DePuy Synthes Mitek Sports Medicine Visualization System Setup & Troubleshooting Guide
2
1 : Visualization Front Side Setup
1 POWER ON TOWER
Turn on the cart’s main power switch. Turn on the front side power switch of each individual component.
2 ENTER PATIENT INFORMATION Select “New Case” on the Image Capture Unit.
A surgical team member will then enter relevant patient data using a Keyboard or other external device.
3 STERILE PROCEDURE
A surgical team member working within the sterile field will pass the equipment-end of the sterilized Camera Head Cable and Light Cable to a surgical team member outside the sterile field for attachment to their respective units.
4 ATTACH CAMERA HEAD TO CCU
Remove protective cap from Camera Head Cable. Attach Camera Head Cable to CCU (Camera Control Unit) by aligning the red dots and pushing into the CCU.
5 ATTACH LIGHT CABLE TO LIGHT SOURCE
Make sure that the light turret is turned so the correct position is on top and that it matches the style of the Light Cable. DePuy Synthes Mitek Sports Medicine Light Cables use the Storz connection here.
CAUTIONEnsure the backside wiring setup is complete and that all backside power switches are turned on.
1
2
3
4
5
3
1 : Visualization Front Side Setup
6 ATTACH COUPLER TO CAMERA HEAD Attach the desired Coupler to the Camera Head
by screwing them together. All surfaces should be dried thoroughly before attaching to ensure no fogging occurs.
NOTE: It is recommended to use 2 different gauze pads with a 1 minute wait time between wiping down all surfaces.
7 ATTACH SCOPE TO COUPLER
Attach the desired Scope to the Coupler. All surfaces should be dried thoroughly with gauze pad before attaching.
NOTE: Press locking mechanism to accept desired Scope. Once inserted, push button in opposite direction to lock Scope into place.
8 ATTACH LIGHT CABLE TO SCOPE
The DePuy Synthes Mitek Sports Medicine Light Cable uses an ACMI connection here. All Scope rings should be removed from the Scope’s light-post to access this connection.
NOTE: DePuy Synthes Mitek Sports Medicine Light Cable Connections
Proximal – Storz / Distal – ACMI
9 WHITE BALANCE THE CAMERA
Make sure the Light Source is turned on and not in Standby Mode. Place several gauze pads approximately 1 inch from the Scope tip and wait 1-2 seconds for shutter to adjust.
Then press-and-hold the “+ / WB” button on the Camera Head for approximately 3 seconds. When the White Balance is complete, a green check mark appears over the WB symbol on the Display.
10 ATTACH SHEATH TO SCOPE
Attach the Sheath to the Scope and the Sheath’s Stopcocks to the appropriate FMS® Fluid Management System inflow and outflow connection tubes.
6
7
8
9
10
9
4
2 : Visualization Backside Setup
DVI > DVI
HDMI > DVI
Remote > GPIO
Connect Camera Control Unit to Display
DVI 1 out > DVI 1 in
Connect Camera Control Unit to
Image Management System
DVI 2 out > DVI in
Connect Image Management System to
Display
HDMI out > DVI 2 in
Connect Camera Control Unit to Image
Management System (connect cable with red band to Remote 2)
Remote 1&2 out > MedX EVO GPIO Port
Connect Image Management System
to Printer
USB out > USB in
Connect Image Management System to
Wireless Imaging Control
USB out > USB in
Ethernet NIC > Ethernet in
Connect Display to Wireless Transmitter (Tx) (see next page for secondary display setup)
DVI out > DVI in
Power to all units from power
strip inside tower
1
2
3
4
5
6
7
8
DIG
ITA
L PR
INTE
RIM
AG
E M
AN
AG
EMEN
T SY
STEM
LIG
HT
SOU
RCE
CA
MER
A C
ON
TRO
L U
NIT
DIS
PLA
YW
IREL
ESS
TRA
NSM
ITTE
R (T
x)
Display Power
Printer Power
Power Splitter (for use w/ Wireless Imaging)
Image Management System Power
USB & Ethernet (wireless control power & network)
5
2 : Visualization Backside Setup – Wireless Imaging/Secondary Display
Wireless Setup – Bonding Wireless Units
VISUALIZATION TOWER
WIRELESS TRANSMITTER (Tx)
SECONDARY DISPLAY
WIRELESS RECIEVER (Rx)
Connect Wireless Receiver (Rx) to Secondary Display
DVI 1 out > DVI 1 in
Power to wireless units utilizing power splitter
off display power
1
2
DVI > DVI Power Splitter (for use w/ Wireless Imaging)
Display Power
1 Press and hold the “bond” button (see image 1 for location) on the transmitter (Tx) unit until its LED begins flashing orange rapidly, then release.
NOTE: Ensure both displays are powered on prior to performing the pairing procedure.
Next press and hold the “bond” button (see image 2 for location) on the receiver (Rx) unit until its LED begins flashing orange rapidly, then release.
3 When the bonding process is complete each unit’s LED will be steady blue.
3
2
1 2
6
3 : Troubleshooting
PROBLEM CAUSES POSSIBLE SOLUTION
Device will not power on
Power cable unplugged Ensure power cable is plugged in
Supply Mains switch OFF Power ON Supply Mains switch
Power switch OFF Power ON Power switch
Fuse is blown Replace fuse
Video cart is not powered ON Power ON video cart
Live video does not display
Camera Head (TCKI) not connected Connect Camera Head (TCKI)
Video cable not connected to the device Connect video cable to device output
Video cable not connected to the Display Connect video cable to Display input
Display is not powered ON Power ON Display
Display cannot accept video resolutionUse a display which complies with the
technical specifications
Supply Mains switch powered OFF Power ON Supply Mains switch
Power switch OFF Power ON Power switch
Display not connected to power source Connect Display to power source
Display set to wrong input Set Display to correct input
LCD screen does not function
Firmware faultReset Camera Control Unit (OVBI) by powering OFF and the ON the Supply
Mains Power switch
Supply Mains switch OFF Power ON Supply Mains switch
Power switch OFF Power ON Power switch
Camera indicator light does not turn on
Camera Head (TCKI) not connectedConnect Camera Head Camera Head
(TCKI) ) to device
Camera Head (TCKI) not recognized by device
Ensure the Camera Head (TCKI) is properly connected – Reset Camera
Control Unit (OVBI)
Supply Mains switch OFF Power ON Supply Mains switch
Power switch OFF Power ON Power switch
7
3 : Troubleshooting
PROBLEM CAUSES POSSIBLE SOLUTION
Loss of video
Video output cable disconnected from device
Connect video cable to device output
Video output cable disconnected from Display
Connect video cable to Display input
Display disconnected from power or turned off
Ensure Display is connected to power and turned on
Display settings changed Ensure Display settings are correctly set
Device disconnected from power or powered off
Ensure device is connected to power with Supply Mains switch powered ON –
Power ON Power switch
Video resolution changedVideo resolution changes cause
temporary video loss If video loss persists, reset device
Video timing changedVideo timing changes cause
temporary video loss If video loss persists, reset device
Light Source power off Turn Light Source on
Color bars display Camera Head (TCKI) disconnected Reconnect Camera Head (TCKI)
Incorrect color representation
Improper White BalanceEnsure screen is white and contains detail
Bright white without detail will not white balance correctly
Device is in Fiberscope SpecialtyFiberscope Specialty may cause
poor color quality
Debris on Endoscope tip Clean Endoscope tip
Debris on Device window or Coupler optics
Remove Coupler from device and clean glass surfaces on device and Coupler
Improper video settingsAdjust video settings or reset settings
to factory defaults
Video too dark
Exposure level too lowIncrease exposure until any area in
the image is too bright
Light Source too low Increase Light Source level
Obstruction in light pathCheck Light Cable and Scope for broken
fibers or light obstruction at Scope tip
Brightness level too low Increase the brightness level
Gamma level too low Increase the gamma level
Video too bright
Brightness level set too high Decrease brightness level
Automatic shutter fault Reset Camera Control Unit (OVBI)
Light Source too bright Decrease Light Source brightness
8
3 : Troubleshooting
PROBLEM CAUSES POSSIBLE SOLUTION
Image is not well centered on display
Endoscope not centered on Coupler
Remove Endoscope fromCoupler and re-attach
Image out of focus
Coupler or Endoscope not properly focused from device
Focus Coupler following Coupler IFU instructions
Endoscope not seated on CouplerRemove Endoscope from Coupler
and re-attach
Debris on Endoscope tip Clean Endoscope tip
Debris on device window or Coupler optics
Remove Coupler from device and clean glass surfaces on device and coupler
Foggy imageMoisture/debris on Endoscope tip Clean Endoscope tip
Moisture/debris on device window or Coupler optics
Remove Coupler from device and clean glass surfaces on device and Coupler
Grainy imageEnhance Level too high Reduce Enhance Level
Firmware fault Reset Camera Control Unit (OVBI)
Buttons on device do not function
Device not connected to Camera Control Unit (OVBI)
Connect device to Camera Control Unit (OVBI)
Firmware fault Reset Camera Control Unit (OVBI)
Buttons damagedConsult Warranty/Service
Return information
Keyboard does not function
Improperly connected Keyboard Connect Keyboard correctly
Invalid Keyboard Connect correct Keyboard
Keyboard or Keyboard cord is damaged Repair or replace Keyboard
Firmware fault Reset Camera Control Unit (OVBI)
Images not storing on flash drive
Improperly connected flash drive Connect flash drive correctly
Invalid flash drive Connect correct flash drive
Camera Head (TCKI) buttons not functioning
See “Buttons on device do not function” problem listed above
Firmware fault Reset Camera Control Unit (OVBI)
Flash drive fullRemove images from flash drive using a
computer and retry
Cannot capture video
Image Management System not functioning
Consult Image Management System IFU
Improperly connectedImage Management System
Connect video output cable and remote cable to Image Management System
and Camera Control Unit (OVBI)
9
3 : Troubleshooting
PROBLEM CAUSES POSSIBLE SOLUTION
Cannot capture video (continued)
Invalid Image Management System Connect correct
Image Management System
Camera Head (TCKI) buttons not functioning
See above and re-attach
LCD screen not functioning Reset Camera Control Unit (OVBI)*
Firmware fault Reset Camera Control Unit (OVBI)*
Printer not printing
Firmware fault Consult Printer IFU
Improperly connected PrinterConnect USB cable to Image
Management System and Camera Control Unit (OVBI)
Printer is out of paper or out of ink Add paper, replace cartridge
Invalid PrinterConnect correct Image Management System
Camera Head (TCKI) buttons not functioning
See “Buttons on device do not function” problem listed on page 8
LCD screen not functioning Reset Camera Control Unit (OVBI)*
Firmware fault Reset Camera Control Unit (OVBI)*
Pictures not storing on videocapture device
Image Management System not functioning
Consult Image Management System IFU
Improperly connected Video Capture Device
Connect video output cable and remote cable to Image Management System
and Camera Control Unit (OVBI)
Invalid Image Management System Connect correct Image Management System
Camera Head (TCKI) buttons not functioning
See “Buttons on device do not function” problem listed on page 8
LCD screen not functioning Reset Camera Control Unit (OVBI)*
Firmware fault Reset Camera Control Unit (OVBI)*
FURTHER TROUBLESHOOTINGFor problems which arise that are not remedied by any of the above means, cycle the device power off and on. If this does not solve the problem, please refer to the individual component IFU’s and/or contact your sales consultant or call the 24-Hour Technical Hotline at 1-800-382-4682.
* Please call the 24-Hour Technical Hotline at the number above for step-by-step instructions on how to reset the Camera Control Unit (OVBl).
© DePuy Synthes Mitek Sports Medicine, a division of DOI 2015. All rights reserved. DSUS/MTK/0515/0580(1) 11/15
DePuy Mitek, Inc.325 Paramount DriveRaynham, MA 02767-0350T. +1 (800) 382-4682
www.depuysynthes.com
Limited Warranty and Disclaimer: DePuy Mitek, Inc. products are sold with a limited warranty to the original purchaser against defects in workmanship and materials. Any other express or implied warranties, including warranties of merchantability or fitness, are hereby disclaimed.