2013 pelletier prp constituents
TRANSCRIPT
74 Clinical Sciences
Pelletier MH et al. Platelet Function and Constituents … Int J Sports Med 2013; 34: 74–80
accepted after revision
May 10 , 2012
Bibliography
DOI http://dx.doi.org/
10.1055/s-0032-1316319
Published online:
August 14, 2012
Int J Sports Med 2013; 34:
74–80 © Georg Thieme
Verlag KG Stuttgart · New York
ISSN 0172-4622
Correspondence
Dr. Matthew H Pelletier, PhD
Surgical & Orthopaedic
Research Labs
University of New South Wales
Level 1
Clinical Sciences Bld
2031 Randwick
Australia
Tel.: +61/293/822 687
Fax: +61/293/822 660
Key words
● ▶ platelet rich plasma
● ▶ blood
● ▶ autologous
● ▶ therapy
● ▶ p-selectin
● ▶ viable
Platelet Function and Constituents of Platelet Rich Plasma
dermal growth factor (EGF). These secreted pro-
teins have important roles in healing and tissue
regeneration [ 1 , 5 , 27 , 40 ] . Although there are
many proteins that may promote wound healing
in PRP, such as fi brinogen, the basic premise for
PRP remains the delivery of a soup of platelet-
derived proteins at concentrations above base-
line [ 13 – 15 , 50 ] .
PRP can be generated by centrifugation of antico-
agulated whole blood and collected with pipette
or syringe, but interest in PRP therapies has led to
the development of several commercial systems
aimed at improving ergonomics, decreasing vari-
ability, and improving platelet recovery in a
closed system. While these devices do produce
PRP with a concentration of platelets above cir-
culating baseline blood levels [ 29 ] , the product
varies between devices, leading Ehrenfest et al.
[ 11 ] to propose a classifi cation to highlight their
properties. Specifi cally, the classifi cation takes
account of diff erences in production including
whether or not the buff y coat (which contains a
near 7-fold increase in leukocytes [ 7 ] ) is col-
lected, fi brin and fi brinogen characteristics,
collection effi ciency, as well as practical consid-
Introduction
▼ Clinical use of Platelet Rich Plasma (PRP) has
been reported since the early 1990s for oral and
maxillofacial surgery as a source of growth fac-
tors to increase the rate and degree of bone for-
mation at early time points [ 31 ] . Since then PRP
has found an increasing use in a variety of clinical
applications. Primarily PRP has been used to sup-
port hard and soft tissue healing and formation
[ 5 , 9 , 35 , 44 ] , but has also extended to other uses
such as on nervous tissue [ 10 , 42 ] , in facial plastic
and cosmetic surgery [ 6 , 28 , 39 ] , burns [ 36 ] and
chronic leg ulcers [ 46 ] . The autologous nature of
PRP also alleviates concerns of transmissible dis-
eases or immunogenic reactions [ 34 ] .
The potential clinical benefi t of PRP relies on the
maintenance of intact platelet -granules. Upon
activation, platelets degranulate and secrete an
array of mitogenic and chemotactic growth factors, which include platelet derived growth
factor (PDGF- , PDGF- , PDGF- ), transform-
ing growth factor (TGF- 1, TGF- 2), vascular
endothelial growth factor (VEGF), fi broblast
growth factor (FGF), interleukin-1 (IL-1), and epi-
Authors M. H. Pelletier 1 , A. Malhotra 1 , T. Brighton 2 , W. R. Walsh 1 , R. Lindeman 2
Affi liations 1 Surgical & Orthopaedic Research Labs, University of New South Wales, Randwick, Australia 2 Department of Haematology, Prince of Wales Hospital, Randwick, Australia
Abstract
▼ Platelet Rich Plasma (PRP) therapies require
blood to be processed prior to application, how-
ever, the full assessment of the output of platelet
sequestration devices is lacking. In this study the
products of the Autologous Fluid Concentrator
(Circle Biologics TM , Minneapolis, MN) and the
Gravitational Platelet Separation System (GPS,
Biomet, Warsaw, IN, USA) were evaluated in
terms of platelet viability and PRP constituents.
The AFC and GPS produced 6.4 ( ± 1.0) ml and
6.3 ( ± 0.4) ml of PRP, with platelet recovery of
46.4 % ( ± 14.7 %) and 59.8 % ( ± 24.2 %) produc-
ing fold increases of platelets of 4.19 ( ± 1.62)
and 5.19 ( ± 1.62), respectively. Fibrinogen con-
centration was increased above baseline PPP
produced with the AFC. pH was lower for both
of the processed samples than for whole blood.
White Blood Cell count was increased around 5
fold. Functional tests showed preserved viabil-
ity with both devices. This represents essential
knowledge that every treating physician should
have before they can confi dently administer PRP
therapy produced by any method. These are the
fi rst published results of platelet function for the
GPS system and the fi rst performance results of
the AFC system. The PRP produced is classifi ed
according to broad classifi cations as Leukocyte-
PRP (L-PRP) for both devices.
75Clinical Sciences
Pelletier MH et al. Platelet Function and Constituents … Int J Sports Med 2013; 34: 74–80
erations such as size, weight of the centrifuge, and ease of use.
The development of this classifi cation will hopefully resolve
variable and sometimes confl icting results regarding PRP thera-
pies in the literature [ 19 , 28 , 30 , 37 , 38 , 44 , 47 ] .
Here we present a fi rst evaluation of the Autologous Fluid Con-
centrator (AFC) (Circle Biologics TM , Minneapolis, MN), a platelet
sequestration device which has recently been granted approval
by the Food and Drug Administration for use in the United States,
and a paired comparison of an established device (GPS, Biomet,
Warsaw, IN, USA).
Materials and Methods
▼ Both devices yield PRP, with the GPS system additionally produc-
ing Platelet Poor Plasma (PPP) and the AFC off ering the option of
PPP or Concentrated Platelet Poor Plasma (c-PPP). An evaluation of
c-PPP, PRP, and whole blood included blood count, platelet count,
pH, white cell count, platelet recovery, and fi brinogen concentra-
tion. Platelet function testing on obtained PRP was performed in a
smaller subset of samples at 0 and 4 h to investigate platelet reac-
tivity and short-term stability. These tests included assessments
of p-selectin levels while resting and upon activation with Adeno-
sine Diphosphate (ADP), reaction to hypotonic shock and agonist-
induced light transmission aggregometry.
Subjects This study conforms to ethical standards in sport and exercise
research as described by Harriss, et al. [ 21 ] . Volunteers were
recruited at the University of New South Wales and Prince of
Wales Hospital following approval of the Human Research Ethics
Committee. A total of 64 subjects were recruited for the study.
Volunteers were excluded if taking anticoagulant, anti-platelet,
or anti-infl ammatory medication, or aspirin. Subjects were de-
identifi ed by assigning a number relating to all subsequent blood
test results.
Blood collection Blood was collected from the medial cubital vein with a 19 G
needle by trained phlebotomists. Blood was collected in a 4 ml
EDTA Vacutainer for Full Blood Count (FBC), a 4.5 ml Li-Heparin
Vacutainer for standard biochemistry, a 3.5 ml Citrate Vacutainer
for fi brinogen concentration, 1 ml heparin syringe for pH, and for
each device 2 × 30 ml syringes preloaded with acid-citrate-dex-
trose formula A (ACD-A) anticoagulant, 10 % ( ● ▶ Fig. 1 ).
Platelet sequestration The centrifuge and AFC devices arrived in sterile packages ready
for use. The AFC is a 2 chamber modular system. The devices
tested in this study utilised a plasma concentration chamber at
the rear and a chamber that, when viewed from the front, resem-
bled an hourglass. Inside the hourglass chamber there is a col-
lection tube with a moveable collection area. The GPS system
uses a buoy of specifi c density within a cylindrical chamber
( ● ▶ Fig. 2 ).
Blood and ACD-A fi lled syringes were emptied into the devices
which were centrifuged at 3 200RPM for 15 min at room tem-
perature. Following centrifugation the plasma, red cell and buff y
coat layers were clearly evident ( ● ▶ Fig. 3 ).
PPP was drawn off with syringes connected to respective ports
and was further processed by 3 fi lter passes for the AFC resulting
in (c-PPP). PRP was drawn off with a moveable collection win-
dow between 2 O-ring seals with the AFC and via a separate port
with the GPS device.
For the collection of PRP, the selector valve was turned back to
“draw”, the collection window was moved so that the top seal
was level with the top of the buff y coat and all available fl uid was
drawn off using a 10 ml syringe connected to port B. A portion of
both PRP, PPP and, where processed, c-PPP was placed in capped
test tubes for FBC (n = 59) and fi brinogen concentration (n = 59)
and a Li Heparin syringe for pH (n = 59). A portion of PRP was
also placed in a capped tube for tests of platelet function (n = 12
for each test).
Blood assays All blood assays were performed at the South Eastern Area Labo-
ratory Services (SEALS) which is accredited by the National
Association of Testing Authorities (NATA), Australia’s govern-
ment-endorsed, international laboratory accreditation body.
59 matched samples of whole blood, PRP and PPP were evalu-
ated for pH, fi brinogen concentration and FBC. pH was measured
on an ABL800 FLEX Radiometer Blood Gas Analyser (Radiometer,
Copenhagen). Full Blood Count was performed on a Roche Sys-
mex Model XE-2 100 (Sysmex Corp., Kobe, Japan). Biochemistry
assays were performed on a Beckman Coulter UniCel DXC880
blood analyzer (Beckman Coulter, USA). Fibrinogen assays were
performed on an STA R Evolution Coagulation Analyzer (Stago,
USA). Platelet recovery was calculated from platelet counts as
100 % *(platelet count of PRP * volume of collected)/(Platelet
count of whole blood * volume of whole blood). Platelet concen-
tration factor was calculated as Platelet count of PRP/whole
blood platelet count.
Surface P-selectin expression (CD62p antigen), as a measure of
platelet activation, was assessed by fl ow cytometry in resting
and activated state samples. Approximately 1 ml of PRP from
each device was set aside for p-selectin assays. Phosphate buff -ered saline (2.92 μl) or ADP agonist (40 μM) was added to 70 ml
subsamples of PRP. These samples were incubated for 15 min at
Fig. 1 Allocation of collected blood. Whole Blood in 2 x30ml syringes loaded
w (ACD-A) 10% foreach device
1 ml: Whole Blood(Heparin syringe)
4 ml: Whole Blood(EDTA vacutainer)
3.5 ml: Whole Blood(Citrate vacutainer)
4.5 ml: Whole Blood(Li-Heparin vacutainer)
AFC or GPSProcessing pH
PRP
Full Blood Count Biochemistry Fibrinogen
PPS; GPSc-PPP; AFC
pHFibrinogen
Full Blood Count
0h p-selectin resting0h p-selectin ADP0h Hypotonic Stress0h Aggregation
4h p-selectin resting4h p-selectin ADP4h Hypotonic Stress4h Aggregation
0h p-selectin resting0h p-selectin ADP0h Hypotonic Stress0h Aggregation
4h p-selectin resting4h p-selectin ADP4h Hypotonic Stress4h Aggregation
pH
Fibrinogen
Full Blood Count
76 Clinical Sciences
Pelletier MH et al. Platelet Function and Constituents … Int J Sports Med 2013; 34: 74–80
37 °C. 4 × 5 μl aliquots were taken and incubated with CD41a-
PerCP antibody and either 1μl CD62-PE or 1 μl mouse IgG1-PE
(isotope control) (DB Biosciences, San Jose, CA). These samples
were incubated in the dark at room temperature for 20 min.
600 ml of Ringer’s solution was added following incubation.
Samples were analyzed in a BD FACS Canto II fl ow cytometer
(Becton, Dickinson and Company, USA). For determination of %
p-selectin expression, forward and light scatter and fl uorescence
were acquired using logarithmic scale. A total of 20 000 platelet
events were gated. The isotope matched control antibody was
used to set threshold for CD62P positivity.
Hypotonic shock response (HSR) measures the platelet’s ability
to recover its normal volume after swelling when exposed to a
hypotonic environment. HSR is an optical method and PRP
required an additional soft spin (900RPM/10 min) to remove
intervening red cells for output from the AFC device. The result-
ing PRP was removed from the red cells with a pipette and plate-
let concentrations adjusted to < 500 × 10 9 /ml and allowed to rest
Fig. 2 Left to Right; Circle Biologics and Biomet
devices, before use (top row) and after centrifuga-
tion (bottom row).
Fig. 3 Close up of buff y coat interface, Circle
Biologics (Left) and Biomet (Right). Black arrows
indicate the PPP portion, the White arrows indicate
the buff y coat, and the double white arrows indi-
cate the red cell pack.
77Clinical Sciences
Pelletier MH et al. Platelet Function and Constituents … Int J Sports Med 2013; 34: 74–80
for 30 min. 6 × 150 μl of PRP from each device was transferred
into the wells of a microtitre plate. 3 × 150 μl of saline was added
to PRP samples and the microtitre plate was placed in a BioTeck
EL808 Ultra Microplate Reader (BioTek, Instruments, Inc.,
Winooski, VT). 150 μl of H 2 O were added with a micropipette to
the remaining 3 samples simultaneously, the lid was closed and
reading commenced. Readings of light transmission was per-
formed at 405 nm for 10 min at 15 s intervals. Results were
recorded with this protocol via Gen5 Data Analysis Software
(Generation 5, Toronto, Canada). Light transmission (T) was con-
verted to optical density according to OD = − log 10 T or T = lO − OD .
The optical density (OD) of the saline samples acted as the base-
line for measurements. The diff erence in the peak OD and the
baseline divided by the diff erence in OD at 10 min and baseline
produced a percent recovery.
Aggregation was performed following preparation to remove
red cells as above. 240 μl samples of PRP, PPP and c-PPP were
tested in a 4 station AggRAM aggregometer (Helena Laborato-
ries, Gateshead, UK). All samples were stirred with magnetic stir
bars at 600 rpm. Following a quality check, the OD was meas-
ured in the matched samples of PPP, which acted as a control.
The PRP was then placed in the aggregometer. Continuous read-
ing of OD continued for 10 min while 10 ul of collagen aggregant
was added to achieve a fi nal concentration of 20 μg/ml and fi nal
volume of 250 μl. Results of all tests were compared with paired
t-tests using Tukey’s criteria and a signifi cance level of 0.05.
Results
▼ Of the 64 blood samples collected for the study, none were
excluded based on irregular biochemistry results, one was lost
to human error, and post processing assays were precluded by
insuffi cient volume in 4 samples resulting in a sample size of
n = 59 for fi brinogen, cell count and pH tests and n = 12 for all
platelet function assays. 54.1 ( ± 0.2) ml (AFC) and 55.7 ( ± 1.5) ml
(GPS) of whole blood plus ACD-A was loaded into the device and
6.4 ( ± 1.0) ml for the AFC and 6.3 ( ± 0.4) ml for the GPS of PRP
was collected. Platelet recovery from whole blood samples for
the AFC and GPS was 46.4 % ( ± 14.7 %) and 59.8 % ( ± 24.2 %),
respectively, representing concentration factors (fold increase)
of 4.19 ( ± 1.62) and 5.19 ( ± 1.62). Fibrinogen concentration was
increased above baseline for c-PPP (AFC only) but not PRP. pH
was lower for both of the processed samples than for whole
blood. Results are summarized in ● ▶ Table 1 .
The results of platelet function assays are summarized in ● ▶ Table 2 .
HSR tests at 0 h and 4 h showed platelet recovery of around 54 %
with no diff erences detected between time points or device.
Tests of the aggregation of PRP revealed total aggregation at
10 min was above 95 % at both time points with no diff erences
found between time or device. Upon in vitro activation with col-
lagen there was a signifi cant increase in surface expression of
CD62p (P selectin) confi rming the collected platelets are func-
tional and capable of activation at 0 and 4 h for both devices.
Discussion
▼ The current study demonstrates that both the AFC and GPS
devices produces platelet and leukocyte enriched plasma in a
closed system with functional viable platelets in volumes suita-
ble for clinical use. Although platelet concentration has been
previously reported, platelet viability has not. Additionally the
AFC performance has not been reported.
Platelet function Damage to platelets during processing can lead to the premature
release and subsequent loss of growth factors [ 1 ] . Viable plate-
lets exhibit several behaviours when activated. Some of these
behaviours can be monitored such as shape change, aggregation
and surface marker exposure. Platelet aggregation is considered
the gold standard in evaluating platelet function [ 41 ] . Mean
platelet aggregation above 95 % in this study represents the
retention of activity. Likewise HSR has been indicated as a very
sensitive marker for viability [ 22 ] , and predictor of in vivo sur-
vival [ 23 ] . During these tests platelet shape recovery was appar-
ent. Granular release is closely related to p-selectin expression
[ 18 ] . Identifying p-selectin on the surface of activated platelets
allows them to be quantitatively assessed as a percentage of all
platelets present. While absolute numbers of p-selectin assays
are shown to vary a great deal between laboratories [ 12 ] , it is a
useful measure to determine activation for matched samples
within a study centre. The results of the current study showed
that platelets could be activated beyond the resting state, indi-
cating preserved function.
The broad classifi cation of PRP encompasses a range of constitu-
ents at varying concentrations. The basic defi nition dictates that
it contains a platelet concentration that is increased above
circulating levels. Marx et al. [ 29 ] describes PRP as a platelet
concentration in a 5 mL volume with 1 000 000 platelets/μL
AFC GPS
Whole Bld c-PPP PRP PPP PRP
fi brinogen (g/L) 2.96 (0.73) 3.30 (0.73) + * 2.97 (0.65) 2.95 (0.65)* 2.94 (0.64)
pH 7.34 (0.03) 7.13 (0.03) + * 7.02 (0.06) + * 7.11 (0.04) + * 7.08 (0.05) + *
platelet concentration ( × 10^9/L) 222 (45.7) 18.7 (12.7) + * 926 (378) + * 11.0 (6.2) + * 1149 (497.7) + *
WBC ( × 10^9/L) 6.36 (1.42) 0.04 (0.135) + 32.2 (11.7) + * 0.01 (0.027) + 31.0 (8.9) + *
*diff erence compared to whole blood value, + diff erence between devices, p < 0.05
Table 1 Blood Assays. Data
summary for whole and processed
blood samples showing mean
(standard deviation).
AFC 0 h AFC 4 h GPS 0 h GPS 4 h
HSR ( %) 54.0 (5.0) 54.1 (6.6) 50.6 (6.8) 55.6 (6.1)
aggregation ( %) 95.2 (7.3) 96.5 (7.1) 97.4 (3.4) 98.5 (2.7)
p-selectin resting ( %) 7.7 (4.2) 6.4 (3.0) + 12.4 (6.5) 9.8 (2.5) +
p-selectin activated ( %) 25.4 (8.1)* 24.3 (9.2)* + 26.5 (6.1)* ̂ 31.9 (3.6)* +^
*diff erence between resting value, + diff erence between devices, ^ diff erence between time points
Table 2 Platelet Function Test-
ing. Data summary for platelet
function assays showing mean
(standard deviation).
78 Clinical Sciences
Pelletier MH et al. Platelet Function and Constituents … Int J Sports Med 2013; 34: 74–80
(approximately a 4–5 fold increase). Gociman et al. [ 17 ] sum-
marized the reported platelet enrichments of 11 platelet-con-
centrating devices. These results showed fold increases ranging
from 1.34 to 6.07. In Gociman’s study the GPS produced a 4.86
fold increase on platelet count, which aligns well with the 5.19
increase seen here. The AFC was capable of increasing platelet
concentrations by a factor of 4.19 which places it roughly in the
middle of Gociman’s results. While many devices report fold
increases of 3–7 fold [ 3 , 6 , 43 ] , it is important to note that highly
concentrated platelet preparations may have an inhibitory eff ect
on healing [ 20 , 47 ] .
The devices were capable of recovering 46.4 % (AFC) and 59.8 %
(GPS) of all available platelets which is in line with previous pub-
lications. Leitner [ 26 ] compared 4 devices, and reported results
of 17 % for the Vivostat PRF Preparation Kit (Vivosolution A/S, Bir-
keroed, Denmark), 70 % for the Harvest SmartPReP 2 APC 60 Proc-
ess (Harvest Technologies Corp., Plymouth MA, USA) 67 % for the
PCCS Platelet Concentrate Collection System (3i PCCS, 3i Implant
Innovations, Palm Beach Gardens FL, USA) and 66 % for the Fibri-
net Autologous Fibrin & Platelet System (Cascade Medical Enter-
prises Ltd, Plymouth, UK). Everts et al. [ 14 ] reported results for
the Autologous Growth Factor Filter (Interpore Cross, Irvine CA,
USA) at 32 %, the Electa Cell-Separator (Sorin Group, Irandola,
Italy) at 48 % but reports a lower value for the GPS Platelet Separa-
tion System (Biomet Biologincs, Inc., Warsaw IN) at 36 %.
Although the ideal concentration of platelets within PRP remains
unknown, the eff ect of PRP is likely to be linked closely to plate-
let concentration. Leitner et al. [ 26 ] has shown that PDGF release
is closely related to platelet count. Likewise Everts et al. [ 14 ]
demonstrated the need for devices to maintain platelet viability
during preparation to maximise growth factor release upon acti-
vation when required. Regardless of the device or method used,
platelet recovery and viability are 2 important factors when
evaluating the PRP product.
Fibrin While platelet concentration and growth factor potential are
obviously essential aspects of PRP, fi brin (the activated form of
fi brinogen) is thought to directly induce angiogenesis by provid-
ing a matrix scaff old supporting cell migration and providing
chemotactic activity. The binding of growth factors such as
PDGF and VEGF to fi brin also supports wound healing [ 8 , 25 ] .
The fi brinogen concentration could therefore be an important
feature not always addressed when comparing diff erent PRP
products.
Passing the PPP repeated through the fi lter increased the con-
centration of fi brinogen, with each successive pass decreasing
the overall volume as water was removed. Our conservative
processing with the AFC showed an increased concentration
compared to whole blood. This process can be repeated any
desired number of times further increasing fi brinogen concen-
tration producing a product that could be used as a scaff old and
when added to PRP may alter the velocity of aggregation [ 24 ] ,
however, this was not examined in the current study.
pH Although there was a diff erence between the whole blood pH
and that of the processed samples, all groups were greater than
6.9 which is well above the level of 6.2 that would indicate a loss
of platelet function [ 22 , 33 ] . The increase in the acidity is likely
due to the addition of citrate to the blood from the ACD-A.
Leukocytes Platelets have a known direct and signifi cant role in immunity
and host defence against pathogenic microorganisms [ 4 , 45 ] .
Moojen, Everts et al. [ 32 ] demonstrated the antimicrobial poten-
tial of PRP in vitro. In addition to platelets, the PRP product col-
lected from the buff y coat layer has been reported to contain
around a 7-fold increase in leukocytes [ 7 ] . The output of both
the devices produced around a 5 fold increase above circulating
levels.
The collection of leukocytes with the PRP product is an area of
uncertainty. Anitua et al. [ 2 ] studied a concentrated platelet
product, Preparation Rich in Growth Factors (PRGF), which
avoids leukocyte content with the intent of avoiding the proin-
fl ammatory eff ects of leukocytes. An increase in leukocytes,
combined with the view of platelets themselves as innate
infl ammatory cells with acute host defence functions [ 16 ] ,
would suggest the PRP product may also be useful against post-
operative infections. The inclusion of leukocytes during process-
ing may also have benefi cial eff ects. Neutrophils have been
suggested to produce additional VEGF [ 48 ] , a protein known to
promote angiogenesis [ 2 , 50 ] . Autologous Conditioned Serum
(ACS) is created by incubating whole blood with glass beads, and
has been shown to increase the concentration of relevant growth
factors considerably without unwanted side eff ects in human
and animal tests [ 49 , 50 ] . A portion of the increase in growth
factor concentration may be attributable to the inclusion of leu-
kocytes during the activation process. Understanding whether
or not to include, and the extent to which WBCs are concen-
trated in the PRP is likely to be relevant for future studies. It has
been suggested that PRP products be described as Leukocyte
Platelet Rich Plasma (L-PRP), and Pure Platelet Rich Plasma
(P-PRP) to diff erentiate between the 2 PRP variations [ 11 ] .
Classifi cation It is commonly understood that not all PRPs are produced equal
[ 30 ] , and as such, PRP from diff erent methods and devices need
to be well characterised in order for the fi nal product to have the
anticipated eff ect. According to the specifi cations laid out by
Ehrenfest et al. [ 11 ] the product of the ACP device is classifi ed as
L-PRP as it contains leukocytes, just as the GPS output is. The
volume collected would be classifi ed as small ( < 25 % blood vol-
ume) except that it can be varied by adding fi brin-rich PPP in the
case of the AFC. Platelet and Leukocyte concentration were good
(40–80 % of all available) and platelets were healthy after collec-
tion. The systems we tested could be activated with any agonist
so fi brin is delivered unaltered, however the ability to concen-
trate the fi brin in the PPP with the option to add this to PRP may
alter the classifi cation.
One aspect that should be noted regarding the output of the AFC
is that the output can be purposefully modifi ed by the user by
changing the height of the collection window while drawing. The
protocol used in the present study collected a portion of the red
cell pack, the buff y coat and the plasma portion resulting from a
single spin. The PRP included some red cells which, while bio-
logically may have little impact in common use, but may be of
interest for future studies depending on the application. It is pos-
sible that the protocol could be altered to allow collections
excluding red cells, and possibly the leukocyte layer, although
this was not studied in the present study. The location of the
drawing window also dictates that a full blood draw be used with
the device so that the buff y coat layer is in the drawing area.
79Clinical Sciences
Pelletier MH et al. Platelet Function and Constituents … Int J Sports Med 2013; 34: 74–80
It is anticipated that the full characterisation of PRP produced by
the untested device and correlation with the existing device will
allow future studies to accurately assess what is being implanted.
Having this information widely available will also aid the practi-
tioner in determining if the output is appropriate for the needs
of their patients.
Acknowledgements
▼ The work presented in this study was supported by funding
from Circle Biologics, Minneapolis, MN. No individual author
received funding from Circle Biologics or other sources in rela-
tion to this study.
References1 Alsousou J , Thompson M , Hulley P , Noble A , Willett K . The biology of
platelet-rich plasma and its application in trauma and orthopaedic surgery: a review of the literature . J Bone Joint Surg Br 2009 ; 91 : 987 – 996
2 Anitua E , Sanchez M , Orive G , Andia I . Shedding light in the contro-versial terminology for platelet rich products . J Biomed Mater Res A 2009 ; 90 : 1262 – 1263
3 Arora N S , Ramanayake T , Ren Y F , Romanos G E . Platelet-rich plasma: a literature review . Implant Dent 2009 ; 18 : 303 – 310
4 Blair P , Flaumenhaft R . Platelet alpha-granules: basic biology and clini-cal correlates . Blood Rev 2009 ; 23 : 177 – 189
5 Borrione P , Gianfrancesco A D , Pereira M T , Pigozzi F . Platelet-rich plasma in muscle healing . Am J Phys Med Rehabil 2010 ; 89 : 854 – 861
6 Cervelli V , Gentile P , Scioli M G , Grimaldi M , Casciani C U , Spagnoli L G , Orlandi A . Application of platelet-rich plasma in plastic surgery: clini-cal and in vitro evaluation . Tissue Eng Part C Methods 2009 ; 15 : 625 – 634
7 Cieslik-Bielecka A , Gazdzik T S , Bielecki T M , Cieslik T . Why the platelet-rich gel has antimicrobial activity? Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007 ; 103 : 303 – 305 author reply 305-306
8 Clark R A . Fibrin and wound healing . Ann NY Acad Sci 2001 ; 936 : 355 – 367
9 de Vos R J , van Veldhoven P L , Moen M H , Weir A , Tol J L , Maff ulli N . Autolo-gous growth factor injections in chronic tendinopathy: a systematic review . Br Med Bull 2010 ; 95 : 63 – 77
10 Ding X G , Li S W , Zheng X M , Hu L Q , Hu W L , Luo Y . The eff ect of platelet-rich plasma on cavernous nerve regeneration in a rat model . Asian J Androl 2009 ; 11 : 215 – 221
11 Dohan Ehrenfest D M , Rasmusson L , Albrektsson T . Classifi cation of platelet concentrates: from pure platelet-rich plasma (P-PRP) to leu-cocyte- and platelet-rich fi brin (L-PRF) . Trends Biotechnol 2009 ; 27 : 158 – 167
12 Dumont L J , VandenBroeke T , Ault K A . Platelet surface P-selectin meas-urements in platelet preparations: an international collaborative study. Biomedical Excellence for Safer Transfusion (BEST) Working Party of the International Society of Blood Transfusion (ISBT) . Trans-fus Med Rev 1999 ; 13 : 31 – 42
13 Eppley B L , Woodell J E , Higgins J . Platelet quantifi cation and growth factor analysis from platelet-rich plasma: implications for wound healing . Plast Reconstr Surg 2004 ; 114 : 1502 – 1508
14 Everts P A , Brown Mahoney C , Hoff mann J J , Schonberger J P , Box H A , van Zundert A , Knape J T . Platelet-rich plasma preparation using three devices: implications for platelet activation and platelet growth factor release . Growth Factors 2006 ; 24 : 165 – 171
15 Everts P A , Hoff mann J , Weibrich G , Mahoney C B , Schonberger J P , van Zundert A , Knape J T . Diff erences in platelet growth factor release and leucocyte kinetics during autologous platelet gel formation . Transfus Med 2006 ; 16 : 363 – 368
16 Flad H D , Brandt E . Platelet-derived chemokines: pathophysiology and therapeutic aspects . Cell Mol Life Sci 2010 , doi: 10.1007/s00018-010-0306-x
17 Gociman B , Agk M , Moran S . Caption: a fi ltration-based platelet con-centration system . Expert Rev Med Devices 2009 ; 6 : 607 – 610
18 Graff J , Klinkhardt U , Schini-Kerth V B , Harder S , Franz N , Bassus S , Kirch-maier C M . Close relationship between the platelet activation marker CD62 and the granular release of platelet-derived growth factor . J Pharmacol Exp Ther 2002 ; 300 : 952 – 957
19 Grageda E , Lozada J L , Boyne P J , Caplanis N , McMillan P J . Bone formation in the maxillary sinus by using platelet-rich plasma: an experimental study in sheep . J Oral Implantol 2005 ; 31 : 2 – 17
20 Graziani F , Ivanovski S , Cei S , Ducci F , Tonetti M , Gabriele M . The in vitro eff ect of diff erent PRP concentrations on osteoblasts and fi broblasts . Clin Oral Implants Res 2006 ; 17 : 212 – 219
21 Harriss D J , Atkinson G . Update – ethical standards in sport and exercise science research . Int J Sports Med 2011 ; 32 : 819 – 821
22 Holme S , Sweeney J D , Sawyer S , Elfath M D . The expression of p-selectin during collection, processing, and storage of platelet concentrates: relationship to loss of in vivo viability . Transfusion 1997 ; 37 : 12 – 17
23 Kim B K , Baldini M G . The platelet response to hypotonic shock. Its value as an indicator of platelet viability after storage . Transfusion 1974 ; 14 : 130 – 138
24 Landolfi R , De Cristofaro R , De Candia E , Rocca B , Bizzi B . Eff ect of fi brinogen concentration on the velocity of platelet aggregation . Blood 1991 ; 78 : 377 – 381
25 Laurens N , Koolwijk P , de Maat M P . Fibrin structure and wound heal-ing . J Thromb Haemost 2006 ; 4 : 932 – 939
26 Leitner G C , Gruber R , Neumuller J , Wagner A , Kloimstein P , Hocker P , Kormoczi G F , Buchta C . Platelet content and growth factor release in platelet-rich plasma: a comparison of four diff erent systems . Vox Sang 2006 ; 91 : 135 – 139
27 Lieberman J R , Daluiski A , Einhorn T A . The role of growth factors in the repair of bone. Biology and clinical applications . J Bone Joint Surg Am 2002 ; 84-A : 1032 – 1044
28 Man D , Plosker H , Winland-Brown J E . The use of autologous platelet-rich plasma (platelet gel) and autologous platelet-poor plasma (fi brin glue) in cosmetic surgery . Plast Reconstr Surg 2001 ; 107 : 229 – 237; discussion 229–238
29 Marx R E . Platelet-rich plasma (PRP): what is PRP and what is not PRP? Implant Dent 2001 ; 10 : 225 – 228
30 Marx R E . Platelet-rich plasma: evidence to support its use . J Oral Maxillofac Surg 2004 ; 62 : 489 – 496
31 Marx R E , Carlson E R , Eichstaedt R M , Schimmele S R , Strauss J E , Georgeff K R . Platelet-rich plasma: Growth factor enhancement for bone grafts . Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1998 ; 85 : 638 – 646
32 Moojen D J , Everts P A , Schure R M , Overdevest E P , van Zundert A , Knape J T , Castelein R M , Creemers L B , Dhert W J . Antimicrobial activity of platelet-leukocyte gel against Staphylococcus aureus . J Orthop Res 2008 ; 26 : 404 – 410
33 Murphy S , Kahn R A , Holme S , Phillips G L , Sherwood W , Davisson W , Buchholz D H . Improved storage of platelets for transfusion in a new container . Blood 1982 ; 60 : 194 – 200
34 Nikolidakis D , Jansen J A . The biology of platelet-rich plasma and its application in oral surgery: literature review . Tissue Eng Part B Rev 2008 ; 14 : 249 – 258
35 Paderni S , Terzi S , Amendola L . Major bone defect treatment with an osteoconductive bone substitute . Musculoskelet Surg 2009 ; 93 : 89 – 96
36 Pallua N , Wolter T , Markowicz M . Platelet-rich plasma in burns . Burns 2010 ; 36 : 4 – 8
37 Peerbooms J C , de Wolf G S , Colaris J W , Bruijn D J , Verhaar J A . No positive eff ect of autologous platelet gel after total knee arthroplasty . Acta Orthop 2009 ; 80 : 557 – 562
38 Peerbooms J C , Sluimer J , Bruijn D J , Gosens T . Positive eff ect of an autolo-gous platelet concentrate in lateral epicondylitis in a double-blind randomized controlled trial: platelet-rich plasma versus corticoster-oid injection with a 1-year follow-up . Am J Sports Med 2010 ; 38 : 255 – 262
39 Rodriguez-Flores J , Palomar-Gallego M A , Enguita-Valls A B , Rodriguez-Peralto J L , Torres J . Infl uence of platelet-rich plasma on the histologic cCharacteristics of the autologous fat graft to the upper Lip of rabbits . Aesthetic Plast Surg 2010 , doi: 10.1007/s00266-010-9640-5
40 Rozman P , Bolta Z . Use of platelet growth factors in treating wounds and soft-tissue injuries . Acta Dermatovenerol Alp Panonica Adriat 2007 ; 16 : 156 – 165
41 Shah U , Ma A D . Tests of platelet function . Curr Opin Hematol 2007 ; 14 : 432 – 437
42 Shen Y X , Fan Z H , Zhao J G , Zhang P . The application of platelet-rich plasma may be a novel treatment for central nervous system diseases . Med Hypotheses 2009 ; 73 : 1038 – 1040
43 Stammers A H , Trowbridge C C , Murdock J D , Yen B R , Klayman M H , Hess W F , Woods E L , Andreychik D A . Establishment of a Quality Control Pro-gram for Platelet Gel Preparation: A Comparison of Four Commercial Devices . Presented at the Society of Cardiovascular Anesthesiolo-gists Meeting: 9 th Annual Update on Cardiopulmonary Bypass. March 14–19, 2004, Snowmass, CO; 2004
80 Clinical Sciences
Pelletier MH et al. Platelet Function and Constituents … Int J Sports Med 2013; 34: 74–80
48 Werther K , Christensen I J , Nielsen H J . Determination of vascular endothelial growth factor (VEGF) in circulating blood: signifi cance of VEGF in various leucocytes and platelets . Scand J Clin Lab Invest 2002 ; 62 : 343 – 350
49 Wright-Carpenter T , Klein P , Schaferhoff P , Appell H J , Mir L M , Wehling P . Treatment of muscle injuries by local administration of autologous conditioned serum: a pilot study on sportsmen with muscle strains . Int J Sports Med 2004 ; 25 : 588 – 593
50 Wright-Carpenter T , Opolon P , Appell H J , Meijer H , Wehling P , Mir L M . Treatment of muscle injuries by local administration of autologous conditioned serum: animal experiments using a muscle contusion model . Int J Sports Med 2004 ; 25 : 582 – 587
44 Sun Y , Feng Y , Zhang C Q , Chen S B , Cheng X G . The regenerative eff ect of platelet-rich plasma on healing in large osteochondral defects . Int Orthop 2010 ; 34 : 589 – 597
45 Tang Y Q , Yeaman M R , Selsted M E . Antimicrobial peptides from human platelets . Infect Immun 2002 ; 70 : 6524 – 6533
46 Villela D L , Santos V L . Evidence on the use of platelet-rich plasma for diabetic ulcer: a systematic review . Growth Factors 2010 ; 28 : 111 – 116
47 Weibrich G , Hansen T , Kleis W , Buch R , Hitzler W E . Eff ect of platelet concentration in platelet-rich plasma on peri-implant bone regenera-tion . Bone 2004 ; 34 : 665 – 671