![Page 1: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/1.jpg)
Rad Tech 110
Circuits
![Page 2: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/2.jpg)
Conditions Necessary X-ray Production
• High voltage– In the kilovoltage range
• Electrons– A supply of electrons– A need to stop the electrons suddenly
![Page 3: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/3.jpg)
Types of Transformers
Shell typeCore type
![Page 4: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/4.jpg)
How do transformers work?
• They can only work using AC current
• On the principle of induction– Mutual– Self
• The basic premise is that AC current ‘induces’ alternating magnetic fields.
![Page 5: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/5.jpg)
Transformers work?
• These alternating magnetic fields are distributed throughout the core of the transformer.
• The alternating magnetic fields ‘induce’ an electrical current in loops (coils) of wire.
![Page 6: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/6.jpg)
![Page 7: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/7.jpg)
What gets transformed?
• Both voltage and current• In a step-up transformer the voltage is
increased– The primary side (incoming) has less coils than
the secondary (outgoing) side.– Because there are more coils on the secondary
side a higher voltage is produced• Increases voltage is reflected as a higher amplitude
of the waveform.
![Page 8: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/8.jpg)
• At the same time as voltage increases the current correspondingly decreases.
• This occurs due to the conservation of energy.– If you increase one aspect, voltage, you have to
decrease the other, current.
– (Voltagep)(currentp)=(voltages)(currents)
![Page 9: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/9.jpg)
Step down transformer
• Here the primary side has more coils than the secondary side.– The result is a decrease in voltage through the
transformer.
![Page 10: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/10.jpg)
The X-ray Circuit
![Page 11: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/11.jpg)
Three sections of the x-ray circuit
• Operating (control) console
• High voltage generator
• X-ray tube
![Page 12: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/12.jpg)
Operating (control) console
• Line voltage compensator
• Autotransformer
• kVp selector
• mA selector
• Timer
• Pre-reading kVp meter
![Page 13: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/13.jpg)
Line voltage compensator
The purpose of the line voltage compensator is to maintain a constant voltage to the system.In the modern system this is done automatically.
![Page 14: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/14.jpg)
Autotransformer
The autotransformer works on the principle of self-induction. It has a single core and is responsible for varying the voltage.Because of its ability to adjust voltage, the autotransformer can be either a step-up or step-down transformer.
![Page 15: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/15.jpg)
kVp selector
• kVp is adjusted at the autotransformer
• Taps are available so that different voltages can be achieved.
![Page 16: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/16.jpg)
Pre-reading kVp meter
• The kVp meter is considered pre-reading because it is not actually measuring kVp but measuring voltage off the autotransformer.– Remember, the autotransformer is used to
adjust voltage.
![Page 17: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/17.jpg)
Timer
• Historically, there have been many types of timers.
• Today, all timers are electronic timers of various types.– mAs– Phototiming (AEC)– Regular electronic
![Page 18: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/18.jpg)
mA selector
• Allows for adjust of the filament voltage– Precision resistors– Choke coil
![Page 19: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/19.jpg)
High Voltage Generator
• Step-up transformer– High voltage transformer– High tension transformer
• Step-down transformer– Filament transformer
• Rectifiers
![Page 20: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/20.jpg)
Step-up Transformer
• This transformer is responsible for producing the high voltages necessary for x-ray production.– Turns ratio of 500:1 or 1000:1
![Page 21: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/21.jpg)
Filament Transformer
• Intended to lower voltage and increase current
• Allows for thermionic emission to occur at the filament
• Ultimately, provides the electrons necessary for x-ray production.– Tube current measured in mA.
![Page 22: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/22.jpg)
Rectifiers• Current traffic cops
– Intended to allow electrons to flow in only one direction
• Provide the mechanism that converts AC to DC current.
• 4 rectifiers are required to fully convert AC to DC.
![Page 23: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/23.jpg)
Types of Circuits
• Single phase– Full wave– Half wave
• Three phase– 6 pulse– 12 pulse
• Medium and High Frequency• Capacitor discharge
– Found predominantly in portables.
![Page 24: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/24.jpg)
Single Phase
• Full
• Half
![Page 25: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/25.jpg)
Three phase
• 6 pulse
• 12 pulse
![Page 26: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/26.jpg)
Three phase transformers
• Three phase units use delta and ‘wye’ transformers
• 6 pulse– 2 ‘wye’ and 1 delta
• 12 pulse– 2 delta and 1 ‘wye’
• They also require 12 rectifiers
![Page 27: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/27.jpg)
Advantages of 3 phase
• Higher beam quantity and quality
• Higher mA stations are also possible
![Page 28: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/28.jpg)
Why higher quantity and quality for three phase?
• Because the voltage ripple is less– Basically when the tube is energized it stays
energized.
full
3 phase 12 pulse
![Page 29: Rad Tech 110 Circuits. Conditions Necessary X-ray Production High voltage –In the kilovoltage range Electrons –A supply of electrons –A need to stop the](https://reader037.vdocuments.mx/reader037/viewer/2022102611/56649c905503460f9494a5ab/html5/thumbnails/29.jpg)
Medium/High Frequency Generator