development of an automated wireless tension force estimation

16
Development of an Automated Wireless Tension Force Estimation System for Cable-stayed Bridges SOOJIN CHO, 1, *JEROME P. LYNCH, 2 JONG-JAE LEE 3 AND CHUNG-BANG YUN 1 1 Department of Civil and Environmental Engineering, KAIST, 373-1 Guseong-dong, Yuseong-gu, Daejeon, 305-701, South Korea 2 Department of Civil and Environmental Engineering, University of Michigan, 2380 G. G. Brown Building Ann Arbor, MI 48109-2125, USA 3 Department of Civil and Environmental Engineering, Sejong University, 98 Gunja-dong, Gwangjin-gu Seoul, 143-747, South Korea ABSTRACT: Cable-supported bridges rely on the use of steel cables to support the bridge deck and load on it. Cable tension forces are monitored during construction to assist the alignment of cables and to ensure no cables are overloaded. Given that the cables are critical load carrying elements, it is prudent to routinely monitor the levels of cable tensions during operation. With current measurement methods being costly and labor-intensive, this article proposes an automated and low-cost wireless sensor system for continuous monitoring of the cable tension based on the vibration signature of the cable. A vibration-based tension force estimation method using a peak picking algorithm is explored by embedding it in the computational core of a wireless sensor. Welch’s method to average Fourier spectra from the segments of a long time history signal is employed to remove the non-stationarity of a short- duration acceleration record, which is a limit of the memory-constrained wireless sensor. A series of laboratory tests are conducted on a slender braided steel cable with a variety of cable sags and tension forces. Excellent agreements have been found between the actual tensions and those estimated by the present wireless system. Key Words: cable tension, wireless sensor, modal analysis, Welch’s method, embedded software. INTRODUCTION C ABLE-STAYED bridges have emerged as economical design choices compared to other long-span bridge concepts including suspension bridges. In general, cable- stayed bridges consist of one or more tall pylons from which steel cables are anchored to support the bridge deck. Recently, a number of impressive cable-stayed bridges have been constructed including Tatara Bridge (Japan) and Sutong Bridge (China) with free spans of 890 and 1088 m, respectively (Yanaka and Kitagawa, 2002; Hua et al., 2005). Monitoring cable tensions during construction of cable-stayed bridges is necessary to align cables properly and to ensure no cable is over- loaded. With cables serving as the primary vertical load carrying elements of the bridge, there is a need to ensure the structural integrity of the cables well after the com- pletion of the bridge (Morgenthal, 1999). Furthermore, small variations in cable tension may cause a dramatic affect on the global response of other parts of the bridge including the deck and pylon. For example, Casas (1994) reports significant variations in pylon bending moments due to small (<10%) variations in the cable tension of a cable-stayed bridge with an inclined pylon. Given the importance of cables for the global integrity of a cable-stayed bridge, continuous monitoring of cables for degradation or anchorage slippage is prudent. Many methods are available for measuring the tension in a bridge cable. During construction, cable tension is closely monitored using load cells or pressure meters at one end of the cable (Wang et al., 1999; Kim and Park, 2007). Another approach is by measuring the elongation as a hydraulic jack is tensioning the cable (Casas, 1994). After construction, direct measurement of cable tensile stress can be made using elastomagnetic (EM) stress sen- sors (Wang et al., 2005). However, one difficulty asso- ciated with EM sensors is the need to wrap the EM sensor solenoid around the cable which can be laborious and costly. Indirect methods of cable tension measure- ment based on the dynamic properties of a cable have been widely used (Zui et al., 1996; Russell and Lardner, 1998). This approach is easy to employ and cost-effective because it requires only an accelerometer to measure the vibration of the cable during ambient loading. *Author to whom correspondence should be addressed. E-mail: [email protected] Figures 13 and 5–13 appear in color online: http://jim.sagepub.com JOURNAL OF INTELLIGENT MATERIAL SYSTEMS AND STRUCTURES, Vol. 21—February 2010 361 1045-389X/10/03 036116 $10.00/0 DOI: 10.1177/1045389X09350719 ß The Author(s), 2010. Reprints and permissions: http://www.sagepub.co.uk/journalsPermissions.nav

Upload: vudan

Post on 01-Jan-2017

218 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Development of an Automated Wireless Tension Force Estimation

Development of an Automated Wireless Tension ForceEstimation System for Cable-stayed Bridges

SOOJIN CHO,1,* JEROME P. LYNCH,2 JONG-JAE LEE3AND CHUNG-BANG YUN

1

1Department of Civil and Environmental Engineering, KAIST, 373-1 Guseong-dong, Yuseong-gu, Daejeon, 305-701, South Korea

2Department of Civil and Environmental Engineering, University of Michigan, 2380 G. G. Brown BuildingAnn Arbor, MI 48109-2125, USA

3Department of Civil and Environmental Engineering, Sejong University, 98 Gunja-dong, Gwangjin-guSeoul, 143-747, South Korea

ABSTRACT: Cable-supported bridges rely on the use of steel cables to support the bridgedeck and load on it. Cable tension forces are monitored during construction to assist thealignment of cables and to ensure no cables are overloaded. Given that the cables are criticalload carrying elements, it is prudent to routinely monitor the levels of cable tensions duringoperation. With current measurement methods being costly and labor-intensive, this articleproposes an automated and low-cost wireless sensor system for continuous monitoring ofthe cable tension based on the vibration signature of the cable. A vibration-based tensionforce estimation method using a peak picking algorithm is explored by embedding it in thecomputational core of a wireless sensor. Welch’s method to average Fourier spectra from thesegments of a long time history signal is employed to remove the non-stationarity of a short-duration acceleration record, which is a limit of the memory-constrained wireless sensor.A series of laboratory tests are conducted on a slender braided steel cable with a variety ofcable sags and tension forces. Excellent agreements have been found between the actualtensions and those estimated by the present wireless system.

Key Words: cable tension, wireless sensor, modal analysis, Welch’s method, embedded software.

INTRODUCTION

CABLE-STAYED bridges have emerged as economicaldesign choices compared to other long-span bridge

concepts including suspension bridges. In general, cable-stayed bridges consist of one or more tall pylons fromwhich steel cables are anchored to support the bridgedeck. Recently, a number of impressive cable-stayedbridges have been constructed including Tatara Bridge(Japan) and Sutong Bridge (China) with free spans of890 and 1088m, respectively (Yanaka and Kitagawa,2002; Hua et al., 2005). Monitoring cable tensionsduring construction of cable-stayed bridges is necessaryto align cables properly and to ensure no cable is over-loaded. With cables serving as the primary vertical loadcarrying elements of the bridge, there is a need to ensurethe structural integrity of the cables well after the com-pletion of the bridge (Morgenthal, 1999). Furthermore,small variations in cable tension may cause a dramaticaffect on the global response of other parts of the bridge

including the deck and pylon. For example, Casas (1994)reports significant variations in pylon bending momentsdue to small (<10%) variations in the cable tension of acable-stayed bridge with an inclined pylon. Given theimportance of cables for the global integrity of acable-stayed bridge, continuous monitoring of cablesfor degradation or anchorage slippage is prudent.

Many methods are available for measuring the tensionin a bridge cable. During construction, cable tension isclosely monitored using load cells or pressure meters atone end of the cable (Wang et al., 1999; Kim and Park,2007). Another approach is by measuring the elongationas a hydraulic jack is tensioning the cable (Casas, 1994).After construction, direct measurement of cable tensilestress can be made using elastomagnetic (EM) stress sen-sors (Wang et al., 2005). However, one difficulty asso-ciated with EM sensors is the need to wrap the EMsensor solenoid around the cable which can be laboriousand costly. Indirect methods of cable tension measure-ment based on the dynamic properties of a cable havebeen widely used (Zui et al., 1996; Russell and Lardner,1998). This approach is easy to employ and cost-effectivebecause it requires only an accelerometer to measure thevibration of the cable during ambient loading.

*Author to whom correspondence should be addressed.E-mail: [email protected] 1�3 and 5–13 appear in color online: http://jim.sagepub.com

JOURNAL OF INTELLIGENT MATERIAL SYSTEMS AND STRUCTURES, Vol. 21—February 2010 361

1045-389X/10/03 0361�16 $10.00/0 DOI: 10.1177/1045389X09350719� The Author(s), 2010. Reprints and permissions:http://www.sagepub.co.uk/journalsPermissions.nav

Page 2: Development of an Automated Wireless Tension Force Estimation

In a cable-stayed bridge, stay cables are anchored inan inclined orientation at their two ends (i.e., at the deckand pylon). If the effect of the sag and flexural rigidity ofthe cable is neglected, the cable tension can be estimatedbased on the taut string theory using the natural fre-quencies of the cable (Irvine, 1981; Kim and Park,2007). However, the taut string theory may give ser-iously inaccurate estimates, if the effect of the sag andflexural rigidity of a cable is significant. In response tothis limitation, Zui et al. (1996) has proposed a set ofpractical formulas that estimate cable tension using thefirst and second modal frequencies of an instrumentedcable. This set of equations has been proven accurate(i.e., within 4%) in estimating the tensions in the staycables of Seohae Bridge in Korea (Kim et al., 2007). Inthis study, the tension estimation equations proposed byZui et al. are adopted for inclusion in an automatedwireless tension monitoring system for cable-stayedbridges.Accelerometers are generally integrated into a perma-

nent structural monitoring system in which shieldedcoaxial wiring is used for data transmission.Unfortunately, a large portion of the cost of installingmonitoring systems is attributed to the installation of itsextensive amount of wiring (Straser and Kiremidjian,1998). As a result, monitoring systems contain fewersensors than desired. For example, the cost of installingaccelerometers on every stay cable in a cable-stayedbridge would be prohibitive; therefore, a monitoringsystem might only instrument a handful of cables. Toreduce the cost and complexity of installing a structuralmonitoring system, wireless sensors have been proposedas a low-cost alternative to the traditional tethered sen-sors (Straser and Kiremidjian, 1998; Spencer et al., 2004;Lynch and Loh, 2006). Wireless sensors have been suc-cessfully tested on a variety of short- and long-spanbridges revealing their accurate and reliable perfor-mance (Lynch et al., 2004a; Lynch et al., 2006;Gangone et al., 2007; Pakzad et al., 2008). For cable-stayed bridges, wireless sensors would be an ideal sen-sing solution for continuously monitoring the staycables at low-cost. Furthermore, the computationalresource included in the design of most wireless sensorscan be leveraged to locally process acceleration data toestimate cable tension using some of the aforementionedindirect estimation methods.In this study, the tension estimation method proposed

by Zui et al. (1996) is modified for embedment in a wire-less sensor. The wireless sensor has been designed expli-citly for structural monitoring applications to offer high-resolution data collection from sensors, local computingand memory resources, and long-range communication.To monitor the vibration response of a stay cable, a low-noise microelectromechanical system (MEMS) acceler-ometer is interfaced to the wireless sensor node.Embedded within the computational core of the wireless

sensor are modal analysis algorithms to convert thetime-history accelerations to the frequency domainwhere modal frequencies can be accurately extracted.Using modal frequencies and a priori geometric infor-mation of the instrumented bridge cable, the set of ten-sion estimation equations proposed by Zui et al. areexecuted to calculate the cable tension. A set of labora-tory experiments are conducted to validate the wirelesssensor-based cable tension estimation system. Aninclined steel cable instrumented with MEMS acceler-ometers and wireless sensors is exposed to broad-bandexcitations. The wireless sensors record the cableresponse and autonomously execute embedded algo-rithms that extract cable frequencies and estimatecable tension. The article concludes with a summary ofthe salient research findings and suggestions for futurework.

AUTOMATED WIRELESS SENSORS FOR

MONITORING STAY CABLES

Wireless sensors for structural health monitoring werefirst proposed by Straser and Kiremidjian (1998). Sincetheir seminal study, a large number of academic (Straserand Kiremidjian, 1998; Lynch et al., 2004a; Wang et al.,2007) and commercial (Spencer et al., 2005; Gangoneet al., 2007; Pakzad et al., 2008) wireless sensor proto-types have been proposed for monitoring civil struc-tures. While differences can be observed among thedifferent wireless sensors proposed, conceptually theyshare similar architectural designs. The design of wire-less sensors can be divided into three major portions.First, an interface that accommodates the interfacingof actual sensing transducers is provided. The primaryrole of the sensing interface is the digitization of sampledsensor data. Second, every wireless sensor has a wirelessinterface that literally frees the sensor from the use ofwires for communications. A variety of different wirelessinterfaces have been previously utilized based on differ-ent protocol standards; however, the field currentlyseems to be converging to ubiquitous use of theIEEE802.15.4 wireless protocol standard which hasbeen designed explicitly for low-power, low data ratewireless sensor networks (IEEE Computer Soceity,2003). The third component of a wireless sensor is itscomputational core which generally includes a micro-controller (or microprocessor) and memory.

The inclusion of computing resources in the form of amicrocontroller represents a significant paradigm-shiftfor the sensing community. While the primary functionof the microcontroller is to coordinate the activities ofthe wireless sensors (e.g., collect, store, and communi-cate data), it can also be used to perform some rudimen-tary processing of raw sensor data prior tocommunication. In essence, the inclusion of computingwith the sensor is effectively pushing intelligence out to

362 S. CHO ET AL.

Page 3: Development of an Automated Wireless Tension Force Estimation

the outer edge of the monitoring system (i.e., at thesensor). Embedded computing is effective in convertinghigh bandwidth data stream into low bandwidth streamsthereby requiring less wireless bandwidth. This in-turnresults in greater power-efficiency (Lynch et al., 2004b),improved system reliability, and scalability of the wire-less monitoring system (i.e., greater nodal densities arepermissible).A variety of embedded algorithms have been pro-

posed for wireless sensors installed in civil structuresfor health monitoring. Early work focused on sensor-based interrogation schemes that utilized the sensor’sown data. Examples included the fast Fourier transform(FFT) (Lynch et al., 2006), autoregressive time-seriesmodeling (Lynch et al., 2004b), and principal compo-nent analysis (Park et al., 2008). More recent work hasbegun to consider network-centric computing architec-tures implemented in wireless sensor networks. In sucharchitectures, data (raw or processed) is shared amongwireless sensors so that the wireless sensors can executecoordinated but decentralized data processing of thesystems data. Examples of algorithms implemented forin-network decentralized data processing includes modalanalysis (i.e., mode shape determination) (Zimmermanet al., 2008), model updating for damage detection(Zimmerman and Lynch, 2010), and decentralizeddamage detection using structural flexibility(Nagayama, 2007). In this study, a sensor-centric com-puting architecture is adopted to embed tension estima-tion algorithms into the computational core of a wirelesssensor prototype proposed by Wang et al. (2007). Thealgorithms require only raw acceleration response datafrom a bridge cable; using this data, frequency domainmethods are employed to extract modal frequenciesfrom which cable tension is estimated. This sectionfirst describes the wireless sensor hardware in whichthe embedded algorithms are embedded. Second, imple-mentation of the tension estimation algorithm proposedby Zui et al. in the wireless sensor is discussed.

Wireless Sensor

The wireless sensor used in this study has beendesigned explicitly for structural monitoring of infra-structure systems (Wang et al., 2007). The wirelesssensor design offers a flexible sensing interface thatcan accommodate a heterogeneous array of sensorsincluding accelerometers, strain gauges, linear variabledifferential transducers (LVDT) among others. The sen-sing interface includes a 4-channel analog-to-digital con-verter (Texas Instruments ADS8341) offering 16-bitresolution and high sample rates (as high as 100 kHz).In the core of the wireless sensor is a low-power 8-bitmicrocontroller (Atmel ATmega128). This microcon-troller has been selected for its rich set of interfaces forcommunication with on-board peripherals such as the

sensing and wireless interfaces. In addition, the micro-controller is endowed with a large memory bank con-sisting of 128 kB of flash memory (for the embedment ofsoftware including data interrogation algorithms).Although the ATmega128 has its own random accessmemory (4 kB of RAM), it is deemed too small for theamount of data anticipated during the monitoring ofcivil infrastructure systems. In response to this limita-tion, an additional 128 kB of RAM is integrated in thewireless sensor design to store measurement and pro-cessed data. To establish a wireless communicationlink with other sensors, the Maxstream 9XCite wirelesstransceiver is integrated in the wireless sensor design.This radio operates on the 900MHz frequency bandand is able to communicate up to 300m, line-of-sightwith an over-the-air data rate of 38.4Kbps (kilobitsper second). The radio’s use of frequency-hoppingspread spectrum encoding ensures communications tobe highly reliable in the field.

The wireless sensor, when assembled, can be poweredby a single battery back consisting of 5 AA lithium-ionbatteries (with a nominal voltage of 7.5V). On average,the wireless sensor can safely operate for more than 30 hbefore battery replacement is necessary. However, duty-cycle use of the unit through the use of the hardware’sultra low-power sleep states can extend the battery lifeexpectancy to more than 1 year. The printed circuitboards and battery supply are packaged in a hardenedcontainer so that the sensor is protected from externalshock and environmental pollution. The wireless sensorprototype is shown in Figure 1 with key hardware com-ponents highlighted. After assembling the wirelesssensor hardware, software including the operatingsystem (OS) and data processing algorithms areembedded into the microcontroller’s flash memoryusing the microcontroller in-system programminginterface.

In this study a commercialized MEMS accelerometeris interfaced to the wireless sensor to measure thedynamic response of bridge stay cables. The CrossbowCXL02LF1 MEMS accelerometer is selected because ofits high sensitivity (1V/g) and exceptionally low noisefloor (1mg root mean square). A more detailed sum-mary of the accelerometer’s performance specificationis provided in Table 1. Since cables in bridges are typi-cally characterized by low natural frequencies resultingfrom their long and slender geometries, the Crossbowaccelerometer’s bandwidth (50Hz) is adequate for theapplication at hand. However, the low amplituderesponse of bridge cables under ambient excitationcould be a challenge for the wireless sensor’s 16-bitanalog-to-digital converter. To allow the wirelesssensor to acquire a higher quality reading from theaccelerometer, the analog sensor output could be ampli-fied before interfacing to the wireless sensor. Therefore,a signal conditioning board with amplification

Automated Wireless Tension Force Estimation System for Cable-stayed Bridges 363

Page 4: Development of an Automated Wireless Tension Force Estimation

capabilities developed by Wang et al. (2007) is used inthis study. The circuit has three primary functions: (1)shift all sensor outputs to have a mean output of 2.5V(which is the center of the wireless sensor interface 5Vinput range); (2) amplification of the sensor output(offers amplification factors of 5, 10, and 20 through a3-way switch); and (3) anti-alias filtering (0.02�25Hz).The signal conditioning board is designed using discreteanalog circuit elements and ordinary operational ampli-fiers. The signal conditioning board shown at Figure 2has been validated in several field applications (Lynchet al., 2006).

Embedded Implementation of Cable Tension Estimation

Algorithm

Embedded within the aforementioned wireless sensoris a multi-threaded OS that is designed to automate theoperation of the wireless sensor within a wireless struc-tural monitoring system (Wang et al., 2007). For exam-ple, the OS is responsible for reading raw data from thesensing interface, storing temporary and permanent datain memory, packetizing data for communication, andoperating the wireless transceiver. Software that imple-ments data interrogation algorithms is written toreside on a layer above the OS within the embedded

software hierarchy. In this study, the engineering algo-rithms embedded in the wireless sensor are divided intotwo portions. First, algorithms that transform time-his-tory data to the frequency domain and extract modalfrequencies are written. Second, cable tension estimationalgorithms that use the empirically derived modal fre-quencies are implemented.

FREQUENCY DOMAIN ANALYSISTo estimate tension in a vibrating stay cable requires a

precise estimate of the cable’s modal frequencies. Forthis reason, the Cooley�Tukey algorithm of the FFThas been implemented within the wireless sensor(Lynch et al., 2006). The FFT algorithm is used to cal-culate the complex-valued Fourier spectrum of accelera-tion time-history data collected by the sensing interface.In the implementation used in this study, the FFT algo-rithm utilizes 4096 points of acceleration time historydata when calculating the Fourier spectrum.

Once the wireless sensor has calculated the Fourierspectrum of the acceleration record, a means of auto-mating the identification of modal frequencies isexplored. In this study, an algorithm is proposed toautomate the identification of modal frequenciesthrough peak-picking in the frequency domain usingthe Fourier spectrum. The implementation of the peakpicking algorithm is assisted by the fact that cable damp-ing is generally low; as a result, their Fourier spectra arecharacterized by high and sharp modal peaks. Based onthe taut string theory, higher modal frequencies, fn, areroughly n times the first modal frequency, f1. These twoobservations can be leveraged to write a robust peak-picking algorithm for cables as follows:

(1) For f1: The initial estimate ( �f1) is taken from the firstpeak greater than the threshold value. Then it isupdated as the frequency of the largest peak in thefrequency range of [ �f1,1.8 �f1]. The same procedure is

4-Channelsenosr

interface

8-bitmicrocontroller

128 kB RAM memory(a) (b)

Figure 1. Wireless sensor prototype with key components highlighted: (a) primary PCB with microcontroller, memory, etc., (b) fully assembledwireless sensor.

Table 1. Specification of MEMS type accelerometer(Crossbow CXL02LF1).

Characteristic Value

Input range ±2 gSensitivity 1000 mV/gNoise level 1.0 mg rmsBandwidth 0 to 50 HzSupply voltage 5 VZero g output 2.5 V

364 S. CHO ET AL.

Page 5: Development of an Automated Wireless Tension Force Estimation

consecutively taken on the updated �f1, till no largerpeak is found in the updated frequency range.

(2) For fn (n> 1): A similar peak searching procedure isperformed for fn in a frequency range of[fn�1þ 0.8 f1, fn�1þ 0.6 f1]. In this study the fre-quency searching is carried out up to the third nat-ural frequency (f3).

Accurate extraction of a cable’s modal frequencies isnecessary to ensure accurate estimation of cable tension.However, ambient excitation on the stay cable may benon-stationary resulting in a non-stationary response ofa short duration. Fourier spectra calculated using ashort-time frame of non-stationary time-history datacould result in inaccurate estimates of the cable modalfrequencies. To overcome this technical challenge,Welch’s method (Welch, 1967) to average Fourier spec-tra calculated from the overlapping windowed segmentsof a time-history record is adopted for inclusion in thewireless sensor’s microcontroller. The method is animprovement over the standard approach of calculating

a single Fourier spectrum for each time-history signalthat has short-duration non-stationarity. Anotherattractive feature of the approach is that it is designedexplicitly for memory-constrained computers as is thecase when implementing on a wireless sensor (Welch,1967).

An attractive feature of the embedded OS of the wire-less sensor used in this study is its multi-tasking capabil-ities. Specifically, data collected by the wireless sensorcan be processed while additional data is collected andstored in memory (Wang et al., 2007). This feature isexploited in this study to implement Welch’s methodin a recursive manner so that long time histories ofdata can be analyzed using only a modest amount ofon-board memory. For example, once 8192 consecutivedata points have been collected, the wireless sensor pro-cesses the data as the next 4096 data points continue tobe collected. After processing the first 8192 points, theresults of the average Fourier spectrum are stored inmemory, the buffer is cleared, and the wireless sensorwaits for the next 8192 data points, composed of the

(a)

High-pass and amplifying filter Low-pass filter(b)

Sensorinput

2.5V Ref

2.5V Ref

1 uF

0.1 uF

0.1 uF840 Ω

2.5V Ref

0.1 uF

44.5 44.539.6 39.6

7.59Ref

Ref

2.5V

2.5V

Ref2.5V

0.1 uF

Sensoroutput

11.2 MΩ

1kΩ kΩkΩ

kΩ10 kΩ

10 kΩ

kΩkΩ

Variableresistor toadjust gain(×5, 10, 20)

Figure 2. (a) Signal conditioning board, (b) circuit schematic for mean-shifting, amplifying, and anti-aliasing sensor output signals.

Automated Wireless Tension Force Estimation System for Cable-stayed Bridges 365

Page 6: Development of an Automated Wireless Tension Force Estimation

last 4096 data points in the previous 8192 data pointsand newly measured 4096 data points, to be written tothe buffer after which the analysis is repeated. Thedetailed procedure on each segment of data is describedas follows:

(1) Collect 8192 points of acceleration time history data,a(k), which are stored in memory.

(2) Divide the time history into five segments (a1(k),a2(k), a3(k), a4(k), and a5(k)) consisting of 4096points with 75% overlap. For example, a1 beginswith a(1) while a2 begins with a(1025).

(3) Window a1(k) with a Hanning window. Using theFFT algorithm, calculate the Fourier spectrum(A1(k)) corresponding to the windowed segment,a1(k). Store the absolute value (|A1(k)|) intomemory.

(4) Calculate the Fourier spectrum corresponding tothe second windowed (Hanning) segment, a2(k).Sum |A1(k)| and |A2(k)| to yield |Asum(k)|. Store|Asum(k)| in place of the original |A1(k)|.

(5) Repeat Steps 3 and 4 for each of the next segmentsuntil the 5th segment has been processed.

(6) Divide |Asum(k)| by 5 to obtain the averaged Fourierspectrum, |Aave(k)|.

(7) Repeat the procedure on the next 8192 point seg-ment of data.

It is important to note that only four segments (a2(k),a3(k), a4(k), and a5(k)) are utilized from the subsequentsegments of data after the first segment. This is becausea5(k) in the first segment is overlapped with a1(k) of thesecond segment. The procedure is visualized in Figure 3.

Tension Force Estimation

Among the several methods available for estimatingthe tension force of a stay cable, the approach proposedby Zui et al. (1996) is selected, because it considersboth the flexural rigidity and sag inherent to the inclinedcable. The method is structured around seven simpleequations that correspond to a cable in different geomet-ric states (e.g., small vs. large sag, among others).As input, the equations consider the properties(i.e., Young’s modulus, E, inertia, I) and geometric con-figuration (i.e., length, l, sag, s) of the cable shown

Long-time history data (N∗ 4096)0.2

0

0 50

1st polling cycle

75%overlapping

100 150 200 250

FFT

Am

plitu

deA

mpl

itude

Am

plitu

deA

mpl

itude

Am

plitu

deA

mpl

itude

Am

plitu

de

Frequency (Hz)

Frequency (Hz)

Frequency (Hz)

FFT

FFT

FFT

FFT

FFT

–0.2

2nd polling cycle

Nth polling cycle

Windowed data

FFT result of each segmented data20

10

00 5 10 15 20 25

0 5 10 15 20 25

0 5 10 15 20 25

0 5 10 15 20 25

0 5 10 15 20 25

0 5 10 15 20 25

0 5 10 15 20 25

20

10

0

20

10

020

10

0

20

5

0

20

10

0

40

20

0

Averaging

Result of welch’s method

Figure 3. An overview of Welch’s method.

366 S. CHO ET AL.

Page 7: Development of an Automated Wireless Tension Force Estimation

in Figure 4. It also uses the first two modal frequencies,f1 and f2 (which are the easiest to extract accurately evenfor ambiently excited cables with large sag) as inputs.The seven equations may be divided into three casesaccording to the sagging condition and length of thecable as:CASE 1 � small sag (�� 3): Using f1

T ¼4w

gð f1l Þ

2 1� 2:20C

f1� 0:550

C

f1

� �2" #

, 17 � �

ð1-aÞ

T ¼4w

gð f1l Þ

2 0:865� 11:6C

f1

� �2" #

, 6 � � � 17

ð1-bÞ

T ¼4w

gð f1l Þ

2 0:828� 10:5C

f1

� �2" #

, 0 � � � 6 ð1-cÞ

CASE 2 � large sag (�� 3): Using f2

T ¼w

gð f2l Þ

2 1� 4:40C

f2� 1:10

C

f2

� �2" #

, 60 � �

ð2-aÞ

T ¼w

gð f2l Þ

2 1:03� 6:33C

f2� 1:58

C

f2

� �2" #

, 17 � � � 60

ð2-bÞ

T ¼w

gð f2l Þ

2 0:882� 85:0C

f2

� �2" #

, 0 � � � 17 ð2-cÞ

CASE 3 � very long cables: Using fn (2� n)

T ¼4w

n2gð fnl Þ

2 1� 2:20nC

fn

� �, 200 � � ð3Þ

where T is cable tension force, f1, f2, and fn are the 1st,2nd, and nth natural frequencies measured, respectively;� ¼ l

ffiffiffiffiffiffiffiffiffiffiffiT=EIp

; � ¼ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiwl=128EA�3 cos5 �

p½ð0:31� þ 0:5Þ=

ð0:31� � 0:5Þ�; w¼weight per unit length;C ¼

ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiEIg=wl4

p; EA¼ extensional rigidity; EI¼ flexural

rigidity; �¼ sag-to-span ratio (¼s/l0); and �¼ inclinationangle as shown in Figure 4.

It is important to note that two parameters (� and �)are needed at the outset to select the appropriate for-mula for tension estimation. In general, � and � may beinitially taken based on the design tension value. Thenthey are revised based on the concurrently identifiedtension force until convergence in T, �, and � isachieved. As discussed in the following section, it hasbeen found that only few iterations are needed (withmany instances requiring no additional iteration) toproperly select the formula to identify the cable tensionvalue. The formulas require only a few lower naturalfrequencies, which can be easily obtained from the ambi-ent acceleration measurements of the cable. Severalnumerical and experimental studies were reported toshow the validity of the method (Zui et al., 1996; Kimand Park, 2007). The embedded software was developedto select a proper formula for cable tension estimationwithout human intervention.

EXPERIMENTS FOR VALIDATION

Experimental Setup

To validate the proposed wireless tension force esti-mation system (WTFES), laboratory tests were carriedout on a scaled-down model of a cable from an actualcable-stayed bridge as shown in Figure 5. For the pur-pose of comparison, the tension force of the cable isdirectly measured using a threaded rod with straingauges, which is connected to the lower end of thecable. The rod was ground to have a square cross sectionso that four metal foil strain gauges could be attachedon four sides. The threaded rod with strain gauges waspreliminarily calibrated by controlled tensile testingusing an MTS load frame; excellent linearity wasfound between the tension force and the average strainfrom the four strain gauges (Figure 6).

Experimental Procedure

A series of validation tests were carried out to checkthe feasibility of the proposed WTFES for various

l0

l0 /2 l0/2

s

θ

l

T

T

E, A, I, w

Figure 4. Conceptual drawing of an inclined stay cable (adaptedfrom Zui et al., 1996).

Automated Wireless Tension Force Estimation System for Cable-stayed Bridges 367

Page 8: Development of an Automated Wireless Tension Force Estimation

sensing locations, tension forces, and sags. Ten sets oftests using FFT of single 4096 data points and eightsets of tests using Welch’s method of 20480(¼5� 4096) data points were carried out with differentcable tension forces (with varying sagging conditions).Impacts were applied at arbitrary locations on the cable.To check the effect of various sensing locations, threewireless sensors are installed on the cable at 0.5 (S1),1.0 (S2), and 1.5m (S3) when measuring from the

lower end of the cable as shown in Figure 5. The detailsof the test cases are shown in Table 2. It can be seen thatthe sag increases as the tension force decreases. It isworthy to note that FFT of single 4096 data pointswas utilized in T1�T10; T1�T4 correspond to thecable with small levels of sags; and Cases T5�T10 arefor cables with large sags. On the other hand, Welch’smethod was used in WT1�WT8 which are cases withlarge sags.

S1(0.5m)

3.52

m

5.22m

33.98°

6.29m

Threaded rodwith strain

gauges

S3(1.5m)

S2(1.0m)

(a) (b)

(c) (d)

Figure 5. Laboratory cable model for validation tests: (a) geometry of laboratory model, (b) cable at laboratory, (c) specially designed cableanchorage, (b) threaded rod with strain gages on it.

0 1 2 3 4 5 6 7 80

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5× 10–4 Tension force–strain curve

Force (kN)

Str

ain

y = 5.6e–005∗x – 1e–007

Measured dataLinear fit

(a) (b)

Figure 6. Calibration of installed strain gages: (a) tensile test of a threaded bar with strain gages using MTS, (b) linear relationship betweentension force and average strain from four strain gages.

368 S. CHO ET AL.

Page 9: Development of an Automated Wireless Tension Force Estimation

EXPERIMENTAL RESULTS

Tests with Small Sag: T1�T4

Test cases T1�T4 were carried out on the cable withsmall levels of sag, where sag-to-span ratios are less than1.51� 10�3 as shown in Table 2. Single 4096 data pointswere collected and processed by embedded FFT moduleto obtain Fourier spectra of the cable. For case T1,Figure 7 shows the Fourier spectra of the measuredaccelerations from the three wireless sensors (S1�S3).The natural frequencies estimated using the embeddedpeak picking algorithm are marked in the figures. Thefrequencies autonomously estimated at three sensorlocations are found to be the same as those obtainedoff-line up to the precision (i.e., floating point precision)of the proposed system. Similarly identified results wereobtained from three sensors for the other test cases(T2�T4). Fourier spectra of the accelerations measuredat S3 for cases T2, T3, and T4 are shown in Figure 7. InTable 3, the estimated frequencies for cases T1�T4 arecompared with the values (marked as ‘Ref.’) obtained byoff-line operations requiring human intervention. Theautonomously estimated natural frequencies are foundto be the same as the reference values for all test cases(Figure 8).For the cable tension estimation, two condition para-

meters (� and �) were initially evaluated using an arbi-trary large tension value of 10 kN, which is much largerthan the true values correspond to all of the test cases aslisted in Table 2. Thus Equation (1-a) was selected at thefirst iteration for the tension estimation for test casesT1�T4. Table 3 shows the estimated cable tension

forces along with the results from the strain gauges(marked as ‘Ref.’). All of the estimated tension valuesare found to be in excellent agreement with the corre-sponding reference values; the discrepancies were <2%.It has been found that no additional iterations were nec-essary for tension estimation in these four test cases. Thepresent results indicate that the embedded software ofthe proposed WTFES is working correctly in a comple-tely automated fashion.

Tests with Large Sag: T5�T10

To investigate the feasibility of the present method forcables with large sags, test cases T5�T10 were carriedout for sag-to-span ratios in the range of 2.43� 10�3 to24.94� 10�3 as shown in Table 2. Example Fourierspectra from the measured 4096 acceleration data andthe peak picked natural frequencies are shown for CasesT5, T7, and T9 in Figure 9. Again, consistent naturalfrequencies were obtained regardless of the sensor loca-tions for all of the test cases. Furthermore, these fre-quencies are the same as those obtained off-line up tothe system’s precision as shown in Table 4.

For cable tension estimation, two condition para-meters were initially evaluated using an arbitrary largetension force of 10 kN as in the previous tests.Accordingly, Equation (2-a) was selected for the firstiteration in cases T5�T10. In Table 4, the estimatedcable tension forces are compared with the valuesobtained using the strain gauges. The estimated tensionforces show larger discrepancies than those obtainedusing the strain gauges: <7% in cases T5�T8 and�12% for case T9. The larger discrepancy in case T9may have come from the relatively large sensitivity of

Table 2. Details of various tests.

Test casesSamplingrate (Hz) Amplification

Sag(mm)

Sag-to-spanratio

Tensionforcea (N)

Single FFT T1 50 5 <3.0 <1�10�3 1162T2 50 5 <3.0 <1�10�3 1023T3 40 5 6.4 1.23�10�3 858T4 40 5 7.9 1.51�10�3 617T5 40 5 12.7 2.43�10�3 522T6 40 5 19.1 3.66�10�3 308T7 40 5 50.8 9.73�10�3 99.5T8 40 5 63.5 12.16�10�3 48.7T9 40 5 95.3 18.26�10�3 28.8

T10 40 5 130.2 24.94�10�3 13.4

Welch’s method WT1 40 5 25.4 4.87�10�3 264WT2 40 5 38.1 7.30�10�3 151WT3 40 5 57.2 10.96�10�3 81.5WT4 40 5 69.9 13.39�10�3 49.4WT5 40 5 82.6 15.82�10�3 33.2WT6 40 5 108.0 20.69�10�3 21.0WT7 40 5 127.0 24.33�10�3 14.4WT8 40 5 152.4 29.19�10�3 8.8

aMeasured directly using strain gauges on threaded rod.

Automated Wireless Tension Force Estimation System for Cable-stayed Bridges 369

Page 10: Development of an Automated Wireless Tension Force Estimation

0 5 10 15 20 250

200

400

600

800

1000Fourier spectrum from S3 (T2)

Frequency (Hz)

Am

plitu

de

(a)

0 5 10 15 200

200

400

600

800

1000Fourier spectrum from S3 (T3)

Frequency (Hz)

Am

plitu

de

(b)

0 5 10 15 200

200

400

600

800

1000Fourier spectrum from S3 (T4)

Frequency (Hz)

Am

plitu

de

(c)

12.15Hz

6.30Hz 17.80Hz

11.24Hz

5.85Hz

16.59Hz

9.70Hz

5.04Hz 14.38Hz

Figure 8. Fourier spectra and natural frequencies (T2�T4).

0 5 10 15 20 250

200

400

600

800

1000Fourier spectrum from S1 (T1)

Frequency (Hz)

Am

plitu

de

(a)

0 5 10 15 20 250

200

400

600

800

1000Fourier spectrum from S2 (T1)

Am

plitu

de

(b)

0 5 10 15 20 250

200

400

600

800

1000Fourier spectrum from S3 (T1)

Frequency (Hz)

Am

plitu

de

(c)

13.05Hz

6.79Hz 19.18Hz

13.05Hz

6.79Hz

6.79Hz

19.18Hz

19.18Hz

13.05Hz

Frequency (Hz)

Figure 7. Fourier spectra and natural frequencies at different loca-tions (T1).

Table 3. Results of tests with small sags (T1�T4).

Natural frequencies Tension force estimation

Test name 1st (Hz) 2nd (Hz) 3rd (Hz) Tension force (N) Error (%) Selected formula

T1 *Reference 6.79 13.05 19.18 1162 � �S1 6.79 13.05 19.18 1173 0.93 1-a

T2 *Reference 6.30 12.15 17.80 1023 � �S1 6.30 12.15 17.80 1004 �1.85 1-a

T3 *Reference 5.85 11.24 16.59 858 � �S1 5.85 11.24 16.59 860 0.23 1-a

T4 *Reference 5.04 9.70 14.38 617 � �S1 5.04 9.70 14.38 628 1.78 1-a

*Reference: Natural frequencies obtained by offline operation and tension force measured using strain gages.

370 S. CHO ET AL.

Page 11: Development of an Automated Wireless Tension Force Estimation

the measured tension force using strain gauges for thecase with a small tension force (in the order of tens ofNewtons). It has been also found that only one or twoiterations were needed for tension estimations: one iter-ation from Equation (2-a) to Equation (2-b) for T5 andT6, and two iterations from Equation (2-a) to Equations(2-b) and (2-c) for T7�T10.

Tests for Welch’s Method with Large Sag: WT1�WT8

To validate the performance of the embedded Welch’smethod (Welch, 1967) for cable tension force estimation,eight more test cases WT1�WT8 were carried out on thecable with large sag conditions, where sag-to-span ratiosare from 4.87� 10�3 to 29.1� 10�3 as shown in Table 2.In each case of WT1�WT8, an acceleration record withtotal 20480 (¼5� 4096) data was collected and pro-cessed concurrently during the measurement byWelch’s method as shown in Figure 3. Total 17 segmentswith 75% overlapping were utilized to obtain an averageFourier spectrum.

Examples of the inaccurate estimates of cable naturalfrequencies from short-duration data is shown inFigures 10 and 11. Two Fourier spectra, one by FFTusing single 4096 data points (FS1) and the other byWelch’s method using 20480 data points (FS2), wereobtained from S1 and S3 in WT5 excited at the sameimpulsive input, and compared with zooming in fre-quency ranges around 4 lower peaks. (Since S3is installed at the nodal point of the fourth vibrationmode, the fourth peak is not clearly shown inFigure 11). FS2s look less noisy than FS1s, and thereare small discrepancies at the second and fourth peaks asshown in the second and fourth zoomed-in figures. Thesecond peak of FS1 is separated into two sharp peaks(3.184 and 3.203Hz), and the first sharp peak (3.184Hz)

Table 4. Results of tests with large sag (T7�T10).

Natural frequencies Tension force estimation

Test name 1st (Hz) 2nd (Hz) 3rd (Hz) Tension force (N) Error (%) Selected formula

T5 *Reference 4.73 9.08 13.49 522 � �S1 4.73 9.08 13.49 494 �5.36 2-b

T6 *Reference 3.80 7.34 10.84 308 � �S1 3.80 7.34 10.84 309 3.34 2-b

T7 *Reference 2.33 4.50 6.33 99.5 � �S1 2.33 4.50 6.33 93.1 6.43 2-c

T8 *Reference 1.88 3.61 5.44 48.7 � �S1 1.88 3.61 5.44 49.4 1.44 2-c

T9*Reference 1.59 3.02 4.57 28.8 � �S1 1.59 3.02 4.57 25.4 11.81 2-c

T10*Reference 1.39 2.68 4.00 13.4 � �S1 1.39 2.68 4.00 13.6 1.49 2-c

*Reference: Natural frequencies obtained by offline operation and tension force measured using strain gages.

0 5 10 15 20

200

0

400

600

800

1000Fourier spectrum from S3 (T5)

Frequency (Hz)

Am

plitu

de(a)

0 5 10 15 200

200

400

600

800

1000Fourier spectrum from S3 (T7)

Frequency (Hz)

Am

plitu

de

(b)

0 5 10 15 200

200

400

600

800

1000Fourier spectrum from S3 (T9)

Frequency (Hz)

Am

plitu

de

(c)

9.08Hz

4.73Hz 13.49Hz

4.50Hz2.33Hz

6.33Hz

3.02Hz1.59Hz

4.57Hz

Figure 9. Fourier spectra and natural frequencies (T5, T7, T9).

Automated Wireless Tension Force Estimation System for Cable-stayed Bridges 371

Page 12: Development of an Automated Wireless Tension Force Estimation

has a little bit larger amplitude then the other (3.203Hz)so that 3.184Hz may be interpreted as the second natu-ral frequency of FS1. However, the other peak of3.203Hz has more power than the first peak in FS2 sothat it should be interpreted as the real second naturalfrequency. This peak discrepancy has resulted in the dis-crepancy of the estimated tension force. As shown inTable 5, Welch’s method estimated the tension force as32.4N with 2.4% error, while FFT of single 4096 datapoints did as 31.7N with 4.5% error.All results from the test cases WT1�WT8 is shown in

Table 5. Examples of averaged Fourier spectra from themeasured record with 20480 data points using Welch’smethod are shown for Cases WT1, WT4, and WT7 inFigure 12. Consistent natural frequencies were obtainedregardless of the sensor locations for all of the test cases,which results in natural frequencies are the same asthose obtained off-line up to the system’s precision asshown in Table 5. The estimated tension forces ofWT1�WT8 show good agreement with the measuredtensions <10% error. Taking into consideration ofsmall values of the tension forces, Welch’s method pro-vides good estimates compared with the measured

tension force. It has been also found that only one ortwo iterations were needed for tension estimations: oneiteration from Equation (2-a) to Equation (2-b) for WT1and WT2, and two iterations from Equation (2-a) toEquations (2-b) and (2-c) for WT3�WT8.

Summary of Estimated Tension Forces

In Figure 13, all of the estimated tension forces usingthe WFTES are plotted against the tension forces mea-sured from the strain gauges. It is interesting to note thatthe sag-to-span ratios of T3�T6 are in the range of1.23� 10�3 to 3.66� 10�3, which are very similar tothe values for the stay-cables of the Seohae GrandBridge (Korea) which are from 0.84� 10�3 to3.62� 10�3 (Ahn et al., 2001). It is clearly shown thatthe present WFTES using the wireless sensors showsexcellent performance for cable tension estimation fora wide range of cable tension force and sag, particularlyfor the cases applicable to the cables in a realistic cable-stayed bridge. If ambient vibration by the passing trafficor wind is measured on the real bridge cable, Welch’s

0 1 2 3 4 5 6 7 8 9 100

50

100

150

200

Frequency (Hz)

Am

plitu

de

Comparison of single FFT and Welch (S1)

Single FFT

Welch

(a)

1.6 1.8 20

5

10

15

20

25

30

35

40

45

50

Frequency (Hz)

Am

plitu

de

3 3.2 3.40

10

20

30

40

50

60

70

80

Frequency (Hz)

Am

plitu

de

4.6 4.8 50

50

100

150

Frequency (Hz)

Am

plitu

de

6.4 6.6 6.80

50

100

150

Frequency (Hz)

Am

plitu

de

(b)

4.822Hz

4.822Hz

6.569Hz

6.608Hz

3.203Hz

3.184Hz

1.650Hz

1.650Hz

Figure 10. Difference of Fourier spectra from S1 for WT5 (single FFT vs. Welch’s method): (a) comparison of the whole Fourier spectra, (b)zoom-in parts.

372 S. CHO ET AL.

Page 13: Development of an Automated Wireless Tension Force Estimation

0 1 2 3 4 5 6 7 8 9 100

50

100

150

200

Frequency (Hz)

Am

plitu

de

Comparison of single FFT and Welch (S3)

Single FFTWelch

1.6 1.8 20

10

20

30

40

50

60

70

80

90

100

Frequency (Hz)

Am

plitu

de

3 3.2 3.40

20

40

60

80

100

120

140

160

180

200

Frequency (Hz)

Am

plitu

de

4.6 4.8 50

20

40

60

80

100

120

140

160

180

Frequency (Hz)

Am

plitu

de

6.4 6.6 6.80

2

4

6

8

10

12

14

16

18

20

Frequency (Hz)

Am

plitu

de

3.184Hz

3.203Hz

1.650Hz

1.650Hz

4.822Hz

4.822Hz

(b)

(a)

Figure 11. Difference of Fourier spectra from S3 for WT5 (single FFT vs. Welch’s method): (a) comparison of the whole Fourier spectra, (b)zoom-in parts.

Table 5. Results of tests using Welch’s method (WT1�WT8).

Natural frequencies Tension force estimation

Test name 1st (Hz) 2nd (Hz) 3rd (Hz) Tension force (N) Error (%) Selected formula

WT1 *Reference 3.604 6.973 10.430 264 � �S1 3.604 6.973 10.430 275 4.17 2-b

WT2 *Reference 2.715 5.283 7.861 151 � �S1 2.715 5.283 7.861 145 �3.97 2-b

WT3 *Reference 2.188 4.180 6.348 81.5 � �S1 2.188 4.180 6.348 76.1 �6.63 2-c

WT4 *Reference 1.816 3.525 5.332 49.4 � �S1 1.816 3.525 5.332 45.6 �7.69 2-c

WT5 *Reference 1.650 3.203 4.824 33.2 � �S1 1.650 3.203 4.824 32.4 �2.41 2-c

WT6 *Reference 1.513 2.920 4.424 21.0 � �S1 1.513 2.920 4.424 21.9 2.41 2-c

WT7 *Reference 1.396 2.725 4.102 14.4 � �S1 1.396 2.725 4.102 15.2 �5.56 2-c

WT8 *Reference 1.309 2.549 3.838 8.8 � �S1 1.309 2.549 3.838 9.6 9.09 2-c

*Reference: Natural frequencies obtained by offline operation and tension force measured using strain gages.

Automated Wireless Tension Force Estimation System for Cable-stayed Bridges 373

Page 14: Development of an Automated Wireless Tension Force Estimation

method is expected to remove the non-stationarity of theshort-duration signal in a memory-constrained smartsensor.

CONCLUSIONS

A low-cost and automated WTFES is developed forsteel cables in long-span bridges. The low-cost hardwaresystem consists of a wireless sensor made of commercialoff-the-shelf components, a cheap commercializedMEMS accelerometer, and a specially designed signalconditioning circuit with amplification, mean-shifting,and anti-aliasing filtering functionality. The softwareembedded in the wireless sensor is composed of apeak picking algorithm for identification of the naturalfrequencies of a cable, Welch’s method to measurelong-time ambient vibration of a cable in the memory-constrained wireless sensing unit, and a modern tension

0 5 10 15 200

100

200

300

400

500Fourier spectrum from S3 (WT1)

Am

plitu

de

Frequency (Hz)

(a)

0 5 10 15 200

50

100

150

200

250Fourier spectrum from S3 (WT4)

Am

plitu

de

Frequency (Hz)

(b)

0 5 10 15 200

50

100

150

200

250Fourier spectrum from S3 (WT7)

Am

plitu

de

Frequency (Hz)

(c)

3.60Hz

10.43Hz6.97Hz

1.82Hz3.53Hz

5.33Hz

1.40Hz

2.73Hz

4.10Hz

Figure 12. Fourier spectra and natural frequencies (WT1, WT4, WT7).

0 0.2 0.4 0.6

Range of sag-to-span ratiosfor Seohae Bridge

0.8 1 1.20

0.2

0.4

0.6

0.8

1

1.2

Tension estimation by WFTES(10 sets (FFT), 8 sets (Welch))

Measured by SG (kN)

Est

imat

ed b

y W

FT

ES

(kN

)

Tension force with FFT

Tension force with Welch

Figure 13. Comparison of estimated and measured tension forces(ten sets of tests using single FFT and eight sets of tests usingWelch’s method).

374 S. CHO ET AL.

Page 15: Development of an Automated Wireless Tension Force Estimation

force estimation method using the identified natural fre-quencies with consideration of the cable sag and bendingstiffness. The system is designed to be fully autonomousrequiring no human intervention.A series of validation tests were carried out on a

scaled down version of a cable from an operationalcable-stayed bridge (Seohae Bridge, Korea). Eighteentest cases were carried out on the cable with varioussag-to-span ratios, which represent small and largesags. The results of the experimental study are summar-ized as follows. (1) Consistent results can be obtained forthe first three natural frequencies (regardless of thesensor location along the cable) using the proposedpeak-picking algorithm. (2) The cable tension forces esti-mated by the present method are very close to the valuesmeasured using strain gauges for the cases with sag-to-span ratios similar to those of the cables in the SeohaeGrand Bridge, which indicates the applicability ofthe present system to realistic cable-supported bridges.(3) Embedding Welch’s method is a good solution formore accurate estimation of the natural frequencies byremoving the non-stationarity of short-duration signals,which is a limit of the memory-constrained wirelesssensors.

ACKNOWLEDGMENTS

This work was jointly supported by the KoreaResearch Foundation Grant funded by the KoreanGovernment (MOEHRD) (KRF-2007-612-D00136)and the Smart Infra-Structure Technology Center(SISTeC) at KAIST sponsored by the Korea Scienceand Engineering Foundation. Additional funding wasoffered by the National Science Foundation (NSF)under Grant Number CMMI-0726812 (ProgramManager: Dr S.C. Liu). The financial supports and theencouragement by Dr Liu have been vital to the successof this work. Technical assisstnace has been provided byMr Andrew T. Zimmerman and Mr R. Andrew Swartzat the University of Michigan, USA. Their valuableassistance is greatly appreciated.

REFERENCES

Ahn, S., Kim, M. and Kim, I. 2001. ‘‘Estimation of Stay CableTension by Ambient Vibration Measurement in Seohae Cable-stayed Bridge,’’ In: Proceedings of KSCE Annual Conference,Seoul, Korea (in Korean).

Casas, J.R. 1994. ‘‘A Combined Method for Measuring Cable Forces:the Cable-stayed Alamillo Bridge, Spain,’’ Structural EngineeringInternational, 4(4):235�240.

Gangone, M.V., Whelan, M.V., Fuchs, M.P. and Janovan, K.D. 2007.‘‘Performance Monitoring of a Short-span Integral-abutmentBridge Using Wireless Sensor Technology,’’ In: Sensor Systemsand Networks: Phenomena, Technology, and Applications forNDE and Health Monitoring, Proceedings of SPIE, San Diego,CA.

Hua, X.G., Hua, X.G., Ni, Y.Q., Zhou, H.F. and Ko, J.M. 2005.‘‘Instrumentation for Durability Monitoring of a Long-span

Cable-stayed Bridge,’’ In: Smart Structures and Materials2005: Sensors and Smart Structures Technologies for Civil,Mechanical, and Aerospace Systems, Proceedings of SPIE,San Diego, CA.

IEEE Computer Society. 2003. Part 15.4: Wireless MediumAccess Control (MAC) and Physical Layer (PHY)Specifications for Low-rate Wireless Personal Area Networks,Institute of Electrical and Electronics Engineers (IEEE),New York.

Irvine, H.M. 1981. Cable Structures, The MIT Press, Cambridge, MA.

Kim, B.H. and Park, T. 2007. ‘‘Estimation of Cable Tension ForceUsing the Frequency-based System Identification Method,’’Journal of Sound and Vibration, 304(3�5):660�676.

Kim, B.H., Park, T., Shin, H. and Yoon, T.Y. 2007. ‘‘A ComparativeStudy of the Tension Estimation Methods for Cable SupportedBridges,’’ International Journal of Steel Structures, 7(1):77�84.

Lynch, J.P., Law, K.H., Kiremidjian, A.S., Carryer, E., Farrar, C.R.,Sohn, H., Allen, D.W., Nadler, B. and Wait, J.R. 2004a. ‘‘Designand Performance Validation of a Wireless Sensing Unit forStructural Monitoring Applications,’’ Structural Engineeringand Mechanics, 17(3�4):393�408.

Lynch, J.P., Sundararajan, A., Law, K.H., Kiremidjian, A.S. andCarryer, E. 2004b. ‘‘Embedding Damage Detection Algorithmsin a Wireless Sensing Unit for Attainment of Operational PowerEfficiency,’’ Smart Materials and Structures, 13(4):800�810.

Lynch, J.P. and Loh, K.H. 2006. ‘‘A Summary Review of WirelessSensors and Sensor Networks for Structural HealthMonitoring,’’ Shock and Vibration Digest, 38(2):91�128.

Lynch, J.P., Wang, Y., Loh, K.H., Yi, J. and Yun, C.-B. 2006.‘‘Performance Monitoring of the Geumdang Bridge Using aDense Network of High-resolution Wireless Sensors,’’ SmartMaterials and Structures, 15(6):1561�1575.

Morgenthal, G. 1999. ‘‘Cable-stayed Bridges � EarthquakeResponse and Passive Contro,’’ PhD Thesis, Imperial College,London, UK.

Nagayama, T. 2007. ‘‘Structural Health Monitoring Using SmartSensors,’’ Ph.D.Thesis, University of Illinois at Urbana-Champaign, Urbana, IL.

Pakzad, S.N., Fenves, G.L., Kim, S. and Culler, D.E. 2008. ‘‘Designand Implementation of Scalable Wireless Sensor Network forStructural Monitoring,’’ Journal of Infrastructure Systems,14(1):89�101.

Park, S., Lee, J.-J., Yun, C.-B. and Inman, D.J. 2008. ‘‘Electro-mechanical Impedance-based Wireless Structural HealthMonitoring Using PCA-Data Compression and k-meansClustering Algorithms,’’ Journal of Intelligent Material Systemsand Structures, 19(4):509�520.

Russell, J.C. and Lardner, T.J. 1998. ‘‘Experimental Determination ofFrequencies and Tension for Elastic Cables,’’ Journal ofEngineering Mechanics, 124(10):1067�1072.

Spencer Jr, B.F., Ruiz-Sandoval, M.E. and Kurata, N. 2004.‘‘Smart Sensing Technology: Opportunities and Challenges,’’Journal of Structural Control and Health Monitoring,11(4):349�368.

Spencer Jr, B.F., Nagayama, T. and Rice, J.A. 2005. ‘‘DecentralizedStructural Health Monitoring Using Smart Sensors,’’ In: SmartStructures and Materials: Sensors and Smart StructuresTechnologies for Civil, Mechanical, and Aerospace Systems,Proceedings of SPIE, San Diego, CA.

Straser, E.G. and Kiremidjian, A.S. 1998. ‘‘A Modular, WirelessDamage Monitoring System for Structures,’’ John A. BlumeEarthquake Engineering Center, Technical Report 128,Stanford, CA.

Wang, D., Liu, J., Zhou, D. and Huang, S. 1999. ‘‘Using PVDFPiezoelectric Film Sensors for In Situ Measurement of Stayed-cable Tension of Cable-stayed Bridges,’’ Smart Materials andStructures, 8(5):554�559.

Wang, M.L., Wang, G. and Zhao, Y. 2005. ‘‘Application of EM StressSensors in Large Steel Cables,’’ Sensing Issues in Civil StructuralHealth Monitoring, Springer, The Netherlands.

Automated Wireless Tension Force Estimation System for Cable-stayed Bridges 375

Page 16: Development of an Automated Wireless Tension Force Estimation

Wang, Y., Lynch, J.P. and Law, K.H. 2007. ‘‘A WirelessStructural Health Monitoring System with MultithreadedSensing Devices: Design and Validation,’’ Structure andInfrastructure Engineering, 3(2):103�120.

Welch, P.D. 1967. ‘‘The Use of Fast Fourier Transform for theEstimation of Power Spectra: a Method Based on TimeAveraging Over Short, Modified Periodograms,’’ IEEETransactions on Audio and Electroacoustics, 15(2):70�73.

Yanaka, Y. and Kitagawa, M. 2002. ‘‘Maintenance of Steel Bridges onHonshu-Shikoku Crossing,’’ Journal of Constructional SteelResearch, 58(1):131�150.

Zui, H., Shinke, T. and Namita, Y. 1996. ‘‘Practical Formulas forEstimation of Cable Tension by Vibration Method,’’ Journal ofStructural Engineering, 122(6):651�656.

Zimmerman, A.T., Shiraishi, M., Swartz, R.A. and Lynch, J.P. 2008.‘‘Automated Modal Parameter Estimation by Parallel Processingwithin Wireless Monitoring Systems,’’ Journal of InfrastructureSystems, 14(1):102�113.

Zimmerman, A.T. and Lynch, J.P. ‘‘A Parallel Simulated AnnealingArchitecture for Model Updating in Wireless Sensor Networks,’’IEEE Sensor Journal, (in press).

376 S. CHO ET AL.