Page 1: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Portable Broadband Seismic Stations -

Checking Data Quality

Cairo, EgyptNovember, 2009

Noel BarstowBruce BeaudoinMarcos Alvarez

Page 2: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Checking Data QualityKeep good records of everything

> Before going to the fieldTest all Instrumentation and Cables

> In the fieldInstall Station and Check: Power, Timing, Sensor Orientation, Signal

Visit stand-alone stations frequently to: Download Data and Review Station Status and Metadata on site

> In the labArchive, Analyze, and Document Data

Page 3: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Hardware Maintenance and QC: Dataloggers

Function test all Dataloggers prior to shipment:Bit weight, oscillator drift, data time stamps

Keep maintenance records in an electronic databaseTest and upgrade firmware when necessary

Page 4: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

PASSCAL’s Hardware Maintenance: Seismometers

Feedback seismometers are Pier-Tested for 4-7 Days and compared to a Reference Sensor:- Amplitude and Phase- Polarity- Spectral Content- Noise Characteristics

Electronic Calibrations:- Corner Period- Damping

Page 5: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Do the signals look the same?

close up of verticals

overlay surface wave

Page 6: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Nightly Plots of Pier Tests

Page 7: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Electronic Step Calibration

Page 8: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

White Noise Calibrations

Input White Noise

Pha

se

Am

plitu

deSensor Output

Sensitivity

Page 9: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Pier Data Comparing Sensor Noise of Different Sensor Types

Ve

rtic

al C

om

pon

en

ts

Page 10: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Streckeisen STS-2

Page 11: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

X = − 2 /3( ) U + 1/6( ) V + 1/6( ) W

Y = 1/2( ) V − 1/2( ) W

Z = 1/3( ) U + 1/3( ) V + 1/3( ) W

STS-2 Element Contribution to Signal

East

North

Vertical

Page 12: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

STS-2 Element V problem

Z

N

E Noisy Sensor

Z

N

E

Page 13: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Triaxial SystemsR

efer

ence

Unfiltered Earthquake Filtered 10 -30 sec period

Z

N

E

Z

N

E

Tes

t S

enso

r:

Drif

t &

DC

off

set

Page 14: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Orthogonal elements2 Horizontal1 Vertical

Guralp CMG-3T

Page 15: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Replace Bent 3T Pivot

Comparing Reference and Problem Sensor

Before

After

Time Scale = 3 hours

Eas

t /

Wes

t C

ompo

nent

Page 16: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Sudden Noise on Guralp 3T Vertical Component

Noise begins

Noisy Spectra

Days:

1

2

3

4

Ver

tical

Dat

a

Spe

ctra

Locking and Unlocking the 3T provides a temporary fix

Page 17: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Amplitude change at DC offset

Time = 8 Days

N/S

E/W

Z

Amplitude Problem: Loss of Differential Output

Das, Sensor, or interconnecting Cables could be the cause

Page 18: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

•PASSCAL develops and maintains software to extract data from the dataloggers, perform in-field quality checks, and manipulate the waveforms into archive formats.

•http://www.passcal.nmt.edu•ftp.passcal.nmt.edu

Data QC & Software

Page 19: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Data Archiving Support PASSCAL acts as an intermediary between

PASSCAL Investigators and the DMC

• Check completeness of data submitted to PASSCAL

• Help evaluate or fix timing

• Verify SEED Format

• Ensure agreement between data and metadata

• Check completeness of data submitted to DMC

Page 20: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

• PASSCAL provides email and phone support for our users before, during, and after experiments to analyze specific problems.

• PASSCAL staff members are available for the initial testing and installation phase of field campaigns.

• Typically there are over 500 broadband and short-period PASSCAL stations deployed at any given time.

• Contact us: [email protected]

Troubleshooting

Page 21: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

After the power point presentation

Live demonstration oflogpeek - plotting of RT130 station log

messagespql - PASSCAL Quick Look - to view

waveforms, fast fourier spectrum, split view, etc.

pql will read mini-seed, sac, and segy formats

Page 22: Portable Broadband Seismic Stations - Checking Data Quality Cairo, Egypt November, 2009 Noel Barstow Bruce Beaudoin Marcos Alvarez

Review Station’s Status

Firmware

Volts

GPS

Timing

Time Scale = 1 WEEK

‘logpeek’Plot of StationLog MessagesFromRT130 DAS