Large Taper Files
![Page 2: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/2.jpg)
Large Taper Files
Engine-driven files
![Page 3: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/3.jpg)
Hand-Using Large Taper Files
![Page 4: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/4.jpg)
A, Lateral view. (SEM, ×50.) B, Cross section. (SEM, ×200.)
![Page 5: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/5.jpg)
![Page 6: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/6.jpg)
Large Taper File Advantages
Easier Only one instrument sequence whatever the canal shape An easy to remember protocol (color-coded)
Faster Only 3 instruments needed in most cases High cutting power
![Page 7: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/7.jpg)
Large Taper File Advantages More efficient increased apical taper for better canal cleaning Better debris removal due to the unique <Multiple Taper>
design of these instruments
Safer Guiding tip with rounded end to minimize canal
deviations These instruments are also available in manual version
to treat anatomically difficult cases or for clinicians who prefer to use hand files for tactile control reasons
![Page 8: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/8.jpg)
How to prepare root canal by using Large
Taper Files ?
![Page 9: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/9.jpg)
D1
DX
D2M2 M1
M3
![Page 10: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/10.jpg)
1. Prepare a straight-line access cavity with no restrictions in the entry path into
the chamber
![Page 11: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/11.jpg)
2. Establish proper access and a glide path with No. 10 and No. 15 steel files to the working lengthor the apical constriction exit.
![Page 12: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/12.jpg)
3. Begin shaping with the D1 using multiple, passive-pressure passes.
Go no deeper than 1/3 of the estimated canal length. Irrigate and recapitulate with a No. 10 hand file, establishing patency to full working length.
With D1, extend the preparation to full working length.
D1 shaping Irrigation recapitulation
NaClOEDTA
![Page 13: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/13.jpg)
4. “Brush” with the DX to improve thestraight-line access in short teeth or to relocate canal access away from furcation in posterior teeth.
![Page 14: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/14.jpg)
5. D2 is now used to full working length.Irrigate, recapitulate
D2 shaping Irrigation recapitulate
NaClOEDTA
![Page 15: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/15.jpg)
6. With M1, passively extend the preparation to within 0.5 mm of the working length.
Withdraw after one second! And only one second!
The M1 has a tip size of 0.20 mm, and if a No. 20 hand instrument is found to be snug, the preparation is finished.
![Page 16: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/16.jpg)
7. If the M1 and the No. 20 hand file are loose, continue the preparation with the M2, which is 0.25 mm diameter at the tip.
If the M2 and the No. 25 hand file are
loose, continue the preparation with the M3, which is 0.30 mm diameter at the tip.
![Page 17: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/17.jpg)
Frequent irrigation and file cleansing are imperative
irrigation and recapitulation!
5.25%NaClO17%EDTAirrigation
![Page 18: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/18.jpg)
The data of engine-driven Large Taper files
![Page 19: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/19.jpg)
How to better use Large Taper Files
Prepare a straight-line access cavity with no restrictions in the entry path into the chamber.
Fill the access cavity with Sod Hypochlorite.
Establish a smooth glide path with No. 10 and No.15 stainless steel hand files.
Use maximum magnification to observe the movement of the rotary instrument. “Seeing” rotary apical movement is safer than simply “feeling” such movement.
![Page 20: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/20.jpg)
How to better use Large Taper Files
Use a torque- and speed-controlled electric motor, powering the hand-piece at 200 to 300 rpm.
Be much gentler than with hand instruments. Always treat in a moist canal. Irrigate frequently!
Slow down! Each instrument should do minimal shaping. Only two, three, or four passes may be required for the file to engage restrictive dentin and carve the shape to the proper depth.
![Page 21: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/21.jpg)
How to better use Large Taper Files
Instruments break when flutes become loaded or when instruments are forced. Check the flutes frequently under magnification and clean them. Cyclic fatigue from overuse, or if the glide path is not well established, also leads to breakage.
![Page 22: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/22.jpg)
How to better use Large Taper Files
Large Taper Files instruments are disposable and, like all endodontic files and reamers, are designed for single-patient use. Sometimes instruments are even changed within the same treatment (eg, in the case of a four-canal molar).
Irrigate with 17% EDTA or a viscous chelator during the ProTaper shaping.
![Page 23: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/23.jpg)
How to disinfect, clean and sterilize files
![Page 24: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/24.jpg)
1. Manual Cleaning or assisted by an ultrasonic device
No visible impurities should be observed on the instruments.
Discard any instruments with large obvious defects (broken, bent, and twisted).
The disinfecting solution should be aldehyde free and without di- or triethanolamines as corrosion inhibitor.
![Page 25: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/25.jpg)
2. Rinsing Abundant rinsing (at least 1 min)
Use quality water in accordance with local regulations.
If a disinfecting solution contains a corrosion inhibitor, it is recommended to rinse the instruments just before the autoclaving.
Dry on a single use non-weaved cloth, or with a drying machine or filtered compressed air.
![Page 26: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/26.jpg)
3. Packaging Place the devices in a kit, support or container to
avoid any contact between instruments and pack the files in “Sterilization pouches”.
Check the validity period of the pouch given by the manufacturer to determine the shelf life.
Use packaging which are resistant up to a temperature of 141°C (286°F) and in accordance with EN ISO 11607.
![Page 27: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/27.jpg)
4. Sterilization Steam sterilization at: 134 °C / 273°F during 18
min.
Use only autoclaves that are matching the requirements of EN 13060, EN 285.
Use a validated sterilization procedure according ISO 17665
Control the efficiency (packaging integrity, no humidity, color change of sterilization indicators).
![Page 28: Large Taper Files. Engine-driven files Hand-Using Large Taper Files](https://reader036.vdocuments.mx/reader036/viewer/2022081504/56649c7b5503460f9492f823/html5/thumbnails/28.jpg)
5. Storage Keep devices in sterilization packaging in a dry and clean environment
Sterility cannot be guaranteed if packaging is open, damaged or wet.
Check the packaging before using them (packaging integrity, no humidity and validity period).