![Page 1: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/1.jpg)
Defibrillator Implants:
Bruce Murdoch PhD CHP CLSOArgonne National LaboratoryArgonne, IL [email protected]
AIHce 2011, Portland OR Roundtable Session RT 201
Occupational Exposure to Electromagnetic Radiation
![Page 2: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/2.jpg)
Typical ICD Device
Dimensions: 2” X 2.5” X 0.2”
AIHce 2011, Portland OR
2
![Page 3: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/3.jpg)
Implant Placement
EM Vulnerabilities Leads Control Unit
Failure Modes False Trigger Stopping
AIHce 2011, Portland OR
3
Source: www.defibrillator-help.com
![Page 4: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/4.jpg)
AIHce 2011, Portland OR
4
CASE STUDY
Where? APS (Advanced Photon Source)
Who? Accelerator Operator w/ICD
Why? Work Clearance
![Page 5: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/5.jpg)
AIHce 2011, Portland OR
6
![Page 6: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/6.jpg)
AIHce 2011, Portland OR
7
APS Control Room
![Page 7: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/7.jpg)
Manufacturer Test Standard
ANSI/AAMA PC69 - 2007
AIHce 2011, Portland OR
8
![Page 8: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/8.jpg)
Typical Manufacturer (Medtronic) Exposure Limit Recommendations Static B-Field B ≤ 0.5 mT (5G)
50/60 Hz B ≤ 0.1 mT (1G) E ≤ 6000 V/m
≥ 10 kHz B ≤ 1.25 mT (12.5 G)
R-F E-Field (≥ 150 kHz) E ≤ 100 V/m
AIHce 2011, Portland OR
9
![Page 9: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/9.jpg)
ACGIH 2010 TLV Book Exposure Limit Recommendations
Static B-Field B ≤ 0.5 mT (5G)
ELF (5 Hz–2kHz) B ≤ 0.1 mT (1G)
VLF (2kHz–400 kHz) No recommendation
R-F (100 kHz–3GHZ) E ≤ 3-10 V/m
AIHce 2011, Portland OR
10
![Page 10: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/10.jpg)
Static B-Fields
MetroLab Model ETM-1
3-Axis Magnetometer
AIHce 2011, Portland OR
11
![Page 11: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/11.jpg)
ELF
Holaday HI-3627ELF Magnetic Field Meterw/ HI-3602 Probe
30 Hz – 2 KHz
AIHce 2011, Portland OR
12
![Page 12: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/12.jpg)
VLF
Holaday HI-3637VLF Magnetic Field Meterw/ HI-3091 Probe
2kHz – 400 KHz
AIHce 2011, Portland OR
13
![Page 13: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/13.jpg)
R-F
Narda NBM-520RF Survey Meterw/ EF3091 E-Field Probe
100 kHz – 3 GHz
AIHce 2011, Portland OR
14
![Page 14: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/14.jpg)
Spectrum Coverage
AIHce 2011, Portland OR
15
RF
![Page 15: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/15.jpg)
AIHce 2011, Portland OR
16
The Instrument Ensemble
![Page 16: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/16.jpg)
Survey Plan
Daily Control Room work area
Weekly maintenance walkthroughs
ESH Coordinator guidance
84 specific survey locations plus area sweeps
Full spectrum survey at each location
AIHce 2011, Portland OR
17
![Page 17: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/17.jpg)
Control Room Panels
AIHce 2011, Portland OR
19
![Page 18: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/18.jpg)
Electrical Distribution RoomELF: B ≤ 0.4 G @ 6 inches from transformerRF: E ≤ 4 V/m
AIHce 2011, Portland OR
22
![Page 19: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/19.jpg)
Mezzanine RF System Freq = 325 MHz E(RF) up to 3.5 V/m @ 6 in.
AIHce 2011, Portland OR
23
![Page 20: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/20.jpg)
Power Control Systems
AIHce 2011, Portland OR
24
![Page 21: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/21.jpg)
Pulser Power Transformers
AIHce 2011, Portland OR
25
![Page 22: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/22.jpg)
RF Distribution Waveguides
Freq = 325 mHz
AIHce 2011, Portland OR
27
![Page 23: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/23.jpg)
Linac KlystronsFreq = 2.8565 GHzE(RF) = 2 V/m @ 1 ft
AIHce 2011, Portland OR
29
![Page 24: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/24.jpg)
Case Study Results
Employee barred from
– Fundamental RF Room: 7 V/m RF
– Electrical Distribution Room: 4 V/m RF
Cautions
– Some instrument racks: 4 V/m RF @ 12”
– Some waveguides: 3.5 V/m RF @ 12”
AIHce 2011, Portland OR
30
![Page 25: Defibrillator Implants: Bruce Murdoch PhD CHP CLSO Argonne National Laboratory Argonne, IL USA btmurdoch@anl.gov AIHce 2011, Portland OR Roundtable Session](https://reader035.vdocuments.mx/reader035/viewer/2022081519/56649ea85503460f94bab542/html5/thumbnails/25.jpg)
Credits
Technical Assistance– Mary Malik– Adam Sauter
Photography– George Joch– Edmund Chang
AIHce 2011, Portland OR
31