HYBRID IMAGING: SPECT / CT
Associate Professor Barry S Elison Director Dept Nuclear Medicine, Wollongong Hospital
Senior Staff Specialist, Bankstown Hospital Clinical Executive Director Clinical Director of Imaging
FRACP, FSCCT
![Page 2: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/2.jpg)
![Page 3: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/3.jpg)
![Page 4: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/4.jpg)
Hybrid Imaging in Cardiac Nuclear Medicine
1. Attenuation Correction 2. May derive some information from CT such as, a) Non-cardiac findings b) Pericardial abnormalities 3. Calcium score. 4. Combined CTCA and SPECT
![Page 5: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/5.jpg)
Why Attenuation Correction
• Anatomical regions usually uneven or heterogeneous and therefore photon attenuation is unavoidable.
• Unavoidable problems do affect quality. • Attenuation correction therefore desirable.
![Page 6: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/6.jpg)
1. PHOTON ATTENUATION • Gamma photons emitted by radiopharmaceuticals
interact with tissues as they pass through the body – Absorbed (attenuated) – Scattered (Compton) – Attenuation is energy-dependent
25
% T
RA
NS
MIT
TED
PH
OTO
NS
0 5 10 15 20 SOURCE DEPTH (cm)
50
75
100
LUNG
WATER
BONE 140 keV
0
![Page 7: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/7.jpg)
1. PHOTON ATTENUATION (CONT)
• Most anatomical regions are heterogeneous – radionuclide data contains errors due to photon attenuation (artifacts)
• Photon attenuation is unavoidable • Photon attenuation seriously affects quality of diagnostic
information
BREAST ATTENUATION IN MPS
DEPTH ATTENUATION b >> a
a b
![Page 8: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/8.jpg)
![Page 9: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/9.jpg)
1. PHOTON ATTENUATION (CONT)
• Exponential relationship for photon attenuation:
xeII µ−= 0
where I0 = incident photon energy I = transmitted photon energy x = tissue thickness (cm) µ = linear attenuation coefficient (cm-1)
e-µa
e-µb
a
b
REF: ZAIDI, 2006
Attenuation Correction is vital to the production of artifact-free data
![Page 10: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/10.jpg)
2. DETERMINATION OF µ-MAP
• Need to generate an attenuation map – Spatial distribution of µ values (µ-map) – Incorporated into reconstruction algorithm to
correct for photon attenuation and scatter
• Patient-specific µ-map can be generated from CT data
![Page 11: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/11.jpg)
2. DETERMINATION OF µ-MAP (CONT)
• But raw CT data is normalised for variations in x-ray spectrum
• E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU
• Need to transform CT numbers back into µ values at correct energy
( ) 1000, xyxCTW
WCT
µµµ −
= where CT = CT number (Hounsfield Units) µCT = LAC of irradiated tissue (cm-1) µW= LAC of water (0.1504 cm-1)
![Page 12: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/12.jpg)
2. DETERMINATION OF µ-MAP (CONT)
• Acquire CT scans of known µ values using phantom • Generate a lookup table using bilinear scaling:
0.05
µ (c
m-1
)
-1000 -500 0 500 CT NUMBER (HU)
0.10
0.15
0.20
140 keV
0
0.25
1000
![Page 13: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/13.jpg)
![Page 14: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/14.jpg)
![Page 15: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/15.jpg)
![Page 16: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/16.jpg)
Radiation Protection Primer
Linear-No-Threshold Model (Any radiation = harmful even at low doses) : controversial ? inaccurate
![Page 17: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/17.jpg)
![Page 18: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/18.jpg)
Radiation Protection Primer Most radiation workers adopt the ALARA principle = As Low As Reasonably Achievable Conventional diagnostic Cardiac Imaging with NM & CT: generally already low radiation exposure & can be justified if clinically indicated (CAD = potentially life-threatening) with CZT SPECT CT: radiation dosage even lower comparable to natural exposure in magnitude With NO compromise on image & diagnostic quality
![Page 19: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/19.jpg)
Calcium Score
• In Hybrid model using SPECT /CT there are different ways to score calcium
• Visual • Qualitative • Quantitative
![Page 20: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/20.jpg)
![Page 21: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/21.jpg)
![Page 22: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/22.jpg)
![Page 23: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/23.jpg)
![Page 24: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/24.jpg)
Coronary Calcium Score -Routinely performed in patients with no known CAD (unless specifically requested not to) -Low dose (<1mSv), non-contrast CT breath-hold, prospective ECG gating, software detection of HU >130 in coronary distribution -CT also used for attenuation correction,
![Page 25: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/25.jpg)
![Page 26: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/26.jpg)
SPECT and calcium scoring – complimentary data Chang et al JACC 11/09
n = 1126 F/U over 6.9 years
![Page 27: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/27.jpg)
JACC 2009; 54: 1872-82
![Page 28: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/28.jpg)
![Page 29: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/29.jpg)
Coronary Calcium Score: Incremental Dx Value over SPECT
![Page 30: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/30.jpg)
49-year-old female -Chest pain, positive EST -Normal stress /rest MIBI -Normal CACS = 0 confirming that she did not have significant CAD.
J Nucl Cardiol 2005;12: 392-400
![Page 31: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/31.jpg)
68-year-old female without known CAD -chest pain -risk factors: age & positive family history. -Normal stress / rest MIBI -CACS =1753, marked calcification in LAD, LCx, LM -reclassified from no significant CAD to significant disease, and aggressive medical therapy was started.
J Nucl Cardiol 2005;12: 392-400
![Page 32: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/32.jpg)
– Simple, fast, no contraindications, no contrast – Screen patients with intermediate risk: Positive CAC increased risk alter therapeutic goals (LDL, BP etc)
– Further stratify higher risk patients with normal stress-rest MIBI
– Identify patients who do not need further cardiac evaluation (CACS = 0, low clinical risk)
– Monitor progression of atherosclerosis – Improve compliance
COMMENTS on CACS
![Page 33: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/33.jpg)
GE Hybrid CZT SPECT-CT Discovery 570C
Better image quality In less time less radiation to patients
![Page 34: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/34.jpg)
Zurich University Hospital
Together with ASIR software (Adaptive Statistical Iterative Reconstruction) Markedly reduced radiation dose in CTCA and also in low dose CT for AC & coronary calcium score
![Page 35: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/35.jpg)
Validation of CT attenuation correction for CZT camera
Normal stress/rest MPI with inferior attenuation artefact corrected with CT AC for CZT Camera
![Page 36: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/36.jpg)
Method: Compare conventional state-of-the-art (Philips Precedence) SPECT-CT hybrid camera with the Discovery 570c SPECT-CT hybrid camera in 25 randomly selected patients from each camera (single day scan protocol). Conclusion: -CZT SPECT-CT with ASIR : major reductions in both radiation dose & scanning time while maintaining high image quality. -An ultralow dose stress only (100MBq) SPECT-CT protocol including calcium score with total effective radiation dose of 1.6mSv is achievable.
Bankstown Hospital Experience
![Page 37: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/37.jpg)
Method: Compare conventional state-of-the-art (Philips Precedence) SPECT-CT hybrid camera with the Discovery 570c SPECT-CT hybrid camera in 25 randomly selected patients from each camera (single day scan protocol). Conclusion: -CZT SPECT-CT with ASIR : major reductions in both radiation dose & scanning time while maintaining high image quality. -An ultralow dose stress only (100MBq) SPECT-CT protocol including calcium score with total effective radiation dose of 1.6mSv is achievable.
Bankstown Hospital Experience
![Page 38: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/38.jpg)
Method: Compare conventional state-of-the-art (Philips Precedence) SPECT-CT hybrid camera with the Discovery 570c SPECT-CT hybrid camera in 25 randomly selected patients from each camera (single day scan protocol). Conclusion: -CZT SPECT-CT with ASIR : major reductions in both radiation dose & scanning time while maintaining high image quality. -An ultralow dose stress only (100MBq) SPECT-CT protocol including calcium score with total effective radiation dose of 1.6mSv is achievable.
Bankstown Hospital Experience
![Page 39: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/39.jpg)
• Her recent chest pains were most likely not due to myocardial ischaemia. Given her age and dyslipidaemia she is however at risk for coronary artery disease. I arranged for her to have an ultralow radiation stress only MIBI scan and a Coronary Artery Calcium Score at Bankstown Hospital. If she has significant coronary artery calcifications I would have a low threshold in commencing her on lipid lowering medication. Obviously if she has evidence of ischaemia I would perform coronary angiography
![Page 40: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/40.jpg)
GE Hybrid CZT SPECT-CT Discovery 570C
![Page 41: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/41.jpg)
The best (or the prettiest) for last
![Page 42: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/42.jpg)
Incremental value of CTCA & software SPECT/CT fusion Assess Stent patency
![Page 43: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/43.jpg)
Incremental value of CTCA & software SPECT/CT fusion over CA Anatomical Anomaly
![Page 44: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/44.jpg)
Incremental value of CTCA & software SPECT/CT fusion over CA
![Page 45: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/45.jpg)
Where will this technology take us
• Clinical relevance ? • Will this technology change or influence
management • Outcomes : will combining technologies- CTCA
and MPS change or improve outcomes • Radiation burden • Cost
![Page 46: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/46.jpg)
1. Attenuation Correction- use it but cant always rely on it. Never ignore the NC data. Caveat: Quantification with commercially available software
2. Calcium score-Is the evidence trending towards doing it routinely with the hybrid model
CONCLUSION
![Page 47: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/47.jpg)
1. Combined anatomical and functional assessment of the heart is a reality with Novel CZT SPECT CT with major reduction in radiation while maintaining high image quality.
2. Ultralow dose stress only SPECT-CT with CAC with total dose <2mSv feasible.
3. Hybrid SPECT / CTCA : Not sure maybe for the low SDS to prevent ICA and provide one stop shop
CONCLUSION
![Page 48: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/48.jpg)
(may miss soft plaque)
![Page 49: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/49.jpg)
Thank You
![Page 50: HYBRID IMAGING: SPECT / CTnucleus.iaea.org/HHW/NuclearMedicine/IMIC_2013/...2013/10/04 · ray spectrum • E.g., Air = -1000 HU, fat = -100 HU, water = 0 HU, bone = 1000 HU • Need](https://reader034.vdocuments.mx/reader034/viewer/2022042409/5f265e5e9266305de0569922/html5/thumbnails/50.jpg)
50 Copyright ©2003 American Heart Association
Hachamovitch, R. et al. Circulation 2003;107:2900-2907
Log hazard ratio for revascularization (Revasc) vs medical therapy (Medical Rx) as a function of % myocardium
ischemic based on final Cox proportional hazards model
•(Shaw et al JACC;1999 33:661-9) •(Shaw et al JACC;1999 33:661-9)