Dual Energy CT Physics:

Hardware and Image Quality Assessment

AAPM 2015

B. Schmidt, C. Hofmann

Siemens Healthcare GmbH

Imaging & Therapy Systems

Page 2: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Outline

1. Physics of Spectral CT Measurements

2. Techniques to Acquire Spectral CT Data

3. Spectral CT Quality

4. What About the Dose?

Page 3: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

The value of color…

• often not needed in daily life

− does not matter

− obvious for known objects

• but there can be surprises …

C. Hofmann, B. Schmidt

3

Page 4: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

PHYSICS OF SPECTRAL CT

MEASUREMENTS

Page 5: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

What Does the Detector Measure?

Polychromatic Attenuation Formula

Polychromatic Lambert-Beer law contains

Input X-ray tube quanta distribution, S(E),

Spectral responsivity of detector, D(E), and

Spectral object attenuation, µ(E,r).

E

E

rdrE

dEEDES

dEeEDES

I

IA

L

)(

)()(,

0

C. Hofmann, B. Schmidt

5

X-ray source

Detector

Page 6: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Principle of Dual Energy CT

Materials show different attenuation at different mean energies: μ(<E>,r)

10-1

100

101

102

10 30 50 70 90 110 130 150 Energy / keV

Att

en

ua

tio

n

Iodine Bone

56 kV 76 kV

Large increase

Small increase

C. Hofmann, B. Schmidt

6

80kV 140kV 100kV 120kV

iodine

bone

fat

plastic

higher CT-value at 80kV: iodine, bone,

metal ...

higher CT-value at 140kV: fat, plastic,

uric acid ...

(almost) same CT-value: water, soft

tissue, blood ...

water/ soft tissue

Page 7: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

TECHNIQUES TO ACQUIRE

SPECTRAL CT DATA

Page 8: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Spectral Difference Generated by X-ray Source

• Dual Spiral:

Two spiral scans at low and

high kV, respectively.

• Slow kV Switching:

Switch kV level typically once per

gantry rotation (sequence or spiral)

• Fast kV switching:

Switch kV level ~ every millisecond

• Dual Source

Simultaneous scan with 2 tubes

• Split filter

The beam of one source “sees” two

different filters

C. Hofmann, B. Schmidt

8

“Dual kV” “Dual Source”

“Split filter”

Page 9: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Spectral Difference Generated by Detector

• Sandwich Detector:

- first detector layer for low

energy photons

- second detector layer for

high energy photons

• Quantum Counters:

- photon absorbed in

semiconductor (CdTe /

CdZnTe)

- photon energy is measured

- number of photon in each

energy bin is counted

C. Hofmann, B. Schmidt

9

“Energy resolving

detector”

“Photon counting”

Page 10: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

scintillator photodiode

X-rays

Dual kV:

One X-Ray Source

Two scans with different kV or kV-switching (fast or slow) during

one scan is performed

C. Hofmann, B. Schmidt

10

Page 11: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Two Scans with Different kV:

Dual Spiral / Slow kV Switching

140 kV

Switch kV

and mA

for equal dose 80 kV

A (partial) scan is performed with one kV-setting (e. g. 140 kV)

kV and mA are switched

A second (partial) scan is performed at the same z-position,

with the other kV-setting (e. g. 80 kV) and the other mA-setting

C. Hofmann, B. Schmidt

11

Page 12: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Two Scans with Different kV:

Benefits

Simplest approach, providing dual energy for standard CT systems

Good spectral separation

Spectral optimization possible (e. g. by selective pre-filtration – Zn)

Full field of view

No cross-scatter problems

Similar radiation dose at 140 kV and at 80 kV by mA - adaptation

low kV high kV

C. Hofmann, B. Schmidt

12

Page 13: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Two Scans with Different kV:

Challenges

Long duration motion artifacts, registration problems – can be

addressed with registration

Applications with contrast agent limited by blood flow dynamics – only

late phase scans lead to reasonable results

Difference image high kV – low kV

without registration with registration

C. Hofmann, B. Schmidt

13

Page 14: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Spiral Dual Energy – Possible Applications:

Gout

• The spectral behavior of uric acid is different from that of bone.

• Left: CT image with color LUT. Blue: bone / green: uric acid.

C. Hofmann, B. Schmidt

14

Vancouver General Hospital, Canada

Page 15: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Spiral Dual Energy – Possible Applications:

Improved Metal Visualization with Monoenergetic

70

ke

V

10

0k

eV

co

urt

esy o

f R

ich

mo

nd

Dia

gn

osti

c Im

ag

ing

, R

ich

mo

nd

, V

icto

ria, A

ustr

alia

C. Hofmann, B. Schmidt

15

Page 16: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Spiral Dual Energy – Possible Applications:

Kidney Stones

• Discriminate between uric acid stones (dissolvable) and other stones

• Uric acid-containing stones are labelled in red, non uric acid-containing

stones are labelled blue

C. Hofmann, B. Schmidt

16

Klinikum Großhadern

Page 17: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Spiral Dual Energy – Possible Applications

Electron Density and eff. Z

Javier Pena / H IM CR

RO D

Eff. Z map

e density map

Calculation of electron density ρe and

effective Z for dose calculation in radiation

treatment planning!

Page 18: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Fast kV-Switching During One Scan

140 kV 80 kV

The tube voltage (kV) is switched between two readings

(e.g. from 140 kV to 80 kV)

Two „interleaved“ data sets with different kV-settings are

simultaneously acquired

Has already been implemented in a medical CT scanner in 1986

C. Hofmann, B. Schmidt

18

Page 19: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Fast kV-Switching During One Scan:

Benefits

Good spectral separation

Full field of view

No cross-scatter problems

Raw-data based evaluation techniques possible

No motion artifacts, no registration problems due to simultaneous

data acquisition

No problems with varying concentrations of contrast agent

C. Hofmann, B. Schmidt

19

Page 20: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Today: switching every 250 - 500 μs slower rotation (≥ 0.5 -1s)

preferred, challenging for fast moving organs such as lungs and heart

Only kV-switching, no mA-switching equal dose problematic

Way out: 1 reading at 140 kV, ~ 2-3 readings at 80 kV

But: reduced total number of readings

Currently, no anatomical dose modulation possible

No spectral optimization by different pre-filtration possible

Fast kV-Switching During One Scan:

Challenges

0 500 1000 1500 20000

50

100

150

200

kV

Time in ms

140 kV

80 kV

Time in μs

kV

20

Page 21: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Source Dual Energy

C. Hofmann, B. Schmidt

21

Page 22: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Source Dual Energy:

Benefits:

(Nearly) simultaneous data acquisition

Same dose at 140 kV and at 80 kV due to mA – adaptation

Good spectral separation, spectral optimization possible

Short rotation times for fast moving organs possible

Applications with contrast agent are possible also in early phase

due to high temporal resolution and fast acquisition times

Challenges:

Data acquisition not fully

Raw-data based evaluation difficult

Reduced field of view of the second detector

Cross-scattered radiation, in particular for larger patients

C. Hofmann, B. Schmidt

22

Page 23: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Source Dual Energy – Possible Applications:

Virtual Unenhanced CT (Liver VNC)

courtesy of Ludwig-Maximilians-Universität, Klinikum Großhadern, Munich, Germany

C. Hofmann, B. Schmidt

23

• With this approach one can calculate the VNC images which represent the patient without the iodine

enhancement.

• Furthermore, it allows to quantify iodine-uptake, revealing important information if a tumor is benign or malignant.

Page 24: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Source Dual Energy – Possible Applications:

Reliable Head Bone Removal

courtesy of Friedrich-Alexander University Erlangen-Nuremberg - Institute of Medical

Physics / Erlangen, Germany 100kV/Sn140kV

C. Hofmann, B. Schmidt

24

Without any user interaction, bone can be subtracted, also in complicated anatomical

situations like carotids in base of the skull, vertebral arteries etc.

Page 25: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Source Dual Energy – Possible Applications:

Lung Perfused Blood Volume (PBV)

Quantification of iodine to visualize perfusion defects in the lung

Avoids registration problems of non-dual energy subtraction methods

Courtesy of Prof. J and M Remy, Hopital Calmette, Lille, France

80/140kV Mixed Image Mixed image + iodine overlay Iodine Image

Embolus

Page 26: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Split Filter

Moveable

Split Filter

Spectrum after filter

Spectrum before filter

Low Energy

High

Energy

Tube voltage:

120kV

Split filter Gold (Au, 0.05mm)

Tin (Sn, 0.6mm)

C. Hofmann, B. Schmidt

26

Page 27: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Split Filter:

Benefits

(Nearly) simultaneous data acquisition

Short rotation times for fast moving organs possible

Full field of view for both high and low energy

dose modulation (reduction) techniques possible (tube current)

Dose neutral compared to 120 kV

Almost the same applications possible as in Dual Source Dual Energy

Split filter 120 kV

Dose neutral:up to 40cm diameter less noise than 120kV

30cm phantom, default abdomen protocol,

same dose

C. Hofmann, B. Schmidt

27

Page 28: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Split Filter :

Challenges

Data acquisition not fully simultaneous –potential registration problems

Spectral separation not so good – must be compensated with

advanced image filters

Cross-scattered radiation, in particular for larger patients

Spiral mode only

pitch factor is limited to 0.5

High tube power (2/3 of the dose is absorbed in the filter)

C. Hofmann, B. Schmidt

28

Page 29: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Sandwich-type detector, two layers per channel

Detection of lower energy quanta in the top layer

Detection of higher energy quanta in the bottom layer

Dual Layer Detectors

X-rays

scintillator photodiode

scintillator photodiode

absorbed upper layer spectrum 1 mm ZnSe

absorbed lower layer spectrum 2 mm UFC

C. Hofmann, B. Schmidt

29

Courtesy of Steffen Kappler

Page 30: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Layer Detectors:

Benefits

Full field of view

No cross-scatter problems

Raw-data based evaluation possible

Perfect registration due to simultaneous data acquisition

No motion artifacts or problems with varying densities

of contrast agent

Access to dual energy with single-kV scans

No low-energy (80kV) dose problems as with kV-switching

C. Hofmann, B. Schmidt

30

Page 31: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Layer Detectors:

Challenges

(Courtesy S. Kappler, Siemens Healthcare)

Complex technical realization

Reduced dual energy performance compared to dual kV – spectral

separation is limited because there is a spectral overlap over the entire

spectral range

Dual layer overlap over the entire spectral range

Dual kV 80 kV / 140 kV

31

Page 32: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

20%

20%

20%

20%

20%

When was the Dual Energy Technique of

“Rapid kV-Switching” First Realized in a CT

Scanner?

1. 1982

2. 1986

3. 1990

4. 1994

5. 1998

10

Page 33: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

When was the Dual Energy Technique of “Rapid

kV-Switching” First Realized in a CT Scanner?

1.1982

2.1986

3.1990

4.1994

5.1998

Rapid kV switching has already been implemented in a medical CT

scanner in 1986.

Reference: Björn J. Heismann, Bernhard T. Schmidt, Thomas Flohr,

“Spectral CT imaging”, SPIE Press, PM226, October 2012

33

Page 34: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

SPECTRAL CT QUALITY

Page 35: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Spectral Separation

• Very critical for good SNR, separation quality of materials and

robustness!

15mg/ml Ultravist

20ml / 20mm diameter

20cm water phantom

SOMATOM Force highkV

lowkV

CTvalue

CTvalueratio

35

Highest Dual Energy ratio

Iodine Ratio

Page 36: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Mixed Images

DE

Imag

es

VNC Iodine

SD: -35% SD and dose: equal

80

/14

0 k

V

10

0/1

40

Sn

kV

Dual Source CT – Spectral Optimization

C. Hofmann, B. Schmidt

36

Page 37: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Importance of Temporal Resolution & Temporal

Coherence

Temporal resolution:

• time to collect enough raw data for one image ( typically rotation

time / 2)

• determines amplitude of motion artifacts

temporal resolution as high as possible for good quality!

necessary for applications with contrast dynamics and cardiac

applications!

Temporal coherence:

• Differences in the high and low kV images originating from patient

motion due to a temporal delay between the high and low kV image

acquisitions

• Result: Visibly different low & high kV images

temporal coherence as high as possible for good quality!

needed for all DE applications

C. Hofmann, B. Schmidt

37

Page 38: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Importance of Temporal Coherence

• Visibly different low & high kV images (breathing motion /

incomplete breathold, bowel movement)

Low kV Low kV

Without registration With registration

Technique Temporal

Coherence

Temporal

Resolution

Dual Spiral Low High

Slow kV

Switching

Medium High

Fast kV

Switching

Very high Low

Dual Source High High

Split Filter Medium High

Sandwich

Detector

Very high High

Quantum

Counter

Very High High

38

Page 39: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

The Importance of Noise Reduction

Active field of research; achieve good result

quality at single energy dose:

• iterative reconstruction

• non-linear image filters

VNC (direct) VNC (filtered) Kidney stone

(direct)

Kidney stone

(filtered)

C. Hofmann, B. Schmidt

40

Dose neutral VNC only possible with advanced filters!

Page 40: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

20%

20%

20%

20%

20%

For a Dual Source System, Which of the

following tube voltage combinations results in

the best Dual Energy performance (DE ratio)?

1. 80 / 140 kV

2. 100 / Sn140 kV

3. 80 / Sn150 kV

4. 90 / Sn150 kV

5. 100 / Sn150 kV

10 C. Hofmann, B. Schmidt

41

Page 41: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

For a Dual Source System, Which of the following

tube voltage combinations results in the best Dual

Energy performance (DE ratio)?

1. 80 / 140 kV

2. 100 / Sn140 kV

3. 80 / Sn150 kV

4. 90 / Sn150 kV

5. 100 / Sn150 kV

Dual Energy ratio increases with decreasing voltages of the low kV

beam and with increasing voltages of the high kV beam, and they

increase when prefiltration (e.g. tin) is added to the high kV beam.

Reference: Bernhard Krauss, Katharine L. Grant, Bernhard T. Schmidt

and Thomas G. Flohr, “ The importance of spectral separation, an

assessment of dual energy spectral separation for quantitative

ability and dose efficiency“, Investigative Radiology, 50(2),

February 2015.

42

Page 42: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

WHAT ABOUT THE DOSE?

Page 43: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dose Efficiency: Tube Current Modulation?

Not available for all DE

techniques!!!

x

X-ray tube

Detector

a.p.

lateral

a.p.

lateral

a.p.

lateral

a.p.

lateral

0

500

1000

1500

2000

2500

3000

3500

4000

0100200300400500600

table position in mm

att

en

uati

on

I_0 / I

0

50

100

150

200

250

300

350

400

tub

e c

urr

en

t

Attenuation

tube current

Technique Tube Current

Modulation

Dual Spiral Yes

Slow kV

Switching Yes

Fast kV

Switching Problematic

Dual Source Yes

Split Filter Yes

Sandwich

Detector Yes

Quantum

Counter

Research topic

C. Hofmann, B. Schmidt

44

Page 44: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Dual Source DE – Fit for Clinical Routine:

No Compromise in Dose

“Dual energy CT of the chest: how about the dose?”

Invest Radiol. 2010 Jun;45(6):347-53.

RESULTS:

• The effective dose measured with thermoluminescent detectors was

2.61, 2.69, and 2.70 mSv, respectively, for the 140/80 kVp, the

Sn140 /100 kVp, and the standard 120 kVp scans.

• Image noise measured in the average images of the phantom scans

was 11.0, 10.7, and 9.9 HU (P > 0.05).

• The CNR of iodine with optimized image blending was 33.4 at 140/80

kVp, 30.7 at 140Sn/100 kVp and 14.6 at 120 kVp.

C. Hofmann, B. Schmidt

45

Page 45: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

Comparison Definition DS SE / Flash DE

(DE composition 0.6)

Noise & Contrast & Dose equivalent to single energy on Definition DS

(for FAST body bone removal protocol)

Ultravist

NH4Cl Bone MgSO4

NaBr

466.8 466.0

315.3 320.2

256.2 258.8

102.5 96.0 98.4 90.0

27.6 24.6

SD SD

Definition DS SE standard

Body Angio (120kV)

Definition Flash DE

100kV/Sn140kV

CTDIvol=8.65mGy

C. Hofmann, B. Schmidt

46

Page 46: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B

More Dose or Less Dose?

same total dose = half dose per spectrum

mixed image has similar image noise & contrast as single energy

image

Dual Energy analysis does not need more dose

may save dose by omitting scans

Body Region Dual Energy (mGy) Single Energy (mGy)

Abdomen (Kidney) 16.3 14.2

Abdomen (Liver) 17.8 14.2

Thorax (LungPBV) 7.3 7.4

Carotid Angio (Bone

Removal)

8.0 8.1

Body Angio (Bone

Removal)

9.2 8.1

Extremity-Hand (Gout) 8.8 6.7

* CTDIvol for default scan protocols on SOMATOM Definition Flash

47

Page 47: Dual Energy CT Physics: Hardware and Image Quality ...amos3.aapm.org/abstracts/pdf/99-28844-359478... · Dual Energy CT Physics: Hardware and Image Quality Assessment AAPM 2015 B