Detection of common organic acids using a compact ion chromatography system coupled with mass spectrometry APPLICATION BRIEF Introduction This application brief demonstrates a method for detecting common organic acids using a Thermo Scientific Dionex Integrion HPIC system with a Thermo Scientific Dionex IonPac AS11-HC-4µm column set and coupled to a Thermo Scientific ISQ EC single quadrupole mass spectrometer. Mass spectrometry (MS), as a detection technique for ion chromatography (IC), has recently gained popularity due to the increasing demand for sensitivity, selectivity, confirmation of identity, and structural interpretation. 1 The ISQ EC single quadrupole mass spectrometer increases analytical confidence by providing sensitive detection and mass confirmation in a product that requires less operator training than a typical mass spectrometer. Figure 1 shows chromatograms of 0.5 mg/L each of pyruvate, succinate, malate, and citrate standard using a suppressed conductivity detector (CD) and an MS detector in selected ion monitoring (SIM) mode. The CD responds to all charged species regardless of their molecular mass. If there is a coelution of multiple ionic components, the CD cannot discern them and presents them as one peak. Authors Beibei Huang, Jingli Hu, and Jeffrey Rohrer Keywords Integrion, RFIC, Dionex IonPac AS11-HC-4µm column, ISQ EC single quadrupole mass spectrometer, IC-MS

Upload: others

Post on 01-Jun-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Detection of Common Organic Acids Using a Compact Ion ... · 1. Weiss J. (2016) Handbook of Ion Chromatography, 3 Volume Set, 4th Edition. Wiley-VCH, Weinheim, Germany. 30,000 18,000

Detection of common organic acids using a compact ion chromatography system coupled with mass spectrometry

APPLICATION BRIEF

IntroductionThis application brief demonstrates a method for detecting common organic acids using a Thermo Scientifi c™ Dionex™ Integrion™ HPIC™ system with a Thermo Scientifi c™ Dionex™ IonPac™ AS11-HC-4µm column set and coupled to a Thermo Scientifi c™ ISQ™ EC single quadrupole mass spectrometer. Mass spectrometry (MS), as a detection technique for ion chromatography (IC), has recently gained popularity due to the increasing demand for sensitivity, selectivity, confi rmation of identity, and structural interpretation.1 The ISQ EC single quadrupole mass spectrometer increases analytical confi dence by providing sensitive detection and mass confi rmation in a product that requires less operator training than a typical mass spectrometer.

Figure 1 shows chromatograms of 0.5 mg/L each of pyruvate, succinate, malate, and citrate standard using a suppressed conductivity detector (CD) and an MS detector in selected ion monitoring (SIM) mode. The CD responds to all charged species regardless of their molecular mass. If there is a coelution of multiple ionic components, the CD cannot discern them and presents them as one peak.

AuthorsBeibei Huang, Jingli Hu, and Jeffrey Rohrer

KeywordsIntegrion, RFIC, Dionex IonPac AS11-HC-4µm column, ISQ EC single quadrupole mass spectrometer, IC-MS

Page 2: Detection of Common Organic Acids Using a Compact Ion ... · 1. Weiss J. (2016) Handbook of Ion Chromatography, 3 Volume Set, 4th Edition. Wiley-VCH, Weinheim, Germany. 30,000 18,000

The MS, on the other hand, is a more universal and selective detector. In SIM mode, it typically yields only one component peak for each target analyte, even in the case of coelution. Under aqueous eluent conditions, succinate and malate coelute, and therefore cannot be identified and quantified with suppressed conductivity detection. However, they can be accurately quantified with mass spectrometric detection because they are shown in different SIM channels, m/z 117 for succinate and m/z 133 for malate, respectively. Moreover, compared with suppressed conductivity detection, MS provides higher sensitivity for the four common organic acids in Figure 1.

Methods

Conditions

IC system Dionex Integrion HPIC system

Columns

Dionex IonPac AS11-HC-4 μm Guard, 2 × 50 mm Dionex IonPac AS11-HC-4 μm Analytical, 2 × 250 mm

Eluent source Thermo Scientific™ Dionex™ EGC 500 KOH Eluent Generator Cartridge with Thermo Scientific™ Dionex™ CR-ATC 600

Gradient 30-60 mM KOH (0-9 min), 60 mM KOH (9-10 min), 30 mM KOH (10.05-11 min)

Flow rate 0.38 mL/min

Injection volume 2.5 µL

Temperature 40 ºC (column compartment), 20 ºC (detector compartment)

System backpressure

~4100 psi

Detection

Suppressed Conductivity, Dionex AERS 500e Electrolytically Regenerated Suppressor (2 mm), AutoSuppression, 57 mA, external water mode via Thermo Scientific™ AXP™ Pump, external water flow rate (at least 2× the eluent flow rate)

Background conductance

~ 0.3 µS

Run time 11 min

Mass Spectrometric Detection

MS detector ISQ EC single quadrupole mass spectrometer

Ionization interface

Electrospray ionization (ESI), negative mode

Gas control Sheath gas pressure: 30 psiAux gas pressure: 2.0 psiSweep gas pressure: 0.0 psi

Source voltage -3000 v

Ion transfer tube temperature

150 ºC

Vaporizer temperature

400 ºC

Scans SIM mode: 87, 117, 133, 191Dwell or Scan Times (sec): 0.2 SIM Widths (amu): 0.5Ion Polarity: NegativeSourceCIDVoltage: 0Chrom. Filter Peak Width (sec): 25.0

Figure 1. Conductivity and SIM chromatograms of four common organic acids.

Inte

nsity

(cou

nts)

Res

pons

e (µ

S)

0 2 4 6 8 10 11

0

350,000

Minutes

0

300,000

0

250,000

0

50000

120,000-1.50

-0.40

RT:3.58 - pyruvateRT:4.79 - Succinate/Malate

Standard mix at 0.5 mg/L level

Pyruvate SIM 87

Succinate SIM 117

Malate SIM 133

Citrate SIM 191

RT:9.46 - Citrate

RT:3.58 - pyruvate - Area:16652

RT:4.77 - Succinate - Area:34718

RT:4.79 - Malate - Area:42018

RT:9.46 - Citrate - Area:55471

Page 3: Detection of Common Organic Acids Using a Compact Ion ... · 1. Weiss J. (2016) Handbook of Ion Chromatography, 3 Volume Set, 4th Edition. Wiley-VCH, Weinheim, Germany. 30,000 18,000

©2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. AB72363-EN 0617S

Find out more at thermofisher.com/ISQEC

This method can also be used for confirmation of organic acid identity in samples. Figure 2 shows that five common organic acids were identified in an animal feed sample: m/z 191 for citrate, m/z 89 for lactate, m/z 133 for malate, m/z 103 for malonate, and m/z 117 for succinate.

Figure 2. The identification of common organic acids in an animal feed sample by the comparison of retention time and m/z.

References1. Weiss J. (2016) Handbook of Ion Chromatography, 3 Volume Set, 4th Edition.

Wiley-VCH, Weinheim, Germany.

30,000

18,000

0

70,000 -5.0

45.0

Citrate, m/z 191

Lactate, m/z 89

0 5 10 15 20 25 30 35 40 45

0

7,000

0

25,000

0

Malate, m/z 133

Malonate, m/z 103

Succinate, m/z 117

Animal Feed Sample

Res

pons

e (µ

S)

Minutes

Inte

nsity

(cou

nts)

System: Dionex Integrion HPIC System ISQ EC single quadrupole mass spectrometerColumn: Dionex IonPac AS11-HC-4µm Guard, 2 × 50 mm Dionex IonPac AS11-HC-4µm Analytical, 2 × 250 mmEluent source: Dionex EGC 500 KOH cartridge with CR-ATC 600Gradient: 1 mM KOH (0-8 min), 1-30 mM (8-30 min), 30-60 mM (30-44 min), 60 mM (44-45 min)Column temperature: 40 °CFlow rate: 0.35 mL/minInj. volume: 2.5 µLDetection: Suppressed conductivity, Dionex AERS 500e suppressor, 2 mm, Autosuppression, 52 mA, external water mode (0.7 mL/min) via ASP pump

MS conditions are the same as shown in “Method: except for as below:SIM mode: 89, 103, 117, 133, 191