AND HOW NOT TO GET USED BY THEM….
The Tools we use
![Page 2: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/2.jpg)
RULE No. 1
Tools supplement clinical thinking and details subjective and objective assessments, they do not replace them.
AKA: Treat the patient, NOT THE MONITOR (or the glucometer, or the pulse ox…)
![Page 3: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/3.jpg)
RULE No. 2
Tools fail if you: Don’t take care of them Don’t check them out EVERY SHIFT Don’t know how to use them Don’t know how to NOT use them Don’t read about the technology behind them…
AKA: Murphy is a son of a bitch waiting to screw with you and your equipment….
![Page 4: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/4.jpg)
What we will cover…
StethoscopeSphygmomanometerGlucometerSPO2/SPCOEKGETCO2
![Page 5: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/5.jpg)
Stethoscope
![Page 6: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/6.jpg)
Stethoscope
Prior to the advent of the stethoscope, there were no reliable, non-invasive (and therefore non-lethal) methods to assess certain parts of the human anatomy.
![Page 7: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/7.jpg)
A Stethoscope is used by Auscultation (listening) Sound is simply interpretation of vibrations by the
specialized nervous system structures that send data to the brain.
Therefore any method that limits outside vibration and noise will increase the fidelity of auscultation… Position to the patent Length and thickness of tubing on stethoscope Heels off ground in moving ambulance Environmental factors
![Page 8: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/8.jpg)
Auscultation
![Page 9: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/9.jpg)
Auscultation
1 1
2 23 3
A4 4
5 5
6 6
H
![Page 10: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/10.jpg)
Variations
ElectronicRecordingDuo-stethascope (for
teaching)Fetoscope, AKA A
Pinard Horn
![Page 11: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/11.jpg)
Sphygmomanometer
Rev. Stephen Hale's experiments to determine the blood pressure of a horse. Circa 1733
![Page 12: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/12.jpg)
Sphygmomanometer
When we talk about blood pressure, we are usually speaking of Non-Invasive Blood Pressure (NIBP).
This is used to estimate the Mean Arterial Pressure (MAP) and thus estimate Aortic Pressure.
![Page 13: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/13.jpg)
Key Point: Blood Pressure (NIBP) is an indirect estimate of central
perfusion pressures
![Page 14: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/14.jpg)
Blood Pressure (NIBP)
Two basic types Manual
Mercury Aneroid
Automatic
![Page 15: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/15.jpg)
Blood Pressure (NIBP)
Systolic Diastolic Mm HgPulse pressureMAP
![Page 16: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/16.jpg)
Blood PressureMeasure of pressure against SYSTEMIC vessel walls Systolic
Pressure created in the arteries Reported first
Diastolic Left ventricle relaxed and refilling Reported second
![Page 17: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/17.jpg)
Pulse Pressure
Generally speaking it is the pressure that produces a pulse that we can feel.
Specifically it is pulse pressure is calculated by subtracting the diastolic pressure from the systolic pressure. SBP – DPB = Pulse Pressure
Significant for: A PP is considered LOW if it is less than 25% of the SBP. A narrowing PP is significant for progressing cardiac
tapenade, blood loss, or Tension Pneumothorax A high PP is significant for cardiac disease, HTN, and cor
pulmonale
![Page 18: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/18.jpg)
Korotkoff Sounds in Blood Pressure Korotkoff sounds
Sounds heard during auscultation of the blood pressure Caused by the jetting of blood through the collapsed
artery. Systolic pressure: the first sounds as blood is pushed
through an artificially constricted artery covered by the BP cuff
Diastolic pressure: the point at which blood flow is no longer forced through the constricted artery (i.e pressure is no longer great enough to collapse the artery).
![Page 19: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/19.jpg)
Blood PressureStroke Volume (SV) x heart Rate (HR) =
Cardiac Output (CO)
Cardiac Output (CO) x Systemic Vascular Resistance (SVR) = Blood Pressure
So B/P is the result of the “Pump” and the
“Tank”, and partly the “Volume”
![Page 20: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/20.jpg)
Mean Arterial Pressure
The MAP is an mathmatical estimate of the current perfusion pressure. A minimum MAP of 50-60 mmHg is needed for most perfusion
sensitive organs. Two main methods of calculation:Diastolic pressure + one third of the patients pulse
pressure: diastolic blood pressure + (pulse pressure/3).
Diastolic pressure times 2 plus the systolic pressure and then divide that number by 3: (diastolic pressure x 2) + systolic pressure] / 3
“Normal MAP” is between 70—110 mmHg A normal MAP does not preclude a serious condition
![Page 21: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/21.jpg)
Q: What is the 80-70-60 rule for blood pressure?
Q: Is it accurate?
![Page 22: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/22.jpg)
The 80-70-60 rule ?
The minimum blood pressure pre dicted by the guidelines was exceeded in only four of 20 patients.
The mean blood pressure and reference range obtained for each group indicate that the guidelines overestimate the systolic blood pressure associated with the number of pulses present.
![Page 23: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/23.jpg)
Orthostatic Blood Pressure
Procedure:Assess the need for orthostatics.Obtain patient’s pulse and blood pressure while supine.Have the patient sit up for one minute; Obtain the patient's blood pressure and pulse while
sitting. If positive orthostatic changes occur while sitting, DO NOT continue
to the standing position. Have patient stand for one minute.
Free standing, not leaning. Obtain patient’s pulse and blood pressure while
standing. Have someone stand on each side of the patient in case of
orthostatic syncope
![Page 24: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/24.jpg)
Orthostatic Blood Pressure
Orthostatics are considered POSITIVE if: The patient was unable to tolerate standing and became severely
symptomatic. If pulse has increased by 20 BPM OR systolic blood pressure
decreased by 20 mmHg, the orthostatics are considered positive. As an alternative, an increase of 10 BPM AND a decrease of 10
mmHg may also be considered positive if symptomatic. Document the time and vital signs for supine and
standing positions on/with the patient care report (PCR).KEY Points:
You need both PULSE and BLOOD PRESSURE You need the patient to be in each position for a full minute. If they cant complete the test, it is considered positive.
![Page 25: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/25.jpg)
Monitoring Blood Glucose
Source: Accu-Chek® Aviva used with permission of Roche Diagnostics.
Blood glucose monitoring kit
![Page 26: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/26.jpg)
Monitoring Blood Glucose
Perhaps the single most important factor in a diabetics health is how well they monitor their own BG.
Severity of diabetic complications depends on patient’s average blood glucose level.
![Page 27: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/27.jpg)
Glucose Monitors Facts
Personal use: Calibrated Monthly (rare)For EMS use: Calibrated WEEKLYMost have a chip which need to be changed
with EVRY BOTTLE OF NEW STRIPS Affected by tempAffected by time (samples must be applied
w/in 30 –60 seconds)
![Page 28: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/28.jpg)
Blood Glucose Monitoring
Should be done daily or more often.When done by EMS, should be done in the
opposite ext. as D50 was given.Venous blood usually runs about 10 mg/dl
higher than capillary blood.
![Page 29: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/29.jpg)
Checking the Blood Glucose Level (BG)
Glucometers are commonly found on EMS units.
Determines the amount of glucose in the blood, the sample usually coming from a finger stick.
Glucose is measured in milligrams per deciliter (mg/dl).
A normal range is 80-120 mg/dl. Hypoglycemia is a BGL <60 mg/dl. Hyperglycemia is a BGL >150 mg/dl.
Typically not significant until greater than 250. Often not symptomatic until greater than 400 (not always true)
![Page 30: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/30.jpg)
What are the numbers?
Low: <60mg/dl (<3.0 mmol/L)Normal: 60-150mg/dl (3.0-8.0 mmol/L)
Some recommend 120 mg/dlHyperglycemic: >150 (>8.0 mmol/L)DKA usually seen at 250-500mg/dl
![Page 31: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/31.jpg)
mmol/L vs. mg/dl
What is an MMOL?It is a unit of measurement commonly used in
chemistry based on the molecular weight of the substance it pertains to.
To convert mmol/l of glucose to mg/dl, multiply by 18. To convert mg/dl of glucose to mmol/l, divide by 18 or multiply by 0.055.
![Page 32: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/32.jpg)
SPO2
Pulse oximetry is a non-invasive method allowing the monitoring of the saturation of a patient's hemoglobin.
Originally invented in the 1930’s, it expaded into Ors in the 1980;s, and ER,s ICU’s, and EMS Units in the early 1990’s. Early models were unreliable.
Current technology is not only reliable, but leading into other assessment parameters , including SPCO, SPMet, and Fluid hydration status. There are still things that can “Spoof” common SPO2
devices.
![Page 33: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/33.jpg)
SPO2
It shines infrared light across the blood flow in the capillary beds. The sensor then reads the light reflected back. The color of the light is representative of the state of the hemoglobin (hgb) that is bound up with O2.
Each hemoglobin molecule can be bound (carry) by up to four oxygen molecules.
The SPO2 is an expression of the percentage (%) of hemoglobin that is bound.
![Page 34: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/34.jpg)
SPO2
SPO2 is always expressed as a percentage.By contrast, PAO2 (arterial pressure of
oxygen) is NOT expressed as a percentage. Even though the normal PAO2 is 80-100 mmhg it can
be over 400 mmhg.
![Page 35: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/35.jpg)
SPO2
In Order to have a reliable reading, you must have: A good site to sample Adequate perfusion to that site Appropriate sensor A visible waveform to assess quality of reading
![Page 36: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/36.jpg)
SPO2 Spoofs
CO PoisoningCyanide PoisoningHypothermiaLow perfusion statesShockNail PolishHigh Light situationsPoor Placement
![Page 37: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/37.jpg)
KEY POINT
Pulse oximetry measures solely hemoglobin saturation, not ventilation, not perfusion, not inspired FIO2, and not physiologic stress.
SPO2 WORKS BEST when combined with other monitoring tools, such as the EKG and especially ETCO2
![Page 38: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/38.jpg)
Hyperoxemia
New discussions on the use of SPO2 to detect and prevent hyperoxemia often “get it wrong”.
Hyperoxemia is PAO2 in excess of 100 mmHgSPO2 parameters (controversial and
debatable) 94% SPO2 -> give O2 94%-99% SPO2 -> give supplemental O2 only if in
distress. 99-100% SPO2-> consider withholding O2 or titrating
down O2 unless in severe failure
![Page 39: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/39.jpg)
SPCO
CO-Oxometry is the non-invasive measurement of Carbon Monoxide (CO) in the capillary beds.
It is a developmental evolution from SPO2 technology, originally designed for “low flow” states in neonates.
The current technology most commonly in use is the RAD series from masimo.
![Page 40: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/40.jpg)
The Rad Series
![Page 41: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/41.jpg)
Benefits
Uses “Signal Extraction technology” (AKA SET)
This new generation can be potentially used for SPO2, SPCO2, SpHg and SPMET.
SET technology was originally developed to be a “low perfusion state” and “Motion resistant” SPO2, whose technology was expanded to CO-Oxometry
Early device specific research papers focused on neonates, alternative sensing locations, and anesthesia settings.
![Page 42: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/42.jpg)
RAD-57 Operation: Sensor Placement
SENSOR PLACEMENT IS VERY IMPORTANT When possible, use ring
finger, non-dominant hand.
Insert finger until the tip of finger hits the STOP Block.
Sensor should not rotate or shift freely on finger.
LED’s (red light) should pass through mid-nail, not cuticle.
There is a top and bottom, cable should be on top (nail side).
Optimal LED path
![Page 43: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/43.jpg)
Carbon Monoxide: Suggested Triage Algorithm
Measure SpCO
0-3% >3%
No further medical evaluation of SpCO
needed
Loss of consciousness or neurological
impairment or SpCO >25%
Yes No
Transport on 100% oxygen for ED
evaluation. Consider transport to hospital
with hyperbaric chamber
SpCO >12 SpCO <12
Transport on 100% oxygen for ED
evaluationSymptoms of CO
exposure?
Transport on 100% oxygen for ED
evaluation
No further evaluation of SpCO needed.
Determine source of CO if nonsmoker
Yes No
SpCO TRIAGE ALGORITHM
![Page 44: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/44.jpg)
How its being used in EMS nationally
Asymptomatic patients – Screening when there is a CO alarm, Hx of
potential exposure, or for rehab situations Asymptomatic Patients with elevated readings
may be screened and reevaluated after 15 minutes of High Flow O2 and Medical Control Consult
Asymptomatic patients without elevated levels may be released on scene. (Controversial)
Symptomatic patients Transport all symptomatic patients
REGARDLESS OF SPCO Screening for severity and diversion to HBO
(Hyperbaric therapy)
![Page 45: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/45.jpg)
SPCO
When used correctly : It is accurate within +/-2% of carboxy-hemoglobin levels to measure CO (the accepted standard)
When not used correctly: It is useless and provide false reassurance.
![Page 46: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/46.jpg)
Downfalls
Very sensitive, can be “spoofed” by strobes and high ambient light. Cover the probe when “Zero”-ing to the patient
Approx. $4,000 retail
![Page 47: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/47.jpg)
Controversy
There is some debate on the accuracy of the technology. Operator error, high light environments, lack of
understanding of the technology all contribute to a high false positive and false negative rate.
Smokers, urban dwellers, and certain occupations have a naturally elevated baseline. Their “Norm” may be in otherwise “elevated CO” ranges.
![Page 48: 2013 aemt the tools we use](https://reader036.vdocuments.mx/reader036/viewer/2022062523/58f0cee41a28ab6c3b8b45b7/html5/thumbnails/48.jpg)
WHEN WE GET BACK WE WILL BE ON THE ETCO2 PRESENTATION….