Product Manual: PXP-60HF Poskom Co., Ltd. DOCUMENT NO.: PM-PXP60HF-0512 REVISION NO.: 1.1 DATE OF ISSUE: 2005-12-09 OPERATION & SERVICE MANUAL MODEL: PXP-60HF (PORTABLE X-RAY UNIT) ROOM 405, 1141-2, UNITECH VILLE VENTURE BLDG., BEAKSUK-DONG, ILSAN-GU, GOYANG, KOREA TEL: 82-31-906-9007, FAX: 82-31-908-4208

Upload: tulus-arisono

Post on 24-Oct-2015

261 views

Category:

Documents


52 download

TRANSCRIPT

Page 1: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd. DOCUMENT NO.: PM-PXP60HF-0512

REVISION NO.: 1.1

DATE OF ISSUE: 2005-12-09

OPERATION & SERVICE MANUAL

MODEL: PXP-60HF

(PORTABLE X-RAY UNIT)

ROOM 405, 1141-2, UNITECH VILLE VENTURE BLDG., BEAKSUK-DONG, ILSAN-GU, GOYANG, KOREA

TEL: 82-31-906-9007, FAX: 82-31-908-4208

Page 2: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

2

CONTENTS PART I. OPERATION MANUAL 1 General Description ---------------------------------------------------------------- 32 Operational Safety ---------------------------------------------------------------- 53 Environmental Safety ---------------------------------------------------------------- 64 Radiation Safety ---------------------------------------------------------------- 65 Electromagnetic Compatiblity ---------------------------------------------------------------- 76 System Configuration ---------------------------------------------------------------- 117 Operation Console Panel and Display ---------------------------------------------------------------- 128 Normal Mode of Operation ---------------------------------------------------------------- 139 Remote Control Function & Operation ---------------------------------------------------------------- 1510 Specifications ---------------------------------------------------------------- 19 TABLES 1 Specifications ---------------------------------------------------------------- 19 ILLUSTRATIONS 1 Operation Console Panel and Display ---------------------------------------------------------------- 122 Dimensions ---------------------------------------------------------------- 203 Labels ---------------------------------------------------------------- 21 PART II. SERVICE MANUAL

1 Notice ----------------------------------------------------------------------- 222 Operation Sheet ----------------------------------------------------------------------- 223 Block Diagram ----------------------------------------------------------------------- 234 Calibration Process ----------------------------------------------------------------------- 245 Calibration Code and Data ----------------------------------------------------------------------- 286 Error code & Countermeasure ----------------------------------------------------------------------- 287 Schematics ----------------------------------------------------------------------- 29 TABLES 1 mA Calibration Code ----------------------------------------------------------------------- 282 Filament Preheat Calibration ----------------------------------------------------------------------- 283 Error Code ----------------------------------------------------------------------- 28

Page 3: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

3

PART I. OPERATION 1. GENERAL DESCRIPTION 1.1. OVERVIEW PXP-60HF is one of PXP-HF portable family generators. It consists with the highest power inverter in PXP-HF family of X-ray generators. Likewise of others, the PXP-60HF is most compact, ultra-light weight and powerful HF inverter type of x-ray systems in this portable category. It is designed especially for nursing home care or home care medical market and small medical clinics as well as veterinary practitioner or battle field use, All PXP-HF family x-ray units are the ripple free HF type of generator in producing 40~50% more powerful radiation output than conventional x-ray units in its class. Therefore, PXP-60HF is the perfect x-ray portable unit for an indoor and outdoor medical diagnostic practitioner in the radiographic application where requires an easy of operation and mobility. Features are; -Ultra light weight and compact size -High frequency type of generator -High output 100kV / 60mA -APR pre-set (8) memory PROM -Remote control software -Built in collimator and laser pointer -Auto line compensation -Digital readout reverse switch -Accessories: metal carrying case, Hand switch, power cord -Power requirement: 220 Volt. 50 / 60 Hz

1.2. CONSOLIDATED MANUAL This consolidated manual consists of operating and service instruction. This operator’s manual provides PXP-60HF operating instructions for X-ray technicians, radiologists and other medical personal trained in the use of X-ray equipment. Service manual provides installation instruction, calibration and routine maintenance for field service personnel. This manual is not designed to replace or substitute for certified training in the radiological or medical field. In reading this manual, the operator should be qualified in the use of equipment of this type. This portable X-ray system is only to be used by licensed practitioners in diagnostic, operative and post-operative procedures.

Page 4: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

4

PXP-60HF portable X-ray can be used in variety of diagnostic radiographic systems such as mobile stand or radiographic table integrated with X-ray tube mounting devices. Although many of the features of these portable systems were developed with radiographic application in mind, operation is described in general terms to permit the use of those features in radiographic application 1.3. SAFTY REQUIREMENT This manual provides guideline for the safe use and operation of the portable system. The operator should be in trained in the use of equipment of this type and should refer to this manual in conjunction with that training. It is the responsibility of the owner to ensure that the system is operated only by properly trained, qualified personnel who have obtained credentials from local, state, and federal authorities where required. Potential hazards exist in the use of medical electronic devices and x-ray systems. All users and operators must understand the safety and emergency procedures, and the operating instructions given in this manual Poskom Co., Ltd. certifies that each system complies with the Radiation Control for Health Safety Act in Title 21, Chapter I, subchapter J of the Code of Federal Regulation in the USA and EN 60601, general requirements for radiation protection in diagnostic X-Ray equipment for IEC 601-1-3 in European Nations. Poskom and distributors assume no responsibility or liability for after-sale operating and safety practices, nor can it be responsible for personal injury or damage resulting form misuse of its portable x-ray systems. Please address a question and comments regarding safety issue to Poskom service organization and/or each distributor in your region.

1.4. CAUTIONS AND WARNING: Possible hazards are associated with unsafe operation of this type of portable X-ray equipment. These hazards are described on the following warning or caution signs with the safety precautions which should be taken.

CAUTION The X-ray radiation could be dangerous for patients and operators unless following safety regulations are strictly observed.

Page 5: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

5

2. OPERATIONAL SAFETY User’s safety, easy of operation and product reliability are our major concern in designing this PXP-HF family of generators. However, it is advisable to observe the following instruction and guideline for your additional safety in operating this portable.

A. This product should be operated only by or under the supervision of legally qualified people. B. PXP-60HF is designed for the radiographic uses and not for fluoroscopy or other associated applications. C. PXP-60HF should be used for the diagnosis, not for the therapy. D. PXP-60HF is specified as Class�type B equipment under the IEC 601-1, 2 standards.

E. Do not modify the equipment at your discretion and in case any modification is required unavoidably, ask the help of Poskom or its authorized dealer for the service. F. This system has been calibrated for optimal operations. G. If you have any troubles with the equipment, power it off immediately, and make contact with Poskom or its authorized dealers for their service and assistance. H. If you want to connect any devices made by other companies to this equipment, ask Poskom or its authorized dealers for their instructions.

THE PORTABLE IS NOT PROTECTED AGAINST INGRESSING WATER

OR LIQUIDS.

Page 6: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

6

3. ENVIRONMENTAL SAFETY 3.1. Avoid following places for the normal operation and safe storage.

A. Where the equipment is exposed to water vapor. B. Where the equipment is exposed to direct sunlight. C. Where the equipment is exposed to dust. D. Where the equipment is exposed to high humidity. E. Where there is a ventilation problem. F. Where the equipment is exposed to salty atmosphere. G. Where the equipment is exposed to chemicals or gas.

3.2. For sound operation of the machine, it should keep away from the place with strong vibration and maintain proper environment and conditions. A. Operation Environment

Temperature range 10°C ~ 40°C,

Relative Humidity Range 30% ~ 75% RH

B. Optimal Temperature and Humidity Temperature range 17°C ~ 23°C,

Relative Humidity Range 40% ~ 60% RH

3.3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere. Environment for storage and transportation

Temperature range -25°C ~ +60°C

Relative Humidity Range 10% ~ 95% RH

4. RADIATION SAFETY

A. Users and operators should wear appropriate protecting devices and clothes. B. Stay distance from the radiant sources and all the possible secondary radiation zones. C. Eliminate all unnecessary objects near the exposure zones. D. The distance from the focus to skin should be kept at least 20 cm. E. For the experimental uses, please apply the lowest possible values about kV, mAs and exposure time. F. Be careful not to exceed the limited radiograms in the exposure area.

CAUTION The ionized radiation could be dangerous for patients and operators unless following safety regulations are strictly observed.

Page 7: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

7

5. Electromagnetic Compatibility (EMC) acc. IEC 60 601-1-2:2002 The unit meets the Collateral Standards of Electromagnetic compatibility – Requirements and tests EN 60601-1-2 (IEC 601-1-2) the limits and methods of measurement of electromagnetic disturbance characteristics of industrial, scientific and medical radio frequency equipment EN 55011 Group 1, Class A and the requirements for interference resistance acc. 50082-1 incisiveness 2 and 5 are in compliance. Medical electrical equipment is subject in regard to the electromagnetic compatibility (EMC) and its special precautionary measure. The unit must in reference to the mentioned EMC-hints in the accompanying documents be installed and operated. Portable and mobile RF – communicating systems (such as cell phones) can have influence to medical electrical equipment. 5.1 General regulation for safety acc IEC 60601-1-2

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

The PXP-60HF is intended for use in the electromagnetic environment specified below. The customer or the user

of the PXP-60HF should assure that it is used in such an environment.

Emission test Compliance Electromagnetic environment - guidance

RF-emissions

Acc. CISPR 11 Group 1

The PXP-60HF uses RF energy only for its internal function.

Therefore, its RF emissions are very low and are not likely to

cause any interference in nearby electronic equipment.

RF – emission

Acc. CISPR 11 Class A

Harmonic emissions

Acc. IEC 61000-3-2 Not applicable

Voltage fluctuations/

Flicker emissions

Acc. IEC 61000-3-3

Voltage fluctuations

complies Flicker not

applicable

The PXP-60HF is suitable for use in all establishments,

including domestic establishments other than domestic and

those directly connected to the public low-voltage power

supply network that supplies buildings used for domestic

purposes.

Page 8: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

8

Guidance and Manufacturer’s Declaration – Electromagnetic Emission

The PXP-60HF is intended for use in the electromagnetic environment specified below. The customer or the user

of the PXP-60HF should assure that it is used in such an environment.

Immunity test IEC 60601

Test level Compliance level

Electromagnetic environment -

guidance

Electrostatic Discharge

(ESD) acc.

IEC 61000-4-2

± 6 kV

Contact

± 8kV

Air

± 6kV

Contact

± 8kV

Air

Floors should be wood, concrete or

ceramic tile. If floors are covered

with synthetic material, the relative

humidity should be at least 30%.

Electrical fast transient/

bursts acc. IEC 61000-4-4

± 2kV

for Power supply

lines

± 1kV

for I/O lines

(input/output)

± 2kV

for Power supply lines

± 1kV

for I/O lines

Mains power quality should be that

of a typical commercial or hospital

environment.

Surges acc.

IEC 61000-4-5

± 1kV

differential mode

± 2kV

common mode

± 1kV

differential mode

± 2kV

Common mode

Mains power quality should be that

of a typical commercial or hospital

environment.

Voltage dips, short

interruptions and voltage

variations on power supply

input lines acc. IEC 61000-

4-11

< 5% Ut

(>95% dip in Ut )

for 0,5 cycle

40% Ut

(60% dip in Ut )

for 5 cycles

70% Ut

(30% dip in Ut )

for 25 cycles

<5% Ut

(>95% dip in Ut )

For 5 s

<5% Ut

(>95% dip in Ut )

for 0,5 cycle

40% Ut

(60% dip in Ut)

for 5 cycles

70% Ut

(30% dip in Ut)

for 25 cycles

<5% Ut

(>95% dip in Ut)

for 5 s

Mains power quality should be that

of a typical commercial or hospital

environment. If the user of the PXP-

60HF requires continued operation

during power mains interruption, it is

recommended that the PXP-60HF

be powered from an interruptible

power supply or a battery.

Power frequency (50/60

Hz) magnetic field acc. IEC

61000-4-8

3 A/m 3 A/m

Power frequency magnetic fields

should be at levels characteristic of

a typical location in a typical

commercial or hospital environment.

NOTE Ut is the AC mains voltage prior to application of the test level.

Page 9: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

9

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

The PXP-60HF is intended for use in the electromagnetic environment specified below. The customer or the user of

the PXP-60HF should assure that it is used in such an environment.

Immunity test IEC 60601

Test level

Compliance

level Electromagnetic environment – guidance

Portable and mobile RF communications

equipment should be used no closer to any part of

the PXP-60HF, including cables, than the

recommended separation distance calculated from

the equation applicable to the frequency of the

transmitter.

Recommended separation distance

Conducted RF

Acc. IEC 61000-4-6

3 Vrms

150 kHz to 80 MHz 3 Vrms d = 1,2 √P

Radiated RF

Acc. IEC 61000-4-3

3 V/m

80 MHz 50 2,5 GHz 3 V/m d = 1,2 √P 80 MHz to 800 MHz

d = 1,2 √P 800 MHz to 2, GHz

Where P is the maximum output power rating of the

transmitter manufacturer and d is the

recommended separation distance in metres (m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey a,

should be less than the compliance level in each

frequency range b.

Interference may occur in the vicinity of equipment

marked with the following symbol :

NOTE 1 : At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and

reflection from structures, objects and people.

Page 10: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

10

a : Field strengths from fixed transmitters, such as base stations for radio ( cellular/ cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with

accuracy. To access the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey

should be considered. If the measured field strength in the location in which the PXP-60HF is used exceeds the

applicable RF compliance level above, the PXP-60HF should be observed to verify normal operation. If abnormal

performance is observed, additional measurements may not necessary such as reorienting or relocation the PXP-

60HF.

b : Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.

Page 11: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

11

6. SYSTEM CONFIGURATION 6.1. PXP-60HF is a radiological device for medical uses. It should be applied for the radiographic

diagnosis and operated by qualified practitioners. Users have to comply with safety and health regulations concerning the ionizing protection and the electrical and mechanical safety of the medical devices.

6.2 PXP-60HF is composed of the following parts and devices: They should be firmly fixed and connected with each other for the sound and normal operation.

A. Operation Board B. Inverter C. HV Tank including X-ray Tube D. Collimator E. Exposure Switch F. Power Cord

6.3. Supplied outfit and documents A. Aluminum Carrying Case B. Operation Manual C. Service Manual D. Power Plug E. Hand Switch (When you push hand switch in order to make radiography, it should be pressed by two steps: step one is ready and step two is the exposure.)

Page 12: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

12

7. OPERATING CONSOLE PANEL AND DISPLAY

Illustraion1. Operation console panel and display

No. Description Function

1 kV indicator Display kV value range: 40 -100kV

2 mAs indicator Display mAs value (range: 0.4-100mAs)

3 Ready lamp Lighted when you push the hand switch one time for preparation

4 Exposure lamp Alert lamp during x-ray exposure

5 Wait lamp Lighted until the second exposure started

6 kV up/down switch Select kV value by up or down buttons

7 mAs up/down switch Select mAs value by up or down buttons

8 APR Switch (1-8) Memory setting is available for 8 APR data

9 Store APR Data Save the selected APR Data

10 Display reversal switch Help to read LED value of kV, mAs reversed

11 Collimator lamp on Switch for operation of the lamp inside the collimator

12 Laser pointer To adjust exposure focus

13 Error lamp Lighted in case of system problem

1 2

3 4 5

6

7

8

9

10

11

12

13

Page 13: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

13

8. NORMAL MODE OF OPERATION

8.1. OPERATING PREPARATION A. Wear a lead apron while you make exposures. B. Stay distance at least 2 m from the unit or extend the cord of hand switch as long as possible in

order to secure enough space for your safety. C. Use proper field size and technical values for each procedure to minimize x-ray exposure dose

and gain the best radiographic result. D. Be careful about the decimal point on the mAs display in order to acquire intended mAs values. E. When you make exposure to the patient on the bed, ask visitors to leave the room first and

keep appropriate distance from the patient. F. Pay attention to the maintenance schedule of the device and keep up with it. G. Cumulative radiation dose usually does not exceed recommended maximum levels. However, if

you make radiographic exposures using high kV and mAs quite often, the evaluation of the specialist is needed to check whether extra protective devices are necessary for the user or not.

8.2. COLLIMATOR AND LASER POINTER CHECK A. Place the loaded cassette on the patient’s backside. B. Arrange the SID (Focal Spot to Image Receptor Distance) using a scale located on the side of

the collimator. C. Turn on the collimator lamp and laser pointer by pushing collimator/laser on switch. D. Adjust the size of the x-ray beam in accordance with the film size using the x-ray field

adjustment knobs. Turn off the collimator and laser pointer after 30 seconds on the timer.

WARNING Please check the input voltage and the power requirement are in accordance with the system labels, which attached on the body of the machine. The input voltage tolerance within + 10% is acceptable for normal operation.

CAUTION Verify thoroughly a power cable and other cable interconnection before operating the portable and making x-ray exposure.

CAUTION When the portable is first used after purchase or x-ray tube is cool, x-ray tube is needed for warming up to secure long life of the tube. Following method of pilot test will relieve the damage on the x-ray tube. Operate the machine with low kV values (50kV, 5mAs) three times consequently and execute with higher values [90kV, 50mAs].

Page 14: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

14

kV mAs

kV mAs

8.3. NORMAL MODE A. Please open the carrying case. B. Connect power plug into the PXP-60HF device. C. Turn on the line switch and the display will be shown up as follows.

D. Select the kV values using the [6] kV up/down switch. E. Select the mAs values using the [7] mAs up/down switch. F. Positioning the machine at the ready condition by one time pushing of the hand switch. G. When the [3] ready LED lighted, press the hand switch again and make exposure. During the x-

ray exposure, x-ray LED will be lighted on. H. After the exposure is ended, [5] “wait LED” sign will be lighted on and off for the waiting time. 7.3. APR MODE A. Please open the carrying case. B. Connect power plug into the PXP-60HF device. C. Turn on the line switch and the display will be shown up as follows.

D. Select the [8] APR switch out of 1~8. E. Select the kV values using the [6] kV up/down switch. F. Select the mAs values using the [7] mAs up/down switch. G. Save the selected kV and mAs figures with pushing the [9] “store” button. H. When the [3] ready LED sign appears, press the hand switch again and make exposure.

During the x-ray exposure, the [4] X-ray LED will be lighted on. I. After the exposure is ended, [5] “wait LED” will be lighted on and off for the waiting time.

CAUTION If you want to make exposures several times in row, wait until the [5] wait lamp is lighted off. If the number of consecutive exposure exceeds three times (at higher than 50mAs), wait for 3 minutes in order to prevent overheating of the machine.

NOTE As it was illustrated in the front pages, the exposure switch should be pressed by two steps. However, you can make immediate exposure in case the filament of the tube is hot enough.

Page 15: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

15

9. REMOTE CONTROL FUCTION & OPERATION 9.1. INTRODUCTION Currently, mAs, kVp and Pre-set memory techniques (PROM memory) can be selected by the operator by depressing push buttons or by scrolling displays on a control panel. Each technique or mode selection depends on an anatomic region or view for radiographic examination. The collimator lamp and laser pointer is turned “on” by the lamp push-button and turned “off” by a 30 second collimator timer run off. Many portable operators and field users have suggested we find a better way to select an exposure technique and parameters along with mAs, kVp and pre-set PROM memory values. Most field operators hold portable X-Ray units with one hand, while holding the exposure hand switch with the other hand. This makes selecting technique factors on the control panel awkward without dropping the portable unit or hand switch. We have modified our X-ray portable software and functions of the Prep. and Exposure hand switch to allow selection of all exposure parameters (mAs & kVp) and control of the collimator lamp and laser beam. The exposure hand switch consists of a two step hand switch. The first (1) stage of the hand switch is called the “Preparation (Prep. or Ready)” stage. The second (2) stage of the hand switch is called the “Exposure (Exp or X-Ray out)“ stage. (See drawing 1.) All functions of the remote control described above are executed by clicking the first stage on the hand switch which is the “Prep.” mode of operation. Each remote control function and technique value selection is described below. (Drawing 1) 9.2. REMOTE CONTROL MODE

1st stage switch = Prep. Switch

2nd stage switch = Exp. Switch

Hand switch housing

Page 16: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

16

The kVp and mAs setting, memory selection and collimator lamp and laser pointer illumination functions are controlled remotely by the hand switch by pressing the “Prep.” stage push-button as follows:

<TRIPLE CLICK>

9.3. MEMORY SELELCTION When you click the PREP stage hand switch quickly three (3) times, the Pre-set Memory station LED (PXP-60HF: 1-8) will blink sequentially as well as the kVp and mAs LED readouts. Make a single (1) click with the hand switch when the desired pre-set memory station lamp begins to blink. This will cause the pre-set memory data to be selected and displayed in the control LCD window.

9.4. kVp SELELCTION One (1) click of the “Prep” switch will cause the kVp LED readout to blink. You will now be in the kVp selection menu mode. The kVp values increase by one (1) kVp unit per click of the “Prep.” switch. If you depress and hold the prep switch for a few second, the kVp readout will increases rapidly up to the maximum value (PXP-60HF: 100 kVp). The kVp readout will circulate from lowest to highest values as long as you hold and press the button. When the desired kV readout appears on the control

ATTENTION Click the “Prep.” hand switch three (3) times consecutively. This will allow you to enter in the “remote hand control” menu mode for operation.

NOTE All pre-set memory data should be entered and stored at each memory station (PXP-60HF: 1-8), prior to performing the above mode of operation.

Page 17: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

17

display panel, release the prep hand switch and wait for a few seconds. The kVp value will set up automatically.

9.5. mAs SELECTION mAs selection is similar to the kVp selection mode. Click the “Prep” hand switch once to make the mAs LED readout blink. You will be in the mAs selection menu mode. The mAs value will increase with each click of the “Prep” switch (0.4 mAs to the maximum mAs of each portable x-ray unit). If you wish to increase the mAs value quickly, simply hold the “Prep” switch down one stage. (PXP-60HF: 100 mAs). The mAs readout will circulate as long as you hold the “Prep” switch. When the desired mAs readout appears on the control display panel, release the prep hand switch and wait for a few seconds. Then mAs value will set automatically. 9.6. COLLIMATOR / LASER BEAM TURNED ON <DOUBLE CLICK> With the exposure technique (mAs & kVp) set with your remote hand switch, you are ready to make a radiographic exposure for examination. It is very important to position the right region of interest on a patient’s anatomy. Therefore, the collimator beam must be turned on to confirm the size of region and size of film covered. The laser beam must also be turned on for checking the precise center of the target prior to making a final X-ray exposure. When the Prep switch is rapidly clicked twice, the collimator lamp and laser beam will turn on simultaneously until the lamp timer (30 seconds) runs off. However, it will turn off automatically when x-ray exposure sequence is completed. 9.7. X-RAY EXPOSURE After the x-ray exposure technique is selected and collimator light beam size is adjusted with the collimator blade knob, an x-ray exposure can be made by an authorized individual of the medical clinic or hospital.

ATTENTION Click the Prep. hand switch rapidly twice. The collimator lamp and laser pointer beam will illuminate for X-ray exposure positioning.

WARNING When an X-ray exposure is made in a room or outdoors, lower radiation will scatter throughout the surrounding area. An authorized operator is expected to take precautions and is highly recommended to ware a lead apron and gloves if necessary for your patient’s and your own protection.

X-RAY EXPOSURE SEQUENCE: A. Press the Prep hand switch (1st stage) and hold B. Press down the 2nd stage hand switch. Exposure will be made automatically by the exposure timer.

Page 18: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

18

< 1ST STAGE> <2ND STAGE> 9.8. REMOTE CONTROL OVERRIDE

NOTE TO THE USER All of the before mentioned remote control hand switch functions and operations can also be accomplished via the one-touch control panel. All exposure functions and parameters can be Overridden through the main control panel prior to final exposure. (See user’s manual for directions on how to use the main control panel)

Page 19: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

19

10. SPECIFICATIONS

[Table 1. Specifications] Output Power 3.2 kW

Voltage 200V-240V

Phase & Frequency Single / 50/60 Hz

Input Power

Current Short time 18A

Tube voltage current mAs

40kV ~ 50kV 60mA 0.4 ~ 100

51kV ~ 70kV 40mA 0.4 ~ 64

61kV ~ 70kV 40mA 80 ~ 100

71kV ~ 80kV 40mA 0.4 ~ 32

71kV ~ 80kV 35mA 40 ~80

81kV ~ 90kV 35mA 0.4 ~ 40

81kV ~ 90kV 30mA 50 ~ 80

91kV ~ 100kV 30mA 0.4 ~ 50

Radiography kV Range in 1kV steps

91kV ~ 100kV 25mA 64 ~ 80

MAs Range 0.4mAs – 100mAs, 25steps

Max. kV Deviation ±7 %

Max. mAs Deviation ±10 %

Indication kV (error code) /mAs: 7-segment LED

Model Name D-205 TOSHIBA

Focal Spot 2.0mmx2.0mm

Target Angle 20 degree

Anode Heat Storage 40 kHU

X-ray Tube

Heat Dissipation 378 HU/sec

Total Filtration 2.5 mm Al eq. @ 100kV

Compensation for the Supply Power

Regulations

Automatic

Type Double slit type, Manual operation

Min. x-ray field size ≦5cm x 5cm @1m SID

Max. x-ray field size 40cm x 40cm @ 75cm SID

Laser Pointer (Option) Class : ⅢA 5mW *Apply only for veterinary use

Timer Push button illuminator with 30 sec timer

Collimator

Lamp 12V / 100W

Page 20: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

20

Dimension See drawing Fig- (460mmx250mmx200mm)

Weight 13 kg

Illustration 2. Dimension (Main Unit & Carrying Case)

1) PXP-60HF

2) CARRING CASE (ALUMINUM)

Page 21: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

21

Illustration 3. Labels PXP-60HF Portable Label Sticker and location

High Voltage Tank MODEL: XXXXXXX SER.NO: XXXXXX Rating: Max. 100kV, 60mA, Max output: 3.2 kVA POSKOM Room 405, Unitech Ville Bldg., 1141-2, Baeksuk-dong Ilsan-gu, Goyang, Korea Warning: Electric shock hazard do not remove cover Reserve Servicing to qualified personnel.

Warning: This X-ray unit maybe dangerous to patient and operator unless safe exposure factors and operation instruction are observed

DIAGNOSTIC X-RAY UNIT MODEL: PXP-60HF SER. NO.: DATE: INPUT Rating: AC200~240V, 50/60 Hz Rating: X-ray Tube Voltage: 40~100kV

X-ray current: 30~60mA Operation condition: Time interval between 3 exposure: 3 min (When the mAs is higher than 50) Filtration: Total filtration: Min 2.5 mmAl Tube: 0.8 mmAl, collimator, 0.5 mmAl, Al Filt.: 1.2mmAl X-ray tube (TOSHIBA), Model: D-205, Serial no.: Focal spot: 1.2mm X 1.2mm

POSKOM Room 405, Unitech Ville Bldg., 1141-2, Baeksuk-dong, Ilsan-gu, Goyang, Korea

Take adequate Precautions to prevent the possibility of any persons carelessly exposing themselves or others to radiation.

Beam Limiting Device (Collimator) MODEL: PCMAX-100 SER.NO: XXXXXXX Input Rating: 24V, 150W Min. X-ray field size: Below 5cm x 5 cm at SID 100cm Max. X-ray field size: Below 35cm x 35cm at SID 65cm Max. Tube Voltage: 100kV, Inherent filtration: 0.5 mmAl POSKOM Room 405, Unitech Ville Bldg., 1141-2, Baeksuk-dong Ilsan-gu,

Page 22: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

22

PART II. SERVICE 1. NOTICE 1.1. If the output calibration is needed after periodical inspections or repairs, please observe following rules and methods. CAUTION

The ionized radiation is dangerous for the operator if the following safety measures are not strictly observed.

1.2. This manual is designed to ensure correct use and operation of PXP-60HF. Please read all the lines thoroughly before you use this equipment. 1.3. Incorrect use and operation exceeding described conditions in this manual may occur damage of the machine and shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein. 1.4. This equipment should be used only by the legally qualified persons and practitioners. 1.5. Regarding the AC cable and software, only the one developed and supplied by Poskom should be used. 1.6. No responsibility is taken by Poskom for any infringement of patients or other rights of the third parties which many results from the use of this manual. 1.7. Always keep the manual at hand for your reference. 2. OPERATION SHEET APR SWITCH 1~8 kV UP/DOWN SWITCH

Page 23: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

23

mAs UP/DOWN SWITCH STORE SWITCH

LASER SWITCH

COLLIMATOR ON/OFF SWITCH

Page 24: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

24

3. BLOCK DIAGRAM

PXP60HF-CTL-001

+12V

CON5

35312-05

1 2 3 4 5

LAM

P_C

ON

T

CON3

35312-11

1 2 3 4 5 6 7 8 9 10 11

* EXPORT : NORMAL

CON335312-05

1 2 3 4 5PXP60HF-PWR

GND

CON10

35312 12

PWR & DRIVE

C1

CON1

53258-05

12345

2

EP

_CO

NT2

CON7

35312 12

20

AGND

THERMAL

12

+5V

TB4

SHIELD LINE

AG

ND

250V/16A

CN235312

1 2

FIL_

FB

CON1

35312-03

1

2

3

COM

LASER

CON2

35312-02

12

575

PLUG AC MALECN1

CONTROL BOARD

FG

T2CON5

35312-03

123

-12V

CON435312-11

1 2 3 4 5 6 7 8 9 10 11

CN135312

1 2

FIL_

CO

NT1

250V/20A

* DOMESTIC : NOT USED

IGBT BOARD

DG

ND

EP(-)

TB1

SHIELD LINE

PXP-60HF-Main Block

<OrgName>

1 1

A3 A<Cage Code>

Size FSCM NO. DWG NO Rev

Sheetof

APPROVED BY DESIGNED BY

CHECKED BY DATE

C2

12V/50W

IP_FB(+)

LAMP

D-205

EP

_CO

NT1

T1

TANK

3V/35mA

OP PANEL

CON9

3531212

READY

X-RAY

IP(+)

FILAMENT DRIVE

HAND SWITCH

PXP-60HF-IGBT

& HT CONT.

CN435312

1 2

/100

V_S

EL

CON6

35312-02

12

L101

3TURNS

FIL_

CO

NT2

TB2

CB1

CON6

53258-02

12

CON8

3531212

SHIELD GND

CON2

35312-05

1 2 3 4 5

TB1T1

HT XFMR

COLLIMATOR DRIVE

CN335312

1 2

TB2

COLLIMATOR

T2

FIL XFMR

3

CON4

53258-04

1234

T

3

TB3

EP_FB(-)

+12.2

LASER POINTER

Page 25: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

ProductManual: PXP-60HF Poskom Co., Ltd.

25

kV mAs

-PREPARATION

Following tools and instruments are needed. a. Tools: Philips screwdriver No.1 & 2, Straight slot screwdriver b. Instruments: DVM, oscilloscope (Tektronics TDS 3032 and more than Equivalent)

4. CALIBRATION PROCESS

4.1. Line Power Supply

-Line switch should be at off position. -Connect power line cord to the power outlet. -Turn on the line switch. -Then the LED on the operation panel illuminates.

4.2. Calibration procedures are composed of three steps. -Tube voltage confirmation and calibration. -Tube current confirmation and calibration. -Filament preheat calibration

A. Calibration and Confirmation of Tube Voltage -Turn off the line switch. -Lift up the control panel from the unit, and then the PCB OP board appears. -All DIP switches on the PCB OP board are turned off. (Note: DIP switch is usually located in this location.)

-Connect the channel one probe of the oscilloscope to TP5 (IP F/B). -Connect the channel another one probe of the oscilloscope to TP6 (EP F/B).

-Connect the ground to TP11 (GND). -Turn on the line switch and the display indicates as follows.

-Set the kV and the mAs at 40kV and 0.4mAs (60mA). Then observe the mA waveform of the oscilloscope channel 1.

-If it is not about 9V, adjust mA by VR3 on the OP board. Then observe the kV waveform of the oscilloscope channel 2.

-If it is not 2V, adjust kV by VR2 on the OP board. After adjustment, set the tube voltage kV to 80KV by the kV up/down switch.

-Set the kV and the mAs at 80kV and 0.4mAs(40mA).Then observe the mAs, kV waveform of the oscilloscope channel 1, 2.

Channel 1 is mA, and 60mA = 9 V Channel 2 is kV, and 40kV = 2V

Page 26: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

ProductManual: PXP-60HF Poskom Co., Ltd.

26

kV mAs

kV mAs

-If it is not about 6V, adjust mA by VR3 on the OP board. -If it is not about 4V, adjust kV by VR2 on the OP board.

B. Calibration and Confirmation of Tube Current

-Turn off the line switch. -Lift up the control panel from the unit, and then the PCB OP board appears. -All DIP switches on the PCB OP board are turned off. (Note: DIP switch is usually in this position.) -Connect the channel one probe of the oscilloscope to TP5 (IP F/B). -Connect the channel another one probe of the oscilloscope to TP6 (EP F/B).

-Connect the ground to TP11 (GND). -Turn on the line switch and the display indicates as follows.

-Set the kV and the mAs at 40kV and 0.4mAs (60mA). Then observe the mA waveform of the oscilloscope channel 1. -If it is not 7V, adjust mA by the calibration data on the OP board. -Turn off the line switch, and wait until the display switch the light off. -DIP switch 1 on the PCB OP board is turned on. -Turn on the line switch and the display indicates as follows. -Then push the mAs up switch (see operation sheet). The display indicates as follows.

C00 is calibration code of tube current (60mA at 40kV~50kV). 179 is calibration data of tube current (60mA at 40kV~50kV). -Adjust mA by the mAs up/down switch (see operation sheet). -After it is adjusted, push the store switch (see operation sheet). -DIP switch 1 on the PCB OP board is turned off then exposure X-ray. -If it is not 9V, set the calibration data again. -DIP switch 1 on the PCB OP board is turned on. -Then push the kV up switch (see operation sheet).

Channel 1 is mA, and 60 mA = 9V Channel 2 is kV, and 40 kV = 2V

Calibration code Calibration data

Page 27: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

ProductManual: PXP-60HF Poskom Co., Ltd.

27

kV mAs

kV mAsPreheat calibration code Preheat calibration data

The display indicates as follows.

C01 is calibration code of tube current (40mA at 51kV~70kV). 153 is calibration data of tube current (40mA at 51kV~70kV). -Adjust mA by the mAs up/down switch (see operation sheet). -After it is adjusted, push the store switch (see operation sheet). -DIP switch 1 on the PCB OP board is turned off then exposure X-ray. -If it is not 6V, set the calibration data again. -Table .1 shows the calibration code and the calibration data against an each tube current.

C. Filament preheat calibration

-This is preheating filament for making tube current. -The overshoot or the undershoot may occur from the tube current if the preheat calibration data or the VR3(preheat gain volume) is set by mistake.

-There are two methods to make preheat calibration setting. The calibration data is possible to be changed in accordance with the tube condition.

-The VR3(preheat gain volume) adjusts the whole range of tube current. Filament preheat calibration data adjusts each tube current.

-Turn off the line switch. -Lift up the control panel from the unit, and then the PCB OP Board appears. -DIP switch 1 on the PCB OP board is turned on. -Turn on the line switch.

-Push APR switch ① and then the display indicates as follows.

P00 is Filament preheat calibration code (Preheat calibration code of 60mA). 17 is Filament preheat calibration data (Preheat calibration data of 60mA). -Adjust data by the mAs up/down switch (see operation sheet). -After it is adjusted, push the store switch (see operation sheet). -DIP switch 1 on the PCB OP board is turned off then exposure X-ray. If overshoot or undershoot occur in the waveform, set the calibration data again. Then push the kV up switch (see operation sheet).

Calibration code Calibration data

Page 28: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

ProductManual: PXP-60HF Poskom Co., Ltd.

28

kV mAs

The display indicates as follows.

P01 is Filament preheat calibration code (Preheat calibration code of 40mA). 37 is Filament preheat calibration data (Preheat calibration data of 40mA).

-Adjust data by the mAs up/down switch ⑥ (on the control panel).

-After it is adjusted, push the Store switch ⑨ (on the control panel).

-DIP switch 1 on the PCB OP board is turned off then exposure X-ray. -If overshoot or undershoot occur in the waveform, set the calibration data again. -Table. 2 shows the calibration code and the calibration data against an each tube current.

The calibration data is possible to be changed in accordance with the tube condition.

Preheat calibration code Preheat calibration data

Page 29: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

ProductManual: PXP-60HF Poskom Co., Ltd.

29

5. CALIBRATION CODE AND DATA

[Table 1. mA Calibration Code and Data]

AC voltage Code Data Tube voltage Tube current mAs

220V C00 235 40kV ~ 50kV 60mA 0.4 ~ 100 220V C01 156 51kV ~ 70kV 40mA 0.4 ~ 64 220V C02 155 51kV ~ 70kV 40mA 80 ~ 100 220V C03 155 71kV ~ 80kV 40mA 0.4 ~ 32 220V C04 136 71kV ~ 80kV 35mA 40 ~80 220V C05 120 81kV ~ 90kV 30mA 0.4 ~ 40 220V C06 115 81kV ~ 90kV 30mA 50 ~ 80 220V C07 115 91kV ~ 100kV 30mA 0.4 ~ 50 220V C08 97 91kV ~ 100kV 25mA 64 ~ 80

[Table 2. Filament Preheat Calibration Code and Data]

AC voltage Code Data Tube current

220V P00 7 60mA 220V P01 60 40mA 220V P02 75 35mA 220V P03 85 30mA 220V P04 95 25mA

6. ERROR CODE & COUNTERMEASURE [Table 3. Code & Measures]

Error Code Content Action

Err. 07 When the line switch is turned on, this is inspected. The ready contactor of the hand switch is closed even it is not pushed.

Change the hand switch

Err. 08 When the line switch is turned on, this is inspected. The x-ray contactor of the hand switch is closed even it is not pushed.

Change the hand switch

Err. 10 It is displayed when the FIL FB signal is not detected in the state of standby.

Replace operation board or Replace HV tank

Err. 11 It is displayed when the EP FB signal is not detected in the state of standby.

Replace operation board or Replace HV tank

Err. 12 It is displayed when the IP FB signal is not detected in the state of standby.

Replace operation board or Replace HV tank

Err. 14 It is displayed when the FIL FB signal is not detected in the state of ready.

Replace main board or Replace power board

Err. 15 It is displayed when the EP FB signal is low in the state of exposure.

Replace operation board

Err. 16 It is displayed when the IP FB signal is low in the state of exposure.

Replace main board or Replace power board

Err. 17 It is displayed when the IP FB and EP FB signal is high in the state of exposure.

Replace main board or replace power board

Err. 19 It is displayed when the mAs is over in the state of exposure.

Check the IP FB level and adjust it Replace operation board

Page 30: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

ProductManual: PXP-60HF Poskom Co., Ltd.

30

Err. 21 It is displayed when the connector CON6 on the main board is not insert.

Check CON3

Page 31: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

31

7. SCHEMATICS

CONTROL BOARD OP PANEL & HT CONT.

HTP-24 PORTABLE X-RAY

3 7

A3 0<Cage Code>

Size FSCM NO. DWG NO Rev

Sheetof

APPROVED BY DESIGNED BY

CHECKED BY DATE

+5V

SEG04_OUTSEG03_OUTSEG02_OUTSEG01_OUT

SEG06_OUTSEG05_OUT

SEG_ASEG_BSEG_CSEG_DSEG_ESEG_FSEG_G

SEG_DP

SEG07_OUT

SEG08_OUT1

R2615 1/4W1 2

LED10 M612R28

15 1/4W1 2

LED14 REVERSE12

LED11 M712

LED13 ERROR12

LED12 M812

SEG6 uD_SR105K37 6 4 2 1 9 10 5

8C

CA B C D E F G

DP

CC

SEG4 uD_SR105K37 6 4 2 1 9 10 5

8C

CA B C D E F G

DP

CC

SEG3 uD_SR105K37 6 4 2 1 9 10 5

8C

CA B C D E F G

DP

CC

SEG2 uD_SR105K37 6 4 2 1 9 10 5

8C

CA B C D E F G

DP

CC

SEG1 uD_SR105K37 6 4 2 1 9 10 5

8C

CA B C D E F G

DP

CC

LED2 READY12

R2315 1/4W1 2

R2215 1/4W1 2

R2115 1/4W1 2

R2715 1/4W1 2

R2515 1/4W1 2

R2415 1/4W1 2

SEG5 uD_SR105K37 6 4 2 1 9 10 5

8C

CA B C D E F G

DP

CC

U8

TD62304

12345678 9

10111213141516

I1I2I3I4I5I6I7GND VCC

Q7Q6Q5Q4Q3Q2Q1

LED9 M512

LED8 M412

LED7 M312

LED6 M212

LED5 M112

LED4 WAIT12

LED3 X-RAY12

(1B-13)(1B-13)(1B-12)(1B-12)(1B-12)(1B-12)

GREEN YELLOW RED GREEN GREEN GREEN GREEN GREEN

GREENREDGREEN GREEN GREEN

(2F-5)(2G-5)(2H-5)(2J-5)(2K-5)(2L-5)(2N-5)(2O-5)

(1B-12)

(1M-6)

HTP24-CTL-001

Page 32: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

32

CONTROL BOARD OP PANEL & HT CONT.

HTP-24 PORTABLE X-RAY

4 7

A3 0<Cage Code>

Size FSCM NO. DWG NO Rev

Sheetof

APPROVED BY DESIGNED BY

CHECKED BY DATE

+5V

+5V

HS_READY

HS_X-RAY

/FAULTIP_FB_IN

/LASER

ISO_PWR

ISO_GND

RA7

4.7K

2 3 4 5

1

CON1

35312-0410

1234

D35

1N4148

12

PC1

TLP620-2

1

2

8

7

3

4

6

5

R29270

1 2

R30270

1 2

D341N4148

12

D331N4148

12

D36

1N4148

12

CON6

35312-0310

1

2

3

CON2

35312-0210

1

2(1M-7)

(1B-10),(7G-13)(1B-10),(5N-13)

(1B-10)

(1B-9)

HTP24-CTL-001

Page 33: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

33

CONTROL BOARD OP PANEL & HT CONT.

HTP-24 PORTABLE X-RAY

5 7

A3 0<Cage Code>

Size FSCM NO. DWG NO Rev

Sheetof

APPROVED BY DESIGNED BY

CHECKED BY DATE

+12V

-12V

+12V

-12V

-12V

+12V

+12V +12V

-12V

IP_CONTIP_REF

EP_CONTEP_REF

EP_FB

IP_FB_IN

IP_FB

IP_FB_ADC

EP_FB_ADC

/HT_CHECK

IP_FB(+)

EP_FB(+)

C25

BOX223

12

C59 MN1041 2

R36 1K

C30 MN10112

C33MN101

12

C32

BOX224

12

R761K

12

R34 4.7k

U10

VFC32

141

51011

713

124

+IN-IN

1SCCPI

COM

FOUTVOUT

+VS-VS

+

- U11ALM358

3

21

84

TP3

EP_REF

1

TP4

IP_REF

1

+

-

U9B

LM358

5

67

84

+

-

U9A

LM358

3

21

84

D39

1N4148

12

D38

1N4148

12

D37

1N4148

12

+

-

U11B

LM358

5

67

84

R33 39K

C58 MN10412

C27

BOX223

12C26

BOX223

12

C21 BOX3321 2

C24 BOX1031 2

D40

1N41481

2

R4110M

12

R3910K 1%1 2

R4410K 1%

12

R3810K 1%

1 2

R4010K 1%1 2

R4210M

12

VR110K

1 3

2R4310K 1%

12

CON3

35312-0510

1

2

3

4

5

R351/2W 200

12

VR2 2K1 3

2

R84R

R37 100K 1%

+

C22

6.8uF/16V tantal

R32

10K 1%

R31 10K 1%

C31 MN1041 2

C28 MN10412

R75 100K 1%

C62

MN10412

C60 39P12

C29 MN10112

C63

MN10412

40mA(0 ~ 8V)

100KV(0 ~ -5V)

(7B-13)

(1A-10)

(1J-5)

(1B-10),(4J-11)

(6B-14),(7B-11)

(1J-5)

(1F-7)(6B-14) (1F-7)

(6C-5)

HTP24-CTL-001

Page 34: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

34

CONTROL BOARD OP PANEL & HT CONT.

HTP-24 PORTABLE X-RAY

6 7

A3 0<Cage Code>

Size FSCM NO. DWG NO Rev

Sheetof

APPROVED BY DESIGNED BY

CHECKED BY DATE

+12V

+12V

+12V

+12V

-12V

+5V

EP_REF

/HV_ON

EP_FB

FIL_FB

FIL_FB

/PREP

IP_REF

FIL_FB_ADC

/LAMP_ON

/100V_SEL

IP_FB

PN2222 DRV1

EP_CONT1

EP_CONT2

FIL_CONT2FIL_CONT1

LAMP_CONT

R70 470K

1 2

C61 BOX104

12

R78 30K

1 2

R77 5.6K

1 2

R79 5.6K

1 2

R80

1K

12

U17PN2222

21

3

R581K

12

U13B14066

11 10

12

R82 10K

1 2

C36

BOX102

12

R81

3.3K

12

C34BOX682

12

C37 BOX10212

C38

1uF/25V TANTAL

12

U12

TL494

1

2

3

5

6

4

71516

12

89

1110

1314

+IN1

-IN1

F.B.

CT

RT

DT

GND-IN2+IN2

VCC

C1E1

C2E2

OCVR

R47 9101 2

R5247K

12

R532K

12

R516.2K 1%

1 2

R49680

1 2

R48470K

1 2R465.6K

1 2

R455.6K

1 2

R50 9101 2

CON4

35312-1110

12

34

56

78

910

11

U13A14066

1 2

13

U15

TL494

1

2

3

5

6

4

71516

12

89

1110

1314

+IN1

-IN1

F.B.

CT

RT

DT

GND-IN2+IN2

VCC

C1E1

C2E2

OCVR

+

-

U14

OP07

3

26

7 14 8

R57

10K 1%1

2

C43BOX473

12

C42 BOX10212

R5610K 1%

1 2

C39BOX473

12C41MN104

12

C40BOX104

12

R6747K

12

R54750

1 2

R682K

12

R645K 1%

1 2

R6310K

1 2

R62470K

1 2

R595.6K

1 2

R5515K

1 2

R661K

12

R651K

12

U13C14066

4 3

5

VR32K

1 3

2

R604.7K

1 2

R615.6K

1 2

(1M-6)

+

(6K-11)

(5L-4)

(1M-6)

(1J-5)

(1M-6)

(5J-9),(7B-11)(5F-4)

(6B-7)

HTP24-CTL-001

Page 35: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

35

CONTROL BOARD OP PANEL & HT CONT.

HTP-24 PORTABLE X-RAY

7 7

A3 0<Cage Code>

Size FSCM NO. DWG NO Rev

Sheetof

APPROVED BY DESIGNED BY

CHECKED BY DATE

+12V

+5V

+12V

+5V

+12V+12V

-12V

IP_FB

EP_FB

/FAULT

TP10

AGND

1

TP9

+12V

1

TP8

DGND

1

TP7

+5V

1

C47MN104

12

C55MN104

12

C52MN104

12

C44

MN104

1 2

+

-

U16B

LM393

5

67

84

TP6

EP_FB1

TP5

IP_FB

1

C49MN104

12

C48MN104

12

+C54100uF/50V

12

+ C53100uF/50V

12

R71100K

1 2

R72100K

1 2

C50MN104

12

H8

HOLE

1

H7

HOLE

1

H6

HOLE

1

H5

HOLE

1

H4

HOLE

1

H3

HOLE

1

H2

HOLE

1

H1

HOLE

1

+

-

U16A

LM393

3

21

84

CON5

35312-0510

12345

C51MN104

12

H11

HOLE

1

H10

HOLE

1

H9

HOLE

1

R85100k

R86100k

C46BOX103

12

C45BOX473

12

TP11

AGND

1

H12

HOLE

1

+ C2310u/16V

+ C3510u/16V

R69

64K1

2

C57MN104

12

+C56100uF/50V

12

R74

100K

12

R7356K

12

TP12

-12V

1

74HC573 27512 TD62304AP TD62304AP 14066

(5I-15)

(5J-9),(6B-14)

(1B-10),(4J-11)

HTP24-CTL-001

Page 36: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

36

R14 1W100

R3420K

R2927K

S1

DRV2A

12345678

D11 1N4937

TB3

1

C29

BOX102

S

+

C23100/16V

+

C54

TG

T1 PQ32/301

7

2

8

4

3

5

6

10

12

9

R12 NTC10D22

R215W 0.2

C12

224/AC250V

R81/2W470K

Q18PN2222

C42

101/3KV

U5 PC817

34 2

1

+15V

-12V

TP5

-12V 20mA

1

R1310K

D13

1N4937

R4147K

S

+24V

C41M472/100V

C45

104

R17 4.7

HS1

Q21 C2235

SIG1(-)

R241/2W470

Q15IRF640

CON9

35312

12

C44 222/AC250V

TB1

1

S

GND

C4822/25V

Q14PN2222

+24V

CON453258-4

1234

R9

68K 1W

Q11BTA26-600

D10

1N4937

+5V

+ C36

47/25V

C13

224/AC250V

CON3

35312-11

123456789

1011

+

C16

330/400V

C19M104/100V

S

R27 10

Q13

50N60

+24V

R355W/0.1

C32

BOX102

+15V

+

C9

ISO_5V

TP2

NP

1

Z6

1N4746+

C332200/16V(13pie)

D141N4937

R38 10K

GND

C17102/AC250V

+C20 47/25V

+12.2V

TP7 +15V 6A1

U7 UC3844

1

3

57

2

4

6

8

Q19IRF640

CAPACITOR BANK BOARD

R311%9.3K

R2020K

R152W24K

R255.1K

TH

C11

GND

Q20PN2907

R28

1W1K

FIL_FB

S1

DRV2A

12345678

100V_SEL

TP1 +5V 0.5A

1

R193.9K

COLIMETER

C47

1/400V

D9 D10LC20U

C18 100/400V

R22

120

+C22

2200u/25V +

C24

100/16V

C14

R18 220

R26 100

GND

C43104

ISO_GND

U478L05

1

2

3 OUT

GN

D

IN

TP4 +24V 0.2A1

+12.2V

+C35

2200/16V

C8102/AC250V

- +

BD1D25XB60

CON8

35312

12

+

C55L2

SC-10-D050

TB2

1

GND

SIG1(+)

C30

BOX472

F1 50T2A

+ C25100/16V

CON7

35312

12

FET GATE DRIVE CIRCUIT

FIL_CONT

Q17PN2907

+24V

R231W680

+

C10

D12

1N4937

S1

DRV2A

12345678

CON235312-5

12345

+C38100uF/25V

Z5

1N4747

HAND SW PWR

TB4

1

R162W 100K

CON10

35312

12

C31

BOX103

U6TL431

23

1

G

+5V

+15V

PXP-60HF PWR & DRV

1 1

A3 A<Cage Code>

Size FSCM NO. DWG NO Rev

Sheetof

APPROVED BY DESIGNED BY

CHECKED BY DATE

S1

DRV2A

12345678

TC7662A1

U3

1

35

72

46

8

R30

20K

HS3

D15 S2L60

+

C53

330u/400V/30mm*40mm

+5V

SIG2(-)

G

+C342200/16V

TP3

GND

1

PXP60HF-PWR

+

C28TT22/16V

R37 4.7KHS4

D6 1N4937

LAMP_CONT

C37M472/100V

t

R365D11

12

CON153258-05

12345

G

DC- OUT

D8 S2L60

R7

68K 1W

+ C3910uF/25V

12

+

C26TT22/16V

TP4

PG_S

1

U2 LM7805CT1 3

2

IN OUT

GN

D

R391%1.8K

GND

+ C4022/25V

C27

BOX104

C21222/1KV

C46104

+

C51

470/25V

Q16C1815

L3SQE20

DC+ OUT

D71N4937

TP6

PG

1

+

C15

+15V

+24V

R33

1W1K

CON5

35312-3

123

G

Z7

1N4746

Q12IXIF 16N90Q

1

23

+24V

R32 10

CON6

53258-2

12

Page 37: РуководÑтво по Poskom PXP60-HF на английÑком Ñзыке

Product Manual: PXP-60HF Poskom Co., Ltd.

37