Mass Spectrometry
Ionization Techniques
and
Mass Analyzers
-
8/6/2019 Source and Analyzer
2/28
Mass Spectrometer
All Instruments Have:
1. Sample Inlet
2. Ion Source
3. Mass Analyzer
4. Detector
5. Data System
http://www.asms.org
-
8/6/2019 Source and Analyzer
3/28
Ionization Techniques
Gas-Phase Methods Electron Impact (EI)
Chemical Ionization (CI)
Desorption Methods Matrix-Assisted Laser Desorption Ionization (MALDI)
Fast Atom Bombardment (FAB)
Spray Methods Electrospray (ESI)
Atmospheric Pressure Chemical Ionization (APCI)
-
8/6/2019 Source and Analyzer
4/28
Electron Impact
http://www.noble.org/PlantBio/MS/ion_tech_main.html
-
8/6/2019 Source and Analyzer
5/28
Electron Impact
Advantages
Well-Established
Fragmentation Libraries No Supression
Insoluble Samples
Interface to GC
Non-Polar Samples
Disadvantages
Parent Identification
Need Volatile Sample Need Thermal Stability
No Interface to LC
Low Mass Compounds
(
-
8/6/2019 Source and Analyzer
6/28
Chemical Ionization
http://www.noble.org/PlantBio/MS/ion_tech_main.html
-
8/6/2019 Source and Analyzer
7/28
Chemical Ionization
Advantages
Parent Ion
Interface to GC Insoluble Samples
Disadvantages
No Fragment Library
Need Volatile Sample Need Thermal Stability
Quantitation Difficult
Low Mass Compounds
(
-
8/6/2019 Source and Analyzer
8/28
FAB
-
8/6/2019 Source and Analyzer
9/28
FAB
Advantages
Parent Ion
High Mass Compounds(10,000 amu)
Thermally Labile
Compounds (R.T.)
Disadvantages
No Fragment Library
Solubility in Matrix(MNBA, Glycerol)
Quantitation Difficult
Needs Highly SkilledOperator
Relatively LowSensitivity
(nanomole)
-
8/6/2019 Source and Analyzer
10/28
MALDI
http://www.noble.org/PlantBio/MS/ion_tech_main.html
-
8/6/2019 Source and Analyzer
11/28
MALDI
Advantages
Parent Ion
High Mass Compounds(>100,000 amu)
Thermally Labile
Compounds (R.T.)
Easy to Operate
Disadvantages
No Fragment Library
Wide variety of matrices Quantitation Difficult
(low femtomole)
-
8/6/2019 Source and Analyzer
12/28
ESI
http://www.noble.org/PlantBio/MS/ion_tech_main.html
-
8/6/2019 Source and Analyzer
13/28
ESI
Advantages
Parent Ion
High Mass Compounds(>100,000 amu)
Thermally LabileCompounds (
-
8/6/2019 Source and Analyzer
14/28
APCI
http://www.rzuser.uni-heidelberg.de/~bl5/ency/ency.html
-
8/6/2019 Source and Analyzer
15/28
APCI
Advantages
Parent Ion
Insensitive to Salts Interface to HPLC
Can use Normal Phase
Solvents
Handles High FlowRates
Disadvantages
Need Volatile Sample
Need Thermal Stability
(high femtomole)
-
8/6/2019 Source and Analyzer
16/28
Mass Analyzers
Double Focusing Magnetic Sector
Quadrupole Mass Filter
Quadrupole Ion Trap
Linear Time-of-Flight (TOF)
Reflectron TOF Fourier Transform Ion Cyclotron
Resonance (FT-ICR-MS)
-
8/6/2019 Source and Analyzer
17/28
Double-Focusing Magnetic Sector
http://www.asms.org
-
8/6/2019 Source and Analyzer
18/28
Advantages
Very High Resolution
(60,000)
High Accuracy (
-
8/6/2019 Source and Analyzer
19/28
Quadrupole Mass Filter
http://www.asms.org
-
8/6/2019 Source and Analyzer
20/28
Advantages
Inexpensive
Easily Interfaced toMany Ionization
Methods
Disadvantages
Low Resolution (100ppm)
MS/MS requires
multiple analyzers
Low Mass Range(
-
8/6/2019 Source and Analyzer
21/28
Quadrupole Ion Trap
http://www.asms.org
-
8/6/2019 Source and Analyzer
22/28
Advantages
Inexpensive
Easily Interfaced toMany Ionization
Methods
MS/MS in one analyzer
Disadvantages
Low Resolution (100ppm)
Space Charging
Causes Mass Shifts
Low Mass Range(
-
8/6/2019 Source and Analyzer
23/28
Linear Time-of-Flight (TOF)
http://www.asms.org
-
8/6/2019 Source and Analyzer
24/28
Advantages
Extremely High Mass
Range (>1 MDa)
Fast Scanning
Disadvantages
Low Resolution (4000)
Low Accuracy(>200ppm)
MS/MS not possible
Linear Time-of-Flight (TOF)
-
8/6/2019 Source and Analyzer
25/28
Reflectron Time-of-Flight (TOF)
Courtesy Bruker Datonics BioTOF users Manual
-
8/6/2019 Source and Analyzer
26/28
Advantages
High Resolution
(>20,000 in some
models)
High Accuracy (
-
8/6/2019 Source and Analyzer
27/28
FT-ICR-MS
http://www.asms.org
-
8/6/2019 Source and Analyzer
28/28
Advantages
Extremely High
Resolution (>500,000)
Very Good Accuracy
(