analytical comparison of three quantitative immunochemical fecal occult blood tests for colorectal...
TRANSCRIPT
Research Article
Analytical Comparison of Three Quantitative ImmunochemicalFecal Occult Blood Tests for Colorectal Cancer Screening
Lydia Guittet1, Elodie Guillaume1, Romuald Levillain2, Philippe Beley2, Jean Tichet2,Olivier Lantieri2, and Guy Launoy1
AbstractBackground: The superiority of several immunochemical fecal occult blood tests (I-FOBT) over guaiac-
based tests in colorectal cancer screening is now established. The aim of this study was to compare the
analytical performance of 3 quantitative I-FOBTs.
Methods: Stool samples from 10 healthy volunteers, initially I-FOBT negative, supplemented with human
blood, were used to compare reproducibility and stability of measurement at varying storage temperatures
(4�C, 10�C, 20�C, and 30�C) and durations before test analysis (1 to 10 days) for 3 I-FOBTs (New Hemtube/
Magstream HT, OC-Auto sampling bottle3/OC-Sensor DIANA, and FOB Gold/SENTiFOB). Concentrations
ranging from 0 to 350 mg Hb/g of feces were evaluated.
Results: The measurement reproducibility of OC-Sensor was superior to Magstream and far superior to
FOB Gold. For all tests, variability was essentially related to sampling. Detected hemoglobin (Hb) levels were
substantially lower for all tests at temperatures above 20�C. At 20�C, this loss in concentration was less
important withOC-Sensor (significant 1.7% daily decrease vs. 7.4% forMagstream and 7.8% for FOBGold). At
30�C, daily loss was 8.6% with OC-Sensor, whereas after 24 hours, only 30% of the original Hb was detected
with FOB Gold, compared to 70% with Magstream. No Hb was detected on day 5 for the latter 2 tests.
Conclusions: About reproducibility and temperature stability, OC-Sensor performed better than Mag-
stream and far better that FOB Gold.
Impact: Independently of the chosen test, the delay between sampling and test processing should be
reduced, the maximal admissible delay depending on ambient temperature. Cancer Epidemiol Biomarkers Prev;
20(7); 1492–501. �2011 AACR.
Introduction
Colorectal cancer is a major public health issue in allindustrialized countries. Screening using guaiac fecaloccult blood tests (G-FOBT) reduces specific mortalityrelated to colorectal cancer (1). Several studies haveconcluded that both the Magstream (Fujirebio) andOC-Sensor (Eiken Chemical Co.) automated immuno-chemical FOBTs (I-FOBT) offer a gain in sensitivity inthe detection of advanced neoplasias, compared with G-FOBT, at a cost of lower specificity (2–4). For both tests,ideal balance between sensitivity and specificity can bereached by variation in hemoglobin (Hb) concentration
cutoff and number of samples (5–8). For both tests, a gainin both sensitivity and specificity for the detection ofadvanced neoplasias was possible (4, 7, 8).
Since it has been established that I-FOBTs do better thanG-FOBTs, these tests are expected to beused in all nationalscreeningprogramsusingFOBT.Accordingly, the use of afecal immunochemical test has been included in U.S.guidelines for colorectal cancer screening (9). However,several I-FOBTs are available and their performance isdifficult to compare because the cutoff provided in studiesis expressed in concentration of Hb in the collecting tubedependingon the concentration ofHb in the feces, but alsoon the volume of buffer in the tube, and on the amount offeces introduced in the tube. Therefore, optimal test,optimal number of samples, or optimal Hb concentrationcutoff are for the moment indeterminate (10). Moreover,seasonal variations in positivity rates of screening pro-gramsusingOC-Sensor orMagstream I-FOBThave raisedthe question of the sensibility to temperature (11–13), andlaboratory analyses have established a decrease in Hbconcentration in OC-Sensor I-FOBTwith increasing delayin the sample (11, 14).
As a summary, the performance of I-FOBT dependsmainly on the test’s sensitivity to Hb, its reproducibility
Authors' Affiliations: 1INSERM ERI3 "Cancers & Populations," Universityof Caen Basse-Normandie (UCBN), University Hospital (CHU de Caen),Caen, France; 2IRSA (Institut inter-R�egional pour la Sant�e), La Riche,France
Corresponding Author: Lydia Guittet, INSERM ERI3 "Cancers & Popu-lations," UFR M�edecine, CHU de Caen, Avenue de la Côte de Nacre,F-14000 Caen, France. Phone: 33-231-06-51-21; Fax: 33-231-53-08-52;E-mail: [email protected]
doi: 10.1158/1055-9965.EPI-10-0594
�2011 American Association for Cancer Research.
CancerEpidemiology,
Biomarkers& Prevention
Cancer Epidemiol Biomarkers Prev; 20(7) July 20111492
American Association for Cancer Research Copyright © 2011 on August 10, 2011cebp.aacrjournals.orgDownloaded from
Published OnlineFirst May 16, 2011; DOI:10.1158/1055-9965.EPI-10-0594
of sampling and measurement, and the stability of Hb inthe collecting tube, in particular with regard to tempera-ture variations and delay from fecal sampling to testanalysis.Our study aimed to compare measurement precision
and reproducibility, together with Hb measurement sta-bility at varying storage temperatures and varying delaysbetween sampling and analysis of 3 I-FOBTs previouslyused in colorectal cancer screening programs: Magstream(New Hemtube) analyzed using a Magstream HT auto-mated instrument (Fujirebio), OC-Sensor (OC-Auto sam-pling bottle3) analyzed using an OC-Sensor DIANAinstrument (Eiken Chemical Co.; distributed by MastDiagnosis), and FOB Gold (distributed by SKD, France)analyzed using a SENTiFOB instrument (Sentinel diag-nostics).
Materials and Methods
I-FOBTsAll 3 tests use polyclonal rabbit antibodies directed
against human HbA. All tests are fully automated(Table 1). Analysis of OC-Sensor and FOB Gold is basedon immunoturbidimetry, which involves a measurementof the absorbance of light through the tube, whichincreases with the importance of Hb antibodies com-plexes. Analysis of Magstream involves the use of anautomated visual measurement of migration of aggluti-nated magnetic particles. In routine use, the crude pixelvalue generated by Magstream is converted into MSRunits, an arbitrary unit proposed by the manufacturer.For the present study, we asked Fujirebio to providesoftware (not routinely integrated in the machine) toallow for the collection of the measurements (pixelvalues). AlthoughMagstream is commercialized by Fujir-ebio as a qualitative test, a quantitative measure is pro-vided by the instrument, so we considered the test asquantitative in the analysis.
Fecal sampling methodFirst, freshly collected stools, obtained from 10
healthy subjects aged less than 50 years, were testedusing all 3 tests to confirm the initial absence of Hb.These stools were mixed and homogenized, thendivided into containers. In each of these containers, avolume of human whole blood lysate, the Hb content ofwhich had previously been measured (Advia 2120, Sie-mens), was added to obtain all prespecified Hb con-centrations in the stool. Each container was vigorouslyshaken after adjunction of blood. For each of the con-centrations, sampling of all 3 tests was done using thesame containers, ensuring that concentration was iden-tical for all tests. Finally, collecting tubes were shakenafter sampling and before analysis. This procedure wasrepeated 2 times leading to 2 distinct stool mixtures, onebeing analyzed to evaluate reproducibility and the otherto evaluate stability to storage.
Experimental planTwo distinct experiments were conducted: the first
one to evaluate and compare the reproducibility of thetests (experiment 1) and the other one to evaluate andcompare their sensitivity to temperature and delay ofstorage (experiment 2). For each of these 2 experiments,a distinct stool mixture was performed, as described asfollows.
Experiment 1(stool mixture n�1). To compare testsfor a given Hb concentration, we initially explored, foreach test, the relationship between concentration in thefeces and the value provided by the instrument for atotal of 10 values of Hb concentration in feces, varyingfrom 0 to 350 mg Hb/g of feces. This range of concen-trations was selected to: (i) cover the range of usual orproposed Hb positivity cutoffs of all tests and (ii) coverthe range of physiological bleeding of colorectal lesions(15). To assess measurement reproducibility, 10 tubeswere collected for each instrument and each concentra-tion, and all tubes were repeatedly analyzed 5 timesleading to a total of 50 readings per concentration andtest. In this experiment, all prepared I-FOBTs werestored for 3 days at 10�C before being analyzed, approx-imating to the conditions of a screening program invol-ving mailed samples.
Experiment 2 (stool mixture n�2). A second experi-mental plan was developed to assess the influence ofstorage temperature and delay between sampling andanalysis for different Hb concentrations in feces. For eachI-FOBT, 4 storage temperatures: 4�C, 10�C, 20�C, and30�C, and 5 delays between sampling and analysis: 1,3, 5, 7, and 10 days were tested for 6 Hb concentrations infeces (0, 20, 75, 100, 150, and 250) mgHb/g of feces. Due tothe large number of combinations of the 3 aforemen-tioned parameters in an exhaustive plan (360 combina-tions) and to the relatively slow operating rate of 1 of themachines (SENTiFOB, Sentinel Diagnostics), we used anoptimal experience plan, determined using the ADX(Analysis and Design of eXperiments) tool developedby SAS software. Four factors were introduced in thisexperimental plan (I-FOBT, storage temperature, delay,and concentration), as well as their 2-level interactions.Optimization of the design was achieved by D-optimaloptimization (maximization of the determinant of theinformation matrix), to do a linear regression evaluatingthe mean daily decrease in standardized concentrations(to avoid the impact of the slope of the relationshipbetween fecal concentration and buffer concentrationof blood) in the collecting tubes. However, due to thenonlinear relationship between concentration of blood inthe feces and the collecting tube (roughly logarithmic)and to the semiquantitative nature of the Magstream test(see Results), each test was assessed individually and nooverall analysis was conducted (see Statistical Methodsnext).
In our experimental plan, 32 combinations of storagetemperature, delay between sampling and analysis, andfecal Hb concentration were used for each test. The
Analytical Comparison of Immunochemical FOBTs
www.aacrjournals.org Cancer Epidemiol Biomarkers Prev; 20(7) July 2011 1493
American Association for Cancer Research Copyright © 2011 on August 10, 2011cebp.aacrjournals.orgDownloaded from
Published OnlineFirst May 16, 2011; DOI:10.1158/1055-9965.EPI-10-0594
Tab
le1.
Tech
nica
lcom
parison
oftheI-FO
BTs
OC-S
enso
rFO
BGold
Mag
stream
Buffervo
lumea
2mL
1.7mL
1mL
Feca
lsam
plingvo
lumea
10mg
10mg
0.3mg
Antibod
iesus
edRab
bitan
ti-hu
man
HbA
(polyc
lona
l)Rab
bitan
ti-hu
man
HbA
(polyc
lona
l)Rab
bitan
ti-hu
man
HbA(polyc
lona
l)
Prove
nan
alytical
rang
e50
to1,00
0ng
/mL(equiva
lent
to10
to20
0mg
/g)
50to
1,00
0ng
/mL
from
20ng
/mL(M
SR
¼1;
commercialized
asaqua
litativetest)
Rea
dingtech
nique
Anti-hu
man
Hban
tibod
iesaread
sorbed
onlatexpartic
les.
Inthepres
ence
ofblood
instoo
ls,they
provo
kethe
antig
en–an
tibod
yreac
tionan
dag
glutination,
cons
eque
ntly
referred
toas
thelatexag
glutinationtest.Cha
nges
insa
mpleturbidity
bylatexag
glutinationare
mea
suredop
tically.
Mag
netic
gelatin
partic
lesareattach
edto
anti-hu
man
Hb
antib
odies.
Collectingtubes
aretilted60
�from
theho
rizon
tal
pos
ition
,en
ablingfree
mag
netic
partic
lesto
slidedow
nthe
slop
eof
thewell,thus
form
ingamea
surable
line.
Thehigh
ertheprese
nceof
human
Hb,thesh
orterthelineis.Adigita
lpicture
ofthelineis
take
nbythemac
hine
.
Pixels:
328
Pixels:
197
Pixels:
132
Neg
ative
Pos
itive
Stron
gpos
itive
Abso
rban
ce57
0nm
Abso
rban
ce66
0nm
Collectingtube
FOB
Gold
OC-A
utosa
mplingbottle
3New
Hem
Tube
Autom
ated
analyzer
SENTiFO
BDiana
Mag
Strea
mHT
Rates
ofread
ing
75tubes
/hou
r28
0tubes
/hou
r90
0tubes
/hou
rUsu
althresh
old
175ng
Hb/m
linthebuffer
100ng
Hb/m
linthebuffer
211pixe
ls(M
SR
¼1.0)
Adjustab
lethresh
olddue
toqu
antitativena
ture
ofthetest.
Theo
retic
ally
nona
djustab
lethresh
oldsinc
ethetest
isco
mmercialized
asaqua
litativetest
bytheman
ufac
turers.
aTa
kenfrom
doc
umen
tsprovided
bytheman
ufac
turer.
Guittet et al.
Cancer Epidemiol Biomarkers Prev; 20(7) July 2011 Cancer Epidemiology, Biomarkers & Prevention1494
American Association for Cancer Research Copyright © 2011 on August 10, 2011cebp.aacrjournals.orgDownloaded from
Published OnlineFirst May 16, 2011; DOI:10.1158/1055-9965.EPI-10-0594
experimental plan can be deducted from Table 2 whichpresents mean measurement obtained for each test,according to temperature, concentration, or delay, andtherefore in which nonempty cells correspond to evalu-ated combinations of temperature*concentration*time*t-est. Ten tubes were collected for each of the includedsituations, each tube being analyzed 5 times.Temperature of storage was monitored using a system
of electronic measurement of temperature every 5 min-utes, and radio frequency recording (AOIP instrument).All tubes, whatever test or concentration, were stored inthe same conditions for each temperature (dedicatedfridge for 4�C and 10�C temperatures with a range ofvariations of temperature of �2�C, air-conditioned roomfor 20�C temperaturewith a range of variations from 20�Cto 22�C, and a dedicated incubator for 30�C temperaturewith a range of variations of �1�C).
Statistical analysisReproducibility of I-FOBTs (experiment 1). For each
test and each of the experimental concentrations, theintertube and intratube variances were determined byusing a random effect model (SAS PROC MIXED). Thisprocedure enabled us to compute the variation coeffi-cients due to sampling (intertube) and reading (intra-tube).Stability to temperature and duration of storage
(experiment 2). In an exploratory analysis, we com-puted the mean measurement obtained for each test,according to temperature, concentration, or delay(Table 2). Then, for each temperature and processingdelay, the mean measurement was standardized on theinitial fecal concentration to allow comparison of var-iations in concentration due to storage duration andtemperature between tests. For FOB Gold and OC-Sensor, the buffer concentration was directly propor-tional to fecal concentration (Fig. 1). The theoreticalrelationship between fecal Hb concentration (Cf),expressed in mg/g, and concentration of Hb in thebuffer (Cb), expressed in ng/mL, is given by the follow-ing formula: Cb ¼ ðCf * qsÞ=vb, where qs is the quantityof stools introduced in the tube (in mg) and vb is thevolume of buffer (in mL). However, as we did notcheck independently the values of qs and vb providedby the manufacturers, the buffer concentration pro-vided by the automated analyzer was simply dividedby the initial theoretical fecal concentration, to stan-dardize data. For Magstream, the pixel value providedby the automated analyzer had an inverse logarithmrelationship to fecal Hb concentration, and the meanpixel value was 330 in the absence of Hb in the sample(Fig. 1). The following transformation was thereforeapplied to data: (330 pixel)/log(initial fecal concentra-tion). However, this transformation was less adapted tosmaller concentrations. For each test, we evaluated themean daily decrease in standardized Hb concentrationaccording to temperature using mixed effect linearmodels.
Statistical analysis was done using SAS software, ver-sion 9.1 (SAS Institute).
Results
Experiment 1Relationship between measurements and fecal concen-
tration of Hb. First, the relationship of the values pro-vided by each automated analyzer (3 days’ storage at10�C) was assessed according to the initial fecal Hbconcentration. A linear relationship between measure-ment and fecal Hb content was observed for OC-Sensorand FOB Gold, but not for Magstream (Fig. 1).
For Magstream, the pixel values were bounded inthe range 130 to 330 pixels. Above a concentration inthe tested stools of 250 mg Hb/g, the pixel measure-ments in the tube did not change with concentration(the mean pixel measurement was 130 pixels). There-fore, the test could only be considered as quantitativewithin a fecal Hb concentration range of 20 to 200 mgHb/g of feces.
Instrument measurements have been plotted in Figure1 and the expected value is represented as a dotted lineaccording to linear modeling for FOB Gold and OC-Sensor, and loess (local polynomial fitting) modelingfor Magstream. We defined overlap between concentra-tions whether no cutoff could perfectly separate instru-ment measurements between 2 successive fecalconcentrations. For Hb concentrations up to 150 mgHb/g of feces, there was no overlap in the range ofmeasurements obtained with the different Hb concentra-tions in feces selected in our protocol for OC-Sensor. Forboth Magstream and FOB Gold, overlap was observedbetween concentrations in the entire range of these testedconcentrations.
ReproducibilityReproducibility could be explored in the entire concen-
tration range for OC-Sensor and FOB Gold (Fig. 2). How-ever, it was only explored for concentrations below 250 mgHb/g of feces for Magtream, because for higher concen-trations, the absence of measurement variation betweenconcentrations using this test would lead to artificiallygood reproducibility.
The best reproducibility (smaller total variation coeffi-cient) was observed with OC-Sensor, with the exceptionof concentrations below 75 mg Hb/g of feces, for whichMagstream offered better reproducibility. However,variability for small concentrations using Magstreamwas reduced due to lack of quantitative measurementsabove 330 pixels. The worst reproducibility was observedwith the FOB Gold test. Indeed, the mean total variationcoefficient observed between 75 and 250 mg Hb/g of feceswas 0.07 for OC-Sensor, 0.10 for Magstream, and 0.18 forFOB Gold.
For all tests, measurement variability involved inter-tube variability rather than intratube variability (thevariation coefficients associated with sampling were far
Analytical Comparison of Immunochemical FOBTs
www.aacrjournals.org Cancer Epidemiol Biomarkers Prev; 20(7) July 2011 1495
American Association for Cancer Research Copyright © 2011 on August 10, 2011cebp.aacrjournals.orgDownloaded from
Published OnlineFirst May 16, 2011; DOI:10.1158/1055-9965.EPI-10-0594
Tab
le2.
Mea
nmea
suremen
tac
cordingto
test*con
centratio
n*temperature*delay
Tes
tOC-S
enso
rFO
BGold
Mag
stream
Con
centratio
nof
hemog
lobin
inthefece
s,mg
/g0
2075
100
150
250
020
7510
015
025
00
2075
100
150
250
Exp
erim
ent
Temperature
Day
s1
10� C
30
8533
449
867
01,08
66
124
552
664
877
1,86
332
129
921
018
614
713
22
4�C
110
84
151
147
30
789
6328
113
213
35
100
464
641
2,22
132
47
1207
1255
259
148
134
1058
762
532
010
� C1
01,28
767
61,09
924
03
579
163
2,12
131
514
55
1,21
94
129
70
101
796
130
1041
970
54
205
152
130
20� C
154
765
716
462
031
513
33
9539
360
092
012
45
984
8868
819
918
67
436
697
331
310
032
659
531
422
330
� C1
340
565
420
23
733
531
228
35
220
438
213
317
311
796
289
612
317
315
108
317
810
315
Guittet et al.
Cancer Epidemiol Biomarkers Prev; 20(7) July 2011 Cancer Epidemiology, Biomarkers & Prevention1496
American Association for Cancer Research Copyright © 2011 on August 10, 2011cebp.aacrjournals.orgDownloaded from
Published OnlineFirst May 16, 2011; DOI:10.1158/1055-9965.EPI-10-0594
greater than those associated with reading). Intratubevariability was low for all tests (the mean coefficient ofvariation due to reading between 75 and 250 mg Hb /g offeces was 0.025, 0.016, and 0.060 for FOB Gold, OC-Sensor, and Magstream, respectively).The difference between these sources of variation was
the highest for FOB Gold, for which the intertube varia-bility was particularly important compared with theother tests. The mean coefficient of variation due tosampling between 75 and 250 mg Hb/g of feces was0.15 for FOB Gold, whereas it was only 0.10 for OC-Sensor and 0.07 for Magstream.Intertube variability associated with FOB gold tended
to decrease as the Hb concentration in feces increased.
Intertube variability associated with OC-Sensor wasstable between 75 and 250 mg Hb/g of feces, but washigh for concentrations below 75, or above 250 mg Hb/gof feces. The latter concentrations were higher than theupper limits of good performance recommended by themanufacturer. Intertube variability using Magstreamincreased as did fecal Hb concentration up to 150 mgHb/g of feces, then decreased due to a nonlinear relation-ship between pixels and concentration.
Experiment 2Stability. Mean measurements according to test,
temperature, concentration, and time are provided inTable 2. For all 3 tests, measurement was stable over
OC-Sensor
Hb
conc
entr
atio
n in
the
buffe
r, ng
/mL
0
0 50 100 150 200 250 300 350
500
1,00
01,
500
FOB Gold
Hb
conc
entr
atio
n in
the
buffe
r, ng
/mL
0 50 100 150
Initial fecal Hb concentration, µg/g
200 250 300 3500
500
1,00
01,
500
2,00
02,
500
3,00
0 Magstream
0 50 100 150 200 250 300 350
150
200
Pix
el v
alue 25
030
0
Figure 1. Value returned by the automated analyzers according to test and fecal Hb concentration. Manufacturer's thresholds are presented as a continuousline (175 ng/mL for FOB Gold; 100 ng/mL for OC-Sensor; and 211 pixels for Magstream). Data modeling is presented as a dotted line (linear modeling:FOB Gold, OC-Sensor; loess modeling: Magstream).
OC-Sensor
0.2
0.15
0.1
Var
iatio
n co
effic
ient
0.05
0
50 100 150 200 250* 300* 350*
0.2
0.15
0.1
Var
iatio
n co
effic
ient
0.05
0
50 100 150* 200*
Initial fecal Hb concentration, µg/g
250* 300* 350*
FOB Gold
0.2
0.15
0.1
Var
iatio
n co
effic
ient
0.05
0
50 100 150 200 250** 300** 350**
Magstream
Figure 2. Variation coefficients due to sampling or reading according to fecal Hb concentration. X, variation coefficients due to sampling. ^, variationcoefficients due to reading. *, variation coefficients total. *, quantification of Hb implied dilution during reading. **, not calculable.
Analytical Comparison of Immunochemical FOBTs
www.aacrjournals.org Cancer Epidemiol Biomarkers Prev; 20(7) July 2011 1497
American Association for Cancer Research Copyright © 2011 on August 10, 2011cebp.aacrjournals.orgDownloaded from
Published OnlineFirst May 16, 2011; DOI:10.1158/1055-9965.EPI-10-0594
time, independently of storage temperature when no Hbwas added to feces. Relationship between instrumentmeasurements (pixel or buffer concentration) and fecalHb concentration was verified (data not shown). Figures3 and 4 plot meanmeasurement permachine according tostorage duration or temperature. A decrease in measure-ment was observed for all tests at increased storagedurations, and at temperatures of 20�C and 30�C, butnot 4�C or 10�C.
Figure 5 provides the mean of ratio of measurementsstandardized on the experimental fecal concentration foreach combination of storage duration and temperatureand for each concentration (see Materials and Methods).The stability of the Hb measurement provided by all 3tests was good at 4�C and 10�C for OC-Sensor and Mag-stream. However, an increase in concentration providedby FOB Gold was observed within the first days, parti-cularly at a storage temperature of 4�C. A similar (yet
smaller) increase in concentration detected by OC-SensorandMagstream over the first days cannot be excluded onthe basis of our data.
At a storage temperature of 20�C, a substantialdecrease in Hb concentration was observed over timeusing all 3 tests. The best stability was observed usingOC-Sensor with a significant daily decrease in measure-ment of 1.7% (P < 0.01) at 20�C, compared with a dailydecrease of 7.4% and 7.8% with Magstream and FOBGold, respectively (P < 10�3). However, the decreaseobserved with Magstream depended on the initial Hbconcentration. It was greater with the smaller concentra-tion of 20 mg Hb/g of feces.
At 30�C, the performance of OC-Sensor was far betterthan that of FOB Gold and Magstream. Indeed, the meandaily decrease in measurement observed with OC-Sensorwas 8.6% (P < 10�3), whereas the mean decrease inmeasurement observed on the first day was 30% for
OC-Sensor
0 5 10 15 20 25 30
Mea
n of
buf
fer
conc
entr
atio
, ng/
mL
020
040
060
080
01,
000
0 5 10 15 20 25 30
Temperature of storage, °C
Mea
n of
buf
fer
conc
entr
atio
n, n
g/m
L
020
040
060
080
01,
000
1,20
0
FOB Gold
0 5 10 15 20 25 30
Mea
n of
pix
el v
alue
150
200
250
300
350
Magstream
Figure 4. Comparison of mean measurements and SD according to test: temperature of storage.
OC-Sensor
0 2 4 6 8 10
020
0
Mea
n of
buf
fer
conc
entr
atio
n, n
g/m
L
400
600
800
1,00
0
0
Duration of storage, d
2 4 6 8 10M
ean
of b
uffe
r co
ncen
trat
ion,
ng/
mL
020
040
060
080
01,
000 FOB Gold
0 2 4 6 8 10
Mea
n of
pix
el v
alue
150
200
250
300
350
Magstream
Figure 3. Comparison of mean measurements and SD according to test: duration of storage.
Guittet et al.
Cancer Epidemiol Biomarkers Prev; 20(7) July 2011 Cancer Epidemiology, Biomarkers & Prevention1498
American Association for Cancer Research Copyright © 2011 on August 10, 2011cebp.aacrjournals.orgDownloaded from
Published OnlineFirst May 16, 2011; DOI:10.1158/1055-9965.EPI-10-0594
Magstream and 70% for FOB Gold. The decreaseobserved with OC-Sensor at 30�C depended on the initialconcentration, being more important for small concentra-tions. On day 7, 38% of an initial concentration of 150 mgHb/g of stool was still detected with OC-Sensor. On day10, 25% of an initial concentration of 250 mg Hb/g of stoolwas still detected by the same test. On the contrary, fromday 5, no Hb content was detected by FOB Gold orMagstream in the samples.
Discussion
Our results show that the precision and the reprodu-cibility of Hb measurements in feces are better with OC-Sensor than with Magstream, and better with Magstream(in the range of concentrations where the test can beconsidered as quantitative) than with FOB Gold. Aboutstability at varying temperatures, our results show thatindependently of the test used, there is a substantial lossin Hb measurement as from 20�C. This loss is moreimportant for FOB Gold than for the other tests. At20�C and 30�C, denaturation of Hb is less importantand occurs less quickly in the buffers used with OC-Sensor than with Magstream devices.Independently of the test used, intratube variability
was lower than intertube variability. Similar findingswere found in an experimental study conducted in U.K., although intratube variabilitywas assessed in solutionof Hb, and intertube in artificially positive stools (16). Theintertube variability estimated in our experiment was aconsequence of both the tube characteristics and thebiologist’s reproducibility in using the test (samplingof feces using the immunochemical tests probes). In realscreening settings, such a tube effect would probably begreater because a patient is certainly less reproducible in
his sampling technique than a biologist, and patients donot mix their stool to homogenize their blood contentbefore performing the test. Moreover, the tube effectmeasured in screening programs including several tubesper patient also includes the effect of the intermittency ofthe bleeding associated with colonic lesions. To the con-trary, the intratube variability measured in our experi-ment should be similar to that which occurs in realscreening settings, as it is solely an effect of the automatedanalyzer itself.
The lower intertube variability when using OC-Sensorcould be explained by a more accurately calibrated quan-tity of stool incorporated in the sample (notmeasurable inour study). In addition, more overlap occurred betweenthe evaluated concentrations with FOB Gold and Mag-stream than with OC-Sensor. Accordingly, the loss inprecision of these tests (including intertube and intratubevariability) could have consequences on positivity ratesin real screening settings.
Temperature-related Hb degradation is expected;however, it can be delayed by the use of a suitablestabilizing agent in the buffer. Such stabilizing agentsare included in all 3 tests. Nevertheless, the stability ofHbmeasurement at varying temperatures and over timewas better with OC-Sensor than with Magstream andfar better than FOB Gold. The superiority of OC-Sensorwas also observed in the NHS evaluation report (16). Ata temperature of 20�C, the decrease observed with FOBGold and Magstream was at least twice that observedwith OC-Sensor. At 30�C, the performance of Mag-stream and FOB Gold was very poor, whereas theOC-Sensor was much better and remained reliable, evenat this high temperature. I-FOBT sensitivity variationsrelated to storage duration have previously beendescribed in both laboratory experiments and genuine
Magstream
Mea
n of
rat
io (
(330
-Pix
el)/
log(
feca
l con
cent
ratio
n))
2 4 6 8 10
510
15
FOB Gold
2 4
Duration of storage, d
6 8 100
24
68
10
Mea
n of
rat
io (
buffe
r co
ncen
trat
ion/
feca
l con
cent
ratio
n)
1
2 4
OC-Sensor
6 8 10
Mea
n of
rat
io (
buffe
r co
ncen
trat
ion/
feca
l con
cent
ratio
n)
23
45
6
Figure 5. Comparison of the evolution of the measured concentration in the buffer according to storage temperature and duration. *, mean ratiotemperature ¼ 4�C; modelization (—) linear (OC-Sensor); polynomial (FOB Gold, Magstream). D, mean ratio temperature ¼ 10�C; modelization (- - -)linear (OC-Sensor); polynomial (FOB Gold, Magstream). þ, mean ratio temperature ¼ 20�C; modelization (. . ..) linear (OC-Sensor); polynomial (FOB Gold,Magstream). �, mean ratio temperature ¼ 30�C; modelization (_ . _) linear (OC-Sensor); polynomial (FOB Gold, Magstream).
Analytical Comparison of Immunochemical FOBTs
www.aacrjournals.org Cancer Epidemiol Biomarkers Prev; 20(7) July 2011 1499
American Association for Cancer Research Copyright © 2011 on August 10, 2011cebp.aacrjournals.orgDownloaded from
Published OnlineFirst May 16, 2011; DOI:10.1158/1055-9965.EPI-10-0594
screening settings (11–14). The interaction between sto-rage duration and temperature has been quantified forthe OC-Sensor test in a laboratory experiment: The dailydecrease in fecal Hbmeasurement was 0.3% at 4�C, 2.2%at 20�C, and 3.7% at 28�C (14). Our findings (2% dailydecrease at 20�C and 9% daily decrease at 30�C) areconsistent with this observation, although the decreaseobserved at 30�C seemed more important. This could beexplained by a difference in initial Hb concentration infeces, the relative decrease being higher for small initialconcentrations. Such differences in sensitivity related tostorage temperature and duration, and affecting thereliability of colorectal screening test programs, couldhave an important impact on screening organizationand test choice, particularly in countries where ambienttemperatures are high. It would seem that, wheneverpossible, the delay between sampling and test proces-sing should be reduced to 3 days, or CRC screeningprograms should be stopped during the summer incountries with long period of very high temperatures(>30�C). In addition, patients should be advised to storefecal samples in the refrigerator at home before for-warding them by post. Acceptability of such a recom-mendation needs further investigations.
Our study has several limitations associated withexperimental conditions and data analysis constraints.Since it was based on laboratory preparation of positivefecal samples, it is possible that these artificial positivesamples behave differently than native positive samples,despite the fact that we used human feces. However,decrease in Hb from real positive samples has also beenshown with OC-Sensor (11). On the one hand, mixed ornonmixed stool samples could behave differently in theapplication of I-FOBT. On the other hand, the mixture ofstool samples from several subjects done in our protocolallowed us to avoid differential bias between concentra-tions and tests in case ofundetectedHbpreviouslypresentin 1 or several fecal samples. In addition, sampling errorwas probably underestimated since the tests were con-ducted by trained biologists, rather than by subjects fromthe general population. For example, no miss manipula-tion of tubes occurred, such as opening the wrong side ofthe FOB Gold test or spilling the OC-Sensor or FOB Goldbuffer. Targeted concentrations were not checked in thesamples by independent Hb measurement methods.Nevertheless, within each of the experiments (for exam-ple, storage evaluation), calibrated stool samples wereobtained from the same negative FOB-tested stools.Finally, the nonlinear relationship between pixel valueand concentration of Hb in the buffer for the Magstreamtest implied transformation of crude data. However, weselected the transformation which minimized underlyingassumptions and independently of the results.
However, the fact that this study was entirely labora-tory-performed enabled every parameter to be controlled,as from the introduction of blood into the stool. This is notthe case for studies exploring the stability of fecal samplessent to the laboratory after having been performed by the
patient at home, measurement of initial Hb concentrationin the stool, and storage duration or temperature fromhome to laboratory being unknown (11, 14). This alsoenabled the 3 tests to be compared on the same stools.Moreover, the use of blood lysate guaranteed the bestpossible conditions for the antibody–antigen reaction.Finally, our laboratory study was designed to fit as mostas possible real screening settings with the choice of arange of concentrations adapted to physiological bleedingof lesions and discussed cutoffs, temperatures close tousual weather conditions, and delays compatibles withmailing of samples bypost to a central analysis center, as itis the case in FOBT-based screening programs.
Our results show better analytical and stability perfor-mances of OC-Sensor or Magstream than FOB Gold.Comparison between OC-Sensor and Magstream wasmade more complex since: (i) the area of optimal perfor-mance was different for both tests, OC-Sensor detectingsmaller Hb concentrations, and (ii) the relationshipbetween measurement of OC-Sensor and Magstreamwas not linear. The improved stability offered by OC-Sensor compared with Magstream and the semiquanti-tative nature of the Magstream offer strong arguments infavor of OC-Sensor. Nevertheless, several studies haveshown the good performance of Magstream in popula-tion surveys, even using only 1 sample at the man-ufacturer’s threshold (17, 18). Costs of the tests shouldalso be considered, together with adaptation of screeningprograms to limit the effect of sensitivity of tests totemperature on their performances (12). Further studiesand cost-effectiveness analysis are needed to compareMagstream and OC-Sensor in real screening settings,appropriately taking into account temperature and dura-tion of storage.
Disclosure of Potential Conflicts of Interest
No potential conflicts of interest were disclosed.
Contributions
L. Guittet, E. Guillaume, R. Levillain, P. Beley, and G. Launoy elabo-rated the experimental plan. P. Beley and R. Levillain performed thelaboratory analyses. L. Guittet and E. Guillaume performed the statisticalanalysis. Results were analyzed and discussed by all authors. Finally, L.Guittet, E. Guillaume, and G. Launoy wrote the article, which was care-fully corrected by all authors.
Grant Support
The study was funded by the French National Cancer Institute (InCA).Instruments were provided on loan by manufacturing companies. Supportingcompanies had no access to crude data. Preliminary analysis was provided tothem for the purposes of information; however, in no manner whatsoever did itinfluence our analysis or the drafting of this article.
The costs of publication of this article were defrayed in part by thepayment of page charges. This article must therefore be hereby markedadvertisement in accordance with 18 U.S.C. Section 1734 solely to indicatethis fact.
Received June 7, 2010; revised April 18, 2011; accepted May 3, 2011;published OnlineFirst May 16, 2011.
Guittet et al.
Cancer Epidemiol Biomarkers Prev; 20(7) July 2011 Cancer Epidemiology, Biomarkers & Prevention1500
American Association for Cancer Research Copyright © 2011 on August 10, 2011cebp.aacrjournals.orgDownloaded from
Published OnlineFirst May 16, 2011; DOI:10.1158/1055-9965.EPI-10-0594
References1. Hewitson P, Glasziou P, Watson E, Towler B, Irwig L. Cochrane syste-
matic review of colorectal cancer screening using the fecal occult bloodtest (hemoccult): an update. Am J Gastroenterol 2008;103:1541–9.
2. Guittet L, Bouvier V, Mariotte N, Vallee JP, Ars�ene D, Boutreux S, et al.Comparison of guaiac based and an immunochemical faecal occultblood test in screening for colorectal cancer in a general average riskpopulation. Gut 2007;56:210–4.
3. Van Rossum LG, Van Rijn AF, Laheij RJ, van OijenMG, Fockens P, vanKrieken HH, et al. Random comparison of guaiac and immunochem-ical fecal occult blood tests for colorectal cancer in a screeningpopulation. Gastroenterology 2008;135:82–90.
4. Hol L, Wilschut JA, van Ballegooijen M, van Vuuren AJ, van der Valk H,Reijerink JC, et al. Screening for colorectal cancer: random compar-ison of guaiac and immunochemical faecal occult blood testing atdifferent cut-off levels. Br J Cancer 2009;100:1103–10.
5. Levi Z, Rozen P, Hazazi R, Vilkin A, Waked A, Maoz E, et al. Aquantitative immunochemical fecal occult blood test for colorectalneoplasia. Ann Intern Med 2007;146:244–55.
6. Grazzini G, Visioli CB, Zorzi M, Ciatto S, Banovich F, Bonanomi AG,et al. Immunochemical faecal occult blood test: number of samplesand positivity cutoff. What is the best strategy for colorectal cancerscreening? Br J Cancer 2009;100:259–65.
7. Guittet L, Bouvier V, Mariotte N, Vallee JP, Levillain R, Tichet J, et al.Performances of immunochemical faecal occult blood test in colorectalcancer screening in average-risk population according to positivitythreshold and number of samples. Int J Cancer 2009;125:1127–33.
8. van Rossum LG, van Rijn AF, Laheij RJ, van Oijen MG, Fockens P,Jansen JB, et al. Cutoff value determines the performance of a semi-quantitative immunochemical faecal occult blood test in a colorectalcancer screening programme. Br J Cancer 2009;101:1274–81.
9. U.S. Preventive Services Task Force. Screening for colorectal cancer:U.S. preventive services task force recommendation statement. AnnInt Med 2008;149:627–37.
10. Lieberman DA. Clinical practice. Screening for colorectal cancer. NEngl J Med 2009;361:1179–87.
11. van Rossum LG, van Rijn AF, van Oijen MG, Fockens P, Laheij RJ,Verbeek AL, et al. False negative fecal occult blood tests due todelayed sample return in colorectal cancer screening. Int J Cancer2009;125:746–50.
12. Grazzini G, Ventura L, Zappa M, Ciatto S, Confortini M, Rapi S, et al.Influence of seasonal variations in ambient temperatures on perfor-mance of immunochemical faecal occult blood test for colorectalcancer screening: observational study from Florence district. Gut2010;59:1511–5.
13. Australian Government, Department of Health and Aging. Nationalbowel cancer screening program: screening retest fact sheet.[accessed 2010 Sept 13]. Available from: www.cancerscreening.gov.au/internet/screening/publishing.nsf/content/bw-retest
14. Rozen P, Waked A, Vilkin A, Levi Z, Niv Y. Evaluation of a desk topinstrument for the automated development and immunochemicalquantification of fecal occult blood. Med Sci Monit 2006;12:MT27–32.
15. Ciatto S, Martinelli F, Castiglione G, Mantellini P, Rubeca T, GrazziniG, et al. Association of FOBT-assessed faecal Hb content with coloniclesions detected in the florence screening programme. Br J Cancer2007;96:218–21
16. NHS Evaluation report. Immunochemical faecal occult blood tests;2009 Nov. Report No.: CEP09042.
17. Morikawa T, Kato J, Yamaji Y, Wada R, Mitsushima T, Shiratori Y. Acomparison of the immunochemical fecal occult blood test and totalcolonoscopy in the asymptomatic population. Gastroenterology2005;129:422–8.
18. Guittet L, Bouvier V, Mariotte N, Vallee JP, Levillain R, Tichet J, et al.Comparison of a guaiac and an immunochemical fecal occult bloodtest for the detection of colonic lesions according to lesion type andlocation. Br J Cancer 2009;100:1230–5.
Analytical Comparison of Immunochemical FOBTs
www.aacrjournals.org Cancer Epidemiol Biomarkers Prev; 20(7) July 2011 1501
American Association for Cancer Research Copyright © 2011 on August 10, 2011cebp.aacrjournals.orgDownloaded from
Published OnlineFirst May 16, 2011; DOI:10.1158/1055-9965.EPI-10-0594