BIO-SIGNALS
*The bio-signals potentials are generate atcellular level and the source of thesepotentials is ionic in nature.
* cell consists of an ionic conductor
separated from the outside environment bya semi permeable membrane which acts as aselective ionic filter to the ions. This means
that some ions can pass through themembrane freely where as other cannot doso.
-
7/30/2019 8-ECG fni$
2/61
*All living matters are composed of cellsdifferent types.
* The bio-signals are EMG, EOG, ECG, orEEG this signals can be analyzed andsensors.
*This signal can monitoring ,rehabilitation ,feedback functions and control gain .
* EMG is electro mayo graph its detect
contraction of the skeletal muscles result ingeneration of action potentials in theindividual muscle fibers.
-
7/30/2019 8-ECG fni$
3/61
*EOG is electro oculo graph its detect thepotentials that generated by the movement
of the eye ball.
The EEG electro cardio gram its detect thebrain waves which indicate the function of
the brain.
-
7/30/2019 8-ECG fni$
4/61
ELECTRICAL SYSTEM OF HEART
-
7/30/2019 8-ECG fni$
5/61
-
7/30/2019 8-ECG fni$
6/61
THE ECG WAVEFORM
-
7/30/2019 8-ECG fni$
7/61
ECG BASICS
*P wave: represents the depolarizationimpulse across the atria
*Q, R and S waves: all these three wavesrepresent the ventricular depolarization
*T wave: represents the repolarization of
the ventricles
-
7/30/2019 8-ECG fni$
8/61
ECG Basics
*P wave : 0.25 mv
*R wave : 1.60 mv
*Q wave : 25% of R-wave
*T wave : 0.1 to 0.5 mv
-
7/30/2019 8-ECG fni$
9/61
ECG Basics
*P-R Interval : 0.12 to 0.20 s
*Q-T Interval : 0.35 to 0.44 s
*S-T Segment : 0.05 to 0.15 s
*P wave : 0.11 s
*QRS Interval : 0.09 s
-
7/30/2019 8-ECG fni$
10/61
Electro cardio graph
*ECG is an instrument which records the electrical
activity of the heart.
Electrical signals from the heart characteristicallyprecede the normal mechanical function and
monitoring of these signals has greatsignificance.
*ECG provides valuable information about widerange of cardiac disorders such as the presence
of an inactive part (infarctions) or anenlargement (cardiac hypertrophy) of the heartmuscle.
-
7/30/2019 8-ECG fni$
11/61
-
7/30/2019 8-ECG fni$
12/61
Block diagram
*The potentials pick up by electrodes are taken to
the lead selector switch. In the lead selector theelectrodes are selected two by two according tothe lead program.
*The signal is connected symmetrically to the pairdeferential amplifier. The pre amplifier is usually athree or four stages the output signal pick topower amplifier.
*Direct writing recorder is usually adequate sincethe ESG signal of the interested has limited hasbandwidth.
-
7/30/2019 8-ECG fni$
13/61
Frequency selective network is an R-C network.Which provides necessary damping of the penmotor and is preset by the manufacture.
*The auxiliary circuit provide a 1mV calibrationsignal and automatic blocking of the amplifierduring a change in the position of the lead switch.
*It may include a speed control circuit for thechart drive motor.
-
7/30/2019 8-ECG fni$
14/61
*Isolated preamplifier :
It had traditional for all ECG to have the right leg(RL) electrode connected to the chassis . And fromthere to the ground.
*this provided a ready path for any groundseeking current through the patient and presentedan electrical hazard.
-
7/30/2019 8-ECG fni$
15/61
ECG LEADS
*Two electrodes placed over different areas of theheart and connected to the galvanometer willpicket up the electrical current resulting from thepotentials difference between them.
*So that any two sides due to electrical activity ofthe heart is called (LEAD)
-
7/30/2019 8-ECG fni$
16/61
-
7/30/2019 8-ECG fni$
17/61
Bipolar lead:
ECG recording by using two electrodes such
that the final trace corresponds to the difference ofelectrical potentials existing between them. Theyare called standard leads
* I lead the electrodes are placed on theright and the left arm (RA&LA).
* II lead the electrodes are placed on theright arm and left leg (RA&LL).
* III lead the electrodes are placed on theleft arm and left leg (LA & LL).
-
7/30/2019 8-ECG fni$
18/61
Unipolar limb leads:
*Also know as augmented limb leads, examine the
composite potentials from all three limbssimultaneously.
*in all three augmented leads, the signals from
two limbs are summed in resistor network andthen applied to the amplifiers inverting input.
*the signal from the remaining limb electrode isapplied to the noninverting input.
-
7/30/2019 8-ECG fni$
19/61
*Lead aVR: RA is connected to the no invertinginput while LA and LL are summed at the invertinginput
*Lead aVL: LA is connected to the no invertinginput, while RA and LL are summed at the
inverting input.
*Lead aVF: LL is connected to the no inverting
input while RA and LA are summed at theinverting input.
-
7/30/2019 8-ECG fni$
20/61
Unipolar chest leads:
V1 through V6 are measured with the signalsfrom certain specified location on the chestapplied to the amplifier no inverting input.
-
7/30/2019 8-ECG fni$
21/61
Leads Combination
* LEAD I :LA+RA
* LEAD II :LL+RA
* LEAD III :LL+LA
* AVR :RA & LA+LF
* AVF :LF & RA+LA
* AVL :LA & RA+LF
-
7/30/2019 8-ECG fni$
22/61
-
7/30/2019 8-ECG fni$
23/61
Electrode Placement
-
7/30/2019 8-ECG fni$
24/61
Electrode Placement
* v1 - the fourth intercostals space on theright side of the sternum
* v2 - the fourth intercostals space on theleft edge of the sternum
* v3 half way between C2 and C4
-
7/30/2019 8-ECG fni$
25/61
Electrode Placement
* v4 - the fifth intercostals space in the leftcentral clavicular line
* v5 - straight line from the C4 pointperpendicularly to the left front auxiliaryline in the intersection point with this line.
* v6 - at the same level as C5, but in leftcentral auxiliary line
-
7/30/2019 8-ECG fni$
26/61
ELECTRODES:
*Bioelectric events have to be picked up
from the surface of the body before theycan be put into the amplifier forsubsequent record or display.
*This done by using electrodes.*Electrodes make a transfer from theionic conduction in the tissue to the
electronic conduction which is necessaryfor making measurements.
-
7/30/2019 8-ECG fni$
27/61
*They are two types of the electrodes areused in practice surface electrodes and thedeep seated electrodes.
*The surface electrodes pick up the potentialdifference from the tissue surface when placedover it without damping the live tissue.
*Whereas the deep-seated electrodes indicatethe electric potential difference arising inside
the live tissue or cell.
-
7/30/2019 8-ECG fni$
28/61
Electrode-Tissue Interface:
*used electrodes in patient monitoring and
related studies are surface electrodes.
*The notable examples are when they are
used for recording ECG, EEG and respiratoryactivity by impedance pneumography.
*The characteristics of a surface electrodes ofa metal electrode and attached to the surfacebody through electrolyte jell depend on thecondition of metal-electrolyte interface .
-
7/30/2019 8-ECG fni$
29/61
Metal electrolyte interface :
*They are tendency for each electrodes todischarge ions to the solution and for ionsin the electrolyte to combine with eachelectrode.
*That creations a charge gradient(difference- potentials)
-
7/30/2019 8-ECG fni$
30/61
Electrodes tissue interface
Tissue Electrolyteskininterface
Electrolyte MetalElectrolyte
Interface
ToInstruments
-
7/30/2019 8-ECG fni$
31/61
ECG electrodes:
1- Limb electrodes
* rectangular surface* German silver
* impedance is 2 to 5 K
* using in surgery
2- Floating electrodes:
* used without jell
* contact impedance 50K
3- Pregelled Disposable Electrodes:* stress testing ,long term monitoring
* reducing possibility of artifacts and drift
-
7/30/2019 8-ECG fni$
32/61
Limb Electrode for ECG
-
7/30/2019 8-ECG fni$
33/61
Floating Electrode
-
7/30/2019 8-ECG fni$
34/61
ECG Machine Block Diagram
-
7/30/2019 8-ECG fni$
35/61
Leads Selector
A multiplexer is a logic circuit which accepts
several data inputs and outputs, but only one ofthem at a time. In essence, it behaves as multi-position switch which operates under the controlof select or address inputs, the figure show
Inputs output
Select input
-
7/30/2019 8-ECG fni$
36/61
Preamplifier
-
7/30/2019 8-ECG fni$
37/61
Operation Amplifier
-
7/30/2019 8-ECG fni$
38/61
Galvanometer:* A galvanometer is a permanent magnet movingcoil (PMMC).
* The pen will at rest in the centre of its travelwhen no current flows in the coil. The movementof the pin is depend on the current supply to thecoils.
they are also other types of writing:
* hot-tip (styles) heated by resistor wire.
* the thermal paper its turn black when its heat
* writing knife edge in the thermal recorder.
-
7/30/2019 8-ECG fni$
39/61
* Dot matrix printer used with the computer its 27pin printer make good quality recording waveformand numerical number like blood pressure,temperature etc.
-
7/30/2019 8-ECG fni$
40/61
Motors:
*The motor is connected to the drive rollerthrough either a drive a chain or a gear train.
*Most machines use ac motors, with tappedwindings to select drive speed.
*Some ac motors provide very accurate drivespeeds that are synchronized to the ac power
mains frequency 60 Hz in the united states.
-
7/30/2019 8-ECG fni$
41/61
*Only few models use dc motors, and thoseregulated speed by using a regulated dc powersupply.
* in some cases an alternator /tachometer onthe motor shaft to provide negative feedback.
-
7/30/2019 8-ECG fni$
42/61
Figure 239 small
-
7/30/2019 8-ECG fni$
43/61
-
7/30/2019 8-ECG fni$
44/61
-
7/30/2019 8-ECG fni$
45/61
-
7/30/2019 8-ECG fni$
46/61
ECG FAULTS AND TROUBLESHOOTING:-
Most common problem that occur in an ECG are:
PROBLEM 1:-
SYMPTOM: machine runs but the thermal tip stylus doesnot writes or writes very lightly.
POSSIBLE CAUSE:(1) Too little heat on the stylus tip .
(2) Insufficient stylus pressure.
TROUBLESHOOTING: Use a screw driver or anyisolated tool to gently press the stylus if a dark line appearson the paper the problem is pressure , but if no dark lineappears the problem is heat.
-
7/30/2019 8-ECG fni$
47/61
ECG FAULTS AND TROUBLESHOOTING:-
SOLUTION:
(1) For no heat check the heater voltage at the stylus
,if voltage is correct change the stylus ,if voltage is
not correct refer the service manual for details on
stylus power drive.(2) Adjust the stylus pressure , use a pressure gage
and refer to the service manual for the correct
value.on some models pressure must be made at a
specific heater voltage.
-
7/30/2019 8-ECG fni$
48/61
ECG FAULTS AND TROUBLESHOOTING:-
PROBLEM 2:-SYMPTOM: Smeared trace.
POSSIBLE CAUSE:
Worn stylus or incorrectly loaded paper.
TROUBLESHOOTING:
Check paper loading and if proper ,check stylus for wear
,pitting and other irregularities.
Incorrect loading is one of the most common fault and
often results from bypassing the paper brake or tension
bar.
-
7/30/2019 8-ECG fni$
49/61
PROBLEM 3:-
SYMPTOMS: Poor recording.
POSSIBLE CAUSE: Electronic or mechanicalproblem , bad switch or patient cable.
TROUBLESHOOTING:(1) Place lead selector switch in STD ,short allelectrodes together and press 1- mv cal button.
(2) If normal calibration pulse appears then problem
is connection to the patient.(3) If problem persists then repeat step 1 using knowngood patient cable .
ECG FAULTS AND TROUBLESHOOTING:-
-
7/30/2019 8-ECG fni$
50/61
ECG FAULTS AND TROUBLESHOOTING:-
60 Hz INTERFERENCE:-
CAUSES: 60 Hz interference appears due to power
mains,another cause of 60 Hz interference if a broken
or loose cable or broken power main ground on the
ECG machine, additionally some dc supply alsocause these artifacts.
SOLUTION:
The problem can be isolated by shorting together all
electrodes of the patient cable and checking eachposition of the lead selector switch.
-
7/30/2019 8-ECG fni$
51/61
ECG FAULTS AND TROUBLESHOOTING:-
(1) If the interference ceases , then the problem is a badelectrode or no electrolytic gel.
(2)If the interference exists on all positions of the lead
selector switch then the problem is internal of the
machine.
(3) If the problem occurs in certain lead position of the
selector switch then suspect an open wire, useohmmeter for continuity.
St ECG
-
7/30/2019 8-ECG fni$
52/61
Stress ECG
*Some potentials dangerous cardiac arrhythmiaand other anomalies shown up only under stressconditions.
*Physicians examine the stress condition by placing
the patient on the treadmill or stair stepper whilemonitoring the patient's ECG waveform on both anoscilloscope monitor and a paper chart.
*Modern stress ECG machines are usually equippedwith a microcomputer that analyzes the waveform .
* h d f h
-
7/30/2019 8-ECG fni$
53/61
*A common method is to perform the stress test,recording the ECG waveform, and then follow thestress test with a thallium scan.
*The radioactive thallium is taken up by healthycardiac cells, so areas of the heart where blood
flow is sufficient appear darker on a gammacamera display then healthy areas.
*These two tests allow the physician to evaluatethe existence, location and extent of cardiacdisease.
-
7/30/2019 8-ECG fni$
54/61
-
7/30/2019 8-ECG fni$
55/61
Manual Controller
The optional manual controller
Allows personal treadmill operation, whilegiving the user control of speed and
elevation functions without interferingwith
medical usage.
-
7/30/2019 8-ECG fni$
56/61
Drive Assembly
The low profile design and simplified layout of
drive assembly components provide easyaccessibility for maintenance and provides ahigh level of performance.
di l
-
7/30/2019 8-ECG fni$
57/61
displaysBody weight, Time, Distance ,Elevation
Total calories, Calories per minutePace (minutes/mile), Heart rate.
-
7/30/2019 8-ECG fni$
58/61
ECG PARAMETER
ECG leads 12 standard Record speed 5/25/50
Sensitivity 2.5/5/10/20
-
7/30/2019 8-ECG fni$
59/61
Operations and Controls
INST: No signal gets recorded when this ispressed
ADV: Pressing this forward the leadposition by one lead.
REV: Pressing this takes the lead position
by one lead.
-
7/30/2019 8-ECG fni$
60/61
TECHNICAL SPECIFICATIONS
GENERAL Dimensions 297 x 210 x 80mm Weight
4.8 kg with battery POWER SUPPLYAC Mains 180 - 260 VAC
Battery Back-up 100 m
-
7/30/2019 8-ECG fni$
61/61
TECHNICAL SPECIFICATIONS
RECORDING Power Thermo-sensitivePaper with Recording Consumption :
50 mm (Recording 40 mm) 25 and 50mm/sec. (switch selection)
LEAD SELECTION 12 lead selection
Selector Positions Test, I, II, III, aVR, aVL,aVF, V