Document & Record Control

For Medical Devices

Willem van den Biggelaar

Quality & Regulatory Consultant

PROCESS VISION

DOWN TO EARTH QUALITY SERVICES

Page 2: Document & Record Control For Medical Devices

2 of 28 © Process Vision – Document & Record Control

Content

• Quality Management System requirements

• Good Documentation Practices

• Tooling

• How to continue

Page 3: Document & Record Control For Medical Devices

3 of 28 © Process Vision – Document & Record Control

Quality Management System requirements

Page 4: Document & Record Control For Medical Devices

4 of 28 © Process Vision – Document & Record Control

Start at the basic: Architecture QMS

Work instructions - guidelines - checklists

templates

Deliverables that are

prescribed by the

processes,

procedures and work

instructions and

provide evidence of

conformity to

requirements of

products and

processes and

evidence of the

effective operation of

the QMS

Page 5: Document & Record Control For Medical Devices

5 of 28 © Process Vision – Document & Record Control

Traceability QMS

Quality Record

Template

Procedure

Process

Ref 1To which process

does a procedure belong?

Ref 2Which templates has this procedure

Ref 3From which template is this record?

Ref 1: Example of a procedure that refers to its process (and owner)

Ref 2: Example of a procedure that refers to its templates, guidelines,.....

Ref 3: By using the latest released template (id+version), reference 3 is assured

Page 6: Document & Record Control For Medical Devices

6 of 28 © Process Vision – Document & Record Control

Quality Records

• Device specific records: • Design History File (DHF)

• Device Master Record (DMR)

• Device History Record (DHR)

• Non specific device records: • Quality System Records (QSR)

– 820.20 (c) Management review records

– 820.22 Audit records

– 820.25 Training records

– 820.72 Calibration records

– 820.100 CAPA records

– 820.198 Complaint records

Page 7: Document & Record Control For Medical Devices

7 of 28 © Process Vision – Document & Record Control

ISO 13485 requirements control docs/records

Controlled Documents

• Readily Identifiable

• Legible

• Readiliy Retrievable

• Defined retention period

• Review & approve

• Change control

• Versioned

Records

• Readily Identifiable

• Legible

• Readily Retrievable

• Defined retention period

• Review & approve if needed

• Audit trail

• No versioning

Records are a special type of controlled document

(ISO 13485-4.2.3)

Page 8: Document & Record Control For Medical Devices

8 of 28 © Process Vision – Document & Record Control

Examples of Controlled documents

• Drawings

• Standard operating procedures (SOPs)

• Plans

• Specifications

• Designs

• Inspection instructions

• Test methods

• Device master records (DMRs)

• Forms

• Labeling including labels for in-process and final devices.

• Inspection and test reports

• Qualification and validation protocols and reports

• Audit reports

Page 9: Document & Record Control For Medical Devices

9 of 28 © Process Vision – Document & Record Control

DHF DMR DHR

product docs

Order

confirmed

End of

production

Start Design

Phase

Design &

Engineering

Production

EngineeringSales Production

Customer

(Service)

product docs order data service data

DHF index

Design

History

File

(DHF)

DMR index

Device

Master

Record

(DMR=TPD)

DHR Index

Device

History

Record

(DHR)

Device

History

Record

(DHR)

As

delivered

DHR Index

Device

History

Record

(DHR)

End of

Service

Start Series

Phase

Start Proto

Phase

Design

Transfer

production data

Abbr. Record Type Description

DHF 820.30 (j)

Design History

File

Compilation of records which described the design history of a finished device. There is a history file per

designed device.

It provides proof that product is designed according project plan.

E.g. System Requirements, System Design, test specifications, test reports, project plan.

DMR

(TPD)

820.81

Device Master

Record

Compilation of records containing procedures and specifications for a finished device. There is a master record

per designed device.

This information needed by manufacturing, end users and service.

E.g. assembly instructions, instructions for use, service manual,

DHR 820.184

Device History

Record

Compilation of records containing the production history of a finished device. There is a history record per

produced device (particular unit or batch of devices). It is the order, production and service history of a device

E.g. confirmed sales order, acceptance records (to DMR) and calibration records.

Page 10: Document & Record Control For Medical Devices

10 of 28 © Process Vision – Document & Record Control

820.184 Content DHR • Number of produced devices including production date

• Number of distributed devices

• Identification device(s) (SN#)

• BOM including (if applicable) component SN#

• Reference to DMR (e.g. Baseline number)

• Id’s of used (calibrated) measuring devices

• Id’s of used labels

• Id’s of used test software (version)

• Test results

• 820.80 Acceptatance results including conclusion, review and approval evidence

• 820.90 Non conforming records

Page 11: Document & Record Control For Medical Devices

11 of 28 © Process Vision – Document & Record Control

ISO 13485 requirements design documents 7.3.3 Design and development outputs

• Shall meet the input requirements for design and development

• Outputs examples: specifications, manufacturing procedures, engineering drawings

7.3.5 Design and development verification • Shall be performed in accordance with planned arrangements to ensure

that the design and development outputs have met the design and development input requirements.

7.3.6 Design and development validation • Shall be performed in accordance with planned arrangements to

ensure that the resulting product is capable of meeting the requirements for the specified application or intended use.

7.3.7 Control of design and development changes • Shall be identified

How to achieve above? See next sheet

Page 12: Document & Record Control For Medical Devices

12 of 28 © Process Vision – Document & Record Control

Requirements and test traceability example maximum

Product Validation

Report

System

Verification Report

Detailed Verification

Report

User Needs

Intended Uses

Regulatory standards

System Requirements

Detailed

Requirements

Implementation

(Integration) Test

ReportSystem Design

Detailed Design

Pro

du

ct

Ris

k M

an

ag

em

en

t F

ile

7.3.5

7.3.5

7.3.5

7.3.6

7.3.5

7.3.5

7.3.5

Page 13: Document & Record Control For Medical Devices

13 of 28 © Process Vision – Document & Record Control

Requirements and test traceability example minimum

Product Validation

Report

System

Verification Report

Detailed Verification

Report

User Needs

Intended Uses

Regulatory standards

System Requirements

Detailed

Requirements

Implementation

(Integration) Test

ReportSystem Design

Detailed Design

Pro

du

ct

Ris

k M

an

ag

em

en

t F

ile

7.3.5

7.3.5

7.3.6

7.3.5

7.3.5

Page 14: Document & Record Control For Medical Devices

14 of 28 © Process Vision – Document & Record Control

Requirements & test traceabiliy

• On requirements level not on document level

• You need to have uniquely identifiable requirements

• You need a tool

• Simple tool: excel

• Complext tool: DOORS

Page 15: Document & Record Control For Medical Devices

15 of 28 © Process Vision – Document & Record Control

DHF DMR DHR traceability

• DHR shows version of DMR with which it is build

• DMR must be traceable to DHF of which it is derived

• DMR must be traceable to used QMS templates

• DHF must be traceable to used QMS templates

Design

History

File

(DHF)

Device

Master

Record

(DMR)

Device

History

Record

(DHR)

Device

Master

Record

(DMR)

QMS templates

Page 16: Document & Record Control For Medical Devices

16 of 28 © Process Vision – Document & Record Control

DMR Template example (DMR DHR ) Reference to

DMR version

Id of DHR

Page 17: Document & Record Control For Medical Devices

17 of 28 © Process Vision – Document & Record Control

DMR example – QMS DMR DHR

Ref to QMS template

Ref to derived product

DMR template

Page 18: Document & Record Control For Medical Devices

18 of 28 © Process Vision – Document & Record Control

Good Documentation Practices

Page 19: Document & Record Control For Medical Devices

19 of 28 © Process Vision – Document & Record Control

Golden rules for document control

• Consistency throughout DHF, DMR and DHR • Same project names

• Same product names

• Same definitions and abbreviations

• Mind sequence of approval • E.g. test report after test spec, design after requirement

• Always use approved input documents

• Correct doc identification and version control

• Good Documentation Practices (see next sheet)

Page 20: Document & Record Control For Medical Devices

20 of 28 © Process Vision – Document & Record Control

Good Documentation Practices (GDP) • Gebruik permanente blauwe inkt

• Eenduidige datum notatie: bv JJJJ-MMM-DD

• Eenduidige naam notaitie: bv initialen + achternaam

• Leesbaar, bv in blokletters

• Geen lege cellen, NA alleen als is toegestaan

• Correcties? Doorhalen met enkele lijn, naam + handtekening + datum + reden

• Gebruik geen “aanhangsels” zonder identificatie

• Afronden getallen alleen bij uitkomst, niet bij bewerking

• Handtekening = verklaring dat je iets hebt gedaan, reden moet helder zijn voor de persoon die tekent

• Handtekening bij afwezigheid alleen door bevoegde personen

• Maak goede templates en train de mensen op GDP

Page 21: Document & Record Control For Medical Devices

21 of 28 © Process Vision – Document & Record Control

GDP bad example

Nummer en verkeerd format naam (had moeten zijn R. vd Boogaerd)

Geen voorgeschreven datum format en ook verkeerd ingevuld (yyyy-mm-dd)

Initials op template i.p.v. handtekening

Page 22: Document & Record Control For Medical Devices

22 of 28 © Process Vision – Document & Record Control

GDP good example of a template GDP Help text

Reason for signing

off

Format help

Page 23: Document & Record Control For Medical Devices

23 of 28 © Process Vision – Document & Record Control

GDP good example of a record

Page 24: Document & Record Control For Medical Devices

24 of 28 © Process Vision – Document & Record Control

Can I use digital signatures?

• For FDA compliancy with CFR part 11 is mandatory

• Part 11 will not apply when: • computers are used to generate paper printouts of electronic records

and

• these paper records meet all predicate rule requirements and

• persons rely on the paper records to perform regulated activities.

• Tools that are CFR part 11 compliant are still costly

• Alternative • Keep orginal paper documents for audit purpose

• Scanned pdf for publication purpose

Page 25: Document & Record Control For Medical Devices

25 of 28 © Process Vision – Document & Record Control

Tooling

Page 26: Document & Record Control For Medical Devices

26 of 28 © Process Vision – Document & Record Control

Tools are essential in document control

• Analyse current landscape

• Define tool requirements, e.g. • Interface possibilities with ERP systems

• FDA CFR 21 (digital signatures) compliancy

• Flows for review / approval

• Interface with change control tooling (or integrated)

• Generate doc id, automatic versioning

• Choose tool

• Make /Buy tool

• Validate tool (on intended use)

• Use tool

• Re-validate if intended use changes

Page 27: Document & Record Control For Medical Devices

27 of 28 © Process Vision – Document & Record Control

Document tool examples

PLM system

PLM system

Document Management System

Document Compliance System

CAPA, NC’s , Non conforming, Doc & Change control

Sharepoint user friendly flows plugin

Paperless manufacturing

Page 28: Document & Record Control For Medical Devices

28 of 28 © Process Vision – Document & Record Control

How to continue?

• Plan Define tooling requirements and choose

• Plan Setup procedures and templates

• Plan Setup tooling environment (inc validation)

• Do Pilot in typical project/product

• Do Train train train people

• Do Transfer of current project/products

• Check Regular Audit on compliance

• Check Train train train people

• Act Adapt if needed based on audits / feedback