Download - Spectraa 240

Transcript
  • Varian AA systems

    Pre-installation manual

    Installation Category II Pollution Degree 2 Safety Class 1 (EN 61010-1)

    8510119300 October 2003

  • Varian AA Pre-installation

    ii Publication date: October 2003

    Varian offices Varian has offices in most countries. The major offices for optical spectroscopy products are listed below:

    Varian Australia Pty Ltd (Manufacturing site) 679 Springvale Road Mulgrave, Victoria 3170 Australia International telephone: + 61 3 9560 7133 International fax: + 61 3 9560 7950

    Varian Instruments 2700 Mitchell Dr Walnut Creek, CA 94598 Telephone: 1 800 926 3000 International telephone: +1.925.939.2400 International fax: +1.925.945.2102

    Varian BV Herculesweg 8, 4338 PL, Middelburg, Netherlands International telephone: +31 118 67 1000 International fax: +31 118 62 3193

    Internet The Varian Internet home page can be found at: http://www.varianinc.com

    Varian Australia Pty Ltd is the owner of copyright on this document and any associated software. Under law, the written permission of Varian Australia Pty Ltd must be obtained before either the documentation or the software is copied, reproduced, translated or converted to electronic or other machine-readable form, in whole, or in part.

    First published 1994 in Australia. Updated 1995, 1997, 1998, 2000, 2001, 2002, 2003. Comments about this manual should be directed to the Marketing Communications Manager, Varian Australia at the address above or to [email protected].

    Varian Australia is ISO9001 certified.

    1994 Varian Australia Pty Ltd (A.B.N. 67 004 559 540) All rights reserved

    Microsoft, Windows and Windows NT are registered trademarks of Microsoft Corporation in the United States and other countries.

  • Varian AA Pre-installation

    Publication date: October 2003 iii

    Contents

    1. Introduction..............................................................................1-1

    1.1 Installation guide 1-1

    1.2 Recommended PC configurations 1-2

    1.3 Warning and Caution messages 1-6

    2. Suitable work area ...................................................................2-1

    2.1 Environment control 2-1

    2.2 Workbench 2-2

    2.3 Weights and dimensions 2-3

    3. Exhaust System........................................................................3-1

    3.1 General information 3-1

    3.2 Varian exhaust system 3-2

    4. Electrical power supplies .........................................................4-1

    4.1 General information 4-1

    5. Gas supplies..............................................................................5-1

    5.1 Storage of pressure cylinders 5-1

    5.2 Gas connection fittings 5-2

    5.3 Permitted gas supplies 5-2

    5.4 Air supply 5-3

    5.5 Nitrous oxide supply 5-5

    5.6 Acetylene supply 5-6

    5.7 Gas supplies for Graphite Tube Atomizers (both Zeeman and non-Zeeman) 5-8

    5.8 Gas supplies for the Vapor Generation Accessory 5-9

    6. Supplementary information ....................................................6-1

    6.1 Waste fluids 6-1

    6.2 Special arrangements for use with organic solvents 6-1

  • Varian AA Pre-installation

    iv Publication date: October 2003

    7. Water cooling system .............................................................. 7-1

    7.1 General information 7-1

    8. Equipment on-site ................................................................... 8-1

    8.1 Insurance after delivery 8-1

    8.2 Transit damage 8-1

    8.3 In-house transit routes 8-2

    9. Guidelines for software installation........................................ 9-1

    10. Operator training ...................................................................10-1

    11. Appendix................................................................................11-1

    11.1 Weights and dimensions of SpectrAA-880/880Z and GTA-100.11-1

    11.2 Electrical power supplies 11-3

    11.3 In-house transit routes 11-4

  • Varian AA Pre-installation

    Publication date: October 2003 1-1

    1. Introduction

    This publication contains information that you need to enable you to prepare your laboratory for the installation of a Varian atomic absorption spectrometer. Described below are facilities that must be provided before the installation can be completed. Detailed operating procedures are provided in the operation manual that is supplied with the instrument or in the multimedia online help, supplied with the instrument software.

    When all preparations are finished, please complete the checklist (strike out those entries not applicable) and send the completed, signed pre-installation checklist to your local Varian sales and service office or Varian agent. When it is received, the Varian field service manager or Varian agent will contact you to arrange a mutually convenient time for installation.

    If you need assistance or details of operator training courses, please contact your Varian representative.

    1.1 Installation guide Allow a minimum of four hours for installation by a Varian customer service representative. If accessories are included, allow a maximum of eight hours. The installation will include:

    Spectrometer installation

    Basic customer training

    Maintenance overview

    Accessory installation

    We recommend a telephone outlet near the instrument. This will make it easier for you if you need to seek assistance in operational procedures or apparent fault diagnosis. This can also facilitate remote troubleshooting and diagnosis where this support option has been included.

    Note Installation time will increase if Qualification services (IQ/OQ) are

  • Varian AA Pre-installation

    1-2 Publication date: October 2003

    required. Contact your local Varian customer service representative for further details.

    1.2 Recommended PC configurations If you are supplying your own PC for use with a Varian AA instrument, the minimum and recommended configurations of that PC are listed below. The minimum configuration represents the absolute minimum on which you can run the Varian AA instrument software. This PC configuration may be out of manufacture, but you may wish to use a PC you already have. The recommended configuration is that which you would buy new.

    Minimum: Intel Pentium III 500 MHz processor, 256 MB RAM, 17 inch Super VGA monitor, graphics card supporting 1024 x 768 resolution, 4 GB hard disk or bigger, CD-ROM drive, sound card, Windows NT 4 (Service Pack 6) or Windows 2000 (Service Pack 2) or Windows XP (Service pack 1) operating system, one RS232 serial port.

    Note The PC must have 1 spare PCI expansion slot for the Varian PCI-IEEE interface card.

    Recommended: Intel Pentium IV 2 GHz processor, 512 MB RAM, 17 inch Super VGA monitor, graphics card supporting 1024 x 768 resolution, 40 GB hard disk or bigger, 24 speed CD-ROM drive, 32 bit sound card and speakers, Windows NT 4 (Service Pack 6) or Windows 2000 (Service Pack 2) or Windows XP (Service pack 1) operating system, two RS232 serial ports.

    Note The PC must have 1 spare PCI expansion slot for the Varian PCI-IEEE interface card.

    The Varian SpectrAA software does not support the AT/ISA PC bus.

  • Varian AA Pre-installation

    Publication date: October 2003 1-3

    1.2.1 Pre-installation checklist Use the following checklist to ensure that you complete each step of the pre-installation procedure. Tick the box to indicate that each requirement is complete.

    Instrument delivered, unpacked and positioned on work bench Any instrument damage has been reported to the company

    responsible for transporting the instrument, a Varian representative was informed of the damage and a copy of the damage report has been sent to the Varian office.

    Computer components on site Brand:

    Processor:

    RAM size in MB:

    Hard disk size in MB:

    Operating system:

    Printer Brand:

    Model number:

    SIPS 10/20 Sample Introduction Pump System Serial number: SPS Autosampler

    Model number:

    Serial number:

    VGA-77 Vapor Generation Accessory Serial number: GTA Accessory Serial number: ETC-60 Accessory Serial number:

  • Varian AA Pre-installation

    1-4 Publication date: October 2003

    Additional accessories Type and model:

    Serial number:

    Type and model:

    Serial number:

    Exhaust system Electrical power supplies Gas supplies Telephone outlet Waste container Water cooling/circulation system

    Brand:

    Model number:

    Serial number:

    1.2.1.1 21 CFR Part 11 Pre-installation requirements The 21 CFR Part 11 booklet has been received and read, and

    action has been taken to ensure section 4 of the 21 CFR Part 11 booklet has been completed

    System Administrator has been notified Arrangements have been made for a System Administrator to be

    present during installation

    All preparations have been completed in accordance with the information provided in the Varian AA pre-installation manual. Please arrange for the installation to be completed as soon as possible. I understand that if the installation site is not prepared in accordance with the enclosed instructions, additional installation charges may be applicable.

    Company name and address _____________________ ___________________________________

  • Varian AA Pre-installation

    Publication date: October 2003 1-5

    Name ______________________________ Position ____________________________________

    Telephone ___________________________ Date ______________________________________

    Preferred installation date _______________ Signature ___________________________________

    1.2.2 Pre-installation complete All preparations have been completed. To arrange installation of the instrument, complete the details on the pre-installation checklist and return the form to the Varian Office shown below. A copy should be retained for Validation purposes.

    Varian office card to be attached here.

  • Varian AA Pre-installation

    1-6 Publication date: October 2003

    1.3 Warning and Caution messages A Warning message is used in the text when failure to observe instructions or precautions could result in death or injury.

    A Caution message is used when failure to observe instructions could result in damage to equipment (Varian supplied and/or other associated equipment).

    A Note message is used to give advice or information.

  • Varian AA Pre-installation

    Publication date: October 2003 1-7

    1.3.1 Warning symbols These symbols are included with warnings in the manual or on the products.

    Electrical shock

    Eye hazard

    Noxious gases

    Hot surfaces

    Fire hazard

    Explosion hazard

    Corrosive chemical

    Moving part

    Heavy weight (Danger to feet)

    Toxic material

    Sharp objects

    Magnetic fields

    Heavy weight (Danger to hands)

  • Varian AA Pre-installation

    1-8 Publication date: October 2003

    This page is intentionally left blank.

  • Varian AA Pre-installation

    Publication date: October 2003 2-1

    2. Suitable work area

    2.1 Environment control The area selected for operation of a Varian AA spectrometer system must be free from drafts, corrosive atmospheres and vibration. Sample preparation areas and materials storage facilities should be located in a separate room.

    The area should have a dust-free, low humidity atmosphere. Air-conditioning is strongly recommended for control of the environment. The instrument should not be located near a window, door or any other area where drafts may cause unstable thermal conditions.

    Recommended environmental conditions:

    For operation: 0 to 853 m (0 to 2800 ft) 853 to 2133 m (2800 to 7000 ft)

    10 to 35 C (50 to 95 F) 10 to 25 C (50 to 77 F)

    8 to 80% relative humidity 8 to 80% relative humidity (non-condensing) (non-condensing)

    For storage: 0 to 2133 m (0 to 7000 ft)

    5 to 45 C (40 to 115 F)

    20 to 80% relative humidity (non-condensing)

    Note For optimum analytical performance, it is recommended that the ambient temperature of the laboratory be maintained between 2025 C, and held constant to within 2 C throughout the entire working day.

  • Varian AA Pre-installation

    2-2 Publication date: October 2003

    2.2 Workbench The workbench must be free from vibration and be stable and strong enough to support the total weight of the equipment to be used. The bench top should be large enough to permit a free circulation of air around the main instrument and each of the accessories.

    The information provided in table 2.1 will make planning easier. Portable or semi-permanent trolleys can be used as workbenches but you must lock the wheels. Accessories such as the SPS Autosampler, can be positioned on a trolley available from Varian (part number 9910056300). To avoid damage through spillage of samples being used, the bench top should be covered with a material which is corrosion resistant and impervious to liquid spillage. Generally, for comfortable working conditions and ease of access to the instruments, the workbench should be approximately 700 mm (28") high.

    The location of the workbench may be constrained by the need for an exhaust flue to remove fumes and vapors from the spectrometer sample compartment (see Chapter 3).

  • Varian AA Pre-installation

    Publication date: October 2003 2-3

    2.3 Weights and dimensions

    Spectrometer System unit Width Depth Height Weight

    SpectrAA-50/55/110/220 instrument only

    790 mm (31")

    585 mm (23")

    575 mm (22.5")

    56 kg (123 lb)

    SpectrAA-50/55/110/220 instrument and shipping box

    1200 mm (47")

    780 mm (31")

    870 mm (34")

    97 kg (213 lb)

    Varian-AA140/240 instrument only 780 mm (31")

    600 mm (24")

    735 mm (29")

    69 kg (152 lb)

    Varian-AA140/240 instrument and shipping box

    1170 mm (46")

    890 mm (35")

    890 mm (35")

    106 kg (234 lb)

    Varian-AA280 instrument only 780 mm (31")

    600 mm (24")

    735 mm (29")

    69 kg (152 lb)

    Varian-AA280 instrument and shipping box

    1170 mm (46")

    890 mm (35")

    890 mm (35")

    106 kg (234 lb)

    Accessories

    GTA-110 Graphite Tube Atomizer (workhead stowed)

    530 mm (21")

    530 mm (21")

    400 mm (16")

    43 kg (95 lb)

    Shipping dimensions 1000 mm (40)

    710 mm (28.4)

    700 mm (28)

    71 kg (156 lb)

    GTA-110 Zeeman

    (including workhead)

    440 mm (18")

    540 mm (22")

    300 mm (12")

    52 kg (114 lb)

    Shipping dimensions 1000 mm (40)

    710 mm (28.4)

    700 mm (28)

    79 kg (174 lb)

    Programmable Sample Dispenser PSD-110

    305 mm (12.2)

    360 mm (14.4)

    310 mm (12.4)

    6.5 kg (14 lb)

    Shipping dimensions 620 mm (24.8)

    420 mm (16.8)

    325 mm (13)

    11.5 kg (25 lb)

    GTA-120 Graphite Tube Atomizer (workhead stowed)

    240 mm (10")

    600 mm (24")

    580 mm (23")

    37 kg (82 lb)

    Shipping dimensions 950 mm (37)

    760 mm (30)

    870 mm (34)

    76 kg (167 lb)

    includes PSD-120

  • Varian AA Pre-installation

    2-4 Publication date: October 2003

    GTA-120 Zeeman

    (including workhead)

    240 mm (10")

    600 mm (24")

    580 mm (23")

    48 kg (106 lb)

    Shipping dimensions 950 mm (37)

    760 mm (30)

    870 mm (34)

    87 kg (192 lb)

    includes PSD-120

    Programmable Sample Dispenser PSD-120

    300 mm (11.8)

    380 mm (15)

    310 mm (12.2)

    6 kg (13.2 lb)

    Shipping dimensions 660 mm (26)

    420 mm (16.5)

    310 mm (12.2)

    10 kg (22 lb)

    SPS2 Autosampler 415 mm (16")

    265 mm (10")

    435 mm (17")

    11 kg (24.2 lb)

    Shipping dimensions 760 mm (29.9)

    500 mm (19.7)

    840 mm (33.1)

    29 kg (63.9 lb)

    SPS3 Autosampler 490 mm (19)

    285 mm (11)

    515 mm (20)

    15 kg (33 lb)

    Shipping dimensions 760 mm (29.9)

    500 mm (19.7)

    840 mm (33.1)

    31 kg (68.3 lb)

    SPS3 Autosampler with Diluter 572 mm (23)

    285 mm (11)

    515 mm (20)

    18 kg (39.6)

    Shipping dimensions 760 mm (23.6)

    500 mm (15.7)

    840 mm (23.6)

    34 kg (74.9 lb)

    Trolley cart dimensions 600 mm (23.6")

    420 mm (16.5")

    775 mm (30.5")

    VGA-77 Vapor Generation 310 mm (13")

    210 mm (8")

    270 mm (11")

    5.5 kg (12 lb)

    Shipping dimensions 385 mm (15)

    340 mm (14)

    400 mm (16)

    8.5 kg (19 lb)

    ETC-60 Electrothermal Controller 260 mm (10")

    240 mm (9.5")

    150 mm (6")

    5 kg (11 lb)

    Shipping dimensions 385 mm (15)

    340 mm (14)

    400 mm (16)

    8.5 kg (19 lb)

  • Varian AA Pre-installation

    Publication date: October 2003 2-5

    SIPS -0/20: SIPS Power Module 225 mm (9)

    100 mm (4)

    385 mm (15.5)

    4.5 kg (10 lb)

    SIPS Pump module 285 mm (11)

    275 mm (10.5)

    215 mm (9)

    4.5 kg (10 lb)

    Shipping dimensions 620 mm (25)

    530 mm (21)

    360 mm (14)

    15 kg (33 lb)

    Personal Computers Consult the manufacturer.

    Printers Consult the manufacturer.

    5BCMF Equipment weights and dimensions for Varian AA instruments.

  • Varian AA Pre-installation

    2-6 Publication date: October 2003

    'JHVSF SpectrAA-50/55/110/220 instrument dimensions

  • Varian AA Pre-installation

    Publication date: October 2003 2-7

    'JHVSFFront view of a Varian-AA280 series instrument with GTA-120 and PSD-120.

    'JHVSFTop view of a Varian-AA280 series instrument, GTA-120 and PSD-120.

  • Varian AA Pre-installation

    2-8 Publication date: October 2003

    'JHVSF Side view of the Varian-AA280 with PSD-120.

    'JHVSF Front of a Varian-AA140/240 series instrument with GTA-120 and PSD-120.

  • Varian AA Pre-installation

    Publication date: October 2003 2-9

    'JHVSF Top view of a Varian-AA140/240 series instrument with GTA-120 and PSD-120.

    'JHVSF Side view of a Varian-AA140/240 with PSD-120.

  • Varian AA Pre-installation

    2-10 Publication date: October 2003

    This page is intentionally left blank.

  • Varian AA Pre-installation

    Publication date: October 2003 3-1

    3. Exhaust System

    'JHVSFSpectrometer and flue position

    3.1 General information The flame operates at a temperature of approximately 3000 C (5400 F), and can generate up to 540 kilojoules (510 BTU) per minute. Exhaust fumes can be toxic or corrosive.

    The spectrometer must be located under a flue, which is vented by an exhaust fan and ducted to an external vent. The exhaust system with flue, ducting and external vent, must provide a minimum flow of 6 cubic metres per minute (200 cfm) at 16 mm water gauge static pressure (free delivery of 11 cubic metres per minute).

    The exhaust system installation must comply with any rules and/or regulations that may be imposed by local authorities responsible for control of facilities and fixtures in the work place.

  • Varian AA Pre-installation

    3-2 Publication date: October 2003

    The exhaust fan should be located at least 3 metres (10 feet) away from the flame. The fan blades must be made of metal. The fan control switch and running indicator lamp should be located close to the instrument.

    Ducting must be corrosion resistant, fireproof and should be kept clear of fire alarms, sprinkler heads, heat sensitive devices and combustible materials. It should rise vertically for at least two metres (6 feet) from the spectrometer and there should be no tight bends. All ducting joints must be fitted the hot exhaust gases may melt soldered joints.

    The external vent must be fitted with a back draft damper and the outlet location must be clear of doors, windows and air-conditioning inlets.

    Warning An exhaust system must be used with Varian AA flame and furnace instruments to remove hazardous and toxic gases.

    3.2 Varian exhaust system The components required for an extraction system may be purchased from Varian either separately or as a kit of parts ready for installation at the instrument site. To allow for personal preferences in the selection of control gear, switch and pilot light assemblies are not included with the items supplied by Varian.

    Exhaust kit Part number Exhaust kit for 240 V, 50 Hz supply 9910028100 Exhaust kit for 115 V, 60 Hz supply 9910028500

    5BCMFExhaust kits.

    Note Mounting hardware for the flue and fan are not included with the Varian supplied exhaust kit.

    Each Varian supplied exhaust kit contains the following items, which must be installed by local fitters before the Varian Customer Service Representative is called.

  • Varian AA Pre-installation

    Publication date: October 2003 3-3

    Item Remarks Part number Qty

    1 Exhaust fan (for 240 V, 50 Hz supply only) 1810000100 1 OR Exhaust fan (for 115 V, 60 Hz supply only) 1810000200 1 2 Flue without mounting brackets 5510000300 1 3 Flexible ducting, 1 metre aluminium 0410000400 3 4 Coupling for joints in ducting 1610000200 2 5 Ducting clamp, 2 per coupling 0810000300 4

    5BCMFExhaust kit components (refer to Figure 3.2).

    Ducting

    Clamps

    Clamps

    Flue

    Coupling

    Coupling

    Exhaust fan

    Remote switch

    'JHVSFExhaust kit components.

    Part number 1810000100 1810000200 Voltage 240 Volts Single Phase 115 Volts Single Phase Current 0.45 Amps 0.7 Amps Frequency 50 Hz 60 Hz Power input 74 Watts 51 Watts Direction of rotation Anti-clockwise (as seen from shaft end) Nominal speed 1300 rpm

    5BCMFExhaust fan motor specifications.

  • Varian AA Pre-installation

    3-4 Publication date: October 2003

    'JHVSFExhaust fan wiring information.

    Note The Varian supplied exhaust fan must be configured to operate in permanent split phase mode as shown in Figure 3.3.

  • Varian AA Pre-installation

    Publication date: October 2003 3-5

    'JHVSF Exhaust fan wiring information for 115VAC, 60Hz motor.

    Note The direction of motor rotation (Clockwise or Counter-Clockwise) is as viewed from the shaft end of the motor. The exhaust fan rotation is counter-clockwise (CCW) and the motor must be wired accordingly

  • Varian AA Pre-installation

    3-6 Publication date: October 2003

    This page is intentionally left blank.

  • Varian AA Pre-installation

    Publication date: October 2003 4-1

    4. Electrical power supplies

    4.1 General information The installation of electrical power supplies must comply with the rules and/or regulations imposed by the local authorities responsible for the use of electrical energy in the work place.

    All Varian AA instruments are supplied with a 2 metre (6 6") mains power cord terminated as indicated in table 4.2.

    All power supplies should be single phase AC, 3 wire system (active, neutral and ground, or two actives and ground) and should be terminated at an appropriate connection receptacle that is within reach of the system power cable assembly. In areas where 208/220/240 Volt supplies are not normally available in a single phase configuration, supplies may be taken from two phases and ground of a three phase system.

    A separate connection receptacle should be provided for each unit in the system (see Table 4.1). Do not use double adapters or extension cords.

    A separate mains circuit individually protected by fuses or circuit breakers must be used for the GTA accessory. It is preferable for the GTA and the instrument to share the same phase.

    Note If the system being installed is a Zeeman system, then two separate mains circuits individually protected by fuses or circuit breakers must be used one each for the instrument and the Zeeman GTA accessory. It is preferable for the GTA and the instrument to share the same phase but separate power supply circuits.

    Avoid using power supplies from a source that may be subject to electrical interference from other services (large electric motors, elevators, welders and air conditioning units etc.).

  • Varian AA Pre-installation

    4-2 Publication date: October 2003

    System unit Required supply voltage Rating

    SpectrAA-50/55/110/220 100 VAC +10% -5% 120, 220 or 240 VAC 10% 230 VAC +14% -6% 230 VAC +6% -14% 50 or 60 Hz 1 Hz

    170 VA

    Varian-AA240 100 VAC +10% -5% 120, 220 or 240 VAC 10% 230 VAC +14% -6% 230 VAC +6% -14% 50 or 60 Hz 1 Hz

    230 VA

    Varian-AA280 100 VAC +10% -5% 120, 220 or 240 VAC 10% 230 VAC +14% -6% 230 VAC +6% -14% 50 or 60 Hz 1 Hz

    230 VA

    Zeeman version

    SpectrAA-220 Zeeman 208-240 VAC 10 % 50 or 60 Hz 1 Hz

    3500 VA **

    Varian-AA240 Zeeman 208-240 VAC 10 % 50 or 60 Hz 1 Hz

    3500 VA **

    Varian-AA280 Zeeman 208-240 VAC 10 % 50 or 60 Hz 1 Hz

    3500 VA **

    GTA Accessory

    GTA-110

    GTA-120

    208, 220 or 240 VAC 10% 230 VAC +14% -6% 230 VAC +6% -14% 50 or 60 Hz 1 Hz

    3500 VA **

    SIPS-10/20 100-240 VAC 10 % 55 VA

    SPS5 Autosampler 100, 120, 220 or 240 VAC 10 % 230 VAC +14% -6%, 230 VAC +6% -14% 50 or 60 Hz 1 Hz

    80 VA

    SPS2/3 Autosampler 100-240 VAC

    50-60 Hz 1 Hz

    50 VA (SPS2)

    75 VA (SPS3)

  • Varian AA Pre-installation

    Publication date: October 2003 4-3

    VGA-77 100, 120, 220 or 240 VAC 10 % 230 VAC +14% -6% 230 VAC +6% -14% 49 61 Hz

    24 VA

    ETC-60 110 120, 220 240 VAC 10 % 50 or 60 Hz 3 %

    550 VA max.

    UltrAA lamp control module

    100, 120, 220 or 240 VAC 10 %

    230 VAC +14% -6%, 230 VAC +6% -14%

    50-60 Hz 1 Hz

    150 VA

    Personal computer (external)

    Consult the manufacturer

    PC monitor Consult the manufacturer

    Printer Consult the manufacturer

    5BCMF Electrical specifications for instruments.

    ** In normal operation, the Zeeman and GTA units will draw surge currents in excess of the nominal rating. Power supplies to these units must be isolated from other supplies to the system and should include delayed action protection devices such as circuit breakers or motor start fuses.

    Surges in the Zeeman instrument occur in the "read" cycle during atomization and may be up to 28 A for up to 10 seconds in any 1 to 2 minute period.

    Surges in the GTA accessory depend on the choice of ramp rate and programmed temperature and may be up to 35 A for approximately 1 second, reducing to about 20 A for up to 10 seconds, perhaps repeating every 1 to 2 minutes.

  • Varian AA Pre-installation

    4-4 Publication date: October 2003

    Instrument & accessories AA Instrument GTA-110/120, Zeeman

    Plug supplied

    Australia -00 10 A, 250 VAC Complies with AS3112

    Clipsal 439D15M

    USA -01 Complies with NEMA 5-15P Complies with NEMA L6-30P (Hubbell #2621)

    Canada 01 Complies with NEMA 5-15P 20 A, 250 VAC, Complies with NEMA L6-20P (Hubbell #2321+)

    Europe -02 Perena 3410 (Complies with CEE 7 Sheet VII or NFC 61.303)

    Kaiser CEBEC 616 VDE (Complies with DIN 49441R2)

    Type of wall socket required

    Australia -00 General purpose 10 A 250 V outlet (HPM 787, Clipsal 15)

    Dedicated circuit, 15 A 250 V outlet (HPM 787/15, Clipsal 15/15

    USA -01 To comply with NEMA 5-15R (15 A supply) (Hubbell IG 5262)

    To comply with NEMA L6-30R (30 A supply) (Hubbell #2626)

    Canada 01 To comply with NEMA 5-15R (15 A supply) (Hubbell IG 5262)

    20 A, 250 VAC, Complies with NEMA L6-20R (Hubbell #2326+)

    Europe -02 To comply with CEE 7 standard No.7 Sheet VII, or Norma Francais C61.303 Sheet V.A.

    No standard known (Kaiser CEBEC 702 type 31/131.5)

    Power supply, current rating and overload protection

    Between 5 and 20 A Between 30 and 40 A

    Power supply Single phase Single phase

    5BCMFSpectrometer power connections.

  • Varian AA Pre-installation

    Publication date: October 2003 5-1

    5. Gas supplies

    All gas supply installations must comply with rules and/or regulations that are imposed by the local authorities responsible for the use of compressed gas energy in the workplace. The following points should be considered.

    5.1 Storage of pressure cylinders Cylinders containing gas under pressure should be firmly secured to a rigid structure, and the storage area must be adequately ventilated. Acetylene cylinders should always be stored and moved in an upright position to avoid acetone saturation of the fibrous safety disc and acetone contamination of the acetylene flow. Acetone contamination will cause erratic analytical results and may damage the instrument. Such damage is not covered by the instrument warranty.

    Never locate gas cylinders near a source of ignition, or in a position that is subject to direct heat. Gas storage cylinders often incorporate a pressure release device which will discharge the gas at a predetermined temperature, usually around 52 C (125 F).

    If gases are to be plumbed from a remote storage area to the instrument site, ensure that the local outlets are fitted with stop valves and pressure gauges which are easily accessible to the instrument operator. Stop valves must be rapid action shut off and should be mounted near the side of the instrument, not behind it and never near or above the sample compartment. The gas outlets must be provided within 1.5 metres (5 feet) of the instrument.

    If a pressure regulator is used at a local outlet, it is necessary to ensure that its input pressure is always at least 40% greater than its output pressure under all combinations of input pressure and gas flow. The effect of supplying multiple instruments through long supply pipelines must be taken into consideration.

  • Varian AA Pre-installation

    5-2 Publication date: October 2003

    5.2 Gas connection fittings Gas fittings may vary among suppliers. If you have difficulty obtaining the correct fittings for direct connection of gas supplies to the spectrometer, please contact your Varian Sales and Service Office for assistance.

    5.3 Permitted gas supplies Only three gases are specified for use in a flame instrument. They are air, nitrous oxide and acetylene. Varian will not install a flame instrument where other gases may be provided for connection to it. The GTA accessory can use air or other gases as an alternate gas but does not use nitrous oxide and acetylene.

    In planning your gas reticulation scheme, please observe the following points:

    Never run rubber tubing through a wall or across a floor. Never use rubber tubing outdoors. Acetylene tubing must be stainless steel or black iron. Use only refrigeration grade copper pipe for other gases

    plumbing grade has oil and grease residues.

    Do not cut off the factory-fitted hose connectors. These are needed to connect to test equipment gas-handling jigs. Their removal will cause extra expense during service calls.

    Use only 3/8" tube or 1/2" pipe to minimize pressure drop.

    Fit pressure gauges on a wall at the supply point, and ensure that they are visible to the operator.

    Place cut-off valves in a position that can be reached safely in case of a fire.

    Warning Varian AA instruments are designed for flame operation with acetylene and air, or acetylene and nitrous oxide, only. The use of unspecified gases for flame operation of the spectrometer is extremely hazardous. You will risk fire and/or explosion, which could result in death or serious injury to personnel, and damage to property. Use only the specified gases and ensure that the installation complies with the relevant rules and regulations as laid down by the appropriate authorities in your region.

  • Varian AA Pre-installation

    Publication date: October 2003 5-3

    5.3.1 Diameters of gas lines All gases should be supplied through gas lines with an internal diameter no smaller than 6.4 mm.

    5.4 Air supply Quality Clean, dry, free of oil Permissible pressure range 245 to 455 kPa (35 to 65 psi) Recommended pressure 350 kPa (50 psi) Normal flow rate 13.5 to 20 litres per minute

    Warning Air enriched with oxygen, or pure oxygen, must NEVER be used with a flame instrument as it will cause a flashback.

    The instrument air hose is fitted with a US Standard Compressed Air Connection, Female Union, Size 9/16" - 18 UNF. The hose is 1.8 metres (6 feet) long. Connection adaptors are supplied to suit different locations as illustrated in figure 5.1.

    The air supply must be regulated to maintain the instrument operating pressure over the full range of normal flow rates. A dual stage regulator should be used. This is most important for instruments with programmable flow rates because a slow response from the regulator may cause the instrument to shut down during a demand for increased flow. A shutdown of this kind is safe and does not imply a fault in the instrument, but will interrupt your work.

    If you choose to use an air compressor, it must be installed in a remote location where its intake is fed from a dry, dust and vapor free environment. The control switch and pilot light should be mounted close to the instrument.

    Where the air supply is derived from a general purpose in-house system or from a dedicated air compressor, a filter unit must be installed. If your location has high humidity, the air supply should include a dryer or condenser device.

  • Varian AA Pre-installation

    5-4 Publication date: October 2003

    'JHVSFAir supply connection adaptors.

    Part number Description 0110265400 230 V 50 Hz Compressor with Air Service Unit 0110265600 115 V 60 Hz Compressor with Air Service Unit 0110209300 Air Service Unit (includes air filter, pressure regulator and emergency air reservoir) 9810006400 Air filter maintenance kit

    'JHVSFGas supplies.

  • Varian AA Pre-installation

    Publication date: October 2003 5-5

    5.5 Nitrous oxide supply Quality Instrument-grade, 99.5% pure

    (minimum) Permissible pressure range 245 to 455 kPa (35 to 65 psi) Recommended pressure 350 kPa (50 psi) Normal flow rate 11 to 16 litres per minute

    Note Instrument grade nitrous oxide must be used. If you choose to use industrial grade, instrument performance is not guaranteed.

    The instrument nitrous oxide hose is fitted with a US Standard Nitrous Oxide Gas Connection, Female Union, Size 3/4" - 16 UNF. The hose is 1.8 metres (6 feet) long. Connection adaptors are supplied to suit different locations as illustrated in Figure 5.3.

    The nitrous oxide supply must be regulated to maintain the instrument operating pressure over the full range of normal flow rates. A dual stage regulator should be used. This is most important for instruments with programmable flow rates because a slow response from the regulator may cause the instrument to shut down during a demand for increased flow. A shutdown of this kind is safe and does not imply a fault in the instrument, but will interrupt your work.

    Nitrous oxide is stored as a liquid. Evaporation across the regulator may cause it to freeze at high flow rates. This can cause erratic flame conditions and in extreme cases may cause a flashback. You can purchase a preheater, which will prevent freezing. Your Varian Sales and Service Office can advise you and can supply these devices:

    Region Description Part number North America Regulator with integral 2759071000 thermostatted heater Continental Europe Heater, for connection 5710007200 between cylinder and regulator

  • Varian AA Pre-installation

    5-6 Publication date: October 2003

    'JHVSFNitrous oxide supply connection adaptors.

    5.6 Acetylene supply Quality Instrument-grade, 99.0% pure (minimum) Permissible pressure range 65 to 100 kPa (9.5 to 14.5 psi) Recommended pressure 75 kPa (11 psi) Normal flow rate 0 to 10 litres per minute Packaging solvent Acetone

    'JHVSFAcetylene supply connection adaptors.

    The instrument acetylene hose is fitted with a US Standard Acetylene Gas Connection, Female Union, Size 9/16" - 18 UNF, Left-hand thread. The hose is 1.8 metres (6 feet) long. Connection adapters are supplied to suit different locations as illustrated in Figure 5.4.

    The acetylene supply must be regulated to maintain the instrument operating pressure over the full range of normal flow rates. A dual stage regulator should be used. This is most important for instruments with programmable flow rates because a slow response from the

  • Varian AA Pre-installation

    Publication date: October 2003 5-7

    regulator may cause the instrument to shut down during a demand for increased flow. A shutdown of this kind is safe and does not imply a fault in the instrument, but will interrupt your work.

    Acetone, the acetylene packaging solvent, can damage neoprene, rubber and plastic components of the instrument. It can also absorb ultraviolet light and may affect analytical performance. High levels of acetone will reduce the flame temperature, which may affect analytical results at many wavelengths. Acetone may be entrained in the acetylene gas at low cylinder pressure.

    Note You must replace the cylinder when its pressure falls to 700 kPa (100 psi) to prevent acetone from being entrained in the acetylene gas and causing damage to the instrument. Acetone may also be entrained if the flow rate exceeds 1/7th of the cylinder contents per hour. Your supplier can advise you on an appropriately rated cylinder.

    Instrument-grade acetylene is free of phosphorus, arsenic and sulphur contaminants, which are residues of the acetylene generation process (calcium carbide and water). Phosphine content is also minimized. Phosphine contamination will depress calcium, strontium and barium absorption signals. Broadband emission from phosphine combustion products can interfere with sodium and potassium analysis.

    Acetylene is unstable as a free gas under pressure and it should not be piped at pressures exceeding 105 kPa (15 psi). Explosive acetylides can form if acetylene contacts unalloyed copper, silver or mercury.

    Warning Never use copper tubing or tubing of alloys with more than 65% copper content. With these materials, copper acetylide could be formed and there is the risk of spontaneous fire or explosion. Always use stainless steel, or black iron pipe for fixed supply lines, and test for leaks at regular intervals. The installation of acetylene gas must comply with the rules and regulations of the relevant local, state and federal authorities.

  • Varian AA Pre-installation

    5-8 Publication date: October 2003

    5.7 Gas supplies for Graphite Tube Atomizers (both Zeeman and non-Zeeman)

    Argon, nitrogen and air are the only gases recommended for use with a Varian GTA. Proprietary mixtures of argon or nitrogen containing up to 5% hydrogen may be used. If your applications require other gases, please contact your Varian Sales and Service Office.

    NORMAL gas inlet quality High purity argon, 99.99% pure (minimum) High purity nitrogen, 99.99% pure (minimum) ALTERNATE gas inlet quality High purity nitrogen, 99.99% pure (minimum) Air, clean, dry and free of oil

    Recommended pressure range 140 kPa to 200 kPa (20 to 30 psi) Maximum pressure 350 kPa (50 psi) Normal flow rate (GTA-110) 0-3.5 litres per minute Normal flow rate (GTA-120) 0-0.3 litres per minute for Internal

    flow plus separate external flow of 0.5 litres per minute. Additional Boost flow of 3.0 litres per minute during atomization and when programmed temperature exceeds 400C.

    The GTA control unit is fitted with standard barbed-tail adapters for connection of gas supplies using reinforced plastic tubing (supplied) of 6 mm (1/4") internal diameter.

    GTA gas supplies must be regulated to maintain the accessory operating pressures over the full range of normal flow rates under dynamic conditions. Dual stage regulators should be used.

  • Varian AA Pre-installation

    Publication date: October 2003 5-9

    5.8 Gas supplies for the Vapor Generation Accessory Argon and nitrogen are the only gases recommended for use with the Varian VGA. If your applications require other gases, please contact your Varian Sales and Service Office.

    Quality High purity argon, 99.99% pure (minimum) or High purity nitrogen, 99.99% pure (minimum)

    Permissible pressure range 300 to 400 kPa (43 to 57 psi) Recommended pressure 300 kPa (43 psi) Normal flow rate 0.05 to 0.1 litres per minute

    The VGA is fitted with 6 mm (1/4") internal diameter reinforced plastic hose for connection to a standard 6 mm (1/4") barbed-tail connector.

    VGA gas supplies must be regulated to maintain the accessory operating pressure at its normal flow rate.

    The VGA gas outlet should be fitted with a shut off valve to prevent the loss of approximately 50 mL/min of inert gas when the VGA is turned off.

  • Varian AA Pre-installation

    5-10 Publication date: October 2003

    This page is intentionally left blank.

  • Varian AA Pre-installation

    Publication date: October 2003 6-1

    6. Supplementary information

    6.1 Waste fluids The Varian AA flame instrument only atomizes a small percentage of the sample taken up. The excess liquids from the spray chamber must be drained into a waste vessel. Suitable tubing is supplied with the spectrometer for use with inorganic solvents. If you use organic solvents, you will need different tubing, suitable for the solvent(s) of choice. See section 6.2 below. The Varian AA instrument also needs a drain or a sump for disposal of waste liquid during rinse cycles when flame or furnace autosamplers are used. A chemically inert container, not glass and not narrow-necked, to hold approximately two litres (4 pints) of waste must be provided by the instrument user. It should be located underneath the sample compartment (or on the right side of the spectrometer), where it is protected by the bench and in full view of the operator.

    6.2 Special arrangements for use with organic solvents If organic solvents are used in a flame system, the following special option fittings are recommended. Item Part number Kit, O-rings (Organics) 9910091300 Tube, butyl, 9 mm i.d., 15 mm o.d. (1 m length) 3710011700

    A suitable outlet for the liquid trap vent pipe must also be supplied. A venting tube should be connected to the vapor vent (the upper nipple) on the liquid trap when you are analyzing organic or toxic liquids. This tubing should be led out from the sample compartment, parallel to the drain tubing, and MUST slope downwards to enable any liquid overflow to drain out and prevent the tubing from becoming blocked. Do NOT lead the vapor tube to the waste vessel. Please note the following points: The venting tube cannot be routed upwards within the sample compartment because the heat from the flame will melt it.

  • Varian AA Pre-installation

    6-2 Publication date: October 2003

    If organic vapors are vented to the exhaust system, they may be ignited by the flame and so cause damage to the exhaust system. Local environmental protection laws may prohibit toxic vapors from being discharged in to the general atmosphere or into a working area. If necessary, an active exhaust system should be used to draw away toxic vapors and neutralize them. If you are not analyzing solutions of a toxic nature, leave the vapor outlet on the spray chamber uncovered.

  • Varian AA Pre-installation

    Publication date: October 2003 7-1

    7. Water cooling system

    7.1 General information The Graphite Tube Atomizer must have a supply of cooling water to remove heat from the furnace workhead.

    Required flow rate 1.5 litres per minute at 180 kPa (27 psi) Maximum pressure 200 kPa (30 psi) Recommended inlet temperature 18 to 25 C (64 to 70 F) Maximum temperature 40 C (104 F)

    Minimum temperature 10 C (50 F)

    A suitable regulator is included in the GTA control unit. It will be adjusted, if necessary, by the installing Customer Service Representative. Pressure regulation is recommended for supplies which may exceed the maximum permissible pressure.

    The temperature of the furnace coolant is detected by an interlock device which will stop operation if the cooling water reaches 40 C (104 F). Where the ambient temperature is high or where a low capacity water cooling system is used, you may need to include some form of heat exchanger to maintain the recommended water inlet temperature.

    Provided that the flow rate can be maintained above 1.5 litres per minute and there is no back pressure at the outlet, the cooling water supply may be taken from an in-house domestic water system, if local regulations permit.

  • Varian AA Pre-installation

    7-2 Publication date: October 2003

    An alternative is to install a small recirculating system, which should have a tank capacity of at least 100 litres (22 Imperial gallons). The tank should be covered to prevent contamination by dust and other impurities and to minimize evaporation loss. Algicide should be used. A heat exchanger may be necessary. Any such installation must be capable of extracting 950 W at 20 C derated by 17% for 50 Hz operation.

    A refrigerated water cooler has the advantage of not needing a large held volume of water. There are many types available. When choosing a water cooler, check that the specifications meet the requirements of the GTA.

    Your local Varian office or representative may be able to supply a suitable refrigerated water cooler for use with the Varian GTA. Contact your local office or representative for details if required.

  • Varian AA Pre-installation

    Publication date: October 2003 8-1

    8. Equipment on-site

    Note The Varian Customer Service Representative cannot start the installation until the instrument is removed from its packing and pallet and located on the intended workbench.

    8.1 Insurance after delivery As the carriers liability ceases when the equipment is delivered, Varian recommends that you arrange separate insurance that will cover transportation from the delivery point to the installation site. The delivery point will vary according to the carrier, the shipping method, and in some cases the terms of sale. Some carriers will only deliver to their own distribution centre, while others may deliver to your off-loading bay. Few will deliver to the actual installation site.

    8.2 Transit damage Transit damage can be obvious or concealed, and in either case will only be admitted by the carrier if it is reported as agreed in the terms of the carrier's agreement. For any claims against damage in transit, the following general rules apply.

    1. Before accepting delivery, you should inspect the packages for signs of obvious damage. The nature of any obvious damage must be noted on the carriers way-bill, and then must be countersigned by a representative of the carrier.

    2. Within the time limits stated in the terms of conditions of carriage, a further inspection must be made for concealed damage. If any damage is found at this stage, the carrier must be notified in writing and all packaging material must be retained for subsequent inspection by a representative of the carrier.

    3. A copy of any damage report must be forwarded to the Varian Sales Office dealing with the supply of the equipment.

  • Varian AA Pre-installation

    8-2 Publication date: October 2003

    Varian instruments are inherently robust, and the packaging is designed to prevent internal damage when shipped or handled in accordance with the labels and warnings. However, it must be remembered that the contents form part of a precision measuring system and all packages should be handled accordingly. In transit, sharp jolts and shocks must be avoided and the packages must not be inverted, or tilted unnecessarily. Markings on the shipping cartons generally indicate which side of the package should be kept at the top.

    Warning Many of the packages are large and heavy. To avoid injury to personnel, or accidental damage to the equipment, always use two or more people when lifting equipment into position. Never attempt to lift packages alone.

    Warning To prevent death or personnel injury from accidental contact with high voltages within the instrument, do not remove any of the instrument covers. Do not connect the instrument to the mains supply.

    8.3 In-house transit routes In-house transit routes must be carefully considered. Vertical, horizontal and turning clearances should be calculated from the shipping crate dimensions of the spectrometer, which is the largest unit in any system arrangement.

    Figure 8.1 provides an indication of the minimum turning clearance and minimum door width required for the spectrometer in its shipping crate.

  • Varian AA Pre-installation

    Publication date: October 2003 8-3

    'JHVSFMinimum clearance required for transportation in the shipping crate (SpectrAA-50/55/110/220 and Varian-AA140/240/280).

    Carefully check the clearance of any doors in the transit route to the laboratory. The required turning and door clearance may need to take into consideration any lifting device used for transporting the instrument (e.g. fork lift, pallet truck or trolley).

  • Varian AA Pre-installation

    8-4 Publication date: October 2003

    This page is intentionally left blank.

  • Varian AA Pre-installation

    Publication date: October 2003 9-1

    9. Guidelines for software installation

    The following information assumes that you are working on a clean, empty hard disk. If you have any other files on the PC hard disk, ensure you make backups of these before continuing.

    Note Varian will not assume responsibility for loss of data files or third party software.

    9.1.1 Installing the Windows operating system For instructions on installing the Windows operating system, please refer to the Windows manuals supplied with the software. It is the responsibility of the customer to ensure that Windows is installed on the PC if they are supplying it.

    9.1.2 Installing Varian SpectrAA software For instructions on installation of Varian AA software please refer to the Operation manual supplied with the instrument.

    Varian SpectrAA BASE and PRO software requires Windows 2000 Service pack 2, Windows NT 4 Service pack 6 or Windows XP PRO (Service pack 1) to operate properly. Windows 95/98/ME operating systems are no longer supported.

    Note To ensure correct operation, the user installing Windows 2000/NT4/XP must have Administrator privileges for correct installation of Varian AA software.

  • Varian AA Pre-installation

    9-2 Publication date: October 2003

    9.1.3 Installing Varian SpectrAA 21CFR11 software Varian SpectrAA 21CFR11 version software requires Windows 2000 Service pack 3 or Windows XP PRO (Service pack 1).

    Varian SpectrAA 21CFR11 software should be installed by the System Administrator and the System Administrator must be on-site during installation of the Varian AA instrument.

    Note Varian SpectrAA 21CFR11 software requires either Varian SpectrAA Base or PRO software to be installed prior to installation.

    Note 21CFR11 version software requires the file system to be NTFS and not FAT32. This is to be set up when installing the Windows operating system.

    Note To ensure correct operation, Windows 2000/Windows NT 4/Windows XP users must have Administrator privileges for correct installation of Varian AA software.

    Refer to the "21 CFR Part 11 Electronic Records, Electronic Signatures Compliance" manual or the 21CFR11 Assistant online Help supplied with the 21CFR11 software for details on how to use the CFR Assistant to configure the Windows operating system and for details on setting up users and assigning user privileges.

  • Varian AA Pre-installation

    Publication date: October 2003 10-1

    10. Operator training

    Operator confidence when using atomic absorption instrumentation is essential to obtain accurate and reliable results. The Varian Customer Service Representative who installs your equipment will demonstrate the basic operating procedures while conducting the installation performance tests. The service representative is not necessarily experienced in more complex analytical procedures and is not authorized to conduct extended training. The controlling software runs under the Microsoft NT4, 2000 or XP operating systems. The service representative will explain elementary operations but is not authorized to conduct extended training on the Windows operating system.

    Your operators will gain most from observing the installation and will be more immediately productive if they have already been trained. Varian recommends that you take advantage of special training courses that are conducted at various locations by the Varian Technical Support Organization.

    In some locations, it may be possible to arrange for operator training after the installation, using your own instrument. Please contact your Varian Sales and Service Office if you would like to arrange this.

  • Varian AA Pre-installation

    10-2 Publication date: October 2003

    This page is intentionally left blank.

  • Varian AA Pre-installation

    Publication date: October 2003 11-1

    11. Appendix

    11.1 Weights and dimensions of SpectrAA-880/880Z and GTA-100.

    This information is provided to assist in the re-location of already installed older SpectrAA instrument and accessories.

    Spectrometer System unit Width Depth Height Weight

    SpectrAA-880/880Z instrument only 1100 mm (44")

    600 mm (24")

    420 mm (17")

    90 kg (200 lb)

    SpectrAA-880/880Z instrument and shipping box

    1425 mm (56")

    800 mm (31")

    750 mm (30")

    110 kg (240 lb)

    Accessories

    SPS5 Sample Preparation System 525 mm (21")

    610 mm (24.4")

    570 mm (22.8")

    19 kg (42 lb)

    GTA100 Graphite Tube Atomizer (workhead stowed)

    530 mm (21")

    530 mm (21")

    400 mm (16")

    43 kg (95 lb)

    Shipping dimensions 1000 mm

    (40)

    710 mm

    (28.4)

    700 mm

    (28)

    71 kg

    (156 lb)

    GTA100 Zeeman 440 mm (18")

    540 mm (22")

    300 mm (12")

    52 kg

    (114 lb)

    Shipping dimensions 1000 mm

    (40)

    710 mm

    (28.4)

    700 mm

    (28)

    79 kg

    (174 lb)

    Autosampler PSD100 305 mm

    (12.2)

    360 mm

    (14.4)

    310 mm

    (12.4)

    6.5 kg

    (14 lb)

    Shipping dimensions 620 mm

    (24.8)

    420 mm

    (16.8)

    325 mm

    (13)

    11.5 kg

    (25 lb)

  • Varian AA Pre-installation

    11-2 Publication date: October 2003

    'JHVSFFront view of a SpectrAA-880 series instrument

    'JHVSFTop view of a SpectrAA-880 series instrument

  • Varian AA Pre-installation

    Publication date: October 2003 11-3

    11.2 Electrical power supplies System unit Required supply voltage Rating

    SpectrAA 880 100, 120, 220 or 240 VAC 10% 50 or 60 Hz 1 Hz

    470 VA

    Zeeman version

    SpectrAA 880Z 208, 220 or 240 VAC 10% 230 VAC +14% -6% 230 VAC +6% -14% 50 or 60 Hz 1 Hz

    3500 VA **

    GTA Accessory 208, 220 or 240 VAC 10% 230 VAC +14% -6% 230 VAC +6% -14% 50 or 60 Hz 1 Hz

    3500 VA **

    SPS5 Autosampler 100, 120, 220 or 240 VAC 10 % 230 VAC +14% -6%, 230 VAC +6% -14% 50 or 60 Hz 1 Hz

    80 VA

    5BCMF Electrical specifications for instruments.

  • Varian AA Pre-installation

    11-4 Publication date: October 2003

    11.3 In-house transit routes

    'JHVSFMinimum clearance required for transportation in the shipping crate (SpectrAA-880/880Z).


Top Related