efficient ?total? extraction of perfluorooctanoate from polytetrafluoroethylene fluoropolymer
TRANSCRIPT
Efficient ‘‘total’’ extraction of perfluorooctanoate frompolytetrafluoroethylene fluoropolymer
Barbara S. Larsen,* Mary A. Kaiser, Miguel A. Botelho, Stanley F. Bachmura and L. William Buxton
Received 15th May 2006, Accepted 17th July 2006
First published as an Advance Article on the web 14th August 2006
DOI: 10.1039/b606801d
To determine the optimum conditions for the complete extraction of perfluorooctanoate (PFO)
from polytetrafluoroethylene fluoropolymers, sample preparation and pressurized solvent
extraction (PSE) conditions were investigated. Solvent extraction temperature, solvent residence
time, relaxation time between extractions, and the effects of heating before PSE showed that
methanol at 150 uC extraction temperature and a 12 min solvent residence time were the most
efficient conditions. Preheating the polymer before extraction at 150 uC for 24 h significantly
enhanced the quantity of PFO removed. Heating above 150 uC resulted in loss of PFO. PFO was
determined by liquid chromatography with tandem mass spectrometry.
Introduction
Ammonium perfluorooctanoate (APFO) is a processing aid
used in the production of many fluoropolymers. Reports have
shown low levels of perfluorooctanoate (PFO) and perfluoro-
octane sulfonate (PFOS) in human blood sera,1–3 with
additional reports showing PFO, PFOS, and other perfluori-
nated compounds in wildlife.4–6 The US Environmental
Protection Agency (US EPA) has identified both commercial7
and consumer use8 of fluoropolymers as possible contributors
to these levels. One specific commercial use is in analytical
instrument systems, due to the chemical inertness and stability
of fluoropolymers. However, background levels of perfluori-
nated materials have been associated with instrument proce-
dures and blanks.9 This complicates low-level detection of
perfluorinated carboxylic acids due to possible contamination
from the analytical systems and procedures used in the
analyses. Since fluorinated materials are found at low levels
in human sera and in wildlife, it is important to know if
exposure is possible from fluoropolymer processing aids
remaining from the polymerization process. Many fluorinated
polymers would need to be examined for this purpose,
therefore optimal analytical conditions must be ascertained.
This study is an extension of a previous study10 designed to
select the most efficient and complete extraction method and
solvent for quantifying total PFO content in polytetrafluoro-
ethylene (PTFE). In that study we showed that methanol was
the best solvent and that pressurized solvent extraction (PSE)
was the most efficient technique. In this study we show the
effect of sample pretreatment and sample ‘relaxation’ on the
efficiency of PFO extractability. Optimization of temperature,
residence time, and relaxation time of the sample for the PSE
with methanol as the solvent was performed on a PTFE
polymer for maximum extraction of PFO.
Experimental
Apparatus
PSE was performed using a pressurized solvent extractor, an
ASE1 series 200 fitted with 11 mL stainless steel cells, stainless
steel frits, cellulose extraction cap filters, and polyetherether-
ketone (PEEK) seals (Dionex, Sunnyvale, CA). 40 mL
I-CHEM glass collection vials were used (Chase Scientific
Glass Inc., Rockwood, TN). Sample evaporation was con-
ducted with Reacti-Therm III and Reacti-Vap III (Pierce
Chemical Co., Rockford, IL) evaporators. The oven of a
model 5890 gas chromatograph (Agilent Technologies,
Wilmington, DE) was used for sample heating. The oven
temperature was verified with a NIST traceable thermometer
(VWR International Inc., Bridgeport, NJ) to be within
¡0.8 uC.
Materials and reagents
Polytetrafluoroethylene fluoropolymer resin was obtained
from a commercial lot (average particle size approximately
500 mm). Analytical grade methanol and standard grade
Ottawa sand was purchased from EMD Chemicals, Inc.
(Darmstadt, Germany). Methanol was furthered purified by
passing it through a C18 reversed-phase liquid chromatography
column. Water was obtained from a Simplicity water
purification system (.18.2 MV cm) (Millipore, Billerica,
MA). Reagent-grade ammonium acetate was obtained from
Sigma–Aldrich (Milwaukee, WI). House nitrogen obtained
from a Zero Air nitrogen generation system backed up by boil-
off from a liquid nitrogen tank was used for evaporation.
Pentadecafluorooctanoic acid (PFOA) and APFO used for
spiking/recovery was purchased from Oakwood Products, Inc.
(West Columbia, SC). A dual 13C enriched PFOA internal
standard was synthesized in-house. Standards for the seven-
point external calibration curve were prepared by aqueous
dilution of a 1000 ppb (mg L21) standard solution prepared
in methanol to concentrations of 0.5, 1, 5, 10, 25, 50, and
100 ppb.
E I du Pont de Nemours and Company, E228, PO Box 80228,Wilmington, DE, 19880-0228, USA.E-mail: [email protected]; Fax: +1 302 695 1351;Tel: +1 302 695 4876
PAPER www.rsc.org/analyst | The Analyst
This journal is � The Royal Society of Chemistry 2006 Analyst, 2006, 131, 1105–1108 | 1105
Publ
ishe
d on
14
Aug
ust 2
006.
Dow
nloa
ded
by U
nive
rsity
of
Prin
ce E
dwar
d Is
land
on
24/1
0/20
14 2
3:28
:36.
View Article Online / Journal Homepage / Table of Contents for this issue
House nitrogen, from time-to-time, was found to contain
quantifiable levels of perfluoroheptanoate, PFO, perfluorono-
nanoate, perfluorodecanoate, perfluoroundecanoate, and
perfluorododecanoate that contributed to background levels
in evaporated samples. Blank water did not show any
quantifiable levels of these anions.
Pressurized solvent extraction procedure
All cells were preconditioned by filling with Ottawa sand, 3 mm
from the top, the top secured, and the cell loaded into the
instrument. The instrument solvent reservoir was filled with
methanol and the lines were rinsed four times prior to use. The
preconditioning cycle parameters were 1600 psi (11.0 kPa),
150 uC, 7 min heating duration, 100% volume flush, and 240 s
purge for 1 cycle. The solvent extract was discarded and a fresh
40 mL collection vial was set in place. Each collection vial was
cleaned before use by soaking twice in water followed by a
methanol soak, and then allowed to air-dry. In each extraction
experiment five preconditioned cells with Ottawa sand were
used, three containing polymer (1.5 g), a recovery check
standard of PFOA and a blank. After extraction each
collection vessel was evaporated to dryness with N2 gas
purging at y45 uC, and PFO determined by LC/MS/MS.10
The thermal stability experiments were performed by
inserting a standard quantity of PFOA or APFO in a
preconditioned cell and heating the cell in the GC oven for
24 hours at 150, 175 or 200 uC. The cells were then extracted
using the PSE. Optimization of the temperature parameter
consisted of three experiments: one at 65 uC, one at 125 uC and
one at 150 uC. The parameters for the extractions were 1600 psi
(11.0 kPa), 12 min heating duration, 100% volume flush, 240 s
purge, and 4 cycles. Using the optimum temperature of 150 uC,
additional residence time experiments were run using duration
settings of 4 min and 30 min.
The sequential extraction of the PTFE polymer (1.5 g) was
performed using the optimum parameter of 150 uC 1600 psi
(11.0 kPa), 12 min heating duration, 100% volume flush, 240 s
purge, and 4 cycles. The extraction procedure was repeated on
each cell a total of six times with no rest time in between. The
relaxation experiment was performed similar to the sequential
experiment with the addition of a 24 h resting time with the
cells maintained at room temperature or 150 uC in between
extractions. The final experiment heated the cells at 150 uC for
24 or 48 hours prior to initial extraction. The resting time at
150 uC for 24 and 48 hours was repeated for six extractions.
Each data point represents an average of four extraction
experiments.
Analytical method
The PFO extract taken after solvent evaporation was
reconstituted by adding 1 mL of methanol to the collection
vial and shaking for 30 min on a wrist-action shaker. The
methanol was then transferred to a 5 mL Class A volumetric
flask and brought to volume with the LC mobile phase A
described below. The dual 13C-enriched standard (final
concentration 50 ppb) was added as an internal standard to
all of the reconstituted samples. The concentration of PFO was
determined using LC coupled with negative ion electrospray
tandem mass spectrometry (LC/MS/MS) (Micromass Quatro
Ultima, Beverly, MA).
Analytes were separated using a model 1100 liquid
chromatograph (Agilent Technologies, Wilmington, DE)
modified with low dead-volume internal tubing. A guard
column, Hypersil C18 2 mm 6 50 mm (Thermo Keystone,
Bellefonte, PA), was installed between the mixer and the
autoinjector. 25 mL of the reconstituted extract was injected
onto a Hypersil ODS 2.1 mm 6 200 mm (Thermo Keystone,
Bellefonte, PA) at a flow rate of 0.3 mL min21 and maintained
at 60 uC. Duplicate injections were made for all samples.
The initial gradient mobile-phase composition was 15%
mobile phase B, where mobile phase A was 2 mM ammonium
acetate–5% methanol and mobile phase B was 100% methanol.
A linear gradient was used from 15%–67% B over sixteen
minutes. The conditions were returned to 15% B for four
additional minutes. Typical elution time for PFOA was
approximately 16.5 min. A typical chromatogram for these
conditions is shown in reference 10. PFO was observed in the
negative ion mode as an anion at 413 amu (CF3(CF2)6 COO2).
The internal standard was observed at 415 amu
(CF3(CF2)513CF2
13COO2).
Selected ion monitoring for the transition of 413 A 369 (loss
of CO2) for the analyte and 415 A 370 (loss of 13CO2) for the
internal standard was used for quantitative analysis. Samples
that fell outside of the calibration range were rediluted and
reanalyzed. A seven-point external calibration curve was
prepared (not including zero) with each calibrant run in
duplicate, bracketing the samples. A typical calibration curve
consisted of all fourteen-calibration points. The acceptance
criterion for the calibration curve requires a correlation
coefficient (R2) ¢ 0.985, with 1/X weighting.
A methanol blank was run after each 100 ppb standard. The
limit of quantitation (LOQ) was defined as 0.5 ppb, the
concentration of the lowest calibration standard. An accep-
tance criterion was set so that the area of the 0.5 ppb standard
must be at least five times greater than any background peak
obtained in the solvent blank. No quantifiable area was
observed in the solvent blank.
Results and discussion
In order to optimize the extraction of PFO from fluoropoly-
mer, it is necessary to understand the thermal stability of
APFO and PFOA at elevated temperatures. Recovery results
from PFOA and APFO spike recovery experiments show that
150 uC is the maximum temperature at which PFO can be
reproducibly recovered (Table 1). The results from 175 uC and
200 uC show recoveries that fall outside the acceptable
recovery range of 70 to 130%. This observation of 150 uC as
Table 1 PFOA/APFO 24 h thermal stability scouting study
StandardConcentration(ppb)
% Recovery150 uC
% Recovery175 uC
% Recovery200 uC
APFO 300 94.5 2 11APFO 30 97.2 0 86PFOA 300 98.5 3 4PFOA 30 78 19 63
1106 | Analyst, 2006, 131, 1105–1108 This journal is � The Royal Society of Chemistry 2006
Publ
ishe
d on
14
Aug
ust 2
006.
Dow
nloa
ded
by U
nive
rsity
of
Prin
ce E
dwar
d Is
land
on
24/1
0/20
14 2
3:28
:36.
View Article Online
the maximum operating temperature for extraction corre-
sponds with data from recent NMR thermolysis studies11,12
that demonstrated significant decomposition of PFOA and
APFO at elevated temperatures. This observation also agrees
with recent studies of commercial PTFE-coated nonstick fry
pans. Fry pans are thermally treated at elevated temperature
during the manufacturing process. In a surface extraction
study no quantifiable PFO was detected.13 In a different total
extraction study of nonspecified PTFE coating that was
physically removed from the pan surface, a very low-level of
PFO was found.14
Determining efficient extraction parameters for PSE
requires selection of optimal temperature and solvent residence
time. In a previous study10 methanol was shown to be a
preferred solvent for efficient extraction of this fluoropolymer.
Table 2 shows the effects of solvent temperature using a
12 minute solvent residence time. Optimization results for
residence time extractions are shown in Table 3. Together these
experiments indicate that the optimal extraction parameters
are 150 uC with a 12 min solvent residence time. These
conditions were used for all ensuing extractions.
The purpose of this study is to determine the total mass of
PFO in a sample of PTFE. In order to accomplish this goal, a
series of sequential extractions was performed. A limiting
value of 5% of the initial mass extracted was set to determine
the end point or last extraction. The results for the sequential
extraction series are displayed graphically in Fig. 1 as the mass
of PFO extracted from PTFE with each subsequent extraction.
The extraction curve for sequential extractions shows an
exponential removal of PFO under these conditions. Note that
two blank samples and two PEEK seal samples demonstrated
that there was no measurable PFO background in the
extraction system, indicating that the low level measured in
the final extraction was most likely coming from the polymer.
In order to determine if relaxation or thermal treatment of
the polymer might help to expedite PFO removal, a series of
relaxation and thermal treatment experiments were performed.
These experiments were designed to demonstrate if relaxation
or thermal treatment of the polymer would help PFO deep
inside the particle migrate towards the surface, making it more
readily available for extraction. Summaries of the room
temperature and 150 uC relaxation/extraction experiments
are displayed in Fig. 2. A comparison of the sequential
extraction (Fig. 1) with the room temperature, 24 h relaxation
experiment (shown in Fig. 2) indicates that the rest period
alone does not aid the extraction. A comparison of the room
temperature relaxation with the 24 h relaxation at 150 uCexperiment shows that the 150 uC relaxation period gives a
much greater mass on subsequent extractions than at
room temperature. This experiment indicates that the 24 h
relaxation period at 150 uC makes the PFO in this polymer
more accessible for extraction. In order to determine if
thermal treatment or relaxation with thermal treatment was
required to optimize PFO extraction, the polymer was
preheated for 24 h at 150 uC (Fig. 3). These results indicate
that thermal treatment alone greatly increased the quantity of
PFO extracted. Preheating for 48 h rather than 24 h showed no
significant enhancement in the mass extracted.
Table 2 Temperature optimization extraction study of PFO fromPTFE polymer
Temperature/uCConc. PFO inPTFE (ppb)a
Mass PFOextracted/ng
Check standard% recoveryb
65 32.5 ¡ 1.6 48.8 ¡ 2.2 98 ¡ 4.8125 67.2 ¡ 4.0 101 ¡ 6.4 107 ¡ 5.9150 119 ¡ 7.6 179 ¡ 12 118 ¡ 3.7a Average of three sample weights. b Average of five check standardsamples.
Table 3 Residence time optimization extraction study of PFO fromPTFE polymer
Residence time/minConc. PFO inPTFE (ppb)a
Mass PFOextracted/ng
Check standard% recovery
4 71.5 ¡ 5.9 107 ¡ 9.6 99.512 119 ¡ 7.6 179 ¡ 12 11830 105 ¡ 6.1 158 ¡ 9.3 125a Average of three sample weights.
Fig. 1 Sequential extraction of PFO from PTFE polymer.
Fig. 2 Comparison of PFO extraction of PTFE polymer with 24 h
relaxation time at room temperature and 150 uC.
This journal is � The Royal Society of Chemistry 2006 Analyst, 2006, 131, 1105–1108 | 1107
Publ
ishe
d on
14
Aug
ust 2
006.
Dow
nloa
ded
by U
nive
rsity
of
Prin
ce E
dwar
d Is
land
on
24/1
0/20
14 2
3:28
:36.
View Article Online
Conclusions
An accurate, reproducible optimized PSE method to determine
quantitatively PFO in PTFE polymer was demonstrated.
Optimum conditions used methanol, a 150 uC extraction
temperature, and a 12 min solvent residence time. Preheating
the polymer to 150 uC for 24 h prior to PSE greatly enhanced
the quantity of PFO extracted. Heating above 150 uC in this
experiment resulted in the loss of PFO. These optimized
conditions will be used for subsequent determination of total
PFO from similar fluoropolymers.
References
1 K. J. Hansen, L. A. Clemen, M. E. Ellefson and H. O. Johnson,Environ. Sci. Technol., 2001, 35, 766.
2 Z. Kuklenyik, J. A. Reich, J. S. Tully, L. L. Needham andA. M. Calafat, Environ. Sci. Technol., 2004, 38, 3698.
3 K. Inoue, F. Okada, R. Ito, S. Kato, S. Sasaki, S. Nakajima,A. Uno, Y. Saijo, F. Sata, Y. Yoshimura, R. Kishi andH. Nakazawa, Environ. Health Perspect., 2004, 112, 1204.
4 M. M. Schultz, D. F. Barofsky and J. A. Field, Environ. Eng. Sci.,2003, 20, 487.
5 K. Kannan, S. Corsolini, J. Falandysz, G. Oehme, S. Focardi andJ. P. Giesy, Environ. Sci. Technol., 2002, 36, 3210.
6 J. W. Martin, M. M. Smithwick, B. M. Braune, P. F. Hoekstra,D. C. G. Muir and S. A. Mabury, Environ. Sci. Technol., 2004, 38,373.
7 US EPA, FMG [Fluoropolymers Manufacturers Group] Letter ofIntent, Office of Pollution Prevention and Toxics; EPA-HQ-OPPT-2003-0012-0012, 2003.
8 US EPA, 3 M Letter of Intent, Office of Pollution Prevention andToxics, EPA-HQ-OPPT-2003-0012-0007, 2003.
9 N. Yamashita, K. Kannan, S. Taniyasu, Y. Horii, T. Okazawa,G. Petrick and T. Gamo, Environ. Sci. Technol., 2004, 38,5522.
10 B. S. Larsen, M. A. Kaiser, M. Botelho, G. R. Wooler andL. W. Buxton, Analyst, 2005, 130, 59.
11 P. J. Krusic, A. A. Marchione and D. C. Roe, J. Fluorine Chem.,2005, 126, 1510.
12 P. J. Krusic and D. C. Roe, Anal. Chem., 2004, 76, 3800.13 C. R. Powley, M. J. Michalczyk, M. A. Kaiser and L. W. Buxton,
Analyst, 2005, 130, 1299.14 T. H. Begley, K. White, P. Honigfort, M. L. Twaroski,
R. Neches and R. A. Walker, Food Addit. Contam., 2005, 22,1023.
Fig. 3 Comparison of PFO extraction from thermally treated PTFE
polymer.
1108 | Analyst, 2006, 131, 1105–1108 This journal is � The Royal Society of Chemistry 2006
Publ
ishe
d on
14
Aug
ust 2
006.
Dow
nloa
ded
by U
nive
rsity
of
Prin
ce E
dwar
d Is
land
on
24/1
0/20
14 2
3:28
:36.
View Article Online