Page 1: Application 325-00 Agilent Extended Natural Gas Analyzer · Application Highlights ... Methods 2177, 2261, 2186, and 2286

Application Highlights• A Flame Ionization Detector (FID) to detect the C1

through C12 n-paraffins to a lower detection limit of1 ppm, except for trace peaks eluting on the tail of amajor component.

• A Thermal Conductivity Detector (TCD) to detect aircomposite, carbon dioxide, C1 through C5 paraffins withan initial C6+ composite backflush to detector.

• System configured to meet Gas Processors AssociationMethods 2177, 2261, 2186, and 2286.

• Approximate analysis time is 30 minutes.

Optional Configurations• Detailed hydrocarbon analysis of extended natural gas

• TCD/FID/FPD or TCD/FID/SCD for extended naturalgas with trace sulfur analysis

• TCD/TCD/FID for extended natural gas with helium orhydrogen

Application 325-00

Agilent Extended Natural Gas Analyzer

For More Information

For more information on our products and services, visit ourWeb site at www.agilent.com/chem.

Technical Overview

1 Methane2 Ethane3 Propane4 Isobutane5 n-Butane6 Isopentane7 n-Pentane8 Hexane9 Benzene10 nC7+

11 Toluene12 nC8+13 Ethylbenzene14 m/p-Xylene15 o-Xylene16 nC9+17 nC10+18 nC11+19 nC12+

FID output

TCD output

1

2

2

11

3

3

4

41

5

5

6

6

7

8

7

8

910

9 10

12

13

14

15

16

17 18 19

1 C6+ backflush2 Air composite3 Methane4 Carbon dioxide5 Ethane6 Propane7 Isobutane8 n-Butane9 Isopentane10 n-Pentane

Page 2: Application 325-00 Agilent Extended Natural Gas Analyzer · Application Highlights ... Methods 2177, 2261, 2186, and 2286

Agilent shall not be liable for errors contained herein or for incidental or consequen-tial damages in connection with the furnishing, performance, or use of thismaterial.

Information, descriptions, and specifications in this publication are subject to changewithout notice.

© Agilent Technologies, Inc. 2002

Printed in the USADecember 6, 20025988-6736EN

www.agilent.com/chem

1 Methane2 Ethane3 Propane4 Isobutane5 n-Butane6 Isopentane7 n-Pentane8 Hexane9 Benzene10 nC7+

11 Toluene12 nC8+13 Ethylbenzene14 m/p-Xylene15 o-Xylene16 nC9+17 nC10+18 nC11+19 nC12+

FID output

TCD output

1

2

2

11

3

3

4

41

5

5

6

6

7

8

7

8

910

9 10

12

13

14

15

16

17 18 19

1 C6+ backflush2 Air composite3 Methane4 Carbon dioxide5 Ethane6 Propane7 Isobutane8 n-Butane9 Isopentane10 n-Pentane

FID and TCD output from the Agilent Extended Natural Gas Analyzer.


Top Related