Non-Invasive Blood Pressure Waveform Monitoring in Mice
By
Andrew McClellan
Laura Miller
![Page 2: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/2.jpg)
The Need for Blood Pressure Measurements• Baseline levels
• Effects of pharmacological drug administration
• Exercise or Food Studies
• Behavioral Response Studies
• Effects of genetic alterations
![Page 3: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/3.jpg)
Need for Continuous Waveform Measurements vs. Single Measurements
• Single measurements only give max and min (systolic / diastolic)
• Continuous waveforms needed for frequency analysis– autonomic nervous system responses– blood pressure variability– baroreceptor function
• Examining blood pressure changes over time, like during behavioral tests
![Page 4: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/4.jpg)
What Currently Exists
Mouse Human
Beat-to-beat Tail Cuff Arm Cuff
Continuous • Arterial Line• Implanted Catheter/
Transmitter
• Finger Cuff (Finapres)
![Page 5: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/5.jpg)
Beat to beat BP measurements: Korotkoff Method
Mouse: inflatable tail cuff
www.TSE-systems.com
Human: inflatable arm cuff
http://www.geocities.com/lorijean_2002/academia/sphygmomanometer.jpg
![Page 6: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/6.jpg)
What Currently Exists
Mouse Human
Beat-to-beat Tail Cuff Arm Cuff
Continuous • Arterial Line•Implanted Catheter/
Transmitter
Finger Cuff (Finapres)
![Page 7: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/7.jpg)
What Currently Exists for Mice: Implantable transmitters•Catheter pressure transducers (Data Sciences International)
•Continuous waveform
•Require surgical implantation into carotid artery
•Mice need at least 3 weeks post-surgery recovery time for stabilization of BP/HR
•Transmitters are expensive and require corresponding receiver hardware and analysis software ($600/transmitter + $100/each battery exchange + $1000s in receivers and software)
Mills P, et al. The design and performance of an implantable device for monitoring blood pressure, heart rate, and movement activity from conscious freely moving laboratory mice. 2nd International Conference on Methods and Techniques in Behavioral Research, 1998.
Mills PA, et al A new method for measurement of blood pressure, heart rate, and activity in the mouse by radiotelemetry. J Appl Physiol. 2000 May;88(5):1537-44
![Page 8: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/8.jpg)
What Currently Exists
Mouse Human
Beat-to-beat Tail Cuff Arm Cuff
Continuous • Arterial Line• Implanted Catheter/
Transmitter
• Finger Cuff (Finapres)
![Page 9: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/9.jpg)
What Currently Exists: finger cuff measurements
Finapres
• Non-invasive
• Continuous waveform
• Finger cuff
Bogert LW, et al. Non-invasive pulsatile arterial pressure and stroke volume changes from the human finger. Exp. Physiol 90.4 pp 437-446. 2005
![Page 10: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/10.jpg)
What Currently Exists
Mouse Human
Beat-to-beat Tail Cuff Arm Cuff
Continuous • Arterial Line• Implanted Catheter/
Transmitter
• Finger Cuff (Finapres)
Problem: There are no blood pressure devices that non-invasively measure a continuous waveform in mice
![Page 11: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/11.jpg)
Our project goal
Non-invasive, continuous blood pressure waveform measurement in mice
![Page 12: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/12.jpg)
Challenges From Adapting a Human Device
• Size – mice are much smaller
• Physiological Parameters –
HR = 500-700bpm, BP = 100-250mmHg
• Cuff – must work for above pressures
• System response – high frequency
![Page 13: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/13.jpg)
Peňáz Method
• Dynamic pulsatile unloading of the arterial walls
• Volume clamp method
• Pressure needed to maintain mean arterial volume equals the blood pressure waveform
![Page 14: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/14.jpg)
Plethysmographic Signal
• Obtained through measuring optical impedance through blood vessel in tail
• Optical impedance corresponds to volume of blood in artery
![Page 15: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/15.jpg)
How it Works
• Blood volume α blood pressure
• Blood volume α optical impedance (plethysmographic signal)
• Use optical impedance to determine necessary pressure to maintain mean arterial volume
![Page 16: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/16.jpg)
Transduction
Blood Volume = Blood Pressure
P-signal (voltage)
(photosensor)
Pressure Wave
invert & offset
(speaker)
(tail cuff)
![Page 17: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/17.jpg)
Tail Cuff with Photosensor
www.TSE-systems.com
![Page 18: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/18.jpg)
Peňáz Method1. Obtain plethysmographic signal2. Ramp up cuff-pressure until Vp-p of plethymographic signal is
max (offset)3. Invert plethysmographic signal4. Convert electrical signal to a pressure waveform in the tail cuff5. Increase gain until p-signal fluctuations in plethysmographic
signal are minimal
max
ramp-up
increase gain
ramp-up
![Page 19: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/19.jpg)
Peňáz Method
max
ramp-up
increase gain
ramp-up
![Page 20: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/20.jpg)
Circuit
buffer gain offset audio amp
![Page 21: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/21.jpg)
Schematic
tail cuff
Harvard plethysmogram / pressure monitor
computer
pump
speaker
amplify offset
plethysmographic signal
pressure signal
analog circuit
![Page 22: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/22.jpg)
Lab Setup
![Page 23: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/23.jpg)
Economic AnalysisDevice:• Tail Cuff with Photosensor $400• Circuit components $10• Speaker $5• Power Source $100• Air pump $50• Pressure Sensor $100
$665Acquisition software/hardware• Windaq $2000
![Page 24: Non-Invasive Blood Pressure Waveform Monitoring in Mice By Andrew McClellan Laura Miller](https://reader030.vdocuments.mx/reader030/viewer/2022032612/56649efd5503460f94c11132/html5/thumbnails/24.jpg)
Plan of Attack• Use existing tail cuff• Obtain a plethysmographic signal• Build an analog circuit to amplify and
offset the signal• Use speaker to transduce voltage to
pressure• Obtain blood pressure waveform• Check waveform against catheter
transducer