![Page 1: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/1.jpg)
Products Solutions Services
Calibration scope of Endress+Hauser(Thailand) Laboratory.Service Department
Slide 1 Service Department (Calibration Laboratory)
![Page 2: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/2.jpg)
Index
• In-house calibration / Parameter / Method
• On-site calibration / Parameter / Method
• Traceability
• Quality insurance
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 2 Service Department (Calibration Laboratory)
![Page 3: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/3.jpg)
Calibration capability of flow laboratory
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 3 Service Department (Calibration Laboratory)
35 m³/h
50 m³/h
60 m³/h
0 10 20 30 40 50 60 70
COMPACT FLOW RIG*ON SITE
MOBILE FLOW RIG*IN-HOUSE
LAB (NONTHABURI)IN-HOUSE
*Not in accreditation ISO17025
Calibration capability of flow laboratory
Master meter premium accuracy ±0.05%o.r.
![Page 4: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/4.jpg)
Calibration methods – secondary calibration method
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 4 Service Department (Calibration Laboratory)
![Page 5: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/5.jpg)
Calibration capability of flow laboratory
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 5 Service Department (Calibration Laboratory)
![Page 6: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/6.jpg)
Calibration capability of flow laboratory
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 6 Service Department (Calibration Laboratory)
![Page 7: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/7.jpg)
Calibration capability of flow laboratory (Rayong)
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 7 Service Department (Calibration Laboratory)
HMI & Data acquisition software
(Flow-X)
Master MeterPromass 300
![Page 8: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/8.jpg)
Calibration capability of pressure laboratory
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 8 Service Department (Calibration Laboratory)
-0.9 *
-0.9
70 *bar g
70 *bar g
10 bar g
-1 9 19 29 39 49 59 69 79
NOT IN ACCREDITATION ISO17025ON SITE
NOT IN ACCREDITATION ISO17025IN-HOUSE
LAB (NONTHABURI)IN-HOUSE
*Not in accreditation ISO17025
Positive gauge pressure Negative gauge pressure
![Page 9: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/9.jpg)
Calibration capability of pressure laboratory
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 9 Service Department (Calibration Laboratory)
UUC
Precision Pressure Indicator (PACE1000)
LoopCalibrator
Pressure Regulator
Nitrogen Supply
PACE1000 Accuracy @ module 10 bar <0.003 bar@ module 200 mbar <0.2 mbar
![Page 10: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/10.jpg)
Calibration capability of temperature laboratory
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 10 Service Department (Calibration Laboratory)
-20 *˚C
-20*˚C
-20˚C
150˚C
200˚C
150˚C
-30 0 30 60 90 120 150 180 210
WITH DRY BLOCK CALIBRATORON SITE
WITH STANDARD PROBEIN-HOUSE
WITH DRY BLOCK CALIBRATORIN-HOUSE
*Not in accreditation ISO17025
![Page 11: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/11.jpg)
Calibration capability of temperature laboratory
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 11 Service Department (Calibration Laboratory)
Uncertainty ±0.23˚C
PRT accuracy ±0.03˚C
![Page 12: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/12.jpg)
• Capability
• Units Under Calibration (UUC)
➢ Flowmeter type of UUC
➢ Coriolis Mass Flowmeter
➢ Magnetic Flowmeter
➢ Vortex Flowmeter
➢ Ultrasonic Flowmeter
➢ Size DN08 to DN100 (3/8” to 4”)
Compact Flow Rig
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Con.>>>
Slide 19 Service Department (Calibration Laboratory)
![Page 13: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/13.jpg)
On-Site calibration
Calibration scope of Endress+Hauser(Thailand) Laboratory.
• Analysis measurement parameter
➢ pH
➢ Conductivity
➢ OPR
➢ Turbidity
➢ Dissolved Oxygen
➢ Chlorine
▪ Level measurement parameter
➢ Comparison with Laser Distance meter or Steel tape
Slide 24 Service Department (Calibration Laboratory)
![Page 14: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/14.jpg)
Customer’s
Equipment
Working
Standards
Secondary Standards
National Standard METAS
Endress+HauserFlowtec AGSwitzerland
NIMT
Promass 83 F DN08
100…2000 kg/h
Promass 83 F DN40
4500…34000 kg/h
Promass 83 F DN50
7000…60000 kg/h
Promass 83 F DN25
1800…11000 kg/h
Promass 83 F DN15
650…6000 kg/h
TPATechnology Promotion Association
(Thailand-Japan).
T-001inlet
T-002outlet
P-001
Flowcal XpertF-011
Pulse/Current signal
Liquid Flow Meter DN08...100 ( 3/8"….4")
Scope Range 100...60,000 kg/h or 100...60,000 l/h
Druck”UPSIII”_LC-001Loop Calibrator
F-001
F-003F-004
F-005F-006
F-007F-008
F-009F-010
Traceability (Flow)
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 25 Service Department (Calibration Laboratory)
![Page 15: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/15.jpg)
Traceability (Flow)
• Master Meter : Coriolis mass flowmeter (Endress+Hauser)
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 26 Service Department (Calibration Laboratory)
![Page 16: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/16.jpg)
Traceability (Pressure & Temperature)
Calibration scope of Endress+Hauser(Thailand) Laboratory.
• Traceable to National International Metrology (THAILAND), NIMT
• Precision standard pressure >>> Rockertek (Thailand)>>>NIMT
• Dry Block Calibrator >>> Unithai >>>NIMT
Slide 27 Service Department (Calibration Laboratory)
![Page 17: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/17.jpg)
Traceability (Pressure)
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 28 Service Department (Calibration Laboratory)
![Page 18: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/18.jpg)
Traceability (Temperature)
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 29 Service Department (Calibration Laboratory)
![Page 19: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/19.jpg)
Quality insurance
• Training Service engineer
• Proficiency Test (Flow, Pressure, Temperature) with Endress&Hauser
World W…. Group
➢ Flow ;Since 2011 to 2020
➢ Pressure ;Since 2016 to 2020
➢ Temperature ;Since 2016 to 2020
➢ Analysis (pH & Conductivity) ; Since 2020
▪ Internal & External audit
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 30 Service Department (Calibration Laboratory)
![Page 20: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/20.jpg)
Proficiency test
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 31 Service Department (Calibration Laboratory)
![Page 21: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/21.jpg)
Proficiency test
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 32 Service Department (Calibration Laboratory)
![Page 22: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/22.jpg)
Quality improvement
• Migration ISO/IEC 17025: 2005 to 2017 (New version)
✓ Submit to TISI on July 17, 2020.
• Go-on to Accreditation for on-site calibration
➢ Flow (mobile & compact flow rig)
➢ Pressure
➢ Temp
➢ pH, Conductivity
Calibration scope of Endress+Hauser(Thailand) Laboratory.
ISO/IEC 17025 Migration
2005 >>> 2017
TISI Accreditation Certificate
Quality
Slide 33 Service Department (Calibration Laboratory)
![Page 23: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/23.jpg)
Mini Wet Flow Calibration Rig
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 34 Service Department (Calibration Laboratory)
• e.g. Promass A
DN Measuring range Test point (min-max)
01 0 to 20 kg/h 5 to 20 kg/h
02 0 to 100 kg/h 10 to 100 kg/h
04 0 to 450 kg/h 50 to 450 kg/h
![Page 24: Calibration scope of Endress&Hauser(Thailand) Laboratory](https://reader031.vdocuments.mx/reader031/viewer/2022020701/61f6f97b01dcf71eb03383b0/html5/thumbnails/24.jpg)
Calibration scope of Endress+Hauser(Thailand) Laboratory.
Slide 35 Service Department (Calibration Laboratory)
Thank you for your attention