dilutions and introduction to spectrophotometry
TRANSCRIPT
-
7/29/2019 Dilutions and Introduction to Spectrophotometry
1/3
Laboratory 2
Dilutions and Introduction to Spectrophotometry
September 15, 2012
-
7/29/2019 Dilutions and Introduction to Spectrophotometry
2/3
I. Introduction
Dilutions are an essential procedure in the microbiology laboratory, whether oneis working with bacterial samples, blood samples, or chemicals, they are used to ensure
that an optimal concentration is used for safety and for the accuracy of the measurements
to be carried out. Dilutions are carried out by taking a small volume of the stock solutionand mixing it with a larger quantity of solvent. The new concentration can then be
calculated by dividing the mass or volume of the solute added by the new volume of the
solution.Beers law provides a useful way to measure the accuracy of a dilution. Using an
instrument known as a photospectrometer that measures the intensity of a particular
wavelength of light passing through a sample, one can calculate the absorbance, which
according to Beers law is linearly related to the concentration of a sample. The equationfor Beers law is:
(1) A = lc
Where A is absorbance, is an absorptivity constant particular to the material, l is thelength of the path that the light travels through the sample, and c is the concentration ofthe sample. (Clark, 2007)
The purpose of this experiment is to practice making dilutions, and to
experimentally confirm Beers law. The hypothesis is that the absorbance of a solutionincreases in a linear trend with increasing concentration, as stated by Beers law.
II. Method
Materials used for this experimentation are: concentrated dye, two 10mL pipettes,
distilled water, and a total of 11 test tubes. The procedure consisted of preparing 11different dilutions of dye in distilled water and measuring their absorbance using aspectrophotometer, then comparing the obtained absorbance with the theoretical value as
predicted by Beers law.
The dilutions were prepared so that the total volume in each test tube was always5mL. Thus the first sample contained 5mL of concentrated dye, the second contained
4mL of dye and 1mL of water, and so on until the last test tube contained 5mL of water.
This test tube containing pure water was used to zero the spectrophotometer (which wasset at a wavelength of 640nm), then each test tube was inserted in turn and the
absorbance was recorded. The theoretical absorbance was calculated by taking the
measured absorbance of the sample of pure concentrated dye and multiplying it times the
percent of concentration.
-
7/29/2019 Dilutions and Introduction to Spectrophotometry
3/3
III. Results
Concentration(% of solute)
TheoreticalAbsorbance
ActualAbsorbance % Error
100 1.460 1.460 0.00
90 1.314 1.315 0.08
80 1.168 1.205 3.17
70 1.022 1.054 3.13
60 0.876 0.936 6.85
50 0.730 0.806 10.41
40 0.584 0.638 9.25
30 0.438 0.520 18.72
20 0.292 0.362 23.97
10 0.146 0.221 51.37
0 0 0.051 -
Table 1: Theoretical and measured absorbances for each concentration. Note the increasing percentage
error as concentration decreases.
Beer's Law
y = 0.0139x + 0.0827
R2 = 0.9985
0.000
0.200
0.400
0.600
0.800
1.000
1.200
1.400
1.600
0 10 20 30 40 50 60 70 80 90 100
Concentration (%)
Absorbance
Graph 1: Plot of the experimental values for absorbance versus the concentration of the samples,
showing a strong linear correlation, as shown by the equation and R2 value.
IV. Summary and Conclusions
The linear trend obtained from plotting absorbance versus concentration coincideswith that predicted by Beers law, supporting the hypothesis. However, the measured
results did show an increasingly significant deviation from the predicted values as the
concentration of solute became smaller, probably due to experimental or systematic error.
V. References:
1. Clark, J. (2007). The Beer-Lambert law. Retrieved from Chemguide website:
http://www.chemguide.co.uk/analysis/uvvisible/beerlambert.html