A novel minimal invasive approach The way forward
Dr. Uditha Hewarathna
Ajith Kularatne, Gnanamoorthi Mayurathan,Uditha Hewarathna, Shanike Karunaratne,Narada Kodithuwakku
Teaching Hospital, Kandy,
Sri Lanka
![Page 2: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/2.jpg)
Mr. K
14 years old
Presented with a infected wound in left leg
Incidentally found to be having polycythaemia (Hb - 19.9 g/dl)
![Page 3: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/3.jpg)
On Examination
◦ Cyanosed
◦ Plethoric
◦ Clubbing in B/L UL and LL
◦ Oxygen saturation 62% on air
CRP - negative
ESR - 3 mm/ 1st hour
Blood picture - secondary polycythemia
![Page 4: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/4.jpg)
Transthoracic 2D echo ◦ Normal size cardiac chambers
◦ EF > 60%
◦ No ASD, VSD or PDA
◦ No evidence of PHT or TAPVD
◦ ? Fistulous connection between pulmonary artery and left atrium
TOE
◦ No ASD or VSD
◦ Normal pulmonary venous drainage
Normal Lung function test
![Page 5: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/5.jpg)
![Page 6: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/6.jpg)
![Page 7: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/7.jpg)
![Page 8: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/8.jpg)
![Page 9: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/9.jpg)
![Page 10: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/10.jpg)
![Page 11: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/11.jpg)
![Page 12: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/12.jpg)
Surgical closure
Transcatheter closure
![Page 13: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/13.jpg)
Major operation
Has potential surgical and anesthetic complications
Patient and mother not given consent for the surgery
![Page 14: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/14.jpg)
Has two options
◦ Use a covered stent across the fistula in the right pulmonary artery
◦ Device AV fistula closure
![Page 15: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/15.jpg)
Mechanism
◦ Fistula will have no feeding
Problems
◦ Covered stent more expensive and patient couldn’t afford
◦ Due to stagnation of blood in AV fistula connection there is a risk of thrombosis
◦ Risk of embolization of the thrombus
![Page 16: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/16.jpg)
Problems
◦ Largest Amplatzer vascular plug device available is 22mm and fistula was 18 ×20mm in size (They recommend 50% oversizing for these type of fistula)
◦ When a larger device is used, it is very long and device will be jutting out into LA and RPA
◦ Risk of thrombus formation
◦ Risk of obstruction
![Page 17: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/17.jpg)
Finally planned to use atrial septal occluder device to close the fistula
Atrial septal occluder device in LA is safe
Available in different sizes
![Page 18: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/18.jpg)
![Page 19: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/19.jpg)
9F long guiding sheath and dilator were advanced over the guide wire through IVC, RA, RV, MPA, RPA, fistulous connection to left atrium
![Page 20: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/20.jpg)
Sending ASO device through delivery system
![Page 21: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/21.jpg)
Deploying 16mm ASO device -unsuccessful
![Page 22: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/22.jpg)
Deploying 20mm ASO device
![Page 23: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/23.jpg)
Deployed 20 mm device obstructing right pulmonary artery
![Page 24: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/24.jpg)
Deployed 20 mm device obstructing right pulmonary artery
![Page 25: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/25.jpg)
Successful deploying of 16mm ASO device
![Page 26: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/26.jpg)
Soon after deploying the device
![Page 27: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/27.jpg)
Soon after deploying the device checking the pulmonary flow and closure of AV fistula
![Page 28: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/28.jpg)
Soon after deploying the device checking the pulmonary flow and closure of AV fistula
![Page 29: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/29.jpg)
Patient’s saturation improved from 62% to 94% on air
Transthoracic echo confirmed there was no turbulence or pressure gradient in the right pulmonary artery
TOE confirmed normal pulmonary venous drainage.
![Page 30: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/30.jpg)
10 min after deploying the device checking the pulmonary flow and closure of AV fistula
![Page 31: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/31.jpg)
10 mm after deploying the device checking the pulmonary flow and closure of AV fistula
![Page 32: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/32.jpg)
10 mm after deploying the device checking the pulmonary flow and closure of AV fistula
![Page 33: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/33.jpg)
Releasing of device by counterclockwise rotation of the delivery cable after checking the stability
![Page 34: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/34.jpg)
After releasing of the device
![Page 35: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/35.jpg)
Bubble contrast echo after the procedure
![Page 36: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/36.jpg)
Dual antiplatelets for 1 month followed by 6 month single antiplatelet used to prevent thrombo-embolism
At 6 months
◦ Patient asymptomatic
◦ Fluoroscopy show ASO device in same position
◦ No shunt in transthoracic echo
![Page 37: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/37.jpg)
Direct communication between the right pulmonary artery and the left atrium is an unusual variation of a pulmonary arteriovenous fistula1
In fact, there are only about 50 cases reported in the literature2
This is one condition in which clinical examination may reveal only cyanosis1
Contrast echocardiography may show only right-to-left shunt, and only angiography is truly diagnostic
1Vladimir AM, Ingo D, Stanislav O, Roland H. Right pulmonary artery to left atrium communication, a rare cause of systemic cyanosis. Texas Heart Inst J 2001;28:122–124. 2Krishnamoorthy KM, Rao S. Pulmonary artery to left atrial fistula. Eur J Cardiothorac Surg 2001;20:1052–1053
![Page 38: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/38.jpg)
Two treatment options were available either surgical closure or transcatheter closure
Most of patient has underwent surgical closure1
Coil embolization, Amplatzer vascular plug device have been used to close the AV fistulae2,3
1Zeebregts CJ, Nijveld A, Lam J, van Oort AM, Lacquet LK. Surgical treatment of a fistula between the right pulmonary artery and left atrium: presentation of two cases and review of literature. Eur J Cardiothorac Surg 1997;11:1056 –1061. 2Slack MC, Jedeikin R, Jones JS., Transcatheter coil closure of a right pulmonary to left atrial fistula in an ill neonate. Catheter Cardiovasc Interv 2000;50:330–333. 3Francis E, Sivakumar K, Kumar RK. Transcatheter closure of fistula between the right pulmonary artery and left atrium using the Amplatzer duct occluder. Cathet Cardiovasc Interv 2004;63:83–86.
![Page 39: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/39.jpg)
Our patient has 18×20mm fistulous connection and largest Amplatzer vascular plug device available is 22mm (They recommend 50% oversizing for these type of fistula)
So, fistulous connection in our patients were too wide to be closed safely and completely using the above devices
We chose ASO because it was the only large enough device available to us to occlude the large fistulous connection seen in our patients.
![Page 40: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/40.jpg)
The other reasons were ◦ Availability in different sizes
◦ Easy implantation
◦ Anchorage stability offered by both the discs
◦ Retrieval as well as repositioning capability before the final release
Our case is unique because the communication was closed in an antegrade fashion without transseptal puncture
![Page 41: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/41.jpg)
Large and short Pulmonary AV fistule can be closed safely, effectively, and nonsurgically with atrial septal occluder device
Placing a atrial septal occluder device in pulmonary artery without obstructing the flow is safe
It is a effective method for our country with poor resource setting
![Page 42: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/42.jpg)
“I cant put into word the comfort and consolation I feel now. It like been in the heaven from hell”
![Page 43: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/43.jpg)
Ajith Kularatne and Gnanamoorthi Mayurathan who are primary operators in this case
My trainers in TH Kandy ◦ Dr. Rohini Tennakoon
◦ Dr. Subhashini Jayawickreme
◦ Dr. Gamini Weerakoon
◦ Dr. Srinath Dolapihilla
All doctors in cardiology unit Kandy
All staff in the Cath lab TH Kandy
![Page 44: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/44.jpg)
Device Size/ Waist Diameter (mm) A
Waist Length (mm) B
Right Atrial Disc Diameter (mm) C
Left Atrial Disc Diameter (mm) D
15 4 25 29
16 4 26 30
17 4 27 31
18 4 28 32
19 4 29 33
20 4 30 34
![Page 45: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/45.jpg)
Device Diameter (mm) A
Unconstrained Device Length (mm) B
12 9
14 10
16 12
18 14
20 16
22 18
![Page 46: A novel minimal invasive approach The way forward](https://reader033.vdocuments.mx/reader033/viewer/2022061503/629e95b488e66a34095d5733/html5/thumbnails/46.jpg)