  • XRF:QuickUserGuideSetup:

    Setupbaseandplugintothewall PlacebatteryintoXRFunit(inthebaseofthehandle) Placesparebatteryintothebatterychargeronthebase Placeunitonbase

    Startup: Pushthegreenbuttononthetopoftheunittoturnonthecomputer Acceptthewarningscreensthatpopupandletunitheatup(about2minutes)

    GetXRFsetupfortesting: Inthemain(Home)screen,selectMode>selectSoils(sothatitishighlightedbluebox)>

    backontheModeScreen,selectModeSetup

    SelectTestConditions>selecttheTest/Tools(blueandredarrowshittingaplate/screwdriver

    andwrench)button

  • > > Setuptesttimes:generallyall3beamsatmax:30isthedefaultweveused>hitOK

    Testing:

    FromtheHomescreenselecttheTestbutton>onthetoprightofthescreen,makesurethatthetriggerisunlockedifitislocked,touchthelocktounlockit

    Test: Locked: Unlocked: Totest,selectthegreenGobuttonorhitthetrigger Takeonereadingonthebasetomakesureyouhavethedesiredresults: Whentestisgoingonmakesurenobodypartsarebelowthebeamendandthereisa1m

    radiusaroundtheXRFclearofpeople(iftheyseethelightsflashing,theyareonthewrongside).XRFdeviceemitsradioactivewavesandcanbeharmfultopeople.

    Whenthetestisdone,thescreenbelowwillappear>keeptrackofthetestnumber(besidethedateingreen(#8inthescreenshotbelow)alongwithwhatacmesamplenumberitcorrespondsto(sample1198784)

    Thereshouldbe3clickswhenthetestisgoingononeafter30seconds(beamchange),oneafter60seconds(beamchange)andoneafter90seconds(beamstopped).Whenlightsstopflashingandthebottomlefthandcornerhasagreenlightandsaysready,theXRFisreadyforthenexttest.

  • CalCheck:

    Ifacalcheckisrequired(generallywhenyouturnontheunitandonceortwiceduringtheday),thecalcheckbuttonwillappearwherethestart/stopbuttonis.InthestatusbartherewillalsobeagreencircleandCalCheckRequiredwrittenthere.

    HittheCalCheckbuttonandwaituntiltestisdone. Ifyoudonotpassthetest,hittheexclamationmark,checkthelogandtestagain Ifyoupassthetest,continueXRFingsamples

    Results:

    SelecttheResultsicon(piechart) SelecttheToolsbutton(screwdriverandwrench) SelecttheoptionsyouwantSelectalltheelements,thesamplesbelowdetection,etc... Whenyouhaveselectedwhatyouareinterestedin,hittheExportbutton

  • Selectthedateoftheresultsthatyouareinterested UsetheToolsbuttontoselecttheparametersyouareexporting.TheSetChemistrybuttonon

    thenextscreenwillallowyoutoselecttheelementsthatyouareexportingalongwiththeexportthreshold.

    Whenyouarereadytogo,hittheExportbutton(paperwiththegreenarrow)

  • TheExportbuttonwillbringyoutotheexportingscreen.Dependingonthesettingsyoumaybeabletorenameyourfile.Thedefaultnamewillbethedateyouareexportingregardlessofhowmanysamplesyouareexportingorfromwhichdays.

    AllowtheexporttocompleteandyouaredonewiththeXRF.

    Gettingresultsontoyourcomputer:

    MakesureWindowsMobileisinstalledonyourlaptop.ThisfreeprogramwillallowyoutoaccesstheXRFsdriverstogetthedata.ThisprogramalsovariesfromWindowsversionsandbitdriverssoyouwillhavetonavigatefromthelinkbeloworGoogleWindowsMobileDeviceCenterWindows(typeyouhave)bit(typeyouhave32or64).http://www.microsoft.com/download/en/windowsMobile.aspx?q=windows+mobile

    PlugtheXRFtoyourcomputer Minimizethewindowsmobilescreenwhenitpromptsyouforsyncing UsingMyComputeraccesstheXRFcomputer. IbelieveyoucanthenfindtheresultsundertheMyDocuments>InnovX>Data>

    Export>thensimplyselectyourdocument,copyandpasteontoyourcomputer.(ThismightnotbethecorrectfilepathasIgenerallyfinditbytrialanderror)

    Checktomakesurethecsvfileiscorrectandthatyouexportedtheelementsthatyouwantinthesettingsthatyouwant.Ifyouhaveanydoubts,gothroughtheExportingprocessagainandedityourselectionsintheToolswindowuntilyouhavetheresultsinadesiredoutlook.

    Shuttingdown: IntheSetupwindowselecttheExitbutton SelectYes,youaresureyouwanttoshutdown Oncetheprogramclosesyoumustshutdowntheoperatingsystem SelectthePowerButtononthetoprighthandsideofthescreenandselectYes,youaresure

    youwanttoshutdown

  • Removethebatteryfromtheunitonceithasshutdown.*NeverstorethebatteryintheXRFUnit*

    Precautionsandhints:

    IftheXRFoverheatsitWILLSHUTDOWNuntilitiscoolenoughtooperatekeepafanonitifitisgettingtoohot

    Ifyouaretestingsomething(Cu)greatthan5%useminingmode.Forsampleslessthan5%usesoilsmode

    TheXRFreadswhatisatthesurfacedirtyorwetsampleswillgivefalsereadingsforthesamplethereforemakesurethesurfaceyouarereadingiscleanofoxides,dust,dirtandliquids(Withthatsaid,thebeamwillstillpenetratethroughthesample.Thebeammaynotbeabletorefractbacktothedetector)Moisture>20%=1:1errorratio

    IncorporateyourownQA/QCtoensuretheXRFisworkingasitissupposetotakereadingofthecalibrationpuck(silverdollartypethingseparateinthecaseorontheXRFchargingbase).

    Whenyouneedtochangethebattery,youhaveover30secondstohotswapitwithouthavingtoshutdownandstartupagain.Makesureyouarereadywithyourreplacementbattery.*Lockthetriggerwhendoingthis*

    PasswordsforFactorysettingwillbeoneofthefollowing:o Username:ADMI Password:1234o Username:ADMINISTRATOR Password:Z

    TheXRFisacomputerifyouarehavingissues,exittheXRFprogramandopenitagain,orrestarttheXRF.Donottakeoutthebatterybeforeyouhaveproperlyshutdowntheunit!

    Ifyoudocannotfindtheanswerinthisguide,refertotheusermanual