development of multi channeldosimeter system foraesj.net/hp/pnst/04/081.pdf · journal of nuclear...

4
Journalof NUCLEARSCIENCE andTECHNOLOGY Supplement 4 p.318-321 (IVl arch20(4) Development of Multi ChannelDosimeter System for Measurement of Dose Disfributi n in Therapeutic Photon Beams Jh inKee K 臥r! 1 3' Bu Gil KIM 2 Jung Soo KIM 1,3, Hyeong Cheol KWON 1 JungHong KIM 4 Jung Ku KANG 5 Young Kee OH 6 Kyo Chul SHIN 7 Jeung Kee KIM 8 Ki Hwan KIM 9 Dong Hyeok JEONG 10 'Research Institute ofClinical Medicine Chonbuk National University Jeorifu 2 Biomedical Chonbuk National University Jeonju 43Radiation OncoloD chonbukNational UniversiOY Hospitaf JEOFdu Physics Chonbuk National University Jeonju 5 Radiation Oncology Chonju Jesus Hospital Jeonju 6BiomedicalEngineering YeungnamUniversity Tae 7 Radiation Oncolo g y Dankook Universi Hospital Cheonan 8Radiation Oncology Dong-AUniversi Hospital Pusan 9Radiation Oncolo g y ChungnamUniversity Hospital Taejon 1U Radiation Oncology Won 'angUniv. Hospital san Korea We investigated the characteristicsof the multi channeldosimeter (MCD)systemfordosedis ibutionby the moving wedge in clinical photon beams. A1 so we developed an elec ometerfor the multi-channel dosimeter system for radiation therapy. 1n dynamic field irradiation ,也evirtual wedge technique and its actionmethods are available through thecomputer controlledasynnnetricindependent collimator.lemodifiedMCDsystem consistsof an electrometer a solid detector and 阻即ayphanto m. The MCD is used for the point dose measurement and the field sizescanning.Thus we evaluated the dosim iccharacteristicsof thevirtual wedgeand the conventionalfixed wedge by the radiation field analyzer (RFA). In our analysis for the 6 d10 MV photon beams the maximum dose in levirtual-wedge d lefixed-wedge factors varied om1. 2 % to 1. 6 % forsqu ecollimator setting rangingom10to20cm respectivel Wecons uctedlemodifiedMCDsystemwiththedeveloped electrometer for radiation therapy to improve on the analytical expressions describing . edose distributions for le virtual wedge and the wedge actionmethod in 6 d10 MV clinical photon beams. KEYWORD S: riable 面砂 pewedge jilte r wedge angle multi-channel detector system 1 . Introduction Themaintargetofradiotherapyisto increasethe cure rate by m dicatingplenty dosein focused region to be treated. An d the final purpose is to minimize the dose fornormaltissue andtheside effectoccurredduring eatmen t. R cen t1 y the intensity modulated radio- therapy (I MRT)methodand 3Dconformal therapy have been developed to achieve these purposesY) These treatment methods are to make irradiation field along the shapeoffocusedregionomseveraldirectionsandto medicate radiation dose for each direction. Wh en we use this3D conformal radiation therapy we need to regulate thedosedis ibutionaccordingtoshapesof irradiation surface andtumor in which we canpartially apply methodsωusedosecompensatororn 01dose of radiation andmulti-leafcollimator or beam intensity modu 1a ted method.Anoth rspecialcomp nsatorcan be appliedto give a changein isodos distribution of radiation path being situated. The commercial compensatorof wedgeformcould beusedasaspecial compensation device. During irradiationweused the special coverwith andwithout such commercialized generalwedge.Inaddition wecouldchangethesome specialcompensationeffectand theisodosedistribution bycon ollingtheirradiationtime. In ordertoapply *Correspondingauthor Te l .82 -63-250 1184 Fax.82-63- 250 1192 E-mail: kmjnki@cho l. com 318 such effect the techniques contro l1i ng the operating time ofasymme y collimatorthroughcomputerhavebeen .l 3-5) usea Among these techniqu s the virtual wedge filter techniquegives theoreticallyachangeofthe isodose distribution curvebyasymmetryco l1i mator. Sincethe radiation dosedistributionintissue is di :ff erentfrom those of the fixed wedge there is no decreasing phenomenon of dose and the scattered dose becomes less. Also itisreportedforthevirtualwedgetobemore us 巴血1thantheexistingfixedwedgetomakea relative direct wedge angleregardless ofdepth. 6 - 8 ) We have developed an auxiliary device for such elaborated radiationtherapymethod inwhichtheexactradiation dose should be irradiated in tumor r gionin radiotherapy. The dose measurement method was repeatedly to measurethe dose forspecificpositionbythe single measuringdevice. Therefore thebulky m asurementfor allirradiationfieldsizesshouldbeaccompaniedtoget the doseinformationin irradiation eld.Therewere overloads ofmachine by long use manpower and energy waste. An dthe radiation dosimetry was quitedi cultin eof .additional device such a variable-type wedge fi 1t er for therapeutic photon beams. 則的 Because of the above problems andnecessity of altemative method we needed researches for the d lamicfield and the MCD system. Variable typewedge fi 1t er is a tool being attached to medicallinear accelerator. We want dto improve the dosimeter for wedge filter and to supplement the fixed wedge filter function and wished

Upload: others

Post on 11-Jul-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Development of Multi ChannelDosimeter System foraesj.net/hp/pnst/04/081.pdf · Journal of NUCLEAR SCIENCE and TECHNOLOGY, Supplement 4 ぅp. 318-321 (IVl arch 20(4) Development

Journal of NUCLEAR SCIENCE and TECHNOLOGY, Supplement 4ぅp.318-321 (IVlarch 20(4)

Development of Multi国 ChannelDosimeter System for Measurement of Dose Disfributi⑪n in Therapeutic Photon Beams

JhinKee K臥r!1,3',Bu Gil KIM2, Jung Soo KIM1,3, Hyeong Cheol KWON1べJungHong KIM4, Jung Ku KANG5

Young Kee OH6, Kyo Chul SHIN7, Jeung Kee KIM8, Ki Hwan KIM9, Dong Hyeok JEONG10

'Research Institute ofClinical Medicine, Chonbuk National University, Jeorifu,

2 Biomedical Engìneer初~g, Chonbuk National University, Jeonju,

43Radiation OncoloD,chonbukNational UniversiOY Hospitaf,JEOFdu,

Physics, Chonbuk National University, Jeonju, 5 Radiation Oncology, Chonju Jesus Hospital, Jeonju,

6BiomedicalEngineering, Yeungnam University, Tae,伊 ,7Radiation Oncology, Dankook UniversiかHospital,Cheonan,

8Radiation Oncology, Dong-A Universi砂Hospital,Pusan,9Radiation Oncology, Chungnam University Hospital, Taejon,

1URadiation Oncology, Won伽 'angUniv. Hospital,及san,Korea

We investigated the characteristics of the multi四channeldosimeter (MCD) system for dose dis仕ibutionby the moving wedge in clinical photon beams. A1so we developed an elec甘ometerfor the multi-channel dosimeter system for radiation therapy. 1n dynamic field irradiation,也evirtual wedge technique and its企actionmethods are available through the computer田controlledasynnnetric independent collimator.百lemodified MCD system consists of an electrometer, a solid detector and阻即ayphantom. The MCD is used for the point dose measurement and the field size scanning. Thus, we evaluated the dosim甜 iccharacteristics of the virtual wedge and the conventional fixed wedge by the radiation field analyzer (RFA). In our analysis for the 6回 d10 MV photon beams, the maximum dose inせlevirtual-wedge叩 d吐lefixed-wedge factors varied仕om1.2 % to 1.6 % for squ紅 ecollimator setting ranging企om10 to 20 cm, respectivelぁ Wecons仕ucted仕lemodified MCD system with the developed electrometer for radiation therapy to improve on the analytical expressions describing白.edose distributions for仕levirtual wedge and the wedge企actionmethod in 6阻 d10 MV clinical photon beams.

KEYWORDS: 拘 riable面砂pewedge jilter, wedge angle, multi-channel detector system

1 . Introduction

The main target of radiotherapy is to increase the

cure rate by m巴dicatingplenty dose in focused region to

be treated. And the final purpose is to minimize the dose

for normal tissue and the side effect occurred during 仕eatment. R巴cent1y,the intensity modulated radio-

therapy (IMRT) method and 3D conformal therapy have been developed to achieve these purposesY) These

treatment methods are to make irradiation field along the

shape of focused region企omseveral directions and to

medicate radiation dose for each direction. When we use this 3D conformal radiation therapy, we need to regulate the dose dis仕ibutionaccording to shapes of irradiation

surface and tumor, in which we can partially apply

methodsωuse dose compensator or∞n仕01dose of

radiation and multi-leaf collimator or beam intensity

modu1ated method. Anoth巴rspecial comp巴nsatorcan be

applied to give a change in isodos巴 distributionof radiation path being situated. The commercial

compensator of wedge form could be used as a special compensation device. During irradiation we used the

special cover with and without such commercialized

general wedge. In addition, we could change the some special compensation effect and the isodose distribution

by con位ollingthe irradiation time. In order to apply

*Corresponding author, Tel. 82・-63-250・1184,Fax. 82-63-250司 1192,E-mail: [email protected]

318

such effect, the techniques control1ing the operating time of asymme位ycollimator through computer have been

...l 3-5) usea

Among these techniqu巴s,the virtual wedge filter technique gives theoretically a change of the isodose

distribution curve by asymmetry col1imator. Since the radiation dose distribution in tissue is di:fferent from those of the fixed wedge, there is no decreasing

phenomenon of dose, and the scattered dose becomes less. Alsoラ itis reported for the virtual wedge to be more us巴血1than the existing fixed wedge to make a relative

direct wedge angle regardless of depth.6-8) We have

developed an auxiliary device for such elaborated

radiation therapy method, in which the exact radiation dose should be irradiated in tumor r巴gionin radiotherapy.

The dose measurement method was repeatedly to

measure the dose for specific position by the single

measuring device. Thereforeラthebulky m巴asurementfor

all irradiation field sizes should be accompanied to get

the dose information in irradiation白eld.There were

overloads ofmachine by long幽 use,manpower and energy waste. And the radiation dosimetry was quite di伍 cultin

田 eof .additional device such a variable-type wedge fi1ter for therapeutic photon beams.則的

Because of the above problems and necessity of altemative methodラ we needed researches for the

d戸lamicfield and the MCD system. Variable同 typewedge fi1ter is a tool being attached to medicallinear accelerator.

We want巴dto improve the dosimeter for wedge filter and

to supplement the fixed wedge filter function, and wished

Page 2: Development of Multi ChannelDosimeter System foraesj.net/hp/pnst/04/081.pdf · Journal of NUCLEAR SCIENCE and TECHNOLOGY, Supplement 4 ぅp. 318-321 (IVl arch 20(4) Development

to accumulate some basic data for development of domestic instruments instead of the foreign expensive equipments. Therefore, we developed partially the MCD, in which the d巴viceswere modified applicable to dynamic wedge method.

II.l¥置aterialsa:nd Methods.

1. Specific property of dosimetry system

Diode detector, ion ionization chamber and thermoluminescence dosimeter (TLD) are being田 edto measure the radiation dose reaching in tumor volume in a phantom or in humari body. In the ionization chamber, we measured by using electrometer the accumulated electric charge gen巴ratedby the irradiated radiation. It was converted to the radiation dos巴bymeans of calibration factor such as temperature and atmospheric pressure. Since the semiconductor detector has usually small volume size with higher space resolution, and betier radiation sensitivity than those ofionization chamber, and thus it is commonly田 edfor the dosimetry ofphoton and electron beams.ll-13)

Since the measurement of dose distribution of buildup region shou1d have high space l'巴solution,we wanted to obtain repeatability for measur巴mentby using a suitab1e semiconductor detector (Scanditronix p-Si detector, Sweden). Our external diode detectors was available for photon beams as follows: detector material is p-type Si1icon, chip thickness of 0.05 mm, effective det巴ctionpoint of 1.5 mm, axia1 directional depend四回 within2%,maximum deviation from linearity in the dose range 0.1 to 0.6 mGy within 1 % dose p巴rpulse, sensitivity variation with temperature 0.4% per degr巴巴 centigrade.We have田 edthe reference ionization chamber (PTW-

W30004, 0.6 m1, PTW, Germany) as standard dosimeter for the absorbed dose verification.

2. Composed detector system and signal transaction

For the dose detector syst巴mwith multi-channels, the basic amplification circuit of detector output h田 beenused by taking an advantage of the mu1tip1ex basic technology. First, we amplified the output of detector, and converted the detected data to eight-bit analog digital (A-D) converter, and applied the data processing program for eight channels, so that the simultaneous measurement could be available selectively. Because of very small detector output, after multiplying 100 times by using current-voltage amplification, we changed the small current to voltage. Basically, it is current圃vo1tageconversion circuit, in which the current f10wing into the input terminal of reversal amplification passes through the feedback resistance. Since we ne巴deda high resistor with good stability, we put a potentiometer into the feedback loop to obtain the high amplification rate, and could mak巴ahigh equivalent resistor. Because the drift

SUPPLEMENT 4う MARCH2004

319

current is also magnified together,ラ weus巴dthe 10w“drift OP amplifier (0トAMP).Resistance value of voltage dividing circuit must be selected carefully since its error is smaller than that of the feedback resistance. Since an OP-AMP must use small input bi田 current,the field effect仕組sistorOP-AMP (3528BM of Barbrown Co・9

Ltd) are used, and the 0宜setvoltage quantity by bias current could be col1ected by making a norトreversalinput. The A-D converter with a ADC0809 (National semiconductor Co.) was used, which is combined with a multiplex of 8 channels in input department. A data processing program of the system was developed using Delphi language, and the module controlling printer port was m:ade with an assembler. The data of system were collect巴dby printer port of computer. After selecting channel, we latched and inf1icted conversion-opening signal of A-D converter. The end of A-D converter was done by inputting A圃 Dconvert巴routput of 3 stat巴bufferoutputs by detecting the end・of-channel(EOC) signal of A-D converter. WI巴madethe data processing program to select each channe1 separatively, and analyzed the radiation dose, the dose rate, the percent-depth-dose (PDD) and so on, which are used commonly in radiation therapy. Also, we could store the measurement data into text file, and used it as beam data in treatment p1an computer.

3. Measurement of dos巴distributionfor wedge field

The depth dose distribution for 6 and 10 MV photon beams was measured and compared with each other by using standard measur巴mentsystem (pTW・UNIDOS-W30004, PTW, Germany) in medical linear accelerator. We constructed th巴 prototypesystem consisted of the multi胸 channeldosimeter system and p-Si semi“conductor detector (with sensitive volume of 0.25 ml, effective m巴asurementpoint of 0.54 mm), and manufactured the auxi1iary device with a f1at style phantom for applying to the commercialized scanning system侭FA・7,Sweden). The scanning system田 edin measurement is consisted of a シD water phantom (with size of 51x51x51cm3), a

dosimetry regulator and a remote陶 controlsystem for positioning the detector. We measured the depth dose distribution for standard

field size about each photon beamsヲ andmeasured the beam f1atness to beam axis in 1.5, 5, 10, 20 cm depths. We also ca1culated the wedge angles企omthe beam f1atness fo1' variable-type wedge fi1ter in 10 cm depth,

and compared them with the fixeιwedge angles.

m. Results and Discussion

For the measurement in the radiation dose distribution of cancer therapy by 6 and 10 MV photon beams, and for adminis仕組onof the various treatment methods, we investigated the direction and mobility of instrumentation system, th巴 beamf1atness and the adaptability of dos巴

Page 3: Development of Multi ChannelDosimeter System foraesj.net/hp/pnst/04/081.pdf · Journal of NUCLEAR SCIENCE and TECHNOLOGY, Supplement 4 ぅp. 318-321 (IVl arch 20(4) Development

320

dis仕ibutionmeasurement by the signal process of multi-channel AlD conversion method. We amplified the output of detector, and converted the detected data into the digital data by 8・bitanalog-digital converter (ADC80 17), and applied data processing program for eight channels so that the simultan巴ous measurement is available selectively for data processing. In order to apply the multi-channel dosimeter system to adaptabi1ity of administration application for radiation therapy system, we manufactured, as shown in Figs.l,2, an adapter for the applicable detector arrangement in RFA 3・Dphantom and the flat-type 2-D phantom, which are used in measurement. A dependence on direction of 0町semiconductor detector EDD・5 (Scandi仕onix,Commercial, Hi予Si)with six channels was within 3%, 4% for 砿 is and for direction of 450 tilt angle, respectively. The detector impedance was more than 350

MQ in 0 V. The measurement data was reproduced

within 2.1 %, as shown in Fig.3, which indicate the dose profiles and the beam flatn回 sfor 6 MV medical photon beam. It was ca1cu1ated in the difference between measured data of standard reference chamber and measured data ofmulti-channel detector. Among these techniques, the virtual wedge filter

technique gives theoretical1y a change of the isodose dis仕ibutioncurve by asymmetry co1limator. Since the radiation dose distribution in tissue is different from those with the fixed wedge fi1tersラ th巴reis no decreasing phenomenon of dose, and the scattered dose become less. Also, the virtual wedge is more useful than the existing fixed wedge to make a wedge angle regardless of depth, as indicated in Figs.4, and Table 1

Fig. 1. QA Tools幽Arrayadapter fo1' scanning system

Fig. 2. QA Tools問 Plateadapter for dose measurement

Relaiive Dose (%)

180

160 一+一 d1.5 cm

14日 --a-d 5 cm 120

100 一合一d10 cm

畠日 ---d 20 cm

昔日

40

20 。。 2 4 6 8 10 12

Off岡田isdistance (cm)

Fig. 3. Measured beam profile for 6 MV photon beam in field size 10xl0 cm2.

Relative Dose (%)

140

120

100

80 。げ+--EI-W ellge 15 40 ー←Wedge30 20 一持一Wedge45

→←Wedge 60 。-80 -40

。40 80

Off-aXIs distance (mm)

Fig. 4. Measured beam profile for 6 MV photon beam in wedge field sIze lOxlO cm2•

It is necessary to define correct1y the tumor region in human body for patient's radiotherapy, and to maximize the cure effect by means of exact amount of radiation dose at the planed region, and also to minimize the side effect by minimizing the radiation dose that is irradiated in normal region. In this case, the isodose distribution in therapy region is much deficient, because ofthe radiation diminution and the abso中tion property by the inhomogeneous body structure, the indeterminate form of tumor shape and limit of radiotherapy method. Actual1y, the randomicity of radiation dose distribution in tumor occurred by these causes is now permitted within 5% change. Except the above elements, an uncertainty of 由巳 radiationdose is increased by inaccuracy of radiation output of radiotherapy equipment, limit of data, dose measurement e町orand the dose calculation error. By these errors, the radiation dos巴thatis irradiated in tumor in radiation treatment is permitted within 3% error from tumor dose.

JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY

Page 4: Development of Multi ChannelDosimeter System foraesj.net/hp/pnst/04/081.pdf · Journal of NUCLEAR SCIENCE and TECHNOLOGY, Supplement 4 ぅp. 318-321 (IVl arch 20(4) Development

Table 1. A comparison of virtual wedge angle measured at 100 cm

SAD, dep仕110cm.

1ljrtual Wedge Augle

150

300

450

600

6MV

15.800

30.030

45.030

61.20。

The program to minimize these errors is applied in

each hospital unit, and also the quality assurance (QA) programs that 巴nforce administration locally and

nationally or interτlationally to confirm almost impossible

systematic error within hospital have been developed by

the Intemational Atomic Energy Agency (IAEA) and

several institutions, and been applied. However, these QA programs put an essential point in syst巴matic巴町orthat

can produce in irradiation to patient. There is danger that the excess or insufficient radiation dose is irradiated

by output fluctuation or miscontrol of radiotherapy equipments, which can be happen in ac同altreatment. To confirm occurring possibility of the various

predictable errors and to irradiate the exact dose, w巴need

the correct dosimetry and the QA adminis仕ationwith

treatrnent method. Therefore, the maximum field size at radiation treatment is usually 40 x 40 cm2.

We need to expand the nurnber of channels of detector

system to 64 channels in order to improve accuracy of

dosimetry including all radiation field sizes which is田 ed

in the actual patient treatment by multi-channel

dosimetry system. If one wants to use our multi-channel

dosimeter system as the common measuring instrument, he should use the conf・idenceamplification circuit to low drift by using the precision parts and through a circuit

improvement, and to low noise in the amplification part ofdet巴ctoroutput to improve accuracy of measuring ins住ument.The detector arrangement for

measurement of the dynamic wedge m巴thodshould be

done by 1 cm within. Our scanning measurement is

available by 0.5 cm within using by moving adapter.

In this researchラ weconstructed an administration

syst巴m about radiation therapy, dosimetry system of variable-type wedg巴出ter,multi-channel dosimetry

system, detector-signal amplification technology and data processing algorithm, which are considered to contribute to the radiometry equipment and the radiation QA

research, and hereafter, our res巴archis assumed to be

applied to development of peripheral devices for radiation therapy.

N. Conclusions

8ince the change of dose distribution is occurred by

SUPPLEMENT 4う MARCH2004

321

as戸nme仕y collimator, there is no decreasing

phenomenon in the internal radiation dose distribution, unlike with the fixed wedge filter. These resu1ts are

considered to be very田 e:fulin applying to the irradiation methods for the relative linear wedge angle, other various wedge angles, method ofmixing the企action,sndso on. According to the arbitrary special shielding effect to be

wanted, we. could give the change of radiation dose distribution with and without the universal fixed wedge during irradiation and by regulating the irradiation time

and by control of the operation time for asymmetry

collimator.

To apply these effects, we investigated the dose distribution and the wedge angles through the measuring

system composed of variable-type wedge filter. 80 we

developed partially MCDラ anddevices are modified to be applicable to dynamic wedge method.

References 1) Convery DJ, Rosenbloom ME.,“The generation of intensity帽

modulated fields for conformal radiotherapy by dynamic collimation,円Phys.Med. Biol., 37, 1359 (1992).

2) Khan FM, Potish RA. Treatment Planning in Radiation Oncology. Williarns & Wilkins, 187 (1998)

3) Klein EE, Low DA, Meigooni A., Purdy JA., "Dosimetry and clinical implementation of dynamic wedge," 1nt. J. Radiat. Oncol. Biol. Phys., 31, 583 (1995).

4) Lydon JM. Rykers KL.,“Beam profiles in the nonw巴:dgeddirection for dynamic wedges," Phys. Med. Biol., 41, 1217 (1996).

5) Leavitt DD, Martin M, Moeller JH, Lee WL.,“Dynamic wedge field techniques computer controlled collimator motion and dose delivery," Med. Phys., 17,87 (1990).

6) Leavitt DD, Larsson L.,“Evaluation of a diode detector array for measurement of dynamic wedge dose distribution," Med. Phys., 20, 381 (1993).

7) Zwicker RD, Shahabi S, Wu A, Sternick ES.,“Effective wedge angles for 6MV wedges 企ommedical linear accelerators," Med. Phys., 12,347 (1985).

8) Wil1iamson JF, Kh阻 FM,Sharma SC.,“Film dosime町 ofmegavo1tage photons bea立国岨Apractical method of isodensity to isodese curve conversion, Med. Phys., 8,94 (1981)

9) Kijewski PK, Chin LM,町amgardBE.,“Wedge四 shapeddose dis仕ibutionsby computer-controlled collimator motion," Med. Phys., 5, 426 (1987).

10) Cross P, Joseph DJ, Cant J, Cooper SG, D巴nhamJW., “Tangential breast irradiation: simple improvements," Int. J.

Radiat. Oncol. Biol. Phys., 23, 433 (1992) 11) Grusell, E. Rikner, G.,“Evaluation oftemperature effects in P勾rpesilicon detectors," Phys. Med. Biol., 31, 527 (1986).

12) Rikner, G., G即日ell,E., "General specifications for silicon semiconductors for use in radiation dos即時句;"Phys‘Med. Biol., 32, 1109 (1987).

13) Slessinger ED, Gerber RL, Harms WB, Klein EE, Purdy JA.,“Independent col1imator dosimetry for a dual photon energy linear accelerator," Int. J. Radiat. Oncol. Biol. Phys., 27,681 (1993).