eddyliner px user manual

Upload: carlos

Post on 03-Jun-2018

253 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/11/2019 Eddyliner Px User Manual

    1/146

    Manual eddyliner

    P2eddyliner

    P3

    and eddylinerP4

    April 16 2004 (v 30e)

    ibg Doc: 0108131e

    eddyliner P

    F2

    START

    F1

    F3

    F4 ESC

    QUIT

    F?

    ?

    RESET

    PRINT

    EXTERN

    SETUP

  • 8/11/2019 Eddyliner Px User Manual

    2/146

    Software and Manual

    Copyright 1994-2004 by ibg Prfcomputer GmbH,Pretzfelder Strae 27, D-91320 Ebermannstadt

    The copyright claimed covers all forms and types of materials and information falling

    under the copyright, which at the present time are legally admitted or granted in the

    following, including material generated without restriction from the software

    programme appearing on the screen, such as symbols, screen displays, etc.

    Exclusion of responsibility

    Great care was taken in compiling texts and figures. Nevertheless, errors cannot be

    ruled out completely. Editors and authors cannot assume any legal responsibility or

    liability for erroneous information and resulting consequences. The editor will readily

    accept any proposals for improvement and hints at errors.

    The unit delivered incorporates the latest state of the art. The manufacturer declines

    any legal responsibility and liability for inspections carried out with the apparatus

    and for resulting consequences. Successful implementation of an inspection

    presupposes, in addition to the control of the unit setting, also a knowledge of the

    product to be tested, its material properties, of possible errors and their underlying

    causes. If these prerequisites are not satisfied, the result will be questionable. In

    order to avoid faulty operation and other errors, training in the manufacturers factory

    or by an independent training organisation is recommended.

    eddylineris a registered trademark of ibg Prfcomputer GmbH.

    EPSON is a registered trademark of Seiko Epson Corporation.

    HP and LaserJet are registered trademarks of Hewlett-Packard Corporation.

    NEC is a registered trademark of Nippon Electric Corporation.

    Trademarks are used without warranting free usability.

  • 8/11/2019 Eddyliner Px User Manual

    3/146

    ibg Headquarter:

    ibg Prfcomputer GmbH

    Pretzfelder Strae 27D-91320 EbermannstadtTel: +49-9194-7384-0Fax: +49-9194-7384-10E-mail: [email protected]

    ibg International:

    ibg NDT Systems Corp.20793 Farmington RoadFarmington Hills, MI 48336Tel: +1-248-478 9490Fax: +1-248-478 9491E-mail: [email protected]

    ibg SWISS AGGalgenried 6CH-6370 Stans

    Tel: +41-41-612 2650Fax: +41-41-612 2651E-mail: [email protected]

    ibg UK Ltd.33 Park View RoadSutton ColdfieldGB-West Midlands B74 4PRTel/Fax: +44-121-352 1188E-mail: [email protected]

  • 8/11/2019 Eddyliner Px User Manual

    4/146

  • 8/11/2019 Eddyliner Px User Manual

    5/146

    Manual eddylinerPx Content

    ibg Doc: 0108131e, Apr 16, 2004

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 0-3

    Content

    Exclusion of responsibility 1-0

    1. Introduct ion .................................................................................................... 1-1

    1.1. The principle of eddy current testing....................................................................1-1

    1.2. Possibilities of eddy current inspection................................................................1-3

    2. General hints .................................................................................................. 2-1

    2.1. Selection of the test coils.....................................................................................2-1

    2.2. Creation of optimum test conditions ....................................................................2-3

    2.3. Wiring...................................................................................................................2-52.4. Elements of the front panel..................................................................................2-6

    2.4.1. Soft keys ....................................................................................................2-9

    2.4.2. Information window..................................................................................2-10

    2.4.3. Help function............................................................................................2-10

    2.5. Test sequence ...................................................................................................2-11

    3. Short operating instructions ......................................................................... 3-1

    4. Operating modes and funct ions ................................................................... 4-1

    4.1. MAIN MENU ........................................................................................................4-14.2. SETUP.................................................................................................................4-2

    4.2.1. COMMUNICATION SETUP.......................................................................4-5

    4.2.1.1. Self-start ........................................................................................4-8

    4.2.1.2. Volume ..........................................................................................4-8

    4.2.1.3. V24-interface.................................................................................4-9

    4.2.1.4. The printer .....................................................................................4-9

    4.2.1.5. Warning .......................................................................................4-10

    4.2.1.6. Sorting data recording .................................................................4-10

    4.2.1.7. The sort impulse..........................................................................4-11

    4.2.1.8. PIP-function (Start of test) ...........................................................4-114.2.1.9. Allocation of coils.........................................................................4-13

    4.2.1.10. END-impulse .............................................................................4-13

    4.2.1.11. Generation of the sum sort outputs OKS/NGS..........................4-14

    4.2.1.12. Coil monitoring...........................................................................4-15

    4.2.2. TEST-SETUP ..........................................................................................4-16

    4.2.2.1. Number of setup..........................................................................4-21

    4.2.2.2. Number of channel ......................................................................4-22

    4.2.2.3. Set name .....................................................................................4-22

    4.2.2.4. Test frequency.............................................................................4-23

    4.2.2.5. Frequency set proposal...............................................................4-24

    4.2.2.6. Size of the tolerance zone...........................................................4-24

  • 8/11/2019 Eddyliner Px User Manual

    6/146

    Content Manual eddylinerPx

    Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    0-4 D-91320 Ebermannstadt Germany

    4.2.2.7. Settling delay .............................................................................. 4-25

    4.2.2.8. Test time and total test time........................................................ 4-26

    4.2.2.9. Interference suppression ............................................................ 4-27

    4.2.2.10. AUTOPIP-Frequency................................................................ 4-27

    4.2.2.11. AUTOPIP threshold value......................................................... 4-28

    4.2.3. Working with the External Setup Module (ESM)..................................... 4-29

    4.2.3.1. Storage of the External Setup Module (ESM)............................. 4-30

    4.2.3.2. Loading from the External Setup Module (ESM) ........................ 4-30

    4.3. MODULE TEST AND OPERATION STATISTICS............................................ 4-31

    4.3.1. MODULE TEST 1.................................................................................... 4-31

    4.3.2. MODULE TEST 2.................................................................................... 4-33

    4.3.3. MODULE TEST 3.................................................................................... 4-34

    4.3.4. OPERATION STATISTICS ..................................................................... 4-35

    4.4. MATERIAL DATA.............................................................................................. 4-354.4.1. RECORDING O.K. PARTS..................................................................... 4-37

    4.4.2. TOLERANCE ZONE ............................................................................... 4-40

    4.4.3. TOLERANCE ZONE CORRECTION ...................................................... 4-42

    4.4.4. DYNAMIC CONTROL............................................................................. 4-44

    4.5. SORTING.......................................................................................................... 4-47

    4.5.1. Bargraph Display..................................................................................... 4-47

    4.5.2. Display of ellipse ..................................................................................... 4-53

    4.5.3. Display of sorting data ............................................................................ 4-54

    4.6. Printing.............................................................................................................. 4-544.6.1. Screen printout........................................................................................ 4-55

    4.6.2. Printout of all tolerance zones................................................................. 4-55

    4.6.3. Protocol printout...................................................................................... 4-55

    5. Data transfer via V24 interface ....................................................................5-56

    5.1. Interconnection with external computer ............................................................ 5-56

    5.2. Format of transferable data and commands..................................................... 5-56

    5.3. Test data set ..................................................................................................... 5-56

    5.4. Interruption data set .......................................................................................... 5-56

    5.5. Commands possible with data transfer running................................................ 5-56

    5.5.1. Request parts counter set ('z')................................................................ 5-56

    5.5.2. Reset of parts counter ('r') ..................................................................... 5-56

    5.5.3. Interruption of data transfer ('x' or 'y') .................................................... 5-56

    5.5.4. Admitting transfer of test data ('v' or 'w')................................................. 5-56

    5.6. Commands only possible when data transfer is interrupted ............................. 5-56

    5.6.1. Common features of multi-byte commands ............................................ 5-56

    5.6.2. Sending of material data of O.K. parts ('tm') .......................................... 5-56

    5.6.3. Loading of material data of O.K. parts ('lm') .......................................... 5-56

  • 8/11/2019 Eddyliner Px User Manual

    7/146

    Manual eddylinerPx Content

    ibg Doc: 0108131e, Apr 16, 2004

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 0-5

    5.6.4. Sending of instrument setting incl. tolzones ('te')...................................5-56

    5.6.5. Loading of instrument setting incl. tolzones ('le')...................................5-56

    5.6.6. Configuration of data sets ('lv')..............................................................5-56

    5.6.7. Remote control by simulation of keyboard input......................................5-565.7. BASIC - program example.................................................................................5-56

    6. Connection w ith automatic sor t systems................................................... 6-56

    6.1. Temporary course of measurement...................................................................6-56

    6.1.1. Start of measurement (PIP) .....................................................................6-56

    6.1.1.1. Triggering by external PIPx-signal...............................................6-56

    6.1.1.2. Continuous measurement (KONTI-PIP) ......................................6-56

    6.1.1.3. Subsequent testing (AFTER-PIP)................................................6-56

    6.1.1.4. Automatic Triggering (AUTOM-PIP) ............................................6-56

    6.1.1.5. Possible errors working with AUTOPIP.......................................6-566.1.2. Settling delay ...........................................................................................6-56

    6.1.3. Total test time and MPRG .......................................................................6-56

    6.1.4. Sort decision............................................................................................6-56

    6.2. Optointerface (POPX)........................................................................................6-56

    6.2.1. Protective wiring of inputs and outputs....................................................6-56

    6.2.2. Signal levels of inputs and outputs ..........................................................6-56

    6.2.3. Inputs "PIPx" (Piece in Probe) .................................................................6-56

    6.2.4. Input "DPIP" (disable PIP) .......................................................................6-56

    6.2.5. Output "MPRG" (Measuring in Progress) ................................................6-566.2.6. Sort outputs "OKx", "OKS" and "NGS" ....................................................6-56

    6.2.6.1. Sort impulse length in "milliseconds"...........................................6-56

    6.2.6.2. Sort impulse length "synchronous"..............................................6-56

    6.2.7. Outputs "ENDx" and "END" (End of Measurement) ................................6-56

    6.2.8. Output "SFTY" (SAFETY, watch dog)......................................................6-56

    6.2.9. Output "ALRM" (ALARM).........................................................................6-56

    6.2.10. Input "ALRS" (ALARM-RESET).............................................................6-56

    6.2.11. Input "DABGL" (disable self-calibration) ................................................6-56

    6.3. Connection Distributor (PEXT) ..........................................................................6-56

    6.4. Integrated voltage supply 24 Vdc, max 1.6 A ....................................................6-56

    6.5. Special functions................................................................................................6-56

    6.5.1. Module test ..............................................................................................6-56

    6.5.2. Self-start ..................................................................................................6-56

    6.5.3. Alarm after xxx rejected parts in series....................................................6-56

    6.5.4. Automatic Self-Calibration .......................................................................6-56

    7. Remote Control with SET-CONTROL plug ................................................. 7-56

    7.1. Signals...............................................................................................................7-56

    7.2. Timing of switching ............................................................................................7-56

  • 8/11/2019 Eddyliner Px User Manual

    8/146

    Content Manual eddylinerPx

    Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    0-6 D-91320 Ebermannstadt Germany

    7.2.1. Setup switching (SETUP)........................................................................ 7-56

    7.2.2. Menu switching (MENU) ......................................................................... 7-56

    7.2.3. Keyboard lock (KEY)............................................................................... 7-56

    7.3. SET-CONTROL plug......................................................................................... 7-567.4. Special functions............................................................................................... 7-56

    7.4.1. Module test ............................................................................................. 7-56

    10. Appendix ...................................................................................................10-56

    10.1. Automatic testing, course of Programme...................................................... 10-56

    10.2. Time diagram................................................................................................ 10-56

    10.3. Electronic diagram of Interfaces ................................................................... 10-56

    10.4. Pin Configuration of POPX and PEXT .......................................................... 10-56

    10.5. Pin Configuration SET-CONTROL plug........................................................ 10-56

    10.6. Wiring examples optointerface...................................................................... 10-56

    10.6.1. Interface proximity switch and internal power supply.......................... 10-56

    10.6.2. Interface to universal sorting device ................................................... 10-56

    10.6.3. PLC-controlled testing using one coil at four test positions ................ 10-56

    10.6.4. PLC-controlled testing with recognition of direction ............................ 10-56

    10.7. Wiring examples SET-CONTROL plug......................................................... 10-56

    10.7.1. Setup-Switching of 15 Setups using a PLC ........................................ 10-56

    10.8. V24-Wiring .................................................................................................... 10-56

    10.9. Example of test protocol ............................................................................... 10-5610.10. Electrical rated values................................................................................. 10-56

  • 8/11/2019 Eddyliner Px User Manual

    9/146

    Manual eddylinerPx Introduction

    ibg Doc: 0108131e, Apr 16, 2004

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 1-1

    1. Introduction

    Congratulation on the decision to use the eddyliner

    Px in your factory. You haveacquired the most advanced eddy current testing device working according to the

    PREVENTIVE MULTI-FREQUENCY method. Successful implementation of a test

    presupposes, in addition to a full command of the instrument settings, also a solid

    knowledge of the product to be tested, its material properties, of the errors likely to

    occur and their underlying causes.

    The operator must be conversant with the limitations and restrictions in error

    detectability as well as with the efficiency of the equipment used and with its proper

    function. It is the proper functioning that needs to be checked prior to every test and

    periodically during operation in order to warrant reliable test results. If these

    requirements are not satisfied, the result will be of questionable value.

    Should you have any further questions or need our advice for an optimum solution

    of your testing problem, please do not hesitate to contact us. Our after-sales service

    is ready to help you at any time.

    The chapter "3. Short operating instructions" as from page 3-1 is mainly intended to

    save the users who are not constantly working with the eddylinerPx the

    troublesome study of a detailed user manual.

    As a new user of the eddylinerPxyou should simply try to perform a test with this

    unit under the guidance of the "3. Short operating instructions" before going through

    the detailed user manual and the fine adjustment of all parameters. You will be

    surprised how quickly you will learn to handle the eddylinerPx.

    Please read these operating instructions carefully in order to avoid operating errors.

    We have tried to keep the instructions as short as possible in an effort not to burden

    you with unnecessary information.

    1.1. The principle of eddy current testing

    When alternating current is sent through a transmitter coil, a magnetic field builds up

    at the coil. If a conductive test part approaches the current-carrying coil, an eddycurrent is induced in the test part. This phenomenon is schematically illustrated in

    Fig. 1.1.

  • 8/11/2019 Eddyliner Px User Manual

    10/146

    Introduction Manual eddylinerPx

    The principle of eddy current testing Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    1-2 D-91320 Ebermannstadt Germany

    u m u s

    si

    u

    u

    2

    1

    wi

    Fig. 1.1

    The magnitude of the eddy current depends on the chemical composition (alloy) and

    on the structural condition (e.g. hardened). By changing the eddy current frequency,

    an electromagnetic "thumbprint" of the structural state is obtained in the coil system.

    This can be picked up by the instrument through the integrated receiver coil and

    evaluated.

    An alteration in the structural state, in the alloy and the like attributable to a mix-up

    of material, a batch mix or faulty thermal treatment causes also the eddy current

    signal to change. The eddylinerPx testing device recognizes such changes and

    indicates deviations. In an analysis, the differences of the transmitter are comparedwith those of the receiver, evaluated and the result is classified.

    The eddylinerPx operates according to the method of PREVENTIVE MULTI-

    FREQUENCY TESTING (PMFT) developed by ibg. The merit of PREVENTIVE

    MULTI-FREQUENCY TESTING is seen in the high probability of detecting even

    unknown errors due to the wide frequency band.

    In contrast to other procedures, ibg uses low field strength values. This means that

    only low currents flow through the coils and that an insignificant permeability

    develops in ferromagnetic materials.

    This method affords the following advantages:

    It operates at the higher penetration depth of the eddy current so that the

    influence exerted by the marginal layers is reduced.

    The reproducibility of the measurements is guaranteed because no magnetic

    alterations develop in the structure of the test part.

    Improved testing stability because no heat builds up in the test part, the

    instrument and in the coil.

  • 8/11/2019 Eddyliner Px User Manual

    11/146

    Manual eddylinerPx Introduction

    ibg Doc: 0108131e, Apr 16, 2004 Possibilities of eddy current inspection

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 1-3

    The outstanding properties of PREVENTIVE MULTI-FREQUENCY TESTING

    (PMFT) are the following:

    Maximum testing reliability - even unknown errors are identified without fail.

    Only good parts must be presented to the instrument.

    RECORDING O.K. PARTS is extremely simple and effected in a very short time.

    1.2. Possibilities of eddy current inspection

    Eddy current inspection can be applied to any type of material being conductive

    either electrically or magnetically. This holds good for all metallic materials.

    The eddyliner

    Pxtakes care of all the fields of application summarily mentioned inTable 1.1.

    Type of parts Parameters to be tested for

    Forged pieces

    Bearing components

    VA-tubes

    Screws

    Brake discs

    Power transmission elements

    Components of belt locks

    Linear guides

    Steel balls

    heat treatment and material mix-up

    hardness and material mix-up

    material mix-up

    heat treatment

    chill depth

    purity and core strength

    hardness and texture

    purity and grinding burn

    material mix-up

    Table 1.1

    Please note that the eddy current inspection procedure is not an absolute measuring

    method but rather a comparative method. The hardness of parts can be measured

    with the eddylinerPx inspection device to within 2 - 3 HRC without any problem.

    However, as it is the case with all eddy current tests, the inspection is based on

    YES/NO answers. The eddy current testing instrument does recognize deviationsfrom existing tolerance zones, but cannot reveal the causes responsible for them. In

    order to obtain precise information on why the test part was classified as NotGood,

    yet other testing procedures such as hardness test and metallurgraphic studies must

    be employed to confirm the findings of the eddy current device.

  • 8/11/2019 Eddyliner Px User Manual

    12/146

  • 8/11/2019 Eddyliner Px User Manual

    13/146

    Manual eddylinerPx General hints

    ibg Doc: 0108131e, Apr 16, 2004

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 2-1

    2. General hints

    The eddyliner

    Pxeddy current inspection device is microprocessor-controlled. Theentire control software is configured hierarchically, so the submenus selectable

    result automatically from each main menu. The eddylinerPxis operated via a so-

    called soft key control system which enables the user to handle the device with a

    restricted number of keys.

    These operating instructions have been arranged so that each step of operation is

    explained in chronological order. All you need to do is to execute the instructions

    step by step as indicated. While proceeding in this manner, you are guided through

    the complete software.

    Note

    The [? ]-is auxiliaryand serves to indicate a context-related help at any time.

    This means that the help needed is offered at any point of the

    software.

    The key combinations to be actuated by the user are written in

    bold capital letters throughout the User Manual.

    Should you nevertheless experience any problem with operation,

    please contact us under the following telephone number:

    ibg Prfcomputer GmbH, Tel. +49-91 94-73 84-0

    2.1. Selection of the test coils

    Selection of the proper test coils and probes is of decisive importance for the test

    result. In addition to a set of standard coils and probes, special coils adapted to the

    job are available. The type of job and the shape of the test part determine the type

    of coils. Here the following criteria should be considered:

    What is the job like?

    e.g. measurement of the case depth of shafts

    Which coil should be used?

    The suitable coil or probe can be taken from Table 2.1.

  • 8/11/2019 Eddyliner Px User Manual

    14/146

    General hints Manual eddylinerPx

    Selection of the test coils Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    2-2 D-91320 Ebermannstadt Germany

    Field of application Frequency range

    P

    robe

    Encir

    clingcoil

    Encir

    clingcoil

    sh

    ielded

    Rectangularcoil

    Flatcoil

    Material mix:

    ferritic material

    austenitic material

    non-ferrous metals

    hard metals

    25 Hz - 100 kHz

    100 Hz - 100 kHz

    100 Hz - 100 kHz

    50 Hz - 100 kHz

    +++++++

    ++++

    ++

    +

    +

    ++++

    ++

    +++++++

    Heat treatment:

    tempering

    surface hardness

    case depth of shafts

    case depth (carburized)

    carburization

    decarburization

    25 Hz - 25 kHz

    100 Hz - 25 kHz

    5 Hz - 400 Hz

    5 Hz - 4 kHz

    5 Hz - 25 kHz

    10 kHz - 100 kHz

    +++

    +++

    ++++++

    +++

    ++++

    +++

    ++++++

    +++

    +++

    +++

    Cast iron:

    nodular cast iron

    ledeburite (spherocast)

    5 Hz - 5 kHz

    25 Hz - 25 kHz

    +

    ++

    ++

    +

    +

    +

    ++

    +

    +

    ++

    Surface:

    grinding abuse 5 kHz - 100 kHz ++ ++

    ++very good +good possible not possible

    Table 2.1

    Our standard coils are classified by frequency ranges. Whereas the frequency range

    of the encircling coils is directly indicated on the name plate, the probes are coded

    as follows:

    1 ring 25 Hz to 10 kHz

    2 rings 250 Hz to 50 kHz

    3 rings 5 kHz to 250 kHz

    As far as encircling coils are concerned, a filling degree of 60 per cent should not be

    fallen below as otherwise the measuring signals become too small and the

    differences, e.g. in the case depth are no longer recognized distinctly. When using

    encircling coils, a second coil is required for compensation. This coil must feature a

    firmly mounted compensation part. Removing or shifting the compensation part

  • 8/11/2019 Eddyliner Px User Manual

    15/146

    Manual eddylinerPx General hints

    ibg Doc: 0108131e, Apr 16, 2004 Creation of optimum test conditions

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 2-3

    leads to wrong test results. The compensation part has no comparing function

    designed to distinguish the OK-parts from the NotGood-parts.

    The basic design of the standard encircling coils is displayed in Fig. 2.1. Each coil

    contains a transmitter and a receiver winding. The receiver winding should alwaysbe arranged as close as possible to the area to be measured. It is very easy to find

    out on what side the receiver or transmitter windings are located. Just take an

    encircling coil and read the name plate. If you can read this plate and when the coil

    plug is on the left side, the receiver winding is below and the transmitter winding is

    above in the coil.

    Section A-A

    Coil plug

    Transmitter winding

    Receiver winding

    AA

    ibgP/NS/NDAT

    25Hz-25kHz

    Frequency range

    Fig. 2.1

    For coil protection, wearing bushes of abrasion-resistant plastic material e.g.

    "Murtfeldt S green" should be used.

    2.2. Creation of optimum test conditions

    Please pay attention to the following points when carrying out a test. Carefully read

    the following hints. Disregarding them is likely to lead to faulty measurements. The

    eddy current procedure is no absolute method. The following parameters may cause

    the results to scatter:

    Non-homogeneities in material

    Dimensional tolerances in the test part

    Dimensional tolerances in coil positioning

    Variations in temperature.

    For optimum results please pay attention to the following points:

  • 8/11/2019 Eddyliner Px User Manual

    16/146

    General hints Manual eddylinerPx

    Creation of optimum test conditions Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    2-4 D-91320 Ebermannstadt Germany

    Make sure that the parts for material data recording comply with the desired

    specifications.

    Feed the batches of material separately into the testing device.

    Make sure that the test parts have a constant test temperature.

    For RECORDING O.K. PARTS feed at least 7 to 10 OK-parts per material group

    into the device.

    Select reasonable tolerances to avoid the risk of obtaining undesirable pseudo

    rejects.

    Pick random samples from the reference parts during the test.

    To protect the device against undesirable interventions, lock it via the key switch.

    Read the operating instructions before putting the device into operation.

    Make sure that constant and reproducible test conditions are maintained.

    See to it that no metallic parts, which are not selected for testing, are located

    near the test coils.

    Carry out the test on a wooden table, if possible. Even steel girders underneath a

    wooden board are likely to affect the test results.

    Keep interferences as current impulses, flux machines active during themeasurement, electric welding and other magnetic or electric interference fields

    away from the coil.

    The following factors may adversely affect the measurement directly:

    Positioning of the test part in the coil

    Dimensional tolerances of the test parts

    Variation in the batches

    Differences in temperature between reference part and test part

    Use of coils with a wrong frequency range

    Lift-off effect due to dust and dirt

    Edge effect caused by changing measuring positions with probe.

  • 8/11/2019 Eddyliner Px User Manual

    17/146

    Manual eddylinerPx General hints

    ibg Doc: 0108131e, Apr 16, 2004 Wiring

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 2-5

    2.3. Wiring

    Prior to putting the eddylinerPxinto operation, all cables must be connected.

    WARNING

    Please take care that all cables are plugged and screwed tight

    beforethe power switch is turned on.

    Also ensure that all plugs are firmly connected and screwed tight.

    If plugs are inserted after connection of the eddylinerPxthe

    various components are likely to be damaged !

    Fig. 2.2 shows the rear side of the eddylinerPx. Put the various cable plugs into

    the respective sockets. Mistaken plugging is not possible because each plug fitsonly into the specifically matched socket. The operating system plug has been firmly

    mounted in the factory and needs replacing only when required, e.g. for the update

    of the operating system. The SET-CONTROL-PLUG requires software version

    Px300 or later.

    Coil connectionPower plug External printer

    Operating systemplug

    Optointerface

    (PLC activation,

    sorting switch etc.)

    Name plate Serial interface

    Fine-wire fuse

    T 2.0A

    90-260V

    47-440Hz

    ibgPrfcomputer GmbHP/NS/NDAT

    Relay activation

    T 2.0A

    Set Control

    plug

    Fig. 2.2

    If coil heads, several pairs of coils or several probes are to be used, a coil switch

    must be connected. For this purpose, the coil connection and the control outlet of

    the relay on the backside of the eddylinerPxare connected to the coil switch. The

    individual coil cables are then connected directly to the coil switch.

  • 8/11/2019 Eddyliner Px User Manual

    18/146

    General hints Manual eddylinerPx

    Elements of the front panel Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    2-6 D-91320 Ebermannstadt Germany

    When encircling coils are used, it does not matter which of the two twin cable plugs

    is connected to what coil.

    Important Notice !The eddylinerPxmust be grounded properly (PE connector of

    the mains plug or grounding screw on the instrument housing, see

    drawing of the back side of the instrument below). Otherwise,

    voltage may constitute on the housing. If the instrument is not

    grounded properly, the safety standard protective ground are not

    obeyed. Furthermore, quality of testing is impaired so that optimum

    test results cannot be achieved.

    2.4. Elements of the front panel

    All control elements are arranged on the front panel of the eddylinerPx. They are

    slightly depressed and a background-lit screen is arranged between the operator

    keys. Fig. 2.3 shows the front side of the eddylinerPx. Screen and soft keys are

    explained in the next chapter.

    eddyliner P

    F2

    START

    F1

    F3

    F4 ESC

    QUIT

    F?

    ?

    RESET

    PRINT

    EXTER

    SETU

    Socket for

    External Setup

    Module (ESM)

    Key switch

    +/-

    function lampSoft keys and function keys

    Screen-brightness -contrast Screen with soft-key designations

    Sort display

    NG (Red) OK (Green)

    top

    left right

    bottom

    Cursor keys

    Cursor

    function lamp

    (a)

    (b)

    (c)

    Fig. 2.3

    Softkey [F1]- [F4]The function soft keys will be dealt with in the next chapter.

  • 8/11/2019 Eddyliner Px User Manual

    19/146

    Manual eddylinerPx General hints

    ibg Doc: 0108131e, Apr 16, 2004 Elements of the front panel

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 2-7

    [?] keyWith the [?] key you can call up a context-related help at any point of the controlsoftware of the eddylinerPx. This means that you will receive help at any point

    of the software. Make it a habit to actuate this [? ]key whenever a problem has tobe clarified. By depressing this [? ]key again, you leave the help window.

    [F?]keyThis [F?]key refers to the soft keys [F1] - [F4]; by pressing the [F?]key youreceive a detailed explanation of the individual soft keys which can be selected in

    both short form and long form. See the respective comments in the next chapter.

    [QUIT]keyActuating the [QUIT]key in TEST-SETUP operating mode causes AUTOPIP tobe adjusted to the empty coil, whilst in SETUP EDITOR mode the letters for the

    set denomination are selected.

    [ESC] keyWhen [ESC] key is actuated, a step back is made in the hierarchy of thesoftware.

    If settings were changed without storing, an interrogation is made for safety

    reasons. In general, with [+]you confirm the changes made and in this manneryou make up for the omitted storage. Via [-]you reject the changes (the old dataare displayed again) and pressing the [ESC]key once more, you interrupt theoperation to have another look at the changes.

    If for some reason you must delete all data from the eddylinerPxand reset it to

    the starting position, you can keep the [ESC]key depressed during start-up ofeddylinerPx.Warning! This deletes definitely all settings, material data, tolerance zones and

    sets !

    [START] keyIndependent of the type of measurement triggering in the

    COMMUNICATION-SETUP via the [START] key the measurement can betriggered manually at any time.

    [+]and [-]key

    The keys [+] and [-]can be used for the setting of parameters, selection fromlists and for answering interrogations. The [+]and [-]key is always active whenthe green function lamp between these two keys lights up.

    Cursor keys, left-right-top-bottomThe cursor can be moved across the screen with these keys. This is required

    whenever various choices are offered. The cursor keys are ready for operation

    when the green function lamp between the cursor keys is alight.

    Key switchThis switch has three positions and the key can be taken out at all three

    positions.

  • 8/11/2019 Eddyliner Px User Manual

    20/146

    General hints Manual eddylinerPx

    Elements of the front panel Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    2-8 D-91320 Ebermannstadt Germany

    (a) The instrument is switched off (separated from the power line).

    (b) The instrument is switched on. All functions are accessible.

    (c) The instrument is switched on. No access to the functions.

    During the test unauthorised intervention is not possible.

    When turning the key switch to position (c), an interrogation window

    appears. The range of access restrictions can be selected. If all keys are

    inhibited, the softkeys will disappear from the display. Otherwise, all

    softkeys remain visible although key switch is in position "blocked". This

    way, reading in of different sets in mode SETUP is possible, but individual

    values cannot be modified.

    [Sun]keyThe [Sun]-key has no function. The brightness is always set to maximum.

    [Moon]keyThis [Moon] key is used for setting the screen contrast. You are guidedautomatically by an adjustment menu on the screen.

    [RESET]keyWith this [RESET]key you can reset the eddylinerPxto the MAIN MENU or tothe SORTING menu. The function of the [RESET]key corresponds to the so-called "warm start" in a standard computer. (When the eddylinerPxis switched

    on, it also performs a warm start). For this, the [RESET] and [START] keymust be pressed simultaneously. In the key switch position (c), a reset is not

    possible.If for some reason you have to delete all the data from the eddylinerPxand to

    reset it to its starting position, you may keep the [ESC] key depressed whenswitching it on. Warning ! This erases all settings, material data, tolerance zones

    and sets !

    [PRINT]keyIf an optional laser printer, ink-jet printer or matrix printer is connected to the

    eddylinerPxboth screen printouts (hard copy) and protocol printouts of sorting

    performed may be made.

    If an interrogation window, a help text or a detailed explanation of the softkeys is

    displayed, a screen printout is done without inquiry after pushing the [PRINT]key. Otherwise, you can select either screen printout or protocol printout in a

    special menu.

    External Setup Module ESMA socket is arranged on the front panel for an external setup module. This is

    necessary for the permanent storage of setting values and/or for the transmission

    of such values to another eddylinerPx.

  • 8/11/2019 Eddyliner Px User Manual

    21/146

    Manual eddylinerPx General hints

    ibg Doc: 0108131e, Apr 16, 2004 Elements of the front panel

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 2-9

    2.4.1. Soft keys

    The entire operation of the eddylinerPx is made via so-called soft keys. The

    function of these keys is defined by the software or by the screen text. In Fig. 2.4 the

    MAIN MENU and the adjacent soft keys [F1]- [F4]are shown. On the right side ofthe screen white letters in black fields can be seen. These four fields contain the

    lettering of the adjacent soft keys [F1] - [F4] in abbreviated form (three capitalletters vertically arranged).

    F2

    START

    F1

    F3

    F4 ESC

    QUIT

    F?

    ?

    Operating mode bar

    Information window

    (identical in all menus)

    Definition of soft keysF1-F4 in abbreviated form

    Soft keysF1-F4

    Fig. 2.4

    With the [F?]key a detailed information on the soft keys can be called up. Pressingthe [F?]key replaces the soft key description in abbreviated form with plain text.The soft keys remain active also during the display of this detailed explanatory text.

    On pressing the [F?] key again, the detailed comment disappears and the under-lying screen half becomes visible again.

  • 8/11/2019 Eddyliner Px User Manual

    22/146

    General hints Manual eddylinerPx

    Elements of the front panel Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    2-10 D-91320 Ebermannstadt Germany

    F2

    START

    F1

    F3

    F4 ESC

    QUIT

    F?

    ?

    Unabbreviated definition ofsoft keys F1-F4

    Soft keysF1-F4

    Fig. 2.5

    2.4.2. Information window

    The information window is located on the screen left. It contains the software

    version, you work with, in the left upper corner. Underneath the current date and

    hour the serial number of the eddylinerPx is indicated. Moreover, the currently

    loaded instrument setting (Set...) together with its denomination can be identified

    and also when this setup was changed for the last time.

    Underneath the information window the operating mode bar is arranged whichindicates in white lettering on black background the current active operating mode,

    in this case the MAIN MENU. See also Fig. 2.4.

    2.4.3. Help funct ion

    The eddylinerPx offers a context-related help. This means that you are offered

    actual assistance at any point of the software. The help function provides for every

    operating mode, every warning or message additional information which is displayed

    on the screen via [? ] key. Fig. 2.6 reproduces the help window from the MAIN

    MENU. The help window is subdivided as follows: on top in the header bar thecurrent theme of the auxiliary function is indicated. On the right side of this header

    bar you find information on the page number selected and the total number of pages

    relating to this theme. Underneath there is the auxiliary text.

  • 8/11/2019 Eddyliner Px User Manual

    23/146

    Manual eddylinerPx General hints

    ibg Doc: 0108131e, Apr 16, 2004 Test sequence

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 2-11

    Current help theme Page No/ Number of pages

    Help text

    Fig. 2.6

    With the cursor keys topand bottomyou can select the corresponding page of the

    help function. By pressing again the [? ] key or the [ESC]key, you quit the helpfunction.

    There is yet the [F?]key with a particular help function. By depressing this key, a

    detailed comment on the soft keys can be called up; in this context, see also Fig. 2.5

    on page 2-10. The soft keys remain active also during the display of the detailed

    description.

    2.5. Test sequence

    The eddylinerPxoperates on the ibg-developed principle of PREVENTIVE MULTI-

    FREQUENCY TESTING (PMFT). After the measurement has been triggered, the

    test frequencies are applied successively to the coil system, starting at the lowest

    test frequency. At each test frequency the response of the test part to the coil

    system is recorded and the electric voltage vector is broken down into its X and Y

    components and stored.

    Upon completion of the test, depending on the operating mode, the values stored

    are either compared with the tolerance zones (sorting) or assigned to the reference

    values (mode RECORDING O.K. PARTS).

    PIP (piece in probe)

    Settling delay

    MPRG (measurementin progress)

    OKx (sort impulse ok)

    ENDx (end ofmeasurement)

  • 8/11/2019 Eddyliner Px User Manual

    24/146

    General hints Manual eddylinerPx

    Test sequence Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    2-12 D-91320 Ebermannstadt Germany

    Start of measurement with "PIP-signal" (Piece In Probe) :

    A measurement must be started (PIP-signal) and requires a certain period of time

    (MPRG-signal). Triggering of measurement is only possible in modes RECORDING

    O.K. PARTS or SORTING; in all other modes of the eddylinerPxa PIP-signal doesnot trigger measurement, an MPRG-signal does not appear.

    Settling delay:

    After triggering of a measurement, but before measurement itself starts, an

    adjustable settling delay may be selected. This settling delay allows the test part to

    physically settle inside the test coil and measurement will not start unless the test

    part has reached its final position in the coil.

    "MPRG-signal" (Measuring in Progress) :

    Working in mode RECORDING O.K. PARTS or SORTING, the MPRG-signal

    (Measuring in Progress) will appear immediately after the PIP-signal (Piece In

    Probe). The settling delay follows, and after termination measuring itself will start at

    the first frequency.

    Sorting decision with "OKx-signal":

    If testing is completed at all frequencies, the sort decision is made and applied to the

    sort output (OKx). In the operating mode RECORDING O.K. PARTS, of course no

    sort signal will be generated.

    The ENDx-signal (end of testing by channel .x):

    If all test frequencies for channel .x have completed testing (and possibly a sort

    decision has been made), the ENDx-signal will be generated.

    If the PIP of the following channel is set to AFTER, settling delay of this channel

    starts immediately, the MPRG-signal remains active.

    If it is the last channel within one block of testing, the MPRG-signal disappears.

    Once the MPRG-signal has disappeared, the eddylinerPxis again ready for a new

    measurement (see also chapter "10.2. Time diagram" on page 10-56).

  • 8/11/2019 Eddyliner Px User Manual

    25/146

  • 8/11/2019 Eddyliner Px User Manual

    26/146

    Short operating instructions Manual eddylinerPx

    Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    3-2 D-91320 Ebermannstadt Germany

    eddyliner P

    F2

    START

    F1

    F3

    F4 ESC

    QUIT

    F?

    ?

    RESET

    PRINT

    EXTER

    SETU

    Socket for

    External Setup

    Module (ESM)

    Key switch

    +/-

    function lampSoft keys and function keys

    Screen-brightness -contrast Screen with soft-key designations

    Sort display

    NG (Red) OK (Green)

    top

    left right

    bottom

    Cursor keys

    Cursor

    function lamp

    (a)

    (b)

    (c)

    Fig. 3.1

    (a) When encircling coils are used:

    Place a part in the compensation coil(s).

    (b) When using self-compensated coils or probes:

    A compensation part is not required.

    Press [F3], MAT (Material data)Thereupon the MATERIAL DATA mode appears. Here the OK-parts are input

    and the tolerance zones are formed.

    Press [F1], NEW (new recording of OK-parts ).Thereupon the RECORDING O.K. PARTS mode appears.

    (a) "Simultaneous" testing at several positions using a coil head or several

    pairs of coils:

    Place the O.K.-parts (reference parts) one after the other into the coil.

    Press the [START]-key on the eddylinerPx after every part. Now alocus curve covering the set frequencies is recorded by every channel for

    every one of the at least 10 O.K.-parts.

    (b) Subsequent testing at several positions using one pair of encircling coils

    or a probe:

    Move the cursor to the leftuntil channel .1 is displayed.

    Move an O.K.-part (reference part) or the coil to test position 1 and press

    the [START]-key on the eddylinerPx. Now, a locus curve covering theset frequencies is recorded by channel .1.

    Move the cursor to the rightuntil channel .2 is displayed.

  • 8/11/2019 Eddyliner Px User Manual

    27/146

    Manual eddylinerPx Short operating instructions

    ibg Doc: 0108131e, Apr 16, 2004

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 3-3

    Move an O.K.-part (reference part) or the coil to test position 2 and press

    the [START]-key on the eddylinerPx. Now, a locus curve covering theset frequencies is recorded by channel .2.

    Proceed in the same manner until a locus curve covering the setfrequencies is recorded by every channel for every one of the at least 10

    O.K.-parts.

    Press [F4], END (end of recording OK-parts)Now the eddylinerPxautomatically forms tolerance zones for the parts entered.

    These zones are used for comparison in the ensuing test.

    Press [F4], OK (all tolerance zones are in order)Now the eddylinerPxautomatically jumps into the SORTING mode. This is the

    moment to start the test proper of all parts.

    SORTING

    (a) "Simultaneous" testing at several test positions using a coil head or

    several pairs of coils:

    Place all parts to be tested one after the other into the coil and press the

    [START]-key on the eddylinerPxfor every single part. When a part isOK on the indicated channel, the sort indicating lamp shows green. If a

    part is NOT GOOD (NG), the sort indicating lamp shows red.

    (b) Subsequent testing at several positions using one pair of encircling coils

    or one probe:

    Move the cursor to the leftuntil channel .1 is displayed.

    Move the part to be tested or the coil to test position 1 and press the

    [START]-key on the eddylinerPx. If the part is OK on channel .1, thesort indicating lamp shows green. If the part is NOT GOOD (NG), the sort

    indicating lamp shows red.

    Move the cursor to the rightuntil channel .2 is displayed.

    Move the part to be tested or the coil to test position 2 and press the

    [START]-key on the eddylinerPx. If the part is OK on channel .2, the

    sort indicating lamp shows green. If the part is NOT GOOD (NG), the sortindicating lamp shows red.

    The part to be tested will be tested in this manner at all test positions.

    Press [PRINT]on the eddylinerPx,for printing out a test protocol on the connected printer. An interrogation window

    appears. Now you can decide on printing out either the screen or a protocol.

    Press [-]on the eddylinerPxNow the connected printer prints out a protocol.

  • 8/11/2019 Eddyliner Px User Manual

    28/146

    Short operating instructions Manual eddylinerPx

    Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    3-4 D-91320 Ebermannstadt Germany

    By pressing the [+]key the screen display (hardcopy) can be printed out, if protocolis not wanted.

    The current SORTING is not affected by the printout, only the screen display

    remains "frozen" as long as the printout lasts.

    In the SORTING menu the parts counters are arranged on top right (see Fig. 3.2). If

    the parts counters are to be reset to zero, proceed as follows:

    Press [F1], RES (reset parts counters)A safety interrogation appears. Pressing the [ESC]key returns to the display ofthe SORTING menu without causing a change in the parts counters.

    Press [+]All parts counters are reset to 0 and the eddylinerPxautomatically goes back to

    the display of the SORTING.

    F2

    START

    F1

    F3

    F4 ESC

    QUIT

    F?

    ?

    parts counter

    Fig. 3.2

  • 8/11/2019 Eddyliner Px User Manual

    29/146

    Manual eddylinerPx Operating modes and functions

    ibg Doc: 0108131e, Apr 16, 2004

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 4-1

    4. Operating modes and functions

    The operation of the eddyliner

    Pxis facilitated by the hierarchical structure of themenus. In the following some basic information is given as a guide to the operator.

    By pressing a soft key either a function can be selected directly or a new menu will

    appear. With the appearance of a new menu, the soft keys specify further new

    functions or operating modes. In this manner a hierarchical structure is set up in the

    form of a decisional tree branching out from top to bottom. Only where the work is

    considerably eased it is possible to jump from one tree configuration branch to the

    other, say, from MATERIAL DATA to SORTING and back again. Fig. 4.1 shows the

    software structure diagram. MODULE TEST 3 is first available with software version

    Px300.

    MODULE TEST 3

    MAIN MENU

    DYNAMICS CONTROL

    CORRECT TOLZONE

    SORTING INTERRUPT SORTING

    SETUP

    OPERATION STATISTICS

    MATERIAL DATA

    RECORDING O.K. PARTS

    TOLERANCE ZONE

    SETUP - EDITOR

    MODULE TEST 2

    TEST SETUP

    COMMUNICATION SETUP

    SORTING DATA

    MODULE TEST 1

    Fig. 4.1

    4.1. MAIN MENU

    The four menus SETUP, MODULTEST, MATERIAL DATA or SORTING are

    selected from the MAIN MENU via soft keys. Fig. 4.2 shows the screen of the MAIN

    MENU. You find the activable menus in a three-letter sign on the right side of the

  • 8/11/2019 Eddyliner Px User Manual

    30/146

    Operating modes and functions Manual eddylinerPx

    Setup Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    4-2 D-91320 Ebermannstadt Germany

    screen, white on black ground. These menus are selected via the adjacent soft keys

    [F1]- [F4].

    F2

    START

    F1

    F3

    F4 ESC

    QUIT

    F?

    ?

    Fig. 4.2

    4.2. SETUP

    All settings necessary for adapting the eddylinerPx to the specific test task are

    carried out in the SETUP menu and the subordinate menus. The adjustable

    parameters of the eddyliner

    Pxare subdivided into two groups:

    The test parameters

    contain all settings depending directly on the parts to be tested. This includes

    four representative OK-parts, not material data. In order to guarantee quick

    change-over from one test part to another, 50 sets of test parameters can be

    stored internally. They are available on pushing a button. Apart from the set

    number (1.. 50), a setup name can be chosen by the user for identification of the

    test parameters.

    Before software version Px215 or CPU Mod F only 10 Setups are available.

    The communication parameters contain all settings depending on the technical

    environment of the eddylinerPx. They are stored internally only once, i.e. when

    changing from one test part to another by choosing another setup number, these

    parameters remain unchanged.

    When the SETUP menu is called up, the significance of the soft keys changes. Fig.

    4.3 gives the detailed definition of the soft keys in the SETUP menu.

  • 8/11/2019 Eddyliner Px User Manual

    31/146

    Manual eddylinerPx Operating modes and functions

    ibg Doc: 0108131e, Apr 16, 2004 Setup

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 4-3

    SETUP

    F

    R

    Q

    Selection of Sets,

    setting of FRQ,

    PipFRQ, delay..

    Set of Pip mode,

    output signals,

    printer, V24, Astart..

    Read test and com-

    munication parameters

    from SETUP module

    Store test and com-

    munication parameters

    on SETUP module

    E

    S

    R

    E

    S

    S

    C

    O

    M

    TEST SETUP

    COMMUNICATION SETUP

    Fig. 4.3

    The soft key [F1]calls up the menu TEST-SETUP. In this menu, the individual testparameters can be checked and changed. The soft key [F4] invokes the menuCOMMUNICATION-SETUP. In this menu, the individual communication parameters

    can be checked and changed.

    In the SETUP menu itself the individual parameters cannot be changed. Stored

    settings can be retrieved quickly. Fig. 4.4 shows the monitor of the SETUP menu.

    F2

    START

    F1

    F3

    F4 ESC

    QUIT

    F?

    ?

    setupInformation window

    Name of the setup

    Date of last change

    Number of the

    Fig. 4.4

    The information window contains general information and indicates the active setup

    presently selected. The various entries are explained in chapter "2.4.2. Information

    window" on page 2-10.

  • 8/11/2019 Eddyliner Px User Manual

    32/146

    Operating modes and functions Manual eddylinerPx

    Setup Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    4-4 D-91320 Ebermannstadt Germany

    Once the setting of the test parameters for every type of part to be tested is stored,

    changing over from one type to another is very simple. Setting of the system to a

    different type of part is done in the MAIN MENU as follows:

    Press [F1], SET,and the SETUP menu is activated.

    Press [+]or [-],to select the relevant number of setup from the total of 50 internally stored setups

    of test parameters. Apart from the number of the selected setup, the date of last

    change is displayed. The name of the setup in the display is intended to facilitate

    the identification of the test task.

    Press [ESC],and an interrogation window will appear.

    Press [+]The setting selected is now active.

    If the 50 internal settings available in the eddylinerPxare not sufficient or if various

    communication parameters are to be used, an External Setup Module offers further

    storage capacity.

    Press soft key [F3], SES, to store the displayed instrument setting in the ExternalSetup Module (ESM). In the ESM, the data stored always comprise test parameters

    and communication parameters. This allows to store the complete setting of the

    eddyliner

    Pxseparately from the test instrument for safety reasons.

    Loading the complete setting from the External Setup Module (ESM) in the MAIN

    MENU is very simple.

    Press [F1], SET,to activate the SETUP menu.

    Press [+]or [-],to select a setup of which the internally stored data can be overwritten.

    Press [F2], ESL,and date and name of the setup stored on the ESM are indicated. The previouslyselected number of setup is not changed.

    Press [ESC],and an interrogation window will appear.

    Press [+],and the test parameters and the communication parameters retrieved from the

    ESM are taken over and activated.

  • 8/11/2019 Eddyliner Px User Manual

    33/146

    Manual eddylinerPx Operating modes and functions

    ibg Doc: 0108131e, Apr 16, 2004 Setup

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 4-5

    As a result of the take-over from the ESM, all test and communication parameters

    which were previously available under the selected setup in the eddylinerPx are

    overwritten.

    Please see chapter "4.2.3. Working with the External Setup Module (ESM)" on page4-29 for further information on the ESM.

    4.2.1. COMMUNICATION SETUP

    All settings depending on the technical environment of the eddylinerPxare done in

    the COMMUNICATION-SETUP menu. These communication parameters are

    internally available only once, i.e. they remain unchanged when selecting a different

    setup for changing over from one type of part to another. Fig. 4.5 shows the

    COMMUNICATION-SETUP menu.

    F2

    START

    F1

    F3

    F4 ESC

    QUIT

    F?

    ?

    Information window

    Fig. 4.5

    To the right of the information window, general communication parameters are

    displayed. At the bottom of the display, one line is reserved for every existing

    channel. Here, the specific communication parameters for every individual channel

    are shown. Usage of the sum outputs OKS/NGS and END is shown in an extra line.

    When the COMMUNICATION menu is called up, the significance of the soft keys

    changes. Fig. 4.6 gives the detailed definition of the soft keys in the

    COMMUNICATION menu.

  • 8/11/2019 Eddyliner Px User Manual

    34/146

    Operating modes and functions Manual eddylinerPx

    Setup Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    4-6 D-91320 Ebermannstadt Germany

    COMMUNICATION SETUP

    D

    E

    F

    Display of

    BASIC SETUP

    on the screen

    Store displayed

    parameters as active

    setting

    Read and display com-

    munication parameters

    from SETUP module

    Store test and com-

    munication parameters

    on SETUP module

    E

    S

    R

    E

    S

    S

    S

    A

    V

    Fig. 4.6

    After calling up the menu COMMUNICATION-SETUP, the active instrument setting

    is displayed, i.e. the displayed setting is activated. All changes are done at first on

    the monitor only and may be cancelled at any time by pressing the softkey [F4],SAV, and pressing [-] on inquiry. Only when the displayed setting is stored bypressing [F4], SAV, and pressing [+]on inquiry, the setting is activated.

    Press [ESC]to quit the menu COMMUNICATION-SETUP. If you have modified theset parameters without saving the changes, an inquiry will appear. Press [+] to storethe changes, [-]to cancel the changes or [ESC]to review the changes.

    Press soft key [F3], SES, to store the displayed communication parameters in theExternal Setup Module (ESM). In the ESM, the data stored always comprise the

    active test- and the displayed communication parameters. This allows to store the

    complete setting of the eddylinerPxseparately from the test instrument for safety

    reasons.

    Proceed in the following manner to read the communication parameters (without test

    parameters) in the COMMUNICATION-SETUP from the External Setup Module

    (ESM):

    Press [F2], ESLThe communication parameters stored on the ESM are read and displayed.

    Press [F4], SAVAn interrogation window appears.

    Press [+]to take over the communication parameters read in from the ESM as active

    instrument setting.

  • 8/11/2019 Eddyliner Px User Manual

    35/146

  • 8/11/2019 Eddyliner Px User Manual

    36/146

    Operating modes and functions Manual eddylinerPx

    Setup Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    4-8 D-91320 Ebermannstadt Germany

    The following chapters contain detailed explanations concerning the function and

    setting of the individual test parameters.

    4.2.1.1. Self-start

    The eddylinerPx features a self-start function following power failure or

    disconnection. With this function in ON position, the eddylinerPxstarts again and

    again with the SORTING, provided that this operating mode had been selected prior

    to power failure or disconnection. If the self-start is in OFF position, the

    eddylinerPxstarts with a warning and activates the ALARM output, whenever the

    SORTING has been active before power failure and disconnection. In all other

    cases the eddylinerPxwill start with the MAIN MENU.

    Move the cursor to "self-start" using the cursor keys. The word is now shown

    inverted.

    Press [+]to switch the self-start from OFF to ON or vice versa.

    Press [F4], SAV,and an interrogation window will appear.

    Press [+]

    The setting selected is now active.

    4.2.1.2. Volume

    The eddylinerPxhas a built-in signal generator. The volume of the internal signal

    generator may be varied between 0 and 100 %. Whenever a key on the

    eddylinerPxis pressed, the built-in signal generator emits a signal.

    Move the cursor to "LOUD" using the cursor keys. The word is now shown

    inverted.

    Press [+]or [-],to select the desired volume. The volume setting can be checked by pressing,

    e.g. the [QUIT] key.

    Press [F4], SAVand an interrogation window will appear.

    Press [+]to store the volume set.

  • 8/11/2019 Eddyliner Px User Manual

    37/146

    Manual eddylinerPx Operating modes and functions

    ibg Doc: 0108131e, Apr 16, 2004 Setup

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 4-9

    4.2.1.3. V24-interface

    Setting of parameters is only necessary if the eddylinerPx is connected to an

    external computer via serial interface (V24 or RS232).

    In this case, all parameters indicated following to "V24 :" (baud rate, parity, number

    of data bits, number of stop bits) have to meet exactly the parameters of the

    computer connected to the interface. Baud rates higher than 9600 Baud are

    available with software version Px300 and PCPR11 or later. See also chapter

    "5.1. Interconnection with external computer" on page 5-56.

    Move cursor to the relevant parameter indicated after "V24 :". The selected

    parameter is now shown inverted.

    Press key [+]or [-] ,

    to set parameter to the requested value.

    Press key [F4], SAVAn interrogation window appears.

    Press [+],to activate modified V24 parameters.

    For more details on data transfer or control of eddylinerPxvia serial interface see

    chapter "5. Data transfer via V24 interface" as from page 5-56.

    4.2.1.4. The printer

    Commercially available laser printers, ink-jet or matrix printers can be connected to

    the eddylinerPxvia the parallel interface. Adapters are available for various printer

    types. With the eddylinerPx you may print a protocol or the screen display

    (hardcopy). Further information is provided by the help windows.

    Position cursor on the denomination "printer" using the cursor keys. The word is

    now shown inverted.

    Keep [+] or [-]depresseduntil the desired printer driver is displayed on the screen.

    Press [F4], SAVand an interrogation window will appear.

    Press [+]The changed printer driver is active now.

  • 8/11/2019 Eddyliner Px User Manual

    38/146

    Operating modes and functions Manual eddylinerPx

    Setup Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    4-10 D-91320 Ebermannstadt Germany

    4.2.1.5. Warning

    This function switches the alarm output on, in case an adjustable number of

    NotGood-parts (NG) is recognized in succession by the eddylinerPx. The test in

    progress is not interrupted. A single OK-part resets the NotGood alarm counter, ifthe threshold has not yet been reached, and counting restarts with the next NG-part

    being counted again as number one. The threshold of the NG alarm counter may be

    set in the range from 1 to 999. The zero value switches off this function.

    Position the cursor on the name "Warn:" using the cursor keys. The word is

    shown inverted now.

    Press [+]or [-],to enter the desired alarm threshold.

    Press [F4], SAVand an interrogation window will appear.

    Press [+]The setting selected is active now.

    4.2.1.6. Sort ing data recording

    In SORTING mode, a certain number of measured data is stored. It may be selected

    whether all data or only NG parts are to be stored.

    Depending on the instrument equipment, a maximum number of 10 parts(eddylinerP with PCPU up to MOD E and without PRAM) or 200 parts

    (eddylinerPwith PRAM or PCPU as of MOD F) may be stored. If this maximum is

    reached the data of the oldest part are overwritten when a new part is stored.

    This function is first available with software version 2.14

    Position the cursor on the name "SDat:" using the cursor keys. The word is

    shown inverted now.

    Press [+]or [-],to enter the desired material data.

    Press [F4], SAV,and an interrogation window will appear.

    Press [+]The setting selected is active now.

  • 8/11/2019 Eddyliner Px User Manual

    39/146

    Manual eddylinerPx Operating modes and functions

    ibg Doc: 0108131e, Apr 16, 2004 Setup

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 4-11

    4.2.1.7. The sort impulse

    Every channel generates a separate OKx-signal (sort output OK of channel .x) (OK1

    to OK4 for eddylinerP4). In the SORTING mode after termination of every test with

    channel .x an OKx-signal may be generated.

    When using sum signals (see chapter "4.2.1.11. Generation of the sum sort outputs

    OKS/NGS" on page 4-14), after termination of the last logically linked test,

    additionally an OKS or NGS-signal (sum sort output OK or NOT GOOD) will be

    generated.

    The length of these signals has to be set in the range from 50 ms to 5,000 ms or to

    one of the SYNC operating modes to be selected via [+]and [-]. With the SYNC 1setting, the sort signal is kept alive until the next PIP-signal is emitted. With the

    SYNC 2 setting, the sort signal remains in force until the next sort decision is made.

    If the SORTING operating mode is quit, the sort signals also disappear.

    Position the cursor on the denomination "sortimplength" using the cursor keys.

    The word is shown inverted now.

    Press [+]or [-]You may choose between the synchronous operating modes SYNC 1 and

    SYNC 2 or a fixed time in milliseconds.

    Press [F4], SAV,and an interrogation window will appear.

    Press [+]The setting selected is active now.

    4.2.1.8. PIP-function (Start of test)

    Testing (RECORDING O.K. PARTS or SORTING) is initiated with the PIP signal.

    PIP is the abbreviation for Piece In Probe. For every channel .x there is a separate

    optically isolated input PIPx available for external test initiation (PIP1 to PIP4 for

    eddylinerP4).

    The PIP signal of test position .1 has four different modes of operation :

    (a) AUTOMatic operation by AUTOPIP when a test part is approaching the

    test coil and when a certain threshold value is reached

    (For this operating mode, the frequency and the threshold value for

    AUTOPIP must be set !)

    (b) MANUAl operation by pushing the [START] key or, if a probe cable withtest initiation button is connected, by this test initiation button.

    (c) PIP1 via external optically isolated PIP1 input.

    (d) KONTInuous operation - measurement after measurement

  • 8/11/2019 Eddyliner Px User Manual

    40/146

    Operating modes and functions Manual eddylinerPx

    Setup Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    4-12 D-91320 Ebermannstadt Germany

    The PIP signals of test positions .2 through .4 have three different operating modes :

    (b) MANUAl operation by pushing the [START]key or, if a probe cable withtest initiation button is connected, by this test initiation button.

    (c) PIPx via external optically isolated PIPx input

    (d) AFTER the end of measurement at the previous test position.

    For using the AUTOPIP please pay attention to the hints in chapter

    "6.1.1.4. Automatic Triggering (AUTOM-PIPcod." on page 6-56).

    Furthermore, the [START]key on the eddylinerPxrespectively the test initiationbutton on the probe cable can be used to start measurement in all PIP operating

    modes. These keys always refer to the block of tests containing the displayed

    channel. If all channels are displayed, they refer to channel .1 .

    The PIP operating mode AFTER generates an internal PIP signal when testing is

    completed at the previous test position. Then, the test result is combined with the

    result of the test position before the previous one. Every "block of tests" (displayed

    between the columns "coil" and "time") requires only one single external PIP signal.

    The total test time comprising all test times of one block of tests is displayed in the

    column "time" at the end of every block. The MPRG signal (Measuring in Progress)

    is not switched off between two test positions connected via AFTER-PIP, it remains

    active for the complete block of tests. Due to this, one block of tests can be

    considered as one single test.

    Position the cursor to the column under "Pip" and in the line of the desired test

    position using the cursor keys. The PIP operating mode is now shown inverted.

    Press [+]or [-]to select the desired PIP operating mode.

    Repeat these steps until all test positions are set to the correct PIP operating

    mode.

    Press [F4], SAV,and an interrogation window will appear.

    Press [+]The PIP operating modes selected are now active.

    A detailed explanation of the various PIP operating modes including setting

    instructions for AUTOPIP is given in chapter "6.1.1. Start of measurement (PIP)", on

    page 6-56.

    4.2.1.9. Allocation of coils

    The eddylinerPx has x channels for subsequent testing. If the eddylinerPx is

    connected to a coil switch, a coil to be used for testing can be selected for every test

  • 8/11/2019 Eddyliner Px User Manual

    41/146

    Manual eddylinerPx Operating modes and functions

    ibg Doc: 0108131e, Apr 16, 2004 Setup

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 4-13

    position. Coil 1 is firmly allocated to test position 1. To the other test positions (.2

    through .4 for the eddylinerP4) any coils or pairs of coils can be allocated. There is

    a choice only of those coils connected to the coil switch.

    In most applications, coil x will be allocated to the test position x. The allocation ofone coil to several test positions may be sensible if, for example:

    - a part is tested by one coil at several positions,

    - different parts are tested by one coil or

    - a part is tested at both ends and/or a recognition of direction is done by

    two coils

    The choice of test coils effects the logic combination of the OKx sort outputs and

    thus - if generated - on the OKS/NGS signals as well. Every combination is

    displayed in the column "Sorting" in front of OKx and means an AND operation of

    the individual OKx signals.

    Move the cursor to the column under "Coil" and to the line of the desired test

    position .2 through .4 . The number of the coil is shown inverted.

    Press [+]or [-]to select the desired coil.

    Repeat these steps until the correct coil is allocated to every test position.

    Press [F4], SAV,

    and an interrogation window will appear.

    Press [+]The allocation of the coils to the test positions is now active.

    4.2.1.10. END-impulse

    Every channel generates a separate ENDx-signal (output end of testing with

    channel .x) (END1 to END4 for eddylinerP4). In the modes RECORDING O.K.-

    PARTS and SORTING after termination of every test with channel .x an ENDx-

    signal is generated.

    When using sum signals (see chapter "4.2.1.11. Generation of the sum sort outputs

    OKS/NGS" on page 4-14) after termination of the last logically linked test,

    additionally to the end signal of the single channel an END-signal will be generated.

    For the signal length a value ranging from 50 ms to 5,000 ms must be selected via

    [+] and [-]. A zero value switches the END- or the ENDx-signal off. When theoperating mode RECORDING O.K. PARTS OR SORTING is quit, the END- or the

    ENDx-signal disappears as well.

  • 8/11/2019 Eddyliner Px User Manual

    42/146

    Operating modes and functions Manual eddylinerPx

    Setup Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    4-14 D-91320 Ebermannstadt Germany

    Move the cursor to the column under "Endimp." and to the line of the desired test

    position. The respective signal length is shown inverted.

    Press [+]or [-]

    to select the desired signal length.

    Repeat these steps until all test positions .x are set to the correct ENDx- signal

    length and - if applicable - the correct END-signal length is set for the logic

    operation of the sum sort outputs.

    Press [F4], SAV,and an interrogation window will appear.

    Press [+]The setting selected is now active.

    4.2.1.11. Generation of the sum sort outputs OKS/NGS

    Every channel provides a separate OKx-signal (sort output OK of channel .x) (OK1

    to OK4 for eddylinerP4).

    In many cases, connection of a sorting gate or a PLC can be facilitated by doing the

    logic operations of the individual test results in the eddylinerPx already. In this

    case, instead of (or in addition to) the OKx and ENDx outputs the sum sort outputs

    OKS/NGS and the output END are used.

    The sum sort outputs OKS/NGS can be generated by an AND or an OR operation

    from the (possibly combined) OKx signals available.

    The AND operation is an obvious choice e.g. for testing one part at several

    positions. OKS will be placed after termination of test at the last test position only if

    every position was OK, otherwise NGS will be activated.

    The OR operation is used e.g. for testing a part at both ends, if a recognition of

    direction is done by two coils or if different parts are sorted. OKS is placed after

    termination of test at the last test position if at least one of the OKx signals used was

    placed, otherwise NGS will be activated.

    Whenever the OKS/NGS signals are used, the sequence of test

    positions must be strictly adhered to !

    When testing different types of parts, a logic combination of the OKx signals may not

    be sensible. In this case, the logic combination should be switched off.

    The choice of logic operation effects not only the generation of OKS and NGS

    signals, but also the logic combination of the test positions when recording, marking,

    deleting and counting parts.

  • 8/11/2019 Eddyliner Px User Manual

    43/146

    Manual eddylinerPx Operating modes and functions

    ibg Doc: 0108131e, Apr 16, 2004 Setup

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 4-15

    Move the cursor to the last line of the column "Sorting". The kind of logic

    operation and the term "OKS" are shown inverted.

    Press [+]or [-]

    to select the desired kind of logic operation.

    Press [F4], SAV,and an interrogation window will appear.

    Press [+]The function selected is now active.

    4.2.1.12. Coil moni tor ing

    During function Coil monitoring: 'ON' the test coils will be checked for short circuit orinterruption. In those cases, the test procedure will be interrupted, the ALARM

    output activated and an error message issued on the screen.

    Move the cursor to "Coil monitoring:" using the cursor keys. The word is now

    shown inverted.

    Press [+]to switch the monitoring from OFF to ON or vice versa.

    Press [F4], SAV,

    and an interrogation window will appear.

    Press [+]the setting selected is now active.

    In order not to extend test time unnecessarily, this monitoring is always done during

    automatic self-calibration (see manual, chapter "6.5.4 Automatic Self-Calibration").

    During sorting or recording of material data (if the PIP is not deactivated), automatic

    self-calibration is done in regular intervals (when the instrument is switched on,

    approx. every minute, later approx. every 10 minutes). For trial purposes, automatic

    self-calibration may be forced as follows:

    1) by calling up the menu TEST SETUP or

    2) by changing the AUTOPIP frequency in this menu (if the AUTOPIP is switched

    on)

    If a coil defect occurs during sorting, all parts are sorted NOK at first because the

    parts do not meet the tolerance fields. In the next automatic self-calibration, the

    defect will be recognized and reported. This prevents that unconsciously, OK parts

    are recorded with defective coils.

  • 8/11/2019 Eddyliner Px User Manual

    44/146

    Operating modes and functions Manual eddylinerPx

    Setup Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    4-16 D-91320 Ebermannstadt Germany

    In special applications, it is recommended to deactivate coil monitoring:

    1) If customized coils with very small or very large impedance are used or

    2) if magnetic noise fields occur

    because in these cases, false alarms may result.

    4.2.2. TEST-SETUP

    In the menu TEST-SETUP all settings which depend directly on the parts to be

    tested are done. In order to guarantee easy change-over to a different type of test

    part, 50 test parameter setups can be stored internally. The number of the setup (1..

    50) and a setup name which can be chosen by the user help to easily identify the

    test parameters. Before software version Px215 or CPU Mod F only 10 Setups areavailable.

    Fig. 4.8 shows the TEST-SETUP.

    F2

    START

    F1

    F3

    F4 ESC

    QUIT

    F?

    ?

    Information window

    Fig. 4.8

    General test parameters and information are displayed underneath the information

    window and in the first two lines on the right of the information window. The test

    parameters which can be set separately for every channel are displayed on the right

    side of the display underneath the setup name.

    When the menu TEST-SETUP is called up, the significance of the softkeys changes.

    Fig. 4.9 gives the detailed definition of the soft keys in the menu TEST-SETUP.

  • 8/11/2019 Eddyliner Px User Manual

    45/146

    Manual eddylinerPx Operating modes and functions

    ibg Doc: 0108131e, Apr 16, 2004 Setup

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 4-17

    TEST SETUP

    D

    E

    F

    Display of

    BASIC SETUP

    on the screen

    Store displayed set as

    active setting

    Read and display

    test parameters from

    EXTERNAL SETUP module

    Store test and com-

    munication parameters

    on SETUP module

    E

    S

    L

    S

    E

    S

    S

    A

    V

    Fig. 4.9

    After the TEST-SETUP menu has been called up, it is invariably the current

    instrument setting that is displayed; the displayed setting is identical to the current

    setting. All changes are done on the screen only at first and can be cancelled at any

    time by pressing [F4], SAV, and pressing [-] on inquiry. Only when the displayedsetting is stored by pressing [F4], SAV, and pressing [+] on inquiry, the setting isactivated.

    You can quit the TEST-SETUP menu only by pressing the [ESC] key. If you havemodified the set parameters (e.g. by selecting another set) without storing them via

    [F4], SAV, an interrogation ensues. Then you must decide on either storing thechanges via [+]key or on rejecting them via [-] key or on having another look at thechanges made via [ESC]key.

    Under every setup number the test parameters for all existing channels are stored,

    i.e. for the eddylinerPxfor the channels .1 to .4 (see fig. 4.10).

    The material data of the parts recorded during RECORDING OF O.K.-PARTS are

    notstored under the setup number ! When calling up a setup number containing

    valid tolerance fields, however, automatically four "O.K.-parts" are generated. Thesefour "parts" represent all formerly existing, O.K.-parts which were not stored.

  • 8/11/2019 Eddyliner Px User Manual

    46/146

    Operating modes and functions Manual eddylinerPx

    Setup Apr 16, 2004, ibg Doc: 0108131e

    Pretzfelder Strae 27

    4-18 D-91320 Ebermannstadt Germany

    stored test parameters

    displayed test parameters

    (10 internal sets)

    Set01.4

    Set01

    Set01.3Set01.2Set01.1

    Set02

    Set03

    Set10

    Fig. 4.10

    The screen of the TEST-SETUP is explained in detail in fig. 4.10 on page 4-18. The

    following parameters are valid for all channels and are stored only once per setup

    number:

    Setup number

    Setup date (generation or last modification of setup)

    Setup name

    Interference suppression

    The following parameters are valid only for the channel indicated on the screen and

    are stored for every channel once (i.e. four times for eddylinerP4) per setup

    number:

    Tolerance fields, if formed

    Tolerance field factors (tolerance field dimensions)

    Test frequencies

    Test time for individual frequency

    Settling delay

    Test time (time for the individual test frequencies plus settling delay)

    The parameters in the AUTOPIP information window (in the bottom part of the

    screen) are displayed only if the PIP of channel .1 is set to AUTOM in the

    communication parameters. The AUTOPIP-frequency and the AUTOPIP threshold

    value are always valid for channel .1 and are stored only once per setup number.

  • 8/11/2019 Eddyliner Px User Manual

    47/146

    Manual eddylinerPx Operating modes and functions

    ibg Doc: 0108131e, Apr 16, 2004 Setup

    Pretzfelder Strae 27

    D-91320 Ebermannstadt Germany 4-19

    If the 50 internal setups provided by the eddylinerPx are not sufficient, more

    storage capacity is available in external setup modules.

    For storage of the displayed test parameters on the external setup module (ESM),

    press [F3], SES. The test parameters displayed and the active communicationparameters will be stored on the ESM. This allows the storage of the complete

    setting of the eddylinerPxseparately for safety reasons.

    In order to load all test parameters (without communication parameters) in the

    TEST-SETUP from the external setup module (ESM), proceed in the following

    manner:

    Move the cursor to the word "Set..". The word is now shown inverted.

    Press [+]or [-]

    to select a setup number, the internally stored data of which c