of the bulk magnetization vector will be

    described by:

    ∂ ~M

    ∂t= γ ~Mrot × ~Be1(t) ~i′ (1.9)

    Notice that equation 1.9 will be equal to zero (i.e the bulk magnetization does not precess)

    if the only field present is the original B0~k′ magnetization. The solution to equation 1.9

    up to the pulse width time τp with the initial condition ~M(0) = M0z is the following:

    Mx′(t) = 0 (1.10)

    My′(t) = M0z sin(α) (1.11)

    Mz′(t) = M0z cos(α) (1.12)

  • Chapter 1. Introduction 7

    Where α is referred to as the flip angle and its given by:

    α =

    ∫ t

    0

    γBe1(u)du (1.13)

    If the envelope function is rectangular then α = γB1t. So in this case the bulk magneti-

    zation tips away from the z′ axis about x′ or precesses (since its tipping but also rotating

    about z in the laboratory frame) at an angular velocity of −γB1. Also notice that as

    long as the area under B1(t) is the same, the bulk magnetization will tip to the same

    final angle.

    1.2.2 Slice selection

    Since the envelope function Be1(t) can have any shape and still produce the desired tip

    angle as long as its area is constant, it can be used to selectively excite a slice of the object

    being imaged. For simplicity we will assume the slice is perpendicular to the z axis. If

    we apply a gradient to the B0 field as a function of the z component of location while

    the spins are being tipped, their positions will be spatially encoded by their resonant

    frequencies with the following frequency selection function;

    p(f) =

    1, |f − fc| < ∆f

    0, otherwise

    (1.14)

    Where:

    fc = f0 +γ

    2πGzz0 (1.15)

  • Chapter 1. Introduction 8

    ∆f =γ

    2πGz∆z (1.16)

    To obtain the time signal needed to produce the corresponding frequency spectrum, the

    Fourier transform can be applied on equation 1.14:

    B1(t) ∝∫

    −∞

    p(f)e−2πift dt = ∆fsinc(π∆ft)e−2πifct (1.17)

    From equation 1.18, we get the envelope function:

    Be1(t) = Asinc(π∆ft) (1.18)

    Where the constant A depends on the desired spin tip angle. This pulse is not possible

    to apply in practice so it is truncated from time 0 to some time τp, with the center of the

    sinc function at τp/2.

    1.2.3 Spatial encoding

    One can use the static field B0 to encode spatial information in the final MRI signal. A

    simple way of achieving this is with a constant gradient field such as:

    ω(x) = ω0 + γGxx (1.19)

    So now the field B0 pointing along the z-axis changes magnitude along x. After tipping

    the spins uniformly along a slice in the z-axis to 90◦ for example, this gradient can be

    used to spatially encode a signal using both phase and frequency. This is needed because

  • Chapter 1. Introduction 9

    the image signal in 2D acquisition is:

    s(t) =

    ∫∫

    m(x, y)e−iγ(Gxx+Gyy)tdxdy (1.20)

    Which is the Fourier transform of the object’s magnetization, m(x, y), on the particular

    slice being imaged:

    M(kx, ky) =

    ∫∫

    m(x, y)e−2πi(kxx+kyy)dxdy (1.21)

    Where the Fourier coordinates for general G gradients are given by:

    kx =γ

    ∫ t

    0

    Gx(u) du (1.22)

    ky =γ

    ∫ t

    0

    Gy(u) du (1.23)

    In the MRI literature Fourier space is referred to as k-space. Phase encoding can be

    achieved as follows: first tip all spins 90◦ in the same direction using the B1(t) field

    discussed in Section 1.2.1, then apply a gradient field which happens along a line in

    the x-y plane (the magnetic field is still pointing in the z-axis) for a small time Tpe

    which stands for phase encoding time. After Tpe seconds, the spins will have a phase

    proportional to their location. For example if the G gradient that happens along the

    y-axis and is constant then the phase of the signal as a function of y is:

    φ(y) = −γGyyTpe (1.24)

  • Chapter 1. Introduction 10

    If the signal is acquired right after Tpe for a very short time interval, we would obtain

    the sum of the magnetizations for each line that contains the same phase. However, if we

    apply yet another gradient field such as Gx while recording the signal every ∆t seconds,

    we will also obtain the magnetization along those lines as a function of the frequency,

    i.e the spatial information along the line has been frequency encoded. This procedure is

    referred to as a pulse sequence. The sequence described above is referred to as spin-wrap

    imaging and it is one of the most commonly used sequences because it results in equally

    spaced points in k-space on a grid which can be easily transformed into an image using

    the fast Fourier transform.

    1.2.4 Signal acquisition and reconstruction

    Equation 1.20 for the MRI signal is obtained from radio waves given off by the object

    being imaged which are measured by coils in the MRI machine. The signal from the

    coils is then converted to a digital signal and its phase and frequency components are

    determined. If the signal is sampled in k-space at uniform intervals such as in spin-warp

    imaging, the signal can be obtained applying the inverse discrete Fourier transform of

    the discrete version of the signal equation. This method can take O(N2) operations.

    The same transform can be performed in O(Nlog2N) operations using the fast Fourier

    transform or FFT. FFT is the reconstruction method most likely used in MRI equipment.

    Other transforms are available for k-space acquisition that is sampled radially such as

    the Radon transform. For more information refer to [9].

    1.2.5 Image resolution

    When the MRI signal is being acquired, k-space is recorded at discrete time intervals.

    Using equations 1.23 and 1.22 the k-space spacing is:

  • Chapter 1. Introduction 11

    ∆ky =γ

    2π∆Gyty (1.25)

    ∆kx =γ

    2πGx,max∆t (1.26)

    Because of the Fourier sampling theorem [9], the field of view of the image should be the

    following:

    FOVx =1

    ∆Kx(1.27)

    Which also applies for the y and z axes. The spatial resolution of the image in the x axis

    is the field of view divided by the number of readout samples:

    ∆x =FOVxNread

    =1

    ∆kxNread(1.28)

    While the spatial resolution in the y axis is the field of view divided by the number of

    phase encoding steps:

    ∆y =FOVyNpe

    =1

    ∆kyNpe(1.29)

    Notice the above relations apply only to spin-warp imaging. They do not apply to other

    pulse sequences that sample k-space non-uniformly.

  • Chapter 1. Introduction 12

    1.2.6 Image artefacts

    Image artefacts arise in MRI acquisition either due to insufficient or inaccurate data or

    both. The motivation of this thesis is to provide a tool to simulate MRI such that the

    cause of image errors can be understood and also to design and test protocols such as the

    resolution and how it affects the measurement accuracy of some anatomic feature similar

    to [12] for example. With this in mind, the following image errors are described together

    with their causes and how they can be simulated in general or in the FE-MRI library.

    Gibbs or truncation artefact

    The Gibbs or truncation artefact is seen as spurious ringing around sharp discontinuities

    in the reconstructed image. If the image is a smooth function, the inverse transform of

    the signal equation uniformly converges to the true image when the number of samples

    N → ∞ for a given FOV . However if the function is discontinuous, the inverse transform

    will have ringing with an overshoot and an undershoot of about 9% of the difference at

    the discontinuity. Surprisingly the overshoot and undershoot percent value is constant

    regardless of N . However the period of the ringing will decrease with higher N values.

    If the signal was not truncated to a finite number of N samples then the ringing period

    would be zero and the exact image would be reconstructed hence the name truncation

    artefact. This artefact can be made to appear on any simulated MRI acquisition simply

    by using a low enough number of N samples.

    Aliasing

    Aliasing happens when k-space is under-sampled. The artefact is a wraparound of the

    image in the direction that is being under-sampled. This can be clearly understood by

    the relationship between the true image function and its Fourier series reconstruction [9]:

  • Chapter 1. Introduction 13

    Î =∞∑

    n=−∞

    I

    (

    x− nx∆kx

    )

    (1.30)

    Where the term 1/∆kx = FOVx. This error can be readily be simulated by simply

    using an FOV value that is smaller than the object size along the same direction. This

    error can be eliminated simply by choosing an FOV value larger than the object being

    imaged.

    Chemical shift artefacts

    Chemical shift artefacts happen when an object with different isochromats is being im-

    aged. Different isochromats will have a different γ value. Since spatial position is fre-

    quency encoded in the readout direction when an MRI signal is being obtained, signals

    coming from two different isochromats in the same location such as water and fat will

    end up in different locations along the frequency encode signal. The artefact will look

    like a region of the object such as fat which is not being explicitly imaged will be shifted

    compared to the rest of the image features. The frequency shift due to the different

    isochromat is given by [9]:

    ∆ωc = γδB0 (1.31)

    The frequency bandwidth of a pixel in k-space is:

    ∆ωx = γG∆x (1.32)

    Since the pixel bandwidth is proportional to the pixel length, the spatial displacement

    in units of a pixel is:

  • Chapter 1. Introduction 14

    δx =∆ωc∆ωx

    =δB0G∆x

    (1.33)

    And the spatial shift will be:

    ∆xc = δx∆x =δB0G

    (1.34)

    Therefore an effective way to reduce the chemical shift is to use a strong readout

    gradient. In practise this will also reduce the signal to noise ratio of the acquired signal.

    Therefore an optimal value for G has to be determined. This error can be simulated

    simply by obtaining the k-space or each isochromat separately and then adding the

    signal.

    Motion artefacts

    A type of very obvious motion artefact that happens when the whole object being scanned

    is in motion is called ‘ghosting’. It can be clearly identified by copies of the object being

    imaged along the direction of motion and with a lower contrast than the actual image.

    This error happens when the object moves along the phase-encoding direction. For

    example, assuming the object moves at a constant speed vx along the x-direction and

    there is a constant phase encoding gradient Gx applied for some time interval τ starting

    from time tn, the phase shift is the following:

    φ = γ

    ∫ tn+τ

    tn

    Gn(x+ vxt) dt

    = γGnxτ + γGnvxτ(tn + τ/2)

    = 2πknx+ 2πknvx(tn + τ/2) (1.35)

  • Chapter 1. Introduction 15

    So the motion induced phase shift is:

    ∆φm = 2πknvxtn + 2πknvxτ/2 (1.36)

    Since the time τ is much smaller than the total acquisition time up to that point, the

    second term of equation 1.36 can in most cases be omitted. Equation 1.36 without the

    second term can be obtained directly by applying the Fourier shift theorem to the MRI

    signal equation. For constant vx, Gx and a negligible τ , the signal equation is:

    sm(k, t) = e−2πiknvxts(k, t) (1.37)

    Equation 1.37 has the same phase shift as the one given in equation 1.36. To simulate

    this error, either equation 1.37 can be applied directly to the simulated k-space map for

    a constant speed vx or the object can be translated a certain distance based on vx in the

    x direction and then its corresponding k-space points calculated to simulate the error. A

    further discussion on other types of motion artefacts is given in [9].

    Flow artefacts

    Flow artefacts are caused by the motion of fluids such as blood in the living tissue being

    imaged. The main feature of these errors is a loss or gain of signal intensity where the

    flow is taking place like inside an artery [13]. The only simulation method that has been

    used successfully for these types of error is to either measure the flow directly using other

    imaging methods or to simulate the flow using computational fluid dynamics (CFD).

    Then a certain number of streamlines are extracted from the flow field. Afterwards the

    individual effect of spins moving along the streamlines is computed to obtain an MRI

  • Chapter 1. Introduction 16

    signal. References [13] and [14] use this method of simulation. Since a set of discrete

    spins are used to simulate a continuum fluid are used, the number of spins needed and

    the related computation time is high.

    1.3 Literature review

    The different MRI simulation methods can be grouped into the following types:

    1. Analytic solutions

    2. Simulators using magnetization maps

    3. Simulators using T1, T2 and proton density maps

    4. Finite element simulators

    5. Bloch equation simulators

    The order on the list is based on increasing physical accuracy of the models in most

    cases at the expense of computation time. Except for some Bloch equation simulators, the

    other types use the Fourier transform of a pre-described signal therefore they technically

    are not simulators in the sense that they are not simulating a full MRI machine. A full

    simulation would require modelling both the object using the Bloch equation and the

    MRI components such as the coils and the B0 magnet. However the methods described

    below will be referred to as simulations since they do still crudely simulate the object

    and parts of the MRI machine. Each type is described in the following sections.

  • Chapter 1. Introduction 17

    1.3.1 Analytic solutions

    In some situations, an analytic solution for k-space can be found. The author co-authored

    a paper, [12], where MRI acquisition of an artery was performed to determine how the

    total arterial wall thickness obtained from the segmentation of the MRI image depends

    on the T2 values of the arterial layers (intima, media and adventitia) and TE . To test

    if this is due to a partial volume effect (a low resolution) or because of the different

    T2 values of each layer, an analytic k-space was constructed for each of the 4 acquired

    images by adding the k-spaces of 4 concentric circles. Then the obtained images were

    segmented. The values used are given in the following table:

    Layer ro(mm) pTE=17ms pTE=34ms pTE=51ms pTE=68ms

    arterial blood 3.7 0.0000 0.0000 0.0000 0.0000

    intima 3.82 1.0242 0.8800 0.7561 0.6496

    media 4.35 1.0000 0.7699 0.5927 0.4563adventitia 4.42 0.7063 0.3539 0.1773 0.0888

    background N/A 0.0000 0.0000 0.0000 0.0000

    Table 1.1: Values used for fig. 1.1

    Where ro and p in Table 1.1 stands the outer radius for the image intensity. The image

    intensity is in arbitrary units; the actual value depends on the specific MRI scanner. The

    thickness changes as a function of TE closely match the experimental results. TE is called

    the echo time and is simply the time where the spins are allowed to relax after a RF

    filed is applied. Then a mono-exponential fit was performed (p = e−TE/T2 , where p is the

    image intensity at a pixel) on the 4 simulated images to get a T2 map; it also closely

    matches the T2 map obtained from the actual MRI images. For the mono-exponential

    fit, if a pixel had p values lower than 1.5 × 10−5 on all the 4 images, it was not taken

    into consideration and was given an arbitrary value of 0. The measured and simulated

    magnetization and T2 maps are shown in Figure 1.1.

    The magnetization image shown has a magnetization normalized to 1. This image

    weighted with different magnetization values based on the echo time and the T2 decay are

  • Chapter 1. Introduction 18

    Image x−axis (mm)

    Image y

    −axis

    (m

    m)

    −6 −4 −2 0 2 4 6

    −6

    −4

    −2

    0

    2

    4

    6

    Image x−axis (mm)

    Image y

    −axis

    (m

    m)

    −6 −4 −2 0 2 4 6

    −6

    −4

    −2

    0

    2

    4

    6

    T2 (m

    s)

    50

    100

    150

    200

    250

    Figure 1.1: Measured (top) and simulated (bottom) magnetization and T2 maps of anartery wall

    used in the paper. To obtain the k-space, corresponding to a simplified artery, 4 circle k-

    spaces were added, this can be done since the Fourier transform is a linear transform [9].

    If the artery is long compared to its radius, a 2D Fourier transform can be used instead

    of the Fourier transform for a cylinder. If image and k-space coordinates are changed to

    cylindrical coordinates: r =√

    (∆x)2 + (∆y)2, θ = tan−1(∆x/∆y), kr =√

    k2x + k2y and

    φ = tan−1(kx/ky). Therefore:

    Kxx+Kyy = krr(cosφcosθ + sinφsinθ) = krrcos(θ − φ) (1.38)

  • Chapter 1. Introduction 19

    Then the Fourier transform of a circle with unit radius and height is given in cylin-

    drical coordinates as follows:

    Fcirclesimple(kr) =

    ∫ 1

    0

    r

    [∫ 2π

    0

    e2πikrrcos(θ−φ)dθ

    ]

    dr (1.39)

    The integral can be simplified in cylindrical coordinates since the function for a cycle

    is circularly symmetric. The solution of equation 1.39 is given in terms of the first order

    Bessel function of the first kind J1 [9] and its the following:

    Fcirclesimple(kr) =J1(2πkr)

    kr(1.40)

    In the case of kr = 0, a Taylor series can be taken of equation 1.40 about kr = 0, and

    simplified to yield:

    Fcirclesimple(0) = π (1.41)

    Applying the scaling and linearity properties of the Fourier transform [9] to equations

    1.40 and 1.41 yield the following equations for an arbitrary image intensity (h) and radius

    (r):

    Fcircle(kr) = hrJ1(2πrkr)

    kr(1.42)

    Fcircle(0) = hr2π (1.43)

  • Chapter 1. Introduction 20

    This k-space is implemented in the FE-MRI library as a python script. An example

    of its use in Unix is given below:

    femri2dcylinderkspace -fov 12.0 12.0 1.0 -fovcenter 0.0 0.0 0.0

    -matrix 32 32 1 -mvalue 1.0 -slicethickness 0.0 -radius 3.0

    -center 0.0 0.0 0.0 --pipe femrikspacetoimage -dimensionality 2

    -ofile circleImage.vti

    Both in the magnetization maps derived from measurements and the analytic model

    it was found that the mean wall thickness obtained by segmentation decreases with echo

    time because of the small T2 time of the adventitia. Therefore a short echo time should be

    included to properly image it. This layer plays an important role in plaque development

    [12].

    1.3.2 Simulators using magnetization maps

    In [15], artefacts due to relaxation, linear flow and motion after an RF-pulse are added

    to the k-space of the magnetization also called proton density map. Some limitations

    of this method are not able to simulate errors due to non-linear gradients and motion

    during and RF-pulse, since they are considered to be momentary.

    1.3.3 Simulators using T1, T2 and proton density maps

    In [2], the T1, T2 and proton density maps are Fourier transformed and changed according

    to limited properties of MRI pulse sequences. Although the method is computationally

    fast and can simulate wraparound and simple motion artefacts accurately, by design it

    cannot simulate errors related to signal generation. This type of simulation is used in [1]

    to optimize a MRI pulse sequence for the characterization of atherosclerotic plaque.

  • Chapter 1. Introduction 21

    1.3.4 Finite element based simulators

    Finite element based simulators also use a magnetization map and can include T1, T2

    maps, however they use the finite element discretization which allows the magnetization

    to vary continuously within the object. The elements used for this thesis are shown

    in Table 1.2. The names given in the figure will be used in the rest of the thesis.

    Both types of simulators have to transform the magnetization map into the Fourier

    domain. The Fourier transformation is then done in two different ways: discretizing the

    object into a set of points and then performing an FFT, or discretizing the object into

    elements with continuous values inside them as stated above and then computing exactly

    or approximating to a given error the Fourier transform for each element. The total

    acquired signal using the finite element discretization is the following:

    Stot(~k) =∑

    m

    [

    fm(T1, T2, ρ)∑

    e

    se(~k)

    ]

    (1.44)

    Where m stands for materials with different T1, T2, or ρ values, and e stands for all

    the elements representing the object. The advantage of this method is that if the error

    incurred by the finite element discretization is low enough, the error between the actual

    k-space values of the object and the simulated values can be either zero or can be made

    as low as is required (depending on the allowable computation time) depending on the

    element type. Also this approach can be faster than Fourier transform simulators if one

    starts with a mesh of a complex object, ideally an unstructured mesh to minimize anti-

    aliasing or “stair-step” effects, and takes into account the interpolation step to obtain

    a structured grid. This last operation scales as O(E) where E is the number of finite

    elements which also scales with the mesh volume [16]. Publications using this approach

    are [17] and [18]. In [17], the linear elements tri3 and tet4 are integrated analytically.

    If the magnetization inside a certain region of an object can be assumed to be constant

  • Chapter 1. Introduction 22

    Gauss’ theorem can be used to transform the signal equation into a surface integral [19]:

    se(~k) =

    ∫∫∫

    element

    me−2πi~k·~xd~x

    = m

    ∫∫∫

    element

    i

    2π~k · ~k∇ · e−2πi~k·~xd~x

    =im

    2π~k · ~k

    ∫∫

    surface

    ~k · n̂(x)e−2πi~k·~xdS (1.45)

    Where dS is an infinitesimal surface area and n̂(x) is the unit normal pointing outwards

    from the closed surface. The normal is constant for a given linear element but varies

    with the global coordinates for quadratic elements. As explained in [19], this method is

    more efficient than volumetric integration of the signal equation in the following three

    ways. No volumetric meshing step is needed where volume elements are constructed

    from the surface mesh. The number of elements over which integration is performed is

    reduced and scales with the surface area of the shape and not its volume. Also fewer

    integration points are needed for surface vs. volume elements for the same integration

    order. The analytic solution for tri3 in [17] can also be adapted to solve equation 1.45

    exactly. This method also reduces the storage requirements for the phantom compared

    to Fourier transform simulation methods that require a structured grid with a resolution

    greater than the nominal element size of the mesh [19]. The drawback of this method is

    that one has to decide in how many parts and object should be segmented so that the

    difference between material properties within each part is kept to a minimum.

    The advantage of FE-MRI over voxel based methods is that if the imaging resolution

    approaches the voxel size then “stair-step” errors will be produced. These error are also

    referred to as partial volume effects. This does not happen if an unstructured mesh

    with continuous changes in material properties within each element are used. These

    property makes this simulation method ideally suited to study how changing MRI imaging

  • Chapter 1. Introduction 23

    parameters affect the image quality. With voxel based methods the image quality might

    show errors that are due to the simulation itself.

    1.3.5 Bloch equation simulators

    Bloch equation simulators such as [20] and [21] allow for close to all possible computation

    of errors seen in MRI such as the errors due to static field inhomogeneity. However, the

    excessive time needed to compute one voxel means that not enough voxels can be used

    such that there will be voxel discretization artefacts [20].

    1.4 MRI acquisition within the FE-MRI library

    The volumetric and surface integration steps described in Subsection 1.3.4 have been

    implemented in the FE-MRI library for all the elements in Table 1.2. The main advantage

    compared to simulators using a structured grid of T1, T2 and proton density values is

    that the interpolation step is not needed if one has an unstructured mesh. If one only has

    a surface mesh and can assume these values are constant, the Fourier transform of the

    surface mesh can be computed directly from it as explained in Subsection 1.3.4. This is a

    fair assumption since organs with the same tissue types will have the same isochromats.

    Also the Fourier transform can be calculated exactly for linear elements such as tri3 and

    tet4, or with a desired image error for quadratic elements tri6 and tet10.

    1.5 Research objectives

    In previous work, [18], an error metric for linear elements was derived and used for an

    adaptive Gauss quadrature to compute 1.44 and 1.45. The error metric is an estimate

    of the error between the calculated and exact signal for a given element and k-space

    value. The error metric is used to estimate the number of points for an adaptive Gauss

  • Chapter 1. Introduction 24

    quadrature that integrates the MRI signal for each element in an unstructured grid.

    This was implemented by prof. Steinman and Luca Antiga in the FE-MRI library. The

    contributions of this thesis are as follows:

    1. A derivation of the error metric found in [18] using the element’s shape functions

    and k-space values and derivations of error metrics for three more element types

    including quadratic elements

    2. An addition to FE-MRI that can estimate errors of Gauss quadratures in many

    dimensions from the error obtained in one dimension using a result in [22] and ad-

    ditions to FE-MRI that allow it to compute the MRI signal using the new elements

    3. A test to see if the most promising of the methods to compute highly oscillatory

    integrals (NSD) could be used instead of adaptive Gauss quadrature which is cur-

    rently used in FE-MRI

    The next Chapters explain in detail each of the above contributions.

  • Chapter 1. Introduction 25

    Element diagram Name Nodes Shorthand

    Linear triangle 3 tri3

    Quadratic triangle 6 tri6

    Linear tetrahedron 4 tet4

    Quadratic tetrahedron 10 tet10

    Table 1.2: Finite element types used in this thesis

  • Chapter 2

    Adaptive Gauss quadrature for

    FE-MRI elements

    2.1 Introduction

    To obtain a signal with a uniform error in k-space, an adaptive Gauss quadrature scheme

    can be used. Gauss quadrature is a method of numerical integration where the function

    to be integrated is evaluated at some point called a quadrature point and multiplied by

    a weight for each point called a quadrature weight. The sum of all these products is an

    estimate of the integral:

    1∫

    −1

    f(x)W (x) dx =N∑

    i=1

    wif(xi) + E (2.1)

    Where f(x) is the function to be evaluated and wi and xi are the quadrature points and

    weights. W (x) is another function that has a constant form. A gauss quadrature is a

    special quadrature for numerical integration that allows the exact integration of f(x) for

    a given W (x) such as 1 or e−x if f(x) is a polynomial of degree 2N − 1 or less. If the

    function is ‘smooth’ in the sense that it can be approximated by a polynomial (such as a

    26

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 27

    sinusoid), the integration error will be low. Gaussian quadrature can be readily extended

    for higher dimensions. These quadratures are referred to as product rules. Although

    these are easily derived from 1D rules, they require more function evaluations than is

    necessary to integrate over higher dimensional domains. Optimal Gauss weights and

    point locations exist for linear tetrahedral domains, but are only available for low inte-

    gration orders and are not easily derived for higher orders. Non-optimal rules use more

    Gauss points and weights than optimal rules but they are easier to extend to arbitrary

    integration orders. The parametric representation of finite elements allows them to be

    integrated in a standard local tetrahedral domain. One can do a coordinate transfor-

    mation referred to as a Duffy transformation [23] that maps the quadrature points and

    weights of the quadrature product rule to the same local standard tetrahedral domain.

    One of the integrals we would like to compute is the MRI signal equation for volumetric

    finite elements which is the following:

    se(~k) =

    ∫∫∫

    element

    m(~x) e−2πi~k·~x d~x

    =

    ∫ 1

    0

    ∫ 1−ξ1

    0

    ∫ 1−ξ1−ξ2

    0

    |J |m(~ξ ) e−2πi~k·~x d~ξ (2.2)

    The term J is the element’s Jacobian which is given by:

    J = det

    ∂x∂ξ1

    ∂y∂ξ1

    ∂z∂ξ1

    ∂x∂ξ2

    ∂y∂ξ2

    ∂z∂ξ2

    ∂x∂ξ3

    ∂y∂ξ3

    ∂z∂ξ3

    (2.3)

    As can be seen from 2.2, the higher the magnitude of the k-space vector, the more

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 28

    oscillatory the integral is which can only be approximated by a higher order polynomial

    for a given approximation error. Therefore on average more quadrature points will be

    needed to compute with a fixed error when the magnitude of the k-space vector is higher.

    The global coordinates in terms of local coordinates are:

    xj =

    n∑

    i=1

    Ni(~ξ ) xij (2.4)

    And the magnetization in local coordinates is given by:

    m(~ξ) =n∑

    i=1

    Ni(~ξ ) mi (2.5)

    The terms Ni(~ξ ) are called the element’s shape functions and they are polynomials

    and n is the number of element nodes which are shown in Table 1.2. The term xij is the

    j component of node i. So the integral is a 3D polynomial multiplied by a sinusoid. The

    MRI signal equation for surface elements as derived in Subsection 1.3.4 is the following:

    se(~k) =im

    2π~k · ~k

    ∫∫

    surface

    ~k · n̂(x)e−2πi~k·~xdS

    =im

    2π~k · ~k

    ∫ 1

    0

    ∫ 1−ξ1

    0

    |J |~k · n̂(x)e−2πi~k·~xd~ξ (2.6)

    =im

    2π~k · ~k

    ∫ 1

    0

    ∫ 1−ξ1

    0

    ~k · ~n(x)e−2πi~k·~xd~ξ (2.7)

    The vector n̂(x) is the unit normal pointing outwards from the element which is part

    of a closed surface. The unit normal is given by:

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 29

    n̂(x) =~n(x)

    |~n(x)| =

    ∂~x

    ∂ξ1× ∂~x

    ∂ξ2∣

    ∂~x

    ∂ξ1× ∂~x

    ∂ξ2

    (2.8)

    The simplification from equation 2.6 to equation 2.7 can be made since the denomina-

    tor in equation 2.8 is also the element’s Jacobian for surface elements. Again the integral

    is a polynomial multiplied by a sinusoid.

    By changing variables to change the integration domain to the standard tetrahedron,

    the same product quadrature rules can be used each time. Each element signal is then

    weighted with the appropriate material properties and added to get the full MRI signal

    as shown in equation 1.44. To find out the number of quadrature points needed for both

    integrals, an estimate of the error for a given number of quadrature points is needed.

    Since the integration is carried out in the same standard domain with a fixed size, the

    error is only related to the maximum local frequency.

    2.2 Gauss quadrature to compute oscillatory inte-

    grals

    The maximum local oscillation of a function will determine the minimum distance be-

    tween quadrature points such that the Nyquist criteria is followed and the equation can

    be approximated properly in that area. We can then use a Gauss quadrature with enough

    points to guarantee a given accuracy at that point and the rest of the function since the

    oscillation everywhere else will be lower. Gauss quadrature also has the advantage of

    integrating the magnetization term exactly since it is described by a polynomial in finite

    elements. A way of estimating the maximum local oscillation is as follows: Consider the

    following 1D function:

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 30

    y = eig(x) (2.9)

    The equation can be approximated at x0 by taking the Taylor series of g(x) at x0 and

    substituting the result back into equation 2.9:

    y = eig(x0)eig′(x0)(x−x0)ei

    1

    2g′′(x0)(x−x0)2 ... (2.10)

    The term that is non-constant and will usually have the the highest oscillation fre-

    quency is eig′(x0)(x−x0). This term produces a frequency of g′(x0). To determine the

    maximum local frequency of y to perform a numerical integration from a standard do-

    main such as 0 to 1 one should therefore determine the following:

    max freq = max|g′(x)|, x ∈ [0, 1] (2.11)

    2.3 Computing oscillatory integrals in higher dimen-

    sions

    In 2D, if a product rule using Gauss quadrature is used with the same number of points

    in both coordinates axes to integrate a function in the unit square, the error will be

    dependant on the maximum local frequency in the unit square in the directions of the

    coordinate axes. Therefore this value will be given by:

    max freq = max

    (∣

    ∂g

    ∂η1

    ,

    ∂g

    ∂η2

    )

    , x ∈ s (2.12)

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 31

    Where η1 and η2 are the unit square coordinates and s is the boundary of the unit

    square.

    2.4 Error estimation for tri3 elements

    The integration in local coordinates for triangular elements in the finite element method

    and FE-MRI is done using a Gauss quadrature product rule for the unit square where

    the point coordinates are transformed using the Duffy coordinate transform [23] to a

    unit triangle. This is the same as transforming g in the triangle to the unit square

    and integrating it with a Gauss quadrature product rule in the unit square. So to

    find the maximum oscillation in the standard triangle one should find how the maximum

    absolute values of the gradients of the function g in the unit square relate to the maximum

    absolute values of the gradients of g in the unit triangle. Although there are symmetric

    quadrature rules for triangles that are optimal in the Gauss-quadrature sense such as

    [24], their computation requires finding a non-linear least square solution for a system of

    polynomial equations which in [24] could only be done numerically up to 20 points per

    dimension due to round-off error. Also the estimation of the quadrature error is very

    non-trivial [25]. For a survey of different cubature (quadrature rules for 2 and higher

    dimensions) formulas over triangles refer to [25].

    To find these relationship lets look more closely at what happens to a plane wave

    (produced by the MRI signal equation with a given k-space vector for example) in local

    coordinates in the standard triangle when it is transformed to the standard square. Such

    a transformation is shown in Figure 2.1.

    In this case the gradient of g from the plane wave is constant in local coordinates

    along both coordinate axes. This gradient is also proportional to the number of cycles in

    the direction of each coordinate axes. The number of cycles is also proportional to the

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 32

    ξ1

    ξ 2

    0 0.2 0.4 0.6 0.8 10

    0.1

    0.2

    0.3

    0.4

    0.5

    0.6

    0.7

    0.8

    0.9

    1

    η1

    η2

    0 0.2 0.4 0.6 0.8 10

    0.1

    0.2

    0.3

    0.4

    0.5

    0.6

    0.7

    0.8

    0.9

    1

    Figure 2.1: Transformation of a plane wave in local coordinates in the standard triangleto the unit square

    number of red ‘crests’ or blue ‘troughs’ for a given length along a given direction. We

    can see that the magnitude of the gradient along η1 remains the same as the gradient

    along ξ1 in the unit square at ξ2 = 0 and decreases to 0 when ξ2 increases to 1. We can

    also see that the magnitude of the gradient along η2 appears to remain the same as the

    gradient along ξ2 in the unit square at ξ1 = 0. However in this case when ξ1 increases

    from 0 to 1, the gradient decreases to 0 and increases again to what appears to be the

    gradient along the edge from (0, 1) to (0, 1) of the standard triangle. These observations

    can be made precise as follows. Based on the derivation in [23] for the standard square

    with (−1, 1) bounds, the Duffy coordinate transform between the standard triangle and

    square used in FE-MRI is the following:

    ξ1 = η1(1− η2) (2.13)

    ξ2 = η2 (2.14)

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 33

    Using the chain rule for a multi-varied function we can relate the gradients of g in

    the unit square to the ones in the unit triangle as follows:

    ∂g

    ∂η1=

    ∂g

    ∂ξ1

    ∂ξ1∂η1

    +∂g

    ∂ξ2

    ∂ξ2∂η1

    =∂g

    ∂ξ1(1− η2) (2.15)

    ∂g

    ∂η2=

    ∂g

    ∂ξ1

    ∂ξ1∂η2

    +∂g

    ∂ξ2

    ∂ξ2∂η2

    =∂g

    ∂ξ2− ∂g

    ∂ξ1η1 (2.16)

    The maximum absolute value of equation 2.15 will happen when η2 is 0 and the

    maximum absolute value of equation 2.16 will happen when η1 is 0 is either 0 or 1. When

    η1 = 0 the gradient is along the same direction as the side from (0, 1) to (1, 0) in the

    standard triangle.

    In the MRI signal equation g is kxx + kyy or K · x. In finite element analysis, for

    triangular linear elements in 2D, the global coordinates are related to local coordinates

    in the following way:

    x =n∑

    i=1

    Ni xi, y =n∑

    i=1

    Ni yi (2.17)

    Where n is number of points in the element xi, yi are the global coordinates of point

    i and Ni are the element’s shape functions. For the tri3 element they are given by:

    N1 = ξ1, N2 = ξ2, N3 = 1− ξ1 − ξ2 (2.18)

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 34

    Using the element’s shape functions the maximum oscillation in the ξ1 and ξ2 direc-

    tions can be determined by:

    ∂g

    ∂ξ1= kx

    ∂x

    ∂ξ1+ ky

    ∂y

    ∂ξ1= K · e31 (2.19)

    ∂g

    ∂ξ2= kx

    ∂x

    ∂ξ2+ ky

    ∂y

    ∂ξ2= K · e32 (2.20)

    Where eij is a vector from element vertices i to j in global coordinates. Using the

    above two results yields:

    ∂g

    ∂ξ2− ∂g

    ∂ξ1= K · e32 −K · e31 = K · e12 (2.21)

    Note that these values are constant in local coordinates since the partial derivatives

    of the shape functions for linear finite elements are constant, therefore checking their

    maximum value in domain s is not necessary. These are the same error metrics used

    in [18]. In [18] the error metric is arrived at by observing that it should be invariant

    with respect to element size and shape and the k-space vector. These results can be

    interpreted as the dot product of the k space vector in local coordinates with each of the

    standard triangle side vectors expressed in terms of ξ1 and ξ2 which give the oscillation

    frequency per unit length in the local coordinate axes. The domain s in global and local

    coordinates is given in Figure 2.2.

    Since the k space vector in global coordinates is the following:

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 35

    y

    x(0,0) (1,0)

    (0,1)

    Figure 2.2: tri3 domain s and vertices in global and local coordinate systems with a givenK space vector

    K =

    kx

    ky

    =

    ∂g∂x

    ∂g∂y

    (2.22)

    The k space vector in local coordinates is given by:

    Kl =

    ∂g∂ξ1

    ∂g∂ξ2

    =

    ∂x∂ξ1

    ∂y∂ξ1

    ∂x∂ξ2

    ∂y∂ξ2

    ∂g∂x

    ∂g∂y

    = JK (2.23)

    Where J is the element’s Jacobian. Using the above result, the maximum oscillation

    estimates can be written as follows:

    ∂g

    ∂ξ1= Kl · (1, 0) = Kl · ξ̂1 (2.24)

    ∂g

    ∂ξ2= Kl · (0, 1) = Kl · ξ̂2 (2.25)

    ∂g

    ∂ξ2− ∂g

    ∂ξ1= Kl · (−1, 1) = Kl · (ξ̂2 − ξ̂1) (2.26)

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 36

    Where ξ̂1, ξ̂2 and ξ̂2 − ξ̂1 are the standard triangle side vectors. For the tri3 surface

    element the same results hold, except that the element sides eij and the k-space vector

    are in three dimensions. In this case the local k-space vector in the still 2D local domain

    is a projection of the original k-space vector.

    2.5 Error estimation for tri6 elements

    For quadratic elements equation 2.12 still applies, however the values of the gradients of

    g with respect to ξ1 and ξ2 in local coordinates will vary linearly with ξ1 and ξ2. The

    boundary s in local coordinates for tri6 is the same as the one for the tri3 element. Since

    the gradient of g for a given local coordinate g′(ξ1, ξ2) is a plane in the space formed by

    ξ2 and ξ2 and the possible values of g′(ξ1, ξ2), therefore the maximum absolute value of

    g′(ξ2, ξ2) will be at one of the vertices of the standard triangular boundary. Therefore

    equation 2.12 will have to be computed 3 times for each gradient for a total of 3 x 3 =

    9 gradients. A plane formed by the gradient of g in one of the 3 coordinates and the

    possible maximum absolute values are shown in Figure 2.3.

    The number of gradients that need to be computed is the number of gradients times

    the number of element vertices which is 9. The computation time will be lower if the

    gradients with respect to ξ1 and ξ2 are computed first and then the results are used to

    compute the gradient given by equation 2.16 with ξ1 = 1.

    These gradients can be computed even faster by storing and reusing some of the

    intermediate results. The needed computations for the tri6 element are as follows. The

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 37

    Figure 2.3: The plane formed by the gradient of g with respect to ξ1 and the location ofits possible maximum absolute values in domain s. In this case the maximum absolutevalue of the gradient is at ξ1 = ξ2 = 0

    element’s shape functions are as follows:

    N1 = ζ(2ζ − 1)

    N2 = ξ(2ξ − 1)

    N3 = η(2η − 1)

    N4 = 4ξζ

    N5 = 4ξη

    N6 = 4ηζ (2.27)

    For the gradient in the ξ1 direction:

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 38

    ∂g

    ∂ξ1= kx

    ∂x

    ∂ξ1+ ky

    ∂y

    ∂ξ1(2.28)

    = kx

    6∑

    i=1

    ∂Ni∂ξ1

    xi + ky

    6∑

    i=1

    ∂Ni∂ξ1

    yi (2.29)

    = K

    [

    x

    y

    ]

    ∂N

    ∂ξ1(2.30)

    Where K is a 1 by 2 matrix, [x y]T is a 2 by 6 matrix with the element’s node

    coordinates and ∂N/∂ξ1 is a 6 by 1 matrix. The same result applies for the gradient in

    the ξ2 direction. The matrices ∂N/∂ξ1 and ∂N/∂ξ2 can be precomputed for all elements

    by evaluating them at the local domain vertices (0,0),(1,0),(0,1). Half of the entries in

    these matrices are 0 therefore it is more useful to calculate the product [x y]T∂N/∂ξ1

    directly.

    2.6 Error estimation for tet4 elements

    For linear tetrahedral elements, to estimate the quadrature error, one has to estimate

    the maximum oscillation frequency in the standard cube given by equation 2.31.

    max freq = max

    (∣

    ∂g

    ∂η1

    ,

    ∂g

    ∂η2

    ,

    ∂g

    ∂η3

    )

    , x ∈ s (2.31)

    Where η1−3 are the unit cube coordinates and s is the boundary of the unit cube.

    The Duffy transformation for the unit cube, again based on the derivation in [23] for the

    standard cube with (−1, 1) bounds is the following:

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 39

    ξ1 = η1(1− η3)[1− η2(1− η3)] (2.32)

    ξ2 = η2(1− η3) (2.33)

    ξ3 = η3 (2.34)

    The derivation can be understood as using the 2D transformation recursively in two

    of the coordinate axes to ‘collapse’ the standard cube into the standard tetrahedron.

    Using equations 2.32, 2.33 and 2.34, the gradients of g with respect to the standard cube

    coordinates are the following:

    ∂g

    ∂η1=

    ∂g

    ∂ξ1(1− η3)[1− η2(1− η3)] (2.35)

    ∂g

    ∂η2= − ∂g

    ∂ξ1η1(1− η3)2 +

    ∂g

    ∂ξ2(1− η3) (2.36)

    ∂g

    ∂η3=

    ∂g

    ∂ξ1η1[2η2(1− η3)− 1]−

    ∂g

    ∂ξ2η2 +

    ∂g

    ∂ξ3(2.37)

    Since the values of ξ1−3 can only vary from 0 to 1, the maximum absolute value of

    the following gradients will determine the maximum oscillation frequency of g:

    ∂g

    ∂ξ1,

    ∂g

    ∂ξ2,

    ∂g

    ∂ξ3(2.38)

    ∂g

    ∂ξ2− ∂g

    ∂ξ1,

    ∂g

    ∂ξ3− ∂g

    ∂ξ2,

    ∂g

    ∂ξ3− ∂g

    ∂ξ1(2.39)

  • Chapter 2. Adaptive Gauss quadrature for FE-MRI elements 40

    ∂g

    ∂ξ3− ∂g

    ∂ξ2− ∂g

    ∂ξ1,

    ∂g

    ∂ξ3− ∂g

    ∂ξ2+

    ∂g

    ∂ξ1(2.40)

    The gradients shown in 2.38 and 2.39 are gradients along all the side edges of the

    standard tetrahedron as shown in the 2D case for the triangle. However the gradients

    shown in 2.40 which are derived from equation 2.37 are not related to any side of the

    standard tetrahedron, and would not have been taken into account in the maximum

    frequency calculation if extended the results for 2D elements (that the gradients along

    all the element side edges determine the maximum oscillation frequency) directly to 3D

    elements.

    2.7 Error estimation for tet10 elements

    For quadratic tetrahedral elements equation 2.31 still applies, however their values will

    vary linearly on ξ1−3 like in the tri6 element. The boundary s for tet10 is the same as the

    one for the tet4 element. To estimate the maximum frequency gradients 2.38, 2.39 and

    2.40 should be calculated at the standard tetrahedron vertices. The gradient calculations

    can be performed faster if all the components of the gradients are computed before hand

    and then used for gradients 2.39 and 2.40. Also the values of some of the intermediate

    steps can be stored and reused like in the case of the tri6 element.

  • Chapter 3

    Algorithm details and

    implementation

    3.1 Introduction

    The FE-MRI algorithm simulates either a 2D or 3D MRI signal acquisition. If a 2D

    signal is required the program computes a given number of k-space planes depending

    on the slice thickness and the needed resolution. If not the program computes a given

    k-space volume. If using an acquisition with a fixed number of quadrature points, the

    program computes the needed Gauss quadrature and uses that quadrature for all elements

    and k-space values. If the adaptive quadrature is used, the program computes all the

    Gauss quadratures requested, then it computes the ones needed for all the element/k-

    space point combinations and uses the appropriate ones in order to keep the k-space

    image error constant. If more quadrature points are needed in some elements the user

    is warned. Then the user can either use the result knowing some points at the edge of

    k-space might have a slightly higher image error or he/she can run the function again

    with more quadrature points. This is a useful feature if the computation time even using

    the adaptive algorithm with all the needed quadratures is prohibitive.

    41

  • Chapter 3. Algorithm details and implementation 42

    3.2 FE-MRI software architecture

    The FE-MRI library is implemented as a set of python functions that can be used through

    a command line interface by typing a script in Linux or Unix. An example of a script

    that can be typed in the Unix command line after installing the FE-MRI library is given

    below:

    femritrisurfmeshkspace -ifile modifiedQuadSphere.vtu -dimensionality 2

    -fov 30 30 30 -fovcenter 0.0 0.0 0.0 -matrix 32 32 1 -useoptimal 1

    -error 0.001 -qorder 59 -sliceprofile quadratic -slicethickness 6

    --pipe femrikspacezeropadding -padding 64 64 64

    --pipe femrikspacetoimage --pipe vmtkimageviewer

    The above script reads an unstructured mesh of a sphere-like closed surface composed

    of tri6 elements. Then a 2D acquisition is performed with a field-of-view of 30, a 32 by

    32 k-space matrix and a quadratic slice profile in k-space. This is done using the optimal

    algorithm which can use a Gauss quadrature up to order 59 or points per dimension. Then

    the k-space is zero-padded and transformed back into an image which is displayed on the

    screen. Since the size of the shape is 40 by 40 by 40 units, the resulting image displays

    an aliasing error. In FE-MRI information from one function can be used into the next

    needed function without writing files by using the –pipe command. Some of the python

    functions call C++ functions to perform the needed computations. Both languages are

    used together so the user is able to both use fast and efficient C++ functions while being

    able to easily customize either the script or the python functions. The C++ functions can

    also be modified if needed however it is usually harder to modify that a python script.

    Both the python and C++ functions use data structures and visualization functions

    from the visualization toolkit or VTK1 and the vascular modelling toolkit or VMTK2

    1Visualization toolkit website: https://www.vtk.org/2Vascular modelling toolkit website: https://www.vmtk.org/

  • Chapter 3. Algorithm details and implementation 43

    libraries. The mesh used in the example script above is in VTK’s .vtu format. The

    FE-MRI acquisition algorithm and its related C++ functions are shown in Figure 3.1.

    The algorithm shown in Figure 3.1 is implemented in the vtkfemriTetVolNumer-

    icKSpaceGenerator C++ class for meshes composed of tet4 and tet10 elements. The

    algorithm can use a mesh with both elements present. The same is the case for a surface

    mesh when it is used by its C++ class vtkfemriTriSurfNumericKSpaceGenerator, it can

    contain both tri3 and tri6 elements. The class contains a set of data structures such as

    the k-space and functions that operate on them. This class is used to declare an object

    (an instance of a class) that then can call other objects declared using the classes vtkfem-

    riVolTetQuadratureOrderCalculator and vtkfemriGaussArbQuadrature which compute a

    list of the cycles per element as described in Chapter 2 and their corresponding num-

    ber of quadrature points to obtain a desired image error and the needed quadratures

    respectively.

    3.3 K-space acquisition algorithm

    The k-space signal acquisition in the flowchart in Figure 3.1 is done using Algorithm

    1. The k-space acquisition algorithm shown in Algorithm 1 starts by calculating the

    maximum cycles/element from all the k-space point and element combinations. This is

    done using a design table generated for a user specified error. The design table used for

    the optimal algorithm is a table that contains all the number of quadrature rules required

    and its corresponding cycles per element for a given error. Then it computes and stores

    all quadratures from orders 1 to the order needed for the maximum cycles/element to get

    the required error if the optimal algorithm will be used. The quadratures are calculated

    by the vtkfemriGaussArbQuadrature C++ class. Then the algorithm loops through all

    the k-space values needed and all the elements in the mesh and uses the given Gauss

    quadrature to compute the MRI signal. One thing to take into account in the algorithm

  • Chapter 3. Algorithm details and implementation 44

    Start

    Read mesh and MRIacquisition parameters

    Dimensionality?

    Find location andboundary of slices

    Set slice magnetization profile

    k-space aquisition

    More slices?

    Output k-space file and/orsend data to next python

    function

    Stop

    2D

    No, or 3D acquisition

    3D

    Yes

    Figure 3.1: Flowchart of the main FE-MRI algorithm

  • Chapter 3. Algorithm details and implementation 45

    is to make sure the magnetization and the Jacobian are being evaluated exactly. If

    the optimal algorithm is used, for k-space vectors with a low magnitude the number of

    quadrature points needed will be lower than the ones needed to compute the polynomial

    composed of the Jacobian multiplied by the magnetization. This is because even if the

    sinusoidal term is close to a straight line at low magnitudes of the k-space vector, needing

    only an order of 1 for example, one still wants enough quadrature order to compute the

    quadrature of the polynomial composed of the Jacobian multiplied by the magnetization.

    Algorithm 1 K-space acquisition algorithm

    if optimal algorithm thenmake design table relating cycles/element to quadrature order at desired image errorcompute Gauss quadratures from order 1 to specified order

    end iffor each needed k-space location k doset S(k) = 0for each element e doif optimal algorithm thencalculate cycles/element from k and eset Gauss quadrature to get desired error for calculated cycles/element usingdesign table

    elseset to user-defined Gauss quadrature

    end ifif order of Gauss quadrature < order of magnetization and Jacobian thenset Gauss quadrature to same order as magnetization and Jacobian

    end ifobtain Gauss points pobtain corresponding Gauss weights wset s(k) = 0for each p doevaluate s(k) = s(k) + w m(p) exp(k, p)

    end forevaluate S(k) = S(k) + f(T1, T2, ρ) s(k)

    end forend for

  • Chapter 3. Algorithm details and implementation 46

    3.4 Cycles per element quadrature table

    The algorithm to find the design table is shown in Algorithm 2.

    Algorithm 2 Design table algorithm

    compute max cycles/element (max cpe) for all k-space points and elementsfor 1 to max number of quadrature points specified, n doset point list, pointList(n) = nset start cpe to 0, cpe = 0set error to 0while error < error threshold doadd cpe interval to cpe, cpe = cpe+ dcpeget exact integral, Ie(cpe)obtain Gauss points pobtain corresponding Gauss weights wset Ic = 0for each p doevaluate Ic = Ic + w exp(cpe, p)

    end forevaluate error = |mag(Ie)−mag(Ic)|dim(e)

    end whileset cpe list, cpeList(n) = cpe

    end forif max cpe > cpeList(max n) thendisplay warning

    end if

    The algorithm is carried out in the vtkfemriQuadratureOrderCalculator C++ class.

    The algorithm first calculates the maximum cycles/element from the mesh and the MRI

    acquisition parameters. Then it loops over all the quadratures from 1 to a number

    specified by the user, storing each number in pointList. Within that loop it cycles

    thought increasing cycle/element values, until the error is above the user defined value.

    Then the cycles/element are stored in the corresponding entry in cpeList. The integral

    being computed is akin to a 1D MRI signal equation:

    I =

    ∫ 1

    0

    e−2πik (3.1)

  • Chapter 3. Algorithm details and implementation 47

    Where k is the cycles/element. The exact solution of its magnitude is:

    mag(Ie) =1

    2πk

    2− 2 cos(2πk) (3.2)

    The error is defined as follows:

    E = |mag(Ie)−mag(Ic)|dim(e) (3.3)

    Where Ic is an approximation of equation 3.1 calculated using Gauss quadrature and

    dim(e) is the dimensionality of the elements that the mesh is composed of. Note that

    the algorithm could have looped over the cycles/element first. However this could have

    caused some integrals computed with the quadrature rule to have a lower error simply

    because they are not being ‘sampled’ properly by the points; the Nyquist criteria is not

    followed with that loop ordering. With the current loop ordering the oscillatory integral

    gets more oscillatory while the number of quadrature points is kept the same, making

    sure that once the error is larger than the threshold there were also enough points to

    sample the function. The only difference between the algorithm when computing it for

    2D or 3D elements is that the 1D error is multiplied by 2 and 3 respectively. This is

    because the error of a product rule if using the same number of quadrature points in

    all dimensions can be estimated to first order as the error in one dimension times the

    number of dimensions of the product rule [22]. So the algorithm finds the error in one

    dimension and multiplies it by the number of dimensions to get an error estimate for the

    product rule which is then used to evaluate the element signal s(k) in Algorithm 1.

  • Chapter 3. Algorithm details and implementation 48

    3.5 Calculation of quadrature rules

    The quadrature rules are computed for the design table and the k-space acquisition cal-

    culations done in the C++ classes vtkfemriVolTetQuadratureOrderCalculator and vtk-

    femriTriSurfNumericKSpaceGenerator. Both classes make an object which is an instance

    of the class vtkfemriQuadratureOrderCalculator. The class can produce Gauss-Laguerre

    and Gauss-Jacobi quadratures. The class also computes all the product rules needed by

    all the classes. These quadratures have different weighting functions W (x) as shown in

    equation 2.1. This allows one to make a product rule that integrates the Jacobian from

    the Duffy transformation exactly. Therefore the only error comes from the transformed

    coordinates in the oscillatory term. That is why the derived error metrics in Chapter

    2 only take into account that term. The details of how the quadratures are computed

    are given in Appendix A. The author has added a simpler design table calculation, all

    the error metrics in Chapter 2 and a new class that can compute an arbitrary number of

    quadrature points and weights for the relevant Gauss quadratures.

    3.6 Slice profile simulation for volumetric meshes

    A method to simulate an arbitrary slice profile could be to adaptively subdivide the

    elements inside the slice to linearly approximate the slice. This is referred to in the

    finite element literature as h-refinement. Another approach is to use a higher polynomial

    order for the elements inside the slice to adaptively get a more accurate simulation.

    This method is referred to as p-refinement. While those methods will give an accurate

    solution, they are computationally expensive. A more practical method that does not

    require modifying the mesh or the elements is to weigh the s(k) signal in Algorithm 1

    after it is evaluated with a quadrature point, by using the location of the quadrature

    point and the required slice profile. For example, if one requires a quadratic slice profile

    along the z-direction, the weighting based on the location of the quadrature point within

  • Chapter 3. Algorithm details and implementation 49

    the slice is the following:

    m = 1− 4t2(pz − z0)2 (3.4)

    Where m is the weighting, t is the slice thickness, z0 is the slice origin location and

    pz is the z-component of the quadrature point. Everywhere else the weighting is 0. An

    ideal slice profile is simulated simply by making the weighting 1 within the slice and

    0 otherwise. Notice this method does not compromise the accuracy of the result given

    by the quadrature since the weighting convoluted with the magnetization distribution

    forms a new function that the quadrature is integrating. The ideal, quadratic and a

    trapezoidal slice simulation was already implemented in the C++ class vtkfemriTetVol-

    NumericKSpaceGenerator of the FE-MRI library.

    3.7 Additions and improvements to current FE-MRI

    functions

    The new error metrics in Chapter 2 have been implemented as C++ functions in the FE-

    MRI library. The code has been improved by removing the normalization and Jacobian

    calculation step for surface elements which are the same value for a given quadrature

    point and cancel each other. The algorithm to find the design table shown in Algorithm

    2 was rewritten using dynamically allocated C++ arrays instead of vtk arrays to make

    its execution faster and to aid in readability. Also instead of having a lookup table for

    the quadrature weights up to a certain number of points, the user can specify how many

    points can be used. The points are calculated from first principles once and stored at the

    beginning of the calculation. The details of the quadrature point and weights calculation

    can be found in Appendix A. If the program needs more points to calculate a certain

  • Chapter 3. Algorithm details and implementation 50

    k-space point there is a warning shown that recommends using more points and running

    the function again. If not the program uses the maximum number of points available.

    The memory to store the points is dynamically allocated and the points are stored in a

    contiguous memory block for faster access. This is done in C++ as follows:

    inline double** newQuadratureMatrix(int numPs)

    {

    double *matrix = new double[numPs*(numPs+1)*(2*numPs+1)/6];

    double **m = new double *[numPs];

    for(int i = 0; i < numPs; i++)

    {

    m[i] = &matrix[i*(i+1)*(2*i+1)/6];

    }

    return m;

    }

    The above function returns an array of pointers in m to the memory block named

    matrix, which contains all the points from 1 to numPs. The actual number of points is

    the sum of the square of the the 1 to numPs series since we are storing Gauss quadrature

    product rules which in this case are for 2D surface elements. The series needed and its

    result is shown below:

    n∑

    x=1

    x2 =n(n+ 1)(2n+ 1)

    6(3.5)

    This type of sum can be found using finite calculus [26]. Also the computation time

    has been cut in close to half by computing only half of k-space using the property [9]:

    se(−k) = s∗e(k) (3.6)

  • Chapter 3. Algorithm details and implementation 51

    Where s∗e(k) is the complex conjugate of the original MRI signal. Equation 3.6 applies

    for the Fourier transform of a real signal. However in actual MRI acquisition this con-

    straint cannot be exploited because of object motion and magnetic field inhomogeneities

    [9]. In the case of a simulated signal object motion can be added later to the simulated

    signal by adding a motion dependent phase to the computed k-space for example.

  • Chapter 4

    Results and discussion

    4.1 Introduction

    This chapter shows the k-space error maps and error histograms for meshes with each

    of the element types discussed in this thesis. This is done to confirm the efficacy of the

    algorithms in Chapter 3. Also the MRI acquisition of an anatomically realistic carotid

    artery was simulated to show the potential uses of the surface FE-MRI algorithm. The

    desired error threshold for all elements tested was 10−3.

    4.2 Validation of optimal algorithm for tri3 elements

    To validate the optimal algorithm for tri3 elements first the k-space error map and a

    histogram of the errors found in Table 4.1 for the adaptive algorithm were computed

    for a deformed unit cube. The deformed cube mesh and its simulated MRI image is

    shown in Table 4.2. The histogram shows a sharp peak at an error of 10−4. A sharp

    peak is expected since the oscillation due to the k-space value is constant within each

    element. The lower value for the error might be due to the error estimate for product

    rules given in [22] being a conservative estimate. The shape was chosen because it was

    easy to generate in the Paraview program starting with a meshed cube. Paraview is

    52

  • Chapter 4. Results and discussion 53

    based on VTK and can produce files that can be read by the FE-MRI library. Then

    the shape was made asymmetric by moving the nodes. This was done to highlight any

    possible errors in the algorithm which might not be evident by imaging a symmetric

    object. The imaging parameters are as follows: FOV = 2.5 x 2.5, image size = 32 x 32

    pixels. The magnetization value is 1 throughout. The exact image was computed using

    the exact algorithm available for surface linear elements in FE-MRI. As an example, the

    exact k-space using the analytic solution for the tri3 element