development of a dynamic cardiac simulator controlled by...

10
Development of a dynamic cardiac simulator controlled by Arduino Poster No.: C-3675 Congress: ECR 2019 Type: Scientific Exhibit Authors: C. K. Bandeira 1 , H. Vieira Neto 2 , M. P. M. M. Vieira 2 ; 1 São José dos Pinhais/BR, 2 Curitiba/BR Keywords: Radiographers, Radioprotection / Radiation dose, Image manipulation / Reconstruction, Education, Technical aspects, Quality assurance DOI: 10.26044/ecr2019/C-3675 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to third- party sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. www.myESR.org Page 1 of 10

Upload: others

Post on 21-Nov-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Development of a dynamic cardiac simulator controlled by ...dainf.ct.utfpr.edu.br/~hvieir/download/ecr19.pdf · The Arduino code developed allowed controlling the movement of the

Development of a dynamic cardiac simulator controlled byArduino

Poster No.: C-3675

Congress: ECR 2019

Type: Scientific Exhibit

Authors: C. K. Bandeira1, H. Vieira Neto2, M. P. M. M. Vieira2; 1São José

dos Pinhais/BR, 2Curitiba/BR

Keywords: Radiographers, Radioprotection / Radiation dose, Imagemanipulation / Reconstruction, Education, Technical aspects,Quality assurance

DOI: 10.26044/ecr2019/C-3675

Any information contained in this pdf file is automatically generated from digital materialsubmitted to EPOS by third parties in the form of scientific presentations. Referencesto any names, marks, products, or services of third parties or hypertext links to third-party sites or information are provided solely as a convenience to you and do not inany way constitute or imply ECR's endorsement, sponsorship or recommendation of thethird party, information, product or service. ECR is not responsible for the content ofthese pages and does not make any representations regarding the content or accuracyof material in this file.As per copyright regulations, any unauthorised use of the material or parts thereof aswell as commercial reproduction or multiple distribution by any traditional or electronicallybased reproduction/publication method ist strictly prohibited.You agree to defend, indemnify, and hold ECR harmless from and against any and allclaims, damages, costs, and expenses, including attorneys' fees, arising from or relatedto your use of these pages.Please note: Links to movies, ppt slideshows and any other multimedia files are notavailable in the pdf version of presentations.www.myESR.org

Page 1 of 10

Page 2: Development of a dynamic cardiac simulator controlled by ...dainf.ct.utfpr.edu.br/~hvieir/download/ecr19.pdf · The Arduino code developed allowed controlling the movement of the

Aims and objectives

According to the World Health Organization, cardiovascular diseases are the leadingcause of death in the world [1]. Thus, it is necessary that radiology professionalsacknowledge the procedures and protocols that involve cardiac exams. Therefore, theaim of this work is to develop a dynamic cardiac simulator, which allows the trainingof students and professionals without the need to expose patients to ionizing radiationunnecessarily.

Page 2 of 10

Page 3: Development of a dynamic cardiac simulator controlled by ...dainf.ct.utfpr.edu.br/~hvieir/download/ecr19.pdf · The Arduino code developed allowed controlling the movement of the

Methods and materials

The dynamic cardiac phantom was developed by using a cardiac simulator consisting oflatex and cannulas that simulate the pulmonary and systemic circulation (Fig.1).

Fig. 1: Cardiac simulator consisting of latex.References: Physics, Universidade Tecnológica Federal do Paraná - Curitiba/BR

The internal components are showed in the Fig. 2. The cardiac simulator has fourchambers that represent two atria and two ventricles. A septum separates the chambersbetween right and left side. Between the atria and the ventricles are located septa thatsimulate the atrioventricular valves, that is, they allow the passage of fluid from the atriato the ventricles and prevent the circulation in the opposite direction.

Page 3 of 10

Page 4: Development of a dynamic cardiac simulator controlled by ...dainf.ct.utfpr.edu.br/~hvieir/download/ecr19.pdf · The Arduino code developed allowed controlling the movement of the

Fig. 2: Internal structures of the cardiac simulator.References: Physics, Universidade Tecnológica Federal do Paraná - Curitiba/BR

To simulate the small circulation, a single lumen catheter was connected from the rightventricle to the left atrium and to represent the systemic circulation a double lumencatheter was connected from the left ventricle to the right atrium. One of these lumenswas used to inject fluid into the right atrium.

Four hobby servomotors (Fig. 3) were positioned next to each of the four cardiacchambers of the cardiac phantom in such a way that the horns attached to eachservomotor axis compress the respective chambers. The servomotors were controlledby PWM signals generated by an Arduino Uno.

Page 4 of 10

Page 5: Development of a dynamic cardiac simulator controlled by ...dainf.ct.utfpr.edu.br/~hvieir/download/ecr19.pdf · The Arduino code developed allowed controlling the movement of the

Fig. 3: Hobby servomotor with horns attached to axis used to compress the phantomchambers.References: Physics, Universidade Tecnológica Federal do Paraná - Curitiba/BR

The amplitude and frequency of rotation of the axis of each servomotor was controlledfrom a code developed in the Arduino Integrated Development Environment. For eachheartbeat the total rotation range is 90 degrees, being 45 degrees clockwise and45 degrees counterclockwise, and the time for 1 degree rotation is 10 milliseconds.Movement occurs synchronously.

Page 5 of 10

Page 6: Development of a dynamic cardiac simulator controlled by ...dainf.ct.utfpr.edu.br/~hvieir/download/ecr19.pdf · The Arduino code developed allowed controlling the movement of the

Results

The phantom developed in this work (Fig. 4) resembles dynamic phantoms presentedin the literature [2,3] that reproduce the heart rate during the simulation of radiologicalprocedures. A differential of the presented simulator is the possibility of changing therotation time of the servomotor horns in the developed code, which allows simulatingpathologies such as cardiac arrhythmias.

Page 6 of 10

Page 7: Development of a dynamic cardiac simulator controlled by ...dainf.ct.utfpr.edu.br/~hvieir/download/ecr19.pdf · The Arduino code developed allowed controlling the movement of the

Fig. 4: Developed cardiac phantom. The four hobby servomotors next to eachchamber and the syringe used to inject fluid test and iodine-based contrast.References: Physics, Universidade Tecnológica Federal do Paraná - Curitiba/BR

A radiographic image (Fig. 5) shows the internal components of the phantom. The imageshowed in the Fig. 6 was recorded in a fluoroscopy equipment and indicated that injectediodine-based contrast filled the cardiac chambers.

Fig. 5: Radiographic image showing the internal components of the cardiac simulatorand the catheters that connect the chambers.References: Physics, Universidade Tecnológica Federal do Paraná - Curitiba/BR

Page 7 of 10

Page 8: Development of a dynamic cardiac simulator controlled by ...dainf.ct.utfpr.edu.br/~hvieir/download/ecr19.pdf · The Arduino code developed allowed controlling the movement of the

Fig. 6: Fluoroscopy image showing the cardiac chambers filled with iodine-basedcontrast.References: Physics, Universidade Tecnológica Federal do Paraná - Curitiba/BR

The Arduino code developed allowed controlling the movement of the horns of eachservomotor independently. The compression of the atriums and ventricles occurred at afrequency of 1 hertz. However, the intensity of compression of the cardiac chambers wasnot enough to completely drain them out, which resulted in low flow velocity. There wasalso a small fluid return through the cannulas that connect the ventricles to the atrium.

Page 8 of 10

Page 9: Development of a dynamic cardiac simulator controlled by ...dainf.ct.utfpr.edu.br/~hvieir/download/ecr19.pdf · The Arduino code developed allowed controlling the movement of the

Conclusion

The developed simulator allows independent control of the compression of each chamberof the cardiac phantom. This can contribute to professional training through simulation,making it possible to prepare students for different situations that can be experiencedin practice, as well as to improve protocols to reduce doses of radiation and errors inprocedures.

Improvements are being made to prevent fluid reflux and also to vary the heart ratefrequency to simulate atrial and ventricular arrhythmias.

Page 9 of 10

Page 10: Development of a dynamic cardiac simulator controlled by ...dainf.ct.utfpr.edu.br/~hvieir/download/ecr19.pdf · The Arduino code developed allowed controlling the movement of the

References

[1] WHO. World Health Organization. Cardiovascular Diseases. [cited 12 Dezember2017]. Available from: <http://www.who.int/news-room/fact-sheets/detail/cardiovascular-diseases-(cvds)>.

[2] MAEDA Y et al. Development of a novel 3D dynamic cardiac phantom simulatedhuman anatomy. J Nucl Med May 1, 59(1), 2018.

[3] TOLEDO JM. Modelos, simuladores, dosimetria e respostas imunológicasradioinduzidas em cintilografia de perfusão do miocárdio. [thesis]. Belo Horizonte (MG):Universidade Federal de Minas Gerais; 2014.

Page 10 of 10